TY - JOUR A1 - Jäger, Andreas A1 - Wegener, Sonja A1 - Sauer, Otto A. T1 - Dose rate correction for a silicon diode detector array JF - Journal of Applied Clinical Medical Physics N2 - Purpose A signal dependence on dose rate was reported for the ArcCHECK array due to recombination processes within the diodes. The purpose of our work was to quantify the necessary correction and apply them to quality assurance measurements. Methods Static 10 × 10 cm\(^2\) 6-MV fields delivered by a linear accelerator were applied to the detector array while decreasing the average dose rate, that is, the pulse frequency, from 500 to 30 MU/min. An ion chamber was placed inside the ArcCHECK cavity as a reference. Furthermore, the instantaneous dose rate dependence (DRD) was studied. The position of the detector was adjusted to change the dose-per-pulse, varying the distance between the focus and the diode closest to the focus between 69.6 and 359.6 cm. Reference measurements were performed with an ion chamber placed inside a PMMA slab phantom at the same source-to-detector distances (SDDs). Exponential saturation functions were fitted to the data, with different parameters to account for two generations of ArcCHECK detectors (types 2 and 3) and both DRDs. Corrections were applied to 12 volumetric modulated arc therapy plans. Results The sensitivity decreased by up to 2.8% with a decrease in average dose rate and by 9% with a decrease in instantaneous dose rate. Correcting the average DRD, the mean gamma pass rates (2%/2-mm criterion) of the treatment plans were improved by 5 percentage points (PP) for diode type 3 and 0.4 PP for type 2. Correcting the instantaneous DRD, the improvement was 8.4 PP for type 3 and 0.9 PP for type 2. Conclusions The instantaneous DRD was identified as the prevailing effect on the diode sensitivity. We developed and validated a method to correct this behavior. The number of falsely not passed treatment plans could be considerably reduced. KW - ArcCHECK KW - correction KW - diode KW - dose rate KW - dosimetry, QA Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260446 VL - 22 IS - 10 ER - TY - JOUR A1 - Richter, Anne A1 - Wegener, Sonja A1 - Breuer, Kathrin A1 - Razinskas, Gary A1 - Weick, Stefan A1 - Exner, Florian A1 - Bratengeier, Klaus A1 - Flentje, Michael A1 - Sauer, Otto A1 - Polat, Bülent T1 - Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems JF - Radiation Oncology N2 - Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. Results The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. Conclusions For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system. KW - whole breast irradiation KW - Halcyon KW - IGRT KW - dose to OARs Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-265704 VL - 16 ER -