@phdthesis{Pawellek2021, author = {Pawellek, Ruben}, title = {Charged Aerosol Detector Performance Evaluation and Development of Optimization Strategies for the Analysis of Amino Acids}, doi = {10.25972/OPUS-24319}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-243197}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2021}, abstract = {The charged aerosol detector (CAD) is an aerosol-based detector employed in liquid chromatography which has become established in the field of pharmaceutical analysis due to its outstanding performance characteristics, e.g. the almost uniform response for nonvolatile analytes. Owing to its principle of detection, the response of the CAD depends on the volatility of a compound and is inherently nonlinear. However, the newly implemented instrumental settings evaporation temperature and power function value (PFV) are valuable tools to overcome some of these drawbacks and can even enhance the detector's capabilities when adjusted properly. This thesis aimed to evaluate the impact of the new instrumental settings on the CAD performance. Additionally, the influence of modern separation techniques for small polar compounds on the CAD was assessed and the applicability of hyphenated UV-CAD techniques explored. The optimization strategies derived from the evaluation procedures and the conjunction of the instrumental and chromatographic techniques investigated were utilized for the challenging impurity profiling of amino acids and amino acid-like drugs. The results of the method validation procedures confirmed the broad applicability of the CAD in the pharmaceutical analysis of nonvolatile compounds, supported by satisfactory sensitivity and reproducibility for meeting the regulatory requirements with respect to the ICH guidelines Q2(R1) and Q3A(R2). The limits of applicability include the analysis of semivolatile compounds, and the method transfer between current and legacy CAD models. Further advances in the definition and standardization of allowed ranges for the instrumental settings and the establishment of general optimization procedures in the method development could lead to a more widespread use of the detection technique in compendial methods.}, subject = {Instrumentelle Analytik}, language = {en} } @phdthesis{Unger2020, author = {Unger, Nina}, title = {Stability of Tryptophan in Parenteral Amino Acid Solutions: Identification of Degradation Products and Development of HPLC Analysis Methods}, doi = {10.25972/OPUS-19982}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-199825}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2020}, abstract = {The stability of Trp in pure solutions and in parenteral AA formulations was evaluated with regard to typically used manufacturing processes, storage conditions and primary packaging. Therefore, thorough stability studies on Trp solutions were conducted beforehand. The applied stressing method, i.e. steam sterilization by autoclave, are chemically seen relatively mild but showed to be efficient to induce Trp degradation in the presence of oxygen. Subsequent identification, separation and characterization were challenging due to similar substance properties, numerous stereoisomers and pairs of diastereomers found amongst them. However, the identified o-aminoacetophenone compounds, Kyn and NFK, are associated with photo reactivity and have photo-oxidizing properties. Thus, best possible protection from UV-light, together with strict oxygen expulsion, are the most important criteria to impede Trp degradation after autoclaving. The identification of Trp degradation products was assisted by the compilation of a substance library, which included manifold reported and chemically plausible Trp degradation substances. The substances were classified for priority and their early or late-stage occurrence. The large number of possible substances and stereoisomers was narrowed down with the information retrieved from LC-UV/MS experiments. However, final identification was achieved by the synthesis of proposed substances as references. The following eight substances were characterized as Trp degradation substances: Kyn, NFK and three pairs of diastereomers R,R/R,S DiOia, R,R/R,S Oia and cis/trans PIC. Fig. 33 shows the proposed degradation pathway and demonstrates the close chemical relationship, which may be an explanation for the conversion of some substances into each other during the storage period. The proposed pathway brings together the results of different Trp stability and stressing studies, respectively [89, 94, 97, 98, 103, 133]. To our knowledge, the simultaneous formation of the identified degradation substances has not been reported before and especially not under the stressing conditions applied. The application of a traditional RP-HPLC method was compared to two developed IP-HPLC methods and a RP-HPLC methods using a modified perfluorinated column. Orthogonal analyses methods and especially the combination of UV and MS detection are necessary in order to indicate potentially undetected degradation substances. Main evaluation criteria were the separation performance, analyses time, reproducibility and feasibility. The best results upon assessment of all Trp degradation products, in both; pure Trp solutions and pharmaceutical formulations, were obtained by a traditional RP-HPLC. The optimized method was validated according to ICH guidelines Q2(R1) and meets the criteria of a stability-indicating HPLC-UV method. The validated method has a sufficient separation performance with an adequate selectivity indicating the Trp degradation substances next to each other and next to other AAs in finished pharmaceutical formulations. The detailed knowledge of Trp degradation and the method presented may be transferred practically to the pharmaceutical industry processing Trp-containing products. In general, the findings might contribute to the quality management of such pharmaceutical products during manufacturing and storage. Additionally, the study results provide basic information for the establishment of an impurity consideration following the ICH guidelines Q3B (R2) (impurities in new drug products) for products containing Trp. However, further development of the method applying more sophisticated detectors or more potent HPLC techniques like e.g. UHPLC and the implication of more sensitive (MS) detectors like ToF-MS would be advantageous with regard to economic and practical aspects.