TY - JOUR A1 - Kampf, Thomas A1 - Reiter, Theresa A1 - Bauer, Wolfgang Rudolf T1 - An analytical model which determines the apparent T1 for Modified Look-Locker Inversion Recovery – Analysis of the longitudinal relaxation under the influence of discontinuous balanced (classical MOLLI) and spoiled gradient echo readouts JF - Zeitschrift für Medizinische Physik N2 - Quantitative nuclear magnetic resonance imaging (MRI) shifts more and more into the focus of clinical research. Especially determination of relaxation times without/and with contrast agents becomes the foundation of tissue characterization, e.g. in cardiac MRI for myocardial fibrosis. Techniques which assess longitudinal relaxation times rely on repetitive application of readout modules, which are interrupted by free relaxation periods, e.g. the Modified Look-Locker Inversion Recovery = MOLLI sequence. These discontinuous sequences reveal an apparent relaxation time, and, by techniques extrapolated from continuous readout sequences, a putative real T1 is determined. What is missing is a rigorous analysis of the dependence of the apparent relaxation time on its real partner, readout sequence parameters and biological parameters as heart rate. This is provided in this paper for the discontinuous balanced steady state free precession (bSSFP) and spoiled gradient echo readouts. It turns out that the apparent longitudinal relaxation rate is the time average of the relaxation rates during the readout module, and free relaxation period. Knowing the heart rate our results vice versa allow to determine the real T1 from its measured apparent partner. T2 - Ein analytisches Modell, das die apparente T1 Zeit für Modfied Look-Locker Inversion Recovery bestimmt-Analyse der longitudinalen Relaxation unter dem Einfluss diskontinuierlicher balanced (klassische MOLLI) und spoiled gradient echo readouts KW - longitudinal relaxation KW - T1 KW - T2 KW - Lock Locker KW - MOLLI KW - balanced steady state free precession KW - spoiled gradient echo Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-325498 VL - 28 ER - TY - JOUR A1 - Kampf, Thomas A1 - Bauer, Wolfgang Rudolf A1 - Reiter, Theresa T1 - Improved post-processing strategy for MOLLI based tissue characterization allows application in patients with dyspnoe and impaired left ventricular function JF - Zeitschrift für Medizinische Physik N2 - Contrast and non-contrast MRI based characterization of myocardium by T1-mapping will be of paramount importance to obtain biomarkers, e.g. fibrosis, which determines the risk of heart failure patients. T1-mapping by the standard post-processing of the modified look-locker inversion recovery (MOLLI) lacks of accuracy when trying to reduce its duration, which on the other hand, is highly desirable in patients with heart failure. The recently suggested inversion group fitting (IGF) technique, which considers more parameters for fitting, has a superior accuracy for long T1 times despite a shorter duration. However, for short T1 values, the standard method has a superior precision. A conditional fitting routine is proposed which ideally takes advantage of both algorithms. Materials and methods All measurements were performed on a 1.5 T clinical scanner (ACHIEVA, Philips Healthcare, The Netherlands) using a MOLLI 5(n)3(n)3 prototype with n(heart beats) being a variable waiting time between inversion experiments. Phantom experiments covered a broad range of T1 times, waiting times and heart rates. A saturation recovery experiment served as a gold standard for T1 measurement. All data were analyzed with the standard MOLLI, the IGF fit and the conditional fitting routine and the obtained T1 values were compared with the gold standard. In vivo measurements were performed in a healthy volunteer and a total of 34 patients with normal findings, dilative cardiomyopathy and amyloidosis. Results Theoretical analysis and phantom experiments provided a threshold value for an apparent IGF determining processing with IGF post processing for values above, or switching to the standard technique for values below. This was validated in phantoms and patients measurements. A reduction of the waiting time to 1 instead of 3 heart beats between the inversion experiments showed reliable results. The acquisition time was reduced from 17 to 13 heart beats. The in vivo measurements showed ECV values between 25% (18–33%; SD 0.03) in the healthy, 30% (22–40%; SD 0.04) in patients with DCM and 45% (30–60%; SD 0.9) in patients with amyloidosis. Conclusion The adopted post-processing algorithm determines long T1 values with high accuracy and short T1 values while maintaining a high precision. Based on reduction of waiting time, and independence of heart rate, it shortens breath hold duration and allows fast T1-mapping, which is frequently a prerequisite in patients with cardiac diseases. KW - T1-mapping KW - ECV KW - MOLLI KW - post-processing Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-325481 VL - 28 ER - TY - JOUR A1 - Vogel, Patrick A1 - Rückert, Martin Andreas A1 - Friedrich, Bernhard A1 - Tietze, Rainer A1 - Lyer, Stefan A1 - Kampf, Thomas A1 - Hennig, Thomas A1 - Dölken, Lars A1 - Alexiou, Christoph A1 - Behr, Volker Christian T1 - Critical Offset Magnetic PArticle SpectroScopy for rapid and highly sensitive medical point-of-care diagnostics JF - Nature Communications N2 - Magnetic nanoparticles (MNPs) have been adapted for many applications, e.g., bioassays for the detection of biomarkers such as antibodies, by controlled engineering of specific surface properties. Specific measurement of such binding states is of high interest but currently limited to highly sensitive techniques such as ELISA or flow cytometry, which are relatively inflexible, difficult to handle, expensive and time-consuming. Here we report a method named COMPASS (Critical-Offset-Magnetic-Particle-SpectroScopy), which is based on a critical offset magnetic field, enabling sensitive detection to minimal changes in mobility of MNP ensembles, e.g., resulting from SARS-CoV-2 antibodies binding to the S antigen on the surface of functionalized MNPs. With a sensitivity of 0.33 fmole/50 µl (≙7 pM) for SARS-CoV-2-S1 antibodies, measured with a low-cost portable COMPASS device, the proposed technique is competitive with respect to sensitivity while providing flexibility, robustness, and a measurement time of seconds per sample. In addition, initial results with blood serum demonstrate high specificity. KW - biochemical assays KW - characterization and analytical techniques KW - magnetic properties and materials KW - nanoparticles Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-300893 VL - 13 ER - TY - JOUR A1 - Aster, Hans-Christoph A1 - Evdokimov, Dimitar A1 - Braun, Alexandra A1 - Üçeyler, Nurcan A1 - Kampf, Thomas A1 - Pham, Mirko A1 - Homola, György A. A1 - Sommer, Claudia T1 - CNS imaging characteristics in fibromyalgia patients with and without peripheral nerve involvement JF - Scientific Reports N2 - We tested the hypothesis that reduced skin innervation in fibromyalgia syndrome is associated with specific CNS changes. This prospective case–control study included 43 women diagnosed with fibromyalgia syndrome and 40 healthy controls. We further compared the fibromyalgia subgroups with reduced (n = 21) and normal (n = 22) skin innervation. Brains were analysed for cortical volume, for white matter integrity, and for functional connectivity. Compared to controls, cortical thickness was decreased in regions of the frontal, temporal and parietal cortex in the fibromyalgia group as a whole, and decreased in the bilateral pericalcarine cortices in the fibromyalgia subgroup with reduced skin innervation. Diffusion tensor imaging revealed a significant increase in fractional anisotropy in the corona radiata, the corpus callosum, cingulum and fornix in patients with fibromyalgia compared to healthy controls and decreased FA in parts of the internal capsule and thalamic radiation in the subgroup with reduced skin innervation. Using resting-state fMRI, the fibromyalgia group as a whole showed functional hypoconnectivity between the right midfrontal gyrus and the posterior cerebellum and the right crus cerebellum, respectively. The subgroup with reduced skin innervation showed hyperconnectivity between the inferior frontal gyrus, the angular gyrus and the posterior parietal gyrus. Our results suggest that the subgroup of fibromyalgia patients with pronounced pathology in the peripheral nervous system shows alterations in morphology, structural and functional connectivity also at the level of the encephalon. We propose considering these subgroups when conducting clinical trials. KW - fibromyalgia syndrome KW - CNS imaging KW - peripheral nerve involvement Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-300562 VL - 12 IS - 1 ER - TY - JOUR A1 - Winter, Patrick M. A1 - Andelovic, Kristina A1 - Kampf, Thomas A1 - Hansmann, Jan A1 - Jakob, Peter Michael A1 - Bauer, Wolfgang Rudolf A1 - Zernecke, Alma A1 - Herold, Volker T1 - Simultaneous measurements of 3D wall shear stress and pulse wave velocity in the murine aortic arch JF - Journal of Cardiovascular Magnetic Resonance N2 - Purpose Wall shear stress (WSS) and pulse wave velocity (PWV) are important parameters to characterize blood flow in the vessel wall. Their quantification with flow-sensitive phase-contrast (PC) cardiovascular magnetic resonance (CMR), however, is time-consuming. Furthermore, the measurement of WSS requires high spatial resolution, whereas high temporal resolution is necessary for PWV measurements. For these reasons, PWV and WSS are challenging to measure in one CMR session, making it difficult to directly compare these parameters. By using a retrospective approach with a flexible reconstruction framework, we here aimed to simultaneously assess both PWV and WSS in the murine aortic arch from the same 4D flow measurement. Methods Flow was measured in the aortic arch of 18-week-old wildtype (n = 5) and ApoE\(^{−/−}\) mice (n = 5) with a self-navigated radial 4D-PC-CMR sequence. Retrospective data