TY - JOUR A1 - Wegener, Sonja A1 - Schindhelm, Robert A1 - Sauer, Otto A. T1 - Implementing corrections of isocentric shifts for the stereotactic irradiation of cerebral targets: Clinical validation JF - Journal of Applied Clinical Medical Physics N2 - Purpose: Any Linac will show geometric imprecisions, including non-ideal alignment of the gantry, collimator and couch axes, and gantry sag or wobble. Their angular dependence can be quantified and resulting changes of the dose distribution predicted (Wack, JACMP 20(5), 2020). We analyzed whether it is feasible to correct geometric shifts during treatment planning. The successful implementation of such a correction procedure was verified by measurements of different stereotactic treatment plans. Methods: Isocentric shifts were quantified for two Elekta Synergy Agility Linacs using the QualiForMed ISO-CBCT+ module, yielding the shift between kV and MV isocenters, the gantry flex and wobble as well as the positions of couch and collimator rotation axes. Next, the position of each field's isocenter in the Pinnacle treatment planning system was adjusted accordingly using a script. Fifteen stereotactic treatment plans of cerebral metastases (0.34 to 26.53 cm3) comprising 9–11 beams were investigated; 54 gantry and couch combinations in total. Unmodified plans and corrected plans were measured using the Sun Nuclear SRS-MapCHECK with the Stereophan phantom and evaluated using gamma analysis. Results: Geometric imprecisions, such as shifts of up to 0.8 mm between kV and MV isocenter, a couch rotation axis 0.9 mm off the kV isocente,r and gantry flex with an amplitude of 1.1 mm, were found. For eight, mostly small PTVs D98 values declined more than 5% by simulating these shifts. The average gamma (2%/2 mm, absolute, global, 20% threshold) was reduced from 0.53 to 0.31 (0.32 to 0.30) for Linac 1 (Linac 2) when including the isocentric corrections. Thus, Linac 1 reached the accuracy level of Linac 2 after correction. Conclusion: Correcting for Linac geometric deviations during the planning process is feasible and was dosimetrically validated. The dosimetric impact of the geometric imperfections can vary between Linacs and should be assessed and corrected where necessary. KW - isocenter KW - quality assurance KW - stereotactic irradiation Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-312906 VL - 23 IS - 5 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Zehner, Leonie A1 - Hartrampf, Philipp E. A1 - Cirsi, Sinan A1 - Wegener, Sonja A1 - Buck, Andreas K. A1 - Flentje, Michael A1 - Polat, Bülent T1 - Dose-escalated salvage radiotherapy for macroscopic local recurrence of prostate cancer in the prostate-specific membrane antigen positron emission tomography era JF - Cancers N2 - Simple Summary Prostate cancer often relapses after initial radical prostatectomy, and salvage radiotherapy offers a second chance of cure for relapsed patients. Modern imaging techniques, especially prostate-specific membrane antigen positron emission tomography/computed tomography (PSMA PET/CT), enable radiation oncologists to target radiotherapy at the involved sites of disease. In a group of patients, PSMA PET/CT imaging can detect a macroscopic local recurrence with or without locoregional lymph node metastasis. In these cases, an escalation of the radiotherapy dose is often considered for controlling the visible tumor mass. As the evidence for dose-escalated salvage radiotherapy for macroscopic recurrent prostate cancer after PSMA PET/CT imaging is still limited, we address this topic in the current analysis. We found that the outcome of patients with dose-escalated salvage radiotherapy for macroscopic prostate cancer recurrence is encouragingly favorable, while the toxicity is very limited. Abstract Background: The purpose of this study was to access the oncological outcome of prostate-specific membrane antigen positron emission tomography (PSMA PET/CT)-guided salvage radiotherapy (SRT) for localized macroscopic prostate cancer recurrence. Methods: Between February 2010 and June 2021, 367 patients received SRT after radical prostatectomy. Out of the 367 screened patients, 111 patients were staged by PSMA PET/CT before SRT. A total of 59 out of these 111 (53.2%) patients were treated for PSMA PET-positive macroscopic prostatic fossa recurrence. Dose-escalated SRT was applied with a simultaneous integrated boost at a median prescribed dose of 69.3 Gy (IQR 69.3–72.6 Gy). The oncological outcome was investigated using Kaplan-Meier and Cox regression analyses. The genitourinary (GU)/gastrointestinal (GI) toxicity evaluation utilized Common Toxicity Criteria for Adverse Events (version 5.0). Results: The median follow-up was 38.2 months. The three-year biochemical