TY - JOUR A1 - Lohr, David A1 - Terekhov, Maxim A1 - Weng, Andreas Max A1 - Schroeder, Anja A1 - Walles, Heike A1 - Schreiber, Laura Maria T1 - Spin echo based cardiac diffusion imaging at 7T: An ex vivo study of the porcine heart at 7T and 3T JF - PLoS ONE N2 - Purpose of this work was to assess feasibility of cardiac diffusion tensor imaging (cDTI) at 7 T in a set of healthy, unfixed, porcine hearts using various parallel imaging acceleration factors and to compare SNR and derived cDTI metrics to a reference measured at 3 T. Magnetic resonance imaging was performed on 7T and 3T whole body systems using a spin echo diffusion encoding sequence with echo planar imaging readout. Five reference (b = 0 s/mm\(^2\)) images and 30 diffusion directions (b = 700 s/mm\(^2\)) were acquired at both 7 T and 3 T using a GRAPPA acceleration factor R = 1. Scans at 7 T were repeated using R = 2, R = 3, and R = 4. SNR evaluation was based on 30 reference (b = 0 s/mm\(^2\)) images of 30 slices of the left ventricle and cardiac DTI metrics were compared within AHA segmentation. The number of hearts scanned at 7 T and 3 T was n = 11. No statistically significant differences were found for evaluated helix angle, secondary eigenvector angle, fractional anisotropy and apparent diffusion coefficient at the different field strengths, given sufficiently high SNR and geometrically undistorted images. R≥3 was needed to reduce susceptibility induced geometric distortions to an acceptable amount. On average SNR in myocardium of the left ventricle was increased from 29±3 to 44±6 in the reference image (b = 0 s/mm\(^2\)) when switching from 3 T to 7 T. Our study demonstrates that high resolution, ex vivo cDTI is feasible at 7 T using commercial hardware. KW - Heart KW - Diffusion tensor imaging KW - Eigenvectors KW - Cardiac ventricles KW - Tractography KW - Magnetic resonance imaging KW - Data acquisition KW - Swine Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-201376 VL - 14 IS - 3 ER -