@article{HennesHuflageGrunzetal.2023, author = {Hennes, Jan-Lucca and Huflage, Henner and Grunz, Jan-Peter and Hartung, Viktor and Augustin, Anne Marie and Patzer, Theresa Sophie and Pannenbecker, Pauline and Petritsch, Bernhard and Bley, Thorsten Alexander and Gruschwitz, Philipp}, title = {An intra-individual comparison of low-keV photon-counting CT versus energy-integrating-detector CT angiography of the aorta}, series = {Diagnostics}, volume = {13}, journal = {Diagnostics}, number = {24}, issn = {2075-4418}, doi = {10.3390/diagnostics13243645}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-355568}, year = {2023}, abstract = {This retrospective study aims to provide an intra-individual comparison of aortic CT angiographies (CTAs) using first-generation photon-counting-detector CT (PCD-CT) and third-generation energy-integrating-detector CT (EID-CT). High-pitch CTAs were performed with both scanners and equal contrast-agent protocols. EID-CT employed automatic tube voltage selection (90/100 kVp) with reference tube current of 434/350 mAs, whereas multi-energy PCD-CT scans were generated with fixed tube voltage (120 kVp), image quality level of 64, and reconstructed as 55 keV monoenergetic images. For image quality assessment, contrast-to-noise ratios (CNRs) were calculated, and subjective evaluation (overall quality, luminal contrast, vessel sharpness, blooming, and beam hardening) was performed independently by three radiologists. Fifty-seven patients (12 women, 45 men) were included with a median interval between examinations of 12.7 months (interquartile range 11.1 months). Using manufacturer-recommended scan protocols resulted in a substantially lower radiation dose in PCD-CT (size-specific dose estimate: 4.88 ± 0.48 versus 6.28 ± 0.50 mGy, p < 0.001), while CNR was approximately 50\% higher (41.11 ± 8.68 versus 27.05 ± 6.73, p < 0.001). Overall image quality and luminal contrast were deemed superior in PCD-CT (p < 0.001). Notably, EID-CT allowed for comparable vessel sharpness (p = 0.439) and less pronounced blooming and beam hardening (p < 0.001). Inter-rater agreement was good to excellent (0.58-0.87). Concluding, aortic PCD-CTAs facilitate increased image quality with significantly lower radiation dose compared to EID-CTAs}, language = {en} } @article{GruschwitzHartungKleefeldtetal.2023, author = {Gruschwitz, Philipp and Hartung, Viktor and Kleefeldt, Florian and Erg{\"u}n, S{\"u}leyman and Lichthardt, Sven and Huflage, Henner and Hendel, Robin and Kunz, Andreas Steven and Pannenbecker, Pauline and Kuhl, Philipp Josef and Augustin, Anne Marie and Bley, Thorsten Alexander and Petritsch, Bernhard and Grunz, Jan-Peter}, title = {Standardized assessment of vascular reconstruction kernels in photon-counting CT angiographies of the leg using a continuous extracorporeal perfusion model}, series = {Scientific Reports}, volume = {13}, journal = {Scientific Reports}, doi = {10.1038/s41598-023-39063-z}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-357912}, year = {2023}, abstract = {This study evaluated the influence of different vascular reconstruction kernels on the image quality of CT angiographies of the lower extremity runoff using a 1st-generation photon-counting-detector CT (PCD-CT) compared with dose-matched examinations on a 3rd-generation energy-integrating-detector CT (EID-CT). Inducing continuous extracorporeal perfusion in a human cadaveric model, we performed CT angiographies of eight upper leg arterial runoffs with radiation dose-equivalent 120 kVp acquisition protocols (CTDIvol 5 mGy). Reconstructions were executed with different vascular kernels, matching the individual modulation transfer functions between scanners. Signal-to-noise-ratios (SNR) and contrast-to-noise-ratios (CNR) were computed to assess objective image quality. Six radiologists evaluated image quality subjectively using a forced-choice pairwise comparison tool. Interrater agreement was determined by calculating Kendall's concordance coefficient (W). The intraluminal attenuation of PCD-CT images was significantly higher than of EID-CT (414.7 ± 27.3 HU vs. 329.3 ± 24.5 HU; p < 0.001). Using comparable kernels, image noise with PCD-CT was significantly lower than with EID-CT (p ≤ 0.044). Correspondingly, SNR