}, subject = {Stabilit{\"a}t}, language = {en} } @phdthesis{Kuehnreich2016, author = {K{\"u}hnreich, Raphael}, title = {Development and Validation of Methods for Impurity Profiling of Amino Acids}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-145718}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2016}, abstract = {The requirements for the impurity profiling of substances for pharmaceutical use have become greater over time. They can be accomplished by the use of modern instrumental analysis techniques, which have been evolved in the last decades. New types of columns with HILIC, mixed-mode and chiral stationary phases are suitable for the separation of all kinds of substances mixtures, that were previously hardly possible with the use of common reversed phase columns. Modern, almost universal detectors like CAD, ELSD and CNLSD can be applied for a sensitive detection of substances without a chromophore. However, in addition to some small individual disadvantages to these methods, the costs are high and applications are still kind of rare. Thus, the introduction of these devices at a broader level has not yet taken place. While this presumably will change over time, there is a need for methods that enable the impurity profiling of challenging substances with widespread analytics devices. Methionine is a substance with hydrophobic and hydrophilic impurities. With the help of a mixed-mode stationary phase, which is a combination of a reversed phase and a strong cationic exchanger, the separation of all putative impurities was found possible with good sensitivity and selectivity. The method requires apart from the column only standard isocratic HPLC equipment and was successfully validated. The evaluation of the enantiomeric purity of amino acids is challenging. Two approaches were made. The first method utilizes CE by means of in-capillary derivation with OPA and the subsequent separation with a cyclodextrin. With the use of OPA/NAC and γ-cyclodextrin, a simple and cost-effective method for the indirect enantioseparation of 16 amino acids was developed. With the second approach, racemic amino acids can be analyzed with HPLC and in-needle derivatization. For this, different columns and chiral thiols were evaluated and the chromatographic parameters were optimized. A method with OPA/NIBLC, a pentafluorophenyl column made the enantioseparation of 17 amino acids feasible. A LOQ of the minor enantiomer down to 0.04 \% can be achieved with UV spectrophotometric detection. A similar method was developed for impurity profiling of L-amino acids. This can be used alternatively for the amino acid analysis performed by the European Pharmacopoeia. A simple, robust, precise and accurate method for the evaluation of impurities in glyceryl trinitrate solution was developed and validated. The four impurities of glyceryl trinitrate are separated by means of an acetonitrile-water gradient and the assay for this substance is also possible.}, subject = {Aminos{\"a}uren}, language = {en} } @phdthesis{Wahl2016, author = {Wahl, Oliver}, title = {Impurity Profiling of Challenging Active Pharmaceutical Ingredients without Chromophore}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-137205}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2016}, abstract = {The impurity profiling of pharmaceutical ingredients can oppose many challenges. The best part of active pharmaceutical ingredients (APIs) and the related substances are detectable by UV detection, a very common detection principle. However, if an API lacks a suitable chromophore other means of detection are necessary. The corona charged aerosol detector (CAD) is a detector capable of detecting substances independent of their chemical structure. This "universal" detector has only one limitation: The analyte has to have a sufficiently low vapor pressure. Another important challenge that comes often together with the lack of a chromophore concerns the separation. These substances (e.g. most amino acids and derivatives) often contain structures that make them difficult to retain on conventional reversed phase columns. Possible solutions to overcome these challenges, like the application of the CAD and the benefit of so-called mixed-mode stationary phases in impurity profiling for pharmacopoeial purposes were explored in this work. The related substances analyzed in this thesis comprise amino acids, inorganic ions, bisphosphonic acids, basic and acidic derivatives of amino acids (esters and amides). The successful development and validation of mixed-mode liquid chromatography methods with CAD detection for carbocisteine and ibandronate sodium might help to increase the acceptance of this versatile detector in the pharmaceutical industry and in official authorities dealing with the determination of related substances. The combination of UV and CAD detection proved very useful during the analysis of Bicisate. Most of the related substances and some unidentified impurities were detectable by CAD whereas a synthesis by-product, a semi-volatile ester, was only detectable in the UV trace. The simple combination covers all relevant impurities in a single analysis. Two truly orthogonal methods regarding separation and detection for the enantiomeric purity of magnesium-L-aspartate helped to find the reason for elevated D aspartic acid content in the drug substance. A very quick and sensitive indirect separation using the OPA derivatization with NAC was developed as a powerful screening tool, whereas the direct separation of D- and L-CBQCA-Asp derivatives confirmed the results. Both methods were optimized in order to do without substances mentioned on the REACH list, like sodium tetraborate which is very frequently applied in standard derivatization protocols and CE separations. The importance of orthogonal detection principles in the determination of related substances of amino acids was discussed in a review article dealing with the revision of amino acid monographs in the Ph. Eur..