TY - JOUR A1 - Li, Mengxia A1 - Winkler, Beate A1 - Pabst, Thomas A1 - Bley, Thorsten A1 - Köstler, Herbert A1 - Neubauer, Henning T1 - Fast MR Imaging of the paediatric abdomen with CAIPIRINHA-accelerated T1w 3D FLASH and with high-resolution T2w HASTE: a study on image quality JF - Gastroenterology Research and Practice N2 - The aim of this study was to explore the applicability of fast MR techniques to routine paediatric abdominopelvic MRI at 1.5 Tesla. "Controlled Aliasing in Parallel Imaging Results in Higher Acceleration-" (CAIPIRINHA-) accelerated contrast-enhanced-T1w 3D FLASH imaging was compared to standard T1w 2D FLASH imaging with breath-holding in 40 paediatric patients and to respiratory-triggered T1w TSE imaging in 10 sedated young children. In 20 nonsedated patients, we compared T2w TIRM to fat-saturated T2w HASTE imaging. Two observers performed an independent and blinded assessment of overall image quality. Acquisition time was reduced by the factor of 15 with CAIPIRINHA-accelerated T1w FLASH and by 7 with T2w HASTE. With CAIPIRINHA and with HASTE, there were significantly less motion artefacts in nonsedated patients. In sedated patients, respiratory-triggered T1w imaging in general showed better image quality. However, satisfactory image quality was achieved with CAIPIRINHA in two sedated patients where respiratory triggering failed. In summary, fast scanning with CAIPIRINHA and HASTE presents a reliable high quality alternative to standard sequences in paediatric abdominal MRI. Paediatric patients, in particular, benefit greatly from fast image acquisition with less breath-hold cycles or shorter sedation. KW - gadoxetic acid KW - breath-hold KW - standard KW - Dixon KW - VIBE KW - clinical application Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-149217 VL - 2015 IS - 693654 ER - TY - JOUR A1 - Petritsch, Bernhard A1 - Köstler, Herbert A1 - Gassenmaier, Tobias A1 - Kunz, Andreas S A1 - Bley, Thorsten A A1 - Horn, Michael T1 - An investigation into potential gender-specific differences in myocardial triglyceride content assessed by \(^{1}\)H-Magnetic Resonance Spectroscopy at 3Tesla JF - Journal of International Medical Research N2 - Objective: Over the past decade, myocardial triglyceride content has become an accepted biomarker for chronic metabolic and cardiac disease. The purpose of this study was to use proton (hydrogen 1)-magnetic resonance spectroscopy (\(^{1}\)H-MRS) at 3Tesla (3 T) field strength to assess potential gender-related differences in myocardial triglyceride content in healthy individuals. Methods: Cardiac MR imaging was performed to enable accurate voxel placement and obtain functional and morphological information. Double triggered (i.e., ECG and respiratory motion gating) \(^{1}\)H-MRS was used to quantify myocardial triglyceride levels for each gender. Two-sample t-test and Mann-Whitney U-test were used for statistical analyses. Results: In total, 40 healthy volunteers (22 male, 18 female; aged >18 years and age matched) were included in the study. Median myocardial triglyceride content was 0.28% (interquartile range [IQR] 0.17–0.42%) in male and 0.24% (IQR 0.14–0.45%) in female participants, and no statistically significant difference was observed between the genders. Furthermore, no gender-specific difference in ejection fraction was observed, although on average, male participants presented with a higher mean ± SD left ventricular mass (136.3 ± 25.2 g) than female participants (103.9 ± 16.1 g). Conclusions: The study showed that \(^{1}\)H-MRS is a capable, noninvasive tool for acquisition of myocardial triglyceride metabolites. Myocardial triglyceride concentration was shown to be unrelated to gender in this group of healthy volunteers. KW - cardiac KW - magnetic resonance imaging KW - 1H-Magnetic resonance spectroscopy (1H-MRS) KW - myocardium KW - triglycerides KW - metabolism KW - gender Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-168808 VL - 44 IS - 3 ER - TY - JOUR A1 - Bluemel, Christina A1 - Linke, Fraenze A1 - Herrmann, Ken A1 - Simunovic, Iva A1 - Eiber, Matthias A1 - Kestler, Christian A1 - Buck, Andreas K. A1 - Schirbel, Andreas A1 - Bley, Thorsten A. A1 - Wester, Hans-Juergen A1 - Vergho, Daniel A1 - Becker, Axel T1 - Impact of \(^{68}\)Ga-PSMA PET/CT on salvage radiotherapy planning in patients with prostate cancer and persisting PSA values or biochemical relapse after prostatectomy JF - EJNMMI Research N2 - Background Salvage radiotherapy (SRT) is clinically established in prostate cancer (PC) patients with PSA persistence or biochemical relapse (BCR) after prior radical surgery. PET/CT imaging prior to SRT may be performed to localize disease recurrence. The recently introduced \(^{68}\)Ga-PSMA outperforms other PET tracers for detection of recurrence and is therefore expected also to impact radiation planning. Forty-five patients with PSA persistence (16 pts) or BCR (29 pts) after prior prostatectomy, scheduled to undergo SRT of the prostate bed, underwent \(^{68}\)Ga-PSMA PET/CT. The median PSA level was 0.67 ng/ml. The impact of \(^{68}\)Ga-PSMA PET/CT on the treatment decision was assessed. Patients with oligometastatic (≤5 lesions) PC underwent radiotherapy (RT), with the extent of the RT area and dose escalation being based on PET positivity. Results Suspicious lesions were detected in 24/45 (53.3 %) patients. In 62.5 % of patients, lesions were only detected by 68Ga-PSMA PET. Treatment was changed in 19/45 (42.2 %) patients, e.g., extending SRT to metastases (9/19), administering dose escalation in patients with morphological local recurrence (6/19), or replacing SRT by systemic therapy (2/19). 38/45 (84.4 %) followed the treatment recommendation, with data on clinical follow-up being available in 21 patients treated with SRT. All but one showed biochemical response (mean PSA decline 78 ± 19 %) within a mean follow-up of 8.12 ± 5.23 months. Conclusions \(^{68}\)Ga-PSMA PET/CT impacts treatment planning in more than 40 % of patients scheduled to undergo SRT. Future prospective studies are needed to confirm this significant therapeutic impact on patients prior to SRT. KW - prostate cancer KW - salvage radiotherapy KW - PSMA KW - PET/CT KW - recurrence Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-147798 VL - 6 IS - 78 ER - TY - JOUR A1 - Sauer, Stephanie A1 - Goltz, Jan P. A1 - Gassenmaier, Tobias A1 - Kunz, Andreas S. A1 - Bley, Thorsten A. A1 - Klein, Detlef A1 - Petritsch, Bernhard T1 - Partial Segmental Thrombosis of the Corpus Cavernosum (PSTCC) diagnosed by contrast-enhanced ultrasound: a case report JF - BMC Urology N2 - Background Partial segmental thrombosis of the corpus cavernosum (PSTCC) is a rare disease predominantly occurring in young men. Cardinal symptoms are pain and perineal swelling. Although several risk factors are described in the literature, the exact etiology of penile thrombosis remains unclear in most cases. MRI or ultrasound (US) is usually used for diagnosing this condition. Case presentation We report a case of penile thrombosis after left-sided varicocele ligature in a young patient. The diagnosis was established using contrast-enhanced ultrasound (CEUS) and was confirmed by contrast-enhanced magnetic resonance imaging (ceMRI). Successful conservative treatment consisted of systemic anticoagulation using low molecular weight heparin and acetylsalicylic acid. Conclusion PSTCC is a rare condition in young men and appears with massive pain and perineal swelling. In case of suspected PSTCC utilization of CEUS may be of diagnostic benefit. KW - MRI KW - priapism KW - corpus cavernosum KW - penile thrombosis KW - contrast-enhanced ultrasound Y1 - 2014 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-126495 VL - 14 IS - 100 ER - TY - JOUR A1 - Tran-Gia, Johannes A1 - Wech, Tobias A1 - Bley, Thorsten A1 - Köstler, Herbert T1 - Model-Based Acceleration of Look-Locker T1 Mapping JF - PLoS One N2 - Mapping the longitudinal relaxation time \(T_1\) has widespread applications in clinical MRI as it promises a quantitative comparison of tissue properties across subjects and scanners. Due to the long scan times of conventional methods, however, the use of quantitative MRI in clinical routine is still very limited. In this work, an acceleration of Inversion-Recovery Look-Locker (IR-LL) \(T_1\) mapping is presented. A model-based algorithm is used to iteratively enforce an exponential relaxation model to a highly undersampled radially acquired IR-LL dataset obtained after the application of a single global inversion pulse. Using the proposed technique, a \(T_1\) map of a single slice with 1.6mm in-plane resolution and 4mm slice thickness can be reconstructed from data acquired in only 6s. A time-consuming segmented IR experiment was used as gold standard for \(T_1\) mapping in this work. In the subsequent validation study, the model-based reconstruction of a single-inversion IR-LL dataset exhibited a \(T_1\) difference of less than 2.6% compared to the segmented IR-LL reference in a phantom consisting of vials with \(T_1\) values between 200ms and 3000ms. In vivo, the \(T_1\) difference was smaller than 5.5% in WM and GM of seven healthy volunteers. Additionally, the \(T_1\) values are comparable to standard literature values. Despite the high acceleration, all model-based reconstructions were of a visual quality comparable to fully sampled references. Finally, the reproducibility of the \(T_1\) mapping method was demonstrated in repeated acquisitions. In conclusion, the presented approach represents a promising way for fast and accurate \(T_1\) mapping using radial IR-LL acquisitions without the need of any segmentation. KW - algorithms KW - cerebrospinal fluid KW - NMR relaxation KW - data acquisition KW - relaxation (physics) KW - relaxation time KW - central nervous system KW - magnetic resonance imaging Y1 - 2015 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-126436 VL - 10 IS - 4 ER - TY - JOUR A1 - Ness, Thomas A1 - Bley, Thorsten A. A1 - Schmidt, Wolfgang A. A1 - Lamprecht, Peter T1 - The Diagnosis and Treatment of Giant Cell Arteritis JF - Deutsches Ärzteblatt International N2 - Background: Giant cell arteritis (GCA) is the most common systemic vasculitis in persons aged 50 and above (incidence, 3.5 per 100 000 per year). It affects cranial arteries, the aorta, and arteries elsewhere in the body, e.g., in the limbs. Methods: We selectively review the pertinent