analysis was used to reconstruct the same dataset either at low spatial and high temporal resolution (PWV analysis) or high spatial and low temporal resolution (WSS analysis). To assess WSS, the aortic lumen was labeled by semi-automatically segmenting the reconstruction with high spatial resolution. WSS was determined from the spatial velocity gradients at the lumen surface. For calculation of the PWV, segmentation data was interpolated along the temporal dimension. Subsequently, PWV was quantified from the through-plane flow data using the multiple-points transit-time method. Reconstructions with varying frame rates and spatial resolutions were performed to investigate the influence of spatiotemporal resolution on the PWV and WSS quantification. Results 4D flow measurements were conducted in an acquisition time of only 35 min. Increased peak flow and peak WSS values and lower errors in PWV estimation were observed in the reconstructions with high temporal resolution. Aortic PWV was significantly increased in ApoE\(^{−/−}\) mice compared to the control group (1.7 ± 0.2 versus 2.6 ± 0.2 m/s, p < 0.001). Mean WSS magnitude values averaged over the aortic arch were (1.17 ± 0.07) N/m\(^2\) in wildtype mice and (1.27 ± 0.10) N/m\(^2\) in ApoE\(^{−/−}\) mice. Conclusion The post processing algorithm using the flexible reconstruction framework developed in this study permitted quantification of global PWV and 3D-WSS in a single acquisition. The possibility to assess both parameters in only 35 min will markedly improve the analyses and information content of in vivo measurements. KW - 4D flow KW - pulse wave velocity KW - wall shear stress KW - radial KW - self-navigation KW - mouse KW - aortic arch KW - atherosclerosis KW - mice KW - flow KW - plaque KW - CMR KW - quantification KW - microscopy Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-259152 VL - 23 IS - 1 ER - TY - JOUR A1 - Andelovic, Kristina A1 - Winter, Patrick A1 - Kampf, Thomas A1 - Xu, Anton A1 - Jakob, Peter Michael A1 - Herold, Volker A1 - Bauer, Wolfgang Rudolf A1 - Zernecke, Alma T1 - 2D Projection Maps of WSS and OSI Reveal Distinct Spatiotemporal Changes in Hemodynamics in the Murine Aorta during Ageing and Atherosclerosis JF - Biomedicines N2 - Growth, ageing and atherosclerotic plaque development alter the biomechanical forces acting on the vessel wall. However, monitoring the detailed local changes in wall shear stress (WSS) at distinct sites of the murine aortic arch over time has been challenging. Here, we studied the temporal and spatial changes in flow, WSS, oscillatory shear index (OSI) and elastic properties of healthy wildtype (WT, n = 5) and atherosclerotic apolipoprotein E-deficient (Apoe\(^{−/−}\), n = 6) mice during ageing and atherosclerosis using high-resolution 4D flow magnetic resonance imaging (MRI). Spatially resolved 2D projection maps of WSS and OSI of the complete aortic arch were generated, allowing the pixel-wise statistical analysis of inter- and intragroup hemodynamic changes over time and local correlations between WSS, pulse wave velocity (PWV), plaque and vessel wall characteristics. The study revealed converse differences of local hemodynamic profiles in healthy WT and atherosclerotic Apoe\(^{−/−}\) mice, and we identified the circumferential WSS as potential marker of plaque size and composition in advanced atherosclerosis and the radial strain as a potential marker for vascular elasticity. Two-dimensional (2D) projection maps of WSS and OSI, including statistical analysis provide a powerful tool to monitor local aortic hemodynamics during ageing and atherosclerosis. The correlation of spatially resolved hemodynamics and plaque characteristics could significantly improve our understanding of the impact of hemodynamics on atherosclerosis, which may be key to understand plaque progression towards vulnerability. KW - atherosclerosis KW - mouse KW - 4D flow MRI KW - aortic arch KW - flow dynamics KW - WSS KW - mapping KW - PWV KW - plaque characteristics Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-252164 SN - 2227-9059 VL - 9 IS - 12 ER - TY - JOUR A1 - Horvat, Sonja A1 - Vogel, Patrick A1 - Kampf, Thomas A1 - Brandl, Andreas A1 - Alshamsan, Aws A1 - Alhadlaq, Hisham A. A1 - Ahamed, Maqusood A1 - Albrecht, Krystyna A1 - Behr, Volker C. A1 - Beilhack, Andreas A1 - Groll, Jürgen T1 - Crosslinked Coating Improves the Signal‐to‐Noise Ratio of Iron Oxide Nanoparticles in Magnetic Particle Imaging (MPI) JF - ChemNanoMat N2 - Magnetic particle imaging is an emerging tomographic method used for evaluation of the spatial distribution of iron‐oxide nanoparticles. In this work, the effect of the polymer coating on the response of particles was studied. Particles with covalently crosslinked coating showed improved signal and image resolution. KW - crosslinked coating KW - imaging agents KW - magnetic properties KW - MPI KW - MPS Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-214718 VL - 6 IS - 5 SP - 755 EP - 758 ER - TY - JOUR A1 - Riedl, Katharina A. A1 - Kampf, Thomas A1 - Herold, Volker A1 - Behr, Volker C. A1 - Bauer, Wolfgang R. T1 - Wall shear stress analysis using 17.6 Tesla MRI: A longitudinal study in ApoE\(^{-/-}\)mice with histological analysis JF - PLoS One N2 - This longitudinal study was performed to evaluate the feasibility of detecting the interaction between wall shear stress (WSS) and plaque development. 