}, subject = {Chromatographie}, language = {en} } @phdthesis{Freitag2011, author = {Freitag, Claudia}, title = {Quantifizierung von Aminos{\"a}uren in Infusionsl{\"o}sungen mittels Hochleistungsfl{\"u}ssigkeitschromatographie-(Tandem) - Massenspektrometrie(HPLC-[MS/]MS) Methodenentwicklung und Validierung}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-65198}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2011}, abstract = {Das Ziel vorliegender Arbeit war die Entwicklung einer HPLC-MS(/MS)-Methode, die im Rahmen der pharmazeutischen Qualit{\"a}tskontrolle zur direkten Quantifizierung von Aminos{\"a}uren (AS) in Infusionsl{\"o}sungen angewendet werden kann. Die Zielset-zung schloss eine Validierung innerhalb der f{\"u}r die Zweckbestimmung vorgesehenen Grenzen ein. Im Rahmen der Methodenentwicklung wurde das ESI-MS/MS-Fragmentierungs-muster von 21 Aminos{\"a}uren, von 20 stabil-isotopenmarkierten Aminos{\"a}uren, die als interne Standards verwendet wurden, sowie von einigen weiteren Substanzen bestimmt. Nach Kenntnis von Precursor- und Produktionen erstellte man eine SRM-Methode zur spezifischen MS/MS-Analyse. Dabei wurden durch das jeweilige Frag-mentierungsmuster bedingte Interferenzen bei den zu untersuchenden Aminos{\"a}uren bestimmt, die bei der zu erarbeitenden HPLC-MS-Methode beachtet werden mussten. Die Methodenentwicklung zur HPLC-MS-Analytik von underivatisierten AS umfasste mit der RP-HPLC unter Verwendung eines Ionenpaarreagenzes (IP) und der hydrophilen Interaktionschromatographie (HILIC) zwei verschiedene chromatographi-sche Ans{\"a}tze. Bei der Anwendung der RP-HPLC ergaben sich Probleme. Die Verwendung eines IP, im vorliegenden Fall TDFHA (Tridecafluorheptans{\"a}ure), f{\"u}hrte zu langen Equilibrierungs-, Re-Equilibrierungs- und Sp{\"u}lzeiten und damit bei zwar relativ kurzer HPLC-Laufzeit zu einem aber insgesamt hohen Zeitaufwand. Gleich-zeitig war die LC-MS-Anlage auf diese Anwendung fixiert, da das Ionenpaarreagenz das Ger{\"a}t stark verschmutzte und dadurch andere Analysen erheblich st{\"o}rte. Zudem waren die Retentionszeiten der Analyten trotz langer Equilibrierungszeiten schlecht reproduzierbar, so dass eine solche Methode im Rahmen der pharmazeutischen Qualit{\"a}tskontrolle schwer validierbar w{\"a}re. Weiterf{\"u}hrende Untersuchungen erfolgten daher nicht. In nachfolgenden Studien mit der HILIC wurden verschiedene Einflussparameter (Anteil organischer Phase im Fließmittel, pH-Wert des Fließmittels, Temperatur der S{\"a}ule, Pufferkonzentration im Fließmittel, Gradientenelution) auf die Trennung der AS an einer ZIC®-HILIC-S{\"a}ule untersucht. Durch Optimierung der Parameter wurde so eine HILIC-HPLC-Methode entwickelt, bei der 21 AS und 20 ihrer isotopen-markierten Referenz-AS innerhalb von 20 min eluierten. Diejenigen AS, bei denen im Rahmen der Fragmentierungsstudien Interferenzen aufgrund gleicher bzw. {\"a}hnlicher Massen der Precursor- bzw. Produktionen aufgetreten waren, wurden chroma-tographisch getrennt. Gleichzeitig hat sich die SIM-Analyse als anwendbar erwiesen. Die Anwendung des spezifischeren SRM-Modus und damit der Tandem-Massenspektrometrie war nicht erforderlich. Im Rahmen der nachfolgenden Studien zur Validierung ergab sich, dass die entwickelte Methode {\"u}ber einen weiten Bereich eine lineare Abh{\"a}ngigkeit zwischen Konzentrations- und Messwerten zeigte. F{\"u}r drei der 21 AS (NAcCys, NAcTyr, Pro) wurde die quadratische Regression mit dem Anpassungstest nach Mandel als geeig-neteres Regressionsmodell ermittelt. Bei Untersuchungen zur Wiederfindung wurde ein Einfluss der Matrix-L{\"o}sung der Infusionsl{\"o}sung festgestellt, der zu Abweichungen hinsichtlich des Quotienten AreaAS / AreaIS f{\"u}hrte, so dass eine Quantifizierung innerhalb der geforderten Grenzen bei Kalibrierung {\"u}ber reine Standardl{\"o}sungen nicht m{\"o}glich war. Die Validierung wurde daher nachfolgend in der Matrixl{\"o}sung durchgef{\"u}hrt. Dabei wurde gezeigt, dass mit der entwickelten HILIC-HPLC-MS-Methode Aminos{\"a}uren in Infusionsl{\"o}sungen mit hoher Pr{\"a}zision und Richtigkeit bestimmt werden k{\"o}nnen. Neun der 21 untersuchten AS konnten im Bereich von 30\% - 350\%, zehn weitere im Bereich von 50\% - 350\% innerhalb der zur Gehaltsbestimmung von pharmazeutischen Formulierungen vorgeschriebenen Grenzen (Wiederfindung Einzelbestimmung: 98\% -102.0\%, Mittelwert einer Dreifachbestimmung: 98.5\% - 101.5\%) quantifiziert werden. F{\"u}r His und Phe gelang allerdings keine Quantifizierung innerhalb der Akzeptanzkriterien, wobei der Grund hierf{\"u}r in weiteren Studien gekl{\"a}rt werden m{\"u}sste. Mit der entwickelten Methode ist damit eine gleichzeitige Quantifizierung verschiedener AS-Infusionsl{\"o}sungsformulierungen m{\"o}glich, die sich bei gleicher Matrix in der Konzentration an AS unterscheiden. Beispielsweise seien hier die Formulierungen „Aminoplasmal® E 5\% / 10\% /15\%" genannt, die mit der validierten Methode erfassbar sind. Die Probenvorbereitung beschr{\"a}nkt sich dabei auf den Zusatz der IS-Formulierung zur Infusionsl{\"o}sung und einen Verd{\"u}nnungsschritt. Die Quanti-fizierung erfolgt {\"u}ber eine 5-Punkt-Kalibriergerade, die aus einer AS- und IS-Standardmischung, nach Zusatz der einfach herzustellenden Elektrolyt-Matrix, erstellt wird. Die Analysenzeit der HPLC-MS-Methode betr{\"a}gt einschließlich Equilibrie-rungszeit 35 min und ist damit deutlich k{\"u}rzer als die 120 min, die bei der nach wie vor zur AS-Analytik allgemein gebr{\"a}uchlichen Ionenaustauschchromatographie mit Ninhydrin-Nachs{\"a}ulenderivatisierung anzusetzen sind.