literature, including guidelines and recommendations from Germany and abroad. Results: The typical symptoms of new-onset GCA are bi-temporal headaches, jaw claudiacation, scalp tenderness, visual disturbances, systemic symptoms such as fever and weight loss, and polymyalgia. The diagnostic assessment comprises laboratory testing (erythrocyte sedimentation rate, C-reactive protein), imaging studies (duplex sonography, high-resolution magnetic resonance imaging, positron-emission tomography), and temporal artery biopsy. The standard treatment is with corticosteroids (adverse effects: diabetes mellitus, osteoporosis, cataract, arterial hypertension). A meta-analysis of three randomized controlled trials led to a recommendation for treatment with methotrexate to lower the recurrence rate and spare steroids. Patients for whom methotrexate is contraindicated or who cannot tolerate the drug can be treated with azathioprine instead. Conclusion: Giant cell arteritis, if untreated, progresses to involve the aorta and its collateral branches, leading to various complications. Late diagnosis and treatment can have serious consequences, including irreversible loss of visual function. KW - systemic vasculitides KW - arteriitis temporal arteriitis KW - erythrocyte sedimentation-rate KW - complication aortic-aneurysm KW - large-vessel vasculitis KW - c-reactive protein KW - polymyalgia-rheomatica KW - corticosteroid treatment KW - double-blind KW - ocular manifestations Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-131676 VL - 110 IS - 21 ER - TY - JOUR A1 - Frey, Anna A1 - Gassenmaier, Tobias A1 - Hofmann, Ulrich A1 - Schmitt, Dominik A1 - Fette, Georg A1 - Marx, Almuth A1 - Heterich, Sabine A1 - Boivin-Jahns, Valérie A1 - Ertl, Georg A1 - Bley, Thorsten A1 - Frantz, Stefan A1 - Jahns, Roland A1 - Störk, Stefan T1 - Coagulation factor XIII activity predicts left ventricular remodelling after acute myocardial infarction JF - ESC Heart Failure N2 - Aims Acute myocardial infarction (MI) is the major cause of chronic heart failure. The activity of blood coagulation factor XIII (FXIIIa) plays an important role in rodents as a healing factor after MI, whereas its role in healing and remodelling processes in humans remains unclear. We prospectively evaluated the relevance of FXIIIa after acute MI as a potential early prognostic marker for adequate healing. Methods and results This monocentric prospective cohort study investigated cardiac remodelling in patients with ST-elevation MI and followed them up for 1 year. Serum FXIIIa was serially assessed during the first 9 days after MI and after 2, 6, and 12 months. Cardiac magnetic resonance imaging was performed within 4 days after MI (Scan 1), after 7 to 9 days (Scan 2), and after 12 months (Scan 3). The FXIII valine-to-leucine (V34L) single-nucleotide polymorphism rs5985 was genotyped. One hundred forty-six patients were investigated (mean age 58 ± 11 years, 13% women). Median FXIIIa was 118 % (quartiles, 102–132%) and dropped to a trough on the second day after MI: 109%(98–109%; P < 0.001). FXIIIa recovered slowly over time, reaching the baseline level after 2 to 6 months and surpassed baseline levels only after 12 months: 124 % (110–142%). The development of FXIIIa after MI was independent of the genotype. FXIIIa on Day 2 was strongly and inversely associated with the relative size of MI in Scan 1 (Spearman’s ρ = –0.31; P = 0.01) and Scan 3 (ρ = –0.39; P < 0.01) and positively associated with left ventricular ejection fraction: ρ = 0.32 (P < 0.01) and ρ = 0.24 (P = 0.04), respectively. Conclusions FXIII activity after MI is highly dynamic, exhibiting a significant decline in the early healing period, with reconstitution 6 months later. Depressed FXIIIa early after MI predicted a greater size of MI and lower left ventricular ejection fraction after 1 year. The clinical relevance of these findings awaits to be tested in a randomized trial. KW - blood coagulation factor XIII KW - ST-elevation myocardial infarction KW - healing and remodelling processes KW - cardiac magnetic resonance imaging Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-236013 VL - 7 IS - 5 ER - TY - JOUR A1 - Rasche, Leo A1 - Kumar, Manoj A1 - Gershner, Grant A1 - Samant, Rohan A1 - Van Hemert, Rudy A1 - Heidemeier, Anke A1 - Lapa, Constantin A1 - Bley, Thorsten A1 - Buck, Andreas A1 - McDonald, James A1 - Hillengass, Jens A1 - Epstein, Joshua A1 - Thanendrarajan, Sharmilan A1 - Schinke, Carolina A1 - van Rhee, Frits A1 - Zangari, Maurizio A1 - Barlogie, Bart A1 - Davies, Faith E. A1 - Morgan, Gareth J. A1 - Weinhold, Niels T1 - Lack of Spleen Signal on Diffusion Weighted MRI is associated with High Tumor Burden and Poor Prognosis in Multiple Myeloma: A Link to Extramedullary Hematopoiesis? JF - Theranostics N2 - Due to the low frequency of abnormalities affecting the spleen, this organ is often overlooked during radiological examinations. Here, we report on the unexpected finding, that the spleen signal on diffusion-weighted MRI (DW-MRI) is associated with clinical parameters in patients with plasma cell dyscrasias. Methods: We investigated the spleen signal on DW-MRI together with clinical and molecular parameters in 295 transplant-eligible newly diagnosed Multiple Myeloma (NDMM) patients and in 72 cases with monoclonal gammopathy of undetermined significance (MGUS). Results: Usually, the spleen is the abdominal organ with the highest intensities on DW-MRI. Yet, significant signal loss on DW-MRI images was seen in 71 of 295 (24%) NDMM patients. This phenomenon was associated with the level of bone marrow plasmacytosis (P=1x10(-10)) and International Staging System 3 (P=0.0001) but not with gain(1q), and del(17p) or plasma cell gene signatures. The signal was preserved in 72 individuals with monoclonal gammopathy of undetermined significance and generally re-appeared in MM patients responding to treatment, suggesting that lack of signal reflects increased tumor burden. While absence of spleen signal in MM patients with high risk disease defined a subgroup with very poor outcome, re-appearance of the spleen signal after autologous stem cell transplantation was seen in patients with improved outcome. Our preliminary observation suggests that extramedullary hematopoiesis in the spleen is a factor that modifies the DW-MRI signal of this organ. Conclusions: The DW-MRI spleen signal is a promising marker for tumor load and provides prognostic information in MM. KW - multiple myeloma KW - diffusion weighted mri KW - spleen KW - tumor burden KW - high risk KW - extramedullary hematopoiesis Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-224982 VL - 9 IS - 16 ER - TY - JOUR A1 - Guggenberger, Konstanze V. A1 - Bley, Thorsten A. A1 - Vogt, Marius L. A1 - Urbach, Horst A1 - Meckel, Stephan T1 - High-Resolution Black Blood Vessel Wall Imaging in COVID-19 Encephalopathy-Is it Really Endotheliitis? JF - Clinical Neuroradiology N2 - No abstract available. KW - Covid-19 KW - encephalopathy KW - endotheliitis Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-264794 VL - 32 IS - 1 ER - TY - INPR A1 - Heidenreich, Julius F. A1 - Gassenmaier, Tobias A1 - Ankenbrand, Markus J. A1 - Bley, Thorsten A. A1 - Wech, Tobias T1 - Self-configuring nnU-net pipeline enables fully automatic infarct segmentation in late enhancement MRI after myocardial infarction N2 - Purpose To fully automatically derive quantitative parameters from late gadolinium enhancement (LGE) cardiac MR (CMR) in patients with myocardial infarction and to investigate if phase sensitive or magnitude reconstructions or a combination of both results in best segmentation accuracy. Methods In this retrospective single center study, a convolutional neural network with a U-Net architecture with a self-configuring framework (“nnU-net”) was trained for segmentation of left ventricular myocardium and infarct zone in LGE-CMR. A database of 170 examinations from 78 patients with history of myocardial infarction was assembled. Separate fitting of the model was performed, using phase sensitive inversion recovery, the magnitude reconstruction or both contrasts as input channels. Manual labelling served as ground truth. In a subset of 10 patients, the performance of the trained models was evaluated and quantitatively compared by determination of the Sørensen-Dice similarity coefficient (DSC) and volumes of the infarct zone compared with the manual ground truth using Pearson’s r correlation and Bland-Altman analysis. Results The model achieved high similarity coefficients for myocardium and scar tissue. No significant difference was observed between using PSIR, magnitude reconstruction or both contrasts as input (PSIR and MAG; mean DSC: 0.83 ± 0.03 for myocardium and 0.72 ± 0.08 for scars). A strong correlation for volumes of infarct zone was observed between manual and model-based approach (r = 0.96), with a significant underestimation of the volumes obtained from the neural network. Conclusion The self-configuring nnU-net achieves predictions with strong agreement compared to manual segmentation, proving the potential as a promising tool to provide fully automatic quantitative evaluation of LGE-CMR. KW - Deep learning KW - CMR KW - Segmentation KW - Myocardial infarction KW - Scar KW - nnU-net Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-323418 UR - https://doi.org/10.1016/j.ejrad.2021.109817 ET - accepted version ER - TY - JOUR A1 - Grunz, Jan-Peter A1 - Pennig, Lenhard A1 - Fieber, Tabea A1 - Gietzen, Carsten Herbert A1 - Heidenreich, Julius Frederik A1 - Huflage, Henner A1 - Gruschwitz, Philipp A1 - Kuhl, Philipp Josef A1 - Petritsch, Bernhard A1 - Kosmala, Aleksander A1 - Bley, Thorsten Alexander A1 - Gassenmaier, Tobias T1 - Twin robotic x-ray system in small bone and joint trauma: Impact of cone-beam computed tomography on treatment decisions JF - European Radiology N2 - Objectives Trauma evaluation of extremities can be challenging in conventional radiography. A multi-use x-ray system with cone-beam CT (CBCT) option facilitates ancillary 3-D imaging without repositioning. We assessed the clinical value of CBCT scans by analyzing the influence of additional findings on therapy. Methods Ninety-two patients underwent radiography and subsequent CBCT imaging with the twin robotic scanner (76 wrist/hand/finger and 16 ankle/foot/toe trauma scans). Reports by on-call radiologists before and after CBCT were compared regarding fracture detection, joint