TY - JOUR A1 - Hock, Michael A1 - Terekhov, Maxim A1 - Stefanescu, Maria Roxana A1 - Lohr, David A1 - Herz, Stefan A1 - Reiter, Theresa A1 - Ankenbrand, Markus A1 - Kosmala, Aleksander A1 - Gassenmaier, Tobias A1 - Juchem, Christoph A1 - Schreiber, Laura Maria T1 - B\(_{0}\) shimming of the human heart at 7T T2 - Magnetic Resonance in Medicine N2 - Purpose Inhomogeneities of the static magnetic B\(_{0}\) field are a major limiting factor in cardiac MRI at ultrahigh field (≥ 7T), as they result in signal loss and image distortions. Different magnetic susceptibilities of the myocardium and surrounding tissue in combination with cardiac motion lead to strong spatio‐temporal B\(_{0}\)‐field inhomogeneities, and their homogenization (B0 shimming) is a prerequisite. Limitations of state‐of‐the‐art shimming are described, regional B\(_{0}\) variations are measured, and a methodology for spherical harmonics shimming of the B\(_{0}\) field within the human myocardium is proposed. Methods The spatial B\(_{0}\)‐field distribution in the heart was analyzed as well as temporal B\(_{0}\)‐field variations in the myocardium over the cardiac cycle. Different shim region‐of‐interest selections were compared, and hardware limitations of spherical harmonics B\(_{0}\) shimming were evaluated by calibration‐based B0‐field modeling. The role of third‐order spherical harmonics terms was analyzed as well as potential benefits from cardiac phase–specific shimming. Results The strongest B\(_{0}\)‐field inhomogeneities were observed in localized spots within the left‐ventricular and right‐ventricular myocardium and varied between systolic and diastolic cardiac phases. An anatomy‐driven shim region‐of‐interest selection allowed for improved B\(_{0}\)‐field homogeneity compared with a standard shim region‐of‐interest cuboid. Third‐order spherical harmonics terms were demonstrated to be beneficial for shimming of these myocardial B\(_{0}\)‐field inhomogeneities. Initial results from the in vivo implementation of a potential shim strategy were obtained. Simulated cardiac phase–specific shimming was performed, and a shim term‐by‐term analysis revealed periodic variations of required currents. Conclusion Challenges in state‐of‐the‐art B\(_{0}\) shimming of the human heart at 7 T were described. Cardiac phase–specific shimming strategies were found to be superior to vendor‐supplied shimming. KW - 7 T KW - B KW - cardiac MRI KW - shimming KW - ultrahigh field Y1 - 2021 UR - https://opus.bibliothek.uni-wuerzburg.de/frontdoor/index/index/docId/21809 UR - https://nbn-resolving.org/urn:nbn:de:bvb:20-opus-218096 VL - 85 IS - 1 SP - 182 EP - 196 ER -