}, subject = {Aminos{\"a}uren}, language = {de} } @phdthesis{Beyer2011, author = {Beyer, Tanja}, title = {Quantitative NMR-Spektroskopie in der pharmazeutischen Analytik -- Identit{\"a}t, Reinheit und Gehalt von Arzneistoffen}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-65091}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2011}, abstract = {Das Ziel der vorliegenden Arbeit war die Kl{\"a}rung der Fragestellung, ob sich die quantitative NMR-Spektroskopie zur Bestimmung von Identit{\"a}t, Reinheit und Gehalt von Arzneistoffen eignet, und wie sich Pr{\"a}zision und Empfindlichkeit dieser Methode im Vergleich zu etablierten chromatographischen Verfahren verhalten. Die quantitative Untersuchung der drei strukturell jeweils verwandten Mehrkomponentengemische Codergocrinmesilat, Clomifencitrat und Flupentixoldihydrochlorid bewies eindrucksvoll die Eignung der 1H-NMR-Spektroskopie als orthogonale, analytische Messmethode im Vergleich zu validierten HPLC-Arzneibuchmethoden. Die im Rahmen einer Validierung der 1H-NMR-Methode ermittelten Ergebnisse erf{\"u}llten bez{\"u}glich Pr{\"a}zision und Richtigkeit die an eine im pharmazeutischen Bereich eingesetzte analytische Methode gestellten Anforderungen; zudem wurden weitere Pr{\"u}fparameter wie Selektivit{\"a}t, Linearit{\"a}t, Robustheit und Stabilit{\"a}t verifiziert. Externe-Standard-Experimente wie "Zwei-R{\"o}hrchen-Methode" und ERETIC-Verfahren best{\"a}tigten die quantitativen Ergebnisse der Internen Standardisierung; jedoch wurde hier -- insbesondere f{\"u}r die ERETIC-Technik -- eine h{\"o}here Fehleranf{\"a}lligkeit und somit eine gr{\"o}ßere Streuung der Einzelergebnisse beobachtet. Am Beispiel von Codergocrinmesilat und Flupentixoldihydrochlorid konnte zudem die Eignung anderer NMR-aktiver Kerne wie 13C und 19F f{\"u}r die quantitative Analyse von komplexen Substanzgemischen aufgezeigt werden. Das Potential der 1H-NMR-Spektroskopie f{\"u}r die Reinheitspr{\"u}fung von Arzneistoffen wurde am Beispiel der Aminos{\"a}ure L-Alanin aufgezeigt. Die zu erwartenden Verunreinigungen Glutamin-, Asparagin-, {\"A}pfel- und Fumars{\"a}ure konnten im Gegensatz zu den "veralteten" Pr{\"u}fmethoden des Europ{\"a}ischen Arzneibuches eindeutig identifiziert und quantifiziert werden; mit einer Bestimmungsgrenze von <= 0.03\% wurden die Vorgaben der ICH-Richtlinie Q3A(R2) erf{\"u}llt. Die deutliche {\"U}bereinstimmung der NMR-spektroskopisch ermittelten Ergebnisse einer quantitativ untersuchten Alanin-Modellmischung mit einer f{\"u}r den Routinebetrieb geeigneten HPLC-Methode unter Einsatz verschiedener Detektoren wie CAD, NQAD, ELSD und MS, sowie der Vergleich wichtiger Pr{\"u}fparameter wie Linearit{\"a}t und Nachweisgrenze best{\"a}tigten die Eignung der 1H-NMR-Spektroskopie im Rahmen der routinem{\"a}ßig durchgef{\"u}hrten Qualit{\"a}tskontrolle. Die Aufdeckung von Arzneimittelf{\"a}lschungen mit Hilfe der NMR-Spektroskopie wurde im Rahmen dieser Arbeit anhand der zwei aktuellen Fallbeispiele Heparin und Glycerin n{\"a}her untersucht. Die in Zusammenhang mit dem Heparin-Skandal verantwortliche Kontaminante OSCS konnte neben Dermatansulfat und weiteren nat{\"u}rlich vorkommenden Glykosaminoglykan-Verunreinigungen im 1H-NMR-Spektrum eindeutig identifiziert und auf 0.1\% OSCS bzw. 0.5\% Dermatansulfat begrenzt werden. Eine pr{\"a}zise und richtige quantitative Bestimmung der beiden Glykosaminoglykane wurde {\"u}ber die N-Acetyl-Resonanzen mit Hilfe der Signalh{\"o}henbestimmung und dem Standard-Additionsverfahren erm{\"o}glicht; deutliche Abweichungen vom "wahren" Gehalt wurden hingegen, bedingt durch starke Signal{\"u}berlagerungen, nach Fl{\"a}chenvergleich beobachtet. Weitere Verunreinigungen, insbesondere L{\"o}sungsmittelr{\"u}ckst{\"a}nde, die w{\"a}hrend des Extraktions- und Reinigungsprozesses des Heparins eingesetzt werden, konnten ebenfalls {\"u}ber charakteristische Resonanzen identifiziert und mit Hilfe der Internen-Standard-Methode quantitativ erfasst werden. Eine umfangreiche Untersuchung von 145 Heparin-API-Mustern mittels NMR-Spektroskopie und weiteren, neuentwickelten Verfahren wie HPLC, CE, IR- und Raman-Spektroskopie konnte die Eignung der entwickelten 1H-NMR-Methode best{\"a}tigen. Potentielle Glycerin-Kontaminanten wie Diethylenglycol und Ethylenglycol konnten ebenso wie eine weitere, nat{\"u}rlich vorkommende Verunreinigung, Propylenglycol, mittels 1H- und 13C-NMR-Spektroskopie identifiziert und quantifiziert werden. Beide Methoden erf{\"u}llten die in der USP beschriebenen Anforderungen, die f{\"u}r pharmazeutisch eingesetztes Glycerin jeweils h{\"o}chstens 0.1\% Diethylenglycol bzw. Ethylenglycol erlaubt. W{\"a}hrend die quantitative Reinheitspr{\"u}fung beim Einsatz der 1H-NMR-Spektroskopie mit einer Messdauer im Bereich von etwa 30 min f{\"u}r den Routineeinsatz geeignet ist, ist die entwickelte quantitative 13C-NMR-Methode beim Einsatz von Spektrometern geringer Magnetfeldst{\"a}rke aufgrund einer geringen Nachweisempfindlichkeit und der NOE-Problematik f{\"u}r den Routinebetrieb nur bedingt anwendbar. Abschließend kann zusammengefasst werden, dass die untersuchten Beispiele die NMR-Spektroskopie als in hohem Maße geeignet f{\"u}r die quantitative Analyse von Arzneimitteln ausweisen.