affliction, comminuted injuries, and diagnostic confidence. An orthopedic surgeon recommended therapy based on reported findings. Surgical reports (N = 52) and clinical follow-up (N = 85) were used as reference standard. Results CBCT detected more fractures (83/64 of 85), joint involvements (69/53 of 71), and multi-fragment situations (68/50 of 70) than radiography (all p < 0.001). Six fractures suspected in radiographs were ruled out by CBCT. Treatment changes based on additional information from CBCT were recommended in 29 patients (31.5%). While agreement between advised therapy before CBCT and actual treatment was moderate (κ = 0.41 [95% confidence interval 0.35–0.47]; p < 0.001), agreement after CBCT was almost perfect (κ = 0.88 [0.83–0.93]; p < 0.001). Diagnostic confidence increased considerably for CBCT studies (p < 0.001). Median effective dose for CBCT was 4.3 μSv [3.3–5.3 μSv] compared to 0.2 μSv [0.1–0.2 μSv] for radiography. Conclusions CBCT provides advantages for the evaluation of acute small bone and joint trauma by detecting and excluding extremity fractures and fracture-related findings more reliably than radiographs. Additional findings induced therapy change in one third of patients, suggesting substantial clinical impact. KW - cone-beamcomputed tomography KW - extremities KW - fractures, bone KW - radiography Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-235233 SN - 0938-7994 VL - 31 ER - TY - JOUR A1 - Guggenberger, Konstanze Viktoria A1 - Torre, Giulia Dalla A1 - Ludwig, Ute A1 - Vogel, Patrick A1 - Weng, Andreas Max A1 - Vogt, Marius Lothar A1 - Fröhlich, Matthias A1 - Schmalzing, Marc A1 - Raithel, Esther A1 - Forman, Christoph A1 - Urbach, Horst A1 - Meckel, Stephan A1 - Bley, Thorsten Alexander T1 - Vasa vasorum of proximal cerebral arteries after dural crossing - potential imaging confounder in diagnosing intracranial vasculitis in elderly subjects on black-blood MRI JF - European Radiology N2 - Objectives Vessel wall enhancement (VWE) may be commonly seen on MRI images of asymptomatic subjects. This study aimed to characterize the VWE of the proximal internal carotid (ICA) and vertebral arteries (VA) in a non-vasculitic elderly patient cohort. Methods Cranial MRI scans at 3 Tesla were performed in 43 patients (aged ≥ 50 years) with known malignancy for exclusion of cerebral metastases. For vessel wall imaging (VWI), a high-resolution compressed-sensing black-blood 3D T1-weighted fast (turbo) spin echo sequence (T1 CS-SPACE prototype) was applied post gadolinium with an isotropic resolution of 0.55 mm. Bilateral proximal intradural ICA and VA segments were evaluated for presence, morphology, and longitudinal extension of VWE. Results Concentric VWE of the proximal intradural ICA was found in 13 (30%) patients, and of the proximal intradural VA in 39 (91%) patients. Mean longitudinal extension of VWE after dural entry was 13 mm in the VA and 2 mm in the ICA. In 14 of 39 patients (36%) with proximal intradural VWE, morphology of VWE was suggestive of the mere presence of vasa vasorum. In 25 patients (64 %), morphology indicated atherosclerotic lesions in addition to vasa vasorum. Conclusions Vasa vasorum may account for concentric VWE within the proximal 2 mm of the ICA and 13 mm of the VA after dural entry in elderly subjects. Concentric VWE in these locations should not be confused with large artery vasculitis. Distal to these segments, VWE may be more likely related to pathologic conditions such as vasculitis. KW - vertebral artery KW - magnetic resonance imaging KW - vasa vasorum KW - large artery vasculitis KW - Atherosclerosis, intracranial arteries Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-266524 SN - 1432-1084 VL - 32 IS - 2 ER - TY - JOUR A1 - Wech, Tobias A1 - Ankenbrand, Markus Johannes A1 - Bley, Thorsten Alexander A1 - Heidenreich, Julius Frederik T1 - A data-driven semantic segmentation model for direct cardiac functional analysis based on undersampled radial MR cine series JF - Magnetic Resonance in Medicine N2 - Purpose Image acquisition and subsequent manual analysis of cardiac cine MRI is time-consuming. The purpose of this study was to train and evaluate a 3D artificial neural network for semantic segmentation of radially undersampled cardiac MRI to accelerate both scan time and postprocessing. Methods A database of Cartesian short-axis MR images of the heart (148,500 images, 484 examinations) was assembled from an openly accessible database and radial undersampling was simulated. A 3D U-Net architecture was pretrained for segmentation of undersampled spatiotemporal cine MRI. Transfer learning was then performed using samples from a second database, comprising 108 non-Cartesian radial cine series of the midventricular myocardium to optimize the performance for authentic data. The performance was evaluated for different levels of undersampling by the Dice similarity coefficient (DSC) with respect to reference labels, as well as by deriving ventricular volumes and myocardial masses. Results Without transfer learning, the pretrained model performed moderately on true radial data [maximum number of projections tested, P = 196; DSC = 0.87 (left ventricle), DSC = 0.76 (myocardium), and DSC =0.64 (right ventricle)]. After transfer learning with authentic data, the predictions achieved human level even for high undersampling rates (P = 33, DSC = 0.95, 0.87, and 0.93) without significant difference compared with segmentations derived from fully sampled data. Conclusion A 3D U-Net architecture can be used for semantic segmentation of radially undersampled cine acquisitions, achieving a performance comparable with human experts in fully sampled data. This approach can jointly accelerate time-consuming cine image acquisition and cumbersome manual image analysis. KW - undersampling KW - cardiovascular magnetic resonance (CMR) KW - deep learning KW - radial KW - semantic segmentation Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-257616 VL - 87 IS - 2 ER - TY - JOUR A1 - Huflage, Henner A1 - Fieber, Tabea A1 - Färber, Christian A1 - Knarr, Jonas A1 - Veldhoen, Simon A1 - Jordan, Martin C. A1 - Gilbert, Fabian A1 - Bley, Thorsten Alexander A1 - Meffert, Rainer H. A1 - Grunz, Jan-Peter A1 - Schmalzl, Jonas T1 - Interobserver reliability of scapula fracture classifications in intra- and extra-articular injury patterns JF - BMC Musculoskeletal Disorders N2 - Background Morphology and glenoid involvement determine the necessity of surgical management in scapula fractures. While being present in only a small share of patients with shoulder trauma, numerous classification systems have been in use over the years for categorization of scapula fractures. The purpose of this study was to evaluate the established AO/OTA classification in comparison to the classification system of Euler and Rüedi (ER) with regard to interobserver reliability and confidence in clinical practice. Methods Based on CT imaging, 149 patients with scapula fractures were retrospectively categorized by two trauma surgeons and two radiologists using the classification systems of ER and AO/OTA. To measure the interrater reliability, Fleiss kappa (κ) was calculated independently for both fracture classifications. Rater confidence was stated subjectively on a five-point scale and compared with Wilcoxon signed rank tests. Additionally, we computed the intraclass correlation coefficient (ICC) based on absolute agreement in a two-way random effects model to assess the diagnostic confidence agreement between observers. Results In scapula fractures involving the glenoid fossa, interrater reliability was substantial (κ = 0.722; 95% confidence interval [CI] 0.676–0.769) for the AO/OTA classification in contrast to moderate agreement (κ = 0.579; 95% CI 0.525–0.634) for the ER classification system. Diagnostic confidence for intra-articular fracture patterns was superior using the AO/OTA classification compared to ER (p < 0.001) with higher confidence agreement (ICC: 0.882 versus 0.831). For extra-articular fractures, ER (κ = 0.817; 95% CI 0.771–0.863) provided better interrater reliability compared to AO/OTA (κ = 0.734; 95% CI 0.692–0.776) with higher diagnostic confidence (p < 0.001) and superior agreement between confidence ratings (ICC: 0.881 versus 0.912). Conclusions The AO/OTA classification is most suitable to categorize intra-articular scapula fractures with glenoid involvement, whereas the classification system of Euler and Rüedi appears to be superior in extra-articular injury patterns with fractures involving only the scapula body, spine, acromion and coracoid process. KW - confidence KW - scapula KW - glenoid KW - fracture KW - classification KW - reliability Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-299795 VL - 23 IS - 1 ER - TY - JOUR A1 - Herz, Stefan A1 - Stefanescu, Maria R. A1 - Lohr, David A1 - Vogel, Patrick A1 - Kosmala, Aleksander A1 - Terekhov, Maxim A1 - Weng, Andreas M. A1 - Grunz, Jan-Peter A1 - Bley, Thorsten A. A1 - Schreiber, Laura M. T1 - Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission JF - PloS One N2 - Background To investigate the effects of B\(_1\)-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI). Methods Stenosis phantoms with different grades of stenosis (0%, 20%, 40%, 60%, 80%, and 100%; 5 mm inner vessel diameter) were produced using 3D printing (clear resin). Phantoms were imaged with four different concentrations of diluted Gd-DOTA representing established arterial concentrations after intravenous injection in humans. Samples were centrally positioned in a thorax phantom of 30 cm diameter filled with a custom-made liquid featuring dielectric properties of muscle tissue. MRI was performed on a 7 T whole-body system. 