TY - THES A1 - Hock, Michael T1 - Methods for Homogenization of Spatio-Temporal B\(_0\) Magnetic Field Variations in Cardiac MRI at Ultra-High Field Strength T1 - Methoden zur Homogenisierung räumlicher und zeitlicher Variationen des B\(_0\)-Feldes in der kardialen Ultrahochfeld-MRT N2 - Cardiovascular disease is one of the leading causes of death worldwide and, so far, echocardiography, nuclear cardiology, and catheterization are the gold standard techniques used for its detection. Cardiac magnetic resonance (CMR) can replace the invasive imaging modalities and provide a "one-stop shop" characterization of the cardiovascular system by measuring myocardial tissue structure, function and perfusion of the heart, as well as anatomy of and flow in the coronary arteries. In contrast to standard clinical magnetic resonance imaging (MRI) scanners, which are often operated at a field strength of 1.5 or 3 Tesla (T), a higher resolution and subsequent cardiac parameter quantification could potentially be achieved at ultra-high field, i.e., 7 T and above. Unique insights into the pathophysiology of the heart are expected from ultra-high field MRI, which offers enhanced image quality in combination with novel contrast mechanisms, but suffers from spatio-temporal B0 magnetic field variations. Due to the resulting spatial misregistration and intra-voxel dephasing, these B0-field inhomogeneities generate a variety of undesired image artifacts, e.g., artificial image deformation. The resulting macroscopic field gradients lead to signal loss, because the effective transverse relaxation time T2* is shortened. This affects the accuracy of T2* measurements, which are essential for myocardial tissue characterization. When steady state free precession-based pulse sequences are employed for image acquisition, certain off-resonance frequencies cause signal voids. These banding artifacts complicate the proper marking of the myocardium and, subsequently, systematic errors in cardiac function measurements are inevitable. Clinical MR scanners are equipped with basic shim systems to correct for occurring B0-field inhomogeneities and resulting image artifacts, however, these are not sufficient for the advanced measurement techniques employed for ultra-high field MRI of the heart. Therefore, this work focused on the development of advanced B0 shimming strategies for CMR imaging applications to correct the spatio-temporal B0 field variations present in the human heart at 7 T. A novel cardiac phase-specific shimming (CPSS) technique was set up, which featured a triggered B0 map acquisition, anatomy-matched selection of the shim-region-of-interest (SROI), and calibration-based B0 field modeling. The influence of technical limitations on the overall spherical harmonics (SH) shim was analyzed. Moreover, benefits as well as pitfalls of dynamic shimming were debated in this study. An advanced B0 shimming strategy was set up and applied in vivo, which was the first implementation of a heart-specific shimming approach in human UHF MRI at the time. The spatial B0-field patterns which were measured in the heart throughout this study contained localized spots of strong inhomogeneities. They fluctuated over the cardiac cycle in both size and strength, and were ideally addressed using anatomy-matched SROIs. Creating a correcting magnetic field with one shim coil, however, generated eddy currents in the surrounding conducting structures and a resulting additional, unintended magnetic field. Taking these shim-to-shim interactions into account via calibration, it was demonstrated for the first time that the non-standard 3rd-order SH terms enhanced B0-field homogeneity in the human heart. However, they were attended by challenges for the shim system hardware employed in the presented work, which was indicated by the currents required to generate the optimal 3rd-order SH terms exceeding the dynamic range of the corresponding shim coils. To facilitate dynamic shimming updated over the cardiac cycle for cine imaging, the benefit of adjusting the oscillating CPSS currents was found to be vital. The first in vivo application of the novel advanced B0 shimming strategy mostly matched the simulations. The presented technical developments are a basic requirement to quantitative and functional CMR imaging of the human heart at 7 T. They pave the way for numerous clinical studies about cardiac diseases, and continuative research on dedicated cardiac B0 shimming, e.g., adapted passive shimming and multi-coil technologies. N2 - Herz-Kreislauf-Erkrankungen zählen zu den häufigsten Todesursachen weltweit und werden bisher in der Regel mittels Echokardiographie, Nuklearkardiologie und Katheterisierung untersucht. Die kardiale Magnetresonanztomographie hat das Potential diese invasiven Bildgebungsmodalitäten zu ersetzen. Dabei können sowohl das kardiovaskuläre System anhand der myokardialen Gewebestruktur sowie der Funktion und Perfusion des Herzens als auch Anatomie und Blutfluss der Koronararterien während einer einzigen Untersuchung charakterisiert werden. Im Gegensatz zu den weit verbreiteten klinischen Magnetresonanztomographie- (MRT) Geräten, welch häufig bei magnetischen Feldstärken zwischen 1.5 und 3T operieren, ermöglichen Feldstärken von 7 Tesla und mehr eine höhere Auflösung und somit eine akkuratere Quantifizierung kardialer Parameter. Die Ultrahochfeld-Magnetresonanztomographie (UHF-MRT) ermöglicht einzigartige Einblicke in die Pathophysiologie des Herzens. Neuartige Kontrastmechanismen und die verbesserte Bildqualität leiden jedoch unter Inhomogenitäten des statischen magnetischen B0-Feldes. Aufgrund der daraus resultierenden falschen räumlichen Registrierung der Voxel und einer Dephasierung des Signals innerhalb eines Voxels erzeugen diese Inhomogenitäten des B0-Feldes eine Vielzahl unerwünschter Bildartefakte, beispielsweise eine künstliche Deformation des Bildes. Die resultierenden makroskopischen Gradienten führen zu Signalverlust und beeinträchtigen die Messung der effektiven transversalen T2*-Relaxationszeit, welche für die Charakterisierung myokardialen Gewebes essentiell ist. Vor allem bei der Bildakquisition mittels der Steady State Free Precession Methode führen Inhomogenitäten des B0-Feldes zu Signalauslöschungen. Die dadurch entstehenden Bildartefakte erschweren die genaue Markierung des Myokards und haben so systematische Fehler bei der Bestimmung der kardialen Funktion zur Folge. Klinische MRT-Geräte sind dabei mit sogenannten Shim-Systemen ausgestattet um die Inhomogenitäten des B0-Feldes zu korrigieren. Für die kardiale UHF-MRT des Herzens sind diese standardisierten Shim-Systeme allerdings nicht mehr ausreichend. Im Fokus stand deshalb die Entwicklung moderner Methoden zur räumlichen und zeitlichen Korrektur der B0-Inhomogenitäten, welche als „Shimming“ bezeichnet wird, für die kardiale UHF-MRT. Es wurde eine neue, herzphasen-spezifische Shimming-Strategie untersucht, welche auf der getriggerten Datenaufnahme, der Optimierung für die Anatomie des Herzens, sowie der kalibrierungsbasierten Modellierung des korrigierenden Magnetfeldes basierte. Zudem wurde der Einfluss technischer Limitationen der Hardware auf das Shimming, insbesondere das dynamische Shimming, in dieser Studie erörtert. Schließlich wurde die entwickelte neuartige Shimming-Strategie in vivo evaluiert, welche zu diesem Zeitpunkt die erste Implementierung einer herzspezifischen Shimming-Strategie in der humanen kardialen UHF-MRT darstellte. Räumlich wies das B0-Feld, welches im Rahmen dieser Studie im Herzen gemessen wurde, lokalisierte Inhomogenitäten im Myokardium auf. Diese variierten zudem in ihrer Größe sowie der Stärke der B0-Inhomogenität zeitlich über den Herzzyklus hinweg und ließen sich mittels anatomisch angepasstem, kalibrierungsbasiertem Shimming deutlich reduzieren. Erzeugt man ein korrigierendes Magnetfeld mittels einer Shim-Spule, so werden jedoch Wirbelströme in nahen leitenden Strukturen und weiterhin ein zusätzliches, unerwünschtes Magnetfeld erzeugt. Berücksichtigt man diese Wechselwirkungen zwischen den verschiedenen Shim-Spulen, konnte erstmalig der Vorteil von korrigierenden Magnetfeldern in der Form von Kugelflächenfunktionen der dritten Ordnung für die kardiale UHF-MRT gezeigt werden. Hierbei waren jedoch die erforderlichen, besonders starken Ströme in den Shim-Spulen zu berücksichtigen, welche über den Herzzyklus hinweg oszillierten und für dynamisches Shimming angepasst werden sollten. Die erste in vivo Anwendung der neu entwickelten Shim-Strategie stimmte gut mit den vorigen Simulationen überein. Die vorgestellten technischen Entwicklungen stellen grundlegende Anforderungen an die quantitative und funktionelle kardialer UHF-MRT dar. Klinische Studien zu kardialen Erkrankungen wie der Herzinsuffizienz erscheinen nun ebenso in Reichweite wie weitere Forschung zu kardialem B0-Shimming basierend auf angepasstem passiven Shimming sowie Multikanal-Spulen. KW - Kernspintomografie KW - Bildgebendes Verfahren KW - 7 T KW - B0 KW - Cardiac MRI KW - Shimming KW - Ultrahigh field Y1 - 2024 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-348213 ER - TY - THES A1 - Lankl, Sebastian T1 - Untersuchung des Einflusses von Myokardinfarkten auf die Wandbewegungsgeschwindigkeit der linken midventrikulären Segmente mittels Phasenkontrast-MRT T1 - Investigation of the influence of myocardial infarction on the wall motion velocity of the left midventricular segments using phase contrast MRI N2 - Die Letalität des Myokardinfarktes ist in Deutschland rückläufig, die Bedeutung von Folgeerkrankungen des Myokardinfarktes nimmt daher zu. Durch pathologische Umbauprozesse (Remodeling) nach Myokardinfarkten kann die Mechanik des linken Ventrikels beeinträchtigt werden, sodass eine ischämische Kardiomyopathie entsteht. Im Rahmen dieser Arbeit wurde der Einfluss von Myokardinfarkten auf die Wandbewegungsgeschwindigkeit des linken Ventrikels mittels Tissue Phase Mapping untersucht. Tissue Phase Mapping ist eine MRT-basierte Untersuchungstechnik, welche die Wandbewegung des linken Ventrikels als Gewebegeschwindigkeit mit hoher zeitlicher und räumlicher Auflösung in drei Dimensionen quantifiziert. Bisher durchgeführte Tissue Phase Mapping-Studien bei Infarktpatienten werden in ihrer Aussagekraft durch eine veraltete Sequenztechnik und ein heterogenes Patientenkollektiv limitiert. In dieser Arbeit wurden daher selektiv Patienten mit stattgehabtem Vorderwandinfarkt mit einem bisher unveröffentlichten aktuellen Tissue Phase Mapping-Protokoll untersucht und mit einer Kontrollgruppe verglichen. Hierbei wurden statistisch signifikante pathologische Veränderungen der lokalen myokardialen Rotation und der diastolischen Expansion in radialer Richtung in postischämisch vernarbten Segmenten identifiziert. Aus anderen MRT-basierten Messmethoden (unter anderem Strain-Encoded Magnetic Resonance und Displacement Encoding With Stimulated Echos) ist bereits bekannt, dass die Rotationsbewegung in postischämisch vernarbten Segmenten pathologisch verändert ist. In dieser Arbeit wurde jedoch erstmals eine Reduktion und zum Teil eine Umkehr der lokalen myokardialen Rotation in vernarbten Segmenten mittels Tissue Phase Mapping nachgewiesen. Limitationen dieser Arbeit sind insbesondere die hohe Messzeit und die Anfälligkeit der Untersuchungstechnik für Bewegungsartefakte. Zudem konnten in anderen Studien Veränderungen der linksventrikulären Mechanik in vernarbten Segmenten mittels Strain-Parametern mit höherer Sensitivität erfasst werden. Nichtsdestotrotz könnten Weiterentwicklungen des Tissue Phase Mappings in Zukunft dazu beitragen, die linksventrikuläre Mechanik im Rahmen des Remodelings besser zu verstehen und die ischämische Kardiomyopathie früher zu diagnostizieren. N2 - The mortality rate of myocardial infarction is declining in Germany, so the relevance of secondary diseases of myocardial infarction is increasing. Pathological remodelling processes after myocardial infarction can impair the mechanics of the left ventricle, resulting in ischaemic cardiomyopathy. In this study, the influence of myocardial infarction on the wall motion velocity of the left ventricle was investigated using tissue phase mapping. Tissue phase mapping is an MRI-based examination technique that quantifies the wall motion of the left ventricle as tissue velocity with high temporal and spatial resolution in three dimensions. Tissue phase mapping studies performed so far in infarct patients are limited in their significance by an outdated sequencing technique and a heterogeneous patient population. In this study, therefore, patients with a previous anterior wall infarction were selectively examined using a previously unpublished, up-to-date tissue phase mapping protocol and compared with a control group. Statistically significant pathological changes in local myocardial rotation and diastolic expansion in the radial direction were identified in post-ischaemic scarred segments. It is already known from other MRI-based imaging techniques (including strain-encoded magnetic resonance and displacement encoding with stimulated echoes) that the rotational motion in post-ischaemic scarred segments is pathologically altered. In this study, however, a reduction and partial reversal of local myocardial rotation in scarred segments was demonstrated for the first time using tissue phase mapping. The main limitations of this study are the long measurement time and the susceptibility of the examination technique to movement artefacts. In addition, other studies have been able to detect changes in left ventricular mechanics in scarred segments using strain parameters with higher sensitivity. Nevertheless, further developments in tissue phase mapping might contribute to a better understanding of left ventricular mechanics in the context of remodelling and to earlier diagnosis of ischaemic cardiomyopathy. KW - Kernspintomografie KW - Kardiale Phasenkontrast-MRT Y1 - 2024 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-346826 ER - TY - THES A1 - Kleineisel, Jonas T1 - Variational networks in magnetic resonance imaging - Application to spiral cardiac MRI and investigations on image quality T1 - Variational Networks in der Magnetresonanztomographie - Anwendung auf spirale Herzbildgebung und Untersuchungen zur Bildqualität N2 - Acceleration is a central aim of clinical and technical research in magnetic resonance imaging (MRI) today, with the potential to increase robustness, accessibility and patient comfort, reduce cost, and enable entirely new kinds of examinations. A key component in this endeavor is image reconstruction, as most modern approaches build on advanced signal and image processing. Here, deep learning (DL)-based methods have recently shown considerable potential, with numerous publications demonstrating benefits for MRI reconstruction. However, these methods often come at the cost of an increased risk for subtle yet critical errors. Therefore, the aim of this thesis is to advance DL-based MRI reconstruction, while ensuring high quality and fidelity with measured data. A network architecture specifically suited for this purpose is the variational network (VN). To investigate the benefits these can bring to non-Cartesian cardiac imaging, the first part presents an application of VNs, which were specifically adapted to the reconstruction of accelerated spiral acquisitions. The proposed method is compared to a segmented exam, a U-Net and a compressed sensing (CS) model using qualitative and quantitative measures. While the U-Net performed poorly, the VN as well as the CS reconstruction showed good output quality. In functional cardiac imaging, the proposed real-time method with VN reconstruction substantially accelerates examinations over the gold-standard, from over 10 to just 1 minute. Clinical parameters agreed on average. Generally in MRI reconstruction, the assessment of image quality is complex, in particular for modern non-linear methods. Therefore, advanced techniques for precise evaluation of quality were subsequently demonstrated. With two distinct methods, resolution and amplification or suppression of noise are quantified locally in each pixel of a reconstruction. Using these, local maps of resolution and noise in parallel imaging (GRAPPA), CS, U-Net and VN reconstructions were determined for MR images of the brain. In the tested images, GRAPPA delivers uniform and ideal resolution, but amplifies noise noticeably. The other methods adapt their behavior to image structure, where different levels of local blurring were observed at edges compared to homogeneous areas, and noise was suppressed except at edges. Overall, VNs were found to combine a number of advantageous properties, including a good trade-off between resolution and noise, fast reconstruction times, and high overall image quality and fidelity of the produced output. Therefore, this network architecture seems highly promising for MRI reconstruction. N2 - Eine Beschleunigung des Bildgebungsprozesses ist heute ein wichtiges Ziel von klinischer und technischer Forschung in der Magnetresonanztomographie (MRT). Dadurch könnten Robustheit, Verfügbarkeit und Patientenkomfort erhöht, Kosten gesenkt und ganz neue Arten von Untersuchungen möglich gemacht werden. Da sich die meisten modernen Ansätze hierfür auf eine fortgeschrittene Signal- und Bildverarbeitung stützen, ist die Bildrekonstruktion ein zentraler Baustein. In diesem Bereich haben Deep Learning (DL)-basierte Methoden in der jüngeren Vergangenheit bemerkenswertes Potenzial gezeigt und eine Vielzahl an Publikationen konnte deren Nutzen in der MRT-Rekonstruktion feststellen. Allerdings besteht dabei das Risiko von subtilen und doch kritischen Fehlern. Daher ist das Ziel dieser Arbeit, die DL-basierte MRT-Rekonstruktion weiterzuentwickeln, während gleichzeitig hohe Bildqualität und Treue der erzeugten Bilder mit den gemessenen Daten gewährleistet wird. Eine Netzwerkarchitektur, die dafür besonders geeignet ist, ist das Variational Network (VN). Um den Nutzen dieser Netzwerke für nicht-kartesische Herzbildgebung zu untersuchen, beschreibt der erste Teil dieser Arbeit eine Anwendung von VNs, welche spezifisch für die Rekonstruktion von beschleunigten Akquisitionen mit spiralen Auslesetrajektorien angepasst wurden. Die vorgeschlagene Methode wird mit einer segmentierten Rekonstruktion, einem U-Net, und einem Compressed Sensing (CS)-Modell anhand von qualitativen und quantitativen Metriken verglichen. Während das U-Net schlecht abschneidet, zeigen die VN- und CS-Methoden eine gute Bildqualität. In der funktionalen Herzbildgebung beschleunigt die vorgeschlagene Echtzeit-Methode mit VN-Rekonstruktion die Aufnahme gegenüber dem Goldstandard wesentlich, von etwa zehn zu nur einer Minute. Klinische Parameter stimmen im Mittel überein. Die Bewertung von Bildqualität in der MRT-Rekonstruktion ist im Allgemeinen komplex, vor allem für moderne, nichtlineare Methoden. Daher wurden anschließend forgeschrittene Techniken zur präsizen Analyse von Bildqualität demonstriert. Mit zwei separaten Methoden wurde einerseits die Auflösung und andererseits die Verstärkung oder Unterdrückung von Rauschen in jedem Pixel eines untersuchten Bildes lokal quantifiziert. Damit wurden lokale Karten von Auflösung und Rauschen in Rekonstruktionen durch Parallele Bildgebung (GRAPPA), CS, U-Net und VN für MR-Aufnahmen des Gehirns berechnet. In den untersuchten Bildern zeigte GRAPPA gleichmäßig eine ideale Auflösung, aber merkliche Rauschverstärkung. Die anderen Methoden verhalten sich lokal unterschiedlich je nach Struktur des untersuchten Bildes. Die gemessene lokale Unschärfe unterschied sich an den Kanten gegenüber homogenen Bildbereichen, und Rauschen wurde überall außer an Kanten unterdrückt. Insgesamt wurde für VNs eine Kombination von verschiedenen günstigen Eigenschaften festgestellt, unter anderem ein guter Kompromiss zwischen Auflösung und Rauschen, schnelle Laufzeit, und hohe Qualität und Datentreue der erzeugten Bilder. Daher erscheint diese Netzwerkarchitektur als ein äußerst vielversprechender Ansatz für MRT-Rekonstruktion. KW - Kernspintomografie KW - Convolutional Neural Network KW - Maschinelles Lernen KW - Bildgebendes Verfahren KW - magnetic resonance imaging KW - convolutional neural network KW - variational network KW - cardiac imaging KW - machine learning KW - local point-spread function KW - resolution KW - g-factor Y1 - 2024 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-347370 ER - TY - THES A1 - Fey, Philipp T1 - KI-gestützte MR-Klassifizierung von Zellen und zellulärer Differenzierung T1 - AI-assisted MR classification of cells and cellular differentiation N2 - Für die Verwendung von zellbasierten Therapeutika ist vor allem die korrekt Identifikation sowohl vom Ausgangsmaterial wie auch dem produziertem Material von zentraler Wichtigkeit. In dieser Arbeit wurde eine Methodik entwickelt, welche eine nicht-invasive Klassifizierung von Zellen und zellulärer Entwicklung aufgrund ihrer zweidimensionalen Magnetresonanz-Korrelationsspektren ermöglichte. Hierzu wurde ein mobiler MR-Scanner mit einer Feldstärke von 0.5T und einem Isozentrum von 1 cm3 verwendet. Aufgrund der kompakten und leichten Bauweise war es möglich, das System in normalen Zellkulturlaboren zu verwenden. Von den Proben wurde ein zweidimensionales T1/T2 -Korrelationsspektrum aufgenommen, anhand dessen die Zellen klassifiziert werden sollten. Mithilfe von Agarose-Dotagraf® -Zell- Phantomen konnte die Stabilität und Reproduzierbarkeit des Messsystems und der verwendeten Sequenz validiert werden. Aufgrund der unter Umständen recht langen Messzeiten der MR-Technologie war auch die Handhabung und Kultur der Zellproben während des Messprozesses von großer Bedeutung. Um hierfür den Durchsatz an Proben zu erhöhen, wurde eine kostengünstige und ebenfalls mobile Robotikanlage entwickelt. Diese basierte auf dem kommerziell erhältlichen Roboterarm Braccio, welcher durch einen Arduino Mega Mikrocontroller gesteuert wurde. Mit bis zu 24 Proben pro Tag konnte durch die Automatisierung der Durchsatz an Proben um den Faktor 3 – 4 gesteigert werden. Durch den entwickelten Prozess war es möglich, eine umfangreiche Datenbank – bestehend aus 362 unabhängigen Messungen (biologische Replikate) – aufzubauen. Die Datenbank enthielt Messungen von zehn unterschiedlichen Zelllinien. Zusätzlich wurden T1/T2 -Korrelationsspektren von mesenchymalen Stromazellen (MSCs) vor und nach deren Differenzierung zu Adipocyten aufgenommen, um ihre zelluläre Entwicklung nicht-invasiv charakterisieren zu können. Die aufgenommenen Daten wurden mithilfe einer geeigneten Support Vector Machine wie auch angepassten künstlichen neuronalen Netzwerken klassifiziert. Mithilfe dieser Methoden konnten die Zelllinien und MSCs anhand ihrer aufgenommenen Korrelationsspektren mit einer Genauigkeit von bis zu 98% klassifiziert werden. Diese hohe Treffsicherheit legte den Schluss nahe, dass die Kombination aus nichtinvasiver, zweidimensionaler T1/T2 -MR-Relaxometrie und der Verwendung von geeigneten Methoden des machine learning und der künstlichen Intelligenz eine effiziente Methodik für die nicht-invasive Klassifizierung von Zellen sowie zellulärer Entwicklung darstellt. N2 - For the use of cell-based therapeutics, the correct identification of both the starting material and the produced material is of central importance. In this work, a methodology was developed that allowed non-invasive classification of cells and cellular development based on their two-dimensional magnetic resonance correlation spectra. For this purpose, a mobile MR scanner with a field strength of 0.5T and an isocenter of 1 cm3 was used. Due to its compact and lightweight design, it was possible to use the system in normal cell culture laboratories. A two-dimensional T1/T2 - correlation spectrum was recorded from the samples, which was used to classify the cells. Agarose-Dotagraf® cell phantoms were used to validate the stability and reproducibility of the measurement system and the sequence used. Due to the possibly quite long measurement times of the MR technology, the handling and culture of the cell samples during the measurement process was also of great importance. To increase the throughput of samples for this purpose, a low-cost and also mobile robotic system was developed. This was based on the commercially available Braccio robotic arm, which was controlled by an Arduino Mega microcontroller. With up to 24 samples per day, the automation increased the throughput of samples by a factor of 3 - 4. Through the developed process it was possible to build an extensive database – consisting of 362 independent measurements (biological replicates) . The database contained measurements from ten different cell lines. In addition, T1/T2 -correlation spectra of mesenchymal stromal cells (MSCs) before and after their differentiation into adipocytes were recorded to characterize their cellular development non-invasively. The recorded data were classified using an appropriate Support Vector Machine as well as adapted artificial neural networks. Using these methods, the cell lines and MSCs could be classified based on their recorded correlation spectra with an accuracy of up to 98 %. This high accuracy suggested that the combination of non-invasive, two-dimensional T1/T2 -MR relaxometry and the use of appropriate machine learning and artificial intelligence methods is an efficient methodology for the non-invasive classification of cells as well as cellular development. KW - MRT KW - NMR KW - Künstliche Ingelligenz KW - Kernspintomografie KW - Stammzelle Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-345164 ER - TY - THES A1 - Gram, Maximilian T1 - Neue Methoden der Spin-Lock-basierten Magnetresonanztomographie: Myokardiale T\(_{1ρ}\)-Quantifizierung und Detektion magnetischer Oszillationen im nT-Bereich T1 - New methods of spin-lock-based magnetic resonance imaging: myocardial T\(_{1ρ}\) quantification and detection of magnetic oscillations in the nT range N2 - Das Ziel der vorliegenden Arbeit war die Entwicklung neuer, robuster Methoden der Spin-Lock-basierten MRT. Im Fokus stand hierbei vorerst die T1ρ-Quantifizierung des Myokards im Kleintiermodell. Neben der T1ρ-Bildgebung bietet Spin-Locking jedoch zusätzlich die Möglichkeit der Detektion ultra-schwacher, magnetischer Feldoszillationen. Die Projekte und Ergebnisse, die im Rahmen dieses Promotionsvorhabens umgesetzt und erzielt wurden, decken daher ein breites Spektrum der Spin-lock basierten Bildgebung ab und können grob in drei Bereiche unterteilt werden. Im ersten Schritt wurde die grundlegende Pulssequenz des Spin-Lock-Experimentes durch die Einführung des balancierten Spin-Locks optimiert. Der zweite Schritt war die Entwicklung einer kardialen MRT-Sequenz für die robuste Quantifizierung der myokardialen T1ρ-Relaxationszeit an einem präklinischen Hochfeld-MRT. Im letzten Schritt wurden Konzepte der robusten T1ρ-Bildgebung auf die Methodik der Felddetektion mittels Spin-Locking übertragen. Hierbei wurden erste, erfolgreiche Messungen magnetischer Oszillationen im nT-Bereich, welche lokal im untersuchten Gewebe auftreten, an einem klinischen MRT-System im menschlichen Gehirn realisiert. N2 - The main goal of the present work was to develop new, robust methods of spin-lock-based MRI. The initial focus was on T1ρ quantification of the myocardium in small animal models. However, in addition to T1ρ imaging, spin-locking offers the possibility of detecting ultra-weak magnetic field oscillations. The projects and results realized and obtained in this PhD project therefore cover a broad spectrum of spin-lock based imaging and can be roughly divided into three areas. The first step was to optimize the basic pulse sequence of the spin-lock experiment by introducing balanced spin-locking. The second step was to develop a cardiac MRI sequence for robust quantification of the myocardial T1ρ relaxation time on a preclinical high-field MRI scanner. In the final step, concepts of robust T1ρ imaging were adapted to spin-lock based magnetic field detection. First successful measurements of magnetic field oscillations in the nT range, which occur locally inside the tissue under investigation, were realized on a clinical MRI system in the human brain. KW - Kernspintomografie KW - Magnetresonanztomographie KW - Kernspinresonanz KW - Spin-Lock KW - T1ρ KW - T1rho KW - Kardio-MRT KW - Rotary Excitation KW - Myokardiale T1ρ-Quantifizierung KW - Felddetektion KW - funktionelle MRT Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-322552 ER - TY - THES A1 - Portmann, Johannes T1 - Accelerated inversion recovery MRI of the myocardium using spiral acquisition T1 - Beschleunigte Inversion-Recovery MR-Bildgebung des Myokards mit spiralen Auslesezügen N2 - This work deals with the acceleration of cardiovascular MRI for the assessment of functional information in steady-state contrast and for viability assessment during the inversion recovery of the magnetization. Two approaches are introduced and discussed in detail. MOCO-MAP uses an exponential model to recover dynamic image data, IR-CRISPI, with its low-rank plus sparse reconstruction, is related to compressed sensing. MOCO-MAP is a successor to model-based acceleration of parametermapping (MAP) for the application in the myocardial region. To this end, it was augmented with a motion correction (MOCO) step to allow exponential fitting the signal of a still object in temporal direction. Iteratively, this introduction of prior physical knowledge together with the enforcement of consistency with the measured data can be used to reconstruct an image series from distinctly shorter sampling time than the standard exam (< 3 s opposed to about 10 s). Results show feasibility of the method as well as detectability of delayed enhancement in the myocardium, but also significant discrepancies when imaging cardiac function and artifacts caused already by minor inaccuracy of the motion correction. IR-CRISPI was developed from CRISPI, which is a real-time protocol specifically designed for functional evaluation of image data in steady-state contrast. With a reconstruction based on the separate calculation of low-rank and sparse part, it employs a softer constraint than the strict exponential model, which was possible due to sufficient temporal sampling density via spiral acquisition. The low-rank plus sparse reconstruction is fit for the use on dynamic and on inversion recovery data. Thus, motion correction is rendered unnecessary with it. IR-CRISPI was equipped with noise suppression via spatial wavelet filtering. A study comprising 10 patients with cardiac disease show medical applicability. A comparison with performed traditional reference exams offer insight into diagnostic benefits. Especially regarding patients with difficulty to hold their breath, the real-time manner of the IR-CRISPI acquisition provides a valuable alternative and an increase in robustness. In conclusion, especially with IR-CRISPI in free breathing, a major acceleration of the cardiovascular MR exam could be realized. In an acquisition of less than 100 s, it not only includes the information of two traditional protocols (cine and LGE), which take up more than 9.6 min, but also allows adjustment of TI in retrospect and yields lower artifact level with similar image quality. N2 - Diese Arbeit behandelt die Beschleunigung der kardiovaskulären MRT zum Erfassen funktioneller Information bei Steady-State-Kontrast und zur Unter- suchung der Vitalität bei Wiederherstellung der Magnetisierung nach ihrer Inversion. Zwei Ansätze werden eingeführt und im Detail diskutiert: MOCO- MAP, welches ein exponentielles Modell nutzt, um dynamische Daten zu rekonstruieren, und IR-CRISPI, welches mit seinem “low-rank plus sparse"- Algorithmus mit Compressed Sensing verwandt ist. MOCO-MAP ist der Nachfolger der modellbasierten Beschleunigung des Parameter-Mappings (MAP) für die Anwendung im Bereich des Myokards. Hierzu wurde es mit einer Bewegungskorrektur (MOCO) versehen, um expo- nentielles Fitten eines unbewegten Objects in Zeitrichtung zu ermöglichen. Das Einbringen dieses physikalischen Vorwissens zusammen mit dem Erzwin- gen von Konsistenz mit den Messdaten wird dazu genutzt, iterativ eine Bildfolge aus Daten einer deutlich kürzeren Messung als herkömmlich zu rekonstruieren (< 3 s gegenüber ca. 10 s). Die Ergebnisse zeigen die Umsetz- barkeit der Methode sowie die Nachweisbarkeit von Delayed Enhancements im Myokard, aber deutliche funktionelle Abweichungen und Artefakte bereits aufgrund von kleinen Ungenauigkeiten der Bewegungskorrektur. IR-CRISPI geht aus CRISPI hervor, welches zur Auswertung von funk- tionellen Echtzeitdaten bei konstantem Kontrast dient. Mit der Rekon- struktion durch getrennte Berechnung von niedrigrangigem und dünnbe- setztem Matrixanteil wird hier bei der Datenrekonstruktion weniger stark eingeschränkt als bei einem strikten exponentiellen Modell. Die pirale Auf- nahmeweise erlaubt hierzu ausreichend effiziente k-Raumabdeckung. Die “low-rank plus sparse"-Rekonstruktion ist kompatibel mit dynamischen und mit Inversion-Recovery-Daten. Eine Bewegungskorrektur ist folglich nicht nötig. IR-CRISPI wurde mit einer Rauschunterdrückung durch räumliche Wavelet- Filterung versehen. Eine Studie, die 10 Patienten einschließt, zeigt die Eignung für die medizinische Anwendung. Der Vergleich mit herkömm- lichen Aufnahmetechniken lässt auf den gewonnenen diagnostischen Nutzen schließen. Besonders für Patienten, die Schwierigkeiten mit dem Luftanhal- ten haben, eröffnet diese Echtzeitaufnahmemethode eine wertvolle Alterna- tive und erhöhte Stabilität. Am Ende konnte gerade mittels IR-CRISPI eine bemerkenswerte Beschleu- nigung der kardiovaskulären MR-Untersuchung verwirklicht werden. Trotz der kurzen Aufnahmezeit von weniger als 100 s für den kompletten linken Ven- trikel schließt es nicht nur die Information zweier herkömmlicher Protokolle mit ein (Cine und LGE), die zusammen mehr als 9,6 min dauern, sondern es erlaubt zusätzlich auch das Einstellen der TI-Zeit im Nachhinein und liefert Ergebnisse mit geringerem Artefaktlevel bei ähnlicher Bildqualität KW - Kernspintomografie KW - Herzfunktion KW - Herzmuskel KW - Bildgebendes Verfahren KW - Echtzeit KW - cine loop KW - late enhancement KW - late gadolinium-enhancement KW - magnetic resonance imaging KW - real-time imaging KW - spiral trajectory Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-302822 ER - TY - THES A1 - Grebe, Sören T1 - Diagnose der linksventrikulären Hypertrophie bei Hämodialyse-PatientInnen anhand von Echokardiographie und EKG im Vergleich zum CMRI T1 - Diagnosis and quantification of left ventricular mass by ecg and echocardiography compared to cardiac magnet resonance imaging in hemodialysis patients N2 - In der Gruppe der Hämodialyse-PatientInnen besteht ein deutlich erhöhtes Risiko an kardiovaskulären Ereignissen zu versterben. Korrespondierend hierzu weisen Hämodia-lyse-PatientInnen eine erhöhte Prävalenz an linksventrikulärer Hypertrophie (LVH) auf. Diese gilt als starker unabhängiger Risikofaktor für kardiovaskuläre Mortalität. Auf-grund der prognostischen Aussagekraft dient die Bewertung des linksventrikulären Massenindex (LVMI) sowie die Diagnose einer LVH vor allem in prospektiven Studien als ein bedeutendes Werkzeug zur Beurteilung des kardiovaskulären Risikos. Die Be-stimmung der LVH kann anhand von bildgebenden Verfahren (u.a. Echokardiographie, CMRI) oder dem EKG erfolgen. Die CMRI-Messung wird gegenwärtig als Goldstan-dard zur Messung der LVH betrachtet. Die 2D geführte M-mode-Methode der Echokardiographie zur Bestimmung der LVM zeichnet sich durch seine einfache und schnelle Durchführbarkeit aus und wird deshalb trotz präziserer Messverfahren wie dem 3D-Verfahren sowie diverser Einschränkungen weiterhin von der American Society of Echocardiography (ASE) als Screening-Methode und zur Untersuchung großer PatientInnenpopulationen empfohlen. Die empfohlene ASE-Formel überschätzt jedoch den LVMI nachweislich im Vergleich zum CMRI-Messverfahren. Die Überschätzung zeigte sich abhängig von der Höhe des LVMI. Es wird vermutet, dass die zunehmende Überschätzung Folge der geometrischen Grundan-nahmen ist, welche den LV vereinfachend als Ellipsoid mit konstantem L/D-Verhältnis annimmt. Dieses Verhältnis scheint sich jedoch bei zunehmender Herzgröße zu verän-dern, was wiederum zu einer Fehleinschätzung des LVMI führt. Die Teichholz (Th)-Formel korrigiert das L/D-Verhältnis mithilfe einer kurvilinearen Anpassung an den linksventrikulären Durchmesser und zeigte kürzlich in einer PatientInnengruppe mit Aor-tenstenose die geringste Tendenz der Überschätzung bei PatientInnen mit LVH. In der vorliegenden Studie wurden die echokardiographischen Formeln – ASE und Th – mit dem CMRI-Messverfahren verglichen. Beide Formeln zeigten eine deutliche Überschät-zung des LVMI. Die Th-Formel demonstrierte jedoch neben einer besseren Überein-stimmung zum CMRI, eine insgesamt geringere Überschätzung des LVMI sowie eine sukzessive Abnahme der Überschätzung mit zunehmendem LVMI. Zusammenfassend kann festgehalten werden, dass die Th-Formel der ASE-Formel in Bezug auf die Be-rechnung des LVMI bei Hämodialyse-PatientInnen insbesondere bei PatientInnen mit LVH überlegen ist. Weitere Studien sind jedoch erforderlich, um die Th-Formel in grö-ßeren Hämodialyse-PatientInnen-Kohorten mit höheren LVMI-Werten zu testen sowie um den prognostischen Wert der Th-Formel im Vergleich zur ASE-Formel zu ermitteln. Die klassischen EKG-Indices und -Scores zur Feststellung einer LVH wiesen, wie be-reits in anderen CMRI-Vergleichsstudien gezeigt, eine schlechte Sensitivität bei guter Spezifität auf. Aufgrund dessen verlor das EKG zunehmend an Bedeutung als Scree-ning-Untersuchung. In dieser Studie wurde der Versuch unternommen die Sensitivität durch zwei Lösungsansätze zu verbessern, einerseits durch die Kombination verschiede-ner EKG-Kriterien und andrerseits durch eine Adjustierung der EKG-Kriterien an den mittels Bioimpedanz gemessenen Fettmassenanteil. Die Kombination verschiedener EKG-Kriterien erzielte eine deutlich erhöhte Sensitivität von >70 %. Auch die Anpas-sung der EKG-Kriterien an den individuellen Fettmassenanteil könnte ein hilfreicher Lösungsansatz zur Verbesserung der Sensitivität bei Adipositas darstellen. N2 - Left ventricular hypertrophy (LVH) is highly prevalent in patients on hemodialysis. LVH, as measured by the left ventricular mass index (LVMI), is a strong predictor of cardiovascular disease (CVD). Consequently, a reliable and valid method to detect LVH is needed for both clinical and scientific implications. For the assessment of left ventricular mass (LVM), cardiac magnetic resonance imaging (CMR) has been established as the most accurate and reproducible method. However, given its limited availability and high cost, CMR is not practical for clinical use in large-scale clinical studies. In contrast, the two-dimensional (2D) targeted M-mode transthoracic echocardiography (TTE) and ecg is preferred in the clinical context because of its widespread availability, low cost, simple handling, and extensive evidence base. Nevertheless, echocardiographic linear measurement and LVM calculation by cube function formulas have their own limitations. The current recommended formula from the American Society of Echocardiography (ASE) is based on special geometric assumptions, which may become inaccurate in the presence of asymmetric hypertrophy, eccentric remodeling, or distortion of left ventricular (LV) geometry and may lead to an incremental overestimation of LVMI. Teichholz et al. designed a formula that includes a volume-correcting function in order to minimize the error inter alia in patients with LVH. A recent CMR study investigating patients with aortic stenosis demonstrated that the Teichholz (Th) formula had a lower tendency to overestimate the value within a population with increased LVMI. Here, we investigated the performance of two echocardiographic formulas, ASE and Th, in calculating LVMI in patients on hemodialysis. TTE and CMR data for 95 hemodialysis patients who participated in the MiREnDa trial were analyzed. The LVMI was calculated by two-dimensional (2D) TTE-guided M-mode measurements employing the American Societ y of Echocardiography (ASE) and Teichholz (Th) formulas, which were compared to the reference method, CMR. LVH was present in 44% of patients based on LVMI measured by CMR. LVMI measured by echocardiography correlated moderately with CMR, ASE: r = 0.44 (0.34–0.62); Th: r = 0.44 (0.32–0.62). Compared to CMR, both echocardiographic formulas overestimated LVMI (mean ΔLVMI (ASE-CMR): 19.5 ± 19.48 g/m2, p < 0.001; mean ΔLVMI (Th-CMR): 15.9 ± 15.89 g/m2, p < 0.001). We found greater LVMI overestimation in patients with LVH using the ASE formula compared to the Th formula. Stratification of patients into CMR LVMI quartiles showed a continuous decrease in ΔLVMI with increasing CMR LVMI quartiles for the Th formula (p < 0.001) but not for the ASE formula (p = 0.772). Bland-Altman analysis showed that the Th formula had a constant bias independent of LVMI. Both methods had good discrimination ability for the detection of LVH (ROC-AUC: 0.819 (0.737–0.901) and 0.808 (0.723–0.892) for Th and ASE, respectively). The ASE and Th formulas overestimate LVMI in hemodialysis patients. However, the overestimation is less with the Th formula, particularly with increasing LVMI. The results suggest that the Th formula should be preferred for measurement of LVMI in chronic hemodialysis patients. Shown in other CMRI comparative studies the ECG indices and scores for detecting LVH had poor sensitivity with good specificity. Because of this, ECG became increasingly less important as a screening method. To improve the sensitivity by two approaches, on the one hand by combining different ECG criteria and on the other hand by adjusting the ECG criteria to the fat mass measured by bioimpedance. The combination of different ECG criteria achieved a significantly increased sensitivity of >70 %. The adjustment of the ECG criteria to the individual fat mass could also be a helpful approach to improve the sensitivity in obesity. KW - Transthorakale Echokardiographie KW - Elektrokardiogramm KW - Linke Herzkammer KW - Kernspintomografie KW - Bioimpedanz KW - Echokardiographie KW - EKG KW - MRT KW - Bioimpedanz KW - Linksventrikuläre Hypertrophie KW - Linksventrikuläre Masse KW - Kombination EKG-Kriterien KW - Teicholz-Formel KW - Fettmassenanteil Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-272115 ER - TY - THES A1 - Eirich, Philipp T1 - Accelerated non-Cartesian cardiovascular MR Imaging at 3T and 7T T1 - Beschleunigte nicht-kartesische MRT Herzbildgebung bei 3T und 7T N2 - In this work, accelerated non-Cartesian Magnetic Resonance Imaging (MRI) methods were established and applied to cardiovascular imaging (CMR) at different magnetic field strengths (3T and 7T). To enable rapid data acquisition, highly efficient spiral k-space trajectories were created. In addition, hybrid sampling patterns such as the twisting radial lines (TWIRL) k-space trajectory were studied. Imperfections of the dynamic gradient system of a MR scanner result in k-space sampling errors. Ultimately, these errors can lead to image artifacts in non-Cartesian acquisitions. Among other reasons such as an increased reconstruction complexity, they cause the lack of spiral sequences in clinical routine compared to standard Cartesian imaging. Therefore, the Gradient System Transfer Functions (GSTFs) of both scanners were determined and used for k-space trajectory correction in post-correction as well as in terms of a pre-emphasis. The GSTF pre-emphasis was implemented as a fully automatic procedure, which enabled a precise correction of arbitrary gradient waveforms for double-oblique slice orientations. Consequently, artifacts due to trajectory errors could be mitigated, which resulted in high image quality in non-Cartesian MRI. Additionally, the GSTF correction was validated by measuring pre-emphasized spiral gradient outputs, which showed high agreement with the theoretical gradient waveforms. Furthermore, it could be demonstrated that the performance of the GSTF correction is superior to a simple delay compensation approach. The developed pulse sequences were applied to gated as well as real-time CMR. Special focus lied on the implementation of a spiral imaging protocol to resolve the beating heart of animals and humans in real time and free breathing. In order to achieve real-time CMR with high spatiotemporal resolution, k-space undersampling was performed. For this reason, efficient sampling strategies were developed with the aim to facilitate compressed sensing (CS) during image reconstruction. The applied CS approach successfully removed aliasing artifacts and yielded high-resolution cardiac image series. Image reconstruction was performed offline in all cases such that the images were not available immediately after acquisition at the scanner. Spiral real-time CMR could be performed in free breathing, which led to an acquisition time of less than 1 minute for a whole short-axis stack. At 3T, the results were compared to the gold standard of electrocardiogram-gated Cartesian CMR in breath hold, which revealed similar values for important cardiovascular functional and volumetric parameters. This paves the way to an application of the developed framework in clinical routine of CMR. In addition, the spiral real-time protocol was transferred to swallowing and speech imaging at 3T, and first images were presented. The results were of high quality and confirm the straightforward utilization of the spiral sequence in other fields of MRI. In general, the GSTF correction yielded high-quality images at both field strengths, 3T and 7T. Off-resonance related blurring was mitigated by applying non-Cartesian readout gradients of short duration. At 7T, however, B1-inhomogeneity led to image artifacts in some cases. All in all, this work demonstrated great advances in accelerating the MRI process by combining efficient, undersampled non-Cartesian k-space coverage with CS reconstruction. Trajectory correction using the GSTF can be implemented at any scanner model and enables non-Cartesian imaging with high image quality. Especially MRI of dynamic processes greatly benefits from the presented rapid imaging approaches. N2 - In der vorliegenden Arbeit wurden Methoden der beschleunigten Magnetresonanztomographie (MRT) etabliert, welche auf nicht-kartesischer Datenaufnahme beruhen. Diese wurden insbesondere in der Herzbildgebung bei verschiedenen Magnetfeldstärken (3T und 7T) angewendet. Der Fokus lag auf der Entwicklung von hocheffizienten spiralförmigen k-Raum Trajektorien, mit dem Zweck sehr kurze Aufnahmezeiten zu ermöglichen. Zusätzlich wurde eine hybride k-Raum Trajektorie untersucht, die sogenannte "twisting radial lines (TWIRL)" k-Raum Trajektorie. Ungenauigkeiten des dynamischen Gradientensystems eines MRT Scanners resultieren in fehlerbehafteter k-Raum Abtastung während der Datenaufnahme. In der nicht-kartesischen Bildgebung kann dies letztendlich zu Artefakten im rekonstruierten Bild führen. Zusammen mit anderen Hemmnissen, wie beispielsweise einer komplexeren Bildrekonstruktion, sind sie verantwortlich dafür, dass noch immer mehrheitlich kartesische Bildgebungssequenzen in der klinischen Routine durchgeführt werden. Aus diesem Grund wurden die Übertragungsfunktionen der Gradientensysteme der verwendeten MRT Scanner (eng. "Gradient System Transfer Function (GSTF)") bestimmt und für k-Raum Trajektorienkorrekturen verwendet. Diese Korrektur wurde sowohl in der Bildrekonstruktion nach bereits erfolgter Datenaufnahme angewendet als auch im Rahmen einer Vorverstärkung bevor die Gradienten ausgespielt werden. Diese Vorverstärkung wurde als vollständig automatisierter Prozess implementiert und ermöglichte eine präzise Korrektur beliebig gewählter Gradientenfunktionen aller Schichtorientierungen. Auf diesem Wege konnten die durch Trajektorienfehler verursachten Bildartefakte kompensiert werden, was zu hoher Bildqualität in der nicht-kartesischen MRT Bildgebung führte. Des Weiteren wurde die Gradientenkorrektur durch Messungen der tatsächlich ausgespielten Gradientenformen validiert. Diese wiesen eine hohe Übereinstimmung mit den theoretisch zu erwarteten Gradientenformen auf. Darüber hinaus wurde gezeigt, dass die auf der Impulsantwort basierende, umfassende Gradientenkorrektur eine höhere Bildqualität ermöglicht als eine einfache Korrektur mittels globaler Zeitverschiebungen. Die entwickelten MRT Sequenzen wurden sowohl in der segmentierten als auch in der Echtzeit-Herzbildgebung angewendet. Im Speziellen lag der Fokus auf der Implementierung eines Protokolls für die spirale MRT Bildgebung, welche das schlagende Herz von Tieren und Menschen in Echtzeit und freier Atmung auflösen kann. Um Echtzeit-Herzbildgebung mit hoher räumlicher und zeitlicher Auflösung zu vereinen, wurde der k-Raum unterabgetastet. In diesem Zusammenhang wurden Strategien zur effizienten und komprimierten Datenaufnahme entwickelt, unter Anwendung der Modell-basierten "Compressed Sensing" (CS)-Technik. Diese Methode reduziert Aliasing-Artefakte in der Bildrekonstruktion von unterabgetasteten Daten und ermöglicht deshalb hochaufgelöste, dynamische Echtzeit-Bilderserien des schlagenden Herzens. Allerdings wurden die gemessenen Daten stets extern rekonstruiert, sodass die Bilder nicht unmittelbar nach der Aufnahme am MRT Scanner verfügbar waren. Die spirale Echtzeit-Herzbildgebung konnte in freier Atmung durchgeführt werden, was eine Messzeit aller Schichten in der kurzen Herzachse in unter 1 Minute ermöglichte. Bei 3T wurden die Ergebnisse mit dem Goldstandard der mittels eines Elektrokardiogramms segmentierten kartesischen Herzbildgebung im Atemstopp verglichen und es konnte gezeigt werden, dass wichtige funktionelle und volumetrische Herzparameter übereinstimmen. Dies ebnet den Weg zur Anwendung des entwickelten Protokolls in der klinischen