}, subject = {NMR-Spektroskopie}, language = {de} } @phdthesis{Almeling2011, author = {Almeling, Stefan}, title = {The use of aerosol-based detection systems in the quality control of drug substances}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-64722}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2011}, abstract = {The work presented in this thesis was mainly targeted at exploring the capabilities of evaporation based LC detectors as well as further alternatives for the control of impurities in substances not exhibiting a suitable chromophore for UV-detection. In the course of the work carried out, several new methods for the identification, impurities control and composition testing of APIs were elaborated. An evaporation based detector that entered into the field of pharmaceutical analysis in the recent years was the Evaporative Light Scattering Detector (ELSD). However, non-reproducible spikes were reported when injecting concentrated test solutions as they are usually required for the control of impurities. The reasons, for the appearance of these spikes as well as possibilities for their avoidance were explored in a systematic study. Moreover, the dependence of the detector sensitivity on different eluent composition, eluent flow-rate and ELSD settings was investigated. In the course of the revision of the Ph.Eur. monographs for aspartic acid and alanine, a C18 reversed phase ion-pair LC method using 1 mmol/L of perfluoroheptanoic acid as an ion-pair reagent and a charged aerosol detector (CAD) was developed and fully validated for the purity control of Asp. The method was capable of separating the organic acids and major amino acids known to occur as process related impurities. With a slight modification, the method was also applicable for the purity control of Ala. Based on the developed LC-CAD method for the impurity control of alanine, a comparative study of the performance characteristics of different evaporation based LC detectors, i.e. ELSD, CAD and the recently developed Nano Quantity Analyte Detector (NQAD) was carried out. Additionally, an MS detector and qNMR were included in this study. It was found that the control of impurities in Alanine at an ICH conform level could be ensured using LC coupled to CAD, MSD and NQAD detection as well as by the use of qNMR. In terms of performance, prize and ease of use CAD and NQAD were found to be the most suitable alternatives. In terms of repeatability and sensitivity, the CAD appeared slightly superior to the NQAD. The quality of streptomycin sulfate is not sufficiently controlled by the current Ph.Eur. monograph in that an appropriate test for the control of the related substances is missing. A study was carried out to develop a C18 reversed phase ion-pair LC method using pentafluoropropionic acid as an ion-pair reagent and a CAD for the identification and control of the related substances. The developed method allowed the separation of 21 impurities from streptomycin. Moreover, coupling of the method to MS allowed the identification of the separated impurities. The method was shown to be sufficiently sensitive to control the related substances with a disregard limit of 0.1\% as it is normally applied in the Ph.Eur. for products derived from fermentation. Currently, the aescin content of horse-chestnut standardized dry extract is determined using a complex and laborious photometric determination. A more selective LC-UV assay determination for beta-aescin has been proposed for the Ph.Eur. draft monograph of horse-chestnut standardized dry extract. Possibilities were explored to further improve the LC-method using detection by CAD. It was demonstrated that by the use of a modified LC-CAD method several problems related to the differences in the UV-response of the various components contained in the active aescin fraction could be eliminated. Moreover the proposed reference standard strategy was reviewed. Eventually, it was demonstrated on the example of two different clusters of pharmacologically active peptides how low energy collision induced dissociation mass spectrometry (low energy CID-MS) can successfully be used for identification testing in pharmacopoeial monographs. In this respect, the combination of a direct confirmation of the molecular mass via the m/z-ratio of the molecule ions with structural sequence information obtained by low energy CID-MS experiments was found to deliver a higher degree of certainty of the identity of a given substance than the set of tests currently described in the monographs. A significant gain in efficiency and throughput and important reduction of the amount of sample consumed during testing were identified as being additional advantages of this approach. Taken together, it could be demonstrated on various examples how recent technological advancements in the field of analytical chemistry can contribute to improve the quality control of APIs.}, subject = {Elektronensprayionisations-Massenspektrometrie}, language = {en} } @phdthesis{Borst2011, author = {Borst, Claudia}, title = {Kapillarelektrophoretische Reinheitsanalytik verschiedener Arzneistoffe des Europ{\"a}ischen Arzneibuchs}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-56243}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2011}, abstract = {Die Kapillarelektophorese (CE), deren Trennprinzip auf der Wanderung geladener Teilchen im elektrischen Feld basiert, ist eine Methode, die in verschiedenen Techniken angewandt werden kann. Sowohl die w{\"a}ssrige Kapillarzonenelektrophorese (CZE) als auch die wasserfreie CE (NACE), aber auch die elektrokinetische Chromatographie mittels Mikroemulsion (MEEKC) wurden in dieser Arbeit f{\"u}r die Reinheitsanalytik der im Europ{\"a}ischen Arzneibuch beschriebenen Wirkstoffe Ethambutol, Quetiapin, Ephedrin sowie Levodopa