@article{ReiterGenslerRitteretal.2012, author = {Reiter, Theresa and Gensler, Daniel and Ritter, Oliver and Weiss, Ingo and Geistert, Wolfgang and Kaufmann, Ralf and Hoffmeister, Sabine and Friedrich, Michael T. and Wintzheimer, Stefan and D{\"u}ring, Markus and Nordbeck, Peter and Jakob, Peter M. and Ladd, Mark E. and Quick, Harald H. and Bauer, Wolfgang R.}, title = {Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures}, series = {Journal of Cardiovascular Magnetic Resonance}, volume = {14}, journal = {Journal of Cardiovascular Magnetic Resonance}, number = {12}, doi = {10.1186/1532-429X-14-12}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-134927}, year = {2012}, abstract = {Background: One of the safety concerns when performing electrophysiological (EP) procedures under magnetic resonance (MR) guidance is the risk of passive tissue heating due to the EP catheter being exposed to the radiofrequency (RF) field of the RF transmitting body coil. Ablation procedures that use catheters with irrigated tips are well established therapeutic options for the treatment of cardiac arrhythmias and when used in a modified mode might offer an additional system for suppressing passive catheter heating. Methods: A two-step approach was chosen. Firstly, tests on passive catheter heating were performed in a 1.5 T Avanto system (Siemens Healthcare Sector, Erlangen, Germany) using a ASTM Phantom in order to determine a possible maximum temperature rise. Secondly, a phantom was designed for simulation of the interface between blood and the vascular wall. The MR-RF induced temperature rise was simulated by catheter tip heating via a standard ablation generator. Power levels from 1 to 6 W were selected. Ablation duration was 120 s with no tip irrigation during the first 60 s and irrigation at rates from 2 ml/min to 35 ml/min for the remaining 60 s (Biotronik Qiona Pump, Berlin, Germany). The temperature was measured with fluoroscopic sensors (Luxtron, Santa Barbara, CA, USA) at a distance of 0 mm, 2 mm, 4 mm, and 6 mm from the catheter tip. Results: A maximum temperature rise of 22.4 degrees C at the catheter tip was documented in the MR scanner. This temperature rise is equivalent to the heating effect of an ablator's power output of 6 W at a contact force of the weight of 90 g (0.883 N). The catheter tip irrigation was able to limit the temperature rise to less than 2 degrees C for the majority of examined power levels, and for all examined power levels the residual temperature rise was less than 8 degrees C. Conclusion: Up to a maximum of 22.4 degrees C, the temperature rise at the tissue surface can be entirely suppressed by using the catheter's own irrigation system. The irrigated tip system can be used to increase MR safety of EP catheters by suppressing the effects of unwanted passive catheter heating due to RF exposure from the MR scanner.}, language = {en} } @article{WinterKampfHelluyetal.2013, author = {Winter, Patrick and Kampf, Thomas and Helluy, Xavier and Gutjahr, Fabian T. and Meyer, Cord B. and Rommel, Eberhard and Bauer, Wolfgang R. and Jakob, Peter M. and Herold, Volker}, title = {Fast retrospectively triggered local pulse-wave velocity measurements in mice with CMR-microscopy using a radial trajectory}, series = {Journal of Cardiovascular Magnetic Resonance}, journal = {Journal of Cardiovascular Magnetic Resonance}, doi = {10.1186/1532-429X-15-88}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-96602}, year = {2013}, abstract = {Background The aortic pulse-wave velocity (PWV) is an important indicator of cardiovascular risk. In recent studies MRI methods have been developed to measure this parameter noninvasively in mice. Present techniques require additional hardware for cardiac and respiratory gating. In this work a robust self-gated measurement of the local PWV in mice without the need of triggering probes is proposed. Methods The local PWV of 6-months-old wild-type C57BL/6J mice (n=6) was measured in the abdominal aorta with a retrospectively triggered radial Phase Contrast (PC) MR sequence using the flow-area (QA) method. A navigator signal was extracted from the CMR data of highly asymmetric radial projections with short repetition time (TR=3 ms) and post-processed with high-pass and low-pass filters for retrospective cardiac and respiratory gating. The self-gating signal was