TY - THES A1 - Obid, Nada T1 - EMG activity of masticatory muscles of patients with different bruxism grades during tasks with submaximal controlled force T1 - Elektromyographie der Kaumuskulatur unter kontrollierter Beißkraft N2 - The main objective of this study was to test whether subjects with different degrees of bruxism differ regarding EMG parameters and whether CES intervention affects those parameters. The hypothesis was that CES influences EMG parameters and after its’ cessation, all EMG parameters return to baseline (exposure–response relationship). For this purpose, forty subjects were examined, 16 men and 24 women, matched for age and gender and assigned randomly in the intervention (N=20) and control group (N=20). The procedure was as follows: 1-week inactive GC (N=40), 2 weeks inactive/active GC (N=20/N=20), 2 weeks inactive GC (N=40). Each interval was followed by a surface EMG recording from eight muscle parts (right and left anterior -, medial -, and posterior masseter and right and left anterior temporalis) under force-controlled feedback (BiteFork®) with three submaximal bite forces. The resulting EMG activity is expressed as RMS % MVC and RMS at MVC. The statistics is performed with t-test, one-way rmANOVA, and Friedman rmANOVA on ranks, according to the distribution of the data. The significance level was set at p≤0.05. The results generated from the within-groups and between-groups comparison were mostly not statistically significant and could therefore not offer clinically relevant conclu-sions. However, it cannot be excluded that a higher submaximal bite force and an extended intervention interval would have rendered different outcomes. The insufficient study sample resulted in a low observed power which makes the findings prone to Type II er-ror. It can be concluded that this study did not find any substantiating differences be-tween the EMG values of participants with various bruxism activity and that CES could not influence the studied EMG parameters in the two weeks intervention time. Our hypothesis which supposes that subjects with high and low bruxism activity differ in RMS % MVC could not be verified. However, with the gained knowledge, it is recom-mended to further elaborate a definite bruxism diagnosis by using portable EMG devices. N2 - Das Hauptziel dieser Studie bestand darin zu prüfen, ob sich Probanden mit unterschie-dlichem Bruxismusgrad hinsichtlich der EMG-Parameter unterscheiden und darüber hinaus, ob bedingte elektrische Stimulation diese beeinflusst. Die Hypothese lautete, dass die bedingte elektrische Stimulation die EMG-Parameter beeinflusst und nach Ein-stellung alle EMG-Parameter zum Ausgangswert zurückkehren. Zu diesem Zweck wurden vierzig Probanden untersucht (16 Männer und 24 Frauen), die nach Alter und Geschlecht gematcht und zufällig in die Interventions- (N=20) und die Kontrollgruppe (N= 20) eingeteilt wurden. Das Studienprotokoll begann mit einer Woche inaktivem GC (N=40), gefolgt von zwei Wochen inaktivem/aktivem GC (N=20/N=20) und endete mit zwei Wochen inaktivem GC (N=40). Nach jedem Intervall erfolgte eine Ober-flächen-EMG-Aufzeichnung der acht Muskelpartien (rechter und linker anteriorer -, medi-aler - und posteriorer Masseter, sowie rechter und linker anteriorer Temporalis) unter Kraft-kontrolliertem Feedback (BiteFork®). Die statistische Auswertung erfolgte mittels t-Test, einseitiger rmANOVA und Friedman rmANOVA, je nach Verteilung der Daten. Das Signifikanzniveau wurde auf p ≤ 0,05 festgelegt. Die Ergebnisse aus den Vergleichen waren statistisch nicht signifikant und konnten da-her keine klinisch relevanten Schlussfolgerungen liefern. Dennoch kann nicht ausgeschlossen werden, dass eine höhere Beißkraft und ein länge-res Interventionsintervall zu anderen Ergebnissen führen könnten. Die kleine Stichprobe führte zu einer geringen