und deren jeweils strukturverwandter Substanzen benutzt. Der Wirkstoff Ethambutol wird in der (S,S)-Form verwendet, die im Ph. Eur. 7 als Dihydrochlorid aufgef{\"u}hrt ist. Um eine Trennmethode f{\"u}r (S,S)-Ethambutol, sein Enantiomer und die achirale meso-Verbindung entwickeln zu k{\"o}nnen, wurden die beiden stereoisomeren Verunreinigungen aus 2-Amino-1-butanol und Diethyloxalat synthetisiert. Zur Trennung dieser drei Ethambutol-Isomere wurde CZE als Methode gew{\"a}hlt. In saurem Phosphatpuffer musste eine hohe Probenkonzentration von 1 mg/ml verwendet werden, um die Substanzen mit UV detektieren zu k{\"o}nnen (λ: 200 nm). In alkalischem Tetraboratpuffer war das Chromophor dank der freien Elektronenpaare der Stickstoff-Molek{\"u}le besser ausgepr{\"a}gt und die Intensit{\"a}t der Peaks deutlich intensiver. Als chirale Selektoren wurden die nativen α-, β- und γ-Cyclodextine (CDs) und verschiedene derivatisierte β-CDs eingesetzt. Die Methode wurde vielfach in Bezug auf Molarit{\"a}t und pH-Wert der Puffer, Konzentration der verschiedenen chiralen Selektoren, Spannung und Temperatur modifiziert. Jedoch konnte keine Trennung der Stereoisomere erreicht werden. Eine CD-modifizierte MEEKC-Methode wurde herangezogen, um die Racemate der Aminos{\"a}uren Dopa, Methyldopa, Tyrosin und Phenylalanin voneinander zu trennen. Dazu wurde eine Mikroemulsion (ME) aus Ethylacetat, SDS, 1-Butanol, Phosphatpuffer, sulf. β-CD und, wenn n{\"o}tig, aus dem organischen Modifier 2-Propanol eingesetzt. F{\"u}r jede DL-Aminos{\"a}ure wurde die Zusammensetzung der ME als auch die Ger{\"a}teeinstellungen (Spannung, Temperatur) optimiert. Die Trennung von DL-Dopa konnte ohne Zugabe eines organischen Modifiers durchgef{\"u}hrt werden. Auf Grundlage dieser individuellen Methoden wurden zwei CD-modifizierte MEEKC-Methoden entwickelt, mit denen alle vier untersuchten Racemate getrennt werden konnten. Die abschließende Validierung in Bezug auf Wiederholpr{\"a}zision (Aufl{\"o}sung, Migrationszeiten, Verh{\"a}ltnis der korrigierten Peakfl{\"a}chen und Anzahl der theoretischen B{\"o}den) und Detektionsgrenzen zeigte, dass die Methoden pr{\"a}zise Ergebnisse liefern. Die Technik der MEEKC wurde auch zur Trennung von Ephedrin-Derivaten genutzt. Wedig et al. konnten die Racemate von Ephedrin, Pseudoephedrin, N-Methylephedrin und Norephedrin mit einer HDAS-β-CD-modifizierten CZE-Methode in einem Lauf basislinientrennen, indem ein 50 mM Phosphatpuffer, pH 3,0 als HGE eingesetzt wurde. Aus diesem HGE und den organischen Bestandteilen, die zur Trennung der Aminos{\"a}uren f{\"u}hrten, wurde eine ME hergestellt. Entgegen der Methode von Wedig et al. konnte mittels HDAS-β-CD keine zufriedenstellende Trennleistung erreicht werden. Durch Austausch des chiralen Selektors gegen sulf. β-CD und Modifizierung des Phosphatpuffers in Ionenst{\"a}rke und pH-Wert konnte f{\"u}r alle vier Epedrin-Derivate eine Basislinientrennung erzielt werden. Diese MEEKC-Methode wurde auf weitere Ephedrin-Derivate angewandt, wodurch das racemische 2-(Dibutylamino)-1-phenyl-1-propanol partiell, die Racemate von Adrenalin, 2-Amino-1-phenylethanol und Diethylnorephedrin vollst{\"a}ndig voneinander getrennt werden konnten. W{\"a}hrend mit der HDAS-β-CD-modifizierten CZE-Methode alle vier Ephedrin-Derivaten in einem Lauf getrennt werden konnten, hat die MEEKC-Methode den Vorteil mit dem kosteng{\"u}nstigeren sulf. β-CD auszukommen. Schlussendlich wurde eine Reinheitsanalytik von Quetiapin und seinen verwandten Substanzen Quetiapindesethanol, Quetiapin-N-Oxid und Quetiapinlactam entwickelt. Da Quetiapinlactam fast ausschließlich in organischen L{\"o}sungsmitteln l{\"o}slich ist, sollte eine wasserfreie CE-Methode (NACE) eingesetzt werden. Zwar konnte eine Methode entwickelt werden, deren HGE aus Methanol, Acetonitril, Ammoniumacetat und Essigs{\"a}ure bestand, und mit der Quetiapin und seine drei verwandten Substanzen sehr gut getrennt werden konnten. Allerdings konnte sie aufgrund von Stromabbr{\"u}chen nicht validiert werden. Alternativ wurde eine w{\"a}ssrige, gut reproduzierbare CZE-Methode gefunden, deren Elektrolytl{\"o}sung aus einem 80 mM Phosphatpuffer, pH 4.0 bestand. Aufgrund der Wasserunl{\"o}slichkeit von Quetiapinlactam konnten so nur Quetiapin und die Verunreinigungen Quetiapindesethanol und Quetiapin-N-Oxid erfasst werden. Abschließend wurde die CZE-Methode validiert, wodurch die hohe Pr{\"a}zision der ermittelten Werte gezeigt werden konnte.}, subject = {Kapillarelektrophorese}, language = {de} } @phdthesis{Kopec2008, author = {Kopec, Susanne}, title = {Trimebutin, Adenosin, Glutathion und Aminos{\"a}uren - Beispiele f{\"u}r Reinheitsanalytik f{\"u}r das Europ{\"a}ische Arzneibuch}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-26410}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2008}, abstract = {Die Qualit{\"a}tskontrolle von pharmazeutisch verwendeten Substanzen ist eine Voraussetzung f{\"u}r die sichere Anwendung von Arzneimitteln. Sie ist eng mit der Bestimmung der Verunreinigungen einer Substanz im Rahmen der Reinheitsanalytik verkn{\"u}pft. Dabei spielt das Europ{\"a}ische Arzneibuch (Ph. Eur.) eine wichtige Rolle. Die in der vorliegenden Arbeit durchgef{\"u}hrten Untersuchungen zur Bestimmung des Verunreinigungsprofils waren auf die Neuerarbeitung oder {\"U}berarbeitung der im Ph. Eur. beschriebenen Pr{\"u}fungen auf „Verwandte Substanzen" ausgerichtet. Im Rahmen der Neuerarbeitung einer Monographie f{\"u}r Trimebutin und Trimebutin-Maleat wurde eine RP-HPLC-Methode entwickelt. Neben den bekannten Verunreinigungen war ein weiterer Peak nachweisbar, der N-Desmethyltrimebutin zugeordnet wurde. Die Trennung zwischen N-Desmethyltrimebutin und dem Hauptpeak sowie zwischen zwei bekannten Verunreinigungen ist kritisch. F{\"u}r beide Peakpaare wurden Systemeignungskriterien