TY - JOUR A1 - Helmer, Philipp A1 - Hottenrott, Sebastian A1 - Rodemers, Philipp A1 - Leppich, Robert A1 - Helwich, Maja A1 - Pryss, Rüdiger A1 - Kranke, Peter A1 - Meybohm, Patrick A1 - Winkler, Bernd E. A1 - Sammeth, Michael T1 - Accuracy and Systematic Biases of Heart Rate Measurements by Consumer-Grade Fitness Trackers in Postoperative Patients: Prospective Clinical Trial JF - Journal of Medical Internet Research N2 - Background: Over the recent years, technological advances of wrist-worn fitness trackers heralded a new era in the continuous monitoring of vital signs. So far, these devices have primarily been used for sports. Objective: However, for using these technologies in health care, further validations of the measurement accuracy in hospitalized patients are essential but lacking to date. Methods: We conducted a prospective validation study with 201 patients after moderate to major surgery in a controlled setting to benchmark the accuracy of heart rate measurements in 4 consumer-grade fitness trackers (Apple Watch 7, Garmin Fenix 6 Pro, Withings ScanWatch, and Fitbit Sense) against the clinical gold standard (electrocardiography). Results: All devices exhibited high correlation (r≥0.95; P<.001) and concordance (rc≥0.94) coefficients, with a relative error as low as mean absolute percentage error <5% based on 1630 valid measurements. We identified confounders significantly biasing the measurement accuracy, although not at clinically relevant levels (mean absolute error<5 beats per minute). Conclusions: Consumer-grade fitness trackers appear promising in hospitalized patients for monitoring heart rate. KW - Withings ScanWatch KW - health tracker KW - smartwatch KW - internet of things KW - personalized medicine KW - photoplethysmography KW - wearable KW - Garmin Fenix 6 Pro KW - Apple Watch 7 KW - Fitbit Sense Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-299679 VL - 24 IS - 12 ER - TY - JOUR A1 - Helmer, Philipp A1 - Rodemers, Philipp A1 - Hottenrott, Sebastian A1 - Leppich, Robert A1 - Helwich, Maja A1 - Pryss, Rüdiger A1 - Kranke, Peter A1 - Meybohm, Patrick A1 - Winkler, Bernd E. A1 - Sammeth, Michael T1 - Evaluating blood oxygen saturation measurements by popular fitness trackers in postoperative patients: a prospective clinical trial JF - iScience N2 - Summary Blood oxygen saturation is an important clinical parameter, especially in postoperative hospitalized patients, monitored in clinical practice by arterial blood gas (ABG) and/or pulse oximetry that both are not suitable for a long-term continuous monitoring of patients during the entire hospital stay, or beyond. Technological advances developed recently for consumer-grade fitness trackers could—at least in theory—help to fill in this gap, but benchmarks on the applicability and accuracy of these technologies in hospitalized patients are currently lacking. We therefore conducted at the postanaesthesia care unit under controlled settings a prospective clinical trial with 201 patients, comparing in total >1,000 oxygen blood saturation measurements by fitness trackers of three brands with the ABG gold standard and with pulse oximetry. Our results suggest that, despite of an overall still tolerable measuring accuracy, comparatively high dropout rates severely limit the possibilities of employing fitness trackers, particularly during the immediate postoperative period of hospitalized patients. Highlights •The accuracy of O2 measurements by fitness trackers is tolerable (RMSE ≲4%) •Correlation with arterial blood gas measurements is fair to moderate (PCC = [0.46; 0.64]) •Dropout rates of fitness trackers during O2 monitoring are high (∼1/3 values missing) •Fitness trackers cannot be recommended for O2 measuring during critical monitoring KW - multidisciplinary KW - health sciences KW - clinical measurement in health technology KW - bioelectronics KW - fitness trackers Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-349913 SN - 2589-0042 VL - 26 IS - 11 ER - TY - JOUR A1 - Herrmann, Johannes A1 - Adam, Elisabeth Hannah A1 - Notz, Quirin A1 - Helmer, Philipp A1 - Sonntagbauer, Michael A1 - Ungemach-Papenberg, Peter A1 - Sanns, Andreas A1 - Zausig, York A1 - Steinfeldt, Thorsten A1 - Torje, Iuliu A1 - Schmid, Benedikt A1 - Schlesinger, Tobias A1 - Rolfes, Caroline A1 - Reyher, Christian A1 - Kredel, Markus A1 - Stumpner, Jan A1 - Brack, Alexander A1 - Wurmb, Thomas A1 - Gill-Schuster, Daniel A1 - Kranke, Peter A1 - Weismann, Dirk A1 - Klinker, Hartwig A1 - Heuschmann, Peter A1 - Rücker, Viktoria A1 - Frantz, Stefan A1 - Ertl, Georg A1 - Muellenbach, Ralf Michael A1 - Mutlak, Haitham A1 - Meybohm, Patrick A1 - Zacharowski, Kai A1 - Lotz, Christopher T1 - COVID-19 Induced Acute Respiratory Distress Syndrome — A Multicenter Observational Study JF - Frontiers in Medicine N2 - Background: Proportions of patients dying from the coronavirus disease-19 (COVID-19) vary between different countries. We report the characteristics; clinical course and outcome of patients requiring intensive care due to COVID-19 induced acute respiratory distress syndrome (ARDS). Methods: This is a retrospective, observational multicentre study in five German secondary or tertiary care hospitals. All patients consecutively admitted to the intensive care unit (ICU) in any of the participating hospitals between March 12 and May 4, 2020 with a COVID-19 induced ARDS were included. Results: A total of 106 ICU patients were treated for COVID-19 induced ARDS, whereas severe ARDS was present in the majority of cases. Survival of ICU treatment was 65.0%. Median duration of ICU treatment was 11 days; median duration of mechanical ventilation was 9 days. The majority of ICU treated patients (75.5%) did not receive any antiviral or anti-inflammatory therapies. Venovenous (vv) ECMO was utilized in 16.3%. ICU triage with population-level decision making was not necessary at any time. Univariate analysis associated older age, diabetes mellitus or a higher SOFA score on admission with non-survival during ICU stay. Conclusions: A high level of care adhering to standard ARDS treatments lead to a good outcome in critically ill COVID-19 patients. KW - COVID-19 KW - ARDS (acute respiratory distress syndrome) KW - intensive care medicine KW - pandemia KW - Germany Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-219834 SN - 2296-858X VL - 7 ER -