progression-free survival rate was 89.1% (95% CI: 81.1–97.8%) and the three-year metastasis-free survival rate reached 96.2% (95% CI: 91.2–100.0%). The cumulative three-year late grade 3 GU toxicity rate was 3.4%. No late grade 3 GI toxicity occurred. Conclusions: Dose-escalated PSMA PET/CT-guided salvage radiotherapy for macroscopic prostatic fossa recurrence resulted in favorable survival and toxicity rates. KW - prostate cancer KW - salvage radiotherapy KW - macroscopic recurrence KW - PSMA PET/CT KW - simultaneous integrated boost Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-290302 SN - 2072-6694 VL - 14 IS - 19 ER - TY - JOUR A1 - Pollmann, Stephan A1 - Toussaint, André A1 - Flentje, Michael A1 - Wegener, Sonja A1 - Lewitzki, Victor T1 - Dosimetric evaluation of commercially available flat vs. self-produced 3D-conformal silicone boluses for the head and neck region JF - Frontiers in Oncology N2 - Background Boluses are routinely used in radiotherapy to modify surface doses. Nevertheless, considerable dose discrepancies may occur in some cases due to fit inaccuracy of commercially available standard flat boluses. Moreover, due to the simple geometric design of conventional boluses, also surrounding healthy skin areas may be unintentionally covered, resulting in the unwanted dose buildup. With the fused deposition modeling (FDM) technique, there is a simple and possibly cost-effective way to solve these problems in routine clinical practice. This paper presents a procedure of self-manufacturing bespoke patient-specific silicone boluses and the evaluation of buildup and fit accuracy in comparison to standard rectangular commercially available silicone boluses. Methods 3D-conformal silicone boluses were custom-built to cover the surgical scar region of 25 patients who received adjuvant radiotherapy of head and neck cancer at the University Hospital Würzburg. During a standard CT-based planning procedure, a 5-mm-thick 3D bolus contour was generated to cover the radiopaque marked surgical scar with an additional safety margin. From these digital contours, molds were 3D printed and poured with silicone. Dose measurements for both types of boluses were performed with radiochromic films (EBT3) at three points per patient—at least one aimed to be in the high-dose area (scar) and one in the lower-dose area (spared healthy skin). Surface–bolus distance, which ideally should not be present, was determined from cone-beam CT performed for positioning control. The dosimetric influence of surface–bolus distance was also determined on slab phantom for different field sizes. The trial was performed with hardware that may be routinely available in every radiotherapy department, with the exception of the 3D printer. The required number of patients was determined based on the results of preparatory measurements with the help of the statistical consultancy of the University of Würzburg. The number of measuring points represents the total number of patients. Results In the high-dose area of the scar, there was a significantly better intended dose buildup of 2.45% (95%CI 0.0014–0.0477, p = 0.038, N = 30) in favor of a 3D-conformal bolus. Median distances between the body surface and bolus differed significantly between 3D-conformal and commercially available boluses (3.5 vs. 7.9 mm, p = 0.001). The surface dose at the slab phantom did not differ between commercially available and 3D-conformal boluses. Increasing the surface–bolus distance from 5 to 10 mm decreased the surface dose by approximately 2% and 11% in the 6 × 6- and 3 × 3-cm2 fields, respectively. In comparison to the commercially available bolus, an unintended dose buildup in the healthy skin areas was reduced by 25.9% (95%CI 19.5–32.3, p < 0.01, N = 37) using the 3D-conformal bolus limited to the region surrounding the surgical scar. Conclusions Using 3D-conformal boluses allows a comparison to the commercially available boluses’ dose buildup in the covered areas. Smaller field size is prone to a larger surface–bolus distance effect. Higher conformity of 3D-conformal boluses reduces this effect. This may be especially relevant for volumetric modulated arc therapy (VMAT) and intensity-modulated radiotherapy (IMRT) techniques with a huge number of smaller fields. High conformity of 3D-conformal boluses reduces an unintended dose buildup in healthy skin. The limiting factor in the conformity of 3D-conformal boluses in our setting was the immobilization mask, which was produced primarily for the 3D boluses. The mask itself limited tight contact of subsequently produced 3D-conformal boluses to the mask-covered body areas. In this respect, bolus adjustment before mask fabrication will be done in the future setting. KW - flat silicone bolus