20 ApoE\(^{-/-}\)mice were separated in 12 mice with Western Diet and 8 mice with Chow Diet. Magnetic resonance (MR) scans at 17.6 Tesla and histological analysis were performed after one week, eight and twelve weeks. Allin vivoMR measurements were acquired using a flow sensitive phase contrast method for determining vectorial flow. Histological sections were stained with Hematoxylin and Eosin, Elastica van Gieson and CD68 staining. Data analysis was performed using Ensight and a Matlab-based "Flow Tool". The body weight of ApoE\(^{-/-}\)mice increased significantly over 12 weeks. WSS values increased in the Western Diet group over the time period; in contrast, in the Chow Diet group the values decreased from the first to the second measurement point. Western Diet mice showed small plaque formations with elastin fragmentations after 8 weeks and big plaque formations after 12 weeks; Chow Diet mice showed a few elastin fragmentations after 8 weeks and small plaque formations after 12 weeks. Favored by high-fat diet, plaque formation results in higher values of WSS. With wall shear stress being a known predictor for atherosclerotic plaque development, ultra highfield MRI can serve as a tool for studying the causes and beginnings of atherosclerosis. KW - phase-contrast MRI KW - flow patterns KW - blood flow KW - apolipoprotein-E KW - atheriosclerosis KW - mouse KW - mice KW - quantification KW - association KW - lesions Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-229318 VL - 15 IS - 8 ER - TY - JOUR A1 - Kurz, Felix T. A1 - Kampf, Thomas A1 - Buschle, Lukas R. A1 - Schlemmer, Heinz-Peter A1 - Bendszus, Martin A1 - Heiland, Sabine A1 - Ziener, Christian H. T1 - Generalized moment analysis of magnetic field correlations for accumulations of spherical and cylindrical magnetic perturbers JF - Frontiers in Physics N2 - In biological tissue, an accumulation of similarly shaped objects with a susceptibility difference to the surrounding tissue generates a local distortion of the external magnetic field in magnetic resonance imaging. It induces stochastic field fluctuations that characteristically influence proton spin dephasing in the vicinity of these magnetic perturbers. The magnetic field correlation that is associated with such local magnetic field inhomogeneities can be expressed in the form of a dynamic frequency autocorrelation function that is related to the time evolution of the measured magnetization. Here, an eigenfunction expansion for two simple magnetic perturber shapes, that of spheres and cylinders, is considered for restricted spin diffusion in a simple model geometry. Then, the concept of generalized moment analysis, an approximation technique that is applied in the study of (non-)reactive processes that involve Brownian motion, allows deriving analytical expressions of the correlation function for different exponential decay forms. Results for the biexponential decay for both spherical and cylindrical magnetized objects are derived and compared with the frequently used (less accurate) monoexponential decay forms. They are in asymptotic agreement with the numerically exact value of the correlation function for long and short times. KW - magnetized sphere/cylinder KW - magnetic susceptibility KW - correlation function KW - diffusion KW - magnetic resonance imaging Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-190604 SN - 2296-424X VL - 4 ER - TY - JOUR A1 - Herold, Volker A1 - Kampf, Thomas A1 - Jakob, Peter Michael T1 - Dynamic magnetic resonance scattering JF - Communications Physics N2 - Dynamic light scattering is a popular technique to determine the size distribution of small particles in the sub micrometer region. It operates in reciprocal space, by analyzing the signal fluctuations with the photon auto correlation function. Equally, pulsed field gradient magnetic resonance is a technique generating data in the reciprocal space of the density distribution of an object. Here we show the feasibility of employing a magnetic resonance imaging system as a dynamic scattering device similar to dynamic light scattering appliances. By acquiring a time series of single data points from reciprocal space, analogue to dynamic light scattering, we demonstrate the examination of motion patterns of microscopic particles. This method allows the examination of particle dynamics significantly below the spatial resolution of magnetic resonance imaging. It is not limited by relaxation times and covers a wide field of applications for particle or cell motion in opaque media. KW - Characterization and analytical techniques KW - Imaging techniques Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-201091 VL - 2 ER -