and CNR were approximately twofold higher for PCD-CT (p < 0.001). Increasing the spatial frequency for PCD-CT reconstructions by one level resulted in similar metrics compared to EID-CT (CNRfat; EID-CT Bv49: 21.7 ± 3.7 versus PCD-CT Bv60: 21.4 ± 3.5). Overall image quality of PCD-CTA achieved ratings superior to EID-CTA irrespective of the used reconstruction kernels (best: PCD-CT Bv60; worst: EID-CT Bv40; p < 0.001). Interrater agreement was good (W = 0.78). Concluding, PCD-CT offers superior intraluminal attenuation, SNR, and CNR compared to EID-CT in angiographies of the upper leg arterial runoff. Combined with improved subjective image quality, PCD-CT facilitates the use of sharper convolution kernels and ultimately bears the potential of improved vascular structure assessability.}, language = {en} } @article{HuflageGrunzPatzeretal.2023, author = {Huflage, Henner and Grunz, Jan-Peter and Patzer, Theresa Sophie and Pannenbecker, Pauline and Feldle, Philipp and Sauer, Stephanie Tina and Petritsch, Bernhard and Erg{\"u}n, S{\"u}leyman and Bley, Thorsten Alexander and Kunz, Andreas Steven}, title = {Potential of unenhanced ultra-low-dose abdominal photon-counting CT with tin filtration: a cadaveric study}, series = {Diagnostics}, volume = {13}, journal = {Diagnostics}, number = {4}, issn = {2075-4418}, doi = {10.3390/diagnostics13040603}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-304122}, year = {2023}, abstract = {Objectives: This study investigated the feasibility and image quality of ultra-low-dose unenhanced abdominal CT using photon-counting detector technology and tin prefiltration. Materials and Methods: Employing a first-generation photon-counting CT scanner, eight cadaveric specimens were examined both with tin prefiltration (Sn 100 kVp) and polychromatic (120 kVp) scan protocols matched for radiation dose at three different levels: standard-dose (3 mGy), low-dose (1 mGy) and ultra-low-dose (0.5 mGy). Image quality was evaluated quantitatively by means of contrast-to-noise-ratios (CNR) with regions of interest placed in the renal cortex and subcutaneous fat. Additionally, three independent radiologists performed subjective evaluation of image quality. The intraclass correlation coefficient was calculated as a measure of interrater reliability. Results: Irrespective of scan mode, CNR in the renal cortex decreased with lower radiation dose. Despite similar mean energy of the applied x-ray spectrum, CNR was superior for Sn 100 kVp over 120 kVp at standard-dose (17.75 ± 3.51 vs. 14.13 ± 4.02), low-dose (13.99 ± 2.6 vs. 10.68 ± 2.17) and ultra-low-dose levels (8.88 ± 2.01 vs. 11.06 ± 1.74) (all p ≤ 0.05). Subjective image quality was highest for both standard-dose protocols (score 5; interquartile range 5-5). While no difference was ascertained between Sn 100 kVp and 120 kVp examinations at standard and low-dose levels, the subjective image quality of tin-filtered scans was superior to 120 kVp with ultra-low radiation dose (p < 0.05). An intraclass correlation coefficient of 0.844 (95\% confidence interval 0.763-0.906; p < 0.001) indicated good interrater reliability. Conclusions: Photon-counting detector CT permits excellent image quality in unenhanced abdominal CT with very low radiation dose. Employment of tin prefiltration at 100 kVp instead of polychromatic imaging at 120 kVp increases the image quality even further in the ultra-low-dose range of 0.5 mGy.}, language = {en} } @article{MuellerKoehlerHendricksetal.2021, author = {M{\"u}ller, Sophie and K{\"o}hler, Franziska and Hendricks, Anne and Kastner, Carolin and B{\"o}rner, Kevin and Diers, Johannes and Lock, Johan F. and Petritsch, Bernhard and Germer, Christoph-Thomas and Wiegering, Armin}, title = {Brain metastases from colorectal cancer: a systematic review of the literature and meta-analysis to establish a guideline for daily treatment}, series = {Cancers}, volume = {13}, journal = {Cancers}, number = {4}, issn = {2072-6694}, doi = {10.3390/cancers13040900}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-228883}, year = {2021}, abstract = {Colorectal cancer (CRC) is the third most common malignancy worldwide. Most patients with metastatic CRC develop liver or lung metastases, while a minority suffer from