@phdthesis{Janssen2023, author = {Janßen, Jan Paul}, title = {Capabilities of a multi-pinhole SPECT system with two stationary detectors for in vivo imaging in rodents}, doi = {10.25972/OPUS-32860}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-328608}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2023}, abstract = {Molecular imaging of rats is of great importance for basic and translational research. As a powerful tool in nuclear medicine, SPECT can be used to visualize specific functional processes in the body, such as myocardial perfusion or bone metabolism. Typical applications in laboratory animals are imaging diagnostics or the development of new tracers for clinical use. Innovations have enabled resolutions of up to a quarter of a millimeter with acceptable sensitivity. These advances have recently led to significantly more interest in SPECT both clinically and preclinically. The objective of this thesis was to evaluate the performance of the new U-SPECT5/CT E-Class by MILabs with a dedicated ultra-high resolution multi-pinhole collimator for rats and its potential for in vivo imaging of rats. The unique features of the U-SPECT are the large stationary detectors and the new iterative reconstruction algorithm. In addition, compared to the conventional system, the "E-Class" uses only two detectors instead of three. First, the sensitivity, maximum resolution, and uniformity were determined as performance parameters. Thereafter, CNRs for different activity levels comparable to those of typical in vivo activities were examined. Finally, two example protocols were carried out for imaging with 99mTc-MIBI and 99mTc-HMDP in healthy rats to evaluate the in vivo capabilities. For this purpose, CNR calculations and an image quality assessment were performed. The focus was on image quality as a function of scan time and post-reconstruction filter across a wide range of realistically achievable in vivo conditions. Performance was reasonable compared to other systems in the literature, with a sensitivity of 567 cps/MBq, a maximum resolution of 1.20 mm, and a uniformity of 55.5\%. At the lower activities, resolution in phantom studies decreased to ≥1.80 mm while maintaining good image quality. High-quality bone and myocardial perfusion SPECTs were obtained in rats with a resolution of ≥1.80 mm and ≥2.20 mm, respectively. Although limited sensitivity remains a weakness of SPECT, the U-SPECT5/CT E-Class with the UHR-RM collimator can achieve in vivo results of the highest standard despite the missing third detector. Currently, it is one of the best options for high-resolution radionuclide imaging in rats.}, subject = {SPECT}, language = {en} } @phdthesis{Hoffmann2023, author = {Hoffmann, Jan Vincent}, title = {Small-animal SPECT with Two Stationary Detectors: Performance Evaluation and Image Quality Assessment of Multi-pinhole Collimators}, doi = {10.25972/OPUS-32819}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-328195}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2023}, abstract = {SPECT as a representative of molecular imaging allows visualization of metabolic processes in vivo. In clinical practice, single photon emission imaging is an established modality for myocardial perfusion imaging or the diagnosis of adrenal or neuroendocrine tumors, to name a few. With technical advances in scanner design and data processing leading to improved spatial resolution and image quality, SPECT has become a serious contender in small animal preclinical imaging. With multi-pinhole collimation, submillimeter spatial resolutions are achieved without limiting sensitivity, which has led to a significant increase of interest in SPECT for preclinical research in recent years. In this dissertation, the potential of a two-detector system through an analysis of three dedicated mouse collimators with multi-pinhole configurations was demonstrated. For this, sensitivity, spatial resolution, and uniformity as key parameters were determined. In the second part of the present work, an evaluation of the image quality at different activity concentrations to allow prediction of the system performance related to in vivo studies was performed. Therefore, a visual evaluation, as well as a calculation of the contrastto-noise ratio, was performed using mini Derenzo phantoms for the respective three mouse collimators. To better classify the results, the study was extended by a comparison with the predecessor system. Due to the absence of the third bottom detector, sensitivity and uniformity are slightly compromised. All three collimators were able to achieve a spatial resolution in the submillimeter range, XUHR-M offers a peak resolution of up to 0.35 mm. In terms of resolution, both evaluated systems performed on an equal level. Visual assessment of image quality indicates a slight advantage of the new two-detector system, and the contrast-to-noise ratio seems to benefit from the improved SROSEM algorithm. However, the differences between the two systems are marginal. The U-SPECT5/CT E-Class is proven to be state-of-the-art for small animal imaging and is a powerful instrument for preclinical molecular imaging research. Improvements in system design compensate well for the reduction in the detection area, allowing excellent imaging even with low activity concentrations.}, subject = {SPECT}, language = {en} }