TY - JOUR A1 - Richter, Anne A1 - Polat, Bülent A1 - Lawrenz, Ingulf A1 - Weick, Stefan A1 - Sauer, Otto A1 - Flentje, Michael A1 - Mantel, Frederick T1 - Initial results for patient setup verification using transperineal ultrasound and cone beam CT in external beam radiation therapy of prostate cancer JF - Radiation Oncology N2 - Evaluation of set up error detection by a transperineal ultrasound in comparison with a cone beam CT (CBCT) based system in external beam radiation therapy (EBRT) of prostate cancer. Methods: Setup verification was performed with transperineal ultrasound (TPUS) and CBCT for 10 patients treated with EBRT for prostate cancer. In total, 150 ultrasound and CBCT scans were acquired in rapid succession and analyzed for setup errors. The deviation between setup errors of the two modalities was evaluated separately for each dimension. Results: A moderate correlation in lateral, vertical and longitudinal direction was observed comparing the setup errors. Mean differences between TPUS and CBCT were (−2.7 ± 2.3) mm, (3.0 ± 2.4) mm and (3.2 ± 2.7) mm in lateral, vertical and longitudinal direction, respectively. The mean Euclidean difference between TPUS and CBCT was (6.0 ± 3.1) mm. Differences up to 19.2 mm were observed between the two imaging modalities. Discrepancies between TPUS and CBCT of at least 5 mm occurred in 58 % of monitored treatment sessions. Conclusion: Setup differences between TPUS and CBCT are 6 mm on average. Although the correlation of the setup errors determined by the two different image modalities is rather week, the combination of setup verification by CBCT and intrafraction motion monitoring by TPUS imaging can use the benefits of both imaging modalities. KW - prostate cancer KW - transperineal ultrasound KW - IGRT KW - setup verification KW - cone beam CT Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-147677 VL - 11 IS - 147 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Lutyj, Paul A1 - Kraft, Johannes A1 - Lisowski, Dominik A1 - Weick, Stefan A1 - Flentje, Michael A1 - Polat, Bülent T1 - Two-Weekly High-Dose-Rate Brachytherapy Boost After External Beam Radiotherapy for Localized Prostate Cancer: Long-Term Outcome and Toxicity Analysis JF - Frontiers in Oncology N2 - Purpose Evaluation of clinical outcome of two-weekly high-dose-rate brachytherapy boost after external beam radiotherapy (EBRT) for localized prostate cancer. Methods 338 patients with localized prostate cancer receiving definitive EBRT followed by a two-weekly high-dose-rate brachytherapy boost (HDR-BT boost) in the period of 2002 to 2019 were analyzed. EBRT, delivered in 46 Gy (DMean) in conventional fractionation, was followed by two fractions HDR-BT boost with 9 Gy (D90%) two and four weeks after EBRT. Androgen deprivation therapy (ADT) was added in 176 (52.1%) patients. Genitourinary (GU)/gastrointestinal (GI) toxicity was evaluated utilizing the Common Toxicity Criteria for Adverse Events (version 5.0) and biochemical failure was defined according to the Phoenix definition. Results Median follow-up was 101.8 months. 15 (4.4%)/115 (34.0%)/208 (61.5%) patients had low-/intermediate-/high-risk cancer according to the D`Amico risk classification. Estimated 5-year and 10-year biochemical relapse-free survival (bRFS) was 84.7% and 75.9% for all patients. The estimated 5-year bRFS was 93.3%, 93.4% and 79.5% for low-, intermediate- and high-risk disease, respectively. The estimated 10-year freedom from distant metastasis (FFM) and overall survival (OS) rates were 86.5% and 70.0%. Cumulative 5-year late GU toxicity and late GI toxicity grade ≥ 2 was observed in 19.3% and 5.0% of the patients, respectively. Cumulative 5-year late grade 3 GU/GI toxicity occurred in 3.6%/0.3%. Conclusions Two-weekly HDR-BT boost after EBRT for localized prostate cancer showed an excellent toxicity profile with low GU/GI toxicity rates and effective long-term biochemical control. KW - prostate