KW - individual silicone bolus KW - 3D conformal silicone bolus KW - 3D printer KW - head and neck cancer KW - fused deposition modeling (FDM) KW - surface dose measurement KW - volumetric modulated arc therapy (VMAT) Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-283156 SN - 2234-943X VL - 12 ER - TY - JOUR A1 - Richter, Anne A1 - Wegener, Sonja A1 - Breuer, Kathrin A1 - Razinskas, Gary A1 - Weick, Stefan A1 - Exner, Florian A1 - Bratengeier, Klaus A1 - Flentje, Michael A1 - Sauer, Otto A1 - Polat, Bülent T1 - Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems JF - Radiation Oncology N2 - Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. Results The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. Conclusions For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system. KW - whole breast irradiation KW - Halcyon KW - IGRT KW - dose to OARs Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-265704 VL - 16 ER - TY - JOUR A1 - Jäger, Andreas A1 - Wegener, Sonja A1 - Sauer, Otto A. T1 - Dose rate correction for a silicon diode detector array JF - Journal of Applied Clinical Medical Physics N2 - Purpose A signal dependence on dose rate was reported for the ArcCHECK array due to recombination processes within the diodes. The purpose of our work was to quantify the necessary correction and apply them to quality assurance measurements. Methods Static 10 × 10 cm\(^2\) 6-MV fields delivered by a linear accelerator were applied to the detector array while decreasing the average dose rate, that is, the pulse frequency, from 500 to 30 MU/min. An ion chamber was placed inside the ArcCHECK cavity as a reference. Furthermore, the instantaneous dose rate dependence (DRD) was studied. The position of the detector was adjusted to change the dose-per-pulse, varying the distance between the focus and the diode closest to the focus between 69.6 and 359.6 cm. Reference measurements were performed with an ion chamber placed inside a PMMA slab phantom at the same source-to-detector distances (SDDs). Exponential saturation functions were fitted to the data, with different parameters to account for two generations of ArcCHECK detectors (types 2 and 3) and both DRDs. Corrections were applied to 12 volumetric modulated arc therapy plans. Results The sensitivity decreased by up to 2.8% with a decrease in average dose rate and by 9% with a decrease in instantaneous dose rate. Correcting the average DRD, the mean gamma pass rates (2%/2-mm criterion) of the treatment plans were improved by 5 percentage points (PP) for diode type 3 and 0.4 PP for type 2. Correcting the instantaneous DRD, the improvement was 8.4 PP for type 3 and 0.9 PP for type 2. Conclusions The instantaneous DRD was identified as the prevailing effect on the diode sensitivity. We developed and validated a method to correct this behavior. The number of falsely not passed treatment plans could be considerably reduced. KW - ArcCHECK KW - correction KW - diode KW - dose rate KW - dosimetry, QA Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260446 VL - 22 IS - 10 ER - TY - JOUR A1 - Wack, Linda J. A1 - Exner, Florian A1 - Wegener, Sonja A1 - Sauer, Otto A. T1 - The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets — Quantification and possible corrections JF - Journal of Applied Clinical Medical Physics N2 - Purpose To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D\(_{98}\), target volume coverage (TVC), conformity (CI), and gradient index (GI). Methods Winston‐Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assurance (QA) plan was applied to the ArcCHECK phantom to assess the impact of isocenter shift corrections on Gamma pass rates. These corrections included gantry sag, distance of collimator and couch axes to the gantry axis, and distance between cone‐beam computed tomography (CBCT) isocenter and treatment beam (MV) isocenter. We applied the shifts via script to the treatment plan in Pinnacle 16.2. In a planning study, isocenter and mechanical rotation axis shifts of 0.25 to 2 mm were applied to stereotactic plans of spherical planning target volumes (PTVs) of various volumes. The shifts determined via WL measurements were applied to 16 patient plans with PTV sizes between 0.22 and 10.4 cm3. Results ArcCHECK measurements of a stereotactic treatment showed significant increases in Gamma pass rate for all three measurements (up to 3.8 percentage points) after correction of measured isocenter deviations. For spherical targets of 1 cm3, CI was most severely affected by increasing the distance of the CBCT isocenter (1.22 to 1.62). Gradient index increased with an isocenter‐collimator axis distance of 1.5 mm (3.84 vs 4.62). D98 (normalized to reference) dropped to 0.85 (CBCT), 0.92 (table axis), 0.95 (collimator