TY - JOUR A1 - Herz, Stefan A1 - Stefanescu, Maria R. A1 - Lohr, David A1 - Vogel, Patrick A1 - Kosmala, Aleksander A1 - Terekhov, Maxim A1 - Weng, Andreas M. A1 - Grunz, Jan-Peter A1 - Bley, Thorsten A. A1 - Schreiber, Laura M. T1 - Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission JF - PloS One N2 - Background To investigate the effects of B\(_1\)-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI). Methods Stenosis phantoms with different grades of stenosis (0%, 20%, 40%, 60%, 80%, and 100%; 5 mm inner vessel diameter) were produced using 3D printing (clear resin). Phantoms were imaged with four different concentrations of diluted Gd-DOTA representing established arterial concentrations after intravenous injection in humans. Samples were centrally positioned in a thorax phantom of 30 cm diameter filled with a custom-made liquid featuring dielectric properties of muscle tissue. MRI was performed on a 7 T whole-body system. 2D-gradient-echo sequences were acquired with an 8-channel transmit 16-channel receive (8 Tx / 16 Rx) cardiac array prototype coil with and without pTX mode. Measurements were compared to those obtained with identical scan parameters using a commercially available 1 Tx / 16 Rx single transmit coil (sTX). To assess reproducibility, measurements (n = 15) were repeated at different horizontal angles with respect to the B0-field. Results B\(_1\)-shimming and pTX markedly improved flip angle homogeneity across the thorax phantom yielding a distinctly increased signal-to-noise ratio (SNR) averaged over a whole slice relative to non-manipulated RF fields. Images without B\(_1\)-shimming showed shading artifacts due to local B\(_1\)\(^+\)-field inhomogeneities, which hampered stenosis quantification in severe cases. In contrast, B\(_1\)-shimming and pTX provided superior image homogeneity. Compared with a conventional sTX coil higher grade stenoses (60% and 80%) were graded significantly (p<0.01) more precise. Mild to moderate grade stenoses did not show significant differences. Overall, SNR was distinctly higher with B\(_1\)-shimming and pTX than with the conventional sTX coil (inside the stenosis phantoms 14%, outside the phantoms 32%). Both full and half concentration (10.2 mM and 5.1 mM) of a conventional Gd-DOTA dose for humans were equally suitable for stenosis evaluation in this phantom study. Conclusions B\(_1\)-shimming and pTX at 7 T can distinctly improve image homogeneity and therefore provide considerably more accurate MR image analysis, which is beneficial for imaging of small vessel structures. KW - stenosis KW - magnetic resonance imaging KW - thorax KW - in vivo imaging KW - coronary arteries KW - image processing KW - 3D printing KW - signal to noise ratio Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-300129 VL - 17 IS - 6 ER - TY - JOUR A1 - Schneider, Sonja Jasmin Maria A1 - Höhne, Christian A1 - Schneider, Martin A1 - Schmitter, Marc T1 - Photoacoustic tomography versus cone-beam computed tomography versus micro-computed tomography: Accuracy of 3D reconstructions of human teeth JF - PloS One N2 - Objectives In this in-vitro study, teeth were imaged using photoacoustic tomography (PAT), cone-beam computed tomography (CBCT), and micro-computed tomography (μ-CT). The study had aim: to identify the best wavelength for PAT images to determine the accuracy of the three imaging methods, and to determine whether PAT images of teeth can achieve acceptable reconstruction quality. Methods Nineteen human mandibular single-rooted incisors were extracted from patients with trauma or periodontitis. To determine the best wavelength for acquiring photoacoustic images, all 19 teeth were scanned in vitro with PAT, using different laser wavelengths between 680 and 960 nm. The images were analyzed using image analysis software. To assess the accuracy of PAT and compare it with the accuracy of CBCT, each tooth was also scanned in vitro using CBCT and the reference standard technique of μ-CT. Subsequently, three different three-dimensional models, one for each imaging technique, were created for each tooth. Finally, the three different three-dimensional models acquired for the same tooth were matched and analyzed regarding volume and surface. Results The highest quality tooth images were achieved using the 680 nm wavelength, which showed the best contrast ratio. The full geometry of the dental root (μ-CT compared with PAT) could be visualized with relative standard deviations of 0.12 mm for the surface and −7.33 mm3 for the volume (n = 19). The full geometry of the dental root (μ-CT compared with CBCT) could be visualized with relative standard deviations of 0.06 mm for the surface and −14.56 mm3 for the volume (n = 19). The difference between the PAT–μ-CT group and CBCT–μ-CT group regarding the total average of the root surface area was not significant (p>0.06). Conclusion Images, which were acquired using PAT at 680nm showed the best contrast ration, enabling the identification of dentin, cementum and the dental pulp. No significant differences were found between the PAT–μ-CT group and CBCT–μ-CT group regarding the total average of the RSA and the total volume. Thus, three-dimensional reconstructions based on in-vitro PAT are already of acceptable reconstruction quality. KW - teeth KW - imaging techniques KW - computer software KW - incisors KW - in vivo imaging KW - caries KW - magnetic resonance imaging KW - tomography Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301437 VL - 17 IS - 12 ER -