2D-gradient-echo sequences were acquired with an 8-channel transmit 16-channel receive (8 Tx / 16 Rx) cardiac array prototype coil with and without pTX mode. Measurements were compared to those obtained with identical scan parameters using a commercially available 1 Tx / 16 Rx single transmit coil (sTX). To assess reproducibility, measurements (n = 15) were repeated at different horizontal angles with respect to the B0-field. Results B\(_1\)-shimming and pTX markedly improved flip angle homogeneity across the thorax phantom yielding a distinctly increased signal-to-noise ratio (SNR) averaged over a whole slice relative to non-manipulated RF fields. Images without B\(_1\)-shimming showed shading artifacts due to local B\(_1\)\(^+\)-field inhomogeneities, which hampered stenosis quantification in severe cases. In contrast, B\(_1\)-shimming and pTX provided superior image homogeneity. Compared with a conventional sTX coil higher grade stenoses (60% and 80%) were graded significantly (p<0.01) more precise. Mild to moderate grade stenoses did not show significant differences. Overall, SNR was distinctly higher with B\(_1\)-shimming and pTX than with the conventional sTX coil (inside the stenosis phantoms 14%, outside the phantoms 32%). Both full and half concentration (10.2 mM and 5.1 mM) of a conventional Gd-DOTA dose for humans were equally suitable for stenosis evaluation in this phantom study. Conclusions B\(_1\)-shimming and pTX at 7 T can distinctly improve image homogeneity and therefore provide considerably more accurate MR image analysis, which is beneficial for imaging of small vessel structures. KW - stenosis KW - magnetic resonance imaging KW - thorax KW - in vivo imaging KW - coronary arteries KW - image processing KW - 3D printing KW - signal to noise ratio Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-300129 VL - 17 IS - 6 ER - TY - JOUR A1 - Woznicki, Piotr A1 - Laqua, Fabian A1 - Bley, Thorsten A1 - Baeßler, Bettina T1 - AutoRadiomics: a framework for reproducible radiomics research JF - Frontiers in Radiology N2 - Purpose Machine learning based on radiomics features has seen huge success in a variety of clinical applications. However, the need for standardization and reproducibility has been increasingly recognized as a necessary step for future clinical translation. We developed a novel, intuitive open-source framework to facilitate all data analysis steps of a radiomics workflow in an easy and reproducible manner and evaluated it by reproducing classification results in eight available open-source datasets from different clinical entities. Methods The framework performs image preprocessing, feature extraction, feature selection, modeling, and model evaluation, and can automatically choose the optimal parameters for a given task. All analysis steps can be reproduced with a web application, which offers an interactive user interface and does not require programming skills. We evaluated our method in seven different clinical applications using eight public datasets: six datasets from the recently published WORC database, and two prostate MRI datasets—Prostate MRI and Ultrasound With Pathology and Coordinates of Tracked Biopsy (Prostate-UCLA) and PROSTATEx. Results In the analyzed datasets, AutoRadiomics successfully created and optimized models using radiomics features. For WORC datasets, we achieved AUCs ranging from 0.56 for lung melanoma metastases detection to 0.93 for liposarcoma detection and thereby managed to replicate the previously reported results. No significant overfitting between training and test sets was observed. For the prostate cancer detection task, results were better in the PROSTATEx dataset (AUC = 0.73 for prostate and 0.72 for lesion mask) than in the Prostate-UCLA dataset (AUC 0.61 for prostate and 0.65 for lesion mask), with external validation results varying from AUC = 0.51 to AUC = 0.77. Conclusion AutoRadiomics is a robust tool for radiomic studies, which can be used as a comprehensive solution, one of the analysis steps, or an exploratory tool. Its wide applicability was confirmed by the results obtained in the diverse analyzed datasets. The framework, as well as code for this analysis, are publicly available under https://github.com/pwoznicki/AutoRadiomics. KW - radiomics KW - radiology KW - machine learning KW - reproducibility KW - workflow KW - image analysis Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-284813 SN - 2673-8740 VL - 2 ER - TY - JOUR A1 - Reichel, Thomas A1 - Herz, Stefan A1 - el Tabbakh, Mohammed A1 - Bley, Thorsten Alexander A1 - Plumhoff, Piet A1 - Rueckl, Kilian T1 - Less than 9.5-mm coracohumeral distance on axial magnetic resonance imaging scans predicts for subscapularis tear JF - JSES International N2 - Background Diagnosis of subscapularis (SSC) tendon lesions on magnetic resonance imaging (MRI) can be challenging. A small coracohumeral distance (CHD) has been associated with SSC tears. This study was designed to define a specific threshold value for CHD to predict SSC tears on axial MRI scans. Methods This retrospective study included 172 shoulders of 168 patients who underwent arthroscopic surgery for rotator cuff tear or glenohumeral instability. Diagnostic arthroscopy confirmed an SSC tear in 62 cases (36.0%, test group a), rotator cuff tear tears other than SSC in 71 cases (41.3%, control group b) and glenohumeral instability without any rotator cuff tear in 39 cases (22.7%, zero-sample group c). All patients had a preoperative MRI of the shoulder (1.5T or 3T). Minimum CHD was measured on axial fat-suppressed proton density-, T2-, or T1-weigthed sequences. Receiver operating characteristics analysis was used to determine the threshold value for CHD, and sensitivity and specificity were calculated. Results CHD measurement had a good interobserver reliability (Intraclass correlation coefficient 0.799). Mean CHD was highly significantly (P < .001) less for test group a (mean 7.3 mm, standard deviation ± 2.2) compared with control group b (mean 11.1 mm, standard deviation ± 2.3) or zero-sample group c (mean 13.6 mm, standard deviation ± 2.9). A threshold value of CHD <9.5 mm had a sensitivity of 83.6% and a specificity of 83.9% to predict SSC tears. Conclusion A CHD <9.5 mm on MRI is predictive of SSC lesions and a valuable tool to diagnose SSC tears. KW - subscapularis tear KW - coracohumeral distance KW - rotator cuff tear KW - coracohumeral impingement Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-259118 VL - 5 IS - 3 ER - TY - JOUR A1 - Grunz, Jan-Peter A1 - Gietzen, Carsten Herbert A1 - Luetkens, Karsten A1 - Wagner, Matthias A1 - Kalb, Karlheinz A1 - Bley, Thorsten Alexander A1 - Lehmkul, Luka A1 - van Schoonhoven, Jörg A1 - Gassenmaier, Tobias A1 - Schmitt, Rainer T1 - The importance of radial multiplanar reconstructions for assessment of triangular fibrocartilage complex injury in CT arthrography of the wrist JF - BMC Musculoskeletal Disorders N2 - Background: Triangular fibrocartilage complex (TFCC) lesions commonly cause ulnar-sided wrist pain and instability of the distal radioulnar joint. Due to its triangular shape, discontinuity of the TFCC is oftentimes difficult to visualize in radiological standard planes. Radial multiplanar reconstructions (MPR) may have the potential to simplify diagnosis in CT wrist arthrography. The objective of this study was to assess diagnostic advantages provided by radial MPR over standard planes for TFCC lesions in CT arthrography. Methods: One hundred six patients (49 women, 57 men; mean age 44.2 ± 15.8 years) underwent CT imaging after wrist arthrography. Two radiologists (R1, R2) retrospectively analyzed three randomized datasets for each CT arthrography. One set contained axial, coronal and sagittal planes (MPR\(_{Standard}\)), while the other two included an additional radial reconstruction with the rotating center either atop the ulnar styloid (MPR\(_{Styloid}\)) or in the ulnar fovea (MPR\(_{Fovea}\)). Readers evaluated TFCC differentiability and condition. Suspected lesions were categorized using Palmer’s and Atzei’s classification and diagnostic confidence was stated on a fivepoint Likert scale. Results: Compared to standard planes, differentiability of the superficial and deep TFCC layer was superior in radial reconstructions (R1/R2; MPR\(_{Fovea}\): p < 0.001; MPRStyloid: p ≤ 0.007). Palmer and Atzei lesions were present in 86.8% (92/106) and 52.8% (56/106) of patients, respectively. Specificity, sensitivity and accuracy for central Palmer lesions did not differ in radial and standard MPR. For peripheral Atzei lesions, sensitivity (MPR\(_{Standard}\) 78.6%/80.4%, MPR\(_{Styloid}\) 94.6%/94.6%, MPR\(_{Fovea}\) 91.1%/89.3%) and accuracy (MPR\(_{Standard}\) 86.8%/86.8%, MPR\(_{Styloid}\) 96.2%/96.2%, MPR\(_{Fovea}\) 94.3%/93.4%) improved with additional styloid-centered (p = 0.004/0.008) and foveacentered (p = 0.039/0.125) reconstructions. No substantial difference was observed between both radial MPR (p = 0.688/0.250). Interrater agreement was almost perfect for each dataset (κ\(_{Standard}\) = 0.876, κ\(_{Styloid}\) = 0.894, κ\(_{Fovea}\) = 0.949). Diagnostic confidence increased with addition of either radial MPR (p < 0.001). Conclusions: Ancillary radial planes improve accuracy and diagnostic confidence for detection of peripheral TFCC lesions in CT arthrography of the wrist. KW - Triangular fibrocartilage KW - Wrist KW - Arthrography KW - Tomography KW - X-ray computed Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-236075 VL - 21 ER - TY - JOUR A1 - Luetkens, Karsten Sebastian A1 - Ergün, Süleyman A1 - Huflage, Henner A1 - Kunz, Andreas Steven A1 - Gietzen, Carsten Herbert A1 - Conrads, Nora A1 - Pennig, Lenhard A1 - Goertz, Lukas A1 - Bley, Thorsten Alexander A1 - Gassenmaier, Tobias A1 - Grunz, Jan-Peter T1 - Dose reduction potential in cone-beam CT imaging of upper extremity joints with a twin robotic x-ray system JF - Scientific Reports N2 - Cone-beam computed tomography is a powerful tool for 3D imaging of the appendicular skeleton, facilitating detailed visualization of bone microarchitecture. This study evaluated various combinations of acquisition and reconstruction parameters for the cone-beam CT mode of a twin robotic x-ray system in cadaveric wrist and elbow scans, aiming to define the best possible trade-off between image quality and radiation dose. Images were acquired with different combinations of tube voltage and tube current–time product, resulting in five scan protocols with varying volume CT dose indices: full-dose (FD; 17.4 mGy), low-dose (LD; 4.5 mGy), ultra-low-dose (ULD; 1.15 mGy), modulated low-dose (mLD; 0.6 mGy) and modulated ultra-low-dose (mULD; 0.29 mGy). Each set of projection data was reconstructed with three convolution kernels (very sharp [Ur77], sharp [Br69], intermediate [Br62]). Five radiologists subjectively assessed the image quality of cortical bone, cancellous bone and soft tissue using seven-point scales. Irrespective of the reconstruction kernel, overall image quality of every FD, LD and ULD scan was deemed suitable for diagnostic use in contrast to mLD (very sharp/sharp/intermediate: 