brain metastases. There is little information available regarding the presentation, treatment, and overall survival of brain metastases (BM) from CRC. This systematic review and meta-analysis includes data collected from three major databases (PubMed, Cochrane, and Embase) based on the key words "brain", "metastas*", "tumor", "colorectal", "cancer", and "malignancy". In total, 1318 articles were identified in the search and 86 studies matched the inclusion criteria. The incidence of BM varied between 0.1\% and 11.5\%. Most patients developed metastases at other sites prior to developing BM. Lung metastases and KRAS mutations were described as risk factors for additional BM. Patients with BM suffered from various symptoms, but up to 96.8\% of BM patients were asymptomatic at the time of BM diagnosis. Median survival time ranged from 2 to 9.6 months, and overall survival (OS) increased up to 41.1 months in patients on a multimodal therapy regimen. Several factors including age, blood levels of carcinoembryonic antigen (CEA), multiple metastases sites, number of brain lesions, and presence of the KRAS mutation were predictors of OS. For BM diagnosis, MRI was considered to be state of the art. Treatment consisted of a combination of surgery, radiation, or systemic treatment.}, language = {en} } @article{HartrampfPetritschBucketal.2020, author = {Hartrampf, Philipp E. and Petritsch, Bernhard and Buck, Andreas K. and Serfling, Sebastian E.}, title = {Pitfalls in PSMA-PET/CT: Intensive bone-marrow uptake in a case with polycythaemia vera}, series = {European Journal of Nuclear Medicine and Molecular Imaging}, volume = {48}, journal = {European Journal of Nuclear Medicine and Molecular Imaging}, issn = {1619-7070}, doi = {10.1007/s00259-020-05072-7}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-235608}, pages = {1669-1670}, year = {2020}, abstract = {No abstract available.}, language = {en} } @article{GrunzPennigFieberetal.2021, author = {Grunz, Jan-Peter and Pennig, Lenhard and Fieber, Tabea and Gietzen, Carsten Herbert and Heidenreich, Julius Frederik and Huflage, Henner and Gruschwitz, Philipp and Kuhl, Philipp Josef and Petritsch, Bernhard and Kosmala, Aleksander and Bley, Thorsten Alexander and Gassenmaier, Tobias}, title = {Twin robotic x-ray system in small bone and joint trauma: Impact of cone-beam computed tomography on treatment decisions}, series = {European Radiology}, volume = {31}, journal = {European Radiology}, issn = {0938-7994}, doi = {10.1007/s00330-020-07563-5}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-235233}, pages = {3600-3609}, year = {2021}, abstract = {Objectives Trauma evaluation of extremities can be challenging in conventional radiography. A multi-use x-ray system with cone-beam CT (CBCT) option facilitates ancillary 3-D imaging without repositioning. We assessed the clinical value of CBCT scans by analyzing the influence of additional findings on therapy. Methods Ninety-two patients underwent radiography and subsequent CBCT imaging with the twin robotic scanner (76 wrist/hand/finger and 16 ankle/foot/toe trauma scans). Reports by on-call radiologists before and after CBCT were compared regarding fracture detection, joint affliction, comminuted injuries, and diagnostic confidence. An orthopedic surgeon recommended therapy based on reported findings. Surgical reports (N = 52) and clinical follow-up (N = 85) were used as reference standard. Results CBCT detected more fractures (83/64 of 85), joint involvements (69/53 of 71), and multi-fragment situations (68/50 of 70) than radiography (all p < 0.001). Six fractures suspected in radiographs were ruled out by CBCT. Treatment changes based on additional information from CBCT were recommended in 29 patients (31.5\%). While agreement between advised therapy before CBCT and actual treatment was moderate (κ = 0.41 [95\% confidence interval 0.35-0.47]; p < 0.001), agreement after CBCT was almost perfect (κ = 0.88 [0.83-0.93]; p < 0.001). Diagnostic confidence increased considerably for CBCT studies (p < 0.001). Median effective dose for CBCT was 4.3 μSv [3.3-5.3 μSv] compared to 0.2 μSv [0.1-0.2 μSv] for radiography. Conclusions CBCT provides advantages for the evaluation of acute small bone and joint trauma by detecting and excluding extremity fractures and fracture-related findings more reliably than radiographs. Additional findings induced therapy change in one third of patients, suggesting substantial clinical impact.