Teststärke, wodurch ein Risiko für einen Fehler 2. Art besteht. Es lässt sich schlussfolgern, dass diese Studie keine begründeten Unterschiede zwi-schen den EMG-Werten von Teilnehmern unterschiedlicher Bruxismusaktivität gefunden hat, und außerdem, dass die bedingte elektrische Stimulation in dem zweiwöchigen In-tervall zu keiner Veränderung der EMG-Parameter geführt hat. Unsere Hypothese, dass sich Probanden mit hoher und niedriger Bruxismusaktivität im EMG (RMS % MVC) voneinander unterscheiden, konnte nicht verifiziert werden. KW - Bruxismus KW - Elektromyographie KW - Fragebogen KW - contingent electrical stimulation KW - bedingte elektrische Stimulation KW - Klinische Studie Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-345883 ER - TY - THES A1 - Rodamer, Michael T1 - Development of practice-oriented LC-MS/MS methods for the determination of important drugs and their application for building PK/PD concepts T1 - Die Verwendung von praxisorientierten LC-MS/MS Methoden für die Bestimmung von wichtigen Arzneistoffen und ihre Verwendung zur Erstellung von PK/PD Konzepten N2 - In this thesis eight robust and reliable LC-MS/MS methods were developed and validated to analyze atorvastatin, clopidogrel, furosemide, itraconazole, loratadine, naproxen, nisoldipine and sunitinib in human plasma. The active metabolites 2-hydroxyatorvastatin, 4-hydroxyatorvastatin, hydroxyitraconazole, descarboethoxy-loratadine, 4-hydroxynisoldipine and N-desethylsunitinib were also included in the corresponding methods. Due to the different physical, chemical and pharmacokinetic properties of the analytes a wide spectrum regarding sample preparation techniques, chromatography and mass spectrometric detection was covered. Protein precipitation methods were developed for furosemide, itraconazole, naproxen, nisoldipine and sunitinib. Liquid-liquid extraction methods were developed for atorvastatin, clopidogrel and loratadine. Criteria to choose protein precipitation or liquid-liquid extraction were the final plasma concentrations of the drugs, which are mainly dependant on the dose, bioavailability and t1/2 and of course cost-effectiveness. Altogether, the methods have a concentration range from 0.001 ng/mL (LLOQ of clopidogrel) to 50000 ng/mL (highest calibration point for naproxen), covering 5 x 107 orders of magnitude. The runtime of the methods ranged from 2 to 4 minutes, facilitating a high sample throughput. All developed methods were validated according to recent guidelines as they were used to analyze sampes from clinical trials. Excellent linearity, intra-day and inter-day precision and accuracy were observed in the validated calibration ranges. Hemolyzed, lipemic and different batches of human plasma as well as sample dilution did not affect the determiantion of the analytes. Clopidogrel, loratadine, nisoldipine and sunitinib and if available their metabolites were subjected to a matrix effect test, resulting in no influence of different batches of human plasma on the analytical methods. Noteworthy is clopidogrel that shows a slight effect on one of the two used mass spectrometers. However, that effect was reproducible and did therefore not affect clopidogrel determination. No evidence of instability during chromatography, extraction and sample storage processes for all analytes except 4-hydroxyatorvastatin was found, for which a significant decrease was observed after three months. During incurred sample reanalysis of study samples 95 % of the samples were within ±15 % with respect to the first analysis. Moreover, the atorvastatin, loratadine and clopidogrel method were compared on two generations of triple quadrupole mass spectrometers, the API 3000™ and the API 5000™. The new ion source and the changes in the ion path of the API 5000™ provided higher sensitivity, the extend depending on the