used for a reconstruction of high-resolution Cine frames of the aortic motion. To assess the local PWV the volume flow Q and the cross-sectional area A of the aorta were determined. The results were compared with the values measured with a triggered Cartesian and an undersampled triggered radial PC-Cine sequence. Results In all examined animals a self-gating signal could be extracted and used for retrospective breath-gating and PC-Cine reconstruction. With the non-triggered measurement PWV values of 2.3±0.2 m/s were determined. These values are in agreement with those measured with the triggered Cartesian (2.4±0.2 m/s) and the triggered radial (2.3±0.2 m/s) measurement. Due to the strong robustness of the radial trajectory against undersampling an acceleration of more than two relative to the prospectively triggered Cartesian sampling could be achieved with the retrospective method. Conclusion With the radial flow-encoding sequence the extraction of a self-gating signal is feasible. The retrospective method enables a robust and fast measurement of the local PWV without the need of additional trigger hardware.}, language = {en} } @article{NordbeckBoenhofHilleretal.2013, author = {Nordbeck, Peter and B{\"o}nhof, Leoni and Hiller, Karl-Heinz and Voll, Sabine and Arias-Loza, Paula and Seidlmaier, Lea and Williams, Tatjana and Ye, Yu-Xiang and Gensler, Daniel and Pelzer, Theo and Ertl, Georg and Jakob, Peter M. and Bauer, Wolfgang R. and Ritter, Oliver}, title = {Impact of Thoracic Surgery on Cardiac Morphology and Function in Small Animal Models of Heart Disease: A Cardiac MRI Study in Rats}, series = {PLoS ONE}, volume = {8}, journal = {PLoS ONE}, number = {8}, doi = {10.1371/journal.pone.0068275}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-130064}, pages = {e68275}, year = {2013}, abstract = {Background Surgical procedures in small animal models of heart disease might evoke alterations in cardiac morphology and function. The aim of this study was to reveal and quantify such potential artificial early or long term effects in vivo, which might account for a significant bias in basic cardiovascular research, and, therefore, could potentially question the meaning of respective studies. Methods Female Wistar rats (n = 6 per group) were matched for weight and assorted for sham left coronary artery ligation or control. Cardiac morphology and function was then investigated in vivo by cine magnetic resonance imaging at 7 Tesla 1 and 8 weeks after the surgical procedure. The time course of metabolic and inflammatory blood parameters was determined in addition. Results Compared to healthy controls, rats after sham surgery showed a lower body weight both 1 week (267.5±10.6 vs. 317.0±11.3 g, n<0.05) and 8 weeks (317.0±21.1 vs. 358.7±22.4 g, n<0.05) after the intervention. Left and right ventricular morphology and function were not different in absolute measures in both groups 1 week after surgery. However, there was a confined difference in several cardiac parameters normalized to the body weight (bw), such as myocardial mass (2.19±0.30/0.83±0.13 vs. 1.85±0.22/0.70±0.07 mg left/right per g bw, p<0.05), or enddiastolic ventricular volume (1.31±0.36/1.21±0.31 vs. 1.14±0.20/1.07±0.17 µl left/right per g bw, p<0.05). Vice versa, after 8 weeks, cardiac masses, volumes, and output showed a trend for lower values in sham operated rats compared to controls in absolute measures (782.2±57.2/260.2±33.2 vs. 805.9±84.8/310.4±48.5 mg, p<0.05 for left/right ventricular mass), but not normalized to body weight. Matching these findings, blood testing revealed only minor inflammatory but prolonged metabolic changes after surgery not related to cardiac disease. Conclusion Cardio-thoracic surgical procedures in experimental myocardial infarction cause distinct alterations upon the global integrity of the organism, which in the long term also induce circumscribed repercussions on cardiac morphology and function. This impact has to be considered when analyzing data from respective animal studies and transferring these findings to conditions in patients.