cancer KW - high-dose-rate (HDR) brachytherapy KW - radiotherapy KW - long-term outcome KW - toxicity KW - external beam radiotherapy (EBRT) KW - biochemical relapse free survival Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-250992 SN - 2234-943X VL - 11 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Razinskas, Gary A1 - Exner, Florian A1 - Richter, Anne A1 - Kessler, Patrick A1 - Weick, Stefan A1 - Kraft, Johannes A1 - Mantel, Frederick A1 - Flentje, Michael A1 - Polat, Bülent T1 - Comparison of treatment plans for hypofractionated high-dose prostate cancer radiotherapy using the Varian Halcyon and the Elekta Synergy platforms JF - Journal of Applied Clinical Medical Physics N2 - Purpose To compare radiotherapy plans between an O-ring and a conventional C-arm linac for hypofractionated high-dose prostate radiotherapy in terms of plan quality, dose distribution, and quality assurance in a multi-vendor environment. Methods Twenty prostate cancer treatment plans were irradiated on the O-ring Varian Halcyon linac and were re-optimized for the C-arm Elekta Synergy Agility linac. Dose-volume histogram metrics for target coverage and organ at risk dose, quality assurance, and monitor units were retrospectively compared. Patient-specific quality assurance with ion chamber measurements, gamma index analysis, and portal dosimetry was performed using the Varian Portal Dosimetry system and the ArcCHECK® phantom (Sun Nuclear Corporation). Prostate-only radiotherapy was delivered with simultaneous integrated boost (SIB) volumetric modulated arc therapy (VMAT) in 20 fractions of 2.5/3.0 Gy each. Results For both linacs, target coverage was excellent and plan quality comparable. Homogeneity in PTVBoost was high for Synergy as well as Halcyon with a mean homogeneity index of 0.07 ± 0.01 and 0.05 ± 0.01, respectively. Mean dose for the organs at risk rectum and bladder differed not significantly between the linacs but were higher for the femoral heads and penile bulb for Halcyon. Quality assurance showed no significant differences in terms of ArcCHECK gamma pass rates. Median pass rate for 3%/2 mm was 99.3% (96.7 to 99.8%) for Synergy and 99.8% (95.6 to 100%) for Halcyon. Agreement between calculated and measured dose was high with a median deviation of −0.6% (−1.7 to 0.8%) for Synergy and 0.2% (−0.6 to 2.3%) for Halcyon. Monitor units were higher for the Halcyon by approximately 20% (p < 0.001). Conclusion Hypofractionated high-dose prostate cancer SIB VMAT on the Halcyon system is feasible with comparable plan quality in reference to a standard C-arm Elekta Synergy linac. KW - acute toxicity KW - dose evaluation KW - Halcyon KW - hypofractionation KW - prostate cancer KW - Synergy Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260722 VL - 22 IS - 9 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Cirsi, Sinan A1 - Kessler, Patrick A1 - Razinskas, Gary A1 - Exner, Florian A1 - Richter, Anne A1 - Polat, Bülent A1 - Flentje, Michael T1 - Cone beam CT-based dose accumulation and analysis of delivered dose to the dominant intraprostatic lesion in primary radiotherapy of prostate cancer JF - Radiation Oncology N2 - Background Evaluation of delivered dose to the dominant intraprostatic lesion (DIL) for moderately hypofractionated radiotherapy of prostate cancer by cone beam computed tomography (CBCT)-based dose accumulation and target coverage analysis. Methods Twenty-three patients with localized prostate cancer treated with moderately hypofractionated prostate radiotherapy with simultaneous integrated boost (SIB) between December 2016 and February 2020 were retrospectively analyzed. Included patients were required to have an identifiable DIL on bi-parametric planning magnetic resonance imaging (MRI). After import into the RayStation treatment planning system and application of a step-wise density override, the fractional doses were computed on each CBCT and were consecutively mapped