60/55/70%) and mULD (0/3/5%). Superior depiction of cortical and cancellous bone was achieved in FD\(_{Ur77}\) and LD\(_{Ur77}\) examinations (p < 0.001) with LD\(_{Ur77}\) scans also providing favorable bone visualization compared to FD\(_{Br69}\) and FD\(_{Br62}\) (p < 0.001). Fleiss’ kappa was 0.618 (0.594–0.641; p < 0.001), indicating substantial interrater reliability. In this study, we demonstrate that considerable dose reduction can be realized while maintaining diagnostic image quality in upper extremity joint scans with the cone-beam CT mode of a twin robotic x-ray system. Application of sharper convolution kernels for image reconstruction facilitates superior display of bone microarchitecture. KW - medical research KW - preclinical research Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-270429 VL - 11 IS - 1 ER - TY - JOUR A1 - Grunz, Jan-Peter A1 - Wenig, Andreas Max A1 - Kunz, Andreas Steven A1 - Veyhl-Wichmann, Maike A1 - Schmitt, Rainer A1 - Gietzen, Carsten Herbert A1 - Pennig, Lenhard A1 - Herz, Stefan A1 - Ergün, Süleyman A1 - Bley, Thorsten Alexander A1 - Gassenmaier, Tobias T1 - 3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT JF - European Radiology Experimental N2 - Background Elbow imaging is challenging with conventional multidetector computed tomography (MDCT), while cone-beam CT (CBCT) provides superior options. We compared intra-individually CBCT versus MDCT image quality in cadaveric elbows. Methods A twin robotic x-ray system with new CBCT mode and a high-resolution clinical MDCT were compared in 16 cadaveric elbows. Both systems were operated with a dedicated low-dose (LD) protocol (equivalent volume CT dose index [CTDI\(_{vol(16 cm)}\)] = 3.3 mGy) and a regular clinical scan dose (RD) protocol (CTDI\(_{vol(16 cm)}\) = 13.8 mGy). Image quality was evaluated by two radiologists (R1 and R2) on a seven-point Likert scale, and estimation of signal intensity in cancellous bone was conducted. Wilcoxon signed-rank tests and intraclass correlation coefficient (ICC) statistics were used. Results The CBCT prototype provided superior subjective image quality compared to MDCT scans (for RD, p ≤ 0.004; for LD, p ≤ 0.001). Image quality was rated very good or excellent in 100% of the cases by both readers for RD CBCT, 100% (R1) and 93.8% (R2) for LD CBCT, 62.6% and 43.8% for RD MDCT, and 0.0% and 0.0% for LD MDCT. Single-measure ICC was 0.95 (95% confidence interval 0.91–0.97; p < 0.001). Software-based assessment supported subjective findings with less “undecided” pixels in CBCT than dose-equivalent MDCT (p < 0.001). No significant difference was found between LD CBCT and RD MDCT. Conclusions In cadaveric elbow studies, the tested cone-beam CT prototype delivered superior image quality compared to high-end multidetector CT and showed a potential for considerable dose reduction. KW - Cancellous bone KW - Cone-beam computed tomography KW - Elbow KW - Elbow joint KW - Multidetector computed tomography Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-229877 VL - 4 ER - TY - JOUR A1 - Huflage, Henner A1 - Karsten, Sebastian A1 - Kunz, Andreas Steven A1 - Conrads, Nora A1 - Jakubietz, Rafael Gregor A1 - Jakubietz, Michael Georg A1 - Pennig, Lenhard A1 - Goertz, Lukas A1 - Bley, Thorsten Alexander A1 - Schmitt, Rainer A1 - Grunz, Jan-Peter T1 - Improved diagnostic accuracy for ulnar-sided TFCC lesions with radial reformation of 3D sequences in wrist MR arthrography JF - European Radiology N2 - Objectives Triangular fibrocartilage complex (TFCC) injuries frequently cause ulnar-sided wrist pain and can induce distal radioulnar joint instability. With its complex three-dimensional structure, diagnosis of TFCC lesions remains a challenging task even in MR arthrograms. The aim of this study was to assess the added diagnostic value of radial reformatting of isotropic 3D MRI datasets compared to standard planes after direct arthrography of the wrist. Methods Ninety-three patients underwent wrist MRI after fluoroscopy-guided multi-compartment arthrography. Two radiologists collectively analyzed two datasets of each MR arthrogram for TFCC injuries, with one set containing standard reconstructions of a 3D thin-slice sequence in axial, coronal and sagittal orientation, while the other set comprised an additional radial plane view with the rotating center positioned at the ulnar styloid. Surgical reports (whenever available) or radiological reports combined with clinical follow-up served as a standard of reference. In addition, diagnostic confidence and assessability of the central disc and ulnar-sided insertions were subjectively evaluated. Results Injuries of the articular disc, styloid and foveal ulnar attachment were present in 20 (23.7%), 10 (10.8%) and 9 (9.7%) patients. Additional radial planes increased diagnostic accuracy for lesions of the styloid (0.83 vs. 0.90; p = 0.016) and foveal (0.86 vs. 0.94; p = 0.039) insertion, whereas no improvement was identified for alterations of the central cartilage disc. Readers' confidence (p < 0.001) and assessability of the ulnar-sided insertions (p < 0.001) were superior with ancillary radial reformatting. Conclusions Access to the radial plane view of isotropic 3D sequences in MR arthrography improves diagnostic accuracy and confidence for ulnar-sided TFCC lesions. KW - joint instability KW - wrist KW - arthrography KW - magnetic resonance imaging KW - triangular fibrocartilage Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-266512 SN - 1432-1084 VL - 31 IS - 12 ER - TY - JOUR A1 - Mihatsch, Patrick W. A1 - Beissert, Matthias A1 - Pomper, Martin G. A1 - Bley, Thorsten A. A1 - Seitz, Anna K. A1 - Kübler, Hubert A1 - Buck, Andreas K. A1 - Rowe, Steven P. A1 - Serfling, Sebastian E. A1 - Hartrampf, Philipp E. A1 - Werner, Rudolf A. T1 - Changing threshold-based segmentation has no relevant impact on semi-quantification in the context of structured reporting for PSMA-PET/CT JF - Cancers N2 - Prostate-specific membrane antigen (PSMA)-directed positron emission tomography/computed tomography (PET/CT) is increasingly utilized for staging of men with prostate cancer (PC). To increase interpretive certainty, the standardized PSMA reporting and data system (RADS) has been proposed. Using PSMA-RADS, we characterized lesions in 18 patients imaged with \(^{18}\)F-PSMA-1007 PET/CT for primary staging and determined the stability of semi-quantitative parameters. Six hundred twenty-three lesions were categorized according to PSMA-RADS and manually segmented. In this context, PSMA-RADS-3A (soft-tissue) or -3B (bone) lesions are defined as being indeterminate for the presence of PC. For PMSA-RADS-4 and -5 lesions; however, PC is highly likely or almost certainly present [with further distinction based on absence (PSMA-RADS-4) or presence (PSMA-RADS-5) of correlative findings on CT]. Standardized uptake values (SUV\(_{max}\), SUV\(_{peak}\), SUV\(_{mean}\)) were recorded, and volumetric parameters [PSMA-derived tumor volume (PSMA-TV); total lesion PSMA (TL-PSMA)] were determined using different maximum intensity thresholds (MIT) (40 vs. 45 vs. 50%). SUV\(_{max}\) was significantly higher in PSMA-RADS-5 lesions compared to all other PSMA-RADS categories (p ≤ 0.0322). In particular, the clinically challenging PSMA-RADS-3A lesions showed significantly lower SUV\(_{max}\) and SUV\(_{peak}\) compared to the entire PSMA-RADS-4 or -5 cohort (p < 0.0001), while for PSMA-RADS-3B this only applies when compared to the entire PSMA-RADS-5 cohort (p < 0.0001), but not to the PSMA-RADS-4 cohort (SUV\(_{max}\), p = 0.07; SUV\(_{peak}\), p = 0.08). SUV\(_{mean}\) (p = 0.30) and TL-PSMA (p = 0.16) in PSMA-RADS-5 lesions were not influenced by changing the MIT, while PSMA-TV showed significant differences when comparing 40 vs. 50% MIT (p = 0.0066), which was driven by lymph nodes (p = 0.0239), but not bone lesions (p = 0.15). SUV\(_{max}\) was significantly higher in PSMA-RADS-5 lesions compared to all other PSMA-RADS categories in \(^{18}\)F-PSMA-1007 PET/CT. As such, the latter parameter may assist the interpreting molecular imaging specialist in assigning the correct PSMA-RADS score to sites of disease, thereby increasing diagnostic certainty. In addition, changes of the MIT in PSMA-RADS-5 lesions had no significant impact on SUV\(_{mean}\) and TL-PSMA in contrast to PSMA-TV. KW - \(^{18}\)F-PSMA-1007 KW - PET/CT KW - staging KW - prostate cancer KW - standardized reporting system KW - PSMA-RADS Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-254782 SN - 2072-6694 VL - 14 IS - 2 ER - TY - JOUR A1 - Fröhlich, Matthias A1 - Serfling, Sebastian A1 - Higuchi, Takahiro A1 - Pomper, Martin G. A1 - Rowe, Steven P. A1 - Schmalzing, Marc A1 - Tony, Hans-Peter A1 - Gernert, Michael A1 - Strunz, Patrick-Pascal A1 - Portegys, Jan A1 - Schwaneck, Eva-Christina A1 - Gadeholt, Ottar A1 - Weich, Alexander A1 - Buck, Andreas K. A1 - Bley, Thorsten A. A1 - Guggenberger, Konstanze V. A1 - Werner, Rudolf A. T1 - Whole-Body [\(^{18}\)F]FDG PET/CT Can Alter Diagnosis in Patients with Suspected Rheumatic Disease JF - Diagnostics N2 - The 2-deoxy-d-[\(^{18}\)F]fluoro-D-glucose (FDG) positron emission tomography/computed tomography (PET/CT) is widely utilized to assess the vascular and articular inflammatory burden of patients with a suspected diagnosis of rheumatic disease. We aimed to elucidate the impact of [\(^{18}\)F]FDG PET/CT on change in initially suspected diagnosis in patients at the time of the scan. Thirty-four patients, who had undergone [\(^{18}\)F]FDG PET/CT, were enrolled and the initially suspected diagnosis prior to [18F]FDG PET/CT was compared to the final diagnosis. In addition, a semi-quantitative analysis including vessel wall-to-liver (VLR) and joint-to-liver (JLR) ratios was also conducted. Prior to [\(^{18}\)F]FDG PET/CT, 22/34 (64.7%) of patients did not have an established diagnosis, whereas in 7/34 (20.6%), polymyalgia rheumatica (PMR) was suspected, and in 5/34 (14.7%), giant cell arteritis (GCA) was suspected by the referring rheumatologists. After [\(^{18}\)F]FDG PET/CT, the diagnosis was GCA in 19/34 (55.9%), combined GCA and PMR (GCA + PMR) in 9/34 (26.5%) and PMR in the remaining 6/34 (17.6%). As such, [\(^{18}\)F]FDG PET/CT altered suspected diagnosis in 28/34 (82.4%), including in all unclear cases. VLR of patients whose final diagnosis was GCA tended to be significantly higher when compared to VLR in PMR (GCA, 1.01 ± 0.08 (95%CI, 0.95–1.1) vs. PMR, 