substance. However, the API 3000™ had very good precision in the performed system comparison. The validated methods showed excellent performance and quality data during routine sample analysis of eight clinical trials. Moreover, they are suitable for high sample throughput due to their short run times. N2 - In dieser Dissertation wurden acht robuste und verlässliche LC-MS/MS-Methoden zur Analyse von Atorvastatin, Clopidogrel, Furosemid, Itraconazol, Loratadin, Naproxen Nisoldipin und Sunitinib in Humanplasma entwickelt und validiert. Außerdem enthalten die Methoden die aktiven Metaboliten 2-Hydroxyatorvastatin, 4-Hydroxyatorvastatin, Hydroxyitraconazol, Descarboethoxyloratadin, 4-Hydroxynisoldipin und N-Desethylsunitinib. Wegen der unterschiedlichen physikalischen, chemischen und pharmakokinetischen Eigenschaften der Analyten, deckt diese Arbeit ein weites Spektrum bezüglich Probenaufarbeitung, Chromatographie und Massenspektrometrie ab. Präzipitationsmethoden wurden für Furosemid, Itraconazol, Naproxen, Nisoldipin und Sunitinib entwickelt. Flüssig-flüssig-Extraktionen wurden für Atorvastatin, Clopidogrel und Loratadin entwickelt. Kriterien für die Auswahl von Präzipitation oder Extraktion waren die erwartete Plasmakonzentration, die im Wesentlichen von der Dosis, Bioverfügbarkeit und Halbwertszeit abhängig ist, und natürlich Kosteneffektivität. Insgesamt erstrecken sich die Methoden über einen Kalibrierbereich von 0.001 ng/mL (LLOQ von Clopidogrel) bis zu 50000 ng/mL (HLOQ von Naproxen), das entspricht 5x107 Größenordnungen. Die Laufzeiten pro Probe liegen im Bereich von zwei bis vier Minuten, was einen sehr hohen Probendurchsatz ermöglicht. Alle in dieser Arbeit entwickelten Methoden wurden gemäß aktueller Richtlinien (FDA, GLP) validiert und verwendet um Proben aus Pharmakokinetikstudien zu analysieren. Ausgezeichnete Linearität, Präzision und Genauigkeit zeichnen diese Methoden aus. Hämolysiertes, lipämisches und verschiedene Batches von Humanplasma, sowie Vorverdünnung hatten bei keiner Methode Einfluss auf die Bestimmung der Analyten. Clopidogrel, Loratadin, Nisoldipin und Sunitinib und gegebenenfalls deren Metabolite wurden einem Matrix-Effekt-Test unterzogen. Dabei wurde festgestellt, dass keine der Methoden durch die Probenmatrix beeinflusst wurde. Erwähnenswert ist Clopidogrel, da an einem der Massenspektrometer ein leichter Effekt beobachtet werden konnte, der sich auf alle untersuchten Matrices gleich auswirkte und somit keinen Einfluss auf die gesamte Methode hatte. Weiterhin fand sich bei keiner der untersuchten Substanzen ein Hinweis auf Instabilität während der Probenlagerung, -aufarbeitung und -messung, außer bei 4-Hydroxyatorvastatin, dessen Konzentration nach drei Monaten signifikant abnahm. Während der Reanalyse von Studienproben (incurred samples) lagen über 95 % der Proben innerhalb von ±15 % im Vergleich zur ersten Messung. Außerdem wurden die Methoden zur Bestimmung von Atorvastatin, Loratadin und Clopidogrel an zwei Generationen von Massenspektrometern verglichen, nämlich dem API 3000™ und dem API 5000™. Die neue Ionenquelle und die Verbesserungen im Ionenpfad beim API 5000™ ermöglichten - abhängig von der analysierten Substanz - höhere Sensitivität. Allerdings konnte das API 3000™ bei den durchgeführten Experimenten mit einer hohen Präzision aufwarten. Die validierten Methoden zeigten im Alltagbetrieb bei der Messung von acht klinischen Studien hervorragende Performance und Qualitätsdaten. Darüber hinaus sind die Methoden aufgrund ihrer kurzen Laufzeiten ideal für Messungen die einen hohen Probendurchsatz erfordern. KW - LC-MS KW - Validierung KW - Tandem-Massenspektrometrie KW - Bioanalytik KW - Klinische Studie KW - LC-MS/MS KW - Tandem Mass Spectrometry KW - Method Validation KW - Pharmacokinetics Y1 - 2011 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-70809 ER -