festgelegt. Zur Definition von Akzeptanzkriterien in den erarbeiteten Monographieentw{\"u}rfen wurden Chargen untersucht. Die Quantifizierung erfolgte {\"u}ber die „Externer-Standard"-Methode mit einer Verd{\"u}nnung der Untersuchungsl{\"o}sung als Referenz. Falls erforderlich, wurden die Korrekturfaktoren in die Berechnung des Gehaltes einbezogen. Die DC-Methode zur Pr{\"u}fung auf „Verwandte Substanzen" von Adenosin sollte gegen eine HPLC-Methode ausgetauscht werden. In {\"U}bereinstimmung mit Ergebnissen des EDQM-Labors haben die Untersuchungen best{\"a}tigt, dass mit einer Ionenpaar-HPLC-Methode die Anwesenheit von Inosin, Guanosin, Uridin und Adenin in Adenosin kontrolliert werden kann. Durch die Festlegung einer Aufl{\"o}sung > 1.5 zwischen Adenin und Inosin als Systemeignungskriterium werden nicht geeignete HPLC-S{\"a}ulen erkannt. Durch die DC-Pr{\"u}fung des Ph. Eur. k{\"o}nnen verschiedene Nucleotide nachgewiesen werden. Die vorgeschlagene HPLC-Methode wurde durch Einf{\"u}hrung eines Gradienten so ge{\"a}ndert, dass Nucleotide, Nucleoside und Adenin gleichzeitig nachgewiesen werden konnten. Die Analyse der Proben best{\"a}tigte die Ergebnisse der DC, dass kleine Mengen Adenosin-5'-monophosphat in den Chargen vorhanden waren. Eine CE-Methode zur Reinheitspr{\"u}fung von Glutathion wurde entsprechend der Kommentare zum in Pharmeuropa ver{\"o}ffentlichten Monographievorschlag {\"u}berarbeitet. Die kritischen Trennungen zwischen dem internen Standard und Cystein sowie zwischen L-\&\#947;-Glutamyl-L-cystein und dem Hauptpeak, die durch den Systemeignungstest {\"u}berpr{\"u}ft werden, waren durch den pH-Wert des Trennpuffers kontrollierbar. Glutathion zeigt beispielhaft, dass das Verunreinigungsprofil fermentativ gewonnener Produkte komplex ist und von den Herstellungs- und Aufreinigungsprozessen abh{\"a}ngt. Gleiches gilt f{\"u}r Aminos{\"a}uren, die vermehrt biotechnologisch gewonnen werden. In den Aminos{\"a}ure-Monographien ist zur Reinheitspr{\"u}fung eine DC auf „Ninhydrin-positive Substanzen" vorgeschrieben. Diese Methode ist auf den Nachweis anderer Aminos{\"a}uren, die durch unvollst{\"a}ndige Abtrennung bei der Extraktion aus Proteinhydrolysaten stammen k{\"o}nnen, ausgelegt. Die Analyse von Aminos{\"a}uren mittels CE nach Derivatisierung mit FMOC-Cl erm{\"o}glicht einen sensitiven und selektiven Nachweis anderer Aminos{\"a}uren. Durch den Einsatz von CBQCA als Derivatisierungsreagenz k{\"o}nnen auch Aminozucker und kleine Peptide erfasst werden. Mit beiden Methoden wurde das Verunreinigungsprofil von Histidin, Isoleucin, Phenylalanin und Serin bestimmt. W{\"a}hrend in den Phe- und Ser-Proben keine Verunreinigungen erfasst wurden, wurden in den Ile-Proben nach Derivatisierung mit FMOC-Cl andere Aminos{\"a}uren gefunden. Alle untersuchten Aminos{\"a}uren zeigten nach Derivatisierung mit CBQCA viele Verunreinigungen. Wegen Peak{\"u}berlagerungen konnten Aminos{\"a}uren mit dem verwendeten Trennpuffer (25 mM Boratpuffer pH 9.20; 25 mM SDS) nicht bestimmt werden. Durch eine Variation des Puffers (20 mM Tetraboratpuffer pH 9.3; 100 mM SDS) war es m{\"o}glich, die CBQCA-derivatisierten Aminos{\"a}uren zu trennen und zuzuordnen. Im Hinblick auf die Anwesenheit anderer Aminos{\"a}uren waren die Ergebnisse der beiden Verfahren vergleichbar. Insbesondere wurden in den Ile-, Phe- und Ser-Proben keine basischen Aminos{\"a}uren sowie Cystein gefunden. Deren FMOC-Derivate comigrierten mit einem Peak, der nach Strukturaufkl{\"a}rung mittels NMR-Spektroskopie als Nebenprodukt der Derivatisierungsreaktion mit FMOC-Cl identifiziert wurde. Als Verunreinigungen von biotechnologisch hergestellten Aminos{\"a}uren kommen organische S{\"a}uren in Frage. Durch eine RP-HPLC unter Zusatz von Heptafluorbutters{\"a}ure als Ionenpaarreagenz wurden Aminos{\"a}uren und organische S{\"a}uren getrennt. Die Detektion erfolgte mittels ELSD. Nach dem Hauptpeak wurden St{\"o}rsignale detektiert, weshalb sich die Anwendung der Methode zur Reinheitspr{\"u}fung als schwierig herausstellte. Die durchgef{\"u}hrten Experimente deuten darauf hin, dass der ELSD mit der großen Substanzmenge {\"u}berlastet ist und es zur Verschleppung von Substanz kommt.}, subject = {Europ{\"a}isches Arzneibuch}, language = {de} } @phdthesis{Meier2005, author = {Meier, Claudia}, title = {Untersuchungen von Einschlusskomplexen aus Cyclodextrinen mit Aminos{\"a}uren und Dipeptiden}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-14149}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2005}, abstract = {Die Cyclodextrin-modifizierte Kapillarelektrophorese (CE) ist eine wichtige chirale analytische Technik geworden, die zur HPLC und zur Gaschromatographie komplement{\"a}r ist und sich deshalb f{\"u}r die Analyse der Abbauprodukte von Aspartam gut eignet. Ausgehend von diesen Abbauprodukten wurden im Arbeitskreis Scriba an der Universit{\"a}t Jena systematische Studien {\"u}ber die Trennung von Enantiomeren verschiedener Dipeptide mit einer Vielzahl von nativen und derivatisierten Cyclodextrinen bei verschiedenen pH-Werten durchgef{\"u}hrt. Bei der Trennung der Enantiomere von z. B. Ala-Phe oder Ala-Tyr mit beta-Cyclodextrin wurde eine Umkehr der Migrationsreihenfolge bei Erh{\"o}hung des Puffer-pH-Werts von 2,5 auf 3,5 festgestellt. Mit Heptakis-(6-sulfato)-beta-cyclodextrin (HS-beta-CD), Heptakis-(2,3-O-diacetyl)-beta-cyclodextrin (Diac-beta-CD) und Heptakis-(2,3-diacetyl-6-sulfato)-beta-cyclodextrin (HDAS-beta-CD) wurde diese Umkehr der Migrationsreihenfolge nicht beobachtet. Das Ziel der vorliegenden Arbeit war, den Mechanismus der Wechselwirkungen zwischen Cyclodextrinen und Aminos{\"a}uren bzw. Dipeptiden gr{\"u}ndlich und umfassend zu untersuchen. Dabei ging es v. a. um die Untersuchung der Mechanismen der chiralen Erkennung durch Cyclodextrine, die mit Hilfe von verschiedensten