0.92 ± 0.1 (95%CI, 0.85–0.99), p = 0.07), but not when compared to PMR + GCA (1.04 ± 0.14 (95%CI, 0.95–1.13), p = 1). JLR of individuals finally diagnosed with PMR (0.94 ± 0.16, (95%CI, 0.83–1.06)), however, was significantly increased relative to JLR in GCA (0.58 ± 0.04 (95%CI, 0.55–0.61)) and GCA + PMR (0.64 ± 0.09 (95%CI, 0.57–0.71); p < 0.0001, respectively). In individuals with a suspected diagnosis of rheumatic disease, an inflammatory-directed [\(^{18}\)F]FDG PET/CT can alter diagnosis in the majority of the cases, particularly in subjects who were referred because of diagnostic uncertainty. Semi-quantitative assessment may be helpful in establishing a final diagnosis of PMR, supporting the notion that a quantitative whole-body read-out may be useful in unclear cases. KW - giant cell arteritis KW - GCA KW - [18F]FDG PET/CT KW - vasculature KW - inflammation KW - polymyalgia rheumatica KW - PMR KW - vasculitis Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-250227 SN - 2075-4418 VL - 11 IS - 11 ER - TY - JOUR A1 - Petritsch, Berhard A1 - Kosmala, Aleksander A1 - Weng, Andreas Max A1 - Bley, Thorsten Alexander T1 - Tin-filtered 100kV ultra-low-dose CT of the paranasal sinus: initial clinical results JF - PLoS ONE N2 - Objectives To investigate the feasibility, diagnostic image quality and radiation dose of 3\(^{rd}\) generation dual-source computed tomography (CT) using a tin-filtered 100 kV protocol in patients with suspected acute inflammatory sinus disease. Methods We retrospectively evaluated 109 consecutive patients who underwent CT (Siemens SOMATOM Force, Erlangen, Germany) of the paranasal sinus with a new tin-filtered scanprotocol (Sn100 kV; tube current 35 mAs) using iterative reconstruction. Two readers independently assessed subjective image quality using a five-point Likert scale (1 = excellent, 5 = non-diagnostic). Inter-observer agreement was calculated and expressed as percentage of agreement. Noise was determined for calculation of signal-to-noise-ratio (SNR). Effective radiation dose (ED) was calculated from the dose-length-product (DLP). Results All examinations showed diagnostic image quality regarding evaluation of inflammatory sinus disease. On average, subjective general image quality was rated moderate (= 3) with a percentage of agreement between the observers of 81%. The mean image noise was 14.3 HU. The calculated median SNR was 6.0 for intraorbital fat, and 3.6 for the vitreous body, respectively. The median DLP was 2.1 mGy*cm, resulting in a median ED of 0.012 mSv. Conclusions Taking the study limitations into account, ultra-low-dose tin-filtered CT of the paranasal sinus at a tube voltage of 100 kV utilizing an iterative reconstruction algorithm provides for reliable exclusion of suspected acute inflammatory sinus disease in 100% of the cases. KW - Computed axial tomography KW - Inflammatory diseases KW - Radiation exposure KW - Diagnostic medicine KW - Fats KW - Mastoid process KW - X-ray radiography KW - Soft tissues Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-204127 VL - 14 IS - 5 ER - TY - JOUR A1 - Heidenreich, Julius F. A1 - Weng, Andreas M. A1 - Donhauser, Julian A1 - Greiser, Andreas A1 - Chow, Kelvin A1 - Nordbeck, Peter A1 - Bley, Thorsten A. A1 - Köstler, Herbert T1 - T1- and ECV-mapping in clinical routine at 3 T: differences between MOLLI, ShMOLLI and SASHA JF - BMC Medical Imaging N2 - Background T1 mapping sequences such as MOLLI, ShMOLLI and SASHA make use of different technical approaches, bearing strengths and weaknesses. It is well known that obtained T1 relaxation times differ between the sequence techniques as well as between different hardware. Yet, T1 quantification is a promising tool for myocardial tissue characterization, disregarding the absence of established reference values. The purpose of this study was to evaluate the feasibility of native and post-contrast T1 mapping methods as well as ECV maps and its diagnostic benefits in a clinical environment when scanning patients with various cardiac diseases at 3 T. Methods Native and post-contrast T1 mapping data acquired on a 3 T full-body scanner using the three pulse sequences 5(3)3 MOLLI, ShMOLLI and SASHA in 19 patients with clinical indication for contrast enhanced MRI were compared. We analyzed global and segmental T1 relaxation times as well as respective extracellular volumes and compared the emerged differences between the used pulse sequences. Results T1 times acquired with MOLLI and ShMOLLI exhibited systematic T1 deviation compared to SASHA. Myocardial MOLLI T1 times were 19% lower and ShMOLLI T1 times 25% lower compared to SASHA. Native blood T1 times from MOLLI were 13% lower than SASHA, while post-contrast MOLLI T1-times were only 5% lower. ECV values exhibited comparably biased estimation with MOLLI and ShMOLLI compared to SASHA in good agreement with results reported in literature. Pathology-suspect segments were clearly differentiated from remote myocardium with all three sequences. Conclusion Myocardial T1 mapping yields systematically biased pre- and post-contrast T1 times depending on the applied pulse sequence. Additionally calculating ECV attenuates this bias, making MOLLI, ShMOLLI and SASHA better comparable. Therefore, myocardial T1 mapping is a powerful clinical tool for classification of soft tissue abnormalities in spite of the absence of established reference values. KW - T1 mapping KW - MOLLI KW - ShMOLLI KW - SASHA KW - Extracellular volume KW - 3 T Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-201999 VL - 19 ER - TY - JOUR A1 - Vogel, Patrick A1 - Markert, Jonathan A1 - Rückert, Martin A. A1 - Herz, Stefan A1 - Keßler, Benedikt A1 - Dremel, Kilian A1 - Althoff, Daniel A1 - Weber, Matthias A1 - Buzug, Thorsten M. A1 - Bley, Thorsten A. A1 - Kullmann, Walter H. A1 - Hanke, Randolf A1 - Zabler, Simon A1 - Behr, Volker C. T1 - Magnetic Particle Imaging meets computed tomography: first simultaneous imaging JF - Scientific Reports N2 - Magnetic Particle Imaging (MPI) is a promising new tomographic modality for fast as well as three-dimensional visualization of magnetic material. For anatomical or structural information an additional imaging modality such as computed tomography (CT) is required. In this paper, the first hybrid MPI-CT scanner for multimodal imaging providing simultaneous data acquisition is presented. KW - Applied physics KW - Biomedical engineering KW - Imaging techniques Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-202501 VL - 9 ER - TY - JOUR A1 - Augustin, Anne Marie A1 - Welsch, Stefan A1 - Bley, Thorsten Alexander A1 - Lopau, Kai A1 - Kickuth, Ralph T1 - Color-coded summation images in the evaluation of renal artery stenosis before and after percutaneous transluminal angioplasty JF - BMC Medical Imaging N2 - Background: Endovascular therapy is the gold standard in patients with hemodynamic relevant renal artery stenosis (RAS) resistant to medical therapy. The severity grading of the stenosis as well as the result assessment after endovascular approach is predominantly based on visible estimations of the anatomic appearance. We aim to investigate the application of color-coded DSA parameters to gain hemodynamic information during endovascular renal artery interventions and for the assessment of the procedures technical success. Methods: We retrospectively evaluated 32 patients who underwent endovascular renal artery revascularization and applied color-coded summation imaging on selected monochromatic DSA images. The differences in time to peak (dTTP) of contrast enhancement in predefined anatomical measuring points were analyzed. Furthermore, differences in systolic blood pressure values (SBP) and serum creatinine were obtained. The value of underlying diabetes mellitus as a predictor for clinical outcome was assessed. Correlation analysis between the patients gender as well as the presence of diabetes mellitus and dTTP was performed. Results: Endovascular revascularization resulted in statistically significant improvement in 4/7 regions of interest. Highly significant improvement of perfusion in terms of shortened TTP values could be found at the segmental artery level and in the intrastenotical segment (p<0.001), significant improvement prestenotical and in the apical renal parenchyma (p<0.05). In the other anatomic regions, differences revealed not to be significant. Differences between SBP and serum creatinine levels before and after the procedure were significant (p=0.004 and 0.0004). Patients ' gender as well as the presence of diabetes mellitus did not reveal to be predictors for the clinical success of the procedure. Furthermore, diabetes and gender did not show relevant correlation with dTTP in the parenchymal measuring points. Conclusions: The supplementary use of color-coding DSA and the data gained from parametric images may provide helpful information in the evaluation of the procedures ' technical success. The segmental artery might be a particularly suitable vascular territory for analyzing differences in blood flow characteristics. Further studies with larger cohorts are needed to further confirm the diagnostic value of this technique. KW - digital subtraction angiography KW - color-coded KW - endovascular KW - renal artery KW - PTA Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-259086 VL - 21 IS - 1 ER - TY - JOUR A1 - Guggenberger, Konstanze Viktoria A1 - Bley, Thorsten Alexander T1 - Imaging in Vasculitis JF - Current Rheumatology Reports N2 - Purpose of Review: Vasculitides are characterized by mostly autoimmunologically induced inflammatory processes of vascularstructures. They have various clinical and radiologic appearances. Early diagnosis and reliable monitoring are indispensable foradequate therapy to prevent potentially serious complications. Imaging, in addition to laboratory tests and physical examination,constitutes a key component in assessing disease extent and activity. This review presents current standards and some typicalfindings in the context of imaging in vasculitis with particular attention to large vessel vasculitides. Recent Findings: Recently, imaging has gained importance in the management of vasculitis, especially regarding large vesselvasculitides (LVV). Recently, EULAR (European League Against Rheumatism) has launched its recommendations concerningthe diagnosis of LVVs. Imaging is recommended as the preferred complement to clinical examination. Color-coded