Routine der Herzbildgebung am MRT. Darüber hinaus wurde das Protokoll in der Echtzeit-Bildgebung von Schlucken und Sprechen bei 3T getestet. Die Ergebnisse waren ebenfalls von hoher Qualität und bestätigen den unkomplizierten Transfer der spiralen Sequenz in andere Bereiche der MRT Bildgebung. Insgesamt lieferte die GSTF-Korrektur Bilder von hoher Qualität bei beiden Feldstärken, 3T und 7T. Eine durch off-Resonanz verursachte Bildunschärfe wurde durch kurze Auslesezeiten der nicht-kartesischen Gradienten abgeschwächt. Allerdings führte B1-Inhomogenität in manchen Fällen zu Bildartefakten bei 7T. Die vorliegende Arbeit stellt einen wesentlichen Beitrag zur Beschleunigung des MRT Bildgebungsprozesses dar, indem effiziente, unterabgetastete nicht-kartesische k-Raum Trajektorien mit der CS-Rekonstruktionstechnik kombiniert wurden. Trajektorien-Korrektur basierend auf der GSTF kann prinzipiell an jedem MRT Scanner implementiert werden und legt den Grundstein für nicht-kartesische Bildgebung mit hoher Bildqualität. Insbesondere die Bildgebung von dynamischen Prozessen profitiert von den hier vorgestellten beschleunigten Methoden zur Datenaufnahme. KW - Kernspintomografie KW - Bildgebendes Verfahren KW - Spirale KW - Artefakt KW - Übertragungsfunktion KW - MRT KW - MRI KW - Herzbildgebung KW - Cardiac imaging KW - Beschleunigte Bildgebung KW - Accelerated imaging KW - Gradient System Transfer Function KW - Echtzeitbildgebung KW - Real-time imaging KW - Compressed sensing Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-253974 ER - TY - THES A1 - Scheler, Maximilian Heinrich Julius T1 - Die operative Versorgung des Thoraxmagens - Eine Langzeitanalyse von 2008-2015 T1 - Surgical treatment of giant hiatal hernias - a long-term analysis from 2008-2015 N2 - Der Thoraxmagen beschreibt eine zirkuläre Schwachstelle der phrenikoösophagealen Membran mit einer schrittweisen Dislozierung der Magenkardia und des Ösophagus nach mediastinal. Die Therapie des Thoraxmagens kann konservativ im Sinne des „watchful waiting“ oder operativ erfolgen. Aufgrund der möglichen Komplikationen wird die elektive Operation durch die amerikanischen Leitlinien empfohlen. Ein zentrales Problem der Hiatushernienchirurgie stellt die hohe Anzahl an Rezidiven dar. Ob die Gründe hierfür in der Zwerchfellrekonstruktion, Speiseröhrenlänge, Fundoplicatio oder Netzaugmentation liegen, wird nach wie vor kontrovers diskutiert. In dieser Arbeit wurde die operative Versorgung des Thoraxmagens von 124 Patienten des Universitätsklinikums Würzburg im Zeitraum von September 2008 bis Juni 2015 untersucht. Hierfür war neben den perioperativen Daten auch die Rezidiv- und Letalitätsrate von Relevanz. Das Patientenkollektiv wurde sowohl in Hinblick auf das Lebensalter als auch auf die verschiedenen Versorgungsarten analysiert. Um die postoperative Lebensqualität zu beurteilen, erfolgte die Patientenbefragung mit Hilfe eines Symptomfragebogens und dem Gastrointestinalen Lebensqualitätsindex nach Eypasch (GIQLI). Zusätzlich wurden 17 Patienten postoperativ mittels MRT untersucht, um eine optimierte MRT-Sequenz zur Beurteilung der Hiatusregion zu evaluieren. Im Vergleich der Altersgruppen zeigte sich trotz einer erhöhten Komorbiditätsrate bei dem Patientenkollektiv ≥ 75 Jahre (p=0,002) kein signifikanter Unterschied bei Betrachtung der intraoperativen Komplikationen. Die Rezidivrate lag unabhängig vom Alter bei 20,2% im Untersuchungszeitraum, jedoch konnte eine verminderte Rezidivrate bei Patienten mit U-Shape Versorgung (p=0,015) festgestellt werden. In der postoperativen Patientenbefragung zeigten sich 87,0% der Patienten, unabhängig vom Alter und der Versorgungsart, zufrieden mit dem Operationsergebnis und beschrieben ihren Zustand im Vergleich zu präoperativ als gebessert. Die Ergebnisse des GIQLI erbrachten in dem untersuchten Patientenkollektiv ein gegenüber der Allgemeinbevölkerung erniedrigten Wert mit 95,4 Punkten. Die optimierte MRT-Sequenz zeichnete sich durch eine hohe diagnostische Konfidenz bei guter Bildqualität, kurzer Untersuchungsdauer und gleichzeitig hoher Akzeptanz der Patienten gegenüber dieser Art der Diagnostik aus. Zusammenfassend stellt die operative Versorgung von Thoraxmägen, unabhängig des Patientenalters, eine sichere Therapieform dar, die zu einer hohen Patientenzufriedenheit führt. Die modifizierte MRT-Untersuchung hat sich als diagnostische Methode bewährt und stellt eine Alternative zu strahlenexponierenden oder von Seiten der Patienten weniger gut tolerierten Untersuchungsmodalitäten dar. N2 - Giant hiatal hernias occur due to a circular weak point of the phrenic esophageal membrane with a gradual dislocation of the gastric cardia and the esophagus in mediastinal direction. Therapy of these giant hiatal hernias can be approached conservative, in the sense of “watchful waiting”, or surgically. Due to the possible complications however, the American guidelines highly recommend elective surgery. A central problem in hiatal hernia surgery is the high number of recurrences. Whether this is due to the art of reconstruction of the diaphragm, the length of the esophagus, the fundoplication or the mesh augmentation is still a matter of controversy. In this thesis, the surgical treatments of 124 patients with giant hiatal hernias were examined at the University Hospital Würzburg from September 2008 to June 2015. In addition to the perioperative data, the recurrence- and mortality rates were also taken into consideration. The patient collective was analyzed in regard to age, as well as the various types of care. In order to assess the postoperative quality of life, the patient survey was conducted with the help of a symptom questionnaire and the gastrointestinal quality of life index according to Eypasch (GIQLI). Furthermore, to evaluate an optimized MRI sequence for assessing the hiatus region, 17 patients were examined postoperatively using MRI. When comparing the different groups of age it appeared, that even with an increased comorbidity rate in the age ≥ 75 (p=0,002), there was no significant difference in intraoperative complications. Regardless of age, the recurrence rate was 20.2% in the study period. However, a reduced recurrence rate was found in patients with U-shape restoration (p=0.015). In the postoperative patient survey, 87.0% of the patients, regardless of age and type of care, stated to be satisfied with the result of the operation and also described their condition as improved in comparison to their preoperatively status. The results of the GIQLI showed a lower value compared to the general population, with 95.4 points in the examined patient group. The optimized MRT sequence was characterized by a high diagnostic confidence with a good image quality, short examination times and high patient acceptance for this diagnostic method. In summary, the surgical treatment of giant hiatal hernias is a safe type of therapy that leads to a high level of patient satisfaction, regardless of the patients age. The modified MRI examination was shown to be a valid diagnostic method and represents an alternative to other examination modalities that lead to radiation exposition or tend to be less tolerated by the patient. KW - Zwerchfellkrankheit KW - Zwerchfellbruch KW - Laparoskopie KW - Kernspintomografie KW - Thoraxmagen KW - Gastrointestinaler Lebensqualitätsindex nach Eypasch (GIQLI) Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-253018 ER - TY - THES A1 - Geier, Bettina T1 - Kernspintomografische Natriumbildgebung in Haut und Muskel T1 - Magnetic resonance imaging of sodium in skin and muscle N2 - Die vorliegende Arbeit untersucht den Natriumgehalt verschiedener Kompartimente des Körpers mittels Magnetresonanztomographie (= MRT). Die Korrelation zwischen erhöhtem Salzkonsum und arterieller Hypertonie ist bereits umfangreich analysiert worden. Für das Verständnis der pathophysiologischen Zustände und deren Regulation, ist eine Quantifizierung von Natriumkonzentrationen in verschiedenen Gewebearten bedeutsam. Die exakte Messung von Natriumkonzentrationen im menschlichen Gewebe ist derzeit experimentell. Im Rahmen der hier vorgelegten Arbeit wurden die Natriumkonzentrationen von Haut und Skelettmuskel mittels 23Na Magnetresonanztomographie (= 23 Na MRT) im menschlichen Körper quantifiziert. Natriummessungen wurden bei Patienten mit primärem Hyperaldosteronismus (= PHA), bei Patienten mit essentieller Hypertonie (= EH), sowie einer gesunden Kontrollgruppe vorgenommen. Die Ergebnisse zeigten, dass Haut und Skelettmuskel Speicherorgane für Natrium im menschlichen Körper darstellen. Durch gezielte Therapie waren die Natriumkonzentrationen in beiden Speicherorganen modulierbar N2 - The present work investigates the sodium content of different compartments of the body by means of magnetic resonance imaging (= MRI). The correlation between increased salt consumption and arterial hypertension has already been extensively analyzed. For the understanding of pathophysiological states and their regulation, quantification of sodium concentrations in different tissue types is significant. Accurate measurement of sodium concentrations in human tissues is currently experimental. In the work presented here, sodium concentrations of skin and skeletal muscle were quantified using 23Na - magnetic resonance imaging (= 23 Na-MRI) in the human body. Sodium measurements were made in patients with primary hyperaldosteronism (= PHA), in patients with essential hypertension (= EH), and in a healthy control group. The results showed that skin and skeletal muscle are storage organs for sodium in the human body. Sodium concentrations in both storage organs could be modulated by targeted therapy. KW - Natrium-23 KW - Kernspintomografie KW - Haut KW - Muskel KW - MRI KW - sodium-23 KW - Skin KW - Muscle Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-249429 ER - TY - THES A1 - Riedl, Katharina Alina T1 - Optimierung und Validierung einer MRT-Messmethode zur Quantifizierung der Wandschubspannung in arteriellen Gefäßen zur Evaluation der atherosklerotischen Plaqueentwicklung T1 - Optimization and validation of an MR imaging method to quantify wall shear stress in arterial vessels to evaluate the atherosclerotic plaque development N2 - Herz-Kreislauf-Erkrankungen stellen weiterhin die Todesursache Nummer eins in Deutschland dar, welche hauptsächlich durch Atherosklerose verursacht werden. Als ein Prädiktor der atherosklerotischen Plaqueentwicklung wird die Wandschubspannung (WSS) diskutiert. Ziel dieser Arbeit war es daher, anhand von Flussphantomen und C57bl/6 Mäusen eine 2D-Gradientenecho-Bildgebungsmethode mit einer 3D-Phasenkontrast-Flusskodierung zu optimieren und anschließend eine longitudinale Kleintierstudie mit ApoE-/- Mäusen durchzuführen, um den Zusammenhang der WSS und der atherosklerotischen Plaqueentwicklung näher zu untersuchen. Zunächst wurden Flussphantome mit einem Schlauchdurchmesser von 4 mm und 1 mm zur Optimierung der Messmethode verwendet. Anschließend wurde die Messmethode weiter angepasst, um in vivo Messungen an C57bl/6 Mäusen durchführen zu können. Nach erfolgter Optimierung wurde eine longitudinale Kleintierstudie mit zwei verschiedenen Diäten, Western Diät und Chow Diät, durchgeführt. Im Rahmen der Studie erfolgten nach einer, acht und zwölf Wochen MR-Messungen sowie histologische Analysen. Es konnte gezeigt werden, dass die Wandschubspannung in Mäusen bei 17,6 Tesla quantifiziert werden kann. Es zeigte sich eine Tendenz, dass Plaqueformationen mit einer höheren Wandschubspannung einhergehen. N2 - Cardiovascular diseases are the number one of death in Germany, which are mainly caused by atherosclerosis. Wall shear stress (WSS) is discussed as a predictor for atherosclerotic plaque development. The aim of this study was the optimization and validation of a 2D gradient echo imaging method with a 3D phase contrast flow encoding using flow phantoms and C57bl/6 mice and to perform a longitudinal study with ApoE-/- mice to examine the context of WSS and atherosclerotic plaque development. First, flow phantoms with a tube diameter of 4 mm and 1 mm were used to optimize the imaging method. Therefore, the imaging method was further adapted to examine in vivo C57bl/6 mice. After optimization, a longitudinal animal study was performed with two different diets, Western diet and Chow diet. MR measurements and histological analyses were performed after one, eight, and twelve weeks. It could be shown that WSS can be quantified in mice using 17.6 Tesla MRI. There was a tendency towards higher values of wall shear stress in vessels affected by plaque formations. KW - Wandschubspannung KW - Kernspintomografie KW - Maus KW - wall shear stress KW - Ultrahochfeldmagnetresonanztomographie KW - ultra highfield magnetic resonance imaging Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-230814 ER - TY - THES A1 - Kunz, Julian T1 - Wertigkeit der Absoluten Quantitativen Perfusionsauswertung nach Myokardinfarkt im Akutstadium und im Langzeitverlauf mittels Magnetresonanztomographie T1 - Validity of perfusion measurement using MRI in the acute stage of myocardial infarction and upon one year follow up N2 - Es wurden Perfusionsmessungen mittels MRT an Infarktpatienten im Akutstadium und im Langzeitverlauf durchgeführt und quantitativ mittels einem Sektormodell ausgewertet. Hierbei zeigte sich, dass sich die Perfusionswerte im Infarktareal und gesunden Myokard nicht signifikant unterschieden und dass sich diese auch im Jahresverlauf nicht signifikant änderten. Es ergab sich auch keine signifikante Korrelation zwischen der Größe des Infarkareales und den gemessenen Perfusionswerten. N2 - Quantitative cardiac perfusion was measured in the acute stage of myocardial infarction by means of MRI, as well as upon one year follow-up. Using a sector based model, measured perfusion values did not differ significantly between infarcted and non-infarcted tissue. Also, a significant change between the perfusion value upon one year follow-up could not be shown. Furthermore, no significant correlation between perfusion values and the size of the infarcted area could be shown. KW - Kernspintomografie KW - Funktionelle Kernspintomografie KW - Herzinfarkt KW - Cardioperfusion Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-220952 ER - TY - THES A1 - Gasparyan, Artur T1 - Quantifizierung pulmonaler Blutflussgeschwindigkeit durch SENCEFUL Magnetresonanztomographie mit bewegter Schichtselektion T1 - Quantification of pulmonary blood flow velocity through SENCEFUL magnetic resonance imaging using moving slice selection N2 - Patienten mit chronischen Lungenerkrankungen leiden unter schwerwiegender Symptomatik und bedürfen regelmäßiger Verlaufskontrollen der Therapie. Dabei sollte zum Schutz der Patienten sowohl auf kanzerogene, ionisierende Strahlung verzichtet als auch der Einsatz potenziell nebenwirkungsreicher Kontrastmittel vermieden werden. Die pulmonale Blutflussgeschwindigkeit im Parenchym stellt einen quantitativen, bildgebenden Biomarker dar, mit dessen Hilfe die Dynamik des Krankheitsgeschehens untersucht werden kann. In dieser Arbeit wurde eine neue Auswertungsmethode vorgestellt, die mit Hilfe kontrastmittelfreier Magnetresonanztomographie die Blutflussgeschwindigkeit im Lungenparenchym quantifizieren kann. Die auf diese Weise bestimmten Ergebnisse entsprechen den Angaben zur Lungenperfusion, wie sie in der Literatur zu finden sind. N2 - Patients with chronic respiratory diseases suffer from severe symptoms and require regular follow ups during treatment. It is important to avoid the use of cancerogenic ionising radiation as well as potentially harmful contrast agents. The pulmonary blood flow velocity within the parenchyma can serve as a quantitative imaging biomarker, which can help analyse the course of the disease. In this work a new method for the quantification of pulmonary blood flow velocity is shown. The results correspond to physiological values in the human lung. KW - Kernspintomografie KW - Lunge KW - Blutflussgeschwindigkeit KW - Quantitativ Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-215693 ER - TY - THES A1 - Ofenhitzer, Kathrin T1 - Stellenwert der präoperativen radiologischen Diagnostik vor Cochlea-Implantat-Operation T1 - The role of preoperative radiological diagnostics before cochlear implantation N2 - Bei Patienten mit hochgradigem sensorineuralem Hörverlust, der nicht mehr ausreichend mittels Hörgeräten kompensiert werden kann, hat sich die CI-Versorgung zur Verbesserung der Hörfähigkeit als Standardmethode etabliert und ihr Indikationsbereich wächst. Zur Beurteilung des auditorischen Systems sowie seiner angrenzenden Nachbarstrukturen zählt die neuroradiologische Diagnostik mittels spezieller CT- und MRT-Techniken zu den entscheidenden Kriterien zur CI-Indikationsstellung und der Operationsplanung. In der Literatur finden sich verschiedenste Meinungen zur Bedeutung dieser präoperativen Bildgebungen vor Cochlea-Implantationen und die Wahl einer CT- und bzw. oder einer MRT-Untersuchung zur präoperativen Diagnostik wird diskutiert. So wurden am Universitätsklinikum Würzburg im Zeitraum von 2005 bis zum Jahr 2015 zur Diagnostik vor der CI-Operation eines Patienten meist sowohl eine CT-Aufnahme des Felsenbeins als auch eine MRT-Aufnahme des Schädels angefertigt. Im Rahmen dieser Arbeit sollte untersucht werden, inwieweit diese beiden präoperativen CT- und MRT-Untersuchungen für die CI-Operation und die Ursachenabklärung der Hörminderung von Relevanz waren und ob eine Empfehlung für die zukünftige Wahl der Diagnostikmethoden abgegeben werden kann. Dazu wurden Anamnesen, Voruntersuchungen, präoperative CT- und MRT-Bildgebungen und die Operationsberichte von 705 CI-Fällen bei 549 Patienten, die in diesem Zeitraum operiert wurden, retrospektiv ausgewertet. Die Darstellbarkeiten verschiedener Pathologien, die mit Hilfe der beiden Bildgebungsverfahren detektiert wurden, wurden miteinander verglichen und es wurde überprüft, ob ein Verfahren möglicherweise besser zur Diagnostik bestimmter Auffälligkeiten geeignet war als das andere. Durch die Auswertung der Operationsberichte wurde evaluiert, ob und wie die Befunde der präoperativen Bildgebungen mit den intraoperativen Situationen übereinstimmten und ob diese Befunde die Operationsplanung beeinflusst hatten. Die Ergebnisse dieser Arbeit zeigten, dass die bildgebenden Verfahren zur Ursachenfindung einer nicht abgeklärten Hörminderung beitragen können. Des Weiteren konnte gefolgert werden, dass ein Großteil von Auffälligkeiten im Bereich des Felsenbeins, die mit sensorineuralen Schwerhörigkeiten assoziiert sind, mit hoher Wahrscheinlichkeit sowohl mittels CT- als auch mittels MRT-Untersuchung diagnostiziert werden kann. Für einzelne Läsionen konnte vermutet werden, dass diese nur durch die CT und umgekehrt bestimmte Läsionen nur durch die MRT detektiert werden können. Außerdem zeigte sich, dass die beiden bildgebenden Verfahren auf bestimmte Schwierigkeiten, die während einer CI-Operation auftreten können, präoperativ hinweisen können. Zusätzlich konnte festgestellt werden, dass die Wahl der Implantations-Seite bei einseitig operierten aber beidseits schwerhörigen Patienten neben den CT- und MRT-Untersuchungen in den meisten Fällen von weiteren Faktoren abhing. Aus den Ergebnissen dieser Arbeit konnte gefolgert werden, dass die Durchführung einer präoperativen CT- und einer MRT-Untersuchung die umfangreichste radiologische Diagnostikmethode ist, um eine CI-Operation zu planen, möglichst alle Auffälligkeiten, die zu Schwierigkeiten in der Operation führen können, zu detektieren und Ursachen für eine nicht abgeklärte Hörminderung zu finden. Dennoch sollte, insbesondere unter Berücksichtigung der Strahlenbelastung, die ein Patient im Rahmen eines Felsenbein-CTs erfährt, die Überlegung zugelassen werden, ob zukünftig nur die Durchführung einer MRT-Untersuchung als routinemäßiges Diagnostikverfahren ausreichen könnte. Eine zusätzliche CT-Untersuchung wäre dann nur unter bestimmten Voraussetzungen (z. B. Traumaanamnese des Schädels oder Hinweise auf ein Cholesteatom) indiziert. Weitere Untersuchungen, die zeigen, welche genauen Kriterien dazu seitens des Patienten gegeben sein müssten, sowie Langzeitergebnisse nach erfolgreichen CI-Operationen sollten Gegenstand weiterer Studien sein. N2 - Cochlear implants are the standard treatment for patients with profound sensorineural hearing loss, which cannot be treated with traditional hearing aids. In the process of preoperative diagnostics, both computed tomography (CT) and magnetic resonance imaging (MRI) are used to evaluate the auditory system. The goal of this study was to determine, which kind of results CT and MRI provided and in how many cases the results of CT or MRI influenced the procedure of cochlear implantation. In addition, this study had the aim to provide a possible recommendation, if both CT and MRI are necessary for preoperative diagnostics or if either one imaging modality may be sufficient. The results of this study showed that both CT and MRI can provide information about possible etiologies of hearing loss. Both imaging modalities proved to be able to detect the majority of abnormalities of the temporal bone and provided findings which influenced the procedure of cochlear implantation. Nevertheless, specific lesions were only detected with either CT or MRI. In particular, MRI was able to detect brain lesions, which cannot be seen in CT. Considering these results in combination with the negative aspect of exposure to radiation of CT imaging, MRI might be sufficient for preoperative diagnostics in a substantial part of cochlear implant cases. KW - Cochlear-Implantat KW - Kernspintomografie KW - Computertomografie KW - Präoperative radiologische Diagnostik KW - Preoperative radiological diagnostics Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-189909 ER - TY - THES A1 - Lykowsky, Gunthard T1 - Hardware- und Methodenentwicklung für die 23Na- und 19F-Magnetresonanztomographie T1 - Hardware and method development for 23Na and 19F magnetic resonance imaging N2 - Neben dem Wasserstoffkern 1H können auch andere Kerne für die Magnetresonanztomographie (MRT) genutzt werden. Diese sogenannten X-Kerne können komplementäre Informationen zur klassischen 1H-MRT liefern und so das Anwendungsspektrum der MRT erweitern. Die Herausforderung bei der X-Kern-Bildgebung liegt zum großen Teil in dem intrinsisch niedrigen Signal-zu-Rauschen-Verhältnis (SNR), aber auch in den spezifischen Kerneigenschaften. Um X-Kern-Bildgebung optimal betreiben zu können, müssen daher Sende-/Empfangsspulen, Messsequenzen und -methoden auf den jeweiligen Kern angepasst werden. Im Fokus dieser Dissertation standen die beiden Kerne Natrium (23Na) und Fluor (19F), für die optimierte Hardware und Methoden entwickelt wurden. 23Na spielte in dieser Arbeit vor allem wegen seiner Funktion als Biomarker für Arthrose, einer degenerativen Gelenkserkrankung, eine Rolle. Hierbei ist insbesondere die quantitative Natriumbildgebung von Bedeutung, da sich mit ihr der Knorpelzustand auch im Zeitverlauf charakterisieren lässt. Für die quantitative Messung mittels MRT ist die Kenntnis des B1-Feldes der eingesetzten MR-Spule entscheidend, denn dieses kann die relative Signalintensität stark beeinflussen und so zu Fehlern in der Quantifizierung führen. Daher wurde eine Methode zur Bestimmung des B1-Feldes untersucht und entwickelt. Dies stellte aufgrund des niedrigen SNR und der kurzen sowie biexponentiellen T2-Relaxationszeit von 23Na eine Herausforderung dar. Mit einer retrospektiven Korrekturmethode konnte eine genaue und zugleich schnelle Korrekturmethode gefunden werden. Für die 1H- und 23Na-Bildgebung am menschlichen Knieknorpel wurden zwei praxistaugliche, doppelresonante Quadratur-Birdcage-Resonatoren entwickelt, gebaut und charakterisiert. Der Vergleich der beiden Spulen bezüglich Sensitivität und Feldhomogenität zeigte, dass der Vier-Ring-Birdcage dem Alternating-Rungs-Birdcage für den vorliegenden Anwendungsfall überlegen ist. Die in vivo erzielte Auflösung und das SNR der 23Na-Bilder waren bei beiden Spulen für die Quantifizierung der Natriumkonzentration im Knieknorpel ausreichend. Hochauflösende anatomische 1H-Bilder konnten ohne Mittelungen aufgenommen werden. In einer umfangreichen Multiparameter-MR-Tierstudie an Ziegen wurde der Verlauf einer chirurgisch induzierten Arthrose mittels 23Na- und 1H-Bildgebungsmethoden untersucht. Hierbei kamen dGEMRIC, T1ρ-Messung und quantitative Natrium-MRT zum Einsatz. Trotz des im Vergleich zum Menschen dünneren Ziegenknorpels, der niedrigen Feldstärke von 1,5 T und den auftretenden Ödemen konnten erstmals diese MR-Parameter über den Studienverlauf hinweg an den gleichen Versuchstieren und zu den gleichen Zeitpunkten ermittelt werden. Die Ergebnisse wurden verglichen und die ermittelten Korrelationen entsprechen den zugrundeliegenden biochemischen Mechanismen. Die im Rahmen dieser Studie entwickelten Methoden, Bildgebungsprotokolle und Auswertungen lassen sich auf zukünftige Humanstudien übertragen. Die mit klinischen Bildgebungssequenzen nicht zugängliche kurze Komponente der biexponentiellen T2*-Relaxationszeit von 23Na konnte mittels einer radialen Ultra-Short-Echo-Time-Sequenz bestimmt werden. Hierzu wurde eine Multi-Echo-Sequenz mit einem quasizufälligen Abtastschema kombiniert. Hierdurch gelang es, die kurze und lange T2*-Komponente des patellaren Knorpels in vivo zu bestimmen. 19F wird in der MRT wegen seiner hohen relativen Sensitivität und seines minimalen, körpereigenen Hintergrundsignals als Marker eingesetzt. Zur Detektion der niedrigen in-vivo-Konzentrationen der Markersubstanzen werden hochsensitive Messspulen benötigt. Für die 19F-Bildgebung an Mäusen wurde eine Birdcage-Volumenspule entwickelt, die sowohl für 19F als auch 1H in Quadratur betrieben werden kann, ohne Kompromisse in Sensitivität oder Feldhomogenität gegenüber einer monoresonanten Spule eingehen zu müssen. Dies gelang durch eine verschiebbare Hochfrequenzabschirmung, mit der die Resonanzfrequenz des Birdcage verändert werden kann. Es konnte weiterhin gezeigt werden, dass die Feldverteilungen bei 1H und 19F im Rahmen der Messgenauigkeit identisch sind und so der 1H-Kanal für die Pulskalibrierung und die Erstellung von B1-Karten für die 19F-Bildgebung genutzt werden kann. Hierdurch kann die Messzeit deutlich reduziert werden. Ein grundsätzliches Problemfeld stellt die Korrelation unterschiedlicher Bildgebungsmodalitäten dar. In der MRT betrifft das häufig die Korrelation von in-/ex-vivo-MR-Daten und den dazugehörigen Lichtbildaufnahmen an histologischen Schnitten. In dieser Arbeit wurde erstmals erfolgreich eine 1H- und 19F-MR-Messung an einem histologischen Schnitt vorgenommen. Durch die Verwendung einer optimierten 1H/19F-Oberflächenspule konnte die 19F-Signalverteilung in einer dünnen Tumorscheibe in akzeptabler Messzeit aufgenommen werden. Da der gleiche Schnitt sowohl mit Fluoreszenzmikroskopie als auch mit MRT gemessen wurde, konnten Histologie und MR-Ergebnisse exakt korreliert werden. Zusammenfassend konnten in dieser Arbeit durch Hardware- und Methodenentwicklung zahlreiche neue Aspekte der 19F- und 23Na-MRT beleuchtet werden und so zukünftige Anwendungsfelder erschlossen werden. N2 - In addition to the hydrogen nucleus 1H, other nuclei can also be used for magnetic resonance imaging (MRI). These so-called X-nuclei can provide complementary information on classical 1H MRI and thus expand the range of applications of MRI. The challenge in X-nucleus imaging is largely due to the intrinsically low signal-to-noise ratio (SNR), but also to the specific properties of the nucleus. In order to optimally perform X-nuclei imaging, transmit/receive coils, imaging sequences and methods must be adapted to the respective nucleus. The two nuclei sodium (23Na) and fluorine (19F) were in the focus of this dissertation and thus optimized hardware and methods were developed for these nuclei. 23Na played a major role in this work, mainly because of its function as a biomarker of osteoarthritis, a degenerative joint disease. In particular, the quantitative sodium imaging is of importance, as it can characterize the cartilage state over time. For quantitative measurements by MRI, the knowledge of the B1 field of the MR coil used is crucial, because this can strongly influence the signal intensity and thus lead to errors in the quantification. Therefore, a method for the determination of the B1 field was developed. This presented a challenge due to the low SNR and the short and biexponential T2 relaxation time of 23Na. Using a retrospective correction method, a precise and at the same time rapid correction method could be found. Two practicable double resonant quadrature birdcage resonators have been developed, constructed and characterized for 1H/23Na imaging on human knee cartilage. The comparison of the two coils in terms of sensitivity and field homogeneity showed that the four-ring birdcage is superior to the alternating-rungs birdcage for the present application. The in vivo resolution and SNR of the 23Na images were sufficient for both coils to quantify the sodium concentration in the knee cartilage. High-resolution 1H anatomical images could be acquired without averaging. In a large multiparameter MRI animal study on goats, the progression of surgically induced osteoarthritis was studied using 23Na and 1H imaging techniques. DGEMRIC, T1ρ and quantitative sodium MRI were used. Despite thinner goat cartilage compared to humans, low field strength of 1.5 T and the occurring edema, it was possible for the first time to determine these MR parameters over the course of the study on the same experimental animals and at the same time points. The correlations of the MR parameters correspond to the underlying biochemical mechanisms. The methods, imaging protocols and evaluations developed in this study can be applied to future human studies. The short component of the biexponential T2* relaxation time of 23Na, which is not accessible with clinical imaging sequences, could be determined by means of a radial ultra-short echo time sequence. For this purpose, a multi-echo sequence was combined with a quasi-random sampling scheme. This enabled the determination of the short and long T2* component of patellar cartilage in vivo. 19F is used as a marker in MRI because of its high relative sensitivity and minimal body’s own background signal. To detect the low in vivo concentrations of the marker substances, highly sensitive measuring coils are required. For 19F imaging of mice, a birdcage volume coil was developed that can be operated in quadrature for both 19F and 1H without compromising sensitivity or field homogeneity compared to monoresonant coils. This is due to a slidable RF shield, which is used to change the resonance frequency of the birdcage. It has also been shown that field distributions at 1H and 19F are identical allowing the 1H channel to be used for pulse calibration and B1 mapping for 19F imaging. This can significantly reduce the acquisition time. A fundamental challenge is the correlation of different imaging modalities. In MRI, this often affects the correlation of in and ex vivo MR data and the associated images of histological sections. In this work, 1H and 19F MR measurements of a histological section were successfully performed for the first time. By using an optimized 1H/19F surface coil, the 19F signal distribution in a thin tumor slice was acquired within an acceptable acquisition time. Since the same section was measured by fluorescence microscopy as well as by MRI, histology and MR results could be correlated exactly. In summary, hardware and method development in this work has highlighted numerous new aspects of 19F and 23Na MRI, opening up future fields of application. KW - Kernspintomografie KW - Fluor-19 KW - Natrium-23 KW - 19F-MRT KW - 23Na-MRT Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-188710 ER - TY - THES A1 - Mendes Pereira, Lenon T1 - Morphological and Functional Ultrashort Echo Time (UTE) Magnetic Resonance Imaging of the Human Lung T1 - Morphologische und funktionelle Magnetresonanztomographie der menschlichen Lunge mit ultrakurzen Echozeiten (UTE) N2 - In this thesis, a 3D Ultrashort echo time (3D-UTE) sequence was introduced in the Self-gated Non-Contrast-Enhanced Functional Lung Imaging (SENCEFUL) framework. The sequence was developed and implemented on a 3 Tesla MR scanner. The 3D-UTE technique consisted of a nonselective RF pulse followed by a koosh ball quasi-random sampling order of the k-space. Measurements in free-breathing and without contrast agent were performed in healthy subjects and a patient with lung cancer. A gating technique, using a combination of different coils with high signal correlation, was evaluated in-vivo and compared with a manual approach of coil selection. The gating signal offered an estimation of the breathing motion during measurement and was used as a reference to segment the acquired data into different breathing phases. Gradient delays and trajectory errors were corrected during post-processing using the Gradient Impulse Response Function. Iterative SENSE was then applied to determine the fully sampled data. In order to eliminate signal changes caused by motion, a 3D image registration was employed, and the results were compared to a 2D image registration method. Ventilation was assessed in 3D and regionally quantified by monitoring the signal changes in the lung parenchyma. Finally, image quality and quantitative ventilation values were compared to the standard 2D-SENCEFUL technique. 3D-UTE, combined with an automatic gating technique and SENCEFUL MRI, offered ventilation maps with high spatial resolution and SNR. Compared to the 2D method, UTE-SENCEFUL greatly improved the clinical quality of the structural images and the visualization of the lung parenchyma. Through‐plane motion, partial volume effects and ventilation artifacts were also reduced with a three-dimensional method for image registration. UTE-SENCEFUL was also able to quantify regional ventilation and presented similar results to previous studies. N2 - In dieser Arbeit wurde eine 3D-UTE (ultrashort echo time) Sequenz mit SENCEFUL-MRI kombiniert. Die Sequenz wurde für einen 3 T MR-Scanner entwickelt und implementiert. Die 3D-UTE-Technik bestand aus einem nichtselektiven HF- Impuls, gefolgt von einer quasi-zufälligen Abtastung des k-Raums. Messungen in freier Atmung und ohne Kontrastmittel wurden bei gesunden Probanden und einem Patienten mit Lungenkrebs durchgeführt. Zur Zuordnung der Daten zu verschiedene Atemphasen wurde eine Technik verwendet, die verschiedene Spulen mit hoher Signalkorrelation kombiniert. Die Ergebnisse wurden in einer in-vivo Messung bewertet und mit einem manuellen Ansatz der Spulenselektion verglichen. Die Technik ermöglichte eine Visualisierung der Atembewegung und wurde als Referenz verwendet, um die erfassten Daten in mehrere Atemphasen zu segmentieren. Gradientenverzögerungen und Trajektorienfehler wurden mit der "Gradient Impulse Response Function - GIRF" korrigiert. Bei der Bildrekonstruktion kam Iteratives SENSE zum Einsatz. Eine 3D-Bildregistrierung erlaubte es, Signaländerungen durch Bewegung zu eliminieren. Es erfolgte ein Vergleich der Ergebnisse mit einem 2D- Bildregistrierungsverfahren. Die Lungenventilation wurde in 3D gemessen und anhand der Signaländerungen im Lungenparenchym quantifiziert. Schließlich, wurden die Werte für die Bildqualität und Lungenventilation mit der Standard-2D-SENCEFUL-Technik verglichen. Die 3D-UTE-Sequenz in Kombination mit einer automatischen Gating-Technik und SENCEFUL-MRI, ermöglichte die Akquise von Ventilationskarten mit hoher räumlicher Auflösung und SNR. Im Vergleich zur 2D-Methode, verbesserte UTE- SENCEFUL die klinische Qualität der Morphologischen Bilder. Bewegung, Partialvolumeneffekte und Ventilationsartefakte wurden ebenfalls mit einer dreidimensionalen Methode zur Bildregistrierung reduziert. Insgesamt konnten mit der 3D-UTE Technik die Ergebnisse vorangegangener Studien reproduziert und die Bildqualität verbessert werden. KW - Kernspintomografie KW - Lunge KW - MRI KW - Ultrashort echo time - UTE KW - Magnetic Resonance Imaging KW - Lung Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-183176 ER - TY - THES A1 - Endres, Ralph Julian T1 - Networks of fear: Functional connectivity of the amygdala, the insula and the anterior cingulate cortex in two subtypes of specific phobia T1 - Netzwerke der Angst: Funktionelle Konnektivität der Amygdala, der Insula und des anterioren cingulären Cortex in zwei Subtypen der spezifischen Phobie N2 - Neuroimaging research has highlighted the relevance of well-balanced functional brain interactions as an essential basis for efficient emotion regulation. In contrast, abnormal coupling of fear-processing regions such as the amygdala, the anterior cingulate cortex (ACC) and the insula could be an important feature of anxiety disorders. Although activity alterations of these regions have been frequently reported in specific phobia, little is known about their functional interactions during phobogenic stimulus processing. To explore these interrelationships in two subtypes of specific phobia – i.e., the blood-injection-injury subtype and the animal subtype – functional connectivity (FC) was analyzed in three fMRI studies. Two studies examined fear processing in a dental phobia group (DP), a snake phobia group (SP) and a healthy control group (HC) during visual phobogenic stimuli presentation while a third study investigated differences between auditory and visual stimuli presentation in DP and HC. Due to a priori hypotheses of impaired interactions between the amygdala, the ACC and the insula, a first analysis was conducted to explore the FC within these three regions of interest. Based on emerging evidence of functionally diverse subregions, the ACC was further divided into a subgenual, pregenual and dorsal ACC and the insula was divided into a ventral-anterior, dorsal-anterior and posterior region. Additionally, an exploratory seed-to-voxel analysis using the amygdala, ACC and insula as seeds was conducted to scan for connectivity patterns across the whole brain. The analyses revealed a negative connectivity of the ACC and the amygdala during phobogenic stimulus processing in controls. This connectivity was predominantly driven by the affective ACC subdivision. By contrast, SP was characterized by an increased mean FC between the examined regions. Interestingly, this phenomenon was specific for auditory, but not visual symptom provocation in DP. During visual stimulus presentation, however, DP exhibited further FC alterations of the ACC and the insula with pre- and orbitofrontal regions. These findings mark the importance of balanced interactions between fear-processing regions in specific phobia, particularly of the inhibitory connectivity between the ACC and the amygdala. Theoretically, this is assumed to reflect top-down inhibition by the ACC during emotion regulation. The findings support the suggestion that SP particularly is characterized by excitatory, or missing inhibitory, (para-) limbic connectivity, reflecting an overshooting fear response based on evolutionary conserved autonomic bottom-up pathways. Some of these characteristics applied to DP as well but only under the auditory stimulation, pointing to stimulus dependency. DP was further marked by altered pre- and orbitofrontal coupling with the ACC and the insula which might represent disturbances of superordinate cognitive control on basal emotion processes. These observations strengthen the assumption that DP is predominantly based on evaluation-based fear responses. In conclusion, the connectivity patterns found may depict an intermediate phenotype that possibly confers risks for inappropriate phobic fear responses. The findings presented could also be of clinical interest. Particularly the ACC – amygdala circuit may be used as a predictive biomarker for treatment response or as a promising target for neuroscience-focused augmentation strategies as neurofeedback or repetitive transcranial magnetic stimulation. N2 - Neurowissenschaftliche Erkenntnisse der letzten Jahre verdeutlichten die Relevanz intakter neuronaler Netzwerke als Grundlage adäquater Emotionsregulationsmechanismen. Funktionelle Dysregulationen zwischen angstverarbeitenden Regionen wie der Amygdala, der Insula oder dem anterioren cingulären Cortex (ACC) könnten hingegen einen wichtigen pathophysiologischen Mechanismus von Angststörungen darstellen. Obwohl Aktivitätsunterschiede dieser Regionen wiederholt für spezifische Phobien beschrieben wurden, sind deren funktionelle Interaktionen während phobischer Stimulusverarbeitung kaum erforscht. Zur Untersuchung dieser Interaktionen in zwei Subtypen der spezifischen Phobie – dem Blut-Spritzen-Verletzungs-Typus und dem Tier-Typus – wurden im Rahmen dieser Arbeit funktionelle Konnektivitäts-Analysen (FK) anhand dreier fMRT- (funktionelle Magnetresonanztomographie) Studien durchgeführt. Zwei Studien untersuchten die neurale Verarbeitung visueller phobischer Stimuli in einer dentalphobischen Gruppe (DP), einer schlangenphobischen Gruppe (SP) sowie einer Kontrollgruppe (KG). Ergänzend verglich eine dritte Studie den Einfluss visueller und akustischer Stimuli für die DP und eine KG. Basierend auf der a priori-Hypothese einer veränderten FK zwischen der Amygdala, der Insula und dem ACC wurden deren spezifische Konnektivitätsmuster untersucht. Aufgrund funktionell unterschiedlicher Subregionen erfolgte eine Untergliederung des ACC in eine subgenuale, perigenuale und dorsalen Region. Analog dazu wurde die Insula in eine ventral-anteriore, dorsal-anteriore und posteriore Region unterteilt. Um darüberhinausgehender Konnektivitätsmuster über das gesamte Gehirn zu ermitteln, wurde eine abschließende Seed-to-Voxel-Analyse mit den Seeds Amygdala, Insula und ACC durchgeführt. In der Auswertung zeigte sich eine negative FK der Amygdala und des ACC während phobischer Stimulusverarbeitung in der KG, die insbesondere auf die ventrale Division des ACC zurückzuführen war. Die phobischen Gruppen hingegen waren im Vergleich zu der Kontrollgruppe durch eine erhöhte Konnektivität der untersuchten Regionen gekennzeichnet. Dieser Effekt war bei der DP spezifisch für die akustische Stimulusmodalität. Bei visueller Stimuluspräsentation zeigten sich hingegen veränderte Konnektivitätsmuster des ACC und der Insula mit prä- und orbitofrontalen Regionen. Insbesondere die negative FK der Amygdala und des ACC, die theoretisch auf einer top-down-Inhibition des ACC über die Amygdala basiert, erscheint einen wichtigen Bestandteil einer effektiven emotionalen Kontrolle darzustellen. In beiden phobischen Gruppen fehlte diese Inhibition. Die erhöhte FK (para-)limbischer Konnektivität der SP könnte hingegen die verstärkte Rekrutierung autonomischer bottom-up-Prozesse als zugrundeliegendem Mechanismus der überschießenden und irrationalen Angstreaktion repräsentieren. Diese Charakteristika konnten in der DP nur für die akustische Stimulusmodalität beobachtet werden. Während der visuellen Stimuluspräsentation war die DP durch Dysregulationen prä- und orbitofrontaler Regionen gekennzeichnet, welche eine beeinträchtigte kognitive Kontrolle über grundlegende Emotionsprozesse widerspiegeln könnte. Dies entspricht der Annahme, dass die DP vor allem durch evaluationsbasierte Furchtreaktionen gekennzeichnet ist, während in der SP als Vertreter des Tier-Typus evolutionär konservierte, limbische Prozesse dominieren. Zusammenfassend bestätigen die Ergebnisse die Bedeutung funktioneller Netzwerke in der spezifischen Phobie, wobei die gefundenen Konnektivitätsmuster einen intermediären Phänotyp darstellen könnten, der möglicherweise das Risiko für das Auftreten dysfunktionaler phobischer Angstreaktionen vermittelt. Von klinischem Interesse ist vor allem die Amygdala – ACC-Vernetzung, die als prädiktiver Biomarker für das Therapieansprechen genutzt oder im Rahmen neuromodulatorischer Therapieansätze wie dem Neurofeedback oder der repetitiven transkraniellen Magnetstimulation gezielt angesteuert werden könnte. KW - Kernspintomografie KW - Psychiatrie KW - Phobie KW - fMRT KW - Funktionelle Konnektivität KW - Spezifische Phobien KW - fMRI KW - Functional Connectivity KW - Specific Phobia KW - Neuroimaging KW - Dental Phobia KW - Zahnbehandlungsphobie KW - Angstverarbeitung Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-180950 ER - TY - THES A1 - Haug, Lukas T1 - Dynamisch-kernspintomographische Definition der Flexions-Extensionsachse des Kniegelenks T1 - Dynamic magnetic resonance imaged definition of the axis about the knee flexes and extends N2 - Es sollte eine dynamische MRT Methode entwickelt werden, welche die Berechnung und Darstellung einer Achse einer komplexen Bewegung ermöglicht. Dabei war die eigens auferlegte Definition von „Dynamik“, dass ein kompletter Bewegungsablauf als dreidimensionaler Datensatz im Zeitverlauf erfasst und ausgewertet werden soll (4 Dimensionen). Diese Methode sollte auf die komplexe Bewegung des Kniegelenks angewendet werden und mit einem Knie-Modell verglichen werden. Schlussendlich sollte die Flexion-Extensionsbewegung als vereinfacht monoaxial verifiziert werden. Für diese Ziele wurden folgende Schritte durchgeführt: Es wurden zunächst 10 Kadaver-Knie präpariert und hierbei u.a. kontrastmittelhaltige Kugeln als Tracker eingebracht. Die Knie wurden dann im Rahmen des Versuchsaufbaus in einer speziellen, pneumatischen Bewegungsschiene befestigt, welche dann automatisiert eine passive Bewegung vollzog. Diese wurde mit einer dynamischen MRT Methode festgehalten. Der entstandene 4D-Datensatz wurde eingelesen und die Kugelpositionen ausgewertet. Die Koordinaten der Positionen dienten dann als Grundlage für spezielle mathematische Algorithmen, welche die Flexion-Extensionsachse und ggf. die Innen- und Außenrotationsachse berechneten und schließlich eine graphische Darstellung ermöglichten. Zur Überprüfung wurde zudem ein Knie-Modell herangezogen und von diesem die gleichen Berechnungen durchgeführt. N2 - A dynamic MRI method should be developed which allowed the calculation and presentation of the axis of a complex motion. The specially imposed definition of "dynamics" was that a complete sequence of movement should be recorded and evaluated over time as a three-dimensional data set (4 dimensions). This method should be applied to the complex movement of the knee joint and compared with a knee model. Finally, the flexion extension movement should be verified as simplified monoaxially. The following steps have been performed: First, 10 cadaver knees were prepared and were, among other steps, provided with contrast media filled spheres as tracker. The knees were then mounted in a special pneumatic movement device as part of the experimental set-up, which then automatically performed a passive movement. This was recorded using a dynamic MRI method. The resulting 4D data set was read in and the positions of the spheres were evaluated. The coordinates of the positions then served as a basis for special mathematical algorithms, which calculated the flexion extension axis and, depending on the algorithm, the inner and outer rotational axis and finally allowed a graphical presentation. For verification, the same calculations were performed with a knee model. KW - Kniegelenk KW - Kernspintomografie KW - dynamische MRT KW - knee joint KW - dynamic magnetic resonance imaging Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-179799 ER - TY - THES A1 - Gutjahr, Fabian Tobias T1 - Neue Methoden der physiologischen Magnet-Resonanz-Tomographie: Modellbasierte T1-Messungen und Darstellung von chemischem Austausch mit positivem Kontrast T1 - Novell Methods for Physiological MRI: Model based T1-Quantification and Positive Contrast Chemical Exchange Measurements N2 - Ziel dieser Arbeit war es, neue quantitative Messmethoden am Kleintier, insbesondere die Perfusionsmessung am Mäuseherz, zu etablieren. Hierfür wurde eine retrospektiv getriggerte T1-Messmethode entwickelt. Da bei retrospektiven Methoden keine vollständige Abtastung garantiert werden kann, wurde ein Verfahren gefunden, das mit Hilfe von Vorwissen über das gemessene Modell sehr effizient die fehlenden Daten interpolieren kann. Mit Hilfe dieser Technik werden dynamische T1-Messungen mit hoher räumlicher und zeitlicher Auflösung möglich. Dank der hohen Genauigkeit der T1-Messmethode lässt sich diese für die nichtinvasive Perfusionsmessung am Mäuseherz mittels der FAIR-ASL-Technik nutzen. Da auf Grund der retrospektiven Triggerung Daten an allen Positionen im Herzzyklus akquiriert werden, konnten T1- und Perfusionskarten nach der Messung zu beliebigen Punkten im Herzzyklus rekonstruiert werden. Es bietet sich an, Techniken, die für die myokardiale Perfusion angewandt werden, auch für die Nierenperfusionsmessung zu verwenden, da die Niere in ihrer Rinde (Cortex) eine ähnlich hohe Perfusion aufweist wie das Myokard. Gleichzeitig führen Nierenerkrankungen oftmals zu schlechter Kontrastmittelverträglichkeit, da diese bei Niereninsuffizienz u.U. zu lange im Körper verweilen und die Niere weiter schädigen. Auch deshalb sind die kontrastmittelfreien Spin-Labeling-Methoden hier interessant. Die FAIR-ASL-Technik ist jedoch an Mäusen in koronaler Ansicht für die Niere schlecht geeignet auf Grund des geringen Unterschieds zwischen dem markierten und dem Vergleichsexperiment. Als Lösung für dieses Problem wurde vorgeschlagen, die Markierungsschicht senkrecht zur Messschicht zu orientieren. Hiermit konnte die Sensitivität gesteigert und gleichzeitig die Variabilität der Methode deutlich verringert werden. Mit Hilfe von kontrastmittelgestützten Messungen konnten auch das regionale Blutvolumen und das Extrazellularvolumen bestimmt werden. In den letzten Jahren hat das Interesse an Extrazellularvolumenmessungen zugenommen, da das Extrazellularvolumen stellvertretend für diffuse Fibrose gemessen werden kann, die bis dahin nichtinvasiven Methoden nicht zugänglich