TY - THES A1 - Ullherr, Maximilian T1 - Optimization of Image Quality in High-Resolution X-Ray Imaging T1 - Optimierung von Bildqualität in der hochauflösenden Röntgenbildgebung N2 - The SNR spectra model and measurement method developed in this work yield reliable application-specific optima for image quality. This optimization can either be used to understand image quality, find out how to build a good imaging device or to (automatically) optimize the parameters of an existing setup. SNR spectra are here defined as a fraction of power spectra instead of a product of device properties. In combination with the newly developed measurement method for this definition, a close correspondence be- tween theory and measurement is achieved. Prior approaches suffer from a focus on theoretical definitions without fully considering if the defined quantities can be measured correctly. Additionally, discrepancies between assumptions and reality are common. The new approach is more reliable and complete, but also more difficult to evaluate and interpret. The signal power spectrum in the numerator of this fraction allows to model the image quality of different contrast mechanisms that are used in high-resolution x-ray imaging. Superposition equations derived for signal and noise enable understanding how polychromaticity (or superposition in general) affects the image quality. For the concept of detection energy weighting, a quantitative model for how it affects im- age quality was found. It was shown that—depending on sample properties—not detecting x-ray photons can increase image quality. For optimal computational energy weighting, more general formula for the optimal weight was found. In addition to the signal strength, it includes noise and modulation transfer. The novel method for measuring SNR spectra makes it possible to experimentally optimize image quality for different contrast mechanisms. This method uses one simple measurement to obtain a measure for im- age quality for a specific experimental setup. Comparable measurement methods typically require at least three more complex measurements, where the combination may then give a false result. SNR spectra measurements can be used to: • Test theoretical predictions about image quality optima. • Optimize image quality for a specific application. • Find new mechanisms to improve image quality. The last item reveals an important limitation of x- ray imaging in general: The achievable image quality is limited by the amount of x-ray photons interacting with the sample, not by the amount incident per detector area (see section 3.6). If the rest of the imaging geometry is fixed, moving the detector only changes the field of view, not the image quality. A practical consequence is that moving the sample closer to the x-ray source increases image quality quadratically. The results of a SNR spectra measurement represent the image quality only on a relative scale, but very reliable. This relative scale is sufficient for an optimization problem. Physical effects are often already clearly identifiable by the shape of the functional relationship between input parameter and measurement result. SNR spectra as a quantity are not well suited for standardization, but instead allow a reliable optimization. Not satisfying the requirements of standardization allows to use methods which have other advantages. In this case, the SNR spectra method describes the image quality for a specific application. Consequently, additional physical effects can be taken into account. Additionally, the measurement method can be used to automate the setting of optimal machine parameters. The newly proposed image quality measure detection effectiveness is better suited for standardization or setup comparison. This quantity is very similar to measures from other publications (e.g. CNR(u)), when interpreted monochromatically. Polychromatic effects can only be modeled fully by the DE(u). The measurement processes of both are different and the DE(u) is fundamentally more reliable. Information technology and digital data processing make it possible to determine SNR spectra from a mea- sured image series. This measurement process was designed from the ground up to use these technical capabilities. Often, information technology is only used to make processes easier and more exact. Here, the whole measurement method would be infeasible without it. As this example shows, using the capabilities of digital data processing much more extensively opens many new possibilities. Information technology can be used to extract information from measured data in ways that analog data processing simply cannot. The original purpose of the SNR spectra optimization theory and methods was to optimize high resolution x-ray imaging only. During the course of this work, it has become clear that some of the results of this work affect x-ray imaging in general. In the future, these results could be applied to MI and NDT x-ray imaging. Future work on the same topic will also need to consider the relationship between SNR spectra or DE(u) and sufficient image quality.This question is about the minimal image quality required for a specific