duplexsonography is considered the first choice imaging test in suspected giant cell arteritis, and magnetic resonance imaging isconsidered the first choice in suspected Takayasu’sarteritis. Summary: Due to diversity of clinical and radiologic presentations, diagnosis and therapy monitoring of vasculitides mayconstitute a challenge. As a result of ongoing technological progress, a variety of non-invasive imaging modalities now playan elemental role in the interdisciplinary management of vasculitic diseases. KW - Vasculitis KW - Large vessel vasculitides (LVV) KW - Giant cell arteritis (GCA KW - Imaging KW - Magnetic resonance imaging(MRI) KW - EULAR guidelines Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-232762 SN - 1523-3774 VL - 22 IS - 34 ER - TY - JOUR A1 - Weng, Andreas M. A1 - Köstler, Herbert A1 - Bley, Thorsten A. A1 - Ritter, Christian O. T1 - Effect of short-term smoking & L-arginine on coronary endothelial function assessed by cardiac magnetic resonance cold pressor testing: a pilot study JF - BMC Cardiovascular Disorders N2 - Background The effect of smoking on coronary vasomotion has been investigated in the past with various imaging techniques in both short- and long-term smokers. Additionally, coronary vasomotion has been shown to be normalized in long-term smokers by L-Arginine acting as a substrate for NO synthase, revealing the coronary endothelium as the major site of abnormal vasomotor response. Aim of the prospective cohort study was to investigate coronary vasomotion of young healthy short-term smokers via magnetic resonance cold pressor test with and without the administration of L-Arginine and compare obtained results with the ones from nonsmokers. Methods Myocardial blood flow (MBF) was quantified with first-pass perfusion MRI on a 1.5 T scanner in healthy short-term smokers (N = 10, age: 25.0 ± 2.8 years, 5.0 ± 2.9 pack years) and nonsmokers (N = 10, age: 34.3 ± 13.6) both at rest and during cold pressor test (CPT). Smokers underwent an additional examination after administration of L-Arginine within a median of 7 days of the naïve examination. Results MBF at rest turned out to be 0.77 ± 0.30 (smokers with no L-Arginine; mean ± standard deviation), 0.66 ± 0.21 (smokers L-Arginine) and 0.84 ± 0.08 (nonsmokers). Values under CPT were 1.21 ± 0.42 (smokers no L-Arginine), 1.09 ± 0.35 (smokers L-Arginine) and 1.63 ± 0.33 (nonsmokers). In all groups, MBF was significantly increased under CPT compared to the corresponding rest examination (p < 0.05 in all cases). Additionally, MBF under CPT was significantly different between the smokers and the nonsmokers (p = 0.002). MBF at rest was significantly different between the smokers when L-Arginine was given and the nonsmokers (p = 0.035). Conclusion Short-term smokers showed a reduced response to cold both with and without the administration of L-Arginine. However, absolute MBF values under CPT were lower compared to nonsmokers independently of L-Arginine administration. KW - MRI KW - myocardial perfusion KW - cold pressor test KW - endothelium KW - L-arginine KW - smoking Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260559 VL - 21 ER - TY - JOUR A1 - Veldhoen, Simon A1 - Behzadi, Cyrus A1 - Lenz, Alexander A1 - Henes, Frank Oliver A1 - Rybczynski, Meike A1 - von Kodolitsch, Yskert A1 - Bley, Thorsten Alexander A1 - Adam, Gerhard A1 - Bannas, Peter T1 - Non-contrast MR angiography at 1.5 Tesla for aortic monitoring in Marfan patients after aortic root surgery JF - Journal of Cardiovascular Magnetic Resonance N2 - Background: Contrast-enhanced cardiovascular magnetic resonance angiography (CE-CMRA) is the established imaging modality for patients with Marfan syndrome requiring life-long annual aortic imaging before and after aortic root replacement. Contrast-free CMRA techniques avoiding side-effects of contrast media are highly desirable for serial imaging but have not been evaluated in the postoperative setup of Marfan patients. The purpose of this study was to assess the feasibility of non-contrast balanced steady-state free precession (bSSFP) magnetic resonance imaging for aortic monitoring of postoperative patients with Marfan syndrome. Methods: Sixty-four adult Marfan patients after aortic root replacement were prospectively included. Fourteen patients (22%) had a residual aortic dissection after surgical treatment of type A dissection. bSSFP imaging and CE-CMRA were performed at 1.5 Tesla. Two radiologists evaluated the images regarding image quality (1 = poor, 4 = excellent), artifacts (1 = severe, 4 = none) and aortic pathologies. Readers measured the aortic diameters at defined levels in both techniques. Statistics included observer agreement for image scoring and diameter measurements and ROC analyses for comparison of the diagnostic performance of bSSFP and CE-CMRA. Results: Both readers observed no significant differences in image quality between bSSFP and CE-CMRA and found a median image quality score of 4 for both techniques (all p > .05). No significant differences were found regarding the frequency of image artifacts in both sequences (all p > .05). Sensitivity and specificity for detection of aortic dissections was 100% for both readers and techniques. Compared to bSSFP imaging, CE-CMRA resulted in higher diameters (mean bias, 0.9 mm; p < .05). The inter-observer biases of diameter measurements were not significantly different (all p > .05), except for the distal graft anastomosis (p = .001). Using both techniques, the readers correctly identified a graft suture dehiscence with aneurysm formation requiring surgery. Conclusion: Unenhanced bSSFP CMR imaging allows for riskless aortic monitoring with high diagnostic accuracy in Marfan patients after aortic root surgery. KW - MR angiography Y1 - 2017 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-158693 VL - 19 IS - 82 ER - TY - JOUR A1 - Richter, Julian A. J. A1 - Wech, Tobias A1 - Weng, Andreas M. A1 - Stich, Manuel A1 - Weick, Stefan A1 - Breuer, Kathrin A1 - Bley, Thorsten A. A1 - Köstler, Herbert T1 - Free‐breathing self‐gated 4D lung MRI using wave‐CAIPI JF - Magnetic Resonance in Medicine N2 - Purpose The aim of this study was to compare the wave‐CAIPI (controlled aliasing in parallel imaging) trajectory to the Cartesian sampling for accelerated free‐breathing 4D lung MRI. Methods The wave‐CAIPI k‐space trajectory was implemented in a respiratory self‐gated 3D spoiled gradient echo pulse sequence. Trajectory correction applying the gradient system transfer function was used, and images were reconstructed using an iterative conjugate gradient SENSE (CG SENSE) algorithm. Five healthy volunteers and one patient with squamous cell carcinoma in the lung were examined on a clinical 3T scanner, using both sampling schemes. For quantitative comparison of wave‐CAIPI and standard Cartesian imaging, the normalized mutual information and the RMS error between retrospectively accelerated acquisitions and their respective references were calculated. The SNR ratios were investigated in a phantom study. Results The obtained normalized mutual information values indicate a lower information loss due to acceleration for the wave‐CAIPI approach. Average normalized mutual information values of the wave‐CAIPI acquisitions were 10% higher, compared with Cartesian sampling. Furthermore, the RMS error of the wave‐CAIPI technique was lower by 19% and the SNR was higher by 14%. Especially for short acquisition times (down to 1 minute), the undersampled Cartesian images showed an increased artifact level, compared with wave‐CAIPI. Conclusion The application of the wave‐CAIPI technique to 4D lung MRI reduces undersampling artifacts, in comparison to a Cartesian acquisition of the same scan time. The benefit of wave‐CAIPI sampling can therefore be traded for shorter examinations, or enhancing image quality of undersampled 4D lung acquisitions, keeping the scan time constant. KW - free‐breathing KW - lung KW - self‐gated KW - wave‐CAIPI Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-218075 VL - 84 IS - 6 SP - 3223 EP - 3233 ER - TY - JOUR A1 - Weng, Andreas M. A1 - Heidenreich, Julius F. A1 - Metz, Corona A1 - Veldhoen, Simon A1 - Bley, Thorsten A. A1 - Wech, Tobias T1 - Deep learning-based segmentation of the lung in MR-images acquired by a stack-of-spirals trajectory at ultra-short echo-times JF - BMC Medical Imaging N2 - Background Functional lung MRI techniques are usually associated with time-consuming post-processing, where manual lung segmentation represents the most cumbersome part. The aim of this study was to investigate whether deep learning-based segmentation of lung images which were scanned by a fast UTE sequence exploiting the stack-of-spirals trajectory can provide sufficiently good accuracy for the calculation of functional parameters. Methods In this study, lung images were acquired in 20 patients suffering from cystic fibrosis (CF) and 33 healthy volunteers, by a fast UTE sequence with a stack-of-spirals trajectory and a minimum echo-time of 0.05 ms. A convolutional neural network was then trained for semantic lung segmentation using 17,713 2D coronal slices, each paired with a label obtained from manual segmentation. Subsequently, the network was applied to 4920 independent 2D test images and results were compared to a manual segmentation using the Sørensen–Dice similarity coefficient (DSC) and the Hausdorff distance (HD). Obtained lung volumes and fractional ventilation values calculated from both segmentations were compared using Pearson’s correlation coefficient and Bland Altman analysis. To investigate generalizability to patients outside the CF collective, in particular to those exhibiting larger consolidations inside the lung, the network was additionally applied to UTE images from four patients with pneumonia and one with lung cancer. Results The overall DSC for lung tissue was 0.967 ± 0.076 (mean ± standard deviation) and HD was 4.1 ± 4.4 mm. Lung volumes derived from manual and deep learning based segmentations as well as values for fractional ventilation exhibited a high overall correlation (Pearson’s correlation coefficent = 0.99 and 1.00). For the additional cohort with unseen pathologies / consolidations, mean DSC was 0.930 ± 0.083, HD = 12.9 ± 16.2 mm and the mean difference in lung volume was 0.032 ± 0.048 L. Conclusions Deep learning-based image segmentation in stack-of-spirals based lung MRI allows for accurate