war. Die bisher in der Literatur verwendeten Quantifizierungsmethoden missachten den Einfluss, den das Hämatokrit auf den ECV-Wert hat. Es wurde eine neue Korrektur vorgeschlagen, die allerdings zusätzlich zur ECV-Messung auch eine RBV-Messung benötigt. Durch gleichzeitige Messung beider Volumenanteile konnte auch erstmals das Extrazellulare-Extravaskuläre-Volumen bestimmt werden. Eine gänzlich andere kontrastmittelbasierte Methode in der MRT ist die Messung des chemischen Austauschs. Hierbei wirkt das Kontrastmittel nicht direkt beschleunigend auf die Relaxation, sondern der Effekt des Kontrastmittels wird gezielt durch HF-Pulse an- und ausgeschaltet. Durch den chemischen Austausch kann die Auswirkung der HF-Pulse akkumuliert werden. Bislang wurde bei solchen Messungen ein negativer Kontrast erzeugt, der ohne zusätzliche Vergleichsmessungen schwer detektierbar war. Im letzten Teil dieser Arbeit konnte eine neue Methode zur Messung des chemischen Austauschs gezeigt werden, die entgegen der aus der Literatur bekannten Methoden nicht Sättigung, sondern Anregung überträgt. Diese Änderung erlaubt es, einen echten positiven chemischen Austausch-Kontrast zu erzeugen, der nicht zwingend ein Vergleichsbild benötigt. Gleichzeitig ermöglicht die Technik, dadurch dass Anregung übertragen wird, die Phase der Anregung zu kontrollieren und nutzen. Eine mögliche Anwendung ist die Unterscheidung verschiedener Substanzen in einer Messung. In der Summe wurden im Rahmen dieser Arbeit verschiedene robuste Methoden eta- bliert, die die Möglichkeiten der quantitativen physiologischen MRT erweitern. N2 - The objective of this dissertation was to develop new methods for physiological magnetic resonance imaging. A new retrospectively triggered T1-method was developed. Due to the retrospectivity, full sampling of k-space can not be warranted. Therefore a model- based interpolation method was developed to reconstruct missing data efficiently. Using this technique, dynamic T1-measurements with high temporal and spatial resolution could be acquired. Due to the high precision of the developed T1-method, perfusion could be quantified using Arterial Spin Labeling. In comparison to the method established previously in our laboratory, the resolution could be doubled. Retrospective triggering enables reconstruc- tion of parameter maps on arbitrary positions in the heart cycle, as data are acquired continuously over several heart cycles. The perfusion measurement benefits from recon- struction on the end systole, as partial volume effects are decreased, due to the increased myocardial wall thickness. This serves as an effective increase in resolution. Furthermore, the data distributed over the whole heart cycle could be used to accelerate and stabilize the measurement. Cardiac and renal diseases can be directly related, as deficiency in one of the organs affects the other one. Additionally several diseases like hypertension or diabetes affect both organs. Moreover, kidneys are highly perfused, similar to the myocardium. Renal insufficiency can also lead to contrast agent intolerance, as clearance rates can be redu- ced. Therefore the FAIR-ASL technique lends itself to kidney perfusion measurements. It can, however, be problematic in small animals in coronal view, as the control-experiment inadvertently labels much of the same tissue and blood, as the labeling experiment. A modified FAIR-ASL measurement could be shown to increase sensitivity and reduce in- ter-measurement-variability by repositioning the inversion slice of the control experiment orthogonally to the measurement slice. The T1-method was used in combination with contrast agent based measurements to quantify the regional blood volume and the extracellular volume fraction. There has been an increased interest in extracellular volume fraction measurements as the extracel- lular volume is used as a proxy for the detection of diffuse fibrosis, which has previously been inaccessible to non-invasive methods. Several correction factors are used in volume fraction quantification, but the influence of hematocrit in ECV measurements has been neglected so far. In mice and rats, the regional blood volume is a major constituent of the ECV, leading to a significant influence of hematocrit. A new correction is proposed to account for the volume fraction taken up by hematocrit. For this ECV hematocrit correction, the RBV has to be measured as well. Using both measurements, the ex- tracellular volume fraction can be corrected and the extracellular-extravascular-volume- fraction quantified. A fundamentally different contrast-mechanism can be utilized using the measurement of chemical exchange. Instead of shortening relaxation times, the contrast provided by chemical exchange agents can be turned on and off using frequency selective rf-pulses. Due to the chemical exchange the effect of these pulses can be accumulated. Measure- ments exploiting this accumulation effect in general produce a negative contrast requiring a control-experiment for further evaluation. In the last part of this dissertation, a new technique transferring excitation instead of saturation could be demonstrated. By ge- nerating a real positive contrast, no control experiment is required. Other properties unavailable to previously published chemical exchange transfer methods can be exploi- ted. One example demonstrated in this dissertation is the separation of simultaneously excited compounds by their respective phase information imprinted by the excitation pulses. In summary, several robust methods could be implemented to further the capabilities of quantitative physiological MRI. KW - Kernspintomografie KW - Physioloische MRT KW - Modellbasierte Rekonstruktion KW - FAIR-ASL KW - Chemischer Austausch KW - Regionales Blutvolumen KW - Extrazellularvolumen KW - T1-Quantifizierung KW - Kernspinresonanz KW - Myokardiale Perfusion KW - Niere KW - Perfusionsmessung Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-161061 ER - TY - THES A1 - Winter, Patrick T1 - Neue Methoden zur Quantitativen Kardiovaskulären MR-Bildgebung T1 - New methods for quantitative cardiovascular magnetic resonance imaging N2 - Herzkreislauferkrankungen stellen die häufigsten Todesursachen in den Industrienationen dar. Die Entwicklung nichtinvasiver Bildgebungstechniken mit Hilfe der Magnetresonanz-Tomografie (MRT) ist daher von großer Bedeutung, um diese Erkrankungen frühzeitig zu erkennen und um die Entstehungsmechanismen zu erforschen. In den letzten Jahren erwiesen sich dabei genetisch modifzierte Mausmodelle als sehr wertvoll, da sich durch diese neue Bildgebungsmethoden entwickeln lassen und sich der Krankheitsverlauf im Zeitraffer beobachten lässt. Ein große Herausforderung der murinen MRT-Bildgebung sind die die hohen Herzraten und die schnelle Atmung. Diese erfordern eine Synchronisation der Messung mit dem Herzschlag und der Atmung des Tieres mit Hilfe von Herz- und Atemsignalen. Konventionelle Bildgebungstechniken verwenden zur Synchronisation mit dem Herzschlag EKG Sonden, diese sind jedoch insbesondere bei hohen Feldstärken (>3 T) sehr störanfällig. In dieser Arbeit wurden daher neue Bildgebungsmethoden entwickelt, die keine externen Herz- und Atemsonden benötigen, sondern das MRT-Signal selbst zur Bewegungssynychronisation verwenden. Mit Hilfe dieser Technik gelang die Entwicklung neuer Methoden zur Flussbildgebung und der 3D-Bildgebung, mit denen sich das arterielle System der Maus qualitativ und quantitativ erfassen lässt, sowie einer neuen Methode zur Quantisierung der longitudinalen Relaxationszeit T1 im murinen Herzen. Die in dieser Arbeit entwickelten Methoden ermöglichen robustere Messungen des Herzkreislaufsystems. Im letzten Kapitel konnte darüber hinaus gezeigt werden dass sich die entwickelten Bildgebungstechniken in der Maus auch auf die humane Bildgebung übertragen lassen. N2 - Cardiovascular diseases are one of the main causes of death in western countries. Hence, the development of non-invasive imaging techniques using Magnetic Resonance Imaging (MRI) is very important for early detection of these illnesses and for examination of the biological mechanisms. In the past years genetically modified mouse models have proven to be great assets, since they allow the development of new imaging techniques and to investigate the progress of cardiovascular diseases in time lapse. The main challenge of murine MRI is the high heart rate und the fast respiration. Hence, synchronization of the measurement with cardiac motion and breathing by using cardiac and respiration signals is required. Most imaging techniques use ECG leads for synchronization with the heartbeat, however, these probes are prone to disturbances at high magnetic field strengths (>3 T). In this work new imaging techniques were developed that do not use external cardiac and respiration signals but the MRI signal itself for motion synchronization. With these techniques new methods for flow quantification und 3D imaging could be developed for qualitative and quantitative measurements in the murine arteries. Furthermore, a new method for quantification of the longitudinal relaxation time T1 in the murine heart could be developed. The methods presented in this work enable more robust measurements of the cardiovascular system. In the last chapter it could be shown that the imaging techniques developed in the mouse can also be transferred to human MRI. KW - Kernspintomografie KW - Kardiovaskuläres System KW - Flussbildgebung KW - 3D-Bildgebung KW - Selbstnavigation KW - T1 KW - UTE KW - Maus KW - Aorta KW - Herzmuskel KW - Herzschlag Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-174023 ER - TY - THES A1 - Munz, Eberhard T1 - Physiological and metabolical high-resolution MRI of plants T1 - Physiologische und metabolische hochaufgelöste Pflanzen-Magnetresonanzbildgebung N2 - The noninvasive magnetic resonance imaging technique allows for the investigation of functional processes in the living plant. For this purpose during this work, different NMR imaging methods were further developed and applied. For the localisation of the intrusion of water into the germinating rape seed with the simultaneous depiction of the lipid-rich tissue via a 3D rendering, in Chap. 5 the technique of interleaved chemical selective acquisition of water and lipid was used in the germinating seed. The utilization of high-resolution MR images of germinated seeds enabled the localization of a predetermined water gap in the lipid-rich aleurone layer, which resides directly under the seed coat. The for a long time in biology prevalent discussion, whether such a gap exists or the seed soaks up the water from all sides, rather like a sponge, could hereby, at least for the rapeseed seed, be answered clearly. Furthermore, the segmentation and 3D visualization of the vascular tissue in the rapeseed seeds was enabled by the high-resolution datasets, a multiply branched structure preconstructed in the seed could be shown. The water is directed by the vascular tissue and thus awakens the seed gradually to life. This re-awakening could as well be tracked by means of invasive imaging via an oxygen sensor. In the re-awakened seeds, the lipid degradation starts, other than expected, not in the lipid-rich cotyledons but in the residual endosperm remaining from seed development and in the aleurone layer which previously protected the embryo. Within this layer, the degradation could be verified in the high-resolution MR datasets. The method presented in Chap. 6 provides a further characteristic trait for phenotyping of seeds and lipid containing plants in general. The visualization of the compounds of fatty acids in plant seeds and fruits could be achieved by the distinct utilization of chemical shift-selective imaging techniques. Via the application of a CSI sequence the fatty acid compounds in an olive were localized in a 2D slice. In conjunction with an individually adjusted CHESS presaturation module Haa85 the high-resolution 3D visualization of saturated and unsaturated fatty acid compounds in different seeds was achieved. The ratio maps calculated from these datasets allow to draw conclusions from the developmental stage or the type of seed. Furthermore, it could be shown that the storage condition of two soybean seeds with different storage time durations lead to no degradation of the fatty acid content. Additional structural information from inside of dry seeds are now accessible via MRI. In this work the imaging of cereal seeds could be significantly improved by the application of the UTE sequence. The hitherto existing depictions of the lipid distribution, acquired with the spin echo sequence, were always sufficient for examinations of the lipid content, yet defects in the starchy endosperm or differences in the starch concentration within the seed remained constantly unseen with this technique. In a direct comparison of the datasets acquired with the previous imaging technique (spin echo) and with UTE imaging, the advantage of data acquisition with UTE could be shown. By investigating the potential seed compounds (starch, proteins, sugar) in pure form, the constituent parts contributing to the signal could be identified as bound water (residual moisture) and starch. The application of a bi-exponential fit on the datasets of the barley seed enabled the separate mapping of magnetization and of relaxation time of two components contributing to the NMR signal. The direct comparison with histological stainings verified the previous results, thus this technique can be used for the selective imaging of starch in dry seeds. Conclusions on the translocation characteristics in plants can be drawn by the technique proposed in Chap. 8. The associated translocation velocities can now, even in the range of several um/h, be determined in the living plant. Based on calculated concentrations of an MR contrast agent, which was taken up by the plant, these translocation velocities were estimated both in longitudinal direction, thus along the vascular bundle, and in horizontal direction, thus out of the bundle. The latter velocity is located below the contrast agent's velocity value of free diffusion. By adjusting a dynamic contrast-enhancing imaging technique (DCE-Imaging, Tof91) the acquisition duration of a T1-map was significantly reduced. By means of these maps, local concentrations of the contrast agent in plant stems and the siliques of the rapeseed plant could be determined. Numerous questions in plant science can only be answered by non-invasive techniques such as MRI. For this reason, besides the experimental results achieved in this work, further NMR methods were tested and provided for the investigation of plants. As an example, the study on the imaging of magnetic exchange processes are mentioned, which provided the groundwork for a possible transfer of CEST experiments (Chemical Exchange Saturation Transfer) to the plant. The results are presented in the bachelor thesis of A. Jäger Jae17, which was performed under my supervision, they find great interest under biologists. The development of new technologies, which extend the possibilities for the investigation of living organisms, is of great importance. For this reason, I have contributed to the development of the currently unpublished method RACETE (Refocused Acquisition of Chemical Exchange Transferred Excitations [Jak17, Reu17, Gut18a]). By rephasing the transferred magnetization the utilization of properties which have not been available in chemical "`exchange"' experiments is enabled. With this method a positive contrast is generated, thus a reference experiment is not mandatory. Furthermore, the image phase, which in classical experiments contains no information about the exchanged protons, can be used for the distinct identification of multiple substances which have been excited simultaneously. This recently at the Department of Experimental Physics V developed method can be used in particular for the identification of lipids and for the localization of sugars and amino acids, thus it can serve the enhancement and improvement of non-invasive analytical methods. N2 - Die nicht-invasive Bildgebungstechnik der Magnetresonanz ermöglicht es, funktionelle Prozesse in Pflanzen am lebenden Objekt zu untersuchen. Hierfür wurden im Rahmen dieser Arbeit verschiedene NMR-Bildgebungsmethoden weiterentwickelt und angewendet. Da Pflanzen ein magnetisch sehr inhomogenes Gewebe besitzen, bedingt durch Lufteinschlüsse und das Vorhandensein verschiedenster gelöster Stoffe im Pflanzengewebe, wurden daher hauptsächlich Spin-Echo-Methoden für die Bildgebung verwendet. Um das erste Eindringen des Wassers in den keimenden Raps-Samen bei gleichzeitiger Darstellung des lipid-reichen Gewebes mittels einer 3D-Visualisierung zu lokalisieren, wurde in Kapitel 5 die Technik der verschachtelten, chemisch selektiven Aufnahme von Wasser und Lipid im keimenden Samen verwendet. Durch Verwendung von hochausgelösten MR-Aufnahmen an gekeimten Samen konnte weiterhin in der lipid-reichen Aleuron-Schicht, die sich direkt unter der Samenschale befindet, ein gezielt angelegter Einlass für das Wasser verortet werden. Die in der Biologie lange Zeit verbreitete Diskussion, ob es einen solchen Einlass gibt oder der keimende Samen das Wasser eher wie ein Schwamm von allen Seiten aufsaugt, konnte hierdurch, zumindest für den Raps-Samen, eindeutig beantwortet werden. Weiterhin konnte durch die hoch-aufgelösten Aufnahmen das vaskuläre Gewebe in den Raps-Samen segmentiert und in 3D veranschaulicht werden, es zeigte sich eine mehrfach verzweigte Struktur, die bereits im Samen angelegt ist. Das Wasser folgt hierbei dem vaskulären Gewebe und erweckt hierdurch den Samen schrittweise zum Leben. Dieses Wieder-Erwachen konnte ebenfalls durch die invasive Bildgebung mittels eines Sauerstoff-Sensors nachverfolgt werden. Im nun erwachten Samen selbst beginnt der Lipid-Abbau, anders als zunächst angenommen, nicht in den lipid-haltigen Kotyledonen sondern im von der Samen-Entwicklung verbliebenden Endosperm und in der den Keimling vormals schützenden Aleuron-Schicht. In dieser konnte der Abbau an gekeimten Samen durch hochaufgelöste MR-Aufnahmen nachgewiesen werden. Die in Kapitel 6 vorgeschlagene Methode liefert ein weiteres Merkmal zur Phenotypisiserung von Samen und lipidhaltigen Pflanzenbestandteilen im Allgemeinen. Die Darstellung der Bestandteile ungesättigter Fettsäuren in Pflanzensamen und -Früchten konnte durch gezielte Verwendung von chemisch selektiven Bildgebungstechniken erreicht werden. Durch die Anwendung einer CSI-Sequenz konnten die Fettsäurebestandteile in Oliven in einer 2D-Schicht lokalisiert werden. In Verbindung mit einem jeweils angepassten CHESS-Vorsättigungsmodul Haa85 wurde die hochaufgelöste 3D-Darstellung von gesättigten und ungesättigten Fettsäurebestandteilen in unterschiedlichen Samen erreicht. Rückschlüsse über das Entwicklungsstadium sowie die Sorte der verwendeten Samen können aus den Verhältnis-Karten, die aus den jeweiligen Datensätzen berechnet wurden, gezogen werden. Dass in diesem Fall die Aufbewahrungsmethode zu keiner Degradation der Fettsäurezusammensetzung geführt hat, konnte weiterhin am Beispiel von zwei Sojasamen mit unterschiedlicher Lagerdauer gezeigt werden. Zusätzliche strukturelle Informationen aus dem Inneren trockener Samen sind nun mittels MRT zugänglich. In dieser Arbeit konnte durch die UTE-Sequenz die Bildgebung von Getreidesamen deutlich vorangebracht werden. Die bisherigen Darstellungen der Lipid-Verteilung, aufgenommen mit einer Spin-Echo Sequenz, waren zwar für die Betrachtung des Lipid-Gehalts stets ausreichend, Defekte im stärkehaltigen Endosperm oder Unterschiede in der Stärke-Konzentration innerhalb des Samen blieben mit dieser Technik jedoch stets verborgen. Im direkten Vergleich der mit der bisherigen Technik (Spin-Echo) und der UTE-Bildgebung aufgenommenen Datensätze konnte der Vorteil der Datenaufnahme mit UTE gezeigt werden. Durch die Untersuchung der möglichen Samenbestandteile (Stärke, Proteine, Zucker) in Reinform konnten die zum Signal beitragen Bestandteile als gebundenes Wasser (Restfeuchte) und Stärke identifiziert werden. Die Verwendung bi-exponentiellen Fits and die Messdaten ermöglichte es im Gersten-Samen, zwei zum Signal beitragende Komponenten in getrennten Karten bezüglich ihrer Magnetisierung und Relaxationszeit zu trennen. Der Vergleich mit histologischen Färbungen bestätigte die bisherigen Ergebnisse, somit kann diese Technik zur selektiven Darstellung von Stärke in trockenen Samen verwendet werden. Rückschlüsse auf das Transportverhalten in Pflanzen können durch die in Kapitel 8 vorgestellte Technik gezogen werden. Die zugehörigen Transportgeschwindigkeiten im lebenden Pflanzenobjekt können nun, selbst im Bereich von wenigen $\mu$m/h, bestimmt werden. Diese wurden anhand von berechneten Konzentrationen eines von der Pflanze aufgenommenen MR-Kontrastmittels sowohl in longitudinaler Richtung, also entlang des Leitgewebebündels, als auch in horizontaler Richtung, also aus dem Leitbündel heraus, abgeschätzt werden; Letztere Geschwindigkeit liegt deutlich unter dem Wert der freien Diffusionsgeschwindigkeit des Kontrastmittels. Hierfür wurden durch Anpassung einer dynamischen Kontrast-erhöhenden Bildgebungstechnik (DCE-Imaging, Tof91) die Aufnahmedauer einer für die weiteren Berechnungen benötigen T1-Karte deutlich reduziert. Mittels dieser Karten konnten die lokalen Konzentrationen des Kontrastmittels in Pflanzenstängeln und Schoten der Rapspflanze bestimmt werden. Zahlreiche Fragen in der Pflanzenforschung können nur durch nicht-invasive Techniken wie MRT beantwortet werden. Deswegen wurden, neben den experimentellen Ergebnissen, die mittels dieser Arbeit erreicht wurden, auch weitere NMR Methoden für die Untersuchung von Pflanzen getestet und zur Verfügung gestellt. Als Beispiel seien hier die Untersuchungen zur Bildgebung von magnetischen Austauschprozessen genannt, welche eine Vorarbeit zur möglichen Übertragung con CEST-Experimenten (Chemical Exchange Saturation Transfer) auf das Modell Pflanze liefern. Die Ergebnisse sind in der Bachelor-Arbeit von A. Jäger \cite{jaeger17}, an deren Durchführung ich als Betreuer maßgeblich beteiligt war, dargestellt und finden großes Interesse bei Biologen. Von besonderer Wichtigkeit sind auch die Entwicklungen neuer Technologien, die die Möglichkeiten zur Untersuchung von lebenden Organismen erweitern können. Deswegen habe ich zu der Entwicklung der bislang unveröffentlichten Methode RACETE (Refocused Acquisition of Chemical Exchange Transferred Excitations [Jak17, Reu17, Gut18a]) beigetragen. Durch das Rephasieren der transferierten Magnetisierung können Eigenschaften, die bislang in chemischen "`Austausch"'-Experimenten nicht zur Verfügung stehen, ausgenutzt werden. Mit dieser Methode wird ein positiver Kontrast erzeugt, sie ist deshalb nicht zwingend auf ein Referenz-Experiment angewiesen. Weiterhin kann die Bildphase, welche in klassichen CEST-Experimenten keine Information über die ausgetauschten Protonen enthält, zur eindeutigen Identifizierung mehrerer parallel angeregter Substanzen verwendet werden. KW - Kernspintomografie KW - Pflanzen KW - Pflanzenbildgebung Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-172518 ER - TY - THES A1 - Kreutner, Jakob T1 - Charakterisierung des Knochens und seiner Mikrostruktur mit hochauflösender 3D-MRT T1 - Characterization of Bone and its Microstructure using High-resolution 3D-MRI N2 - Neue Therapieansätze durch Tissue Engineering erfordern gleichzeitig angepasste Diagnosemöglichkeiten und nicht-invasive Erfolgskontrollen. Speziell die 3D-MR-Bildgebung ist ein vielversprechendes Instrument, um Parameter mit hoher räumlicher Präzision zu quantifizieren. Vor diesem Hintergrund wurden im Rahmen dieser Arbeit neue Ansätze für die hochauflösende 3D-MRT in vivo entwickelt und deren Eignung im Bereich des Tissue Engineerings gezeigt. Welchen Vorteil die Quantifizierung von Parametern bietet, konnte im Rahmen einer prä-klinischen Studie an einem Modell der Hüftkopfnekrose gezeigt werden. Der Therapieverlauf wurde zu verschiedenen Zeitpunkten kontrolliert. Trotz der niedrigen räumlichen Auflösung, konnten durch eine systematische Auswertung der Signalintensitäten von T1- und T2-FS-gewichteten Aufnahmen Rückschlüsse über Veränderungen in der Mikrostruktur gezogen werden, die darüber hinaus in guter Übereinstimmung mit Ergebnissen von ex vivo µCT-Aufnahmen waren. Dort konnte eine Verdickung der Trabekelstruktur nachgewiesen werden, welche sehr gut mit einer Signalabnahme in den T1-gewichteten Aufnahmen korrelierte. Die radiale Auswertung der Daten erlaubte dabei eine komprimierte Darstellung der Ergebnisse. Dadurch wurde eine effiziente Auswertung der umfangreichen Daten (verschiedene Tiere an mehreren Zeitpunkten mit einer Vielzahl an Einzelaufnahmen) ermöglicht und eine unabhängige Bewertung erreicht. Um die Limitationen der begrenzten Auflösung von 2D-Multi-Schichtaufnahmen aufzuheben, wurden neue Ansätze für eine hochaufgelöste 3D-Aufnahme entwickelt. Hierfür wurden Spin-Echo-basierte Sequenzen gewählt, da diese eine genauere Abbildung der Knochenmikrostruktur erlauben als Gradienten-Echo-basierte Methoden. Zum einen wurde eine eigene 3D-FLASE-Sequenz entwickelt und zum anderen eine modifizierte 3D-TSE-Sequenz. Damit an Patienten Aufnahmen bei klinischer Feldstärke von 1,5 T mit einer hohen räumlichen Auflösung innerhalb einer vertretbaren Zeit erzielt werden können, muss eine schnelle und signalstarke Sequenz verwendet werden. Eine theoretische Betrachtung bescheinigte der TSE-Sequenz eine um 25 % höhere Signaleffizienz verglichen mit einer FLASE-Sequenz mit identischer Messzeit. Dieser Unterschied konnte auch im Experiment nachgewiesen werden. Ein in vivo Vergleich der beiden Sequenzen am Schienbein zeigte eine vergleichbare Darstellung der Spongiosa mit einer Auflösung von 160 × 160 × 400 µm. Für die Bildgebung des Hüftkopfs mit der neuen Sequenz waren jedoch aufgrund der unterschiedlichen Anatomie weitere Modifikationen notwendig. Um längere Messzeiten durch ein unnötig großes Field-of-View zu vermeiden, mussten Einfaltungsartefakte unterdrückt werden. Dies wurde durch die orthogonale Anwendung der Anregungs- und Refokussierungspulse in der TSE-Sequenz effizient gelöst. Technisch bedingt konnte jedoch nicht eine vergleichbare Auflösung wie am Schienbein realisiert werden. Der Vorteil der 3D-Bildgebung, dass Schichtdicken von deutlich weniger als 1 mm erreicht werden können, konnte jedoch erfolgreich auf den Unterkiefer übertragen werden. Der dort verlaufende Nervus Mandibularis ist dabei eine wichtige Struktur, deren Verlauf im Vorfeld von verschiedenen operativen Eingriffen bekannt sein muss. Er ist durch eine dünne knöcherne Wand vom umgebenden Gewebe getrennt. Im Vergleich mit einer 3D-VIBE-Sequenz zeigte die entwickelte 3D-TSE-Sequenz mit integrierter Unterdrückung von Einfaltungsartefakten eine ähnlich gute Lokalisierung des Nervenkanals über die gesamte Länge der Struktur. Dies konnte in einer Studie an gesunden Probanden mit verschiedenen Beobachtern nachgewiesen werden. Durch die neue Aufnahmetechnik konnte darüber hinaus die Auflösung im Vergleich zu bisherigen Studien deutlich erhöht werden, was insgesamt eine präzisere Lokalisierung des Nervenkanals erlaubt. Ein Baustein des Tissue Engineerings sind bio-resorbierbare Materialien, deren Abbau- und Einwachsverhalten noch untersucht werden muss, bevor diese für die klinische Anwendung zugelassen werden. Die durchgeführten in vitro µMR-Untersuchungen an Polymerscaffolds zeigten die reproduzierbare Quantifizierung der Porengröße und Wandstärke. Darüber hinaus wurde eine inhomogene Verteilung der Strukturparameter beobachtet. Die Ergebnisse waren in guter Übereinstimmung mit µCT-Aufnahmen als Goldstandard. Unterschiedliche Varianten der Scaffolds konnten identifiziert werden. Dabei bewies sich die MR-Bildgebung als zuverlässige Alternative. Insgesamt zeigen die Ergebnisse dieser Arbeit, welche Vorteile und Anwendungsmöglichkeiten die 3D-MRT-Bildgebung bietet, und dass auch mit klinischer Feldstärke in vivo Voxelgrößen im Submillimeterbereich für alle Raumrichtungen erreichbar sind. Die erzielten Verbesserungen in der räumlichen Auflösung erhöhen die Genauigkeit der verschiedenen Anwendungen und ermöglichen eine bessere Identifikation von kleinen Abweichungen, was eine frühere und zuverlässigere Diagnose für Patienten verspricht. N2 - New tissue engineering based therapies require adjusted diagnostic methods as well as non-invasive therapy monitoring. Especially 3D MR imaging is a promising tool for parameter quantification at high spatial precision. To serve that need new approaches for high resolution in vivo 3D MRI were developed and their applications in combination with tissue engineering have been demonstrated. The advantages of parameter quantification have been demonstrated in a preclinical study of a femoral heck necrosis model in a large animal. Therapy progress has been monitored at different time points. Despite a commonly used 2D imaging protocol a systematic evaluation of signal intensities from T1 and T2-FS weighted images allowed to draw conclusions about changes in bone microstructure. These results were in good agreement with ex vivo µCT images. The observed increase of trabecular thickness were highly correlated with a signal decrease in the T1 weighted images. The radial evaluation of the data allowed a compressed representation of the results. This lead to an efficient evaluation of numerous data (different animals at various time points with huge number of images each) and allowed an observer independent evaluation. To overcome the limitations from the limited spatial resolution in 2D multi slice images, new approaches for a high-resolution 3D imaging were developed. The focus was on spin echo based sequences due to their better representation of bone microstructure compared to gradient echo based sequences. On one hand a 3D FLASE sequence was developed and on the other hand a modified 3D TSE sequence. To achieve a high resolution in vivo at clinical field strength of 1.5 T within a reasonable scan time, a fast and signal intense sequence is strongly required. A theoretical evaluation of signal equations attributed an increase of 25 % to the TSE sequence compared to the FLASE sequence at identical scan time and resolution. This difference was also observed in experimental results. An in vivo comparison of both sequences at the distal tibia showed a comparable depiction of bone microstructure at a resolution of 160 × 160 × 400 µm. To apply this sequence for high resolution imaging of the femoral head, further modifications were necessary due to the different anatomy. A large field of view had to be avoided to reduce the overall scan time, thus aliasing artifacts had to be suppressed. This was achieved by orthogonal application of excitation and refocusing pulses in the TSE sequence. However, due to technical limitations the achievable resolution was lower than at the distal tibia. A slice thickness much smaller than 1 mm is one of the biggest advantages of 3D MRI and this sequence was successfully applied to imaging of the mandible. The course of the mandibular canal must be known before many surgeries, in order to avoid damaging this structure. The canal is separated from the surrounding only by a small bony wall. In comparison to a 3D VIBE sequence the developed 3D TSE sequence with incorporated aliasing suppression showed a comparable good localization of the canal across the full length of the structure. This was demonstrated in a study with various healthy volunteers and different observers. In comparison to previous results the new imaging technique allowed an increase of spatial resolution to a isotropic voxel size of 0.5 mm, which in total provides a higher precision for localizing the nerve canal. One important element in tissue engineering are bio resorbable materials. Their degradation and ingrowth process must be evaluated before they can be approved for clinical application. The performed in vitro µMRstudies at polymer scaffolds showed a reproducible quantification of pore size and wall thickness for different samples. Additionally, an inhomogeneous distribution of parameters in some samples was observed. The results were in good agreement with data based on µCT images, which are considered to be gold standard for this evaluation and showed significant differences between different groups of scaffolds. The results of this work demonstrate the advantages and possible applications of 3D MRI in clinical applications. Even at clinical field strength it is possible to achieve submillimeter resolution for all three spatial dimension within reasonable scan time. The achieved improvements in spatial resolution allow for an improved precision of the different applications as well as a better identification of small local deviations, which promises an earlier and more reliable diagnosis for patients. KW - Kernspintomografie KW - Mikrostruktur KW - Knochen KW - hochauflösende Bildgebung KW - 3D-Bildgebung KW - Knochenstruktur KW - Spin-Echo KW - Trabekel KW - Hüftkopfnekrose KW - Tissue Engineering Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-168858 ER - TY - THES A1 - Riekert, Maximilian T1 - Wertigkeit klinischer, instrumenteller und bildgebender Untersuchungsverfahren der Kiefergelenksdiagnostik bei Patienten mit juveniler idiopathischer Arthritis T1 - Validity of clinical, instrumental and imaging examination methods of temporomandibular joint diagnosis in patients with juvenile idiopathic arthritis N2 - Zusammenfassung Ziel: Pathomorphologische Veränderungen der Kiefergelenke treten im Rahmen einer JIA häufig auf. In dieser Arbeit sollten anhand von einer kieferorthopädischen zweidimensionalen Röntgenbildgebung (Orthopantomographie - OPG) pathologische Veränderungen der Kiefergelenke bei JIA-Patienten differenziert und Asymmetrien des Unterkiefers in Abhängigkeit vom Grad der Kondylendestruktion bestimmt werden. Darüber hinaus sollte geprüft werden, wie sich die Krankheitsdauer auf den Befall der Kiefergelenke auswirkt. Zusätzlich wurden zur differenziellen Analyse der Pathomorphologie die klinische Funktionsanalyse (FAL), die Joint Vibration Analysis (JVA) sowie die 3D-Stereophotogrammetrie (3d-Scan) eingesetzt. Ziel war mittels non-invasiver, kostengünstiger und schnell verfügbarer Untersuchungsmethoden eindeutige Parameter zum Befall der Kiefergelenke zu detektieren. Patienten und Methodik: In dieser Arbeit wurden 46 Patientin (28 weiblich; 18 männlich) kaukasischer Herkunft mit diagnostizierter JIA eingeschlossen. Die Kiefergelenke (n = 92) wurden einzeln nach dem Grad ihrer Kondylendestruktion (Grad 0 - 4 nach Billiau et. al. [78]) befundet und in eine leicht betroffene Gruppe 1 (Grad 0, 1 und 2 nach Billiau: röntgenologisch unauffällig, Kondylenerosionen, Kondylenabflachungen) und in eine schwer betroffene Gruppe 2 (Grad 3 und 4 nach Billiau: Kondylusabflachungen mit Erosionen, Komplettverlust des Kondylus) eingeteilt. Zur Quantifizierung von Unterkieferasymmetrien wurde das Seitenverhältnis aus Kondylus-, Ramus- und Mandibulahöhe bestimmt. Der Vergleich der einzelnen klinischen, instrumentellen und bildgebenden Untersuchungsverfahren (OPG, FAL, JVA, 3d-Scan) erfolgte jeweils durch die Gegenüberstellung der schwer betroffenen und der leicht betroffenen Patientengruppe. � Ergebnisse: Erkrankungsdauer: Anhand des Grades der Kondylendestruktion wurden 36 Patienten in die leicht betroffene Gruppe 1 und 10 Patienten in die schwer betroffenen Gruppe 2 eingeteilt. Dabei war die Erkrankungsdauer in der schwer betroffenen Patientengruppe (8,9 ± 5,2 Jahre) signifikant länger, als in der leicht betroffenen Patientengruppe (4,6 ± 4,7; Jahre) (p = 0,031). FAL: Die Ergebnisse der FAL zeigten ausgeprägtere funktionelle Abweichungen in der schwer betroffenen Patientengruppe (Gruppe 2). Es wurde jedoch kein signifikanter Unterschied zur Gruppe 1 ermittelt. Die schwer betroffene Patientengruppe zeigten einen höheren Prozentsatz an von Schmerzen bei Palpation der Kiefergelenke (Gruppe 2: 70,0 % vs. Gruppe 1: 61,1 %) oder Schmerzen bei Mundöffnung (Gruppe 2: 10,0% vs. Gruppe 1: 8,3 %), Deflexionen des Unterkiefers (Gruppe 2: 50,0% vs. Gruppe 1: 33,3 %), Gelenkgeräuschen (Gruppe 2: 80,0 % vs. Gruppe 1: 63,9 %) und Mundöffnungseinschränkungen (Gruppe 2: 60,0 % vs. Gruppe 1: 25,0 %). Die durchschnittliche Mundöffnung betrug in Gruppe 2 40,6 mm, während sie in Gruppe 1 43,5 mm betrug. Bei Patienten mit einer Mundöffnung < 40 mm wurde in Gruppe 2 eine durchschnittliche Mundöffnung von 35,3 mm und in Gruppe 1 von 34,1 mm gemessen. JVA: Sowohl in der gelenkbezogenen, als auch in der patientengruppenbezogenen Analyse der JVA deuteten die Messparameter in der schwer betroffenen Patientengruppe vermehrt auf chronisch-degenerative oder bestehende Ergüsse im Kiefergelenk hin. In der gelenkbezogenen Auswertung zeigte sich dies insbesondere durch eine reduzierte Signalstärke in der schwer betroffenen Patientengruppe (Total power: p = 0,005; Power < 300 Hz: p = 0,006; Power > 300 Hz: 0,003;) sowie in einer signifikant erhöhten Peak Frequenz (p = 0,036). OPG: In der Auswertung der OPGs war die Ratio von Kondylus-, Ramus- und Mandibulahöhe in der schwer Patientengruppe (Ratio 79,6 %, 85,9 %, 86,5 %) signifikant kleiner (Kondylushöhe: p = 0,0005; Ramushöhe: p = 0,0030; Mandibulahöhe: p = 0,0002), als in der leicht betroffenen Patientengruppe (Ratio 93,8 %, Ratio 96,0 %, 95,6 %). Somit ergaben sich in der schwer betroffenen Patientengruppe signifikant stärker ausgeprägte Unterkieferasymmetrien, als in der leicht betroffenen Patientengruppe. 3d-Scan: Im 3d-Scan kam es bei Patienten mit schwer betroffenen Kiefergelenken signifikant häufiger zu Abweichungen des Weichteilkinns von der Medianebene (Gruppe 2: 3,0 mm vs. Gruppe 1: 1,2 mm; p = 0,041) und zu Asymmetrien des Unterkiefers (Gruppe 2: 62,5 % vs. Gruppe 1: 14,8 %; p = 0,015) als bei Patienten mit leicht betroffenem Kiefergelenk. Schlussfolgerung: Es zeigte sich, dass mittels einfacher und schnell verfügbarer Untersuchungsmethoden wie der klinischen Funktionsanalyse, der Joint Vibration Analysis und der OPG-Aufnahme eine Pathologie im Kiefergelenk dargestellt werden kann. Die Methoden können als wichtige Referenz zur Kontrolle der Krankheitsprogression bei Patienten mit JIA dienen. Darüber hinaus ist eine Klassifikation der Kondylen in schwer und leicht betroffene Kiefergelenke mittels pathomorphologischer Analyse möglich. Dabei ist von einem direkten Zusammenhang zwischen Grad der Destruktion, Ausmaß der Unterkieferasymmetrie und Dauer der Erkrankung bei Patienten mit JIA auszugehen. Insgesamt konnte die Wertigkeit klinischer, instrumenteller und bildgebender Untersuchungsverfahren der Kiefergelenksdiagnostik bei Patienten mit juveniler idiopathischer Arthritis dargestellt werden. N2 - Summary Aim: Pathomorphological changes of the temporomandibular joints occur frequently in patients with JIA. In this study, orthodontic two-dimensional X-ray imaging (orthopantomography - OPG) was used to differentiate pathological changes of the temporomandibular joints in JIA patients and to determine asymmetries of the mandible in accordance to the degree of condyle destruction. In addition, it should be examined how the disease duration affects the involvement of the temporomandibular joints. In addition, clinical analysis (FAL), joint vibration analysis (JVA) and 3D stereophotogrammetry (3d-scan) were used for the specific analysis of pathomorphology. The aim was to detect unambiguous parameters for affected temporomandibular joints by means of non-invasive, cost-effective and rapidly available examination methods. Patients and methods: In this study, 46 patients (28 female, 18 male) of Caucasian origin were diagnosed with JIA. The temporomandibular joints (n = 92) were individually evaluated based on the degree of their condylar destruction (Grades 0 - 4 according to Billiau et al. [78]) and divided into a slightly affected group 1 (Grades 0, 1 and 2 according to Billiau: radiologically unremarkable, Condylar erosions, condyle flattening) and into a severely affected group 2 (grade 3 and 4 according to Billiau: condyle flattening with erosions, complete loss of the condyle). To quantify mandibular asymmetries, the ratio of condyle, ramus and mandibular height was determined. The comparison of the individual clinical, instrumental and imaging examination procedures (OPG, FAL, JVA, 3d-Scan) was performed by comparing the severely affected and the slightly affected patient group. Results: Disease duration: Based on the degree of condylar destruction, 36 patients were divided into the slightly affected group 1 and 10 patients to the severely affected group 2. The disease duration was significantly longer in the severely affected patient group (8.9 ± 5.2 years) than in the slightly affected patient group (4.6 ± 4.7 years) (p = 0.031). FAL: The results of the FAL showed more functional deviations in the severely affected patient group (group 2). However, no significant difference to Group 1 was found. The severely affected patients showed a higher percentage of pain in palpation of the temporomandibular joints (group 2: 70.0% vs. group 1: 61.1%) or mouth opening pain (group 2: 10.0% vs. group 1: 8.3%), mandibular deflections (group 2: 50.0% vs. group 1: 33.3%), joint noises (group 2: 80.0% vs. group 1: 63.9%) and mouth opening restrictions (Group 2: 60.0% vs. Group 1: 25.0%). The average mouth opening in group 2 was 40.6 mm, while in group 1 it was 43.5 mm. In patients with a mouth opening <40 mm, an average mouth opening of 35.3 mm was measured in group 2 and 34.1 mm in group 1. JVA: In the joint-related as well as in the patient-group-related analysis of the JVA, the measurement parameters in the severely affected patient group increasingly pointed to chronic degenerative or existing effusions in the temporomandibular joint. In the joint-related evaluation, this was demonstrated in particular by a reduced signal strength in the severely affected patient group (total power: p = 0.005, power <300 Hz: p = 0.006, power> 300 Hz: 0.003;) and in a significantly increased peak frequency (p = 0.036). OPG: In the evaluation of the OPGs, the ratio of condyle, ramus and mandibular height was significantly lower in the severe patient group (ratio 79.6%, 85.9%, 86.5%) (condyle height: p = 0.0005; p = 0.0030, mandibular height: p = 0.0002), than in the slightly affected Patient group (ratio 93.8%, ratio 96.0%, 95.6%). Thus, significantly more pronounced mandibular asymmetries were found in the severely affected patient group than in the slightly affected patient group. 