@phdthesis{Leistner2023, author = {Leistner, Adrian Dieter}, title = {Improving the quality analysis of monographed drugs - dapsone, baclofen, acarbose and other selected APIs}, doi = {10.25972/OPUS-30331}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:20-opus-303318}, school = {Universit{\"a}t W{\"u}rzburg}, year = {2023}, abstract = {All presented studies aimed on the improvement of the quality analysis of already monographed drugs. Thereby different LC methods were applied and coupled to i.e., the UV/VIS detector, the CAD or a hyphenation of these detectors, respectively. The choice of the chromatographic system including the detector was largely dependent on the physicochemical properties of the respective analytes. With the risk-assessment report on the API cetirizine we presented an exemplary tool, that can help to minimize the risk of the occurrence of unexpected impurities. An in- deep analysis of each step within synthesis pathway by means of reaction matrices of all compounds was performed. It is essential to understand the complete impurity profile of all reactants, solvents, and catalysts and to include them in the matrix. Finally, the API of this synthesis was checked if all impurities are identified by this tool. Of note, a shortcoming of such a targeted approach is that impurities can still occur, but they are not captured. This disadvantage can be partially compensated by non-targeted approaches if they are performed in parallel with the other studies that represent most of the impurities. However, this work also shows that even in a supposedly simple synthesis, potentially hundreds of by-products can be formed. For each of them, it must be decided individually whether their formation is probable or how their quantity can be minimized in order to obtain APIs, that are as pure as possible. In the dapsone project it was aimed to replace the existing old Ph. Eur. TLC method with a modern RP-HPLC method. This was successful and since Ph. Eur. 10.6, the method developed in this work, became a valid monograph. Within the revision process of the monograph, the individual limits for impurities were tightened. However, this new method needs HPLC instrumentation, suitable to perform gradients. As this is not always available in all control laboratories, we also developed an alternative, more simple method using two different isocratic runs for the impurity analysis. The obtained batch results of both, the new pharmacopoeial method and the more simple one, were in a comparable order of magnitude. Furthermore, within the method development stage of the Ph. Eur. method, we could identify one unknown impurity of the impurity reference by high-resolution MS/MS analysis. Also, in the baclofen project it was aimed to replace the existing Ph. Eur. method with the introduction of an additional impurity to be quantified. A corresponding method was developed and validated. However, due to the harmonization process of the pharmacopoeias, it is currently not used. In addition, we tried to find further, non- 116 SUMMARY chromophoric impurities by means of the CAD. However, except for one counterion of an impurity, no further impurities were found. Also, the aforementioned new impurity could not be detected above the reporting threshold in the batches analyzed. As the only individually specified impurity A is also present at a low level, it can be concluded that the examined batches of baclofen are very pure. The use of universal detectors, such as the CAD can be particularly interesting for compounds with no chromophore or those with only a weak chromophore. Therefore, we decided to take a closer look at the impurity profile of acarbose. Currently, acarbose and its impurities are being studied by low wavelength UV detection at 210 nm. Therefore, the question arose whether there are no other impurities in the API that do not show absorption at this wavelength. CAD, which offers consistent detection properties for all non-volatile compounds, is ideally suited for this purpose. However, it was not so easy to use the CAD together with the UV detector, for example, as a hyphenated detection technique, because the Ph. Eur. method uses phosphate buffers. However, this is non-volatile and therefore inappropriate for the CAD. Therefore, an attempt was made to replace the buffer with a volatile one. However, since this did not lead to satisfactory results and rather the self-degradation process of the stationary phase used could be observed by means of the CAD, it was decided to switch to alternative stationary phases. A column screening also revealed further difficulties with acarbose and its impurities: they show an epimerization reaction at the end of the sugar chain. However, since one wanted to have uniform peaks in the corresponding chromatograms, one had to accelerate this reaction significantly to obtain only one peak for each component. This was best achieved by using two stationary phases: PGC and Amide-HILIC. Impurity-profiling methods could be developed on each of the two phases. In addition, as expected, new impurities could be detected, albeit at a low level. Two of them could even be identified by spiking experiments as the sugar fragments maltose and maltotriose. Taken together, it can be concluded, that this work has contributed significantly to the improvement of the quality analysis of monographed drugs. In addition to the presented general tool for the identification of potential impurities, one of the methods developed, had already been implemented to the Ph. Eur. In an effort to improve the CAD's universal detection capabilities, additional methods have also been developed. Further, new improved methods for the impurity profiling are ready to use.}, subject = {Instrumentelle Analytik}, language = {en} }