Analysenmethoden, vor allem potentiometrische Titrationen und spektroskopische Methoden, wie der NMR-, UV- und CD-Spektroskopie durchgef{\"u}hrt werden sollten. Damit sollte auch die beobachtete Umkehr der Migrationsreihenfolge bei Erh{\"o}hung des pH-Wertes des Laufpuffers in der CE erkl{\"a}rt werden. Die potentiometrische Titrationsmethode lieferte sinnvolle Bindungskonstanten f{\"u}r Cyclodextrin-Einschlusskomplexe mit Aminos{\"a}uren. Eine Analyse der Struktur-Aktivit{\"a}tsbeziehungen f{\"u}r Aminos{\"a}uren und Cyclodextrine ergab, dass ein gewisses Volumen der Aminos{\"a}ure-Seitenkette und damit ein gutes Ausf{\"u}llen der Cyclodextrin-Kavit{\"a}t n{\"o}tig ist, um den vollen Nutzen aus den hydrophoben Wechselwirkungen zwischen der Aminos{\"a}ure-Seitenkette und der Cyclodextrin-Kavit{\"a}t zu ziehen. Eine Verl{\"a}ngerung des hydrophilen Restes, der aus der Kavit{\"a}t herausragt, wie bei den untersuchten Dipeptiden Ala-Phe und Ala-Tyr der Fall, f{\"u}hrt zu der M{\"o}glichkeit der Ausbildung von Wasserstoff-Br{\"u}cken mit den Hydroxylgruppen am breiteren Rand der Cyclodextrin-Kavit{\"a}t und damit zu einer st{\"a}rkeren Bindung an das Cyclodextrin. Um den chiralen Erkennungsprozess von beta-CD und einigen seiner Derivate, n{\"a}mlich HS-beta-CD, Diac-beta-CD und HDAS-beta-CD, zu verstehen, wurden NMR-Experimente durchgef{\"u}hrt, und zwar wurden „durch Komplexbildung induzierte Verschiebungen der chemischen Verschiebungen" (complexation induced chemical shifts, CICS) gemessen und mittels ROESY-Experimenten die Komplexgeometrie untersucht. Betrachtet man die CICS, die f{\"u}r die Paare Diac-beta-CD/Ala-Phe, HDAS-beta-CD/Ala-Phe, Diac-beta-CD/Ala-Tyr und HDAS-beta-CD/Ala-Tyr auftreten, dann zeigt sich, dass sie relativ klein sind und demnach auf eine eher schwache Wechselwirkung des jeweiligen Gastmolek{\"u}ls mit dem Wirt hindeuten. Die CICS f{\"u}r beta-CD- und HS-beta-CD-Dipeptid-Komplexe best{\"a}tigten einen Einschluss des aromatischen Restes in die Cyclodextrin-Kavit{\"a}t. Es konnte gezeigt werden, dass bei pH 2.5 das DD-Enantiomer von Ala-Tyr tiefer in die Kavit{\"a}t von beta-CD eintaucht als das LL-Enantiomer. Außerdem ist bei pH 3.5 die Eintauchtiefe in die Kavit{\"a}t geringer als bei pH 2.5, was durch die Ergebnisse der ROESY-Experimente best{\"a}tigt werden konnte. Um einen besseren Einblick in die Bindungsmodi der Enantiomere von Ala-Phe und Ala-Tyr mit beta-CD bei unterschiedlichen pH-Werten zu erhalten, wurden Molek{\"u}ldynamik-(MD-)-Simulationen durchgef{\"u}hrt. Die Simulationen wurden mit jedem Enantiomer von Ala-Phe und Ala-Tyr in jedem m{\"o}glichen Protonierungszustand durchgef{\"u}hrt, d. h. Kation, Zwitterion und Anion. Zum ersten Mal wurden MD-Simulationen f{\"u}r eine gr{\"o}ßere Serie von unterschiedlichen Komplexen von Enantiomeren in verschiedenen Protonierungszust{\"a}nden systematisch {\"u}ber den langen Zeitraum von 1 ns (=1000 ps) ausgef{\"u}hrt. Die Eintauchtiefe der untersuchten Dipeptide wurde mit Hilfe einer in die Cyclodextrin-Kavit{\"a}t eingepassten Ebene berechnet. Auf diese Weise konnten Informationen {\"u}ber das unterschiedliche Einschlussverhalten der untersuchten Dipeptide erhalten werden. Die angewendete Methode l{\"a}sst sich leicht auf andere Wirt-Gast-Komplexe {\"u}bertragen und erleichtert die Datenerfassung auch in anderen F{\"a}llen. Es konnte gezeigt werden, dass bei pH 2.5 das DD-Enantiomer von Ala-Phe tiefer in die Kavit{\"a}t von beta-Cyclodextrin eintaucht als das LL-Enantiomer, wohingegen bei pH 3.5 der umgekehrte Fall vorliegt. Betrachtet man das Dipeptid Ala-Tyr, dann dringt bei pH 2.5 das DD-Enantiomer tiefer ein, wohingegen keine klare Aussage {\"u}ber das Eindringverhalten der Enantiomere bei pH 3.5 gemacht werden kann. Die CICS und die Ergebnisse der Kapillarelektrophorese weisen jedoch auf ein tieferes Eindringen des LL-Enantiomers hin.}, subject = {Cyclodextrine}, language = {de} } @phdthesis{Novatchev2002, author = {Novatchev, Nikolai}, title = {Untersuchung des Verunreinigungsprofils von Aminos{\"a}uren aus fermentativer Herstellung mittels Kapillarelektrophorese}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-4106}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2002}, abstract = {Gegenstand der vorliegenden Arbeit war die Untersuchungen von Verunreinigungsprofilen von Aminos{\"a}uren aus biotechnologischer Herstellung. Dazu sollten die Aminos{\"a}uren Arg, His, Ile, Lys, Phe, Pro, Ser und Trp von verschiedenen Herstellern und aus unterschiedlichen Batches genauer unter die Lupe genommen werden. Mit der im Europ{\"a}ischen Arzneibuch beschriebenen d{\"u}nnschichtchromatographischen Methode (DC-Methode) f{\"u}r „mit Ninhydrin nachweisbare Substanzen" k{\"o}nnen ausschließlich Fremdaminos{\"a}uren nachgewiesen werden und dies nur, wenn der jeweilige Gehalt relativ hoch ist. Andere Stoffgruppen, die aus der biotechnologischen Herstellung stammen, werden aufgrund der Trennbedingungen sowie der Detektionsart der DC-Methode nicht erfasst. Deshalb mussten neue Methoden entwickelt werden. Laut Vorschriften der International Conference of Harmonisation (ICH) m{\"u}ssen unbekannte Verunreinigungen in Wirkstoffen f{\"u}r orale Therapeutika auf 0,1 \% w/w begrenzt werden k{\"o}nnen. In die Untersuchungen wurden neben den Fremdaminos{\"a}uren auch Peptide, Aminozucker und Nucleins{\"a}uren als potentielle Verunreinigungen, die aus der Herstellung bzw. aus dem Reinigungsprozess kommen, einbezogen. Diese zus{\"a}tzlichen Stoffgruppen k{\"o}nnen aus den „Nebenaktivit{\"a}ten" der Mikroorganismen entstehen. Aminos{\"a}uren, Peptide und Aminozucker wurden versucht, kapillarelektrophoretisch zu quantifizieren. F{\"u}r die Bestimmung von Nucleins{\"a}uren wurde zus{\"a}tzlich die UV-Spektroskopie eingesetzt.}, subject = {Aminos{\"a}uren}, language = {de} }