onto the planning CT via a deformation vector field derived from deformable image registration. Fractional doses were accumulated for all CBCTs and interpolated for missing CBCTs, resulting in the delivered dose for PTV\(_{DIL}\), PTV\(_{Boost}\), PTV, and the organs at risk. The location of the index lesions was recorded according to the sector map of the Prostate Imaging Reporting and Data System (PIRADS) Version 2.1. Target coverage of the index lesions was evaluated and stratified for location. Results In total, 338 CBCTs were available for analysis. Dose accumulation target coverage of PTV\(_{DIL}\), PTV\(_{Boost}\), and PTV was excellent and no cases of underdosage in D\(_{Mean}\), D_95%, D_02%, and D_98% could be detected. Delivered rectum D\(_{Mean}\) did not significantly differ from the planned dose. Bladder mean DMean was higher than planned with 19.4 ± 7.4 Gy versus 18.8 ± 7.5 Gy, p < 0.001. The penile bulb showed a decreased delivered mean DMean with 29.1 ± 14.0 Gy versus 29.8 ± 14.4 Gy, p < 0.001. Dorsal DILs, defined as DILs in the posterior medial peripheral zone of the prostate, showed a significantly lower delivered dose with a mean DMean difference of 2.2 Gy (95% CI 1.3–3.1 Gy, p < 0.001) compared to ventral lesions. Conclusions CBCT-based dose accumulation showed an adequate delivered dose to the dominant intraprostatic lesion and organs at risk within planning limits. Cautious evaluation of the target coverage for index lesions adjacent to the rectum is warranted to avoid underdosage. KW - adaptive radiotherapy KW - deformable image registration KW - dominant intraprostatic lesion KW - dose accumulation KW - prostate cancer KW - prostate Imaging Reporting and Data System Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-265656 VL - 16 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Zehner, Leonie A1 - Hartrampf, Philipp E. A1 - Cirsi, Sinan A1 - Wegener, Sonja A1 - Buck, Andreas K. A1 - Flentje, Michael A1 - Polat, Bülent T1 - Dose-escalated salvage radiotherapy for macroscopic local recurrence of prostate cancer in the prostate-specific membrane antigen positron emission tomography era JF - Cancers N2 - Simple Summary Prostate cancer often relapses after initial radical prostatectomy, and salvage radiotherapy offers a second chance of cure for relapsed patients. Modern imaging techniques, especially prostate-specific membrane antigen positron emission tomography/computed tomography (PSMA PET/CT), enable radiation oncologists to target radiotherapy at the involved sites of disease. In a group of patients, PSMA PET/CT imaging can detect a macroscopic local recurrence with or without locoregional lymph node metastasis. In these cases, an escalation of the radiotherapy dose is often considered for controlling the visible tumor mass. As the evidence for dose-escalated salvage radiotherapy for macroscopic recurrent prostate cancer after PSMA PET/CT imaging is still limited, we address this topic in the current analysis. We found that the outcome of patients with dose-escalated salvage radiotherapy for macroscopic prostate cancer recurrence is encouragingly favorable, while the toxicity is very limited. Abstract Background: The purpose of this study was to access the oncological outcome of prostate-specific membrane antigen positron emission tomography (PSMA PET/CT)-guided salvage radiotherapy (SRT) for localized macroscopic prostate cancer recurrence. Methods: Between February 2010 and June 2021, 367 patients received SRT after radical prostatectomy. Out of the 367 screened patients, 111 patients were staged by PSMA PET/CT before SRT. A total of 59 out of these 111 (53.2%) patients were treated for PSMA PET-positive macroscopic prostatic fossa recurrence. Dose-escalated SRT was applied with a simultaneous integrated boost at a median prescribed dose of 69.3 Gy (IQR 69.3–72.6 Gy). The oncological outcome was investigated using Kaplan-Meier and Cox regression analyses. The