3d scan: In the 3d scan, deviations of the soft tissue chin from the median plane (group 2: 3.0 mm vs. group 1: 1.2 mm, p = 0.041) and mandibular asymmetry were significantly more frequent in patients with severely affected temporomandibular joints (group 2: 62.5% vs. Group 1: 14.8%, p = 0.015) than in patients with slightly affected temporomandibular joints. Conclusion: It has been shown that it is possible to visualize pathology in the temporomandibular joint by means of simple and easily available examination methods such as clinical functional analysis, joint vibration analysis and OPG imaging. The methods can serve as an important reference for controlling disease progression in patients with JIA. In addition, a classification of the condyles in severe and slightly affected temporomandibular joints by means of pathomorphological analysis is possible. There is a direct correlation between the degree of destruction, the extent of mandibular asymmetry and the duration of the disease in patients with JIA. Overall, the value of clinical, instrumental and imaging examination methods of temporomandibular joint diagnosis in patients with juvenile idiopathic arthritis was demonstrated. KW - Kiefergelenkkrankheit KW - Juvenile idiopathische Arthritis KW - Kiefergelenk KW - Orthopantomogramm KW - Kernspintomografie KW - Kiefergelenksarthritis KW - Imaging KW - TMJ KW - Klinische Funktionsanalyse KW - Joint Vibration Analysis KW - Stereophotogrammetrie Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-163762 ER - TY - THES A1 - Slawig, Anne T1 - Reconstruction methods for the frequency-modulated balanced steady-state free precession MRI-sequence T1 - Rekonstruktionsmethoden für die frequenz-modulierte balanced steady-state free precession MRT-Sequenz N2 - This work considered the frequency-modulated balanced steady-state free precession (fm-bSSFP) sequence as a tool to provide banding free bSSFP MR images. The sequence was implemented and successfully applied to suppress bandings in various in vitro and in vivo examples. In combination with a radial trajectory it is a promising alternative for standard bSSFP applications. First, two specialized applications were shown to establish the benefits of the acquisition strategy in itself. In real time cardiac imaging, it was shown that the continuous shift in frequency causes a movement of the bandings across the FOV. Thus, no anatomical region is constantly impaired, and a suitable timeframe can be found to examine all important structures. Furthermore, a combination of images with different artifact positions, similar to phase-cycled acquisitions is possible. In this way, fast, banding-free imaging of the moving heart was realized. Second, acquisitions with long TR were shown. While standard bSSFP suffers from increasing incidence of bandings with higher TR values, the frequency-modulated approach provided banding free images, regardless of the TR. A huge disadvantage of fm-bSSFP, in combination with the radial trajectory, is the decrease in signal intensity. In this work a specialized reconstruction method, the multifrequency reconstruction for frequency-modulated bSSFP (Muffm), was established, which successfully compensated that phenomena. The application of Muffm to several anatomical sites, such as inner ear, legs and cardiac acquisitions, proofed the advantageous SNR of the reconstruction. Furthermore, fm-bSSFP was applied to the clinically highly relevant task of water-fat separation. Former approaches of a phase-sensitive separation procedure in combination with standard bSSFP showed promising results but failed in cases of high inhomogeneity or high field strengths where banding artifacts become a major issue. The novel approach of using the fm-bSSFP acquisition strategy with the separation approach provided robust, reliable images of high quality. Again, losses in signal intensity could be regained by Muffm, as both approaches are completely compatible. Opposed to conventional banding suppression techniques, like frequency-scouts or phase-cycling, all reconstruction methods established in this work rely on a single radial acquisition, with scan times similar to standard bSSFP scans. No prolonged measurement times occur and patient time in the scanner is kept as short as possible, improving patient comfort, susceptibility to motion or physiological noise and cost of one scan. All in all, the frequency-modulated acquisition in combination with specializes reconstruction methods, leads to a completely new quality of images with short acquisition times. N2 - In dieser Arbeit wird eine Modifikation der balanced steady-state free precession (bSSFP) Sequenz betrachtet. Die frequenzmodulierte bSSFP-Sequenz (fm-bSSFP) kann die sonst typischen Band-Artefakte in bSSFP-MR-Bildern verhindern. Die Sequenz wurde im Rahmen der Arbeit am MR-Scanner implementiert und erfolgreich in verschiedenen in-vitro- und in-vivo-Beispielen angewendet. In Kombination mit einer radialen Trajektorie erwies es sich als eine vielversprechende Alternative für alle Standard-bSSFP Anwendungen. Zuerst wurden zwei spezialisierte Anwendungen gezeigt, um die Vorteile der Akquisitionsstrategie an sich darzustellen. Am Beispiel der Echtzeit-Herzbildgebung konnte mit Hilfe der kontinuierlichen Frequenzverschiebung eine Bewegung der Bänder über das FOV erzeugt werden. Somit wird keine anatomische Region ständig von Artefakten überlagert und für jeden Bereich kann ein geeigneter Zeitrahmen gefunden werden, um die wichtigen Strukturen darzustellen und zu untersuchen. Darüber hinaus ist eine Kombination von Bildern mit verschiedenen Artefaktpositionen möglich, ähnlich zu mehreren Aufnahmen mit verschiedenen Phasenzyklen. Auf diese Weise wurde eine schnelle Bildgebung des sich bewegenden Herzens ohne Bandartefakte realisiert. Zusätzlich wurden Aufnahmen mit langen Repetitionszeiten (TR) untersucht. Während in der Standard-bSSFP die Häufigkeit von Bandartefakten mit steigendem TR-Wert zunimmt, lieferte der frequenzmodulierte Ansatz Banding-freie Bilder unabhängig vom TR. Ein großer Nachteil von fm-bSSFP in Kombination mit der radialen Trajektorie ist der Verlust von Signalintensität bei der Rekonstruktion. In dieser Arbeit wurde eine spezielle Rekonstruktionsmethode namens Muffm (mulitfrequency reconstruction for frequency-modulated bSSFP) etabliert, die diesen Verlust erfolgreich kompensieren kann. Die Anwendung von Muffm an verschiedenen anatomischen Strukturen, wie Innenohr, Bein und Herzaufnahmen, bestätigte das vorteilhafte Signal-zu-Rausch-Verhältnis, dass durch die spezielle Rekonstruktion gewonnen werden kann. Darüber hinaus wurde die fm-bSSFP auf die klinisch interessante Wasser-Fett-Trennung angewandt. Frühere Ansätze eines phasenempfindlichen Trennverfahrens in Kombination mit Standard-bSSFP zeigten vielversprechende Ergebnisse, scheiterten jedoch in Fällen hoher Inhomogenität oder hoher Feldstärken an den auftretenden Bandartefakten. Der neue Ansatz, diesen Separationsalgorithmus mit der fm-bSSFP-Akquisitionsstrategie zu verbinden, lieferte robuste, zuverlässige Bilder von hoher Qualität. Auch hier konnten entstehende Verluste in der Signalintensität durch Muffm zurückgewonnen werden, da beide Ansätze vollständig kompatibel sind. Im Gegensatz zu herkömmlichen Bandunterdrückungstechniken, wie Frequenz-Scouts oder die Aufnahme mehrerer Bilder mit verschiedenen Phasenzyklen, beruhen alle in dieser Arbeit etablierten Rekonstruktionsverfahren auf einer einzigen radialen Aufnahme. Die Messzeiten sind daher identisch zur Aufnahme einer Standard-bSSFP Messung. Das Verfahren ermöglicht eine deutliche Verkürzung der Aufenthaltsdauer im Scanner bei einer gleichzeitigen Garantie ein artefaktfreies Bild zu erhalten. Damit ist es insbesondere für Patienten von Vorteil, die unter Platzangst oder sonstigen Beschwerden leiden, die ein langes Stillliegen erschweren. Außerdem werden Bewegungsartefakte, physiologisches Rauschen und nicht zuletzt die Kosten eines Scans minimiert. Insgesamt bietet die frequenzmodulierte bSSFP Aufnahme in Kombination mit spezialisierten Rekonstruktionsverfahren neue Möglichkeiten zur schnellen Aufnahme von Bildern ohne Bandartefakte. KW - Kernspintomografie KW - Magnetic resonance imaging KW - MRI KW - MRT KW - bSSFP KW - Rekonstruktion KW - reconstruction KW - frequency modulation KW - water fat separation KW - Wasser Fett Trennung KW - balanced steady state free precession Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-162871 ER - TY - THES A1 - Ponce Garcia, Irene Paola T1 - Strategies for optimizing dynamic MRI T1 - Strategien zur Optimierung der dynamischen MR Bildgebung N2 - In Magnetic Resonance Imaging (MRI), acquisition of dynamic data may be highly complex due to rapid changes occurred in the object to be imaged. For clinical diagnostic, dynamic MR images require both high spatial and temporal resolution. The speed in the acquisition is a crucial factor to capture optimally dynamics of the objects to obtain accurate diagnosis. In the 90’s, partially parallel MRI (pMRI) has been introduced to shorten scan times reducing the amount of acquired data. These approaches use multi-receiver coil arrays to acquire independently and simultaneously the data. Reduction in the amount of acquired data results in images with aliasing artifacts. Dedicated methods as such Sensitivity Encoding (SENSE) and Generalized Autocalibrating Partially Parallel Acquisition (GRAPPA) were the basis of a series of algorithms in pMRI. Nevertheless, pMRI methods require extra spatial or temporal information in order to optimally reconstruct the data. This information is typically obtained by an extra scan or embedded in the accelerated acquisition applying a variable density acquisition scheme. In this work, we were able to reduce or totally eliminate the acquisition of the training data for kt-SENSE and kt-PCA algorithms obtaining accurate reconstructions with high temporal fidelity. For dynamic data acquired in an interleaved fashion, the temporal average of accelerated data can generate an artifact-free image used to estimate the coil sensitivity maps avoiding the need of extra acquisitions. However, this temporal average contains errors from aliased components, which may lead to signal nulls along the spectra of reconstructions when methods like kt-SENSE are applied. The use of a GRAPPA filter applied to the temporal average reduces these errors and subsequently may reduce the null components in the reconstructed data. In this thesis the effect of using temporal averages from radial data was investigated. Non-periodic artifacts performed by undersampling radial data allow a more accurate estimation of the true temporal average and thereby avoiding undesirable temporal filtering in the reconstructed images. kt-SENSE exploits not only spatial coil sensitivity variations but also makes use of spatio-temporal correlations in order to separate the aliased signals. Spatio-temporal correlations in kt-SENSE are learnt using a training data set, which consists of several central k-space lines acquired in a separate scan. The scan of these extra lines results in longer acquisition times even for low resolution images. It was demonstrate that limited spatial resolution of training data set may lead to temporal filtering effects (or temporal blurring) in the reconstructed data. In this thesis, the auto-calibration for kt-SENSE was proposed and its feasibility was tested in order to completely eliminate the acquisition of training data. The application of a prior TSENSE reconstruction produces the training data set for the kt-SENSE algorithm. These training data have full spatial resolution. Furthermore, it was demonstrated that the proposed auto-calibrating method reduces significantly temporal filtering in the reconstructed images compared to conventional kt-SENSE reconstructions employing low resolution training images. However, the performance of auto-calibrating kt-SENSE is affected by the Signal-to-Noise Ratio (SNR) of the first pass reconstructions that propagates to the final reconstructions. Another dedicated method used in dynamic MRI applications is kt-PCA, that was first proposed for the reconstruction of MR cardiac data. In this thesis, kt-PCA was employed for the generation of spatially resolved M0, T1 and T2 maps from a single accelerated IRTrueFISP or IR-Snapshot FLASH measurement. In contrast to cardiac dynamic data, MR relaxometry experiments exhibit signal at all temporal frequencies, which makes their reconstruction more challenging. However, since relaxometry measurements can be represented by only few parameters, the use of few principal components (PC) in the kt-PCA algorithm can significantly simplify the reconstruction. Furthermore, it was found that due to high redundancy in relaxometry data, PCA can efficiently extract the required information from just a single line of training data. It has been demonstrated in this thesis that auto-calibrating kt-SENSE is able to obtain high temporal fidelity dynamic cardiac reconstructions from moderate accelerated data avoiding the extra acquisition of training data. Additionally, kt-PCA has been proved to be a suitable method for the reconstruction of highly accelerated MR relaxometry data. Furthermore, a single central training line is necessary to obtain accurate reconstructions. Both reconstruction methods are promising for the optimization of training data acquisition and seem to be feasible for several clinical applications. N2 - Dynamische Bildgebung mithilfe der Magnetresonanztomographie stellt eine besondere Herausforderung dar. Für klinische Anwendungen benötigt man Bilder mit hoher räumlicher und bei schnellen Bewegungen auch zeitlicher Auflösung. Technologische Fortschritte in den letzten Jahrzehnten konnten MRT-Experimente erheblich beschleunigen. Ein wichtiger Beitrag lieferte die parallele Bildgebung (pMRT), die durch die Entwicklung von Spulenarrays für den Empfang des MR-Signals ermöglicht wurde. In paralleler Bildgebung wird nur ein Teil der eigentlich benötigten Daten aufgenommen. Diese Unterabtastung des k-Raum führt zu Bildern mit Aliasing-Artefakten. Verschiedenste Algorithmen wurden entwickelt, um mittels der zusätzlichen räumlichen Informationen der Spulenarrays anschließend Bilder zu rekonstruieren. Heute spielen Sensitivity Encoding (SENSE) und Generalized Autocalibrating Partially Parallel Acquisition (GRAPPA) eine große und bilden eine Grundlage für eine Vielzahl anderer Algorithmen. Einen Großteil aller pMRT Methoden erfordern für optimale Ergebnisse zusätzliche räumliche oder zeitliche Informationen zur Kalibrierung. Diese Kalibrations- oder Trainingsdaten werden in der Regel durch einen zusätzlichen Scan erzeugt oder in die beschleunigte Messung eingebettet aufgenommen. Das ist eine unerwünschte Verlängerung der Messzeit die Folge. In dieser Arbeit konnten wir kt-SENSE und kt-PCA Rekonstruktionen dynamischer MRT Daten mit hoher zeitlicher Genauigkeit erzielen bei gleichzeitiger Reduktion bzw. sogar Beseitigung der benötigten Menge an Trainingsdaten. Um die in beiden Methoden benötigten Spulensensitivitäten zu berechnen, kann bei bestimmten Abtastschemata mit dem Mittelwert der dynamischen Daten ein weitgehend Artefakt-freies Bild erzeugt werden. Dieser zeitliche Mittelwert enthält jedoch kleine Fehler, die durch die Anwendung von Methoden wie kt-SENSE zu Signalauslöschungen im Spektrum der rekonstruierten Daten führen können. Es konnte gezeigt werden, dass die Anwendung eines GRAPPA Filter auf den zeitlichen Mittelwert die Fehler in den Spulensensitivitäten reduziert und die Rekonstruktion von Daten aller Frequenzen ermöglicht. Eine weitere aufgezeigte Möglichkeit ist die Verwendung einer radialen Akquisition, die dank der inkohärenten Aliasing-Artefakte ebenfalls zu einer erheblich genaueren Abschätzung des zeitlichen Mittelwerts führt. Dies verhindert zeitliche Ungenauigkeiten in den rekonstruierten Bildern. Zusätzliche zu Spulensensitivitäten verwenden Rekonstruktionsmethoden wie kt-SENSE Vorkenntnisse über räumlich-zeitliche Korrelationen, um Artefakte zu entfernen. Informationen darüber werden gewöhnlich aus voll aufgenommenen Trainingsdaten mit geringer Auflösung extrahiert. Die separate Akquisitions dieser Trainingsdaten verursacht eine unerwünschte Verlängerung der Messzeit. In dieser Arbeit wurde gezeigt, dass die niedrige Auflösung der Trainingsdaten zu zeitlichen Filterungseffekten in den rekonstruierten Daten führen kann. Um dies zu verhindern und um die zusätzliche Aufnahme von Trainingsdaten zu vermeiden, wurde eine Autokalibrierung für kt-SENSE vorgeschlagen und untersucht. Hierbei werden die benötigten Trainingsdaten in einem ersten Schritt durch eine TSENSE Rekonstruktion aus den unterabgetasteten Daten selbst erzeugt. Dank der vollen Auflösung dieser Trainingsdaten kann das Auftreten eines zeitlichen Filters verhindert werden. Die Leistung der Auto-kalibration wird lediglich durch einen Einfluss des SNRs der TSENSE Trainingsdaten auf die finalen Rekonstruktionen beeinträchtigt. Ein weiteres Verfahren für die dynamische MRT ist kt-PCA, das zunächst für die Rekonstruktion von MR-Herzdaten vorgeschlagen wurde. In dieser Arbeit wurde kt-PCA für die neurologische MR Relaxometrie verwendet. Hierbei konnten aus beschleunigten IRTrueFISP und IR-Snapshot-FLASH Messungen genaue M0, T1 und T2 Karten gewonnen werden. Im Gegensatz zur Herzbildgebung weisen MR Relaxometrie Datensätze Signal auf alles zeitlichen Frequenzen auf, was ihre Rekonstruktion mit konventionellen Methoden erschwert. Andererseits können die zeitlichen Signalverläufe mit einigen wenigen Parametern dargestellt werden und die Rekonstruktion mittels kt-PCA vereinfacht sich erheblich aufgrund der geringen Anzahl benötigter Hauptkomponenten (PC). Weiter wurde gezeigt, dass aufgrund der hohen Redundanz ein Trainingsdatensatz bestehend aus einer einzigen Zeile ausreicht, um alle relevanten Informationen zu erhalten. In dieser Thesis wurde demonstriert, dass mit dem Ansatz einer auto-kalibrierten kt-SENSE Rekonstruktion Bilder mit hoher zeitlicher Genauigkeit aus beschleunigten Datensätzen des Herzens gewonnen werden können. Dies vermeidet die gewöhnlich benötigte zusätzliche Aufnahme von Trainingsdaten. Weiterhin hat sich kt-PCA als geeignetes Verfahren zur Rekonstruktion hochbeschleunigter MR Relaxometrie Datensätze erwiesen. In diesem Fall war ein Trainingsdatensatz bestehend aus einer einzelnen Zeile ausreichend für Ergebnisse mit hoher Genauigkeit. KW - Kernspintomografie KW - Dynamische Messung KW - Magnetic resonance KW - Magnetische Resonanz KW - Dynamic magnetic resonance imaging KW - Dynamische MR Bildgebung KW - DNMR-Spektroskopie KW - Bildgebendes Verfahren Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-162622 ER - TY - THES A1 - Kraus, Philip T1 - Verbesserung von Echoplanarer Bildgebung durch Phasenkorrektur T1 - Improvement of echoplanar imaging by phase correction N2 - Die Arbeit liefert eine Übersicht zu möglichen Korrekturen dynamischer Off-Resonanzen in dichtegewichteten und kartesischen echoplanaren funktionellen MRT Sequenzen. N2 - This scientific work gives an overview about potential corrections of dynamic off-resonances in k-space in density weighted and cartesian MRI sequences. KW - Kernspintomografie KW - Echoplanare Bildgebung KW - Phasenkorrektur KW - MRT KW - Off.Resonanzen Y1 - 2017 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-154462 ER - TY - THES A1 - Carinci, Flavio T1 - Quantitative Characterization of Lung Tissue Using Proton MRI T1 - Quantitative Charakterisierung des Lungengewebes mithilfe von Proton-MRT N2 - The focus of the work concerned the development of a series of MRI techniques that were specifically designed and optimized to obtain quantitative and spatially resolved information about characteristic parameters of the lung. Three image acquisition techniques were developed. Each of them allows to quantify a different parameter of relevant diagnostic interest for the lung, as further described below: 1) The blood volume fraction, which represents the amount of lung water in the intravascular compartment expressed as a fraction of the total lung water. This parameter is related to lung perfusion. 2) The magnetization relaxation time T\(_2\) und T� *\(_2\) , which represents the component of T\(_2\) associated with the diffusion of water molecules through the internal magnetic field gradients of the lung. Because the amplitude of these internal gradients is related to the alveolar size, T\(_2\) und T� *\(_2\) can be used to obtain information about the microstructure of the lung. 3) The broadening of the NMR spectral line of the lung. This parameter depends on lung inflation and on the concentration of oxygen in the alveoli. For this reason, the spectral line broadening can be regarded as a fingerprint for lung inflation; furthermore, in combination with oxygen enhancement, it provides a measure for lung ventilation. N2 - Die Magnetresonanztomographie (MRT) stellt ein einzigartiges Verfahren im Bereich der diagnostischen Bildgebung dar, da sie es ermöglicht, eine Vielzahl an diagnostischen Informationen ohne die Verwendung von ionisierenden Strahlen zu erhalten. Die Anwendung von MRT in der Lunge erlaubt es, räumlich aufgelöste Bildinformationen über Morphologie, Funktionalität sowie über die Mikrostruktur des Lungengewebes zu erhalten und diese miteinander zu kombinieren. Für die Diagnose und Charakterisierung von Lungenkrankheiten sind diese Informationen von höchstem Interesse. Die Lungenbildgebung stellt jedoch einen herausfordernden Bereich der MRT dar. Dies liegt in der niedrigen Protondichte des Lungenparenchyms begründet sowie in den relativ kurzen Transversal- Relaxationszeiten T\(_2\) und T� *\(_2\) , die sowohl die Bildau� ösung als auch das Signal-zu-Rausch Verhältnis beeinträchtigen. Des Weiteren benötigen die vielfältigen Ursachen von physiologischer Bewegung, welche die Atmung, den Herzschlag und den Blut� uss in den Lungengefasen umfassen, die Anwendung von schnellen sowie relativ bewegungsunemp� ndlichen Aufnahmeverfahren, um Risiken von Bildartefakten zu verringern. Aus diesen Gründen werden Computertomographie (CT) und Nuklearmedizin nach wie vor als Goldstandardverfahren gehandhabt, um räumlich aufgelöste Bildinformationen sowohl über die Morphologie als auch die Funktionalität der Lunge zu erhalten. Dennoch stellt die Lungen- MRT aufgrund ihrer Flexibilität sowohl eine vielversprechende Alternative zu den anderen Bildgebungsverfahren als auch eine mögliche Quelle zusätzlicher diagnostischer Informationen dar. ... KW - Lung KW - MRI KW - Kernspintomografie KW - Lunge Y1 - 2017 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-151189 ER - TY - THES A1 - Breuer [geb. Hemberger], Kathrin R. F. T1 - Effiziente 3D Magnetresonanzbildgebung schnell abfallender Signale T1 - Efficient 3D Magnetic Resonance Imaging of fast decaying signals N2 - In der vorliegenden Arbeit wird die Rotated-Cone-UTE-Sequenz (RC-UTE), eine 3D k-Raum-Auslesetechnik mit homogener Verteilung der Abtastdichte, vorgestellt. Diese 3D MR-Messtechnik ermöglicht die für die Detektion von schnell abfallenden Signalen notwendigen kurzen Echozeiten und weist eine höhere SNR-Effizienz als konventionelle radiale Pulssequenzen auf. Die Abtastdichte ist dabei in radialer und azimutaler Richtung angepasst. Simulationen und Messungen in vivo zeigen, dass die radiale Anpassung das T2-Blurring reduziert und die SNR-Effizienz erhöht. Die Drehung der Trajektorie in azimutale Richtung ermöglicht die Reduzierung der Unterabtastung bei gleicher Messzeit bzw. eine Reduzierung der Messzeit ohne Auflösungsverlust. Die RC-UTE-Sequenz wurde erfolgreich für die Bildgebung des Signals des kortikalen Knochens und der Lunge in vivo angewendet. Im Vergleich mit der grundlegenden UTE-Sequenz wurden die Vorteile von RC-UTE in allen Anwendungsbeispielen aufgezeigt. Die transversalen Relaxationszeit T2* des kortikalen Knochen bei einer Feldstärke von 3.0T und der Lunge bei 1.5T und 3.0T wurde in 3D isotroper Auflösung gemessen. Außerdem wurde die Kombination von RC-UTE-Sequenz mit Methoden der Magnetisierungspräparation zur besseren Kontrasterzeugung gezeigt. Dabei wurden die Doppel-Echo-Methode, die Unterdrückung von Komponenten mit langer Relaxationszeit T2 durch Inversionspulse und der Magnetisierungstransfer-Kontrast angewendet. Die Verwendung der RC-UTE-Sequenz für die 3D funktionelle Lungenbildgebung wird ebenfalls vorgestellt. Mit dem Ziel der umfassenden Charakterisierung der Lungenfunktion in 3D wurde die simultane Messung T1-gewichteter Bilder und quantitativer T2*-Karten für verschiedene Atemzustände an sechs Probanden durchgeführt. Mit der hier vorgestellten Methode kann die Lungenfunktion in 3D über T1-Wichtung, quantitative T2*-Messung und Rekonstruktion verschiedener Atemzustände durch Darstellung von Ventilation, Sauerstofftransport und Volumenänderung beurteilt werden. N2 - In this thesis the Rotated-Cone-UTE-sequence (RC-UTE), a 3D k- space sampling scheme with uniform sampling density, is presented. 3D RC-UTE provides short echo times enabling the detection of fast decaying signals with higher SNR-efficiency than conventional UTE sequences. In RC-UTE the sampling density is adapted in radial and azimuthal direction. It is shown in simulations and measurements that the density adaption along the radial dimension reduces T2-blurring. By twisting the trajectory along the azimuthal direction fewer projections are needed to fulfill the Nyquist criterion. Thereby, undersampling artefacts or the measurement time is reduced without loss of resolution. RC-UTE has been successfully applied in vivo in cortical bone and the lung. It was shown that the RC-UTE sequence outperforms the standard UTE sequence in all presented applications. In addition, the transversal relaxation time T2* of cortical bone at field strength of 3.0T and the human lung at 1.5T und 3.0T was measured in 3D isotropic resolution. Moreover, the combination of RC-UTE with magnetization preparation techniques for improved image contrast was shown. To this end strategies such as double-echo readout, long T2 suppression by inversion pulses and magnetization transfer contrast imaging were employed. Furthermore, the application of RC-UTE for 3D functional lung imaging is presented. In order to provide broad information about pulmonary function T1-weighted images and quantitative T2*-maps in different breathing states were simultaneously measured in six healthy volunteers. The presented methodology enables the assessment of pulmonary function in 3D by indicating ventilation, oxygen transfer and lung volume changes during free breathing. KW - Kernspintomografie KW - Relaxationszeit KW - Dreidimensionale Bildverarbeitung KW - T2* KW - Ulrakurze Echozeit KW - T1-Wichtung KW - dichteangepasste k-Raum Abtastung KW - Lunge KW - Relaxation KW - Lungenfunktion Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-150750 ER - TY - THES A1 - Schneckenburger, Manuel T1 - Was bedeutet eine Änderung der Kontrastmittelaufnahme in niedrig malignen Gliomen bei Kindern? T1 - Significance of contrast enhancement in low grade gliomas in children N2 - Bei niedriggradigen Gliomen WHO I° und II° ist das Kontrastmittelverhalten variabel und nicht mit der Prognose assoziiert. Andererseits wurde bei spontanen Regressionen von niedriggradigen Gliomen bei Patienten sowohl mit als auch ohne eine Neurofibromatose Typ I eine Abnahme der Kontrastmittelaufnahme berichtet. Vielleicht deswegen verleitet eine Zunahme des Enhancements oder ein neu aufgetretenes Enhancement nicht selten zur Diagnose einer Tumorprogression. Es stellt sich also die Frage, ob eine Kontrastmittelaufnahme bei LGGs mit einer Größenzunahme somit einem Tumorprogress assoziiert ist. Es lässt sich eine überdurchschnittliche Assoziation einer Änderung der Kontrastmittelaufnahme mit dem Wachstumsverhalten bei LGGs nachweisen. Allerdings ist das Maß für eine Progression oder Regression weiterhin die Größenänderung des Tumors. Auch eine neue Kontrastmittelaufnahme innerhalb eines Tumors hat keine Bedeutung für das aktuelle Staging. Sie findet sich zwar häufiger bei Wachstum aber auch bei regredienten Tumoren. N2 - Low grade gliomas show a variable contrast enhancement which does not correlate with the prognosis. On the other hand, there are several case reports which show a spontaneous regression of tumor size with a decreasing enhancement. The purpose of this study is to answer the question, if there is a correlation between increasing contrast enhancement and tumorprogression. We showed an association between change in contrast enhancement and tumorgrowth, growing tumors tended to show an increasing contrast enhancement and vice versa. But the change in tumorsize was too little to be catgorized as progression/regression. A spot of new contrast enhancement had no prognostic value since it appeared in progredient and regredient tumors. We could show, that there is a significant correlation between change in size and contrast enhancement, but an increasing enhancement without increasing size showed no prognostic value. Tumor progression is defined in increasing volume, increasing contrast enhancement has no impact for the further prognosis. KW - Gliom KW - Kontrastmittel KW - Kernspintomografie KW - Hirntumor KW - Enhancement KW - niedrigmaligne KW - Kinder Y1 - 2014 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-145403 ER - TY - THES A1 - Panjwani, Priyadarshini T1 - Induction, Imaging, Histo-morphological and Molecular Characterization of Myocarditis in the Rat to Explore Novel Diagnostic Strategies for the Detection of Myocardial Inflammation T1 - Induktion, Bildgebung und, Histo-morphologische sowie Molekulare Charakterisierung der Myokarditis im Rattenmodell zur Entwicklung neuer diagnostischer Strategien zum Nachweis von Herzmuskelentzündungen N2 - Fulminant myocarditis is rare but a potentially life-threatening disease. Acute or mild myocarditis following acute ischemia is generally associated with a profound activation of the host’s immune system. On one hand this is mandatory to protect the host’s heart by fighting the invading agents (i.e., bacteria, viruses or other microbial agents) and/or to induce healing and repair processes in the damaged myocardium. On other hand, uncontrolled activation of the immune system may result in the generation of auto-reactive (not always beneficial) immune cells. Myocarditis or inflammatory cardiomyopathy is characterized by focal or diffuse infiltrates, myocyte necrosis and/or apoptosis and subsequent fibrotic replacement of the heart muscle. In humans, about 30% of the myocarditis-patients develop dilated cardiomyopathy. As the clinical picture of myocarditis is multifaceted, it is difficult to diagnose the disease. Therefore, the main goal of the present work was to test and further develop novel non-invasive methods for the detection of myocardial inflammation by employing both contrast enhanced MRI techniques as well as novel nuclear tracers that are suitable for in vivo PET/ SPECT imaging. As a part of this thesis, a pre-clinical animal model was successfully established by immunizing female Lewis rats with whole-porcine cardiac myosin (CM). Induction of Experimental Autoimmune Myocarditis (EAM) is considered successful when anti-myosin antibody titers are increased more than 100-fold over control animals and pericardial effusion develops. In addition, cardiac tissues from EAM-rats versus controls were analyzed for the expression of various pro-inflammatory and fibrosis markers. To further exploit non-invasive MRI techniques for the detection of myocarditis, our EAM-rats were injected either with (1) ultra-small Paramagnetic iron oxide particles (USPIO’s; Feraheme®), allowing for in vivo imaging , (2) micron sized paramagnetic iron oxide particles (MPIO) for ex vivo inflammatory cell-tracking by cMRI, or (3) with different radioactive nuclear tracers (67gallium citrate, 68gallium-labeled somatostatin analogue, and 68gallium-labeled cyclic RGD-peptide) which in the present work have been employed for autoradiographic imaging, but in principle are also suitable for in vivo nuclear imaging (PET/SPECT). In order to compare imaging results with histology, consecutive heart sections were stained with hematoxylin & eosin (HE, for cell infiltrates) and Masson Goldner trichrome (MGT, for fibrosis); in addition, immuno-stainings were performed with anti-CD68 (macrophages), anti-SSRT2A (somatostatin receptor type 2A), anti-CD61 (β3-integrins) and anti-CD31 (platelet endothelial cell adhesion molecule 1). Sera from immunized rats strongly reacted with cardiac myosin. In immunized rats, echocardiography and subsequent MRI revealed huge amounts of pericardial effusion (days 18-21). Analysis of the kinetics of myocardial infiltrates revealed maximal macrophage invasion between days 14 and 28. Disappearance of macrophages resulted in replacement-fibrosis in formerly cell-infiltrated myocardial areas. This finding was confirmed by the time-dependent differential expression of corresponding cytokines in the myocardium. Immunized animals reacted either with an early or a late pattern of post-inflammation fibrosis. Areas with massive cellular infiltrates were easily detectible in autoradiograms showing a high focal uptake of 67gallium-citrate and 68gallium labeled somatostatin analogues (68Ga DOTA-TATE). Myocardium with a loss of cardiomyocytes presented a high uptake of 68gallium labeled cyclic RGD-peptide (68Ga NOTA-RGD). MRI cell tracking experiments with Feraheme® as the contrast-agent were inconclusive; however, strikingly better results were obtained when MPIOs were used as a contrast-agent: histological findings correlated well with in vivo and ex vivo MPIO-enhanced MRI images. Imaging of myocardial inflammatory processes including the kinetics of macrophage invasion after microbial or ischemic damage is still a major challenge in, both animal models and in human patients. By applying a broad panel of biochemical, histological, molecular and imaging methods, we show here that different patterns of reactivity may occur upon induction of myocarditis using one and the same rat strain. In particular, immunized Lewis rats may react either with an early or a late pattern of macrophage invasion and subsequent post-inflammation fibrosis. Imaging results achieved in the acute inflammatory phase of the myocarditis with MPIOs, 67gallium citrate and 68gallium linked to somatostatin will stimulate further development of contrast agents and radioactive-nuclear tracers for the non-invasive detection of acute myocarditis and in the near future perhaps even in human patients. N2 - Eine fulminant verlaufende Myokarditis ist eine seltene aber potentiell lebensbedrohliche Erkrankung. Akute oder chronische Myokarditis gehen generell mit einer starken Aktivierung des Immunsystems der Betroffenen einher. Zum einen ist dies notwendig, um das Herz durch Bekämpfung der Eindringlinge (z.B. Bakterien, Viren oder andere mikrobielle Erreger) zu schützen und/oder Heilungs- und Reparaturprozesse im geschädigten Myokard einzuleiten. Zum anderen kann eine unkontrollierte Aktivierung des Immunsystems aber auch zur Entstehung von (nicht immer vorteilhaften) auto-reaktiven Immunzellen führen. Eine Myokarditis oder entzündliche Kardiomyopathie ist charakterisiert durch fokale oder diffuse Infiltrate, Nekrose und/oder Apoptose der Myozyten und einen fortschreitenden fibrotischen Ersatz des Herzmuskelgewebes. Beim Menschen entwickeln etwa 30% der Myokarditis-Patienten eine dilatative Kardiomyopathie. Da das klinische Bild der Myokarditis sehr vielfältig sein kann, ist die Diagnosestellung dieser Erkrankung schwierig. Deshalb war es das Kernziel dieser Arbeit, nicht-invasive Methoden zum Nachweis myokardialer Entzündungen zu testen, und dabei neue Bildgebungsverfahren unter Einsatz von neuen MRT-Kontrastmitteln sowie neuen nuklearen Tracern, die auch für PET/SPECT geeignet wären, zu entwickeln. Diese Verfahren wurden von uns zunächst an einem human-analogen Ratten-Modell evaluiert, mit dem Ziel später evtl. auch einmal beim Menschen eingesetzt werden zu können. Für unser präklinisches Tiermodell wurden weibliche Lewis-Ratten mit kardialem Myosin aus Schweinen immunisiert. Die erfolgreiche Induktion einer „Experimentellen Autoimmunen Myokarditis (EAM)“ wurde durch einen signifikanten Anstieg der Anti-Myosin Antikörpertiter in immunisierten Tieren und die Ausbildung eines Perikardergusses (Echokardiographie) bestätigt. Zusätzlich wurde aus apikalem kardialem Gewebe RNA isoliert und die Expression verschiedener pro-inflammatorischer und pro-fibrotischer molekularer Marker untersucht. Um die Bildgebung mittels kontrast-verstärktem cMRT zu optimieren, wurden den Tieren entweder kleine Eisenoxid-Nanopartikel (Ultra small paramagnetic iron oxide particles, USPIO; Feraheme®), oder sog. ,,Micronsized paramagnetic iron oxide particles (MPIO)‘‘ für das Tracking inflammatorischer Zellen injiziert. Im daraufolgenden Schritt wurden radioaktive nukleare Tracer (67Gallium-Citrat, 68Gallium-markierte Somatostatin-Analoga und 68Gallium-markierte zyklische RGD-Peptide) injiziert, um dann Autoradiogramme von Herzschnitten zu gewinnen. KW - Ratte KW - inflammation KW - Myokarditis KW - Kernspintomografie KW - myocarditis KW - MRI KW - cardiovascular KW - Entzündung Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-122469 ER - TY - THES A1 - Ratz, Valentin T1 - Entwicklung einer funktionellen 3D Magnetresonanz-Defäkographie T1 - Development of a functional 3D magnetic resonance defecography N2 - Epidemiologische Studien schätzen die Inzidenz chronischer Obstipation auf bis zu 27% der Gesamtbevölkerung. Betroffenen Patienten ist die Stuhlentleerung nicht oder nur unter großer Anstrengung und nicht selten nur unter Zuhilfenahme der Hand möglich. Häufig sind funktionelle Pathologien, welche sich nur während der Defäkation ausbilden, hierfür verantwortlich. Daher ist für die Diagnose und Evaluation dieser Pathologien ein bildgebendes Verfahren notwendig, welches die dynamische Darstellung der Defäkation ermöglicht. Der Goldstandard zur Untersuchung von Patienten mit funktionellen Beckenbodenstörungen ist die Entero-Colpo-Cysto-Defäkographie (ECCD). Diese Durchleuchtungsmethode erfordert die Applikation ionisierender Strahlung im Bereich des Beckens. Außerdem müssen für die Untersuchung Rektum und Vagina mit bariumhaltigem Kontrastmittel, der Dünndarm mit barium- und iodhaltigem Kontrastmittel und zusätzlich die Blase mit iodhaltigem Kontrastmittel gefüllt werden. Bei der MR-Defäkographie hingegen ist keine ionisierende Strahlung notwendig und nur eine rektale Füllung mit Ultraschallgel als Kontrastmittel erforderlich. Zudem ermöglichen statische Aufnahmen aufgrund des hohen Weichteilkontrasts der MR-Bildgebung eine detaillierte Darstellung des gesamten Beckenbodens. Die MR-Bildgebung ist jedoch im Vergleich zu anderen Bildgebungsmodalitäten, wie beispielsweise der radiographischen Durchleuchtung, langsam. Besonders zur Darstellung dynamischer Prozesse ist daher eine starke Beschleunigung des Akquisitionsprozesses notwendig. Bei der Standard 2D MR-Defäkographie wird für die Beschleunigung der Datenakquisition eine regelmäßige zweifache Unterabtastung des k-Raums vorgenommen. Hierdurch lassen sich aber nur drei räumlich voneinander getrennte 68 2D Schichten mit einer zeitlichen Aktualisierungsrate der drei Schichten von ca. 1s akquirieren. Dadurch ist aber besonders die Diagnose lateral lokalisierter Pathologien eingeschränkt oder gar nicht möglich. Daher wurde in dieser Arbeit eine 3D MR-Defäkographie zur dynamischen Darstellung der Defäkation innerhalb eines vollständigen 3D Volumens entwickelt, implementiert und anhand von 9 Patientenmessungen optimiert. Die letzten 4 Patienten wurden mit den optimierten Sequenzparametern untersucht. Ausgehend von der kartesischen Datenakquisition der bestehenden 2D MRDefäkographie wurden zunächst dreidimensionale kartesische Trajektorien zur Datenakquisition und dafür geeignete Algorithmen zur Datenrekonstruktion untersucht. In diesem Zusammenhang wurde ein GRAPPA Centric-Out Akquisitionsschema in Kombination mit einer GRAPPA Datenrekonstruktion vorgestellt. Es zeigte sich jedoch, dass eine Stack-of-Stars Trajektorie in Bezug auf die stabile, rauscharme, dynamische Darstellung der Defäkation, vorteilhaft gegenüber der untersuchten kartesischen GRAPPA Centric-Out Trajektorie ist. Zur weiteren Optimierung der Messsequenz wurden daher drei radiale Stackof-Stars Akquisitionsschemata untersucht: Das Standard Stack-of-Stars Schema sowie zwei mit View-Sharing und zwei unterschiedlichen Dichtegewichtungen modifizierte Stack-of-Stars Schemata (DW-Sampling 1 und DW-Sampling 2). Das View-Sharing ermöglicht durch die Umstellung der Reihenfolge der akquirierten Partitionen nahezu eine Verdopplung der rekonstruierten Zeitpunkte der dynamisch gemessenen Zeitserie. Die Dichtegewichtung bewirkt, dass in den zentralen Partitionen mehr radiale Speichen gemessen werden und damit das k-Raum Zentrum dichter abgetastet wird als in den äußeren Partitionen. Beim Dichtegewichtungsschema DW-Sampling 2 ist der Abfall der Anzahl der innerhalb einer Partition gemessenen Speichen stärker als beim DW-Sampling 1. Trotzdem führte das mit View-Sharing und DW-Sampling 2 modifizierte Stackof-Stars Akquisitionsschema in Verbindung mit der FISTA Compressed Sensing Datenrekonstruktion zum besten Kompromiss zwischen erreichbarer räumlicher 69 und zeitlicher Auflösung. Dieses optimierte Setup ermöglicht die dynamische Darstellung der Defäkation in 7 Schichten eines vollständigen 3D Volumens mit einer Volumenaktualisierungsrate von 1,3s. Im Vergleich zur standardmäßig durchgeführten 2D MR-Defäkographie ist daher eine mehr als doppelt so große Abdeckung mit einer vergleichbaren zeitlichen Aktualisierungsrate und einer etwas geringeren räumlichen Auflösung gewährleistet. Hierdurch lassen sich zusätzlich zu den gewöhnlichen zentral gelegenen Pathologien auch lateral ausgeprägte Pathologien besser abdecken und diagnostizieren. N2 - Epidemiological studies estimated that the incidence of chronic obstructive diseases is up to 27% of the general population. Affected patients suffer from incomplete obstructed defecation and in many cases defecation is only possible with manual maneuvers. Functional pathologies that only arise during the defecation process are responsible for that. Therefore, for the diagnosis and evaluation of these pathologies an imaging technique is necessary, that allows to visualize the dynamic of the defecation process. The gold-standard to evaluate patients with functional pelvic floor disorders is the radiographic method of Entero-Colpo-Cysto Defecography (ECCD). Two major disadvantages of ECCD are the application of ionizing radiation in the pelvic floor region and the displeasing measurement procedure for the patients. Barium containing contrast agent has to be administered to rectum and vagina of the patient, barium and iodine containing contrast agents have to be administered to the small intestine and in addition the bladder has to be filled with iodine containing contrast agent. However MR-defecography uses no ionizing radiation and only requires a rectal filling with sonographic gel as contrast agent. In addition static MR images provide a detailed overview over the whole pelvic floor due to the high soft tissue contrast. But compared to other imaging modalities like ECCD MR imaging is slow and thus a high acceleration of the data acquisition is necessary to visualize the fast dynamic defecation process. In the standard 2D MR-defecography a regular twofold undersampling of the kspace is performed for acceleration. Hereby only 3 spatially separated 2D slices with a temporal update rate of about 1s can be acquired. Thus especially the detection of lateral localized pathologies is limited or even impossible. 72 Therefore, a 3D MR-defecography was developed, implemented and optimized by means of a study in 9 patients. The last 4 patients were investigated with the optimized parameters. Based on the Cartesian data acquisition of the standard 2D MR-defecography, 3D Cartesian trajectories for data acquisition and suitable algorithms for data reconstruction were evaluated. In this context a GRAPPA Centric-Out acquisition scheme in combination with a GRAPPA data reconstruction was presented. It turned out however, that a Stack-of-Stars trajectory is preferable to the proposed Cartesian GRAPPA Centric-Out trajectory in regard to a stable, low-noise and dynamic visualization of the defecation. Therefore, to further optimize the measurement sequence, three different Stackof-Stars data acquisition schemes were investigated: The standard Stack-ofStars scheme as well as the two with View-Sharing and Density-Weighting modified Stack-of-Stars schemes DW-Sampling 1 and DW-Sampling 2. ViewSharing allows to almost double the number of reconstructed time frames by changing the order of the acquired partitions. The Density-Weighting results in more radial spokes in the central partitions than in the outer partitions and thereby assures a denser sampling of the k-space center than in the kspace periphery. The undersampling factors in the outer partitions are higher in DW-Sampling 2 than in DW-Sampling 1. Nevertheless the Stack-of-Stars acquisition scheme modified with View-Sharing and DW-Sampling 2 in combination with a FISTA Compressed Sensing data reconstruction led to the best trade-off between achievable spatial and temporal resolution. This optimized setup allows the dynamic visualization of the defecation within 7 slices and a volume update rate of 1,3s. Compared to the standard 2D MR-defecography a more than doubled spatial coverage and a comparable temporal update rate are provided. This offers the possibility to cover and evaluate lateral localized pathologies additionally to the standard central localized pathologies. KW - Kernspintomografie KW - Defäkographie KW - Dynamische MR-Bildgebung KW - Radiale MR-Bildgebung KW - Stack-of-Stars Sequenz KW - Dichtegewichtung KW - Schnelle MR-Bildgebung KW - Dynamic MR imaging KW - radial MR imaging KW - fast MR imaging KW - Stack-of-Stars sequence KW - density weighting Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-139762 ER - TY - THES A1 - Triphan, Simon T1 - T1 und T2*-Quantifizierung in der menschlichen Lunge T1 - T1 and T2* quantification in the human lung N2 - In dieser Arbeit werden für die Anwendung in der menschlichen Lunge optimierte Methoden zur Bestimmung von T1- und T2*-Karten diskutiert: Dc-Gating ermöglicht die Quantifizierung in freier Atmung, wobei für die T1-Quantifizierung mittels Inversion Recovery eine Korrektur des dc-Signals entwickelt wurde. Dies hat den Vorteil, dass Parameterkarten aus mehreren Messungen anhand ihrer dc-Signale passend überlagert werden können. Da T1 und T2* auf unterschiedliche Art und Weise von der Sauerstoffkonzentration abhängen, verbessert dies die Möglichkeit, ΔT1- und ΔT2*- Differenzkarten aus Messungen mit unterschiedlichen O2-Konzentrationen im Atemgas zu erstellen. Die Parameterquantifizierung ist in erster Linie für die Beobachtung von Krankheitsverläufen interessant, da T1 und T2* absolute, vergleichbare Zahlen sind. Da T2* deutlich vom Atemzustand abhängt, ist es auch hierfür sinnvoll, durch Gating identische Atemzustände abzubilden. Um die unterschiedlichen Einflüsse des Sauerstoffs auf T1 und T2* besser vergleichbar zu machen, wurde in dieser Arbeit weiterhin eine kombinierte Messung für beide Parameter implementiert: Da auch diese in freier Atmung stattfindet, profitieren nicht nur die Differenzkarten von der Überlagerung der Bilder, sondern auch der Vergleich der ΔT1- und ΔT2*-Karten untereinander. Messungen mit einer konventionellen kartesischen Methode an COPD-Patienten unter Raumluft- und 100% Sauerstoffatmung ergaben bei Verwendung identischer Atemmasken ein deutlich geringeres ΔT1 als in gesunden Probanden. Dass T1 in der Lunge nicht nur von der Sauerstoffkonzentration sondern auch von der Gewebezusammensetzung und insbesondere auch dem Blutvolumenanteil abhängt, zeigte sich hierbei aber auch an den bei COPD im Mittel sehr viel kürzeren T1-Zeiten bei Raumluft. Die aufgrund emphysematischer Veränderung noch zusätzlich reduzierte Protonendichte im Parenchym kranker Lungen macht diese Messungen allerdings besonders schwierig. Die oben erwähnten Optimierungen der T1-Quantifizierung zielen daher auch darauf ab, das Signal aus der Lunge zu maximieren, um Patientenmessungen einfacher zu machen: Messungen in freier Atmung sind für Patienten nicht nur einfacher, sondern erlauben effektiv auch längere Messzeiten. Insbesondere wurde aber durch die Entwicklung einer radialen Methode die Echozeit zur Messung reduziert, um die kurze T2*-Zeit in der Lunge auszugleichen. Schließlich wurde durch Implementation einer 2D UTE Sequenz die Messung bei der kürzesten vom Scanner erlaubten Echozeit ermöglicht. Die Messungen bei ultrakurzen Echozeiten in Probanden zeigten allerdings deutlich kürzere T1-Zeiten als die zuvor gefundenen oder in der Literatur dokumentierten. In weiteren Experimenten wurde das sichtbare T1 zu mehreren Echozeiten mit Hilfe der zur kombinierten Quantifizierung entwickelten Methode bestimmt. Dabei ergab sich eine Zunahme des gemessenen T1 mit der Echozeit. Aus diesem Verhalten sowie den gefundenen kürzesten und längsten T1 lässt sich schließen, dass das intra- und extravaskuläre Lungenwasser, also Blut bzw. das umgebende Gewebe, mit unterschiedlichen T1- und T2*-Zeiten zum Signal und damit auch dem effektiven T1 beitragen. Dass das TE der Messung die Gewichtung dieser Kompartimente bestimmt, hat dabei mehrere Auswirkungen: Einerseits bedeutet dies, dass beim Vergleich von T1-Messungen in der Lunge stets auch das TE mitbetrachtet werden muss, bei dem diese durchgeführt wurden. Andererseits lässt sich die Möglichkeit, die Messung auf die unterschiedlichen Kompartimente abzustimmen, potentiell ausnutzen, um zusätzliche diagnostische Informationen zu gewinnen: Da T1 vom Blutvolumenanteil und der Gewebezusammensetzung abhängt, könnte dieser Effekt helfen, diese beiden Einflüsse zu differenzieren. Während die in dieser Arbeit beschriebenen Experimente die TE-Abhängigkeit des sichtbaren T1 in Probanden aufzeigen, liefern sie allerdings noch keine genaue Erklärung für die möglichen Ursprünge dieses Effekts. Um diese weiter zu untersuchen, könnten allerdings gezielte Phantom- und in vivo-Experimente Aufschluss geben: Ein Aufbau, der die Feldverzerrung durch luftgefüllte Alveolen in Lösungen mit entsprechenden verschiedenen Suszeptibilitäten nachbildet, reduziert den Unterschied zwischen den Kompartimenten auf T1 und χ. Eine in vivo-Messung mit möglichst großer Differenz zwischen Ex- und Inspiration hingegen könnte den Einfluss der Abstände der Kompartimente vom Gasraum aufzeigen, da die Alveolarwände in tiefer Inspiration am weitesten gedehnt und daher am dünnsten sind. N2 - In this work, methods for the local measurement of T1 and T2* maps optimized for the application in the human lungs are discussed: Quantification during free breathing was enabled by applying dc-gating, where a correction for the dc-signal acquired during T1-quantification using a inversion recovery was introduced. This is especially useful to achieve parameter maps in identical breathing states from multiple measurements using their dc-signals. Since T1 and T2* depend on the oxygen concentration through different mechanisms, this is especially interesting to produce ΔT1- and ΔT2*-difference maps at varying O2-concentrations in the breathing gas. Parameter quantification is primarily interesting for the monitoring of the courses of disease or therapy since T1 and T2* are absolute, comparable numbers. As T2* depends significantly on the respiratory state, ensuring identical states via gating is relevant there as well. To further improve the comparison of oxygen influence on T1 and T2* a method for the combined measurement of both parameters was implemented: Since this is also employs gating, not only the difference maps benefit from image coregistration, but the comparison of the ΔT1 and ΔT2* maps to each other as well. Measurements using the conventional cartesian method on COPD patients under room air and pure oxygen conditions resulted in much lower ΔT1 than in healthy volunteers when using identical oxygen masks. The much lower average T1 times at room air found there demonstrate that T1 in the lungs not only depends on the oxygen concentration but also on tissue composition and especially the blood volume fraction. Proton densities that were reduced even further due to emphysematous destruction made these measurements additionally difficult. Accordingly, the optimizations for T1 quantification mentioned above are intended to maximize signal from the lung parenchyma to improve patient measurements: Measurements during free breathing are not only easier for patients but effectively also allow for longer acquisition times. In particular the developement of a radial method provides a shorter echo time to help compensate for the short T2* in the lungs. Finally, the implementation of a 2D UTE sequence enables the measurement at the shortest echo time available on the scanner hardware. However, the measurements at ultra short echo times in volunteers showed significantly shorter T1 times than those found previously and those reported in the literature. In further experiments, the observable T1 was determined at multiple echo times using the method developed for simultaneous quantification. This revealed a gradual increase of the measured T1 with the echo time. From this behaviour as well as the shortest and longest times found, it can be concluded that the intra- and extravascular compartments of lung water, essentially blood and the surrounding tissue, contribute with different T1 and T2* times to the MR signal and thus also the effective T1. That the echo time of the measurement determines the weighting of these compartments has multiple consequences: Firstly, this means that when comparing T1 measurements in the lungs, the echo time that was used to acquire them also has to be considered. Secondly, the possiblity to focus the measurement on these different compartments might be used to gain additional diagnostic information: Since T1 depends on blood volume content and tissue composition, this effect might help to differentiate these two influences. While the experiments described in this work demonstrate the echo time dependence of the observed T1 in volunteers, they do not yet provide an explanation for the exact origins of this effect. To examine these further, appropriate phantom and in vivo experiments could be insightful: A phantom design that simulates the field distortion caused by air-filled alveoli in solutions with suitable susceptibilites would reduce the difference between the compartments to T1 and χ. A in vivo measurement with an especially large difference between ex- and inspiration could help to show the influence of the distance of the compartments from the gas space, since the alveolar walls are most dilated and thus thinnest during deep inspiration. KW - Kernspintomografie KW - Lunge KW - T2*-Relaxation KW - T1-Relaxtion KW - funktionelle Lungenbildgebung KW - MRT der Lunge KW - Spin-Gitter-Relaxation Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-139621 ER - TY - THES A1 - Schmitt, Peter T1 - MR imaging of tumors: Approaches for functional and fast morphological characterization T1 - MR-Bildgebung von Tumoren: Ansätze zur funktionellen und schnellen morphologischen Charakterisierung N2 - The subject of this work was to develop, implement, optimize and apply methods for quantitative MR imaging of tumors. In the context of functional and physiological characterization, this implied transferring techniques established in tumor model research to human subjects and assessing their feasibility for use in patients. In the context of the morphologic assessment and parameter imaging of tumors, novel concepts and techniques were developed, which facilitated the simultaneous quantification of multiple MR parameters, the generation of “synthetic” MR images with various contrasts, and the fast single-shot acquisition of purely T2-weighted images. N2 - Gegenstand dieser Arbeit war die Entwicklung, Implementierung, Optimierung und Anwendung von Methoden für die quantitative MR-Bildgebung an Tumoren. In Bezug auf eine funktionelle und physiologische Charakterisierung wurden in der Forschung an Tumormodellen etablierte Verfahren für den Einsatz am Menschen adaptiert und ihre Anwendbarkeit zur Untersuchung von Tumoren wurde an Patienten erforscht. Im Bereich der morphologischen Untersuchung und Parameterbildgebung an Tumoren wurden neue Konzepte und Verfahren entwickelt, welche die simultane Quantifizierung mehrerer MR-Parameter, die Generierung "synthetischer" MR-Bilder mit unterschiedlichen Kontrasten, sowie die schnelle "Single-Shot"-Akquisition rein T2-gewichteter Bilder ermöglichen. KW - Kernspintomografie KW - Tumor KW - MR imaging KW - Tumors KW - Relaxometry KW - IR-TrueFISP KW - synthetic MRI KW - TOSSI Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-135967 ER - TY - THES A1 - Bachschmidt, Theresa T1 - Magnetic Resonance Imaging in Proximity to Metal Implants