TY - JOUR A1 - Werner, Rudolf A. A1 - Marcus, Charles A1 - Sheikhbahaei, Sara A1 - Solnes, Lilja B. A1 - Leal, Jeffrey P. A1 - Du, Yong A1 - Rowe, Steven P. A1 - Higuchi, Takahiro A1 - Buck, Andreas K. A1 - Lapa, Constantin A1 - Javadi, Mehrbod S. T1 - Visual and Semiquantitative Accuracy in Clinical Baseline 123I-Ioflupane SPECT/CT Imaging T2 - Clinical Nuclear Medicine N2 - PURPOSE: We aimed to (a) elucidate the concordance of visual assessment of an initial I-ioflupane scan by a human interpreter with comparison to results using a fully automatic semiquantitative method and (b) to assess the accuracy compared to follow-up (f/u) diagnosis established by movement disorder specialists. METHODS: An initial I-ioflupane scan was performed in 382 patients with clinically uncertain Parkinsonian syndrome. An experienced reader performed a visual evaluation of all scans independently. The findings of the visual read were compared with semiquantitative evaluation. In addition, available f/u clinical diagnosis (serving as a reference standard) was compared with results of the human read and the software. RESULTS: When comparing the semiquantitative method with the visual assessment, discordance could be found in 25 (6.5%) of 382 of the cases for the experienced reader (ĸ = 0.868). The human observer indicated region of interest misalignment as the main reason for discordance. With neurology f/u serving as reference, the results of the reader revealed a slightly higher accuracy rate (87.7%, ĸ = 0.75) compared to semiquantification (86.2%, ĸ = 0.719, P < 0.001, respectively). No significant difference in the diagnostic performance of the visual read versus software-based assessment was found. CONCLUSIONS: In comparison with a fully automatic semiquantitative method in I-ioflupane interpretation, human assessment obtained an almost perfect agreement rate. However, compared to clinical established diagnosis serving as a reference, visual read seemed to be slightly more accurate as a solely software-based quantitative assessment. KW - Single-Photon-Emissions-Computertomographie KW - SPECT KW - Parkinson’s disease KW - Parkinsonism KW - DaTscan KW - 123I-Ioflupane KW - SPECT KW - SPECT/CT Y1 - 2018 UR - https://opus.bibliothek.uni-wuerzburg.de/frontdoor/index/index/docId/16818 UR - https://nbn-resolving.org/urn:nbn:de:bvb:20-opus-168181 SN - 1536-0229 VL - 44 IS - 1 ER -