}, language = {en} } @article{PetritschPannbeckerWengetal.2021, author = {Petritsch, Bernhard and Pannbecker, Pauline and Weng, Andreas M. and Grunz, Jan-Peter and Veldhoen, Simon and Bley, Thorsten A. and Kosmala, Aleksander}, title = {Split-filter dual-energy CT pulmonary angiography for the diagnosis of acute pulmonary embolism: a study on image quality and radiation dose}, series = {Quantitative Imaging in Medicine and Surgery}, volume = {11}, journal = {Quantitative Imaging in Medicine and Surgery}, number = {5}, doi = {10.21037/qims-20-740}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-231456}, pages = {1817-1827}, year = {2021}, abstract = {Background: Computed tomography (CT) pulmonary angiography is the diagnostic reference standard in suspected pulmonary embolism (PE). Favorable results for dual-energy CT (DECT) images have been reported for this condition. Nowadays, dual-energy data acquisition is feasible with different technical options, including a single-source split-filter approach. Therefore, the aim of this retrospective study was to investigate image quality and radiation dose of thoracic split-filter DECT in comparison to conventional single-energy CT in patients with suspected PE. Methods: A total of 110 CT pulmonary angiographies were accomplished either as standard single-energy CT with automatic tube voltage selection (ATVS) (n=58), or as split-filter DECT (n=52). Objective [pulmonary artery CT attenuation, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR)] and subjective image quality [four-point Likert scale; three readers (R)] were compared among the two study groups. Size-specific dose estimates (SSDE), dose-length-product (DLP) and volume CT dose index (CTDIvol) were assessed for radiation dose analysis. Results: Split-filter DECT images yielded 67.7\% higher SNR (27.0 vs. 16.1; P<0.001) and 61.9\% higher CNR (22.5 vs. 13.9; P<0.001) over conventional single-energy images, whereas CT attenuation was significantly lower (344.5 vs. 428.2 HU; P=0.013). Subjective image quality was rated good or excellent in 93.0\%/98.3\%/77.6\% (R1/R2/R3) of the single-energy CT scans, and 84.6\%/82.7\%/80.8\% (R1/R2/R3) of the split-filter DECT scans. SSDE, DLP and CTDIvol were significantly lower for conventional single-energy CT compared to split-filter DECT (all P<0.05), which was associated with 26.7\% higher SSDE. Conclusions: In the diagnostic workup of acute PE, the split-filter allows for dual-energy data acquisition from single-source single-layer CT scanners. The existing opportunity to assess pulmonary "perfusion" based on analysis of iodine distribution maps is associated with higher radiation dose in terms of increased SSDE than conventional single-energy CT with ATVS. Moreover, a proportion of up to 3.8\% non-diagnostic examinations in the current reference standard test for PE is not negligible.}, language = {en} } @article{PetritschKoestlerGassenmaieretal.2016, author = {Petritsch, Bernhard and K{\"o}stler, Herbert and Gassenmaier, Tobias and Kunz, Andreas S and Bley, Thorsten A and Horn, Michael}, title = {An investigation into potential gender-specific differences in myocardial triglyceride content assessed by \(^{1}\)H-Magnetic Resonance Spectroscopy at 3Tesla}, series = {Journal of International Medical Research}, volume = {44}, journal = {Journal of International Medical Research}, number = {3}, doi = {10.1177/0300060515603884}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-168808}, pages = {585-591}, year = {2016}, abstract = {Objective: Over the past decade, myocardial triglyceride content has become an accepted biomarker for chronic metabolic and cardiac disease. The purpose of this study was to use proton (hydrogen 1)-magnetic resonance spectroscopy (\(^{1}\)H-MRS) at 3Tesla (3 T) field strength to assess potential gender-related differences in myocardial triglyceride content in healthy individuals. Methods: Cardiac MR imaging was performed to enable accurate voxel placement and obtain functional and morphological information. Double