at 3 Tesla T1 - Magnetresonanzbildgebung nahe metallischer Implantate bei 3 Tesla N2 - Magnetic resonance imaging is derogated by the presence of metal implants and image quality is impaired. Artifacts are categorized according to their sources, the differences in susceptibility between metal and tissue and the modulation of the magnetic radiofrequency (RF) transmit field. Generally, these artifacts are intensified at higher field strength. The purpose of this work is to analyze the efficiency of current methods used for metal artifact reduction at 3T and to investigate improvements. The impact of high-bandwidth RF pulses on susceptibility-induced artifacts is tested. In addition, the benefit of a two-channel transmit system with respect to shading close to total hip replacements and other elongated metal structures in parallel to the magnetic field is analyzed. Local transmit/receive coils feature a higher peak B1 amplitude than conventional body coils and thus enable high-bandwidth RF pulses. Susceptibility-induced through-plane distortion relates reciprocally to the RF bandwidth, which is evaluated in vitro for a total knee arthroplasty. Clinically relevant sequences (TSE and SEMAC) with conventional and high RF pulse bandwidths and different contrasts are tested on eight patients with different types of knee implants. Distortion is rated by two radiologists. An additional analysis assesses the capability of a local spine transmit coil. Furthermore, B1 effects close to elongated metal structures are described by an analytical model comprising a water cylinder and a metal rod, which is verified numerically and experimentally. The dependence of the optimal polarization of the transmit B1 field, creating minimum shading, on the position of the metal is analyzed. In addition, the optimal polarization is determined for two patients; its benefit compared to circular polarization is assessed. Phantom experiments confirm the relation of the RF bandwidth and the through-plane distortion, which can be reduced by up to 79% by exploitation of a commercial local transmit/receive knee coil at 3T. On average, artifacts are rated “hardly visible” for patients with joint arthroplasties, when high-bandwidth RF pulses and SEMAC are used, and for patients with titanium fixtures, when high-bandwidth RF pulses are used in combination with TSE. The benefits of the local spine transmit coil are less compared to the knee coil, but enable a bandwidth 3.9 times as high as the body coil. The modulation of B1 due to metal is approximated well by the model presented and the position of the metal has strong influence on this effect. The optimal polarization can mitigate shading substantially. In conclusion, through-plane distortion and related artifacts can be reduced significantly by the application of high-bandwidth RF pulses by local transmit coils at 3T. Parallel transmission offers an option to substantially reduce shading close to long metal structures aligned with the magnetic field. Effective techniques dedicated for metal implant imaging at 3T are introduced in this work. N2 - Metallimplantate beeinträchtigen die Funktionsweise der Magnetresonanztomographie und verschlechtern die Bildqualität. Die Artefakte werden entsprechend ihres Ursprungs kategorisiert, in einerseits Suszeptibilitätsunterschiede zwischen Metall und Gewebe und andererseits die Modulation des B1-Feldes. Im Allgemeinen verstärken sich diese Artefakte bei höheren Feldstärken. Das Ziel dieser Arbeit ist es, die Effizienz vorhandener Methoden zur Artefaktreduktion bei 3T zu bewerten und mögliche Verbesserungen herauszuarbeiten. Der Einfluss von breitbandigen Hochfrequenz-Pulsen (HF-Pulsen) auf Suszeptibilitätsartefakte wird untersucht. Zusätzlich wird der Einfluss eines Zwei-Kanal Sendesystems auf Abschattungen analysiert, die in der Nähe von Hüftimplantaten und anderen länglichen Implantaten auftreten, welche parallel zu B0 liegen. Im Gegensatz zu konventionellen Ganzkörper-Sendespulen erlauben lokale Sende-/Empfangsspulen eine höhere maximale B1-Amplitude, die breitbandigere HF-Pulse ermöglicht. Die reziproke Abhängigkeit der Suszeptibilitätsartefakte in Schichtrichtung zur HF-Bandbreite wird in vitro für eine Kniegelenkplastik evaluiert. An acht Patienten mit verschiedenen Knieimplantaten werden klinisch relevante Sequenzen (TSE und SEMAC) mit konventionellen und breitbandigen HF-Pulsen in verschiedenen Kontrasten getestet und die Verzerrungen werden von zwei Radiologen bewertet. Eine weitere Studie untersucht das Potenzial einer lokalen Sendespule für die Wirbelsäule. Darüberhinaus werden B1-Effekte nahe länglicher Metallstrukturen durch ein analytisches Modell beschrieben, das numerisch und experimentell überprüft wird. Des Weiteren wird die Abhängigkeit der optimalen Polarisation des B1-Feldes, die minimale Abschattung verursacht, von der Position des Metalls untersucht. Für zwei Patienten wird die optimale Polarisation bestimmt und deren Vorteil gegenüber der zirkularen Polarisation analysiert. Phantomversuche bestätigen die Abhängigkeit zwischen HF-Bandbreite und der Schichtverzerrung, die durch die Verwendung einer lokalen Kniespule mit Sende- und Empfangsfunktion bei 3T um 79% reduziert werden kann. Die Artefakte bei Patienten mit Vollimplantaten, bzw. Titanimplantaten, werden als "kaum sichtbar" bewertet, wenn SEMAC, bzw. TSE, mit breitbandigen HF-Pulsen kombiniert appliziert wird. Im Vergleich zur lokalen Kniespule fallen die Vorteile der lokalen Wirbelsäulen-Sendespule geringer aus; dennoch kann die 3,9-fache HF-Bandbreite der Ganzkörpersendespule erreicht werden. Die B1-Modulation aufgrund von Metall wird im dargestellten Modell gut wiedergegeben und die Position des Metalls im Objekt hat großen Einfluss auf den Effekt. Die Verwendung der optimalen Polarisation kann Abschattungen stark reduzieren. Zusammenfassend können Artefakte aufgrund von Schichtverzerrungen durch die Verwendung lokaler Sendespulen und breitbandiger HF-Pulse bei 3T stark abgeschwächt werden. Die individuelle Wahl der Polarisation des B1-Feldes bietet eine gute Möglichkeit, Abschattungen in der Nähe von länglichen Metallstrukturen zu reduzieren, soweit diese näherungsweise parallel zu B0 ausgerichtet sind. Somit werden in dieser Arbeit wirksame Methoden zur Metallbildgebung bei 3T eingeführt. KW - Kernspintomografie KW - Metallimplantat KW - Artefakt KW - 3 Tesla KW - Implantat KW - Verzerrung KW - Abschattung Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-135690 ER - TY - THES A1 - Ott, Martin T1 - Lautstärkereduzierte Magnetresonanztomographie T1 - Acoustic noise reduced MRI N2 - Messungen mit Magnetresonanztomographen sind seit jeher mit hohen Lautstärken verbunden. Deshalb wird das Gerät im Volksmund auch als „laute Röhre“ bezeichnet. Bisher wurde das Problem mit Kopfhörern, Ohrenstöpseln und akustischer Dämmung des MRT-Scanners angegangen. Auch in der Fachliteratur wird das Problem als gegeben angesehen und es werden kaum wissenschaftliche Lösungsansätze zur Lautstärkereduktion beschrieben. Das Ziel der vorliegenden Arbeit war es, Bildgebungs-Sequenzen für schwer‑optimierbare Bildkontraste und sogenannte Standard-Kontraste aus dem klinischen Umfeld hinsichtlich der Lautstärke zu optimieren. Viele dieser Kontraste können bereits mit einfachen Algorithmen wie dem Gradientenglättungsalgorithmus erfolgreich in Hinblick auf die Lautstärke optimiert werden. Allerdings existieren auch Sequenzen beziehungsweise Kontraste, die aufgrund ihrer Eigenschaften nicht von einem solchen Algorithmus profitieren können. Die Optimierungen und Änderungen sollten software-seitig erfolgen, das heißt durch Änderung der Gradientenformen und Datenakquisition. In der Arbeit wurden die grundlegenden Zusammenhänge zwischen den verwendeten Geräteparametern und der Lautstärke untersucht und zudem die physikalischen Ursachen der Lautstärkeentwicklung hergeleitet. Diese konnten anhand der Lorentz-Kräfte quantitativ beschrieben werden. Somit konnten die Hauptursachen der Lautstärkeentwicklung identifiziert werden. Diese sind abhängig von der Gradienten-Steig-Rate, aber auch von der Amplitude der Gradienten. Es konnte gezeigt werden, dass eine Minimierung dieser Gradientenparameter zu einer geringeren Lautstärkeentwicklung führt. Allerdings führt diese Minimierung in den meisten Fällen auch zu einer systematischen Verlangsamung des Sequenzablaufs, was das Erreichen bestimmter Echozeiten und Bildkontraste unmöglich macht. Zu den problematischen Kontrasten bezüglich der Lautstärkereduktion zählten der T1- und PD‑Kontrast einer Turbo-Spin-Echo-Sequenz. Durch die Kombination von mehreren Maßnahmen, wie der Adaption der k-Raum-Akquisition, der HF-Pulse-Parameter und den Gradientenformen, war es möglich, die Lautstärke in Beispielmessungen um bis zu 16,8 dB(A) zu reduzieren. Wie bei der kürzlich veröffentlichten Methode zur Reduktion für die T2‑gewichteten Kontraste, wurde dies zulasten einer Messzeitverlängerung von bis zu 50% erreicht. Die Endlautstärke betrug dabei circa 81 dB(A). Mit der Lautstärkeoptimierung der klinisch bedeutsamen T1- und PD‑Kontraste wurde die Palette an leisen, mit der Turbo-Spin-Echo‑Sequenz erzielbaren, Standard-Kontrasten (T1, T2 und PD) nun vervollständigt. In einem anderen Ansatz wurde die Anwendbarkeit des CAT-Konzepts auf die Lautstärkereduktion untersucht. Beim CAT-Konzept wird die Messung in Einzelmessungen mit verschiedenen Parametern unterteilt. Bisher wurde dieser Ansatz zur SAR-Reduktion verwendet. Das Zentrum des k-Raums wird mit einer SAR-intensiven, kontrastgebenden Messung aufgenommen. Der verbleibende Teil des k-Raums wird mit einer SAR-reduzierten, bildstrukturrelevanten Messung aufgenommen. In dieser Arbeit wurde die Übertragung des CAT-Konzepts auf die Lautstärkereduktion untersucht. Anstelle von SAR-intensiven und SAR‑reduzierten Messungen, wurde hier die Unterteilung in „laute“ und „leise“ Messungen untersucht. Dabei wurden Überlegungen angestellt, die es für eine Vielzahl an Messungen ermöglichen, einen großen Teil der Messung leise zu gestalten ohne die Bildqualität oder den Bildkontrast zu verändern. In einem weiteren Schritt wurden Überlegungen für die Lautstärkereduktion der lauten Messungen vorgestellt. Anschließend wurden für eine GRE- und TSE-Sequenz Optimierungsschritte evaluiert und die Lautstärke gemessen. Der hinsichtlich der Lautstärkeoptimierung herausforderndste Bildkontrast ist die diffusionsgewichtete Bildgebung. Diese besitzt eine Diffusions-Präparation zur Sichtbarmachung der Diffusivität, bei der die maximal mögliche Gradienten-Amplitude verwendet wird. Ebenso werden nach der Präparation die Daten mit einem EPI‑Akquisitionsmodul mit Blip-Gradienten akquiriert, das mit einem charakteristischem „Pfeifton“ einhergeht. Zum einen wurden die Gradientenformen konsequent angepasst. Zum anderen wurde eine Segmentierung der k-Raum-Akquisition in Auslese-Richtung verwendet, um die Gradienten‑Steig-Raten zu reduzieren. Auch hier konnte eine deutliche Lautstärkereduktion von bis zu 20,0 dB(A) erzielt werden. Dies wurde zulasten einer Messzeitverlängerung von 27% ‑ 34% im Vergleich zur Standard-Sequenz erreicht. Durch eine weitere Messzeitverlängerung um bis zu 23% kann die Lautstärke um weitere 0,9 dB(A) reduziert werden. Dabei hängt die genaue Messzeitverlängerung vom verwendeten GRAPPA-Faktor und der Anzahl der Auslese-Segmente ab. Die entstandene Sequenz wurde in mehreren Kliniken erfolgreich erprobt. Bisher mussten bei MRT-Messungen stets Kompromisse zwischen „hoher Auflösung“, „hohem SNR“ und „geringer Messzeit“ getroffen werden. Als Anschauung dafür wurde das „Bermuda‑Dreieck der MRT“ eingeführt. Da alle drei Größen sich gegenseitig ausschließen, muss stets ein Mittelweg gefunden werden. Einige der in dieser Arbeit erzielten Erfolge bei der Lautstärkereduktion wurden auf Kosten einer verlängerten Messzeit erreicht. Daher ist es naheliegend, das „Bermuda-Dreieck der MRT“ um die Dimension der „geringen Lautstärke“ zu einer „Bermuda-Pyramide der MRT“ zu erweitern. Damit muss die Lautstärkeentwicklung in die Mittelweg‑Findung miteinbezogen werden. Die in dieser Arbeit erzielten Lautstärken liegen in der Größenordnung zwischen 80 ‑ 85 dB(A). Somit können Messungen bei Verwendung von Gehörschutz angenehm für den Patienten durchgeführt werden. Durch neue Techniken der Zukunft wird es wahrscheinlich sein, höhere Auflösungen, höheres SNR oder kürzere Aufnahmedauern zu erzielen, beziehungsweise stattdessen diese in eine geringe Lautstärke „umzuwandeln“. Ebenso werden möglicherweise auf der hardware-technischen Seite Fortschritte erzielt werden, so dass in neueren MRT-Scannergenerationen mehr Wert auf die Lärmdämmung gelegt wird und somit der softwarebasierten Lautstärkereduktion einen Schritt entgegen gekommen wird. Damit könnten zukünftige Patienten-Messungen gänzlich ohne störenden Gehörschutz durchgeführt werden. N2 - Magnetic resonance imaging (MRI) measurements have always been related to high acoustic noise. Therefore, in common parlance MRI is referred to as the “loud tube”. Until now, the acoustic noise was mitigated by the use of headphones and ear plugs as well as acoustic dampening of the MR system. In literature, the problem is more or less acknowledged and solutions to the acoustic noise are rarely provided. The aim of this work was to optimize MR sequences, which generate so-called standard clinical MRI contrasts, for acoustic noise. Many of these contrasts could be optimized for acoustic noise by a gradient smoothing algorithm. Nevertheless, there are sequences and contrasts which cannot benefit from such algorithms and therefore need manual optimization. Software-based optimizations are performed by adapting the gradient waveforms and data acquisition. In this work, the main relationships between parameter settings of the MRI machine and acoustic noise were explored. The physical origin of acoustic noise in the form of Lorentz forces was derived from fundamental equations. The main acoustic noise sources are gradient slew rate and gradient amplitude. It was shown that minimization of these quantities leads to reduced acoustic noise. However, this is mostly accompanied by slowing down the sequence and thus certain echo times and contrasts cannot be reached. T1- and PD-weighted contrasts, acquired with a turbo spin-echo sequence, are problematic contrasts regarding acoustic noise reduction. This problem was tackled by a combination of several approaches such as an adaption of the k-space acquisition, changes to the RF-pulse parameters, and modifications of the gradient waveform. An acoustic noise reduction of up to 16.8 dB(A) was achieved. As for the previously published method for acoustic noise reduction in T2-weighted contrasts, this success came at the cost of an increase of measurement time by 50%. The target acoustic noise level was around 81 dB(A). With this optimization, the palette of quiet standard clinical contrasts, consisting of T1-, T2- and PD-weighted contrasts, can be realized with the turbo spin-echo sequence. In a different approach, the Combined-Acquisition (CAT) concept was applied to acoustic noise reduction. In implementing the CAT concept, each measurement is divided into two measurements with different parameters. This approach was previously used for SAR reduction. The center of k space is acquired using a high-SAR measurement in which contrast is relevant. The remaining k-space area is acquired using a low SAR, contrast-irrelevant measurement. In this work, the CAT concept was applied to acoustic noise reduction. Each measurement was divided into ‘quiet’ and ‘loud’ segments instead of dividing into high-SAR and low-SAR measurements. Considerations allowed for acoustic noise reduction without disrupting the image quality or contrast. In successive steps, the approach was applied to the remaining loud segment of the measurement. This process was executed for a GRE and a TSE sequence. Corresponding acoustic noise measurements were performed. One of the most challenging contrasts in terms of acoustic noise reduction is diffusion weighted imaging. It employs maximum gradient amplitudes in the preparation pulses which sensitize the MR signal to diffusivity. Data acquisition is performed by an EPI readout including blipped gradients. This readout is known for its whistling sound. Therefore, the gradient waveforms were consequently adapted. A k-space segmentation in the readout direction was employed to reduce the gradient slew rates. In this work, an acoustic noise reduction of up to 20.0 dB(A) could be achieved using an adapted readout segmented EPI sequence. This reduction in acoustic noise came at the cost of an increase of measurement time by 27% to 34% compared to the standard sequence. Spending additional 23% of acquisition time can further reduce the acoustic noise by 0.9 dB(A). The exact increase in measurement time depends on the employed GRAPPA factor and the number of readout segments. The optimized sequence was successfully validated in various clinical sites. Until now, compromises had to be made between high resolution, high SNR, and short acquisition time. This compromise can be described as the “Bermuda triangle of MRI”. Trade-offs exist between all three quantities. A compromise has to be chosen in all cases. In this work, some of the achieved acoustic noise reductions came at the cost of increased measurement time. Therefore, the dimensionality of the Bermuda triangle is extended with the addition of low acoustic noise. This yields the “Bermuda triangular pyramid of MRI”. Thus, acoustic noise has to be included in achieving a balance of desired properties in MR image acquisition. In this work, the obtained acoustic noise levels were on the order of 80–85 dB(A). Upon the use of ear protection, measurements became comfortable for the patients. As further advancements in imaging technology are made, it may be likely to achieve higher resolution, higher SNR, or shorter acquisition times, which could instead be traded for lower acoustic noise levels. In addition, it is possible that MRI machine manufacturers will put more effort into hardware based acoustic noise dampening of the devices in order to meet software-based acoustic noise reduction. Therefore, patient measurements could be possible without the need for additional acoustic noise protection in the future. KW - Kernspintomografie KW - Biophysik KW - Magnetische Kernresonanz KW - Lautstärkereduktion KW - Lärm KW - Krach KW - Patientenkomfort KW - Lärmbelastung KW - Geräuschminderung Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-133921 ER - TY - THES A1 - Kartäusch, Ralf T1 - Spektroskopische Flussmessung an Pflanzen mittels mobilem Magnetresonanztomographen T1 - Spectroscopic flow measurements in plants using a mobile magnetic resonance system N2 - The main objective of this dissertation was the development of a flow sensor which is specialized on flow measurements of plants. Hence, an accessible mobile magnet and the receiver/transfer hardware have been developed. Additionally, software to control the MR-console has been written. The AC-method was advanced to acquire slow flow profiles. This enables acquiring flow in plants. Additionally, in cooperation with the working group “Lipid Motobolism” of the IPK-Gatersleben studies have been carried out to measure the influence of the ear of wheat on the water transport mechanism. Furthermore, a new technique based on the Bloch-Siegert-effect has been developed which reduces the influence of eddy currents. This simplifies flow measurements that suffer heavily from eddy currents. Hardware development An accessible mobile magnet with a field strength of 0.42 T has been build. The field homogeneity is 0.5 ppm in 1 cm³. In comparison to the existing closed magnet system at the chair EP5 this is an improvement of a factor 40. Those enhancements have been achieved by an adjusted design of the magnet which has been optimized by computer simulations. The implementation of ferrite pole shoes reduced the eddy currents by a factor 7 in comparison to the usually used iron pole shoes. Therefore, phase sensitive flow measurements using fast switching magnet field gradients could be carried out. A foldable coil has been refined to achieve an accessible receiver system. This coil has been used as a transmit/receiver unit. Furthermore, the SNR of measurements in thin plant stalks was enhanced by a constructed system that could be directly wrapped around the stalk. Additionally, two systems to reduce noise in plant measurements have been developed. Those systems can reduce the noise by a factor 92. This was necessary because the longish plant stems guides electric noise from outside of the case into the receiver coil. Both noise reduction systems, the electromagnetic shielding and the common mode rejection, removed the noise to the same level. Flow measurement In the present work a refinement of the AC-method [36] enabled for the first time acquiring quantitative flow profiles. Hence, it was possible to measure slow velocity in the range of 200 µm/s. The precondition was the replacement of the sinusoidal gradient profile by a trapezoid gradient shape. Those allowed increasing the slew rate of the gradients and therefore shorten the total duration of the ramp which finally allows higher encoding strengths. Additionally, due to intervals without applied gradients, more efficient RF-pulses can be used and more data points can be acquired in an echo. The measured flow profiles correlated to the simulation results. The accurate flow profiles have been achieved by a new evaluation technique and a phase correction mechanism. The newly developed extension to imaging enabled spatially encoded spectral flow measurements. Therefore, the location of xylem and phloem can be spatially separated. In the measurement of the black alder this becomes apparent. Here the shape of dicotyledonous plants, which is described in chapter 5.1, is visible. Additionally, due to the spatial separation of the flow directions (up/down) qualitative flow measurements are possible. In pixels where opposite flow directions can spatially be resolved the difference between the left and the right side of the flow spectra yields the total flow without static water. Due to the phase corrections technique in combination with the automatically frequency calibration, long term flow measurements were possible. Therefore, the response of plants on influences like changes in the illumination have been observed in measurements over a duration of nine days. Here flow changes below 200 µm/s can be detected. Bloch-Siegert phase encoding In this work a new spatial phase encoding technique (BS-SET) using a B1-gradient in combination with far off-resonant radio frequency pulses has been demonstrated. Based on the Bloch-Siegert Shift an eddy current free B1-gradient was used to encode images and apply flow encoding. The BS-gradient induces a phase shift which depends on B1² using a constant gradient. Therefore, adapted reconstructions have been developed that provide undistorted images using this nonlinear encoding. Alternatively, a B1-gradient has been developed where the profile of the B1-field follows a square root shape. This supplies a linear phase encoding removing the need for an adapted reconstruction and enables using this technique for flow encoding. N2 - Das Ziel der Promotion war die Entwicklung eines Flusssensors mit dem Fokus auf Flussmessungen an Pflanzen. Dazu musste zunächst die Hardware in Form eines räumlich zugänglichen Magneten und einer Sende- und Empfangseinheit entworfen werden. Um die MR-Konsole ansteuern zu können, musste eine Software entwickelt werden. Die AC-Methode wurde für Flussmessungen mit niedrigen Geschwindigkeiten angepasst und die entsprechende Theorie dazu erweitert. Mit dieser weiterentwickelten AC-Methode wurde die Flussmessung an Pflanzen demonstriert. Dafür wurden im Rahmen einer Kooperation mit der Arbeitsgruppe „Lipid Motobolism“ der IPK-Gatersleben Flussstudien an Weizenpflanzen durchgeführt. Darüber hinaus wurde in dieser Arbeit eine neue Technik zur Wirbelstromvermeidung bei Permanentmagneten entwickelt, um Problemen mit diesen bei Flussmessungen entgegenzuwirken. Sensorbau Es wurde ein zugänglicher, mobiler Magnet mit einer Feldstärke von 0,42 T gebaut. Die Feldhomogenität beträgt 0,5 ppm in 1 cm³. Im Vergleich zu dem am Lehrstuhl der EP5 bestehenden, geschlossenen, mobilen Magnetsystem erreicht das in dieser Arbeit gebaute System ein 40fach homogeneres Magnetfeld. Erzielt wurden diese Verbesserungen durch ein spezielles Design, welches durch Computersimulationen sukzessiv optimiert wurde. Durch angepasste Polschuhe konnte darüber hinaus die Induktion von Wirbelströmen im Mittel um einen Faktor 7 reduziert werden, wodurch phasensensitive Flussmessungen ermöglicht wurden. Um die Zugänglichkeit zu dem Innenraum der HF-Spulen zu gewährleisten, wurde eine Klappspule weiterentwickelt und als Sende- und Empfangseinheit für den Tomographen gebaut. Ferner wurde ein System gebaut, dass direkt um die Pflanze gewickelt werden kann und sich somit für besonders dünne Pflanzenstängel eignet. Weiterhin wurden zwei Systeme zur Rauschunterdrückung für die Messungen an Pflanzen entwickelt. Dadurch konnte das Rauschen um einen Faktor 92 gesenkt werden. Dies war notwendig, weil die länglichen Pflanzen durch ihre Ausdehnung über das Gehäuse hinweg ein Rauschen in die Empfangsspule induziert haben. Die beiden Rauschunterdrückungssysteme, die elektrische Schirmung und die Gleichtaktunterdrückung, entfernten das Rauschen dabei gleichermaßen. Flussmessung Die im Rahmen der Arbeit erfolgte Weiterentwicklung der AC-Methode [102] erlaubte es erstmals mit der Methode quantitative Flussprofile aufzunehmen. In Folge dessen war es außerdem möglich Geschwindigkeiten unter 200 µm/s zu messen. Die Vorrausetzung dafür war die Implementierung von trapezförmigen Gradienten, welche kürzere Rampzeiten und eine stärkere Kodierung zulassen. Dadurch sind außerdem Intervalle ohne Gradienten realisierbar, die effizientere Refokussierungspulse und die Aufnahme mehrerer Datenpunkte ermöglichen. Die zu erwartenden und simulierten Flussprofile entsprachen den gemessenen Profilen durch die Verwendung einer neuen Auswertungstechnik. Die neu entwickelte Erweiterung zur Bildgebung ermöglicht die ortsaufgelöste, spektroskopische Flussmessung und so können die Bereiche von Xylem und Phloem voneinander getrennt werden. Dies wurde durch Messungen einer Schwarzerle gezeigt, bei der die im Abschnitt 5.1 beschriebene Struktur dikotyler Pflanzen aufgelöst werden konnte. Zusätzlich können qualitativ genauere Aussagen über die Flussgeschwindigkeit getroffen werden. Bei Messungen an Pflanzen konnte mit der optimierten AC-Methode die Flussänderungen aufgrund äußerer Einflüsse, wie der Beleuchtung, beobachtet werden. Langzeitmessungen über 9 Tage zeigten einen der Beleuchtung folgenden Flussverlauf - auch bei sehr geringen mittleren Flussänderungen von unter 200 µm/s. Bloch-Siegert Phasenkodierung Um eine Phasenkodierung ohne die Induktion von Wirbelströmen zu erhalten, wurde im Rahmen der Arbeit die ortsabhängige Phasenkodierung mittels B1-Gradienten entwickelt. Diese Technik basiert auf HF-Wechselfeldern und benutzt den sogenannten BS-Shift um einen B1-feldabhängigen Frequenzshift zu induzieren. Zwei Rekonstruktionstechniken wurden entwickelt, um die Rekonstruktion von entzerrten Bildern zu ermöglichen. Dies war notwendig, da die Kodierung mittels BS-Shift von B1² abhängt. Infolgedessen wird bei der Verwendung von konstanten HF-Gradienten eine vom Quadrat des Ortes abhängige Phasenkodierung induziert. Als Alternative zu diesem Verfahren wurde ein Gradient entwickelt, der einen wurzelförmigen Feldverlauf hat und somit die lineare Kodierung ohne angepasste Rekonstruktionstechniken ermöglicht. KW - Kernspintomografie KW - Wassertransport KW - Spektroskopische Flussmessung KW - AC Gradients KW - Pflanzen Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-125820 ER - TY - THES A1 - Weick, Stefan T1 - Retrospektive Bewegungskorrektur zur hochaufgelösten Darstellung der menschlichen Lunge mittels Magnetresonanztomographie T1 - Retrospective Motion Correction for High Resolution Magnetic Resonance Imaging of the Human Lung N2 - Ziel dieser Arbeit war es, das gesamte Lungenvolumen in hoher dreidimensionaler Auflösung mittels der MRT darzustellen. Um trotz der niedrigen Protonendichte der Lunge und der geforderten hohen Auflösung ausreichend Signal für eine verlässliche Diagnostik zu erhalten, sind Aufnahmezeiten von einigen Minuten nötig. Um die Untersuchung für den Patienten angenehmer zu gestalten oder auf Grund der eingeschränkten Fähigkeit eines Atemstopps überhaupt erst zu ermöglichen, war eine Anforderung, die Aufnahmen in freier Atmung durchzuführen. Dadurch entstehen allerdings Bewegungsartefakte, die die Diagnostik stark beeinträchtigen und daher möglichst vermieden werden müssen. Für eine Bewegungskompensation der Daten muss die auftretende Atembewegung detektiert werden. Die Bewegungsdetektion kann durch externe Messgeräte (Atemgurt oder Spirometer) oder durch eine zusätzliche Anregungen erfolgen (konventionelle Navigatoren) erfolgen. Nachteile dieser Methoden bestehen darin, dass die Bewegung während der Atmung nicht direkt verfolgt wird, dass elektronische Messgeräte in die Nähe des Tomographen gebracht werden und das die Patienten zusätzlich vorbereitet und eingeschränkt werden. Des Weiteren erfordert eine zusätzliche Anregung extra Messzeit und kann unter Umständen die Magnetisierung auf unterwünschte Weise beeinflussen. Um die angesprochenen Schwierigkeiten der Bewegungsdetektion zu umgehen, wurden in dieser Arbeit innerhalb einer Anregung einer 3d FLASH-Sequenz sowohl Bilddaten- als auch Navigatordaten aufgenommen. Als Navigator diente dabei das nach der Rephasierung aller bildgebenden Gradienten entstehende Signal (DC Signal). Das DC Signal entspricht dabei der Summe aller Signale, die mit einem bestimmten Spulenelement detektiert werden können. Bewegt sich beispielsweise die Leber bedingt durch die Atmung in den Sensitivitätsbereich eines Spulenelementes, wird ein stärkeres DC Signal detektiert werden. Je nach Positionierung auf dem Körper kann so die Atembewegung mit einzelnen räumlich lokalisierten Spulenelementen nachverfolgt werden. Am DC Signalverlauf des für die Bewegungskorrektur ausgewählten Spulenelementes sind dann periodische Signalschwankungen zu erkennen. Zusätzlich können aus dem Verlauf Expirations- von Inspirationszuständen unterschieden werden, da sich Endexpirationszustände im Regelfall durch eine längere Verweildauer auszeichnen. Grundsätzlich kann das DC Signal vor oder nach der eigentlichen Datenaufnahme innerhalb einer Anregung aufgenommen werden. Auf Grund der kurzen Relaxationszeit T∗2 des Lungengewebes fällt das Signal nach der RF Anregung sehr schnell ab. Um möglichst viel Signal zu erhalten sollten, wie in dieser Arbeit gezeigt wurde, innerhalb einer Anregung zuerst die Bilddaten und danach die Navigatordaten aufgenommen werden. Dieser Ansatz führt zu einer Verkürzung der Echozeit TE um 0.3 ms und damit zu einem SNR Gewinn von etwa 20 %. Gleichzeitig ist das verbleibende Signal nach der Datenakquisition und Rephasierung der bildgebenden Gradienten noch ausreichend um die Atembewegung zu erfassen und somit eine Bewegungskorrektur der Daten (Navigation) zu ermöglichen. Um eine retrospektive Bewegungskorrektur durchführen zu können, müssen Akzeptanzbedingungen (Schwellenwerte) für die Datenauswahl festgelegt werden. Bei der Wahl des Schwellenwertes ist darauf zu achten, dass weder zu wenige noch zu viele Daten akzeptiert werden. Akzeptiert man sehr wenige Daten, zeichnen sich die Rekonstruktionen durch einen scharfen Übergang zwischen Lunge und Diaphragma aus, da man sehr wenig Bewegung in den Rekonstruktionen erlaubt. Gleichzeitig erhöht sich allerdings das Risiko, dass nach der Navigation Linien fehlen. Dies führt zu Einfaltungsartefakten, die in Form von gestörten Bildintensitäten in den Rekonstruktionen zu sehen sind und die diagnostische Aussagekraft einschränken. Um Einfaltungsartefakte zu vermeiden sollte der Schwellenwert so gewählt werden, dass nach der Datenauswahl keine Linien fehlen. Aus dieser Anforderung lässt sich ein maximaler Schwellenwert ableiten. Akzeptiert man dagegen sehr viele Daten, zeichnen sich die Rekonstruktionen durch erhöhtes Signal und das vermehrte Auftreten von Bewegungsartefakten aus. In diesem Fall müsste der Arzt entscheiden, ob Bewegungsartefakte die Diagnostik zu stark beeinflussen. Wählt man den Schwellenwert so, dass weder Linien fehlen noch zu viel Bewegung erlaubt wird, erhält man Rekonstruktionen die sich durch einen scharfen Diaphragmaübergang auszeichnen und in denen noch kleinste Gefäße auch in der Nähe des Diaphragmas deutlich zu erkennen sind. Hierfür haben sich Schwellenwerte, die zu einer Datenakzeptanz von ca. 40 % führen als günstig erwiesen. Um Einfaltungsartefakte auf Grund der retrospektiven Datenauswahl zu verhindern, muss das Bildgebungsvolumen mehrfach abgetastet werden. Dadurch wird gewährleistet, dass für die letztendliche Rekonstruktion ausreichend Daten zur Verfügung stehen, wobei mehrfach akzeptierte Daten gemittelt werden. Dies spielt auf Grund der niedrigen Protonendichte der Lunge eine wesentliche Rolle in der Rekonstruktion hochaufgelöster Lungendatensätze. Weiterhin führt das Mitteln von mehrfach akzeptierten Daten zu einer Unterdrückung der sogenannten Ghost Artefakte, was am Beispiel der Herzbewegung in der Arbeit gezeigt wird. Da die Messungen unter freier Atmung durchgeführt werden und keine zusätzlichen externen Messgeräte angeschlossen werden müssen, stellte die Untersuchung für die Patienten in dieser Arbeit kein Problem dar. Im ersten Teil dieser wurde Arbeit gezeigt, dass sich mit Hilfe des DC Signales als Navigator und einer retrospektiven Datenauswahl das gesamte Lungenvolumen in hoher dreidimensionaler Auflösung von beispielsweise 1.6 x 1.6 x 4 mm3 innerhalb von 13 min. darstellen lässt. Die Anwendbarkeit der vorgestellten Methode zur Bewegungskorrektur wurde neben Probanden auch an Patienten demonstriert. Da wie bereits beschrieben das Bildgebungsvolumen mehrfach abgetastet werden muss, wiederholt sich auch die Abfolge der für die Bildgebung verantwortlichen Gradienten periodisch. Da sich der Atemzyklus aber auch periodisch wiederholt, kann es zu Korrelationen zwischen der Atmung und den wiederholten Messungen kommen. Dies führt dazu, dass auch nach vielen wiederholten Messungen immer noch größere Bereiche fehlender Linien im k-Raum bleiben, was zu Artefakten in den Rekonstruktionen führt. Dies konnte im Falle der konventionellen Bewegungskorrektur in den Gatingmasken, die die Verteilung und Häufigkeit der einzelnen akzeptierten Phasenkodierschritte im k-Raum zeigen, beobachtet werden. Da eine vorsätzliche Unterbrechung der Atemperiodizität (der Patient wird dazu angehalten, seine Atemfrequenz während der Messung absichtlich zu variieren) zur Vermeidung der angesprochenen Korrelationen nicht in Frage kommt, musste die Periodizität in der Datenaufnahme unterbrochen werden. In dieser Arbeit wurde dies durch eine quasizufällige Auswahl von Phasen- und Partitionskodiergradienten erreicht, da Quasizufallszahlen so generiert werden, dass sie unabhängig von ihrer Anzahl einen Raum möglichst gleichförmig ausfüllen. Die quasizufällige Datenaufnahme führt deshalb dazu, das sowohl akzeptierte als auch fehlende Linien nach der Bewegungskorrektur homogen im k-Raum verteilt auftreten. Vergleicht man das auftreten von Ghosting zeichnen sich die quasizufälligen Rekonstruktionen im Vergleich zur konventionellen Datenaufnahme durch eine verbesserte Reduktion von Ghost Artefakten aus. Dies ist auf die homogene Verteilung mehrfach akzeptierter Linien im k-Raum zurückzuführen. Die homogenere Verteilung von fehlenden Linien im k-Raum führt weiterhin zu einer wesentlich stabileren Rekonstruktion fehlender Linien mit parallelen MRT-Verfahren (z.B. iterativem Grappa). Dies wird umso deutlicher je höher der Anteil fehlender Linien im k-Raum wird. Im Falle der konventionellen Datenaufnahme werden die zusammenhängenden Bereiche fehlender Linien immer größer, was eine erfolgreiche Rekonstruktion mit iterativem Grappa unmöglich macht. Im Falle der quasizufälligen Datenaufnahme dagegen können auch Datensätze in denen 40% der Linien fehlen einfaltungsartefaktfrei rekonstruiert werden. Im weiteren Verlauf der Arbeit wurde gezeigt, wie die Stabilität der iterativen Grappa Rekonstruktion im Falle der quasizufälligen Datenaufnahme für eine erhebliche Reduktion der gesamten Messzeit genutzt werden kann. So ist in einer Messzeit von nur 74s die Rekonstruktion eines artefaktfreien und bewegungskorrigierten dreidimensionalen Datensatzes der menschlichen Lunge mit einer Auflösung von 2 x 2 x 5 mm3 möglich. Des Weiteren erlaubt die quasizufällige Datenaufnahme in Kombination mit iterativem Grappa die Rekonstruktion von Datensätzen unterschiedlicher Atemphasen von Inspiration bis Expiration (4D Bildgebung). Nach einer Messzeit von 15min. wurden 19 unterschiedliche Atemzustände rekonstruiert, wobei sich der Anteil der fehlenden Linien zwischen 0 und 20 % lag. Im Falle der konventionellen Datenaufnahme wäre eine wesentlich längere Messzeit nötig gewesen, um ähnliche Ergebnisse zu erhalten. Zum Schluss soll noch ein Ausblick über mögliche Weiterentwicklungen und Anwendungsmöglichkeiten, die sich aus den Erkenntnissen dieser Arbeit ergeben haben, gegeben werden. So könnte das quasizufällige Aufnahmeschema um eine Dichtegewichtung erweitert werden. Hierbei würde der zentrale k-Raum Bereich etwas häufiger als die peripheren Bereiche akquiriert werden. Dadurch sollte die iterative Grappa Rekonstruktion noch stabiler funktionieren und Ghost Artefakte besser reduziert werden. Die Verteilung der Linien sollte allerdings nicht zu inhomogen werden, um größere Lücken im k-Raum zu vermeiden. Darüber hinaus könnte die vorgestellte Methode der Bewegungskompensation auch für die Untersuchung anderer Organe oder Körperteile verwendet werden. Voraussetzung wäre lediglich das Vorhandensein dezidierter Spulenanordnungen, mit denen die Bewegung nachverfolgt werden kann. So ist beispielsweise eine dynamische Bildgebung des frei und aktiv bewegten Knies möglich, wobei zwischen Beugung und Streckung durch die erste Ableitung des zentralen k-Raum Signales unterschieden werden kann. Dies kann zusätzliche Diagnoseinformationen liefern oder für Verlaufskontrollen nach Operationen benutzt werden [15]. Eine Weiterentwicklung mit hohem klinischen Potential könnte die Kombination der in dieser Arbeit vorgestellten retrospektiven Bewegungskorrektur mit einer Multi- Gradienten-Echo Sequenz darstellen. Hierzu musste die bestehende Sequenz lediglich um eine mehrfache Abfolge von Auslesegradienten innerhalb einer Anregung erweitert werden. Dies ermöglicht eine bewegungskorrigierte voxelweise Bestimmung der transversalen Relaxationszeit T∗2 in hoher räumlicher Auflösung. Unter zusätzlicher Sauerstoffgabe kann es zu einer Veränderung von T∗2 kommen, die auf den sogenannten BOLD Effekt (Blood Oxygen Level Dependent) zurückzuführen ist. Aus dieser Änderung könnten Rückschlüsse auf hypoxische Tumorareale gezogen werden. Da diese eine erhöhte Strahlenresistenz aufweisen, könnte auf diese Bereiche innerhalb des Tumors eine erhöhte Strahlendosis appliziert und so möglicherweise Behandlungsmisserfolge reduziert werden. Gleichzeitig kann durch die 4D Bildgebung eine mögliche Tumorbewegung durch die Atmung erfasst und diese Information ebenfalls in der Bestrahlungsplanung benutzt werden. Die Lungen MRT könnte somit um eine hochaufgelöste dreidimensionale funktionelle Bildgebung erweitert werden. N2 - The goal of this work was to depict the whole lung volume by MRI in high spatial resolution. To obtain sufficient signal for a reliable diagnosis despite the inherently low proton density of the lung and the requested high spatial resolution, total acquisition times of a few minutes are mandatory. Simultaneously, the measurements should be performed under free breathing conditions making patient examinations more comfortable or possible for patients with limited breath holding capabilities. However, free breathing leads to motion artifacts which can severely influence the diagnostic value of the images and hence have to be avoided. To compensate for motion the prevalent breathing pattern has to be detected. This can be achieved by external measurement devices such as a respiration belt or a spirometer or by conventional navigator echoes using an additional excitation pulse. Drawbacks of these methods are that the respiratory motion is detected only indirectly, that electronic devices have to be used near the MRI machine and the patients have to be prepared and are strongly restricted. Furthermore, additional excitation pulses will prolong the total acquisition time and may affect the magnetization adversely. To overcome these limitations of motion detection in the present work, the image as well as the navigator data was acquired within one excitation of a FLASH sequence. The resulting central k-space signal (DC signal) after rephasing of all imaging gradients was used as a navigator signal. The DC signal represents the sum of all signals that can be detected with a single receiver coil element. If the liver is for example moving in the sensitivity area of one coil element due to breathing, an increased DC signal will be detected. Depending on their local position on the body the locally confined coil elements are able to track respiratory motion. The time course of the DC signal of the selected coil element for respiratory motion compensation will depict periodic signal variations accordingly. Additionally, respiratory phases of expiration can be distinguished from inspiratory phases because the resting times in end-expiratory phases are usually longer compared to end-inspiratory phases. The DC signal can be acquired either before or after the actual image data acquisition within one excitation. The short T2* of the human lung tissue leads to a rapid signal decay after the excitation. As shown in this thesis, the DC signal should be acquired after the image data within one excitation. This approach allows for echo time (TE) reduction of 0.3 ms leading to a signal benefit of approximately 20 %. Simultaneously, the remaining signal after image data acquisition and rephasing of all imaging gradients is still sufficient to track respiratory motion and can therefore be used for motion compensation of the acquired data. In order to compensate for motion retrospectively, threshold values for data acceptance have to be defined. Setting the threshold value, neither too less nor too much data should be accepted. Accepting very few data leads to sharp transition between the lung and the diaphragm because not much motion is allowed in the reconstruction process. On the other hand, disturbed signal intensity can be observed because of under-sampling artifacts due to missing lines after gating. These artifacts can restrict the diagnostic value of the reconstructions. Therefore, the selected threshold value should lead to a fully sampled k-space after gating. This requirement can be used to define the maximum threshold value for data acceptance. On the contrary, accepting very much data leads to higher signal intensity but also to more distinctive motion artifacts. In this case, the physician has to decide whether the motion artifacts affect his diagnosis too much. A moderate threshold value leads to a fully sampled k-space as well as good motion artifact compensation. This results in reconstructions that are characterized by a sharp depiction of small vessels even near the diaphragm. For this, threshold values leading to a data acceptance of about 40 % turned out to be beneficial. To avoid under-sampling artifacts because of retrospective gating, the imaging volume has to be acquired several times. This ensures that enough data is available for the final reconstruction whereas multiple accepted data is averaged. Averaging is essential for the reconstruction of high resolution data sets because of the inherently low proton density of the lung. Furthermore it leads to the reduction of ghost artifacts as is shown using the example of heart motion in this work. As no external measurement devices were used and the data was acquired under free breathing conditions the examinations posed no problem for the patients within this work. It was shown so far that the DC signal in combination with retrospective gating can be used to reconstruct high resolution 3d lung data sets with a resolution of 1.6 x 1.6 x 4 mm3 within 13 min., for instance. The applicability of the presented method for motion compensation was shown for volunteers as well as patients. Since as already described the imaging volume must be acquired several times, the series of gradients for spatial encoding are repeated periodically. As the respiratory cycle is periodically as well, correlations between the repeated measurements and the breathing cycle can occur. Therefore, even after many repeated measurements large areas of missing k-space lines can remain, leading to artifacts in the reconstructions. This can be observed in the gating masks, showing the distribution of accepted and missing lines in k-space, in case of conventional motion compensation used in this work so far. To avoid the aforementioned correlations, the periodicity in the repeated acquisitions has to be interrupted because of suspending the periodic breathing pattern of patients deliberately would be a serious intervention and is therefore ineligible. This was accomplished by a quasi-random selection of the phase and partition encoding gradients as quasi-random numbers are generated to fill the space as uniformly as possible regardless of their number. Therefore, accepted lines as well as missing lines are uniformly distributed in k-space after retrospective gating. A more uniform distribution of multiple accepted k-space lines in case of quasirandom sampling leads to an improved reduction of Ghost-Artifacts compared to conventional sampling. Furthermore, the more uniform distribution of missing kspace lines leads a considerably more stable reconstruction of missing lines using parallel imaging techniques (as iterative Grappa for example). This is getting more distinct the higher the proportion of missing k-space lines is. The contiguous areas of missing k-space lines are becoming increasingly large in case of conventional sampling, making a successful reconstruction using iterative Grappa impossible. In contrast, quasi-random sampling enables for the successful reconstruction of artifact free images even when 40 % of the acquired lines were missing after retrospective gating. In addition, the stability of the iterative GRAPPA reconstructions in case of quasirandom sampling allows for a substantial reduction of the total acquisition time. Thus, an artifact free motion compensated data set of 2 x 2 x 5 mm3 resolution could be reconstructed for a measurement time of only 74s. Furthermore, quasi-random sampling in combination with iterative Grappa enables for the reconstruction of data sets of different respiratory phases from inspiration to expiration (4d imaging). Accordingly, 19 different respiratory phases could be reconstructed after 15min of data acquisition. The percentage of missing lines was between 0 and 20 %. Hence, in case of conventional sampling a considerably longer measurement time would have been required to achieve similar results. KW - Kernspintomografie KW - Retrospektive Bewegungskorrektur KW - Magnetresonanztomographie KW - Lungenbildbgebung KW - freie Atmung KW - Retrospective Motion Compensation KW - DC-Gating KW - Lung Imaging KW - free breathing KW - Lunge Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-124084 ER - TY - THES A1 - Sturm, Volker Jörg Friedrich T1 - \(^{19}F\) Magnetresonanztomographie zur Bildgebung von Infektionen im Zeitverlauf T1 - \(^{19}F\) magnetic resonance imaging to monitor the timecourse of bacterial infections in vivo N2 - Im Rahmen dieser Arbeit sollten die Möglichkeiten der MR Tomographie erkundet werden bakterielle Infektionen im Zeitverlauf darzustellen. Genauer gesagt sollte das Potential der MR Tomographie anhand eines durch eine Infektion induzierten lokalisierten Abszesses unter Verwendung dreier unterschiedlicher MRT Methoden untersucht werden: Mittels nativem \(T_2\) Kontrast; der Verwendung von superparamagnetischen Eisenoxid Partieln (USPIO) als \(T_2^*\) Kontrastmittel; und dem Einsatz von Perfluorkarbonen (PFC) als \(^{19}F\) MRT Marker (siehe Kapitel 3). Wie erwartet führte die durch die Infektion hervorgerufene Entzündung zu veränderten \(T_2\)-Zeiten, welche auf \(T_2\)-gewichteten MR Bildern eine Lokalisierung des Abszessbereiches erlauben. Jedoch eigneten sich diese Daten aufgrund der graduellen Änderung der \(T_2\)-Zeiten nicht, um eine klare Grenze zwischen Abszess und umliegendem Gewebe zu ziehen. Superparamagnetische Eisenoxidpartikel andererseit haben als MRT Kontrastmittel bereits in den letzten Jahren ihre Fähigkeit unter Beweis gestellt Entzündungen [53, 58, 64] darzustellen. Die Anreicherung dieser Partikel am Rande des Abszesses [53], wie sie auch in unseren MR Daten zu beobachten war, erlaubte eine relativ scharfe Abgrenzung gegenüber dem umgebenden Gewebe in der chronischen Phase der Infektion (Tag 9 p.i.). Hingegen genügte die nur sehr spärlichen Anreicherung von USPIO Partikeln in der akuten Phase der Infektion (Tag 3 p.i.) nicht für eine entsprechende Abgrenzung [58]. Aufgrund der sehr geringen biologischen Häufigkeit und den sehr kurzen Relaxationszeiten von endogenem Fluor eignen sich Perfluorkarbone als Markersubstanz in der MR Tomographie von biologischen Systemen. Insbesondere da PFC Emulsionen durch phagozytierende Zellen aufgenommen werden und im Bereich von Entzündungen akkumulieren [30, 59]. In dieser Arbeit konnte anhand der erhaltenen MRT Daten eine