axis), and 0.98 (gantry sag), with similar but smaller changes for larger targets. Applying measured shifts to patient plans lead to relevant drops in D\(_{98}\) and TVC (7%) for targets below 2 cm\(^3\) treated on linac 1. Conclusion Mechanical deviations during gantry, collimator, and table rotation may adversely affect the treatment of small stereotactic lesions. Adjustments of beam isocenters in the treatment planning system (TPS) can be used to both quantify their impact and for prospective correction of treatment plans. KW - isocenter KW - quality assurance KW - stereotactic radiotherapy KW - Winston‐Lutz test Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-218146 VL - 21 IS - 5 ER - TY - JOUR A1 - Wegener, Sonja A1 - Herzog, Barbara A1 - Sauer, Otto A. T1 - Detector response in the buildup region of small MV fields JF - Medical Physics N2 - Purpose: The model used to calculate dose distributions in a radiotherapy treatment plan relies on the data entered during beam commissioning. The quality of these data heavily depends on the detector choice made, especially in small fields and in the buildup region. Therefore, it is necessary to identify suitable detectors for measurements in the buildup region of small fields. To aid the understanding of a detector's limitations, several factors that influence the detector signal are to be analyzed, for example, the volume effect due to the detector size, the response to electron contamination, the signal dependence on the polarity used, and the effective point of measurement chosen. Methods: We tested the suitability of different small field detectors for measurements of depth dose curves with a special focus on the surface‐near area of dose buildup for fields sized between 10 × 10 and 0.6 × 0.6 cm\(^{2}\). Depth dose curves were measured with 14 different detectors including plane‐parallel chambers, thimble chambers of different types and sizes, shielded and unshielded diodes as well as a diamond detector. Those curves were compared with depth dose curves acquired on Gafchromic film. Additionally, the magnitude of geometric volume corrections was estimated from film profiles in different depths. Furthermore, a lead foil was inserted into the beam to reduce contaminating electrons and to study the resulting changes of the detector response. The role of the effective point of measurement was investigated by quantifying the changes occurring when shifting depth dose curves. Last, measurements for the small ionization chambers taken at opposing biasing voltages were compared to study polarity effects. Results: Depth‐dependent correction factors for relative depth dose curves with different detectors were derived. Film, the Farmer chamber FC23, a 0.13 cm\(^{3}\) scanning chamber CC13 and a plane‐parallel chamber PPC05 agree very well in fields sized 4 × 4 and 10 × 10 cm\(^{2}\). For most detectors and in smaller fields, depth dose curves differ from the film. In general, shielded diodes require larger corrections than unshielded diodes. Neither the geometric volume effect nor the electron contamination can account for the detector differences. The biggest uncertainty arises from the positioning of a detector with respect to the water surface and from the choice of the detector's effective point of measurement. Depth dose curves acquired with small ionization chambers differ by over 15% in the buildup region depending on sign of the biasing voltage used. Conclusions: A scanning chamber or a PPC40 chamber is suitable for fields larger than 4 × 4 cm\(^{2}\). Below that field size, the microDiamond or small ionization chambers perform best requiring the smallest corrections at depth as well as in the buildup region. Diode response changes considerably between the different types of detectors. The position of the effective point of measurement has a huge effect on the resulting curves, therefore detector specific rather than general shifts of half the inner radius of cylindrical ionization chambers for the effective point of measurement should be used. For small ionization chambers, averaging between both polarities is necessary for data obtained near the surface. KW - buildup region KW - diode KW - dosimetry KW - microionization chambers KW - percent depth dose curves Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-214228 VL - 47 IS - 3 ER - TY - JOUR A1 - Wegener, Sonja A1 - Sauer, Otto A. T1 - The effective point of measurement for depth-dose measurements in small MV photon beams with different detectors JF - Medical Physics N2 - Purpose: The effective point of measurement (EPOM) of cylindrical ionization chambers differs from their geometric center. The exact shift depends on chamber construction details, above all the chamber