genitourinary (GU)/gastrointestinal (GI) toxicity evaluation utilized Common Toxicity Criteria for Adverse Events (version 5.0). Results: The median follow-up was 38.2 months. The three-year biochemical progression-free survival rate was 89.1% (95% CI: 81.1–97.8%) and the three-year metastasis-free survival rate reached 96.2% (95% CI: 91.2–100.0%). The cumulative three-year late grade 3 GU toxicity rate was 3.4%. No late grade 3 GI toxicity occurred. Conclusions: Dose-escalated PSMA PET/CT-guided salvage radiotherapy for macroscopic prostatic fossa recurrence resulted in favorable survival and toxicity rates. KW - prostate cancer KW - salvage radiotherapy KW - macroscopic recurrence KW - PSMA PET/CT KW - simultaneous integrated boost Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-290302 SN - 2072-6694 VL - 14 IS - 19 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Zehner, Leonie A1 - Hartrampf, Philipp A1 - Lisowski, Dominik A1 - Kneitz, Susanne A1 - Cirsi, Sinan A1 - Razinskas, Gary A1 - Flentje, Michael A1 - Polat, Bülent T1 - Salvage nodal radiotherapy as metastasis-directed therapy for oligorecurrent prostate cancer detected by positron emission tomography shows favorable outcome in long-term follow-up JF - Cancers N2 - Simple Summary Patients, who suffer from oligorecurrent prostate cancer with limited nodal involvement, may be offered positron emission tomography (PET)-directed salvage nodal radiotherapy to delay disease progression. This current analysis aimed to access salvage radiotherapy for nodal oligorecurrent prostate cancer with simultaneous integrated boost to PET-involved lymph nodes as metastasis-directed therapy. A long-term oncological outcome was favorable after salvage nodal radiotherapy and severe toxicity rates were low. Androgen deprivation therapy plays a major role in recurrent prostate cancer management and demonstrates a positive influence on the rate of biochemical progression in patients receiving salvage nodal radiotherapy. The present long-term analysis may help clinicians identify patients who would benefit from salvage nodal radiotherapy and androgen deprivation therapy, as a multimodal treatment strategy for oligorecurrent prostate cancer. Abstract Background: The study aimed to access the long-term outcome of salvage nodal radiotherapy (SNRT) in oligorecurrent prostate cancer. Methods: A total of 95 consecutive patients received SNRT for pelvic and/or extrapelvic nodal recurrence after prostate-specific membrane antigen (PSMA) or choline PET from 2010 to 2021. SNRT was applied as external beam radiotherapy with simultaneous integrated boost up to a median total dose of 62.9 Gy (EQD2\(_{1.5Gy}\)) to the recurrent lymph node metastases. The outcome was analyzed by cumulative incidence functions with death as the competing risk. Fine–Gray regression analyses were performed to estimate the relative hazards of the outcome parameters. Genitourinary (GU)/gastrointestinal (GI) toxicity evaluation utilized Common Toxicity Criteria for Adverse Events (v5.0). The results are as follows: the median follow-up was 47.1 months. The five-year biochemical progression rate (95% CI) was 50.1% (35.7–62.9%). Concomitant androgen deprivation therapy (ADT) was adminstered in 60.0% of the patients. The five-year biochemical progression rate was 75.0% (42.0–90.9%) without ADT versus 35.3% (19.6–51.4%) with ADT (p = 0.003). The cumulative five-year late grade 3 GU toxicity rate was 2.1%. No late grade 3 GI toxicity occured. Conclusions: Metastasis-directed therapy through SNRT for PET-staged oligorecurrent prostate cancer demonstrated a favorable long-term oncologic outcome. Omittance of ADT led to an increased biochemical progression. KW - metastasis-directed therapy KW - long-term outcome KW - oligorecurrence KW - prostate cancer KW - salvage radiotherapy KW - PSMA Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-286064 SN - 2072-6694 VL - 14 IS - 15 ER -