measurement task. N2 - Das in dieser Arbeit entwickelte Modell und die Messmethode für SNR Spektren ergeben zuverlässige anwendungsspezifische Optima für die Bildqualität. Diese Optimierung kann verwendet werden, entweder um Bildqualität zu verstehen, um herauszufinden wie ein gutes Bildgebungsgerät gebaut werden kann oder um die Parameter eines existierenden Aufbaus (automatisch) festzulegen. ... KW - Bildqualität KW - Bildgebendes Verfahren KW - Computertomografie KW - x-ray imaging KW - x-ray microscopy KW - image quality KW - signal to noise ratio KW - computed tomography KW - x-ray inline phase contrast Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-231171 ER - TY - THES A1 - Schielein, Richard T1 - Analytische Simulation und Aufnahmeplanung für die industrielle Röntgencomputertomographie T1 - Analytical simulation and acquisition planning for industrial x-ray computed tomography N2 - Röntgencomputertomographie (CT) hat in ihrer industriellen Anwendung ein sehr breites Spektrum möglicher Prüfobjekte. Ziel einer CT-Messung sind dreidimensionale Abbilder der Verteilung des Schwächungskoeffizienten der Objekte mit möglichst großer Genauigkeit. Die Parametrierung eines CT-Systems für ein optimales Messergebnis hängt stark vom zu untersuchenden Objekt ab. Eine Vorhersage der optimalen Parameter muss die physikalischen Wechselwirkungen mit Röntgenstrahlung des Objektes und des CT-Systems berücksichtigen. Die vorliegende Arbeit befasst sich damit, diese Wechselwirkungen zu modellieren und mit der Möglichkeit den Prozess zur Parametrierung anhand von Gütemaßen zu automatisieren. Ziel ist eine simulationsgetriebene, automatische Parameteroptimierungsmethode, welche die Objektabhängigkeit berücksichtigt. Hinsichtlich der Genauigkeit und der Effizienz wird die bestehende Röntgensimulationsmethodik erweitert. Es wird ein Ansatz verfolgt, der es ermöglicht, die Simulation eines CT-Systems auf reale Systeme zu kalibrieren. Darüber hinaus wird ein Modell vorgestellt, welches zur Berechnung der zweiten Ordnung der Streustrahlung im Objekt dient. Wegen des analytischen Ansatzes kann dabei auf eine Monte-Carlo Methode verzichtet werden. Es gibt in der Literatur bisher keine eindeutige Definition für die Güte eines CT-Messergebnisses. Eine solche Definition wird, basierend auf der Informationstheorie von Shannon, entwickelt. Die Verbesserungen der Simulationsmethodik sowie die Anwendung des Gütemaßes zur simulationsgetriebenen Parameteroptimierung werden in Beispielen erfolgreich angewendet beziehungsweise mittels Referenzmethoden validiert. N2 - Industrial X-ray computed tomography (CT) can be applied to a large variety of different specimens. The result of a CT measurement is a three-dimensional image containing the position-dependent attenuation coefficient of the specimen. For an optimal imaging CT-measurement parameters depend on both the properties of the CT-System and the specimen. To predict such an optimal parameterization both the physical interactions with X-rays of the CT-System and the specimen, must be taken into account. This thesis sets out to address the modelling of the interactions as well as the automatization of the parameter finding. The latter is based on a figure of merit for CT-measurements. Aim is a simulation-based, automatic parameter optimization method which includes the object-dependency on distinct specimens. The currently existing X-ray simulation methods are enhanced with respect to accuracy and efficiency. Therefore a method for the calibration of the simulation to a real CT-system is presented. Additionally, a model for second order X-ray scattering is developed in order to calculate the specimen-scattered radiation. This is done using an analytical ansatz and no Monte-Carlo method has to be applied. So far, no universal definition of a figure of merit for CT-results has been given in literature. Using Shannon's information theory such a definition is developed. The improvements of the simulation method and the application of the figure of merit for simulation-based parameter optimization are used in examples or are validated using reference methods. KW - Computertomografie KW - Zerstörungsfreie Werkstoffprüfung KW - Optimierung KW - Aufnahmeplanung KW - Analytische Simulation KW - CT KW - Computersimulation Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-169236 ER - TY - THES A1 - Dremel, Kilian T1 - Modellbildung des Messprozesses und Umsetzung eines modellbasierten iterativen Lösungsverfahrens der Schnittbild-Rekonstruktion für die Röntgen-Computertomographie T1 - Modeling of the process of measurement and development of a model-based iterative reconstruction for X-ray computed tomography N2 - In der computertomographischen Schnittbildgebung treten Artefakte, also Anteile des Ergebnisses auf, die nicht Teil des gemessenen Objekts sind und die somit die Auswertbarkeit der Ergebnisse beeinflussen. Viele dieser Artefakte sind auf die Inkonsistenz des Modells der Rekonstruktion zur Messung zurückzuführen. Gerade im Hinblick auf Artefakte durch die Energieabhängigkeit der rekonstruierten Schwächungskoeffizienten und Abweichungen der Geometrieinformation des