}, language = {en} } @article{DawoodBreuerStebanietal.2023, author = {Dawood, Peter and Breuer, Felix and Stebani, Jannik and Burd, Paul and Homolya, Istv{\´a}n and Oberberger, Johannes and Jakob, Peter M. and Blaimer, Martin}, title = {Iterative training of robust k-space interpolation networks for improved image reconstruction with limited scan specific training samples}, series = {Magnetic Resonance in Medicine}, volume = {89}, journal = {Magnetic Resonance in Medicine}, number = {2}, doi = {10.1002/mrm.29482}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-312306}, pages = {812 -- 827}, year = {2023}, abstract = {To evaluate an iterative learning approach for enhanced performance of robust artificial-neural-networks for k-space interpolation (RAKI), when only a limited amount of training data (auto-calibration signals [ACS]) are available for accelerated standard 2D imaging. Methods In a first step, the RAKI model was tailored for the case of limited training data amount. In the iterative learning approach (termed iterative RAKI [iRAKI]), the tailored RAKI model is initially trained using original and augmented ACS obtained from a linear parallel imaging reconstruction. Subsequently, the RAKI convolution filters are refined iteratively using original and augmented ACS extracted from the previous RAKI reconstruction. Evaluation was carried out on 200 retrospectively undersampled in vivo datasets from the fastMRI neuro database with different contrast settings. Results For limited training data (18 and 22 ACS lines for R = 4 and R = 5, respectively), iRAKI outperforms standard RAKI by reducing residual artifacts and yields better noise suppression when compared to standard parallel imaging, underlined by quantitative reconstruction quality metrics. Additionally, iRAKI shows better performance than both GRAPPA and standard RAKI in case of pre-scan calibration with varying contrast between training- and undersampled data. Conclusion RAKI benefits from the iterative learning approach, which preserves the noise suppression feature, but requires less original training data for the accurate reconstruction of standard 2D images thereby improving net acceleration.}, language = {en} } @article{GotschyBauerWinteretal.2017, author = {Gotschy, Alexander and Bauer, Wolfgang R. and Winter, Patrick and Nordbeck, Peter and Rommel, Eberhard and Jakob, Peter M. and Herold, Volker}, title = {Local versus global aortic pulse wave velocity in early atherosclerosis: An animal study in ApoE\(^{-/-}\) mice using ultrahigh field MRI}, series = {PLoS ONE}, volume = {12}, journal = {PLoS ONE}, number = {2}, doi = {10.1371/journal.pone.0171603}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-171824}, year = {2017}, abstract = {Increased aortic stiffness is known to be associated with atherosclerosis and has a predictive value for cardiovascular events. This study aims to investigate the local distribution of early arterial stiffening due to initial atherosclerotic lesions. Therefore, global and local pulse wave velocity (PWV) were measured in ApoE\(^{-/-}\) and wild type (WT) mice using ultrahigh field MRI. For quantification of global aortic stiffness, a new multi-point transit-time (TT) method was implemented and validated to determine the global PWV in the murine aorta. Local aortic stiffness was measured by assessing the local PWV in the upper abdominal aorta, using the flow/area (QA) method. Significant differences between age matched ApoE\(^{-/-}\) and WT mice were determined for global and local PWV measurements (global PWV: ApoE\(^{-/-}\): 2.7 ±0.2m/s vs WT: 2.1±0.2m/s, P<0.03; local PWV: ApoE\(^{-/-}\): 2.9±0.2m/s vs WT: 2.2±0.2m/s, P<0.03). Within the WT mouse group, the global PWV correlated well with the local PWV in the upper abdominal aorta (R\(^2\) = 0.75, P<0.01), implying a widely uniform arterial elasticity. In ApoE\(^{-/-}\) animals, however, no significant correlation between individual local and global PWV was present (R\(^2\) = 0.07, P = 0.53), implying a heterogeneous distribution of vascular stiffening in early atherosclerosis. The assessment of global PWV using the new multi-point TT measurement technique was validated against a pressure wire measurement in a vessel phantom and showed excellent agreement. The experimental results demonstrate that vascular stiffening caused by early atherosclerosis is unequally distributed over the length of large vessels. This finding implies that assessing heterogeneity of arterial stiffness by multiple local measurements of PWV might be more sensitive than global PWV to identify early atherosclerotic lesions.}, language = {en} }