size, and to some degree on the field-size and beam quality. It generally decreases as the chamber dimensions get smaller. In this work, effective points of measurement in small photon fields of a range of cylindrical chambers of different sizes are investigated, including small chambers that have not been studied previously. Methods: In this investigation, effective points of measurement for different ionization chambers (Farmer type, scanning chambers, micro-ionization chambers) and solid state detectors were determined by measuring depth-ionization curves in a 6 MV beam in field sizes between 2 9 2 cm2 and 10 9 10 cm2 and comparing those curves with curves measured with plane-parallel chambers. Results: It was possible to average the results to one shift per detector, as the results were sufficiently independent of the studied field sizes. For cylindrical ion chambers, shifts of the EPOM were determined to be between 0.49 and 0.30 times the inner chamber radius from the reference point. Conclusions: We experimentally confirmed the previously reported decrease of the EPOM shift with decreasing detector size. Highly accurate data for a large range of detectors, including new very small ones, were determined. Thus, small chambers noticeably differ from the 0.5-times to 0.6-times the inner chamber radius recommendations in current dosimetry protocols. The detector-individual EPOMs need to be considered for measurements of depth-dose curves. KW - depth dose curves KW - effective point of measurement KW - ionization chambers KW - micro-chambers Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-206148 VL - 46 IS - 11 ER - TY - THES A1 - Wegener, Sonja T1 - Dosimetrie unter Nicht-Gleichgewichtsbedingungen T1 - Dosimetry in non-equilibrium situations N2 - Für die Dosimetrie in der Strahlentherapie sind eine Reihe von Detektoren unterschiedlicher Bauform und Funktionsweise erhältlich. Detektoreigenschaften wie die Größe des aktiven Volumens, energieabhängiges Ansprechen und Feldstörungen durch Bauteile beeinflussen ihr Signal, so dass kein idealer, universell einsetzbarer Detektor existiert. Insbesondere unter Messbedingungen, bei denen sich die Teilchenfluenz am Ort der Messung stark ändert, können die Detektorsignale stark von den wahren Dosisverhältnissen abweichen, z.B. in kleinen Feldern. Im Rahmen dieser Arbeit wurde das Ansprechen verschiedener Detektortypen in solchen Extremsituationen analysiert. Dioden und Ionisationskammern verschiedener Bauformen und Größen wurden in verschiedenen Experimenten gegen Gafchromic-EBT3-Film verglichen. Das Ansprechen auf Streustrahlung konnte durch Ausblockung der Feldmitte untersucht werden, wobei zusätzlich geometrisch der Volumeneffekt korrigiert wurde. Dabei zeigte sich teils ein starkes Überansprechen. Ferner wurde gezeigt, dass die bei der Messung von Querprofilen, also sowohl in der Feldmitte, in Bereichen starker Dosisgradienten und außerhalb des Nutzfeldes, auftretenden Abweichungen durch die Verwendung einer Detektorkombination kompensiert werden können. Somit verbessert sich auch die Übereinstimmung mit den auf Film gemessenen Profilen. Für Ionisationskammern wurden effektive Messpunkte bestimmt, wobei die notwendigen Verschiebungen teils deutlich geringer waren als in den gängigen Dosimetrieprotokollen empfohlen. Insbesondere für kleinvolumige Ionisationskammern mit geringen Signalstärken kam es bei der Verwendung von im Bestrahlungsraum positionierten Elektrometern zu Störeinflüssen durch Streustrahlung. Diese Effekte konnten durch Reduzierung der das Elektrometer erreichenden Streustrahlung verringert werden. Anschließend ließ sich das Ansprechen im Aufbaubereich vergleichen. Hier zeigten sich insbesondere Unterschiede zwischen den Detektortypen, aber auch zwischen den verwendeten Polaritäten der Kammerspannung. Durch die Verwendung einer Bleifolie wurde der Einfluss von Elektronenkontamination herausgefiltert. Zusätzlich wurden das Ansprechen verschiedener Detektoren im oberflächennahen Bereich auch bei angelegten magnetischen Feldern von Feldstärken bis zu 1,1 T untersucht. In allen Fällen wurden Detektorgebrauchsgrenzen aufgezeigt. Die Erkenntnisse ermöglichen es, in den verschiedenen Extremsituationen geeignete Detektoren zu wählen, und eine Abschätzung der residualen Abweichungen durchzuführen. Gezeigt wurde auch, wo eine Detektorkombination die Genauigkeit verbessern kann. N2 - A multitude of different detectors is available for relative dosimetry in radiotherapy. Those detectors differ in their design and working principle. Depending on the detector details, the main drawbacks are the volume averaging effect or energy-dependent response. Consequently, there is no universally usable ideal detector. Especially under conditions