Rekonstruktionsmodells wird häufig der Weg einer Nachbearbeitung der Messdaten beschritten, um Rekonstruktionsartefakte zu vermeiden. Im Zuge dieser Arbeit wird ein Modell der computertomographischen Aufnahme mit Konzentration auf industrielle und materialwissenschaftliche Systeme erstellt, das nicht genutzt wird um die Messdaten zu verändern, sondern um das Rekonstruktionsmodell der Aufnahmerealität anzupassen. Zunächst werden iterative Rekonstruktionsverfahren verglichen und ein passender Algorithmus ausgewählt, der die gewünschten Modifikationen des Aufnahmemodells erlaubt. Für diese Modifikationen werden bestehende Methoden erweitert und neue modellbasierte Ansätze entwickelt, die in den Rekonstruktionsablauf integriert werden können. Im verwendeten Modell werden die Abhängigkeiten der rekonstruierten Werte vom polychromatischen Röntgenspektrum in das Simulationsmodell des Rekonstruktionsprozesses eingebracht und die Geometrie von Brennfleck und Detektorelementen integriert. Es wird gezeigt, dass sich durch die verwendeten Methoden Artefakte vermeiden lassen, die auf der Energieabhängigkeit der Schwächungskoeffizienten beruhen und die Auflösung des Rekonstruktionsbildes durch Geometrieannahmen gesteigert werden kann. Neben diesen Ansätzen werden auch neue Erweiterungen der Modellierung umgesetzt und getestet. Das zur Modellierung verwendete Röntgenspektrum der Aufnahme wird im Rekonstruktionsprozess angepasst. Damit kann die benötigte Genauigkeit dieses Eingangsparameters gesenkt werden. Durch die neu geschaffene Möglichkeit zur Rekonstruktion der Kombination von Datensätzen die mit unterschiedlichen Röntgenspektren aufgenommen wurden wird es möglich neben dem Schwächungskoeffizienten die Anteile der Comptonabsorption und der photoelektrischen Absorption getrennt zu bestimmen. Um Abweichungen vom verwendeten Geometriemodell zu berücksichtigen wird eine Methode auf der Basis von Bildkorrelation implementiert und getestet, mit deren Hilfe die angenommene Aufnahmegeometrie automatisch korrigiert wird. Zudem wird in einem neuartigen Ansatz zusätzlich zur detektorinternen Streustrahlung die Objektstreustrahlung während des Rekonstruktionsprozesses deterministisch simuliert und so das Modell der Realität der Messdatenaufnahme angepasst. Die Umsetzung des daraus zusammengesetzten Rekonstruktionsmodells wird an Simulationsdatensätzen getestet und abschließend auf Messdaten angewandt, die das Potential der Methode aufzeigen. N2 - In computed tomography, parts of the result which are not features of the measured object -- so called artifacts -- occur and thus impair the evaluability of the results. Reconstruction methods require a model of the measurement. Many artifacts are induced by the inconsistency between the model of reconstruction and the measurement. Especially with regard to artifacts due to the energy dependence of the reconstructed attenuation coefficients and deviations of the geometry information of the reconstruction model, a frequently used method is the postprocessing of the measurement data to avoid reconstruction artifacts. In this thesis a model of computed tomography measurements with focus on systems used for industrial and material science purposes is developed that is not used to change the measured data, but to adapt the reconstruction model to the reality of measurement. Firstly, iterative reconstruction methods are compared and a suitable algorithm is selected that allows the desired modifications of the model. Therefore existing methods are extended and new model-based approaches are developed that can be integrated in the reconstruction process. The dependencies of the reconstructed values ??from the polychromatic X-ray spectrum are incorporated into the simulation model of the reconstruction process and the geometry of the focal spot and detector elements are integrated. Thereby artefacts caused by the energy-dependency of the attenuation coefficients are shown to be reduced and the resolution of the resulting data is shown to be increased by geometric modelling. Alongside these approaches of modeling new methods are developed and implemented. The X-ray spectrum used for the modeling is adapted during the reconstruction. Thereby the accuracy needed for this input parameter is lowered. Due to possibility of the combination of data sets scanned using different spectra the reconstruction of the Compton- and photoelectric parts of the attenuation coefficient becomes possible. To consider deviations of the geometry model used in the reconstruction a correlation-based method is implemented and tested to automatically correct these aberrations. In addition to radiation scattered within the detector, a new method is developed to simulate the object scattering during the reconstruction process and the model is therefore adapted to the reality of the measurement. The implementation of the reconstruction model composed therefrom is tested on simulation data sets and finally applied to measurement data which show the potential of the method. KW - Dreidimensionale Rekonstruktion KW - Computertomografie KW - Modellbasierte Rekonstruktion Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-157718 ER -