Akkumulation von Perfluorkarbonen nicht nur in der chronischen Phase, sondern auch in der akuten Phase nachgewiesen werden. Diese Daten erlauben somit zu allen untersuchten Zeitpunkten eine Abgrenzung zwischen Infektion und umliegenden Gewebe. Aufgrund der besagten Vorteile wurden die Perfluorkarbone gewählt, um die Möglichkeiten der MR Tomographie zu testen, quantitative Informationen über die schwere der Infektion zu liefern. Als Referenz für die Bakterienbelastung wurden die Biolumineszenzbildgebung (BLI) [49, 50] und die Standardmethode zur Bestimmung der Bakterienbelastung cfu (koloniebildenden Einheiten) herangezogen. Eine Gegenüberstellung der zeitlichen Verläufe der durch die Biolumineszenzbildgebung und durch die cfu erhaltenen Daten liefert eine qualitative Übereinstimmung mit den durch die 19F MR Tomographie erhaltenen Daten. Dies trifft hierbei sowohl auf die über den gesamten Infektionsbereich hinweg summierten Signalamplituden, als auch auf das Volumen zu, in dem Fluor am Ort der Infektion akkumuliert wurde. Im Gegensatz zur Methode der cfu Bestimmung sind die MR Tomographie und die Biolumineszenzbildgebung nicht invasiv und erlauben die Verfolgung des Infektionsverlaufes an einem einzelnen Individuum. Hierzu benötigt, im Gegensatz zur MR Tomographie, die Methode der Biolumineszenzbildgebung jedoch einen speziellen Pathogenstamm. Darüber hinaus ist hervorzuheben, dass die MR Tomographie zudem die Möglichkeit bietet auch morphologische Informationen über den Infektionsbereich und seine Umgebung zu akquirieren. Gerade weil jede dieser Methoden die mit der Infektion einhergehenden Prozesse aus einer leicht anderen Blickrichtung betrachtet, erscheint es sinnvoll diese etablierte Untersuchungsplattform bestehend aus MRT, BLI und cfu über die in dieser Arbeit bearbeitete Fragestellung hinaus näher zu untersuchen. Insbesondere der Aspekt inwieweit die drei Methoden sich gegenseitig ergänzen, könnte einen tieferen Einblick in die Wechselwirkung zwischen Pathogen und Wirt erlauben. Auch wenn für die betrachtete Fragestellung bereits der hierdurchgeführte semiquanitative Ansatz zur Bestimmung der relativen Fluormengen am Ort der Infektion ausreichte, so ist doch im Allgemeinen wünschenswert probenbezogen die Sensitivität der Spule und damit die Güte der Spulenabstimmung zu bestimmen. Hierzu ist jedoch die Aufnahme von \(B_1\)-Karten unabdingbar und wird entsprechend im Kapitel 4 \(Bloch-Siegert B_1^+-Mapping\) näher addressiert. Der Schwerpunkt liegt hierbei, wie der Kapitelname bereits andeutet, auf der Bloch-Siegert Methode, die insbesondere in der präsentierten Implementierung in einer Turbo/ Multi Spin Echo Sequenz eine effiziente Nutzung der relativ langen \(T_\)2-Zeiten der Perfluorkarbone erlaubt. Da zudem die Bloch-Siegert-Methode eine rein phasenbasierte Methode ist, kann neben der aus den Daten erzeugten \(B_1\)-Karte zugleich ein unverfälschtes Magnitudenbild generiert werden, wodurch eine sehr effiziente Nutzung der vorhandenen Messzeit ermöglicht wird. Diese Eigenschaft ist insbesondere für \(^{19}F\) Bildgebung von besonderem Interesse, da hier für jede Messung, aufgrund der üblicherweise relativ geringen Konzentration an Fluoratomen, lange Messzeiten benötigt werden. Zusammenfassend konnte anhand des untersuchten Tiermodells sowohl die Fähigkeit der MR Tomographie nachgewiesen werden Infektionen im Zeitverlauf darzustellen, als auch die Fähigkeit der MR Tomographie quantitative Informationen über den Verlauf der Infektion zu liefern. Desweiteren konnte eine Möglichkeit aufgezeigt werden, welche das Potential hat in vertretbarem Zeitrahmen auch in vivo B1+-Karten auf dem Fluorkanal zu erstellen und so einen zentralen Unsicherheitsfaktor, für Relaxometry und absolute Quantifizierung von \(^{19}F\) Daten in vivo, zu beseitigen. N2 - The main focus of this work is to investigate the potential of magnetic resonance imaging (MRI) to monitor the timecourse of bacterial infections in vivo. More specifically, it focuses on the ability to localize and assess an infection-induced localized bulky abscess using three different MRI methods: the utilization of native \(T_2\) contrast; the usage of super paramagnetic iron oxide nanoparticles (USPIO) as MRI \(T_2^*\) contrast agents; and the application of perfluorcarbons (PFC) as \(^{19}F\) MRI marker (see chapter 3). Study results demonstrated that, as expected the altered \(T_2\) values present in the abscess area permit localization of the infection when using \(T_2\) weighted data. The precise boundary of the abscess, however, could not be determined due to the gradual change of the \(T_2\) values in the area of the infection. Conforming to other studies [53, 58], the MR-detected accumulation of USPIO particles along the abscess rim allowed definition of a fairly exact demarcation line between the abscess and surrounding tissue during the chronic phase of the infection (day 9 p.i.). During the acute phase of the infection (day 3 p.i.), however, the particle accumulation at the abscess rim was too sparse for precise boundary definition [58]. Because of their extremely low biological abundance and the very short relaxation times of endogenous fluorine, PFCs can be imaged background-free in a biological system. Moreover, as emulsified PFCs were taken up by phagocytosing cells and accumulated at the site of inflammation [30, 59], the acquired MRI data showed PFC accumulation during both the chronic and acute phases of infection. It was thus possible to differentiate between the abscess and surrounding tissue at each examined time point. Due to the described advantages, PFCs were chosen to evaluate with MRI the infection severity. As a bacterial burden reference, colony forming units (cfu) and bioluminescence imaging (BLI) [49, 50] were selected. Observation of BLI, cfu and \(^{19}F\) MRI data showed qualitative correlation during the investigated time course. This was true for the accumulated \(^{19}F\) MR signal in the area of infection and for the \(^{19}F\) MR signal volume. Additionally, unlike the cfu method MRI and BLI are non-invasive and thus data can be gathered at multiple time points. However, contrary to BLI, MRI does not require a special pathogen strain. Moreover, it can provide morphological data from an abscess and the surrounding tissue. Because the data delivered by each of these three methods (MRI, BLI and cfu), are based on alternative approaches, additional examinations of the established platform are suggested. For example, the extent to which the methods supplement each other may provide deeper insight into the interaction between pathogen and host. Even though the chosen semi quantitative approach was sufficient in the context of the evaluated issues to estimate the relative fluorine amount at the site of infection, it is in general desirable for each quantification to determine the sensitivity of the coil per sample. To address this issue the Bloch Siegert (BS) based \(B_1\) mapping method implemented in a turbo/ multi spin echo (TSE/MSE) sequence is presented in Chapter 4 Bloch-Siegert \(B_1^+\)-Mapping. Such a sequence allows effective use of the relatively long PFC \(T_2\) times and encodes BS information solely into the phase data. Thus, a \(B_1\) map can be created in addition to the unaltered TSE/MSE magnitude image. In the context of \(^{19}F\) imaging, this is of special interest due to the usually low amounts of fluorine resulting in long measurement times. In conclusion, it was shown that MRI not only enables visualization of the temporal behavior of infections on the investigated animal model, but it can also provide quantitative information about the progress of the infection. Additionally, a method potentially allowing in vivo B1+ mapping was introduced. This is an important step to improve the reliability of relaxometry and absolute quantification of in vivo \(^{19}F\) MRI. KW - Kernspintomografie KW - Bakterielle Infektion KW - 19F MR KW - Perfluorkarbon KW - Infektionsbildgebung KW - Bloch Siegert KW - B1 Mapping KW - Kontrastmittel Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-122851 ER - TY - THES A1 - Hopfgartner, Andreas T1 - Magnetresonanztomographie in der Zahnheilkunde - hochauflösende zahnmedizinische Anwendungen in der MRT mit einer Entwicklung zur Bewegungskorrektur T1 - Magnetic Resonance Imaging in Dentistry – high-resolution dental applications in MRI with development of a method for motion correction N2 - Die zahnmedizinische Behandlung von Erkrankungen der Zähne oder im Bereich der Mundhöhle erfolgt bei Weitem nicht immer aus optischen Gründen. Diese Erkrankungen werden auch mit ernsthaften Erkrankungen in Zusammenhang gebracht. Studien haben gezeigt, dass einige Erkrankungen im Mund- und Zahnbereich zu Herz- und Lungenkrankheiten oder Diabetes führen können. Oftmals erstreckt sich die Pathologie oder Symptomatik von Mund- und Zahnerkrankungen über einen weiten Bereich. In der zahnmedizinischen Klinik kommen daher viele verschiedene diagnostische Apparate zum Einsatz. Allerdings zählt die Magnetresonanztomographie, die sich in anderen Bereichen bereits zum wichtigsten bildgebenden Diagnosetool entwickelt hat, dort noch nicht zu den Standardverfahren. Dabei liegen ihre Vorteile auf der Hand: sie ist bekannt für sehr gute Bildkontraste vor allem zwischen verschiedenen Weichgewebsarten und kommt ohne gefährliche ionisierende Strahlung aus. Wahrscheinlich ist ersteres der Grund, warum die MRT in der Zahnmedizin noch nicht sonderlich vertreten ist, kommt es dort oft auf die kontrastreiche Darstellung von Hartgeweben an. Neueste Entwicklungen und Studien belegen jedoch die vielseitigen Vorteile der MRT auch in diesem Bereich. Ziel dieser Arbeit von der applikativen Seite betrachtet, war es, das enorme Potential der MRT in den vielseitigen Bereichen der Zahnmedizin weiterhin aufzuzeigen. Viele dieser Anwendungen stellen jedoch sehr hohe Anforderungen an die Systeme. Meist sind die darzustellenden Strukturen sehr klein und erfordern eine hohe Auflösung. Während man beim Röntgenverfahren beispielsweise die Energie des Strahles (Dosis) steigern kann, bedeutet dies in der MRT (ohne das Gerät zu wechseln) eine Verlängerung der Messzeit. Gerade im Bereich des Kopfes kommt es oft zu ungewollten Bewegungen, die das Ergebnis und die Reproduzierbarkeit der gewonnenen diagnostischen Informationen verschlechtern oder gänzlich unbrauchbar machen. Die grösste Herausforderung dabei ist die dreidimensionale Abformung von Zahnoberflächen in der Prothetik. Dieses Verfahren kann eine aufwändige und unangenehme manuelle Abformung der Zähne und die Herstellung eines Zwischengipsmodells ersetzen und ein direktes dreidimensionales Modell der Zahnoberflächen produzieren. Durch die moderne CAD-/CAM-Technik kann daraus vom Zahntechniker direkt eine Zahnrestauration erstellt werden. Daher war ein wichtiger Bestandteil des Projekts dentale MRT die Entwicklung einer Methode zur Erkennung und gleichzeitiger Korrektur von Bewegungen. Verschiedenste Anforderungen waren an die Methode gestellt. Zum einen muss die Methode bereits Bewegungen im Bereich von ~100 µm erkennen, um die Anforderungen an die finale Bildauflösung zu unterschreiten. Bei der dentalen Abformung wird eine 1-Kanal-Empfängerspule verwendet und je nach Messung kann der Patient dabei auf dem Bauch oder Rücken liegen. Weiterhin muss die Bewegungserkennung ohne zusätzliche externe Geräte wie Kameras, deren Sicht z.B. durch den Patienten verdeckt ist, durchführbar sein. Die vorliegende Arbeit deckt also zwei größere Themenblöcke ab. Zum einen wurden in der Arbeit neue Applikationen entwickelt oder weiterentwickelt, um verschiedenen Bereichen der Zahnmedizin den Zugang zu MRTUntersuchungen zu eröffnen. Kapitel 4 beschreibt die Möglichkeit, die Bewegung des Kiefergelenks dynamisch zu erfassen. Es stellte sich in der Arbeit heraus, dass sowohl die Bewegung von Weichgewebeanteilen darstellbar waren, als auch der intraartikuläre Abstand im Kiefergelenk unter Kaubelastung in Echtzeit vermessen werden konnte. Dabei wurde die Bildgebungssequenz und der zugehörige Rekonstruktionsalgorithmus so entwickelt, dass die Daten flexibel und ohne Vorwissen akquiriert und aufbereitet werden können. Hierbei konnten verschiedenen Pathologien anhand der dynamischen Bilder sichtbar gemacht werden und die dynamische MRT konnte Erkrankungen erkennen, die mit anderen Mitteln nicht sichtbar waren. Die vielen diagnostischen Möglichkeiten, die dadurch entstehen sind bisher noch nicht untersucht und sollten durch großangelegte Studien untersucht und belegt werden. Kapitel 5 beschreibt die Ergebnisse einer großangelegten Studie im Bereich der dentomaxillären Bildgebung . Die diagnostischen Möglichkeiten der MRT für die kieferorthopädische Anwendung liegen klar auf der Hand. Die typischen Patienten in der Kieferorthopädie sind Kinder und Jugendliche. Die Abwesenheit von gewebsschädigender Strahlung ist hier ein besonderer Vorteil der MRT. Eine Messung dauert zudem nach diversen Weiterentwicklungen der Methode nur noch 2 (bzw. 4) Minuten. Die Auflösung in den gerenderten Bildern beträgt 0.25x0.25x0.5 mm. Mit der Methode konnte unter anderem die Geminisierung einer Zahnwurzel und der Abstand des Zahnmarks zur Zahnoberfläche (Zahnschmelz) dargestellt und vermessen werden. Kapitel 6 stellt Neuentwicklungen im Bereich der dentalen Abformung von Zahnoberflächen dar. Hier wurde eine neue Methode entwickelt um den Patientenkomfort bei der Messung zu steigern und so Bewegungen im Vorhinein zu unterbinden. Bei der alten Methode liegt der Patient auf dem Bauch und ein großer Teil der Mundhöhle ist mit Kontrastmittel befüllt. Durch die Verwendung einer präparierten Tiefziehschiene kann das Kontrastmittel nun lokal appliziert werden und eine Messung in Rückenlage das Patienten ist somit problemlos möglich. Die damit verbundene Reproduzierbarkeit der Abformungsergebnisse wäre durch eine großangelegte Studie zu zeigen. Die Hauptaufgabe der vorliegenden Dissertation war es, eine Methode zur Bewegungskorrektur zu entwickeln, die es ohne eine große Anzahl an Zusatzgeräten ermöglicht, die Bewegung eines Subjekts während der Messung zu erfassen und dementsprechend zu korrigieren. Diese neue Methode, gestützt auf einer Messung eines MRT-aktiven Markers der am Subjekt angebracht wird, beruht außer der Verwendung des Markers nur auf MRT-Hardware. Die Methode wird in Kapitel 8 vorgestellt. Da es sich bei der Methode um eine Neuentwicklung handelt, war es in erster Linie wichtig, die Einflüsse der verschiedenen Parameter, die sich auf die Positionierungsgenauigkeit auswirken, abzuschätzen und letzten Endes festzulegen. Dies wurde in mehreren Vorstudien, Experimenten und Computersimulationen abgehandelt. In der Arbeit konnte durch Validierungsexperimente gezeigt werden dass sich mit dem bildbasierten Navigator Bewegungen im Genauigkeitsbereich von ~50 µm (Translation) und ~0.13◦(Rotation) detektieren lassen. Mit den Positionsinformationen lassen sich MRT-Daten retrospektiv korrigieren oder idealerweise das Bildgebungsvolumen in Echtzeit anpassen um Inkonsistenzen in den Daten im Vorhinein vorzubeugen. Durch Bewegung beeinträchtigte in-vivo Daten konnten so mit der Methode korrigiert werden und anhand eines geeigneten Phantoms konnte die Verbesserung der Erkennung von Kanten, wie sie beispielsweise bei der dentalen Abformung angewandt wird, gezeigt werden. Die kontinuierlichen Entwicklungen in den Bereichen Hard-, Software und Algorithmik ermöglichen weitere hochauflösende Anwendungen. In Kapitel 9 sind die Ergebnisse einer Studie gezeigt, die sich mit der Analyse der Handbewegungen während einer Messung beschäftigt. Für eine hochauflösenden Darstellung der Handanatomie bei 7 T ist eine Unterbindung der Handbewegung sehr wichtig. Um ein geeignetes Design für eine Empfängerspule zu entwerfen, die Bewegungen der Hand unterbindet, wurde eine qualitative Bewegungsanalyse der Hand in mehreren verschiedenen Positionen durchgeführt. Durch Vergleich der Ergebnisse konnte so auf geeignete Designs zurückgeschlossen werden. N2 - The treatment of the teeth or diseased of the oral cavity is by far not only administered for aesthetic reasons. These diseases are sometimes also associated with other serious diseases. Studies have shown that some diseases of the mouth, the gingiva or the surrounding area can lead to heart and lung disease or diabetes. Oftentimes the pathology or symptomatology of dental or oral diseases extends to a wide area. In the dental clinic many different diagnostic devices are used. However, magnetic resonance imaging, which has developed in other areas as the most important diagnostic imaging tool, is not frequently used in dentistry to the present day, although their advantages are obvious: it is known for excellent image contrast, mainly between different soft tissues and comes without hazardous ionizing radiation. The former is probably the reason why the MRI is not yet a standard method in dentistry: here in most cases the contrast of hard tissues is of relevance. However, recent developments and studies demonstrated the versatile advantages of MRI in this area. The aim of this work as seen from the perspective of application, was to continuously show the enormous potential of MRI in the diverse areas of dentistry. However, many of these applications put very high requirements on the systems. Usually structures to display are very small and require very high resolution. To improve the resolution while using the X-ray method, e.g., one can increase the beam energy (dose). In MRI (without changing the MRT scanner) this results in an extension of measurement time. Especially in the area of the head this oftentimes leads to unwanted movements during the measurement time that worsen the outcome and reproducibility of the obtained diagnostic information or making it completely useless. The biggest challenge is the measurement of a three-dimensional impression of the tooth surfaces in prosthetics. This process can replace a complex and unpleasant manual impression of the teeth and avoid the production of an intermediate plaster model. Using MRT techniques, a direct three- dimensional model of the tooth surfaces can be produced. By modern CAD/CAM technology, a dental restoration can be directly manufactured by the dental technician using the digital 3D model. Therefore, an important task of the project was the development of a dental MRT method for the detection and correction of movements. Various requirements were imposed on the method. Firstly, the method must be able to detect movements in the range of ~100 µm to fall below the requirements of the final image resolution. For the acquisition of the contrast agent’s signal, a 1-channel receiver coil is used and depending on the measurement, the patient can lie prone or supine. Furthermore, the motion detection system must work without extensive external devices such as cameras, whose direct vision may be obscured by the patient, e.g. This thesis covers two major subject areas. Firstly, new applications and methods have been developed and further developed in order to provide the various fields of dentistry access to MRT techniques. Chapter 4 describes the possibility to image the motion of the temporomandibular joint dynamically in real-time. In this work it turned out that both, the movement of the soft tissue components were represented, as well as the intra-articular distance in the TMJ could be measured during mastication (under load) in real-time. Here, the imaging sequence and the corresponding reconstruction algorithm were designed such that the data can be acquired without a prioiri knowledge and processed flexibly. MRT showed different pathologies in the images and dynamic MRT could detect some diseases that could not be diagnosed by other means. The emerging diagnostic possibilities should be investigated and the results verified by large-scale studies. Chapter 5 describes the results of dento-maxillary MRT imaging, supported by a large-scale study. The diagnostic capabilities of MRI for orthodontic applications are obvious. The typical patient in orthodontics are children and adolescents. The absence of tissue-damaging radiation is a particular advantage of MRI here. After various developments, the acquisition time of a measurement lasted depending on the method only 2 (4) minutes. The resolution in the rendered images was 0.25x0.25x0.50 mm3. Using the proposed method, among other things a geminisation of a tooth root could be shown and the distance of the dental pulp to the tooth surface (enamel) measured. Chapter 6 presents new developments in the field of digital impressions of tooth surfaces. Here, a new method was developed in order to increase patient comfort during the measurement. This approach helps to prevent movements of the subject in advance. With the old method, the patient lies prone and a large part of the oral cavity is filled with contrast agent. By using a prepared dental cast, the contrast agent can be applied locally and hence the patient may lay supine during the measurement. The associated reproducibility of dental impressions should be shown through a large-scale study. The main task of this thesis was to develop a method for motion correction that allows to detect the movement of a subject during the measurement without a large number of additional devices and correct the acquired data accordingly. This new navigator method, based on the measurement of a MRT-active marker attached to the subject, makes use of MRT hardware only, except for the additional marker. The method is described in chapter 8. Since this is a new development, it was important to primarily estimate the effects of the various parameters and their impact on the positioning accuracy. This has been evaluated in several preliminary studies, experiments and computer simulations. By validation experiments it was shown in the studies that the image-based navigator detects movements with an accuracy of ~50 µm(translation) and ~0.13◦ (rotation). With the position information obtained from the navigator, the MRT data can be corrected retrospectively or the volume of interest can be adjusted in real-time during the imaging process to prevent inconsistencies in the data in advance. In-vivo MRT data impaired by motion of a subject during the measurement could be corrected using the MoCoLoCo method. By using an appropriate phantom and simulation a movement, it could be shown that using the proposed method, the quality of edge detection (as used in dental impressions, e.g.) could be restored. Various new high-resolution applications emerged due to the continuous development in hardware, software and algorithms. In chapter 9, the results of a study are presented, which deals with the analysis of shivering movements of the hand during a measurement. For a high-resolution depiction of hand anatomy at 7 T, a suppression of the hand movement is very important. In order to develop an optimal design for a hand receiver coil, a qualitative analysis of the hand movement in several different positions was performed. By comparison of the results, a suitable coil design could be developed. KW - Kernspintomografie KW - Kernspintomografie KW - Zahnmedizin Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-122557 ER - TY - THES A1 - Alkonyi, Balint T1 - Differential imaging characteristics and dissemination potential of pilomyxoid astrocytomas versus pilocytic astrocytomas T1 - Pilomyxoides Astrozytom und pilozytisches Astrozytom: Vergleich anhand der Bildgebung N2 - Background and Aims: PMA is a recently described rare tumor entity occuring most often in young children. Due the worse outcome of PMA-patients as compared to children with pilocytic astrocytoma (PA), it has received a grade II assignment in the latest WHO classification. Nevertheless, increasing evidence suggests that the two tumor types are indeed pathologically and genetically related. The radiological differentiation of PMAs from PAs is challenging and the limited available data could not yet provide unequivocal distinguishing imaging features. Furthermore, it is not completely clarified whether PMA cases are associated with a higher rate of CSF dissemination compared to similarly young patients with PA. The aim of our study was firstly to compare MR/CT imaging features of these tumors, and secondly, to evaluate the occurrence of CSF dissemination. Material and Methods: The study population included 15 children with PMA and 32 children with PA. A third group consisted of eight children with PAs with focal pilomyxoid features. All cases had been registered in the German multicenter SIOP/HIT-LGG trials. The initial MRIs (and CT scans, if available) at establishing the diagnosis were retrospectively analyzed according to standardized criteria and the findings compared between PMAs and PAs. Furthermore, we compared the occurrence of imaging evidences of CSF tumor dissemination between children with PMA and PA, respectively. Results: The imaging appearance of PMAs and PAs was very similar. However, PAs tended to show more frequently cystic components (p=0.03). As opposed to PAs, PMAs did not have large tumor cysts. We did not find differences with respect to tumor size and tumor margin. Gadolinium enhancement of PMAs was significantly more frequently homogeneous (p=0.006). PMAs appeared to show more often intratumoral hemorrhages (p=0.047). Furthermore, suprasellar PMAs tended to have a more homogeneus texture on T2-weighted MR images (p=0.026). Within the subgroup < 6 years of age the PMA histology tended to have a larger effect on the occurrence of CSF dissemination than the age (p=0.05 vs.0.12). Conclusions: Although the radiological appearance of PMAs and PAs is similar, some imaging features, like enhancement pattern or presence of cysts or hemorrhage may help differentiating these low-grade gliomas. Our results corroborate previous scarce data suggesting higher rate of CSF dissemination in PMAs, even in the youngest patient population. Thus, in young children with a chiasmatic-hypothalamic tumor suggestive of a PMA, an intensive search for CSF dissemination along the entire neuraxis should be performed. N2 - Hintergrund und Ziele: Das PMA ist eine seltene benigne Tumorentität (WHO Grad II nach der Klassifikation von 2007), welche vorwiegend bei Kleinkindern vorkommt. Dem PMA wird eine schlechtere Prognose als dem PA zugeschrieben. Die radiologische Unterscheidung zwischen PMA und PA erscheint schwierig. Zudem ist unklar inwieweit das PMA mit einer höheren Meningeoserate assoziiert sein könnte, verglichen zum PA bei Patienten vergleichbaren Alters. Das Ziel unserer Studie war es, bildgebende Kriterien sowie das Auftreten einer Liquordissemination zwischen beiden Tumorarten zu vergleichen. Material und Methoden: Das untersuchte Kollektiv umfasste 15 Kinder mit PMA sowie 32 Kinder mit PA. Zudem eine dritte Gruppe umfasste 8 Kinder mit PA mit fokalen pilomyxoiden Merkmale. Die initialen MRT-Untersuchungen bei Diagnosestellung wurden anhand standardisierter bildgebender Kriterien analysiert. Anhand der MRT-Bildgebung wurde des Weiteren das Auftreten einer Liquordissemination bei Diagnose und im Verlauf verglichen. Ergebnisse: PMAs und PAs stellten sich bildmorphologisch ähnlich dar. PAs zeigten öfters zystische Tumorkomponente (p=0.03). Im Gegensatz zu PAs, wiesen PMAs keine großen Tumorzysten auf. Auch die Kontrastmittelanreicherung war häufiger homogen in PMAs (p=0.006). Intratumorale Hämorrhagien konnten häufiger in PMAs beobachtet werden (p=0.047). Desweiteren haben sich supraselläre PMAs in der T2-Wichtung häufiger homogener dargestellt (p=0.026). Die histologische Diagnose PMA schien einen größeren Effekt (p=0.05) auf das Auftreten einer Liquordissemination zu haben als das Alter selbst (p=0.12). Schlussfolgerungen: Obwohl sich PMA und PA in der Bildgebung sehr ähnlich darstellen, identifizierten wir einige Kriterien welche beitragen könnten diese Tumorentitäten zu unterscheiden. Bei radiomorphologischen Kriterien, die für ein PMA sprechen, sollte bei den jungen Kindern gezielt nach einer Liquordisseminierung gesucht werden. KW - Astrozytom KW - Kernspintomografie KW - Pilomyxoid astrocytoma KW - Pilocytic astrocytoma KW - Imaging KW - Dissemination KW - Pilomyxoides Astrozytom KW - Pilozytisches Astrozytom KW - Bildgebung KW - Dissemination Y1 - 2014 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-116062 ER - TY - THES A1 - Ye, Yuxiang T1 - Molecular and Cellular Magnetic Resonance Imaging of Myocardial Infarct Healing T1 - Molekulare und zelluläre Magnetresonanztomographie der Wundheilung nach Myokardinfarkt N2 - Myokardinfarkte (MI) sind eine der häufigsten Todesursachen weltweit. Eine rechtzeitige Wiederherstellung des koronaren Blutflusses im ischämischen Myokard reduziert signifikant die Sterblichkeitsrate akuter Infarkte und vermindert das ventrikuläre Remodeling. Überlebende MI-Patienten entwickeln jedoch häufig eine Herzinsuffizienz, die mit einer reduzierten Lebensqualität, hohen Sterblichkeitsrate (10% jährlich), sowie hohen Kosten für das Gesundheitssystem einhergeht. Die Entwicklung der Herzinsuffizienz nach einem MI ist auf den hohen Verlust kontraktiler Kardiomyozyten, während der Ischämie-Reperfusion zurückzuführen. Anschließende komplexe strukturelle und funktionelle Veränderungen resultieren aus Modifikationen des infarzierten und nicht infarzierten Myokards auf molekularer und zellulärer Ebene. Die verbesserte Überlebensrate von Patienten mit akutem MI und das Fehlen effizienter Therapien, die die Entwicklung und das Fortschreiten des ventrikulären Remodelings verhindern, führen zu einer hohen Prävalenz der Herzinsuffizienz. Die kardiale Magnetresonanztomographie (MRT) ist eine wichtige Methode zur Diagnose und Beurteilung des Myokardinfarktes. Mit dem technologischen Fortschritt wurden die Grenzen der MRT erweitert, so dass es heute möglich ist, auch molekulare und zelluläre Ereignisse in vivo und nicht-invasiv zu untersuchen. In Kombination mit kardialer Morphologie und Funktion könnte die Visualisierung essentieller molekularer und zellulärer Marker in vivo weitreichende Einblicke in den Heilungsprozess infarzierter Herzen liefern, was zu neuen Erkenntnissen für ein besseres Verständnis und bessere Therapien des akuten MI führen könnte. In dieser Arbeit wurden Methoden für die molekulare und zelluläre kardiale MRT-Bildgebung der Inflammation und des Kalziumstroms im Heilungsprozess des akuten Myokardinfarktes in vivo in einem Rattenmodel mit klinischer Relevanz etabliert. N2 - Myocardial infarction (MI) is a leading cause of death worldwide. Timely restoration of coronary blood flow to ischemic myocardium significantly reduces acute infarct mortality and attenuates ventricular remodeling. However, surviving MI patients frequently develop heart failure, which is associated with reduced quality of life, high mortality rate (10% annually), as well as high healthcare expenditures. The main processes involved in the evolution of heart failure post-MI are the great loss of contractile cardiomyocytes during ischemia-reperfusion and the subsequent complex structural and functional alterations, which are rooted in modifications at molecular and cellular levels in both the infarcted and non-infarcted myocardium. However, we still lack efficient treatments to prevent the development and progression of left ventricular remodeling. The improved survival rate of acute MI patients combined with the lack of effective therapy for post-MI remodeling contributes to the high prevalence of heart failure. Cardiac Magnetic Resonance Imaging (MRI) is an important tool for diagnosis and assessment of MI. With the advancement of this technology, the frontier of MRI has been extended to probing molecular and cellular events in vivo and non-invasively. In combination with assessment of morphology and function, the visualization of essential molecular and cellular markers in vivo could provide comprehensive, multifaceted views of the healing process in infarcted hearts, which might give new insight for the treatment of acute MI. In this thesis, molecular and cellular cardiac MRI methods were established to visualize and investigate inflammation and calcium flux in the healing process of acute MI in vivo, in a clinically relevant rat model. KW - Kernspintomografie KW - Herzinfarkt KW - magnetic resonance imaging KW - myocardial infarction KW - Wundheilung KW - NMR-Tomographie Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-72514 ER - TY - THES A1 - Benkert, Thomas T1 - Neue Steady-State-Techniken in der Magnetresonanztomographie T1 - Novel Steady-State Techniques for Magnetic Resonance Imaging N2 - Die bSSFP-Sequenz kombiniert kurze Akquisitionszeiten mit einem hohen Signal-zu-Rausch-Verhältnis, was sie zu einer vielversprechenden Bildgebungsmethode macht. Im klinischen Alltag ist diese Technik jedoch bisher - abgesehen von vereinzelten Anwendungen - kaum etabliert. Die Hauptgründe hierfür sind Signalauslöschungen in Form von Bandingartefakten sowie der erzielte T2/T1-gewichtete Mischkontrast. Das Ziel dieser Dissertation war die Entwicklung von Methoden zur Lösung der beiden genannten Limitationen, um so eine umfassendere Verwendung von bSSFP für die MR-Diagnostik zu ermöglichen. Magnetfeldinhomogenitäten, die im Wesentlichen durch Suszeptibilitätsunterschiede oder Imperfektionen seitens der Scannerhardware hervorgerufen werden, äußern sich bei der bSSFP-Bildgebung in Form von Bandingartefakten. Mit DYPR-SSFP (DYnamically Phase-cycled Radial bSSFP) wurde ein Verfahren vorgestellt, um diese Signalauslöschungen effizient zu entfernen. Während für bereits existierende Methoden mehrere separate bSSFP-Bilder akquiriert und anschließend kombiniert werden müssen, ist für die Bandingentfernung mittels DYPR-SSFP lediglich die Aufnahme eines einzelnen Bildes notwendig. Dies wird durch die neuartige Kombination eines dynamischen Phasenzyklus mit einer radialen Trajektorie mit quasizufälligem Abtastschema ermöglicht. Die notwendigen Bestandteile können mit geringem Aufwand implementiert werden. Des Weiteren ist kein spezielles Rekonstruktionsschema notwendig, was die breite Anwendbarkeit des entwickelten Ansatzes ermöglicht. Konventionelle Methoden zur Entfernung von Bandingartefakten werden sowohl bezüglich ihrer Robustheit als auch bezüglich der notwendigen Messzeit übertroffen. Um die Anwendbarkeit von DYPR-SSFP auch jenseits der gewöhnlichen Bildgebung zu demonstrieren, wurde die Methode mit der Fett-Wasser-Separation kombiniert. Basierend auf der Dixon-Technik konnten so hochaufgelöste Fett- sowie Wasserbilder erzeugt werden. Aufgrund der Bewegungsinsensitivät der zugrunde liegenden radialen Trajektorie konnten die Messungen unter freier Atmung durchgeführt werden, ohne dass nennenswerte Beeinträchtigungen der Bildqualität auftraten. Die erzielten Ergebnisse am Abdomen zeigten weder Fehlzuordnungen von Fett- und Wasserpixeln noch verbleibende Bandingartefakte. Ein Nachteil der gewöhnlichen Dixon-basierten Fett-Wasser-Separation ist es, dass mehrere separate Bilder zu verschiedenen Echozeiten benötigt werden. Dies führt zu einer entsprechenden Verlängerung der zugehörigen Messzeit. Abhilfe schafft hier die Verwendung einer Multiecho-Sequenz. Wie gezeigt werden konnte, ermöglicht eine derartige Kombination die robuste, bandingfreie Fett-Wasser-Separation in klinisch akzeptablen Messzeiten. DYPR-SSFP erlaubt die Entfernung von Bandingartefakten selbst bei starken Magnetfeldinhomogenitäten. Dennoch ist es möglich, dass Signalauslöschungen aufgrund des Effekts der Intravoxeldephasierung verbleiben. Dieses Problem tritt primär bei der Bildgebung von Implantaten oder am Ultrahochfeld auf. Als Abhilfe hierfür wurde die Kombination von DYPR-SSFP mit der sogenannten z-Shim-Technik untersucht, was die Entfernung dieser Artefakte auf Kosten einer erhöhten Messzeit ermöglichte. Die mit DYPR-SSFP akquirierten radialen Projektionen weisen aufgrund des angewendeten dynamischen Phasenzyklus leicht verschiedene Signallevel und Phasen auf. Diese Tatsache zeigt sich durch inkohärente Bildartefakte, die sich jedoch durch eine Erhöhung der Projektionsanzahl effektiv reduzieren lassen. Folglich bietet es sich in diesem Kontext an, Anwendungen zu wählen, bei denen bereits intrinsisch eine verhältnismäßig hohe Anzahl von Projektionen benötigt wird. Hierbei hat sich gezeigt, dass neben der hochaufgelösten Bildgebung die Wahl einer 3D radialen Trajektorie eine aussichtsreiche Kombination darstellt. Die in der vorliegenden Arbeit vorgestellte 3D DYPR-SSFP-Technik erlaubte so die isotrope bandingfreie bSSFP-Bildgebung, wobei die Messzeit im Vergleich zu einer gewöhnlichen bSSFP-Akquisition konstant gehalten werden konnte. Verbleibende, durch den dynamischen Phasenzyklus hervorgerufene Artefakte konnten effektiv mit einem Rauschunterdrückungsalgorithmus reduziert werden. Anhand Probandenmessungen wurde gezeigt, dass 3D DYPR-SSFP einen aussichtsreichen Kandidaten für die Bildgebung von Hirnnerven sowie des Bewegungsapparats darstellt. Während die DYPR-SSFP-Methode sowie die darauf beruhenden Weiterentwicklungen effiziente Lösungen für das Problem der Bandingartefakte bei der bSSFP-Bildgebung darstellen, adressiert die vorgestellte RA-TOSSI-Technik (RAdial T-One sensitive and insensitive Steady-State Imaging) das Problem des bSSFP-Mischkontrasts. Die Möglichkeit der Generierung von T2-Kontrasten basierend auf der bSSFP-Sequenz konnte bereits in vorausgehenden Arbeiten gezeigt werden. Hierbei wurde die Tatsache ausgenutzt, dass der T1-Anteil des Signalverlaufs nach Beginn einer bSSFP-Akquisition durch das Einfügen von Inversionspulsen in ungleichmäßigen Abständen aufgehoben werden kann. Ein so akquiriertes Bild weist folglich einen reinen, klinisch relevanten T2-Kontrast auf. Die im Rahmen dieser Arbeit vorgestellte Methode basiert auf dem gleichen Prinzip, jedoch wurde anstelle einer gewöhnlichen kartesischen Trajektorie eine radiale Trajektorie in Kombination mit einer KWIC-Filter-Rekonstruktion verwendet. Somit können bei gleichbleibender oder sogar verbesserter Bildqualität aus einem einzelnen, mit RA-TOSSI akquirierten Datensatz verschiedene T2-Wichtungen als auch gewöhnliche T2/T1-Wichtungen generiert werden. Mittels Variation der Anzahl der eingefügten Inversionspulse konnte ferner gezeigt werden, dass es neben den besagten Wichtungen möglich ist, zusätzliche Kontraste zu generieren, bei denen verschiedene Substanzen im Bild ausgelöscht sind. Diese Substanzen können am Beispiel der Gehirnbildgebung Fett, graue Masse, weiße Masse oder CSF umfassen und zeichnen sich neben den reinen T2-Kontrasten durch eine ähnlich hohe klinische Relevanz aus. Die mögliche Bedeutung der vorgestellten Methode für die klinische Verwendung wurde durch Messungen an einer Gehirntumorpatientin demonstriert. Zusammenfassend lässt sich sagen, dass die im Rahmen dieser Dissertation entwickelten Techniken einen wertvollen Beitrag zur Lösung der eingangs beschriebenen Probleme der bSSFP-Bildgebung darstellen. Mit DYPR-SSFP akquirierte Bilder sind bereits mit bestehender, kommerzieller Rekonstruktionssoftware direkt am Scanner rekonstruierbar. Die Software für die Rekonstruktion von RA-TOSSI-Datensätzen wurde für Siemens Scanner implementiert. Folglich sind beide Methoden für klinische Studien einsetzbar, was gleichzeitig den Ausblick dieser Arbeit darstellt. N2 - The bSSFP sequence combines short acquisition times with a high signal-to-noise ratio, and is therefore a promising imaging technique. However, aside from a few applications, this method is hardly established in the clinical routine. The main reasons for this are signal voids that arise as banding artifacts and the obtained T2/T1-weighted mixed contrast. The goal of this dissertation was to develop strategies to overcome these limitations and allow for a more widespread use of bSSFP for MR diagnostics. In bSSFP imaging, magnetic field inhomogeneities, which are mainly caused by susceptibility differences and imperfections of the scanner hardware, manifest as banding artifacts. In order to efficiently remove these artifacts from the image, DYnamically Phase-cycled Radial bSSFP (DYPR-SSFP) was proposed. While existing methods rely on the acquisition and subsequent combination of several separate bSSFP images, banding removal with DYPR-SSFP requires the acquisition of only a single data set. This is achieved by combining a dynamic phase-cycle with a radial trajectory and a quasi-random acquisition scheme. The individual components of this method can be implemented with little effort. Furthermore, no specific reconstruction scheme is necessary, guaranteeing the broad applicability of the developed approach. DYPR-SSFP outperformed conventional methods for banding removal both in robustness and scan time. In order to demonstrate the applicability of DYPR-SSFP beyond conventional imaging, the method was also applied to fat-water separation. Based on the Dixon technique, fat and water images were generated with high resolution. Due to the motion robustness of the underlying radial trajectory, measurements could be performed during free-breathing, without notable degradation of image quality. Abdominal images showed neither regional fat-water flipping nor residual banding artifacts. A drawback of standard Dixon-based fat-water separation is the fact that several separate images with different echo times have to be acquired, therefore prolonging the respective scan time. This can be overcome by using a multiecho sequence. It was demonstrated that the combination of such multiecho sequence and Dixon DYPR-SSFP allows for robust, banding-free fat-water separation in clinically acceptable scan times. DYPR-SSFP guarantees removal of banding artifacts even for strong magnetic field inhomogeneities. However, signal voids may remain due to intravoxel dephasing. This problem primarily arises when imaging metallic implants or when moving to ultra-high field strengths. To address this issue, the combination of DYPR-SSFP with the so-called z-shim technique was investigated, allowing the removal of these artifacts at the expense of an increased measurement time. Due to the applied dynamic phase-increment, radial projections which are acquired with DYPR-SSFP exhibit slightly different signal levels and phases. This results in incoherent artifacts, that can be effectively reduced by increasing the number of acquired projections. Therefore, DYPR-SSFP should be preferably applied when many projections are intrinsically necessary. It has been demonstrated that, besides high resolution imaging, the choice of a 3D radial trajectory is a promising combination. The proposed 3D DYPR-SSFP technique allowed isotropic banding-free bSSFP imaging without any expense of additional scan time compared to a conventional bSSFP acquisition. Residual artifacts caused by the dynamic phase-cycle could be effectively mitigated by applying a denoising algorithm. Volunteer measurements showed that 3D DYPR-SSFP is a promising candidate for imaging of the cranial nerves and the musculoskeletal system. While DYPR-SSFP and all presented resulting methods constitute an efficient solution for banding artifacts in bSSFP imaging, the proposed RAdial T-One sensitive and insensitive Steady-State Imaging (RA-TOSSI) method addresses the problem of the mixed contrast in bSSFP imaging. The possibility to generate T2-contrast with bSSFP has been shown before. The previous approach is based on the fact that T1-relaxation during the transient phase of a bSSFP acquisition can be suppressed by inserting unequally spaced inversion pulses. Thus, the resulting image shows a clinically relevant T2-contrast. The method which was presented as part of this dissertation relies on the same principle. However, instead of the originally proposed Cartesian trajectory, a radial trajectory in combination with a KWIC-filter reconstruction was applied. This allows the generation of several T2-weighted images as well as T2/T1-weighted images from a single RA-TOSSI dataset, while image quality remains comparable or even improves compared with existing methods. It could further be shown that varying the number of inversion pulses allows the generation of additional contrasts, where different tissue types are attenuated in the image. In the case of brain imaging for instance, these tissues comprise fat, gray matter, white matter, and CSF and offer high clinical relevance similar to T2-weighted images. Measurements of a patient with a brain tumor demonstrate the possible impact of the proposed method. In conclusion, the techniques developed as part of this dissertation present a valuable contribution to the solution of various problems which are associated with bSSFP imaging. Images acquired with DYPR-SSFP can be reconstructed directly at the scanner using existing, commercial reconstruction software. The software for the reconstruction of RA-TOSSI data was implemented for Siemens scanners. Therefore, both methods can be directly employed for clinical studies which remain as future work. KW - Kernspintomografie KW - Radiale Bildgebung KW - Steady-State-Sequenzen KW - balanced SSFP KW - Nicht-kartesische Bildgebung KW - Radial Imaging KW - Steady-State Sequences KW - balanced SSFP KW - Non-Cartesian Imaging KW - Magnetische Kernresonanz KW - Biophysik Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-115381 ER - TY - THES A1 - Choli, Morwan T1 - Hybridmethoden zur Reduzierung der spezifischen Absorptionsrate für neuroradiologische MRT-Untersuchungen an Hochfeldsystemen T1 - Hybrid methods for reducing specific Absorption rate in neuroradiologic MRI-examinations at high-field MR-systems N2 - Die klinische Magnetresonanztomografie (MRT) operiert meist bei einer Magnetfeldstärke von 1,5 Tesla (T). Es halten jedoch immer mehr 3T MRT-Systeme Einzug im klinischen Alltag und seit kurzem auch 7T Ganzkörper-MRT-Systeme in die Grundlagenforschung. Höhere Magnetfeldstärken führen grundsätzlich zum einem verbesserten Signal-zu-Rausch- Verhältnis, welches sich gewinnbringend in eine erhöhte Ortsauflösung oder schnellere Bildaufnahme äußert. Ein Nachteil ist aber die dabei im Patienten deponierte Hochfrequenz-Energie (HF-Energie), welche quadratisch mit ansteigender Feldstärke zusammenhängt. Charakterisiert wird diese durch die spezifische Absorptionsrate (SAR) und ist durch vorgegebene gesetzliche Grenzwerte beschränkt. Moderne, SAR-intensive MRT-Techniken (z.B. Multispinecho-Verfahren) sind bereits bei 1,5T nahe den zulässigen SAR-Grenzwerten und somit nicht unverändert auf Hochfeld-Systeme übertragbar. In dieser Arbeit soll das Potential modularer Hybrid-MRT-Techniken genutzt werden, um das SAR bei besonders SAR-intensiven MRT-Verfahren ohne signifikante Einbußen in der Bildqualität erheblich zu verringern. Die Hybrid-Techniken sollen in Verbindung mit zusätzlichen Methoden der SAR-Reduzierung den breiteren Einsatz SAR-intensiver MRT-Techniken an hohen Magnetfeldern ermöglichen. Ziel dieser Arbeit ist es, routinefähige und SAR-reduzierte MRT-Standard-Protokolle für neuroanatomische Humanuntersuchungen mit räumlicher Höchstauflösung bei Magnetfeldern von 3T und 7T zu etablieren. N2 - Spin echo based MRI sequences builds one of the main priorities in the medical/-morphological MRI imaging. Especially T2-weighted spin-echo sequences and the multi-spin-echo RARE imaging sequence represents one of the basic methods, which allows to minimize measurement times down to minutes or seconds. Modern MRI systems usually have a magnetic field strength beyond 1.5T. In order to acquire high resolution images with acceptable acquisition times, high RF power is needed. In many cases the intrinsic safety limits of the radiated power are already exeeded at field strengths of 3T. This leads often to restrictions for the full performance. In research systems of 7T and more T2-weighted images are only feasible with significant amount of time. Especially in RARE imaging methods, which require a large number of refocusing pulses, this is a significant restriction factor. Therefore, in recent years there were further development of hybrid sequences which combines different acquisition methods together into one, to exploit their fully advantages of the basic methods and to minimize ... KW - Kernspintomografie KW - spezifische Absorptionsrate KW - Magnetresonanztomografie KW - Hybridbildgebung KW - Absorption Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-100023 ER - TY - THES A1 - Neumann, Daniel T1 - Advances in Fast MRI Experiments T1 - Neue Methoden in der MR-Bildgebung N2 - Magnetic Resonance Imaging (MRI) is a non-invasive medical imaging technique, that is rou- tinely used in clinical practice for detection and diagnosis of a wide range of different diseases. In MRI, no ionizing radiation is used, making even repeated application unproblematic. This is an important advantage over other common imaging methods such as X-rays and Computer To- mography. One major drawback of MRI, however, are long acquisition times and associated high costs of experiments. Since the introduction of MRI, several important technical developments have been made to successfully reduce acquisition times. In this work, novel approaches were developed to increase the efficiency of MRI acquisitions. In Chapter 4, an improved radial turbo spin-echo (TSE) combined acquisition and reconstruction strategy was introduced. Cartesian turbo spin-echo sequences [3] are widely used especially for the detection and diagnosis of neurological pathologies, as they provide high SNR images with both clinically important proton density and T2 contrasts. TSE acquisitions combined with radial sampling are very efficient, since it is possible to obtain a number of ETL images with different contrasts from a single radial TSE measurement [56–58]. Conventionally, images with a particular contrast are obtained from both radial and Cartesian TSE acquisitions by combining data from different echo times into a single image. In the radial case, this can be achieved by employing k-space weighted