without charged particle equilibrium, detector signals can deviate substantially from the real dose ratios, e.g. in small fields. In this work, the behavior of detectors of different types was analyzed for a range of extreme situations. Diodes and ionization chambers were compared against EBT3 film in several experiments. Detector response to scattered radiation was analyzed by blocking the central part of photon fields. The geometric volume averaging effect was corrected for simultaneously. Overresponse was observed, and the magnitude depended on the exact detector type. For profile measurements, detector response in the field center, the high gradient regions and out of the field can offset the results. A way to compensate the drawbacks of individual detectors by means of measurements with a detector combination was shown to improve the agreement of the measured profiles with profiles measured on EBT3 film. For ionization chambers, effective points of measurements were determined, which decreased with detector radius and were mostly smaller than the shifts usually recommended in dosimetry protocols. Especially for ionization chambers with very small volumes and, therefore, low signals, an offset current produced by irradiation of an electrometer positioned in the treatment room distorted the measured signals. The effect was greatly reduced by limiting the dose to the electrometer, e.g. by increasing the distance between the electrometer and the radiation source or by shielding sensitive electrometer parts with lead. Afterwards, the detector response in the build-up region was studied. Here, differences were observed between the detector types as well as the signs of the biasing voltage. Electron contamination was filtered out using a lead foil close to the collimator. In addition to that, the response of different detectors near the surface in the presence of magnetic fields was studied up to magnetic field strengths of 1.1 T. In all cases, detector limitations became obvious. This knowledge allows the choice of the proper detector for a given situation as well as an idea of the magnitude of the remaining error, as studied under the extreme model conditions described above. It was also shown how a combination of two detectors, for example, a diode and an ionization chamber, can improve the accuracy of the obtained data. KW - Dosimetrie KW - Strahlentherapie KW - Nichtgleichgewicht Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-184431 ER - TY - JOUR A1 - Bratengeier, Klaus A1 - Holubyev, Kostyantyn A1 - Wegener, Sonja T1 - Steeper dose gradients resulting from reduced source to target distance—a planning system independent study JF - Journal of Applied Clinical Medical Physics N2 - Purpose: To quantify the contribution of penumbra in the improvement of healthy tissue sparing at reduced source‐to‐axis distance (SAD) for simple spherical target and different prescription isodoses (PI). Method: A TPS‐independent method was used to estimate three‐dimensional (3D) dose distribution for stereotactic treatment of spherical targets of 0.5 cm radius based on single beam two‐dimensional (2D) film dosimetry measurements. 1 cm target constitutes the worst case for the conformation with standard Multi‐Leaf Collimator (MLC) with 0.5 cm leaf width. The measured 2D transverse dose cross‐sections and the profiles in leaf and jaw directions were used to calculate radial dose distribution from isotropic beam arrangement, for both quadratic and circular beam openings, respectively. The results were compared for standard (100 cm) and reduced SAD 70 and 55 cm for different PI. Results: For practical reduction of SAD using quadratic openings, the improvement of healthy tissue sparing (HTS) at distances up to 3 times the PTV radius was at least 6%–12%; gradient indices (GI) were reduced by 3–39% for PI between 40% and 90%. Except for PI of 80% and 90%, quadratic apertures at SAD 70 cm improved the HTS by up to 20% compared to circular openings at 100 cm or were at least equivalent; GI were 3%–33% lower for reduced SAD in the PI range 40%–70%. For PI = 80% and 90% the results depend on the circular collimator model. Conclusion: Stereotactic treatments of spherical targets delivered at reduced SAD of 70 or 55 cm using MLC spare healthy tissue around the target at least as good as treatments at SAD 100 cm using circular collimators. The steeper beam penumbra at reduced SAD seems to be as important as perfect target conformity. The authors argue therefore that the beam penumbra width should be addressed in the stereotactic studies. KW - radiotherapy KW - stereotactic irradiation KW - penumbra KW - leaf width KW - virtual isocenter Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-177424 VL - 20 IS - 1 ER -