triggered (i.e., ECG and respiratory motion gating) \(^{1}\)H-MRS was used to quantify myocardial triglyceride levels for each gender. Two-sample t-test and Mann-Whitney U-test were used for statistical analyses. Results: In total, 40 healthy volunteers (22 male, 18 female; aged >18 years and age matched) were included in the study. Median myocardial triglyceride content was 0.28\% (interquartile range [IQR] 0.17-0.42\%) in male and 0.24\% (IQR 0.14-0.45\%) in female participants, and no statistically significant difference was observed between the genders. Furthermore, no gender-specific difference in ejection fraction was observed, although on average, male participants presented with a higher mean ± SD left ventricular mass (136.3 ± 25.2 g) than female participants (103.9 ± 16.1 g). Conclusions: The study showed that \(^{1}\)H-MRS is a capable, noninvasive tool for acquisition of myocardial triglyceride metabolites. Myocardial triglyceride concentration was shown to be unrelated to gender in this group of healthy volunteers.}, language = {en} } @article{PetritschGoltzHahnetal.2011, author = {Petritsch, Bernhard and Goltz, Jan Peter and Hahn, Dietbert and Wendel, Frank}, title = {Extensive craniocervical bone pneumatization}, series = {Diagnostic and Interventional Radiology}, volume = {17}, journal = {Diagnostic and Interventional Radiology}, number = {4}, doi = {10.4261/1305-3825.DIR.4299-11.2}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-139349}, pages = {308-310}, year = {2011}, abstract = {We report a case of extensive abnormal craniocervical bone pneumatization accidentally found in a patient without any history of trauma or surgery. The patient had only mild unspecific thoracic pain and bilateral paresthesia that did not correlate with computed tomography findings.}, language = {en} } @article{OderUeceylerLiuetal.2016, author = {Oder, Daniel and {\"U}ceyler, Nurcan and Liu, Dan and Hu, Kai and Petritsch, Bernhard and Sommer, Claudia and Ertl, Georg and Wanner, Christoph and Nordbeck, Peter}, title = {Organ manifestations and long-term outcome of Fabry disease in patients with the GLA haplotype D313Y}, series = {BMJ Open}, volume = {6}, journal = {BMJ Open}, doi = {10.1136/bmjopen-2015-010422}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-161210}, pages = {e010422}, year = {2016}, abstract = {Objectives: The severity of Fabry disease is dependent on the type of mutation in the α-galactosidase A (AgalA) encoding gene (GLA). This study focused on the impact of the GLA haplotype D313Y on long-term organ involvement and function. Setting and participants: In this monocentric study, all participants presenting with the D313Y haplotype between 2001 and 2015 were comprehensively clinically investigated at baseline and during a 4-year follow-up if available. Five females and one male were included. Primary and secondary outcome measures: Cardiac, nephrological, neurological, laboratory and quality of life data. Results: AgalA enzyme activity in leucocytes (0.3±0.9 nmol/min/mg protein (mean±SD)) and serum lyso-Gb3 (0.6±0.3 ng/mL at baseline) were in normal range in all patients. Cardiac morphology and function were normal (left-ventricular (LV) ejection fraction 66±8\%; interventricular septum 7.7±1.4 mm; LV posterior wall 7.5±1.4 mm; normalised LV mass in MRI 52±9 g/m2; LV global longitudinal strain -21.6±1.9\%) and there were no signs of myocardial fibrosis in cardiac MRI. Cardiospecific biomarkers were also in normal range. Renal function was not impaired (estimated glomerular filtration rate MDRD 103±15 mL/min; serum-creatinine 0.75±0.07 mg/dL; cystatin-c 0.71±0.12 mg/L). One female patient (also carrying a Factor V Leiden mutation) had a transitory ischaemic attack. One patient showed white matter lesions in brain MRI, but none had Fabry-associated pain attacks, pain crises, evoked pain or permanent pain. Health-related quality of life analysis revealed a reduction in individual well-being. At long-term follow-up after 4 years, no significant change was seen in any parameter. Conclusions: The results of the current study suggest that the D313Y genotype does not lead to severe organ manifestations as seen in genotypes known to be causal for classical FD."}, language = {en} }