image contrast (KWIC) reconstruction. In KWIC, the center region of k-space is filled exclusively with data belonging to the desired contrast while outer regions also are assembled with data acquired at other echo times. However, this data sharing leads to mixed contrast contributions to both Cartesian and radial TSE images. This is true especially for proton density weighted images and therefore may reduce their diagnostic value. In the proposed method, an adapted golden angle reordering scheme is introduced for radial TSE acquisitions, that allows a free choice of the echo train length and provides high flexibility in image reconstruction. Unwanted contrast contaminations are greatly reduced by employing a narrow-band KWIC filter, that restricts data sharing to a small temporal window around the de- sired echo time. This corresponds to using fewer data than required for fully sampled images and consequently leads to images exhibiting aliasing artifacts. In a second step, aliasing-free images are obtained using parallel imaging. In the neurological examples presented, the CG-SENSE algorithm [42] was chosen due to its stable convergence properties and its ability to reconstruct arbitrarily sampled data. In simulations as well as in different in vivo neurological applications, no unwanted contrast contributions could be observed in radial TSE images reconstructed with the proposed method. Since this novel approach is easy to implement on today’s scanners and requires low computational power, it might be valuable for the clinical breakthrough of radial TSE acquisitions. In Chapter 5, an auto-calibrating method was introduced to correct for stimulated echo contribu- tions to T2 estimates from a mono-exponential fit of multi spin-echo (MSE) data. Quantification of T2 is a useful tool in clinical routine for the detection and diagnosis of diseases as well as for tis- sue characterization. Due to technical imperfections, refocusing flip angles in a MSE acquisition deviate from the ideal value of 180○. This gives rise to significant stimulated echo contributions to the overall signal evolution. Therefore, T2 estimates obtained from MSE acquisitions typically are notably higher than the reference. To obtain accurate T2 estimates from MSE acquisitions, MSE signal amplitudes can be predicted using the extended phase graph (EPG, [23, 24]) algo- rithm. Subsequently, a correction factor can be obtained from the simulated EPG T2 value and applied to the MSE T2 estimates. However, EPG calculations require knowledge about refocus- ing pulse amplitudes, T2 and T1 values and the temporal spacing of subsequent echoes. While the echo spacing is known and, as shown in simulations, an approximate T1 value can be assumed for high ratios of T1/T2 without compromising accuracy of the results, the remaining two parameters are estimated from the data themselves. An estimate for the refocusing flip angle can be obtained from the signal intensity ratio of the second to the first echo using EPG. A conventional mono- exponential fit of the MSE data yields a first estimate for T2. The T2 correction is then obtained iteratively by updating the T2 value used for EPG calculations in each step. For all examples pre- sented, two iterations proved to be sufficient for convergence. In the proposed method, a mean flip angle is extracted across the slice. As shown in simulations, this assumption leads to greatly reduced deviations even for more inhomogeneous slice profiles. The accuracy of corrected T2 values was shown in experiments using a phantom consisting of bottles filled with liquids with a wide range of different T2 values. While T2 MSE estimates were shown to deviate significantly from the spin-echo reference values, this is not the case for corrected T2 values. Furthermore, applicability was demonstrated for in vivo neurological experiments. In Chapter 6, a new auto-calibrating parallel imaging method called iterative GROG was pre- sented for the reconstruction of non-Cartesian data. A wide range of different non-Cartesian schemes have been proposed for data acquisition in MRI, that present various advantages over conventional Cartesian sampling such as faster acquisitions, improved dynamic imaging and in- trinsic motion correction. However, one drawback of non-Cartesian data is the more complicated reconstruction, which is ever more problematic for non-Cartesian parallel imaging techniques. Iterative GROG uses Calibrationless Parallel Imaging by Structured Low-Rank Matrix Completion (CPI) for data reconstruction. Since CPI requires points on a Cartesian grid, it cannot be used to directly reconstruct non-Cartesian data. Instead, Grappa Operator Gridding (GROG) is employed in a first step to move the non-Cartesian points to the nearest Cartesian grid locations. However, GROG requires a fully sampled center region of k-space for calibration. Combining both methods in an iterative scheme, accurate GROG weights can be obtained even from highly undersampled non-Cartesian data. Subsequently, CPI can be used to reconstruct either full k- space or a calibration area of arbitrary size, which can then be employed for data reconstruction with conventional parallel imaging methods. In Chapter 7, a new 2D sampling scheme was introduced consisting of multiple oscillating effi- cient trajectories (MOET), that is optimized for Compressed Sensing (CS) reconstructions. For successful CS reconstruction of a particular data set, some requirements have to be met. First, ev- ery data sample has to carry information about the whole object, which is automatically fulfilled for the Fourier sampling employed in MRI. Additionally, the image to be reconstructed has to be sparse in an arbitrary domain, which is true for a number of different applications. Last, data sam- pling has to be performed in an incoherent fashion. For 2D imaging, this important requirement of CS is difficult to achieve with conventional Cartesian and non-Cartesian sampling schemes. Ra- dial sampling is often used for CS reconstructions of dynamic data despite the streaking present in undersampled images. To obtain incoherent aliasing artifacts in undersampled images while at the same time preserving the advantages of radial sampling for dynamic imaging, MOET com- bines radial spokes with oscillating gradients of varying amplitude and alternating orientation orthogonal to the readout direction. The advantage of MOET over radial sampling in CS re- constructions was demonstrated in simulations and in in vivo cardiac imaging. MOET provides superior results especially when used in CS reconstructions with a sparsity constraint directly in image space. Here, accurate results could be obtained even from few MOET projections, while the coherent streaking artifacts present in the case of radial sampling prevent image recovery even for smaller acceleration factors. For CS reconstructions of dynamic data with sparsity constraint in xf-space, the advantage of MOET is smaller since the temporal reordering is responsible for an important part of incoherency. However, as was shown in simulations of a moving phantom and in the reconstruction of ungated cardiac data, the additional spatial incoherency provided by MOET still leads to improved results with higher accuracy and may allow reconstructions with higher acceleration factors. N2 - Die Magnetresonanztomographie (MRT) ist ein wichtiges nicht-invasives medizinisches Bildge- bungsverfahren, das im klinischen Alltag zur Entdeckung und Diagnose einer Vielzahl von Krank- heiten verwendet wird. Im Gegensatz zu anderen Methoden wie Röntgen und Computertomo- graphie kommt die MRT ohne den Einsatz ionisierender Strahlung aus, was selbst häufige An- wendungen ohne gesundheitliche Risiken erlaubt. Einer der größten Nachteile der MRT sind lange Messzeiten, die in Kombination mit der teuren Technik hohe Untersuchungskosten bedin- gen. Obwohl in der Vergangenheit durch die Entwicklung von sowohl verbesserter Hardware als auch neuen Rekonstruktionsverfahren bereits bedeutende Fortschritte in Bezug auf die Akquisi- tionsdauer erzielt werden konnten, ist eine weitere Beschleunigung nach wie vor ein wichtiges Forschungsgebiet im Bereich der MRT. Ziel dieser Arbeit war daher die Entwicklung neuer An- sätze zur Steigerung der Effizienz von MRT Experimenten. In Kapitel 4 wurde eine kombinierte Akquisitions- und Rekonstruktionsstrategie für radiale Turbo Spin-Echo (TSE) Experimente vorgestellt. Im klinischen Alltag sind kartesische TSE Sequenzen zur Untersuchung diverser Krankheitsbilder weit verbreitet, da sie ein hohes SNR aufweisen und die Aufnahme der klinisch wichtigen Bilder mit Protonendichte- und T2-Kontrast erlauben. Im Gegensatz zu kartesischem Abtasten, wo aus einem Datensatz lediglich ein Bild mit bes- timmtem Kontrast erzeugt wird, sind radiale TSE Akquisitionen hocheffizient, da hier aus einem Datensatz mehrere Bilder mit verschiedenem Kontrast gewonnen werden können. In beiden Fällen wird in konventionellen Rekonstruktionsmethoden jedes Bild eines definierten Kontrasts durch das Zusammensetzen eines vollständig abgetasteten k-Raums mit Daten von verschiedenen Echozeiten erzeugt. Im radialen Fall geschieht dies durch die sogenannte "k-space weighted im age contrast" (KWIC) Rekonstruktion. Hierbei wird das Zentrum des k-Raums ausschließlich mit zum gewünschten Kontrast gehörigen Daten gefüllt, während die äußeren Bereiche des k-Raums auch Daten von anderen Echozeiten enthalten. Obwohl der Kontrast von MRT Bildern haupt- sächlich von den Daten im k-Raum Zentrum dominiert wird, führt die Kombination von Daten verschiedener Echozeiten in sowohl radialen als auch kartesischen TSE Bildern zu einem uner- wünschten Mischkontrast. Dieser Effekt wird vor allem in protonendichtegewichteten Bildern sichtbar und kann somit deren diagnostischen Wert deutlich verringern. Ein unerwünschter Mischkontrast kann verhindert werden, indem die Bandbreite des KWIC- Filters auf ein kleines zeitliches Fenster um die angestrebte Echozeit herum eingeschränkt wird. Um eine freie Wahl der Echozuglänge und hohe Flexibilität in der Bildrekonstruktion zu er- möglichen, wurde für die radiale TSE Akquisition ein angepasstes Abtastschema unter Verwen- dung des goldenen Winkels vorgestellt. Da bei einem KWIC-Filter mit reduzierter Bandbre- ite für jedes Bild weniger Daten zur Verfügung stehen als für einen voll abgetasteten k-Raum benötigt, weisen rekonstruierte Bildern Einfaltungsartefakte auf. Diese werden in einem zweiten Schritt durch die Anwendung paralleler Bildgebung beseitigt. In den gezeigten Beispielen wurde dazu der CG-SENSE Algorithmus verwendet, da er stabile Konvergenz aufweist und für die Rekonstruktion von Daten mit irregulären Abtastschema angewandt werden kann. Anschließend werden bestehende Korrelationen der Bilderserie zur Reduktion verbleibender Artefakte und zu einer Verbesserung des SNR ausgenutzt. Wie mittels Simulationen gezeigt und für neurologische Daten bestätigt, weisen radiale TSE Bilder, die mit dieser Methode rekonstruiert wurden, keinen sichtbaren Mischkontrast mehr auf. Die erreichte Bildqualität ist hierbar vergleichbar mit kon- ventionellen Rekonstruktionsmethoden. Da die vorgestellte Rekonstruktion einfach auf heutigen Scannern implementiert werden kann und lediglich niedrige Rechenkapazitäten benötigt, könnte sie einen wichtigen Beitrag für den klinischen Durchbruch radialer TSE Akquisitionen darstellen. In Kapitel 5 wurde eine selbstkalibrierende Methode zur Korrektur von aus Multi Spin-Echo (MSE) Bildern gewonnenen T2 Karten vorgestellt. In der klinischen Anwendung spielt die Quan- tifizierung von T2 unter anderem bei der Diagnose von Krankheiten sowie bei der Klassifizierung von Gewebe eine wichtige Rolle. Eine MSE Sequenz verwendet mehrere RF-Pulse, um ein einzelnes Spin-Echo wiederholt zu refokussieren. Idealerweise betragen die Flipwinkel der Re- fokussierungspulse hierbei exakt 180○, um einen exponentiellen Signalabfall zu erhalten. Auf- grund technischer Ungenauigkeiten weichen die Werte der Flipwinkel von Refokussierungspulsen jedoch grundsätzlich von 180○ ab. Niedrigere Flipwinkel führen zu stimulierten Echos, die wesentlich zu den einzelnen Echoamplituden beitragen und den Signalabfall entlang des Echozugs deutlich verlängern können. Somit weisen auch T2 Werte, die aus solchen Bilderserien berech- net werden, eine teilweise deutliche Erhöhung auf. Um exakte Werte zu erhalten, kann der MSE Signalverlauf mittels des "extended phase graph" (EPG) Algorithmus abgeschätzt und so ein Kor- rekturfaktor ermittelt werden. Hierzu müssen die Flipwinkel der Refokussierungspulse, T1 und T2 Werte sowie der zeitliche Abstand der Echos (ESP) bekannt sein. Wie in Simulationen gezeigt wurde, kann T1 für hohe Werte des Quotienten T1/T2 abgeschätzt werden, ohne an Genauigkeit der T2 Ergebnisse einzubüßen. Abschätzungen der verbleibenden benötigten Parameter können direkt aus den Daten selbst gewonnen werden. Während der Flipwinkel aus den Intensitäten der ersten beiden Echos berechnet wird, liefert ein mono-exponentieller Fit der MSE Bilderserie eine erste Näherung für T2. Die Korrektur für die T2 Werte kann anschließend aus den EPG Sig- nalverläufen berechnet werden. Durch Aktualisierung von T2 und erneuter Ausführung des EPG-Algorithmus wird die Genauigkeit der Korrektur iterativ erhöht, wobei schon eine sehr geringe Zahl von Iterationen zu Konvergenz führt. Wie in Simulationen und in Phantomexperimenten für verschiedenste T2-Werte gezeigt, weisen korrigierte T2 Werte eine hohe Genauigkeit auf. Dies gilt auch für niedrigere nominelle Flipwinkel als 180○ und ist somit von speziellem Interesse bei höheren Feldstärken B0, wo Grenzwerte der spezifischen Absorptionsrate die Einstrahlung einer Vielzahl von RF-Pulsen hoher Amplitude verbietet. In Kapitel 6 wurde iteratives GROG, eine neue selbstkalibrierende iterative parallele Bildge- bungsmethode für die Rekonstruktion von nichtkartesischen Daten vorgestellt. Es sind eine Vielzahl nichtkartesischer Trajektorien für MRT Messungen bekannt, die zahlreiche Vorteile gegenüber kartesischer Bildgebung bieten. Dazu gehören unter anderem eine schnellere Akquisi- tion, verbesserte dynamische Bildgebung sowie die Möglichkeit zur intrinischen Bewegungskor- rektur. Ein Nachteil nichtkartesischer Daten jedoch ist eine aufwendigere Rekonstruktion, sowohl bei voll abgetasteten Datensätzen als insbesondere auch in der parallelen Bildgebung. Itera- tives GROG verwendet Calibrationless Parallel Imaging by Structured Low-Rank Matrix Com- pletion (CPI) zur Rekonstruktion fehlender Daten. Diese Methode benötigt Daten auf karte- sischen Gitterpunkten und kann nicht direkt für nichtkartesische Experimente angewandt wer- den. Stattdessen werden die nichtkartesischen Daten zunächst mittels Grappa Operator Gridding (GROG) in einem ersten Schritt auf ein kartesisches Gitter verschoben. GROG basiert auf paral- leler Bildgebung und benötigt einen voll abgetasteten Teil des k-Raums zur Kalibrierung. Erste Kalibrationsdaten können gewonnen werden, indem die nichtkartesischen Punkte ohne Änderung auf die nächsten kartesischen Gitterpunkte verschoben werden und eine CPI-Rekonstruktion eines zentralen k-Raum Bereichs durchgeführt wird. Anschließend wird GROG angewandt um exakte Werte der kartesischen Gitterpunkte zu erhalten und der Prozess wird iteriert. Nach Kon- vergenz können entweder Kalibrationsdaten gewünschter Größe für eine konventionelle parallele Bildgebungsmethode erzeugt oder artefaktfreie Bilder mit CPI rekonstruiert werden. In Kapitel 7 wurde ein neues Abtastungsschema für die 2D Bildgebung vorgestellt, das aus Multiplen Oszillierenden Effizienten Trajektorien (MOET) besteht und optimierte Compressed Sensing (CS) Rekonstruktionen ermöglicht. Für eine erfolgreiche Anwendung von Compressed Sensing müssen einige Voraussetzungen erfüllt sein. Erstens muss jeder Datenpunkt Informa- tionen über das ganze Objekt enthalten, was bei der MRT aufgrund der Datenakquisition im Fourier-Raum automatisch erfüllt ist. Weiterhin muss das gemessene Objekt in einer beliebigen Basis sparse sein. Dies ist für viele verschiedene Anwendungen in der MRT der Fall. Drittens muss für CS Rekonstruktionen die Datenakquisition im k-Raum einem inkohärenten Muster fol- gen. Diese wichtige Voraussetzung ist in der zweidimensionalen Bildgebung mit konventionellen kartesischen und nicht-kartesischen Abtastschemata nur schwer zu erreichen. Deshalb wird für CS Rekonstruktionen häufig eine radiale Trajektorie eingesetzt, trotz der kohärenten streaking- Artefakte in unterabgetasteten Bildern. MOET verwendet daher eine Kombination von radialen Projektionen zusammen mit oszillierenden Gradienten auf der zur Ausleserichtung orthogonalen Achse. So erhält man inkohärente Aliasing-Artefakte und bewahrt gleichzeitig die Vorteile der radialen Bildgebung für die dynamische MRT. Die Überlegenheit von MOET gegenüber radi- aler Bildgebung für CS Rekonstruktionen konnte in Simulationen sowie in der Herzbildgebung aufgezeigt werden. Dies gilt insbesondere für CS Rekonstruktionen direkt im Bildraum, wo MOET gute Resultate liefert während die kohärenten Artefakte bei radialer Bildgebung eine genaue Bildwiederherstellung verhindert. Bei Rekonstruktionen dynamischer Daten, wo Sparsität im xf-Raum ausgenutzt wird, ist der Vorteil von MOET weniger ausgeprägt, da hier bere- its die zeitliche Anordnung der Projektionen einen wesentlichen Beitrag zur Charakteristik der Aliasingartefakte liefert. Wie in Simulationen und für die in vivo Herzbildgebung gezeigt werden konnte, erlaubt die zusätzliche räumliche Inkohärenz von MOET jedoch auch in diesem Fall eine höhere Genauigkeit sowie Rekonstruktionen von Daten höherer Beschleunigung. KW - Kernspintomografie KW - Parallele Bildgebung KW - nichtkarthesische Bildgebung KW - Turbo Spin-Echos KW - Compressed Sensing KW - Parallel Imaging KW - non-Cartesian Imaging KW - Compressed Sensing KW - MRI KW - MRT KW - NMR-Tomographie Y1 - 2014 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-108165 ER - TY - THES A1 - Lother, Steffen Reiner T1 - Entwicklung eines 3D MR-Tomographen zur Erdfeld- und multimodalen MR-MPI-Bildgebung T1 - Development of a 3D MRI-System for Earth Field MRI and the Combination MRI-MPI N2 - Das Ziel dieser Arbeit war die Entwicklung und die Anfertigung eines 3D Erdfeld-NMR Tomographen, um damit die benötigte Technik der MR eines MR-MPI-Tomographen am Lehrstuhl zu etablieren. Daraufhin wurden alle nötigen Komponenten für ein komplettes 3D Erdfeld-NMR-System entwickelt, gebaut und getestet. Mit diesem Wissen wurde in enger Zusammenarbeit mit der MPI-Arbeitsgruppe am Lehrstuhl ein multimodaler MR-MPI-Tomograph angefertigt und die prinzipielle Machbarkeit der technischen Kombination dieser zwei Modalitäten (MRT/MPI) in einer einzigen Apparatur gezeigt. Auf diesem Entwicklungsweg sind zusätzlich innovative Systemkomponenten entstanden, wie der Bau eines neuen Präpolarisationssystems, mit dem das Präpolarisationsfeld kontrolliert und optimiert abgeschaltet werden kann. Des Weiteren wurde ein neuartiges 3D Gradientensystem entwickelt, das parallel und senkrecht zum Erdmagnetfeld ausgerichtet werden kann, ohne die Bildgebungseigenschaften zu verlieren. Hierfür wurde ein 3D Standard-Gradientensystem mit nur einer weiteren Spule, auf insgesamt vier Gradientenspulen erweitert. Diese wurden entworfen, gefertigt und anhand von Magnetfeldkarten ausgemessen. Anschließend konnten diese Ergebnisse mit der hier präsentierten Theorie und den Simulationsergebnissen übereinstimmend verglichen werden. MPI (Magnetic Particle Imaging) ist eine neue Bildgebungstechnik mit der nur Kontrastmittel detektiert werden können. Das hat den Vorteil der direkten und eindeutigen Detektion von Kontrastmitteln, jedoch fehlt die Hintergrundinformation der Probe. Wissenschaftliche Arbeiten prognostizieren großes Potential, die Hintergrundinformationen der MRT mit den hochauflösenden Kontrastmittelinformationen mittels MPI zu kombinieren. Jedoch war es bis jetzt nicht möglich, diese beiden Techniken in einer einzigen Apparatur zu etablieren. Mit diesem Prototyp konnte erstmalig eine MR-MPI-Messung ohne Probentransfer durchgeführt und die empfindliche Lokalisation von Kontrastmittel mit der Überlagerung der notwendigen Hintergrundinformation der Probe gezeigt werden. Dies ist ein Meilenstein in der Entwicklung der Kombination von MRT und MPI und bringt die Vision eines zukünftigen, klinischen, multimodalen MR-MPI-Tomographen ein großes Stück näher. N2 - Developement of an 3D MRI System for Earth Field MRI and the Kombination with MPI KW - NMR-Spektroskopie KW - Erdmagnetismus KW - Kernspintomografie KW - 3D Erdfeld-NMR Tomograph KW - MR-MPI-Tomograph KW - Präpolarisationssystems KW - Gradientensystem KW - Magnetresonanztomographie KW - Systembau KW - Erdfeld KW - Magnetpartikelbildgebung Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-99181 ER - TY - THES A1 - Joseph, Arun Antony T1 - Real-time MRI of Moving Spins Using Undersampled Radial FLASH T1 - Echtzeit MRI von bewegten Spins mithilfe der unterabgetasteten radialen FLASH sequenz N2 - Nuclear spins in motion is an intrinsic component of any dynamic process when studied using magnetic resonance imaging (MRI). Moving spins define many functional characteristics of the human body such as diffusion, perfusion and blood flow. Quantitative MRI of moving spins can provide valuable information about the human physiology or of a technical system. In particular, phase-contrast MRI, which is based on two images with and without a flow-encoding gradient, has emerged as an important diagnostic tool in medicine to quantify human blood flow. Unfortunately, however, its clinical usage is hampered by long acquisition times which only provide mean data averaged across multiple cardiac cycles and therefore preclude Monitoring the immediate physiological responses to stress or exercise. These limitations are expected to be overcome by real-time imaging which constitutes a primary aim of this thesis. Short image acquisition times, as the core for real-time phase-contrast MRI, can be mainly realized through undersampling of the acquired data. Therefore the development focused on related technical aspects such as pulse sequence design, k-space encoding schemes and image reconstruction. A radial encoding scheme was experimentally found to be robust to motion and less sensitive to undersampling than Cartesian encoding. Radial encoding was combined with a FLASH acquisition technique for building an efficient real-time phase-contrast MRI sequence. The sequence was further optimized through overlapping of gradients to achieve the shortest possible echo time. Regularized nonlinear inverse reconstruction (NLINV), a technique which jointly estimates the image content and its corresponding coil sensitivities, was used for image reconstruction. NLINV was adapted specifically for phase-contrast MRI to produce both Magnitude images and phase-contrast maps. Real-time phase-contrast MRI therefore combined two highly undersampled (up to a factor of 30) radial gradient-echo acquisitions with and without a flow-encoding gradient with modified NLINV reconstructions. The developed method achieved real-time phase-contrast MRI at both high spatial (1.3 mm) and temporal resolution (40 ms). Applications to healthy human subjects as well as preliminary studies of patients demonstrated real-time phase-contrast MRI to offer improved patient compliance (e.g., free breathing) and immediate access to physiological variations of flow parameters (e.g., response to enhanced intrathoracic pressure). In most cases, quantitative blood flow was measured in the ascending aorta as an important blood vessel of the cardiovascular circulation system commonly studied in the clinic. The performance of real-time phase-contrast MRI was validated in comparison to standard Cine phase-contrast MRI using studies of flow phantoms as well as under in vivo conditions. The evaluations confirmed good agreement for comparable results. As a further extension to real-time phase-contrast MRI, this thesis implemented and explored a dual-echo phase-contrast MRI method which employs two sequential gradient echoes with and without flow encoding. The introduction of a flow-encoding gradient in between the two echoes aids in the further reduction of acquisition time. Although this technique was efficient under in vitro conditions, in vivo studies showed the influence of additional motion-induced Phase contributions. Due to these additional temporal phase information, the approach showed Little promise for quantitative flow MRI. As a further method three-dimensional real-time phase-contrast MRI was developed in this thesis to visualize and quantify multi-directional flow at about twice the measuring time of the standard real-time MRI method, i.e. at about 100 ms temporal resolution. This was achieved through velocity mapping along all three physical gradient directions. Although the method is still too slow to adequately cover cardiovascular blood flow, the preliminary results were found to be promising for future applications in tissues and organ systems outside the heart. Finally, future developments are expected to benefit from the adaptation of model-based reconstruction techniques to real-time phase-contrast MRI. N2 - Die Bewegung der Kernspins ist eine wesentliche Eigenschaft von dynamischen Vorgängen, die mit Hilfe der Magnetresonanztomographie (MRT) untersucht werden. Bewegte oder fließende Spins charakterisieren viele Funktionen des menschlichen Körpers, wie z.B. die Gewebeperfusion und den Blutfluss in den Gefäßen. Die quantitative MRT von bewegten Spins kann daher wertvolle Informationen über die menschliche Physiologie oder auch über ein technisches System geben. Insbesondere die Phasenkontrast-MRT, die auf der Aufnahme von zwei Bildern mit und ohne flusskodierenden Gradienten basiert, hat sich als ein wichtiges diagnostisches Werkzeug in der Medizin entwickelt, um den Blutfluss funktionell zu quantifizieren. Die klinische Nutzung ist jedoch durch die langen Messzeiten eingeschränkt, da die Daten über mehrere Herzzyklen gemittelt werden müssen und damit die Untersuchung unmittelbarer physiologischer Reaktionen auf Stress und/oder Muskelbelastung ausgeschlossen ist. Ein primäres Ziel dieser Arbeit war es, diese Einschränkungen durch die Entwicklung einer MRT-Flussmessung in Echtzeit zu überwinden. Entscheidende Grundlage jeder Echtzeit-MRT sind kurze Aufnahmezeiten, die vor allem durch eine Reduktion der aufgenommenen Daten (Unterabtastung) realisiert werden. Daher konzentrierte sich die hier vorgestellte Entwicklung auf die damit verbundenen technischen Aspekte wie die MRT-Sequenz zur Datenaufnahme, das räumliche Kodierungsschema, und die Bildrekonstruktion. Experimentell erwies sich ein radiales Kodierungsschema als robust gegenüber Bewegungen und relativ unempfindlich gegenüber milder Unterabtastung. Dieses Kodierungsschema wurde mit der FLASH Aufnahmetechnik für eine effiziente Phasenkontrast-Sequenz in Echtzeit kombiniert. Zusätzlich wurde die Sequenz durch Überlappung von Gradienten hinsichtlich einer kurzen Echozeit optimiert. Für die Bildrekonstruktion wurde die regularisierte nichtlineare inverse Rekonstruktion (NLINV) verwendet, bei der die Bildinformation und die entsprechenden pulensensitivitäten gleichzeitig geschätzt werden. NLINV wurde speziell für die Phasenkontrast-MRT angepasst, um sowohl Betragsbilder als auch robuste Phasenkontrast-Karten mit hoher raumzeitlicher Genauigkeit zu berechnen. Das erarbeitete Verfahren der Phasenkontrast-MRT in Echtzeit kombiniert daher zwei stark unterabgetastete (bis zu einem Faktor von 30) und unterschiedlich flusskodierte, radiale Gradientenecho-Aufnahmen mit einer modifizierten NLINV Rekonstruktion. Mit dieser Methode wurde sowohl eine gute räumliche Auflösung (1.3 mm), als auch eine hohe zeitliche Auflösung (40 ms) erreicht. Bei Anwendungen an gesunden Probanden sowie vorläufigen Untersuchungen von Patienten konnte nachgewiesen werden, dass die Phasenkontrast-MRT in Echtzeit einen verbesserten Komfort für die Patienten (z.B. freie Atmung) und unmittelbaren Zugang zu physiologischen Veränderungen der Flussparameter bietet (z.B. Reaktion auf erhöhten Druck im Brustraum). In den meisten Fällen wurden quantitative Blutflussmessungen in der aufsteigenden Aorta, einem klinisch wichtigen Gefäß des Herz-Kreislauf-Systems, vorgenommen. Die Messungen mit der Phasenkontrast-MRT in Echtzeit wurden mit der EKG-getriggerten Cine Phasenkontrast-MRT (klinischer Standard) an einem Flussphantom und unter in vivo Bedingungen verglichen. Die Ergebnisse zeigten unter vergleichbaren Bedingungen gute Übereinstimmung. Im Rahmen dieser Arbeit wurde zusätzlich eine Doppelecho-Variante der Phasenkontrast-MRT in Echtzeit implementiert. Das Einfügen eines flusskodierenden Gradienten zwischen den beiden Echos führte zu einer weiteren Reduzierung der Messzeit. Obwohl sich diese Technik unter in vitro Bedingungen als tauglich erwies, zeigten sich bei in vivo Studien störende Einflüsse durch bewegungsinduzierte Phasenbeiträge, die wenig Erfolg für quantitative Flussmessungen versprechen. Als weitere Methode wurde in dieser Arbeit eine dreifach kodierte Sequenz zur Phasenkontrast-MRT entwickelt, um multidirektionalen Fluss zu untersuchen. Die Geschwindigkeitskodierung entlang aller drei physikalischen Gradientenrichtungen führte zu einer verlängerten Messzeit (zeitliche Auflösung � 100 ms) gegenüber der Echtzeit-Flussmessung in nur einer Richtung. Obwohl das Verfahren noch zu langsam ist, um den kardiovaskulären Blutfluss adäquat zu beschreiben, waren vorläufige Ergebnisse in Körperregionen außerhalb des Herzens für zukünftige klinische Anwendungen sehr vielversprechend. Es ist zu erwarten, dass entsprechende Weiterentwicklungen von modellbasierten ekonstruktionsverfahren profitieren werden. KW - Kernspintomografie KW - Real-time MRI Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-94000 ER - TY - THES A1 - Zeller, Mario T1 - Dichtegewichtete Magnetresonanz-Bildgebung mit Multi-Echo-Sequenzen T1 - Density Weighted Magnetic Resonance Imaging with Multi-Echo Sequences N2 - Das Signal-zu-Rausch-Verhältnis (SNR) stellt bei modernen Bildgebungstechniken in der Magnetresonanz-Tomographie heutzutage oftmals die entscheidende Limitation dar. Eine Verbesserung durch Modifikation der Hardware ist kostspielig und führt meistens zu einer Verstärkung anderer Probleme, wie zum Beispiel erhöhte Energiedeposition ins Gewebe. Im Gegensatz dazu ist Dichtegewichtung eine Methode, die eine SNR-Erhöhung durch Modifikation der Aufnahmetechnik ermöglicht. In der MR-Bildgebung erfolgt oftmals eine retrospektive Filterung des aufgenommenen Signalverlaufs, beispielsweise zur Artefaktreduktion. Damit einhergehend findet eine Veränderung der Modulationstransferfunktion (MTF) bzw. ihrer Fouriertransformierten, der räumlichen Antwortfunktion (SRF), statt. Optimales SNR wird nach dem Matched Filter-Theorem erzielt, wenn die nachträgliche Filterung dem aufgenommenen Signalverlauf proportional ist. Dies steht dem Ziel der Artefaktreduktion entgegen. Bei Dichtegewichtung steht durch nicht-kartesische Abtastung des k-Raums mit der k-Raum-Dichte ein zusätzlicher Freiheitsgrad zur Verfügung. Dieser ermöglicht es, im Falle eines konstanten Signalverlaufs eine gewünschte MTF ohne Filterung zu erreichen. Bei veränderlichem Signalverlauf kann ein SNR Matched Filter angewendet werden, dessen negative Einflüsse auf die MTF durch Dichtegewichtung kompensiert werden. Somit ermöglicht Dichtegewichtung eine vorgegebene MTF und gleichzeitig ein optimales SNR. In der vorliegenden Arbeit wurde Dichtegewichtung erstmals bei den schnellen Multi-Echo-Sequenzen Turbo-Spin-Echo und Echoplanar-Bildgebung (EPI) angewendet. Im Gegensatz zu bisherigen Implementierungen muss hier der Signalabfall durch T2- bzw. T2*-Relaxation berücksichtigt werden. Dies führt dazu, dass eine prospektiv berechnete dichtegewichtete Verteilung nur bei einer Relaxationszeit optimal ist. Bei Geweben mit abweichenden Relaxationszeiten können sich wie auch bei den kartesischen Varianten dieser Sequenzen Änderungen an SRF und SNR ergeben. Bei dichtegewichteter Turbo-Spin-Echo-Bildgebung des Gehirns konnte mit den gewählten Sequenzparametern ein SNR-Vorteil von 43 % gegenüber der kartesischen Variante erzielt werden. Die Akquisition wurde dabei auf die T2-Relaxationszeit von weißer Substanz optimiert. Da die meisten Gewebe im Gehirn eine ähnliche Relaxationszeit aufweisen, blieb der visuelle Gesamteindruck identisch zur kartesischen Bildgebung. Der SNR-Gewinn konnte in der dichtegewichteten Implementierung zur Messzeithalbierung genutzt werden. Dichtegewichtete EPI weist eine hohe Anfälligkeit für geometrische Verzerrungen, welche durch Inhomogenitäten des Hauptmagnetfeldes verursacht werden, auf. Die Verzerrungen konnten erfolgreich mit einer Conjugate Phase-Methode korrigiert werden. Dazu muss die räumliche Verteilung der Feldinhomogenitäten bekannt sein. Dazu ist zusätzlich zur eigentlichen EPI-Aufnahme die zeitaufwendige Aufnahme einer sogenannten Fieldmap erforderlich. Im Rahmen dieser Arbeit konnte eine Methode entwickelt werden, welche die zur Erlangung einer Fieldmap notwendige Aufnahmedauer auf wenige Sekunden reduziert. Bei dieser Art der Fieldmap-Aufnahme müssen jedoch durch Atmung hervorgerufene Effekte auf die Bildphase berücksichtigt werden. Die Fieldmap-Genauigkeit kann durch Aufnahme unter Atempause, Mittelung oder retrospektiver Phasenkorrektur erhöht werden. Für die gewählten EPI-Sequenzparameter wurde mit Dichtegewichtung gegenüber der kartesischen Variante ein SNR-Gewinn von 14 % erzielt. Anhand einer funktionellen MRT (fMRI)-Fingertapping-Studie konnte demonstriert werden, dass die SNR-Steigerung auch zu einer signifikant erhöhten Aktivierungsdetektion in Teilen der Hirnareale führt, die bei der Fingerbewegung involviert sind. Die Verwendung von zusätzlicher EPI-Phasenkorrektur und iterativer Optimierung der dichtegewichteten k-Raum-Abtastung führt zu weiteren Verbesserungen der dichtegewichteten Bildgebung mit Multi-Echo-Sequenzen. N2 - Magnetic resonance imaging (MRI) is often limited by the signal to noise ratio (SNR). In standard Cartesian acquisition methods, the SNR can be improved by applying a so-called matched filter to the acquired raw data, which correlates with the anticipated signal profile. Unfortunately, this filter changes the spatial response function (SRF), which characterizes the imaging properties of the imaging method, in an undesired way. For example, a matched filter often amplifies undesired image artifacts and is thus normally omitted. In contrast, filters which change the SRF are typically applied, e.g., for artifact reduction. These however do not provide an optimal SNR. Density weighting is a method which allows a desired SRF and an optimal SNR at the same time. This is achieved by introducing a new degree of freedom to the SRF; the density of the acquisition steps in k-space. In this work, density weighting was adapted to turbo spin echo (TSE) and echo planar imaging (EPI). In contrast to earlier implementations of density weighting, signal relaxation has to be taken into consideration with these multi-echo sequences. As a result, the desired SRF and SNR are only obtained for one prospectively determined relaxation time. For deviating relaxation times, changes in SRF and SNR may occur. In density weighted TSE brain imaging, an average SNR gain of 43 % over Cartesian imaging could be achieved for the chosen sequence parameters. The density weighted acquisition was optimized for the T2 relaxation time of white matter. Since the relaxation times of most other tissues in the brain did not significantly differ, the overall visual impression of density weighted and Cartesian images was identical. The achieved SNR gain could be used to halve the acquisition time of the density weighted implementation. Density weighted EPI is especially prone to geometric distortions caused by inhomogeneities of the main magnetic field. The distortions could be successfully corrected with a conjugate phase method. For these methods, a time-consuming acquisition of a so-called field map is typically required. A method could be developed which greatly reduces the field map acquisition time to a few seconds. It was found that phase changes caused by respiration influence the field map accuracy of this and similar methods. A significantly higher accuracy could be achieved by an acquisition under breath-hold or by retrospective phase correction or averaging. It was demonstrated in an fMRI group study that an average SNR gain of 14 % for density weighted EPI resulted in an increased detection power in the activated brain areas. First results involving additional EPI phase correction and iterative k-space sampling optimization demonstrate further improvements of density weighted imaging with multi-echo sequences. KW - Kernspintomografie KW - Störabstand KW - Dichtegewichtung KW - Echoplanar-Bildgebung KW - Turbo Spin Echo KW - B0-Korrektur KW - Density Weighting KW - Point Spread Function KW - Echo Planar Imaging KW - Turbo Spin Echo KW - B0 correction KW - Echo planar imaging KW - NMR-Tomographie Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-84142 ER - TY - THES A1 - Hölscher, Uvo Christoph T1 - Relaxations-Dispersions-Bildgebung in der Magnetresonanztomographie T1 - Relaxation Dispersion Magnetic Resonance Imaging N2 - Das Ziel dieser Promotion ist der Aufbau eines dreMR Setups für einen klinischen 1,5T Scanner, das die Relaxations-Dispersions-Bildgebung ermöglicht, und die anschließende Ergründung von möglichst vielen Anwendungsfeldern von dreMR. Zu der Aufgabe gehört die Bereitstellung der zugrunde liegenden Theorie, der Bau des experimentellen Setups (Offset-Spule und Stromversorgung) sowie die Programmierung der nötigen Software. Mit dem gebauten Setup konnten zwei große Anwendungsfelder — dreMR Messungen mit und ohne Kontrastmitteln — untersucht werden. N2 - The goal of this dissertation is the design of a dreMR setup for a clinical 1.5T whole body scanner and the subsequent exploration of possible application fields for the dreMR method. This task includes the investigation of the underlying theory, the design and construction of the dreMR setup (offset-coil and current driver) and the preparation of required software. Two major application fields have been demonstrated: dreMR with and without contrast agents. KW - Kernspintomografie KW - MRT KW - Dispersion KW - dreMR KW - MRI KW - dreMR KW - NMR-Tomographie KW - Kontrastmittel KW - Bilderzeugung Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-79554 ER - TY - THES A1 - Wichmann, Tobias T1 - Spulen-Arrays mit bis zu 32 Empfangselementen für den Einsatz an klinischen NMR-Geräten T1 - Coil-Arrays with up to 32 receive channels for the use on clinical NMR systems N2 - In dieser Arbeit wurden für spezielle Anwendungen an klinischen MR-Geräten optimierte Phased-Array-Spulen entwickelt. Das Ziel war, durch die Verwendung neuer Spulen entweder neue Anwendungsgebiete für klinische MR-Geräte zu eröffnen oder bei bestehenden Applikationen die Diagnosemöglichkeiten durch eine Kombination von höherem SNR und kleineren g-Faktoren im Vergleich zu bestehenden Spulen zu verbessern. In Kapitel 3 wurde untersucht, ob es durch den Einsatz neu entwickelter, dedizierter Kleintierspulen sinnvoll möglich ist, Untersuchungen an Kleintieren an klinischen MR-Geräten mit einer Feldstärke von 1,5T durchzuführen. Der Einsatz dieser Spulen verspricht dem klinischen Anwender Studien an Kleintieren durchführen zu können, bei denen er den gleichen Kontrast wie bei einer humanen Anwendung erhält und gleichzeitig Kontrastmittel sowie Sequenzen, die klinisch erprobt sind, einzusetzen. Durch die gewählten geometrischen Abmessungen der Spulen ist es möglich, Zubehör von dedizierten Tier-MR-Geräten, wie z. B. Tierliegen oder EKG- bzw. Atemtriggereinheiten, zu verwenden. Durch Vorversuche an für Ratten dimensionierten Spulen wurden grundlegende Zusammenhänge zwischen verwendetem Entkopplungsmechanismus und SNR bzw. Beschleunigungsfähigkeit erarbeitet. Für Ratten wurde gezeigt, dass in akzeptablen Messzeiten von unter fünf Minuten MR-Messungen des Abdomens in sehr guter Bildqualität möglich sind. Ebenfalls gezeigt wurde die Möglichkeit durch den Einsatz von paralleler Bildgebung sowie Kontrastmitteln hochaufgelöste Angiographien durchzuführen. Es stellte sich heraus, dass bei 1,5T dedizierte Mäusespulen bei Raumtemperatur von den SNR-Eigenschaften am Limit des sinnvoll Machbaren sind. Trotzdem war es möglich, auch für Mäuse ein 4-Kanal-Phased-Array zu entwickeln und den Einsatz bei kontrastmittelunterstützten Applikationen zu demonstrieren. Insgesamt wurde gezeigt, dass durch den Einsatz von speziellen, angepassten Kleintierspulen auch Tieruntersuchungen an klinischen MR-Geräten mit niedriger Feldstärke durchführbar sind. Obwohl sich die Bestimmung der Herzfunktion an MR-Geräten im klinischen Alltag zum Goldstandard entwickelt hat, ist die MR-Messung durch lange Atemanhaltezyklen für einen Herzpatienten sehr mühsam. In Kapitel 4 wurde deswegen die Entwicklung einer 32-Kanal-Herzspule beschrieben, welche den Komfort für Patienten deutlich erhöhen kann. Schon mit einem ersten Prototypen für 3T war es möglich, erstmals Echtzeitbildgebung mit leicht reduzierter zeitlicher Auflösung durchzuführen und damit auf das Atemanhalten komplett zu verzichten. Dies ermöglicht den Zugang neuer Patientengruppen, z. B. mit Arrythmien, zu MR-Untersuchungen. Durch eine weitere Optimierung des Designs wurde das SNR sowie das Beschleunigungsvermögen signifikant gesteigert. Bei einem Beschleunigungsfaktor R = 5 in einer Richtung erhält man z. B. gemittelt über das gesamte Herz ein ca. 60 % gesteigertes SNR zu dem Prototypen. Die Kombination dieser Spule zusammen mit neuentwicklelten Methoden wie z. B. Compressed- Sensing stellt es in Aussicht, die Herzfunktion zukünftig in der klinischen Routine in Echtzeit quantifizieren zu können. In Kapitel 5 wurde die Entwicklung einer optimierten Brustspulen für 3T beschrieben. Bei Vorversuchen bei 1,5T wurden Vergleiche zwischen der Standardspule der Firma Siemens Healthcare und einem 16-Kanal-Prototypen durchgeführt. Trotz größerem Spulenvolumen zeigt die Neuentwicklung sowohl hinsichtlich SNR als auch paralleler Bildgebungseigenschaften eine signifikante Verbesserung gegenüber der Standardspule. Durch die Einhaltung aller Kriterien für Medizinprodukte kann diese Spule auch für den klinischen Einsatz verwendet werden. Mit den verbesserten Eigenschaften ist es beispielsweise möglich, bei gleicher Messdauer eine höhere Auflösung zu erreichen. Aufgrund des intrinsischen SNR-Vorteils der 3 T-Spule gegenüber der 1,5 T-Spule ist es dort sogar möglich, bei höheren Beschleunigungsfaktoren klinisch verwertbare Schnittbilder zu erzeugen. Zusammenfassend wurden für alle drei Applikationen NMR-Empfangsspulen entwickelt, die im Vergleich zu den bisher verfügbaren Spulen, hinsichtlich SNR und Beschleunigungsvermögen optimiert sind und dem Anwender neue Möglichkeiten bieten. N2 - Purpose of this work was to develop optimized phased array coils for clinical magnetic resonance imaging (MRI) systems for applications were dedicated coils were not readily available. Chapter 3 evaluates the use of dedicated small animal coils on clinical MR scanners with a field strength of 1,5T instead of using special animal-systems with higher intrinsic signal-to-noise ratio. Advantage of the clinical system is the availability and the portability of the results of animal studies to human applications because sequences can easily be adopted. The available contrast is similar and clinically tested contrast agents can directly be used. Comparisons of different array decoupling methods with respect to SNR and parallel imaging performance have been conducted on coils with the standard size of rat-coils on animal scanners as part of this work. This geometry made it possible to directly use accessories of these systems like animal beds and monitoring systems. It showed that it is possible to acquire images of the abdomen of the rat in under five minutes in very good image quality with such setup. It was also used for high resolution angiographie in very short scanning time due to the use of parallel imaging techniques. However it has shown that the use of dedicated mouse coils is at the very limit of SNR at 1.5 T. Nevertheless a four channel phased array coil was built and tested. The results are described within this work. Another application which can benefit of novel dedicated coils is the assessment of cardiac function. Especially for heart patients it can be very exhausting to hold breath for a longer period of time, which is required by the current standard protocol for cardiac imaging. The combination of 3T and many available receive channels is a very promising combination to shorten the scan time. Chapter 4 describes the development of a 32 channel cardiac phased array coil for 3T to investigate this idea. Starting with an existing coil for 1.5T a first prototype was developed which was the first coil to demonstrate real-time cardiac imaging with only slightly reduced temporal resolution. A further optimization of this coil led to a completely new coil with higher SNR performance and better parallel imaging abilities and was a further step towards real-time imaging of the heart in clinical routine. Chapter 5 describes the development of an optimized 16 channel breast coil for 3T which can be used in clinical routine. Tests at 1.5T were conducted to find the best coil element layout . It was also possible to compare the prototypes at this field strength to an existing breast coil of Siemens Healthcare. Better SNR and parallel imaging performance could be achieved due to the possibility of adjusting the coil size to different breast sizes and therefore optimizing the filling factor. These improved qualities will allow to have higher resolution in the same scan time compared to the current standard in clinical routine. In conclusion it has been shown that these applications can benefit from dedicated array coils due to better SNR and parallel imaging performance. KW - Kernspintomografie KW - NMR-Tomographie KW - Spulen-Array KW - magnetic resonance imaging KW - coil-array KW - Magnetspule KW - Magnetische Kernresonanz Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-79358 ER - TY - THES A1 - Basse-Lüsebrink, Thomas Christian T1 - Application of 19F MRI for in vivo detection of biological processes T1 - Anwendung der 19F MRT zur in-vivo Detektion von biologischen Prozessen N2 - This thesis focuses on various aspects and techniques of 19F magnetic resonance (MR). The first chapters provide an overview of the basic physical properties, 19F MR and MR sequences related to this work. Chapter 5 focuses on the application of 19F MR to visualize biological processes in vivo using two different animal models. The dissimilar models underlined the wide applicability of 19F MR in preclinical research. A subsection of Chapter 6 shows the application of compressed sensing (CS) to 19F turbo-spin-echo chemical shift imaging (TSE-CSI), which leads to reduced measurement time. CS, however, can only be successfully applied when a sufficient signal-to-noise ratio (SNR) is available. When the SNR is low, so-called spike artifacts occur with the CS algorithm used in the present work. However, it was shown in an additional subsection that these artifacts can be reduced using a CS-based post processing algorithm. Thus, CS might help overcome limitations with time consuming 19F CSI experiments. Chapter 7 deals with a novel technique to quantify the B+1 profile of an MR coil. It was shown that, using a specific application scheme of off resonant pulses, Bloch-Siegert (BS)-based B+1 mapping can be enabled using a Carr Purcell Meiboom Gill (CPMG)-based TSE sequence. A fast acquisition of the data necessary for B+1 mapping was thus enabled. In the future, the application of BS-CPMG-TSE B+1 mapping to improve quantification using 19F MR could therefore be possible. N2 - Diese Arbeit handelt von verschiedenen Aspekten und Techniken der 19F Magnet Resonanz Tomographie (MRT). In den ersten Kapiteln wird auf grundlegenden physikalischen Eigenschaften der MRT, die 19F MRT und MRT Sequenzen eingegangen. Kapitel 5 behandelt die Anwendung von 19F MRT zur in vivo Visualisierung von biologischen Prozessen. Dazu wurden zwei verschiedene Tiermodelle benützt. Diese stark unterschiedlichen Modelle markieren die breite Anwendungsmöglichkeit der 19F MR Bildgebung in der präklinischen Forschung. In einem Unterabschnitt des Kapitels 6 wurde gezeigt, dass Compressed Sensing (CS) zur Beschleunigung von 19F Turbo-Spin-Echo Chemical Shift Imaging (TSE-CSI) Experimenten beitragen kann. Allerdings kann CS nur erfolgreich angewendet werden, wenn ein ausreichendes Signal-Rausch-Verhältnis (SNR) vorhanden ist. Denn ist das nicht der Fall und wird der CS Algorithmus dieser Arbeit verwendet, dann entstehen sogenannte spike Artefakte. In einem weiteren Unterabschnitt wurde aber gezeigt, dass diese Artefakte mit einem CS basierten Algorithmus in der Nachbearbeitung der Daten reduziert werden. Zusammenfassend lässt sich sagen, dass CS, die Beschränkungen durch zeitaufwändigen 19F CSI Experimenten überwinden kann. Kapitel 7 handelt von einer neuartigen Technik um das B+1 Profil einer MR Spule quantitativ auszumessen. Es wurde gezeigt, dass mit einem bestimmten Anwendungsschema von offresonanten Pulsen das Bloch-Siegert (BS)-basiertes B+1 Mapping mit Hilfe einer Carr Purcell Meiboom Gill (CPMG) basierten TSE Sequenz betrieben werden kann. Somit wurde eine schnelle Aufnahme der Daten, die für das B+1 Mapping benötigt werden, erreicht. In der Zukunft könnte das BS-CPMG-TSE B+1 Mapping möglicherweise dazu beitragen, die Quantifizierung mittels 19F MRI zu verbessern. KW - Kernspintomografie KW - Fluor-19 KW - Bloch-Siegert KW - Compressed Sensig KW - 19F-MR KW - Rekonstruktion KW - NMR-Tomographie KW - NMR-Bildgebung Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-77188 ER - TY - THES A1 - Wech, Tobias T1 - Compressed Sensing in der funktionellen kardialen Magnetresonanztomographie T1 - Compressed sensing in functional cardiac magnetic resonance imaging N2 - Die MRT des Herzens wird aufgrund hoher Reproduzierbarkeit und geringer Variabilität als Referenzstandard für die Bestimmung der kardialen Funktion betrachtet. Auch in der präklinischen Forschung bietet die MRT eine ausgezeichnete Charakterisierung der kardialen Funktion und ermöglicht eine exzellente Analyse modellierter Krankheitsbilder. In beiden Fällen besteht jedoch weiterhin Optimierungsbedarf. Die klinische Herz-MRT stellt ein aufwendiges Verfahren mit relativ langer Messzeit dar und ist dadurch mit hohen Untersuchungskosten verbunden. In der präklinischen Kleintierbildgebung müssen zum Erreichen der notwendigen höheren Orts- und Zeitauflösung ebenfalls lange Aufnahmezeiten in Kauf genommen werden. Um die kardiale MRT dort routinemäßig in großen Studienkollektiven anwenden zu können, ist eine schnellere Bildgebung essentiell. Neben einer Verbesserung der Tomographen-Hardware und der Optimierung von Bildgebungssequenzen standen im letzten Jahrzehnt vermehrt informationstheoretische Ansätze zur Beschleunigung der MR-Datenakquisition im Fokus der Entwicklung. Während zu Beginn des Jahrtausends die Parallele Bildgebung (PI) einen Forschungsschwerpunkt repräsentierte, spielte sich in den letzten fünf Jahren vermehrt die von Donoho und Candès eingeführte Compressed Sensing (CS) Theorie in den Vordergrund. Diese ermöglicht eine Signalrekonstruktion aus unvollständig gemessenen Koeffizienten einer linearen Messung (z.B. Fouriermessung) unter Ausnutzung der Sparsität des Signals in einer beliebigen Transformationsbasis. Da sich die MRT hervorragend für den Einsatz von CS eignet, wurde die Technik in der Forschung bereits vielfach angewendet. Die zur Rekonstruktion unterabgetasteter Aufnahmen nötigen CS-Algorithmen haben jedoch eine signifikante Veränderung des Bildgebungsprozesses der MRT zur Folge. Konnte dieser zuvor in guter Näherung als linear und stationär betrachtet werden, so repräsentiert die CS-Rekonstruktion eine nichtlineare und nichtstationäre Transformation. Objektinformation wird nicht mehr ortsunabhängig und proportional zur Intensität in die Abbildung transportiert. Das Bild ist viel mehr das Ergebnis eines Optimierungsprozesses, der sowohl die Konsistenz gegenüber der unterabgetasteten Messung als auch die Sparsität des Signals maximiert. Der erste Teil dieser Dissertation beschreibt eine Methode, die eine objektive Einschätzung der Bildqualität CS-rekonstruierter MR-Bilder ermöglicht. Die CS-Beschleunigung verspricht eine Verkürzung der Messzeit ohne Verlust an Bildqualität, wobei letztere bisher größtenteils qualitativ bzw. quantitativ nur unzureichend beurteilt wurde. Konnte der Bildgebungsprozess der klassischen MRT (linear und stationär) durch die Bestimmung einer Punktspreizfunktion (PSF) robust und effektiv validiert und optimiert werden, erlauben die CS-Algorithmen aufgrund ihres nichtlinearen und nichtstationären Verhaltens ohne Weiteres keine äquivalente Analyse. Um dennoch eine entsprechende Evaluierung des CS-Bildgebungsprozesses zu ermöglichen, wurde die Anwendung einer lokalen Punktspreizfunktion (LPSF) für den in der Folge verwendeten Iterative Soft Thresholding Algorithmus untersucht. Die LPSF berücksichtigt die Ortsabhängigkeit der CS-Rekonstruktion und muss daher für jeden Ort (Pixel) eines Bildes bestimmt werden. Darüber hinaus wurde die LPSF im linearen Bereich der CS-Transformation ermittelt. Dazu wurde das zu bewertende Bild nach Anwenden einer kleinen lokalen Störung rekonstruiert. Die Breite des Hauptmaximums der LPSF wurde schließlich verwendet, um ortsaufgelöste Auflösungsstudien durchzuführen. Es wurde sowohl der Einfluss typischer Unterabtastschemata für CS als auch der Einsatz diskreter Gradienten zur Sparsifizierung eines Phantombildes untersucht. Anschließend wurde die Prozedur zur Bestimmung der räumlichen und zeitlichen Auflösung in der Herzbildgebung getestet. In allen Beispielen ermöglichte das vorgeschlagene Verfahren eine solide und objektive Analyse der Bildauflösung CS-rekonstruierter Aufnahmen. Wurde zuvor meist ausschließlich auf Vergleiche mit einer vollständig abgetasteten Referenz zur Qualitätsbeurteilung zurückgegriffen, so stellt die vorgestellte Auflösungsbestimmung einen Schritt in Richtung einer standardisierten Bildanalyse bei der Verwendung der Beschleunigung mittels CS dar. Die Analyse der Abtastmuster zeigte, dass auch bei der Anwendung von CS die Berücksichtigung der nominell höchsten Frequenzen k_max unerlässlich ist. Frühere Publikationen schlagen Abtastfolgen mit einer teils starken Gewichtung der Messpunkte zum k-Raum-Zentrum hin vor. Die Ergebnisse der vorliegenden Arbeit relativieren ein derartiges Vorgehen, da zumindest bei den durchgeführten Untersuchungen ein Auflösungsverlust bei analoger Vorgehensweise zu verzeichnen war. Ebenso zeigten sich dynamische Aufnahmen, die unter Verwendung des x-f-Raums als sparse Basis rekonstruiert wurden, durchaus anfällig für zeitliches Blurring. Dieses resultiert aus der Unterdrückung hoher zeitlicher Frequenzen und konnte durch die ortsaufgelösten Auflösungskarten sichtbar gemacht werden. Neben der Auflösung ist für eine umfassende Analyse der Bildqualität auch die Untersuchung potentieller Aliasing-Artefakte sowie des Signal-zu-Rausch-Verhältnisses (SNR) notwendig. Während Aliasing mit Hilfe der Einträge der LPSF außerhalb des Hauptmaximums untersucht werden kann, wurde in Kap. 5 eine Modifikation der Multi-Replika-Methode von Robson et al. zur Rauschanalyse bei Verwendung nichtlinearer Algorithmen vorgestellt. Unter Einbeziehung aller genannten Qualitätsparameter ist eine robuste Bewertung der Bildqualität auch bei einer Verwendung von CS möglich. Die differenzierte Evaluierung ebnet den Weg hin zu einem objektiven Vergleich neuer Entwicklungen mit bisherigen Standard-Techniken und kann dadurch den Einzug von CS in die klinische Anwendung vorantreiben. Nach den theoretischen Betrachtungen der Bildqualität behandelt die Dissertation die erstmalige Anwendung von CS zur Beschleunigung der funktionellen Herzdiagnostik in der präklinischen MR-Kleintierbildgebung. Diese Studien wurden in Zusammenarbeit mit der British Heart Foundation Experimental Magnetic Resonance Unit (BMRU) der University of Oxford durchgeführt. Die Algorithmen für eine Beschleunigung mittels der CS-Theorie wurden anhand der dort am 9,4T Tomographen gemessenen (unterabgetasteten) Datensätze entwickelt und optimiert. Zunächst wurde eine Beschleunigung ausschließlich mittels CS untersucht. Dazu wurde die segmentierte, EKG- und Atemgetriggerte kartesische Cine-Aufnahme in Phasenkodierrichtung unterabgetastet und mittels CS rekonstruiert. Die sparse Darstellung wurde durch Ermitteln zeitlicher Differenzbilder für jede Herzphase erhalten. Durch Variation der Abtastmuster in der zeitlichen Dimension konnte ein vollständig abgetastetes zeitliches Mittelbild bestimmt werden, das anschließend von jedem einzelnen Herzphasenbild subtrahiert wurde. In einer Validierungsphase wurden an der Maus vollständig aufgenommene Cine-Akquisitionen retrospektiv unterabgetastet, um die maximal mögliche Beschleunigung mittels CS zu ermitteln. Es wurden u.a. funktionelle Herz-Parameter für jede Gruppe des jeweiligen Beschleunigungsfaktors bestimmt und mittels einer statistischen Analyse verglichen. Die Gesamtheit aller Ergebnisse zeigte die Möglichkeit einer dreifachen Beschleunigung ohne eine Degradierung der Genauigkeit der Methode auf. Die ermittelte Maximalbeschleunigung wurde in einer unterabgetastet gemessenen Bilderserie mit anschließender CS-Rekonstruktion validiert. Die Abtastschemata wurden dazu mit Hilfe der Transformations-Punktspreizfunktion weiter optimiert. In einer Erweiterung der Studie wurde zum Zweck einer noch höheren Beschleunigung die CS-Technik mit der PI kombiniert. Erneut fand eine Unterabtastung der Phasenkodierrichtung einer kartesischen Trajektorie statt. Die Messungen erfolgten mit einer 8-Kanal-Mäusespule an einem 9,4T Tomographen. Um das Potential beider Beschleunigungstechniken auszunutzen, wurden die Methoden CS und PI in serieller Weise implementiert. Für die PI-Beschleunigung wurde der vollständig abgetastete k-Raum zunächst gleichmäßig unterabgetastet. Auf dem resultierenden Untergitter wurde zusätzlich eine Unterabtastung nach Pseudo-Zufallszahlen durchgeführt, um eine Beschleunigung mittels CS zu ermöglichen. Die entwickelte Rekonstruktion erfolgte ebenfalls seriell. Zunächst wurde mittels CS das äquidistante Untergitter rekonstruiert, um anschließend mittels GRAPPA die noch fehlenden Daten zu berechnen. Um eine zusätzliche Messung zur Kalibrierung der GRAPPA-Faktoren zu umgehen, wurde das äquidistant unterabgetastete Untergitter von Herzphase zu Herzphase um je einen Phasenkodierschritt weitergeschoben. Dieses Vorgehen erlaubt die Ermittlung eines vollständig abgetasteten k-Raums mit einer geringeren zeitlichen Auflösung, der die notwendige Bestimmung der Wichtungsfaktoren ermöglicht. Folgende Kombinationen von Beschleunigungsfaktoren wurden mittels retrospektiver Unterabtastung eines vollständig aufgenommenen Datensatzes untersucht: R_CS x R_PI = 2 x 2, 2 x 3, 3 x 2 und 3 x 3. Die Analyse des Bildrauschens, des systematischen Fehlers und der Auflösung führte zu dem Schluss, dass eine sechsfache Beschleunigung mit Hilfe der hybriden Rekonstruktionstechnik möglich ist. Während mit steigender CS-Beschleunigung der systematische Fehler leicht anstieg, führte ein höherer PI-Beschleunigungsfaktor zu einer leichten Verstärkung des statistischen Fehlers. Der statistische Fehler zeigte jedoch ebenfalls eine Verringerung bei steigender Beschleunigung mittels CS. Die Fehler waren allerdings stets auf einem Niveau, das durchaus auch Beschleunigungen bis R_CS x R_PI =3 x 3 zulässt. Die LPSF-Analyse zeigte einen Verlust der räumlichen Auflösung von ca. 50 % bei R=6 sowie einen mittleren Verlust von 64 % bei R=9. Offensichtlich ging die ebenfalls beobachtete Minimierung des Bildrauschens durch den CS-Algorithmus im Falle der relativ stark verrauschten Kleintieraufnahmen zu Lasten der Bildauflösung. Die mit zunehmender Beschleunigung stärker geblurrten Grenzen zwischen Blutpool und Myokardgewebe erschweren die Segmentierung und stellen eine mögliche Fehlerquelle dar. Unter Beachtung aller Ergebnisse ist eine sechsfache Beschleunigung (R_CS x R_PI = 2 x 3, 3 x 2) vertretbar. Die Hinzunahme der PI ermöglicht somit im Vergleich zur alleinigen Verwendung von CS eine weitere Beschleunigung um einen Faktor von zwei. Zusammenfassend ermöglicht der Einsatz von CS in der präklinischen funktionellen Herzbildgebung am Kleintier eine deutliche Reduktion der Messzeit. Bereits ohne Vorhandensein von Mehrkanalspulen kann die notwendige Datenmenge ohne signifikante Beeinflussung der Messergebnisse auf ein Drittel reduziert werden. Ist der Einsatz von Spulenarrays möglich, kann die mit PI mögliche dreifache Beschleunigung um einen weiteren Faktor zwei mittels CS auf R=6 erweitert werden. Dementsprechend kann CS einen wesentlichen Beitrag dazu leisten, dass das Potential Herz-MRT am Kleintier in großen Studienkollektiven effektiver abgerufen werden kann. Im letzten Teil der Arbeit wurde eine Technik für die funktionelle klinische MR-Herzbildgebung entwickelt. Hier wurde eine Beschleunigung mittels CS verwendet, um die Aufnahme des gesamten Herzens innerhalb eines Atemstillstandes des Patienten zu ermöglichen. Bei der derzeitigen Standardmethode werden üblicherweise 10-15 2D-Schichten des Herzens akquiriert, wobei jede einzelne Aufnahme einen Atemstillstand des Patienten erfordert. Für die notwendige Beschleunigung wurde eine unterabgetastete 3D-Trajektorie verwendet. Durch Phasenkodierung einer Richtung sowie radiale Projektionen in den beiden anderen Dimensionen konnte eine effiziente Aufnahme unterhalb des Nyquist-Kriteriums erreicht werden. Die Sparsifizierung erfolgte, wie bereits in der beschriebenen präklinischen Anwendung, durch die Subtraktion eines zeitlichen Mittelbildes. In einer Simulation anhand eines retrospektiv unterabgetasteten Datensatzes konnte die theoretische Funktionalität der Rekonstruktionstechnik bei einer Beschleunigung bezüglich der Nyquist-Abtastung von R ~ 10 validiert werden. Die Unterschiede zum vollständig abgetasteten Datensatz waren vernachlässigbar klein, so dass die vorgeschlagene Abtastfolge am Tomographen implementiert wurde. Mit dieser Sequenz wurde anschließend eine funktionelle Bilderserie an einem gesunden Probanden mit vollständiger Herzabdeckung innerhalb eines Atemstopps aufgenommen. Fehlende Daten wurden analog zur Simulation mit Hilfe des vorgeschlagenen Algorithmus rekonstruiert. Im Vergleich zur Simulation ergaben sich aufgrund des Schichtprofils der 3D-Slab-Anregung zusätzliche Aliasing-Artefakte in den äußeren Partitionen. Die für radiale Aufnahmen typischen Streifenartefakte waren im rekonstruierten Bild, wenn auch mit sehr geringer Amplitude, noch erkennbar. Davon abgesehen wurde die Dynamik jedoch über das gesamte Herz hinweg gut dargestellt. Der hohe Kontrast zwischen Myokard und Blutpool bescheinigt den Bildern eine hervorragende Eignung für die Bestimmung funktioneller Herzparameter mittels einer Segmentierung. Zusammengefasst erlaubt die entwickelte Methode aufgrund der drastischen Reduktion der notwendigen Atemstopps des Patienten einen deutlich erhöhten Patientenkomfort sowie einen schnelleren Durchsatz aufgrund der verkürzten Messzeit. N2 - Because of its high reproducibility and its low variability, magnetic resonance imaging (MRI) of the heart is considered the gold-standard for assessing the cardiac function. In preclinical research, magnetic resonance imaging equally provides an accurate characterization of the cardiac function and enables an excellent analysis of modeled diseases. However, there is still a need for improvement in both applications. Clinical cardiac MRI represents a sophisticated procedure featuring long scan times. This renders the examination comparatively expensive. In preclinical imaging of small animals, long scan times have to be accepted to obtain the required high spatial and temporal resolution. Fast imaging is thus essential for an effective application of cardiac MRI in large collectives. Besides the improvement of the scanner hardware and the optimization of imaging sequences, research in the last decade concentrated on procedures to accelerate MR-data acquisition by exploiting information theory. While numerous publications were associated with parallel imaging (PI) at the beginning of this millennium, the compressed sensing theory (CS) recently gained more and more interest. The latter technique enables the reconstruction of signals from undersampled linear measurements (e.g. the Fourier basis) by exploiting the sparsity of the signal in any known transform domain. As MRI is perfectly qualified for an application of CS, a lot of publications already report on dedicated research. However, the algorithms needed for the reconstruction of undersampled data significantly alter the imaging process of MRI. Classical MRI could be assumed to be linear and stationary in a sufficiently good approximation. The introduction of CS into MRI means a change towards a non-linear and non-stationary transformation. Object information is no longer transferred into an image independently from its location and proportional to its intensity. The image is rather the result of an optimization process maximizing both the fidelity to measured data as well as the sparsity of the signal. The first chapter of this thesis describes a method to objectively evaluate the image quality of MR images reconstructed by CS algorithms. The acceleration with CS promises a reduction of scan time while preserving the image quality. The latter, however, has only been assessed qualitatively or in an insufficient quantitative manner. While classical (linear and stationary) MRI could be validated robustly and effectively by determining a point spread function (PSF), CS algorithms prohibit a corresponding analysis in an analogous manner due to their non-linear and non-stationary behavior. Therefore, the application of a local point spread function (LPSF) was investigated for the iterative soft thresholding (IST) algorithm used in this thesis, to enable a comparative evaluation for imaging systems including CS. The LPSF considers the local dependency of the CS algorithm and thus has to be determined in every location (pixel) of an image. In addition, the LPSF was defined in the linear part of the CS transformation. Small local perturbation on the image to be evaluated were reconstructed for this purpose. The width of the main lobe of the LPSF was used to perform spatially resolved studies on the resolution. The influence of typical undersampling schemes for CS as well as the usage of a discrete gradient transform for a further sparsification were investigated. Subsequently, the procedure was used to assess the spatial and temporal resolution in cardiac MRI. For all CS reconstructions performed in this work, the method allowed a solid and objective analysis of the image resolution. While up to now, comparisons to a fully sampled reference are widely used for a quality assessment, the proposed resolution evaluation represents a step towards a standardized analysis of images obtained by exploiting CS acceleration. The study on sampling schemes revealed, that also for CS accelerated acquisitions, the highest frequencies of the desired k-space have to be included. Former publications proposed undersampling patterns which partly featured a strong weight towards the center of k-space. The results of this thesis put these findings into perspective, as a loss in resolution has been observed for according approaches, at least for the simulations performed in this work. The dynamic acquisitions which were reconstructed exploiting x-f-sparsity proved to be prone to temporal blurring. This is substantiated by the suppression of high temporal frequencies and was analyzed by means of spatially resolved maps. Besides the resolution, an investigation of potential aliasing artifacts as well as the signal-to-noise-ratio (SNR) is essential for a comprehensive quality evaluation. While aliasing may also be investigated by means of the entries of the LPSF outside the main lobe, a modification of the multi-replica method proposed by Robson et al. was presented in chapter 5 to analyze the noise in CS reconstructed images. Taking into account all quality parameters, a robust evaluation of image quality is possible, even when CS is included in the imaging process. This allows a more objective comparison between new developments and present standard procedures and thus may aid the introduction of CS in clinical imaging. After the theoretical analysis on image quality, the next part of this thesis reports on the first application of CS to accelerate functional cardiac MRI of small animals. The studies were performed in cooperation with the British Heart Foundation Experimental Magnetic Resonance Unit (BMRU) of the University of Oxford. The algorithms needed for the CS acceleration were developed and optimized by means of the data acquired by the BMRU at their 9,4 T scanner.\\ An acceleration solely based on CS was investigated first. For this purpose, an ECG- and respiratory gated Cartesian cine acquisition was undersampled in phase encoding direction and reconstructed using CS. The dynamic time series was sparsified by determining temporal difference images for every time frame. A fully sampled temporal average image was obtained by varying the sampling pattern in the temporal dimension. Subsequently, this average image was subtracted from the images of individual heart phases, yielding the sparse temporal difference images. In the validation stage of the study, fully sampled cine acquisitions of mouse hearts were retrospectively undersampled in order to figure out the maximum possible CS acceleration. Cardiac functional parameters were determined for each group of a certain undersampling factor and compared by a statistical analysis. It was shown that a three-fold acceleration is possible without any degradation in the accuracy of the method. This undersampling factor was then validated in an accelerated measurement with a subsequent CS reconstruction. For this purpose, the sampling patterns were further optimized using the transform point spread function. In the subsequent chapter, the CS theory was combined with PI to further increase the acceleration. Again, the phase encoding direction of a Cartesian trajectory was undersampled. The acquisitions were performed using a 9,4 T scanner equipped with an 8 channel mouse coil. In order to exploit the potential of both techniques, CS and PI were combined in a serial manner. First, the k-space was equidistantly undersampled to enable the application of PI. An additional undersampling according to pseudo random numbers was then performed on the resulting sub-grid to allow an acceleration by CS. In consequence, the reconstruction was performed in a serial manner, too. CS was first applied to reconstruct the equidistantly undersampled sub-grid. GRAPPA was used subsequently to compute the still missing data. The equidistantly undersampled sub-grid was shifted from heart phase to heart phase in order to obtain a fully sampled low temporal resolution k-space for a calibration of the GRAPPA-weights. This procedure spares the acquisition time of a separate calibration scan. The following combinations were investigated by retrospectively undersampling a fully sampled cine dataset: R_CS x R_PI = 2 x 2, 2 x 3, 3 x 2 and 3 x 3. The analysis of the noise behavior, the systematic error and the resolution leads to the conclusion that a six-fold acceleration is possible using the proposed hybrid technique. While an increasing factor of the CS acceleration resulted in a slightly larger systematic error, a higher PI acceleration factor led to a slight noise enhancement. However, noise was suppressed for increasing CS acceleration at the same time. In summary, the deviations were at a level which allowed accelerations of up to R_CS x R_PI = 3 x 3. The determination of LPSFs showed a loss in spatial resolution of approximately 50% for a six-fold and up to 64% for nine-fold acceleration. Obviously, the observed suppression of noise was paid by a reduced image resolution for the comparatively noisy acquisitions in small animals. The increased blurring at the endocardial border impedes the segmentation and represents a possible source of error. Taking into consideration all results, a six-fold acceleration (R_CS x R_PI = 2 x 3, 3 x 2) seems reasonable. The additional usage of PI thus enables a further acceleration by a factor of 2 in comparison to an exclusive application of CS. In summary, CS enables a distinct reduction of scan time in preclinical functional cardiac MRI of small animals. Even if no phased-array-coils are available, the necessary amount of data can be reduced to one third without impairing the accuracy of left-ventricular volumes and mass measurements. For acquisitions performed with phased-array-coils, the three-fold acceleration by PI can be extended by an additional two-fold CS acceleration to a joint factor of R=6. Therefore, CS may contribute to an effective application of cardiac MRI in small animals for large collectives. In the last part of the thesis, a modality for clinical functional MRI of the heart was developed. CS was used to enable the acquisition of the whole heart in a single breath-hold of the patient. The current method of choice usually acquires 10-15 2D-slices of the heart, while each measurement requires a separate breath-hold. An undersampled 3D-trajectory was used to reach the necessary acceleration. Phase-encoding in one direction and radial projections in the two remaining ones allowed for an effective acquisition below the Nyquist-criterion. The sparsification of the image series was achieved by subtracting a temporal average image as performed for the preclinical studies. The functionality of the reconstruction technique at an acceleration factor of R ~ 10 was validated in a simulation based on a retrospectively undersampled dataset. The differences between the CS reconstructed and the fully sampled dataset were negligible and thus, the proposed trajectory was implemented at the scanner. An image series depicting the cardiac function with coverage of the full heart was acquired in a single breath-hold of a healthy volunteer using this sequence. Data not covered by the trajectory were reconstructed by the algorithm developed in the validation stage. Due to the slice profile of the fast excitation pulses, additional aliasing artifacts were present in the outer partitions with respect to the images obtained in the simulation. Streaking artifacts of a low intensity were still visible. Apart from that, the dynamics of the heart were excellently captured. The high contrast between the blood pool and the myocardium perfectly qualifies the images for the assessment of cardiac functional parameters. Therefore, the method allows for a higher patient comfort and throughput compared to the gold standard by drastically reducing the amount of necessary breath-holds to a single one. KW - Kernspintomografie KW - Magnetresonanztomographie KW - Kardiale MR-Bildgebung KW - Compressed Sensing KW - magnetic resonance imaging KW - cardiac magnetic resonance imaging KW - compressed sensing KW - Signalregenerierung KW - NMR-Tomographie Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-77179 ER - TY - THES A1 - Ehses, Philipp T1 - Development of new Acquisition Strategies for fast Parameter Quantification in Magnetic Resonance Imaging T1 - Entwicklung neuer Aufnahmeverfahren zur schnellen Parameterbestimmung in der Magnetresonanztomographie N2 - Magnetic resonance imaging (MRI) is a medical imaging method that involves no ionizing radiation and can be used non-invasively. Another important - if not the most important - reason for the widespread and increasing use of MRI in clinical practice is its interesting and highly flexible image contrast, especially of biological tissue. The main disadvantages of MRI, compared to other widespread imaging modalities like computed tomography (CT), are long measurement times and the directly resulting high costs. In the first part of this work, a new technique for accelerated MRI parameter mapping using a radial IR TrueFISP sequence is presented. IR TrueFISP is a very fast method for the simultaneous quantification of proton density, the longitudinal relaxation time T1, and the transverse relaxation time T2. Chapter 2 presents speed improvements to the original IR TrueFISP method. Using a radial view-sharing technique, it was possible to obtain a full set of relaxometry data in under 6 s per slice. Furthermore, chapter 3 presents the investigation and correction of two major sources of error of the IR TrueFISP method, namely magnetization transfer and imperfect slice profiles. In the second part of this work, a new MRI thermometry method is presented that can be used in MRI-safety investigations of medical implants, e.g. cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). One of the major safety risks associated with MRI examinations of pacemaker and ICD patients is RF induced heating of the pacing electrodes. The design of MRI-safe (or MRI-conditional) pacing electrodes requires elaborate testing. In a first step, many different electrode shapes, electrode positions and sequence parameters are tested in a gel phantom with its geometry and conductivity matched to a human body. The resulting temperature increase is typically observed using temperature probes that are placed at various positions in the gel phantom. An alternative to this local thermometry approach is to use MRI for the temperature measurement. Chapter 5 describes a new approach for MRI thermometry that allows MRI thermometry during RF heating caused by the MRI sequence itself. Specifically, a proton resonance frequency (PRF) shift MRI thermometry method was combined with an MR heating sequence. The method was validated in a gel phantom, with a copper wire serving as a simple model for a medical implant. N2 - Die Magnetresonanztomographie (MRT) zeichnet sich als medizinisches Bildgebungsverfahren dadurch aus, dass sie ohne ionisierende Strahlung auskommt und nicht-invasiv einsetzbar ist. Ein weiterer wichtiger - wenn nicht der wichtigste - Grund für die weite und wachsende Verbreitung der MRT in der klinischen Praxis ist ihr interessantes und hoch-flexibles Kontrastverhalten, und damit die gute Darstellbarkeit biologischen Gewebes. Die Hauptnachteile der MRT sind die, verglichen mit z.B. Computer-Tomographie (CT), langen Messzeiten und die damit direkt verbundenen hohen Untersuchungskosten. Der erste Teil dieser Arbeit beschreibt Verbesserungen der IR TrueFISP Methode zur MR-Parameterbestimmung. IR TrueFISP ist eine schnelle Methode zur gleichzeitigen Quantifizierung der Protonendichte, der longitudinalen Relaxationszeit T1, sowie der transversalen Relaxationszeit T2. In Kapitel 2 dieser Arbeit wird eine Methode zur Beschleunigung der IR TrueFISP Quantifizierung vorgestellt, die es erlaubt einen kompletten Relaxometrie-Datensatz in unter 6 s pro Schicht aufzunehmen. Weiterhin werden in Kapitel 3 zwei allgemeine Fehlerquellen der IR TrueFISP Methode untersucht und Korrekturverfahren vorgestellt. Im zweiten Teil dieser Arbeit werden neuartige MR-Thermometrie Methoden vorgestellt, die sich besonders zur Untersuchung der MR-Sicherheit von medizinischen Implanten, insbesondere Herzschrittmachern und implantierbaren Kardioverter-Defibrillatoren (ICDs), eignen. Momentan sind in den allermeisten Fällen MRT Untersuchungen an Herzschrittmacher- und ICD-Patienten aufgrund der damit verbundenen Risiken kontraindiziert. Das dabei am schwierigste in den Griff zu bekommende und damit größte Risiko ist die mögliche Schädigung des Myokards, hervorgerufen durch die von den geschalteten HF-Feldern induzierten Ströme in den Schrittmacherelektroden. Um eine MR-sichere Elektrode und/oder sichere Messprotokole zu entwickeln ist es notwendig viele verschiedene Elektroden, Elektrodenpositionen und Messparameter-Einstellungen in einem körperähnlichen Gel-Phantom untersucht. Die bei der jeweiligen Messung auftretenden Erhitzungen werden dabei meist mit Hilfe fiberoptischer Thermometer an verschiedenen Positionen im Gel gemessen. Eine Alternative ist die Aufnahme einer globalen Karte der Temperaturerhöhung mit Hilfe der MR-Thermometrie. In dieser Arbeit wird eine Messmethode vorgestellt, die MR-Thermometrie mit HF induziertem Heizen kombiniert. Diese Methode wurde an einem Kupferdraht im Gelphantom validiert und mit fiberoptisch gemessenen Temperaturanstiegen verglichen. KW - Kernspintomografie KW - Messprozess KW - Relaxometrie KW - Thermometrie KW - nicht-kartesische Bildgebung KW - Relaxometry KW - Thermometry KW - non-Cartesian imaging KW - Optimierung KW - MRI KW - NMR-Tomographie Y1 - 2011 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-72531 ER - TY - THES A1 - Gutberlet, Marcel T1 - K-Raum-Symmetrie und dichtegewichtete Bildgebung: Optimierung der Magnet-Resonanz-Bildgebung hinsichtlich Signal-zu-Rauschverhältnis, Abbildungsqualität und Messzeit T1 - K-space symmetry and density weighted imaging:Optimization of magnetic resonance imaging with regard to signal-to-noise ratio, image quality and acquisition time N2 - Die Magnet-Resonanz (MR)-Bildgebung ist mit vielfältigen Anwendungen ein nicht mehr wegzudenkendes Instrument der klinischen Diagnostik geworden. Dennoch führt die stark limitierte Messzeit häufig zu einer Einschränkung der erzielbaren räumlichen Auflösung und Abdeckung, einer Beschränkung des Signal-zu-Rauschverhältnis (Signal-to-Noise Ratio) (SNR) sowie einer Signalkontamination durch benachbartes Gewebe. Bereits bestehende Methoden zur Reduktion der Akquisitionszeit sind die partielle Fourier (PF)-Bildgebung und die parallele Bildgebung (PPA). Diese unterscheiden sich zum einen im Schema zur Unterabtastung des k-Raums und zum anderen in der verwendeten Information zur Rekonstruktion der fehlenden k-Raum-Daten aufgrund der beschleunigten Akquisition. Während in der PPA die unterschiedlichen Sensitivitäten einer Mehrkanal-Empfangsspule zur Bildrekonstruktion verwendet werden, basiert die PF-Bildgebung auf der Annahme einer langsamen Variation der Bildphase. Im ersten Abschnitt dieser Arbeit wurde das Konzept der Virtuellen Spulendekonvolutions (Virtual Coil Deconvolution) (VIDE)-Technik vorgestellt, das das gleiche Schema der Unterabtastung des k-Raums wie die konventionelle PPA verwendet, aber anstelle der Spulensensitivität die Bildphase als zusätzliche Information zur Herstellung der fehlenden Daten der beschleunigten Bildgebung verwendet. Zur Minimierung der Rekonstruktionsfehler und der Rauschverstärkung in der VIDE-Technik wurde ein optimiertes Akquisitionsschema entwickelt. Die Kombination der PPA und PF-Bildgebung zur Beschleunigung der MR-Bildgebung wird durch das unterschiedliche Unterabtastschema erschwert. Wie Blaimer et al. in ihrer Arbeit gezeigt haben, kann das Prinzip der VIDE-Technik auf Mehrkanal-Spulen übertragen werden, sodass mit dieser Methode die PPA und die PF-Bildgebung optimal vereint werden können. Dadurch kann die Rauschverstärkung aufgrund der Spulengeometrie ohne zusätzliche Messungen deutlich reduziert werden. Obwohl die Abtastung des k-Raums in der MR-Bildgebung sehr variabel gestaltet werden kann, wird bis heute nahezu ausschließlich die regelmäßige k-Raum-Abtastung in der klinischen Bildgebung verwendet. Der Grund hierfür liegt, neben der schnellen Rekonstruktion und der einfachen Gestaltung der Variation des Bild-Kontrasts, in der Robustheit gegen Artefakte. Allerdings führt die regelmäßige k-Raum-Abtastung zu einer hohen Signalkontamination. Die Optimierung der SRF durch nachträgliches Filtern führt jedoch zu einem SNR-Verlust. Die dichtegewichtete (DW-) Bildgebung ermöglicht die Reduktion der Signal-Kontamination bei optimalem SNR, führt aber zur einer Reduktion des effektiven Gesichtsfelds (FOV) oder einer Erhöhung der Messzeit. Letzteres kann durch eine Kombination der PPA und DW-Bildgebung umgangen werden. Der zweite Teil dieser Arbeit befasste sich mit neuen Aufnahme- und Rekonstruktionsstrategien für die DW-Bildgebung, die eine Erhöhung des FOVs auch ohne Einsatz der PPA erlauben. Durch eine Limitierung der minimalen k-Raum-Abtastdichte konnte durch eine geringfügige Reduktion des SNR-Vorteils der DW-Bildgebung gegenüber der kartesischen, gefilterten Bildgebung eine deutliche Verringerung der Artefakte aufgrund der Unterabtastung in der DW-Bildgebung erreicht werden. Eine asymmetrische Abtastung kann unter der Voraussetzung einer homogenen Bildphase das Aliasing zusätzlich reduzieren. Durch die Rekonstruktion der DW-Daten mit der Virtuelle Spulendekonvolution für die effektive DW-Bildgebung (VIDED)-Bildgebung konnten die Artefakte aufgrund der Unterabtastung eliminiert werden. In der 3d-Bildgebung konnte durch Anwendung der modifizierten DW-Bildgebung eine Steigerung des FOVs in Schichtrichtung ohne Messzeitverlängerung erreicht werden. Die nicht-kartesische k-Raum-Abtastung führt im Fall einer Unterabtastung zu deutlich geringeren, inkohärenten Aliasingartefakten im Vergleich zur kartesischen Abtastung. Durch ein alternierendes DW-Abtastschema wurde eine an die in der MR-Mammografie verwendete Spulengeometrie angepasste k-Raum-Abtastung entwickelt, das bei gleicher Messzeit die räumliche Auflösung, das SNR und das FOV erhöht. Im dritten Teil dieser Arbeit wurde die Verallgemeinerung der DW-Bildgebung auf signalgewichtete Sequenzen, d.h. Sequenzen mit Magnetisierungspräparation (Inversion Recovery (IR), Saturation Recovery (SR)) sowie Sequenzen mit einer Relaxation während der Datenaufnahme (Multi-Gradienten-Echo, Multi-Spin-Echo) vorgestellt, was eine Steigerung der Bildqualität bei optimalem SNR erlaubt. Die Methode wurde auf die SR-Sequenz angewendet und deren praktischer Nutzen wurde in der Herz-Perfusions-Bildgebung gezeigt. Durch die Verwendung der in dieser Arbeit vorgestellten Technik konnte eine Reduktion der Kontamination bei einem SNR-Gewinn von 16% im Vergleich zur konventionellen, kartesischen Abtastung bei gleicher Messzeit erreicht werden. N2 - Magnetic resonance (MR) imaging has become a powerful tool in clinical diagnostics. However, long acquisition times used in MR imaging limit the available signal-to-noise Ratio (SNR), spatial resolution and coverage and cause signal contamination from neighboring tissue. Two established methods used to reduce the scan time of MR imaging are partial parallel acquisition (PPA) and partial fourier (PF) imaging. These methods use different schemes to undersample k-space and use a different kind of information to reconstruct the missing data resulting from the accelerated acquisition. While in PPA the varying sensitivities of a multi-channel receiver coil are used in the image reconstruction, PF imaging is based on the assumption of a smoothly varying image phase. In the first section of this work, the concept of virtual coil deconvolution (VIDE) imaging is proposed. This method uses the identical acquisition scheme for the accelerated measurement of k-space as PPA. However, in contrast to PPA, VIDE imaging uses the image phase instead of the varying coil sensitivities to recover the missing data of the accelerated acquisition. Reconstruction errors and noise amplification of VIDE imaging were minimized by an optimized acquisition scheme. VIDE imaging allows an acceleration of MR imaging by a factor of two. The different sampling schemes used in PF imaging and PPA are disadvantageous for the combination of PF imaging and PPA to increase the acceleration of MRI. Blaimer, Gutberlet et al. showed that the concept of VIDE imaging can be extended to multi-channel receiver coils. This allows an optimal combination of PF imaging and PPA. The noise amplification caused by the coil geometry could be significantly decreased without lengthening the scan time. Although k-space can be measured in a variety of sampling schemes, almost exclusively a Cartesian trajectory is used in clinical imaging. Reasons are the fast and simple reconstruction, the robustness against artifacts and the well-defined contrast of Cartesian imaging. However, the Cartesian acquisition results in increased signal contamination. Post-processing filtering allows reduction of contamination but at the expense of SNR. Density weighted (DW) imaging allows optimization of the spatial response function (SRF) at maximum SNR but results in a reduced effective field of view (FOV) or a lengthening of the scan time. This disadvantage of DW imaging can be eliminated by the application of PPA. In the second section new acquisition and reconstruction methods were presented allowing an increase of the effective FOV in DW imaging even without the use of PPA. The limitation of the minimum sampling density in DW imaging resulted in a significant reduction of aliasing. Moderate filtering to correct the k-space weighting resulted in low reduction of the SNR gain in comparison to Cartesian imaging with the identical scan time. On condition of a homogeneous image phase, the aliasing can be additionally reduced by using an asymmetric DW sampling. Using virtual coil deconvolution for effective density weighted (VIDED) imaging for reconstruction, aliasing could be eliminated. By applying the developed DW method, the spatial coverage of 3D imaging was increased even without a lengthening of the scan time. In case of undersampling k-space, DW acquisition results in significantly reduced incoherent aliasing in comparison to Cartesian imaging. Alternating DW sampling revealed an optimized sampling scheme for the coil geometry used in MR-mammography. In experiments the effective FOV or the spatial resolution in slice direction could be increased significantly. In the third section the extension of DW imaging to signal-weighted sequences, i.e. sequences with magnetization preparation (inversion recovery or saturation recovery) or with relaxation between the acquired echoes (multigradient echo, multi-spin echo), was presented. The method provided increased image quality at optimum SNR in comparison to Cartesian imaging. By applying the new technique to SR-sequences, its practical use could be shown in myocardial perfusion imaging. The method presented in this work allowed an optimization of the SRF with an SNR gain of 16% compared to conventional Cartesian sampling at identical scan time. KW - Kernspintomografie KW - Paralleler Prozess KW - Messprozess KW - Dichtegewichtete Bildgebung KW - Parallele Bildgebung KW - VIDE KW - VIDED KW - density weighted imaging KW - density weighting KW - parallel imaging KW - VIDE KW - VIDED KW - Optimierung KW - NMR-Bildgebung KW - NMR-Mammographie KW - Koronarperfusion Y1 - 2011 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-71834 ER - TY - THES A1 - Weber, Daniel T1 - Morphologische und funktionelle MRT-Infarktcharakterisierung und Entwicklung einer diffusionsgewichteten MRT-Methode T1 - Morphological and functional MRI infarct characterization and development of a diffusion-weighted MRI method N2 - Diffusionstensorbildgebung im Vergleich zu anderen Parametermethoden für die Infarktcharakterisierung Ziel dieses Teils der Arbeit war die Klärung der Frage, welches Potential verschiedene MR-Parametersequenzen bei der Charakterisierung eines myokardialen Infarkts sowohl im akuten als auch im chronischen Fall haben. Dazu wurde eine Studie mit akut und chronisch infarzierten Rattenherzen durchgeführt. Untersucht wurden die Parameter T1, T2 und T2* sowie die aus der Diffusionstensorbildgebung berechneten Parameter ADC, FA, cs, cp und cl . Es zeigte sich, dass es kein Analogon zum bei einer cerebralen Ischämie bekannten Mismatch-Konzept gibt. Weder im akuten noch im chronischen war Fall eine ausgewiesene Differenz im diagnostizierten Infarktareal zwischen verschiedenen Sequenzen feststellbar. Alles in allem eignen sich zur detaillierten Charakterisierung der Infarktnarbe am besten eine T2*- oder eine Diffusionstensorsequenz. Die T2*-Sequenz liefert optisch das aufschlussreichere Bild, die aufwendigere Diffusionstensorsequenz dagegen bietet aufgrund der vielfachen Darstellungsmöglichkeiten im Postprocessing ein Mehr an Information und zeigt dazu eine Veränderung der Narbe im Zeitverlauf. Oxygenierungsmessung am Mäuseherz in vivo Die Charakterisierung einer Infarktnarbe kann auch über die Darstellung morphologischer Strukturen hinaus erfolgen. Die Oxygenierung ist ein komplexer Parameter, der funktionelle Auskunft über die Vaskularisierung und Viabilität des Gewebes geben kann. Zugang zu diesem Parameter erhält man über T2*-Messungen, da der Parameter T2* sensitiv auf chemisch gebundenen Sauerstoff reagiert. Hier wurden der Einfluss von reiner Sauerstoffatmung im Gegensatz zu normaler Raumluftatmung auf die Oxygenierung bei gesunden und infarzierten Mäusen untersucht. Die Messungen wurden trotz der Schwierigkeiten, die durch die Bewegung durch Atmung und Herzschlag entstehen, in vivo bei 17,6 Tesla implementiert und durchgeführt. Die Auflösung war ausreichend, um auch nach Infarkt extrem ausgedünnte Myokardwände gut auflösen und charakterisieren zu können. Der Effekt auf das Oxygenierungslevel ist stark unterschiedlich zwischen normalen und infarzierten Herzen, woraus auf eine noch nicht weit fortgeschrittene Revaskularisierung der Narbe eine Woche nach Infarzierung geschlossen werden kann. Die Methode wurde darüber hinaus an einem 7,0 Tesla-Magneten zur Verwendung an Ratten implementiert und auf das im Gegensatz zur Maus veränderte Atmungsverhalten der Ratte angepasst. Zum einen kann dadurch der Einfluss des hohen Magnetfeldes auf die Oxygenierungsmessung untersucht werden, zum anderen ist das Herz als zu untersuchendes Objekt bei der Ratte größer. Diffusionswichtung mittels Hole-Burning Die in dieser Arbeit zur Charakterisierung des Herzens verwendete Diffusionsmethode kann im Grenzfall von kurzen T2-Relaxationszeiten an ihre Grenzen stoßen: Bei den verwendeten starken Magnetfeldern klingt das messbare Signal aufgrund der Relaxationszeit T2 oft sehr schnell ab. Daher wurde eine Methode entwickelt, die einen völlig neuen Ansatz zur diffusionsgewichteten Bildgebung verfolgt, bei dem die Informationen über die Diffusion unabhängig von der limitierenden T2-Zeit gewonnen werden können. Die sog. Hole-Burning-Diffusionssequenz verwendet in einem Vorexperiment lediglich die Longitudinalmagnetisierung zur Diffusionswichtung. Das Signal wird dann mit einer schnellen Auslesesequenz akquiriert. Bei der Präparation werden zunächst auf Subvoxel-Niveau Streifen "gebrannt", d.h. die Magnetisierung wird dort gesättigt. Bis zur nächsten Sättigung ist das Verhalten der Magnetisierung abhängig von der T1-Relaxation in diesem Bereich und vom Diffusionsverhalten. Durch rasches Wiederholen des selektiven Pulszugs wird schließlich eine Gleichgewichtsmagnetisierung erreicht, die von der Diffusionskonstanten D und der T1-Relaxationszeit abhängt. Im Rahmen dieser Arbeit wurden die Abhängigkeiten verschiedener Sequenzparameter untersucht und diese mittels Simulationen optimiert. Außerdem wurde die Sequenz an einem Scanner implementiert und erste Experimente damit durchgeführt. Mit Hilfe von Simulationen konnten dazu Lookup-Tabellen generiert werden, mit denen in bestimmten Bereichen (insbesondere bei nicht zu kurzen T1-Relaxationszeiten) sowohl die Diffusionskonstante D als auch die T1-Relaxationszeit quantifiziert werden konnte. N2 - Diffusion tensor imaging for the characterization of myocardial infarction in comparison to other methods The aim of this part of this work was to evaluate the potential of different MR sequences for the characterization of myocardial infarction in both the acute and chronic case. Therefore a study of acute as well as chronic infarcted rat hearts was performed, and the parameters T1, T2, T2* and the parameters ADC, FA, cs, cp and cl calculated from the diffusion tensor images were investigated. It turned out that there is no equivalent to the ischemia. Neither in the acute nor in the chronic case, a notably difference inside the affected area was detectable between different sequences. All in all, for detailed characterization of the infarct scar a T2* or a diffusion tensor sequence are most suitable. The T2* sequence provides a more informative visual image, whereas the more time-consuming diffusion tensor sequence provides a surplus of information due to the multiple display options in post-processing and shows the remodelling of the scar tissue over time. Oxygen level measurements in mouse hearts in vivo The characterization of an infarct scar can also go beyond the representation of morphological structure. The oxygenation is a complex parameter that can provide functional information of the vascularization and viability of the tissue. Access to this parameter is obtained by T2*-measurements, as the parameter T2* is sensitive to chemically bound oxygen. The influence of pure oxygen breathing in contrast to normal room air breathing on the oxygenation level in healthy and infarcted mice have been explored. Despite the difficulties caused by the movement due to respiration and heartbeat the measurements were implemented and carried out at 17.6 Tesla in vivo. The resolution was sufficient to resolve and investigate extremely thinned heart walls after infarction. The effect on the oxygenation level varies considerably between normal and infarcted hearts; that may be caused by a not yet advanced revascularization of the scar. In addition, the method was implemented to a 7.0 Tesla magnet for use in rats and adapted to the respiration of rats, which is different to the respiration of mice. The first reason was that the influence of the higher magnetic field on the measurement of the oxygenation level could be examined. Second, the heart as the examined object is larger in rats. Diffusion weighting using hole burning The MR diffusion method used in this work for the characterization of myocardial infarctions could be limited by extremely short T2 relaxation times. With the strong magnetic fields used here the measurable signal decays very fast due to the relaxation time T2. Therefore, a method for a completely new approach to diffusion-weighted imaging was developed, where the diffusion weighting can be obainted without being limited by the time constant T2. The so-called hole-burning diffusion sequence uses only the longitudinal magnetization for the diffusion weighting in a preliminary experiment. The signal is then acquired with a fast read-out sequence. During the preparation stripes will be "burned" into the magnetization on a subvoxel level, i.e. the magnetization is saturated there. Until the next saturation pulse the behavior of the magnetization depends first on the T1 relaxation time in this area and second on the diffusion. By rapidly repeating the selective pulse train a steady state magnetization dependend on the diffusion constant D and the T1 relaxation time is reached. In this work the dependencies between different sequence parameters were investigated and optimized using simulations. In addition, the sequence was implemented on a MR scanner and first experiments were carried out. With simulated lookup-tables we were able to quantify both the diffusion coefficient D and the T1 relaxation time in the case of not too short relaxation times T1. KW - Kernspintomografie KW - Infarkt KW - MRI KW - infarct KW - characterization KW - diffusion KW - hole-burning KW - NMR-Tomographie KW - Anisotrope Diffusion KW - Diffusion KW - Spektrales Lochbrennen KW - Herzinfarkt Y1 - 2011 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-71157 ER -