estimation of lung volumes and fractional ventilation values and promises to replace the time-consuming step of manual image segmentation in the future. KW - MRI KW - lung KW - deep learning KW - image segmentation Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260520 VL - 21 ER - TY - JOUR A1 - Thurner, Annette A1 - Augustin, Anne Marie A1 - Bley, Thorsten Alexander A1 - Kickuth, Ralph T1 - 2D-perfusion angiography for intra-procedural endovascular treatment response assessment in chronic mesenteric ischemia: a feasibility study JF - BMC Medical Imaging N2 - Background Endovascular revascularization has become the first-line treatment of chronic mesenteric ischemia (CMI). The qualitative visual analysis of digital subtraction angiography (DSA) is dependent on observer experience and prone to interpretation errors. We evaluate the feasibility of 2D-Perfusion Angiography (2D-PA) for objective, quantitative treatment response assessment in CMI. Methods 49 revascularizations in 39 patients with imaging based evidence of mesenteric vascular occlusive disease and clinical signs of CMI were included in this retrospective study. To assess perfusion changes by 2D-PA, DSA-series were post-processed using a dedicated, commercially available software. Regions of interest (ROI) were placed in the pre- and post-stenotic artery segment. In aorto-ostial disease, the inflow ROI was positioned at the mesenteric artery orifice. The ratios outflow to inflow ROI for peak density (PD), time to peak and area-under-the-curve (AUC) were computed and compared pre- and post-interventionally. We graded motion artifacts by means of a four-point scale. Feasibility of 2D-PA and changes of flow parameters were evaluated. Results Motion artifacts due to a mobile vessel location beneath the diaphragm or within the mesenteric root, branch vessel superimposition and inadequate contrast enhancement at the inflow ROI during manually conducted DSA-series via selective catheters owing to steep vessel angulation, necessitated exclusion of 26 measurements from quantitative flow evaluation. The feasibility rate was 47%. In 23 technically feasible assessments, PD\(_{outflow}\)/PD\(_{inflow}\) increased by 65% (p < 0.001) and AUC\(_{outflow}\)/AUC\(_{inflow}\) increased by 85% (p < 0.001). The time to peak density values in the outflow ROI accelerated only minimally without reaching statistical significance. Age, BMI, target vessel (celiac trunk, SMA or IMA), stenosis location (ostial or truncal), calcification severity, plaque composition or the presence of a complex stenosis did not reach statistical significance in their distribution among the feasible and non-feasible group (p > 0.05). Conclusions Compared to other vascular territories and indications, the feasibility of 2D-PA in mesenteric revascularization for CMI was limited. Unfavorable anatomic conditions contributed to a high rate of inconclusive 2D-PA results. KW - 2D-perfusion angiography KW - chronic mesenteric ischemia KW - endovascular treatment KW - mesenteric stenting Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301131 VL - 22 ER - TY - JOUR A1 - Petritsch, Bernhard A1 - Pannbecker, Pauline A1 - Weng, Andreas M. A1 - Grunz, Jan-Peter A1 - Veldhoen, Simon A1 - Bley, Thorsten A. A1 - Kosmala, Aleksander T1 - Split-filter dual-energy CT pulmonary angiography for the diagnosis of acute pulmonary embolism: a study on image quality and radiation dose JF - Quantitative Imaging in Medicine and Surgery N2 - Background: Computed tomography (CT) pulmonary angiography is the diagnostic reference standard in suspected pulmonary embolism (PE). Favorable results for dual-energy CT (DECT) images have been reported for this condition. Nowadays, dual-energy data acquisition is feasible with different technical options, including a single-source split-filter approach. Therefore, the aim of this retrospective study was to investigate image quality and radiation dose of thoracic split-filter DECT in comparison to conventional single-energy CT in patients with suspected PE. Methods: A total of 110 CT pulmonary angiographies were accomplished either as standard single-energy CT with automatic tube voltage selection (ATVS) (n=58), or as split-filter DECT (n=52). Objective [pulmonary artery CT attenuation, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR)] and subjective image quality [four-point Likert scale; three readers (R)] were compared among the two study groups. Size-specific dose estimates (SSDE), dose-length-product (DLP) and volume CT dose index (CTDIvol) were assessed for radiation dose analysis. Results: Split-filter DECT images yielded 67.7% higher SNR (27.0 vs. 16.1; P<0.001) and 61.9% higher CNR (22.5 vs. 13.9; P<0.001) over conventional single-energy images, whereas CT attenuation was significantly lower (344.5 vs. 428.2 HU; P=0.013). Subjective image quality was rated good or excellent in 93.0%/98.3%/77.6% (R1/R2/R3) of the single-energy CT scans, and 84.6%/82.7%/80.8% (R1/R2/R3) of the split-filter DECT scans. SSDE, DLP and CTDIvol were significantly lower for conventional single-energy CT compared to split-filter DECT (all P<0.05), which was associated with 26.7% higher SSDE. Conclusions: In the diagnostic workup of acute PE, the split-filter allows for dual-energy data acquisition from single-source single-layer CT scanners. The existing opportunity to assess pulmonary “perfusion” based on analysis of iodine distribution maps is associated with higher radiation dose in terms of increased SSDE than conventional single-energy CT with ATVS. Moreover, a proportion of up to 3.8% non-diagnostic examinations in the current reference standard test for PE is not negligible. KW - dual-energy KW - CT-angiography KW - vascular KW - pulmonary arteries KW - embolism/thrombosis Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-231456 VL - 11 IS - 5 ER - TY - JOUR A1 - Gietzen, Carsten Herbert A1 - Kunz, Andreas Steven A1 - Luetkens, Karsten Sebastian A1 - Huflage, Henner A1 - Christopoulos, Georgios A1 - van Schoonhoven, Jörg A1 - Bley, Thorsten Alexander A1 - Schmitt, Rainer A1 - Grunz, Jan-Peter T1 - Evaluation of prestyloid recess morphology and ulnar-sided contrast leakage in CT arthrography of the wrist JF - BMC Musculoskeletal Disorders N2 - Background In wrist arthrograms, aberrant contrast material is frequently seen extending into the soft tissue adjacent to the ulnar styloid process. Since the prestyloid recess can mimic contrast leakage in CT arthrography, this study aims to provide a detailed analysis of its morphologic variability, while investigating whether actual ulnar-sided leakage is associated with injuries of the triangular fibrocartilage complex (TFCC). Methods Eighty-six patients with positive wrist trauma history underwent multi-compartment CT arthrography (40 women, median age 44.5 years). Studies were reviewed by two board-certified radiologists, who documented the morphology of the prestyloid recess regarding size, opening type, shape and position, as well as the presence or absence of ulnar-sided contrast leakage. Correlations between leakage and the presence of TFCC injuries were assessed using the mean square contingency coefficient (r\(_{ɸ}\)). Results The most common configuration of the prestyloid recess included a narrow opening (73.26%; width 2.26 ± 1.43 mm), saccular shape (66.28%), and palmar position compared to the styloid process (55.81%). Its mean length and anterior–posterior diameter were 6.89 ± 2.36 and 5.05 ± 1.97 mm, respectively. Ulnar-sided contrast leakage was reported in 29 patients (33.72%) with a mean extent of 12.30 ± 5.31 mm. Leakage occurred more often in patients with ulnar-sided TFCC injuries (r\(_{ɸ}\) = 0.480; p < 0.001), whereas no association was found for lesions of the central articular disc (r\(_{ɸ}\) = 0.172; p = 0.111). Conclusions Since ulnar-sided contrast leakage is more common in patients with peripheral TFCC injuries, distinction between an atypical configuration of the prestyloid recess and actual leakage is important in CT arthrography of the wrist. KW - prestyloid recess KW - arthrography KW - tomography KW - x-ray computed KW - triangular fibrocartilage complex Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301113 VL - 23 IS - 1 ER - TY - JOUR A1 - Augustin, Anne Marie A1 - Wolfschmidt, Franziska A1 - Elsässer, Thilo A1 - Sauer, Alexander A1 - Dierks, Alexander A1 - Bley, Thorsten Alexander A1 - Kickuth, Ralph T1 - Color-coded summation images for the evaluation of blood flow in endovascular aortic dissection fenestration JF - BMC Medical Imaging N2 - Background To analyze the benefit of color-coded summation images in the assessment of target lumen perfusion in patients with aortic dissection and malperfusion syndrome before and after fluoroscopy-guided aortic fenestration. Methods Between December 2011 and April 2020 25 patients with Stanford type A (n = 13) or type B dissection (n = 12) and malperfusion syndromes were treated with fluoroscopy-guided fenestration of the dissection flap using a re-entry catheter. The procedure was technically successful in 100% of the cases and included additional iliofemoral stent implantation in four patients. Intraprocedural systolic blood pressure measurements for gradient evaluation were performed in 19 cases. Post-processed color-coded DSA images were obtained from all DSA series before and following fenestration. Differences in time to peak (dTTP) values in the compromised aortic lumen and transluminal systolic blood pressure gradients were analyzed retrospectively. Correlation analysis between dTTP and changes in blood pressure gradients was performed. Results Mean TTP prior to dissection flap fenestration was 6.85 ± 1.35 s. After fenestration, mean TTP decreased significantly to 4.96 ± 0.94 s (p < 0.001). Available systolic blood pressure gradients between the true and the false lumen were reduced by a median of 4.0 mmHg following fenestration (p = 0.031), with significant reductions in Stanford type B dissections (p = 0.013) and minor reductions in type A dissections (p = 0.530). A moderate correlation with no statistical significance was found between dTTP and the difference in systolic blood pressure (r = 0.226; p = 0.351). Conclusions Hemodynamic parameters obtained from color-coded DSA confirmed a significant reduction of TTP values in the aortic target lumen in terms of an improved perfusion in the compromised aortic region. Color-coded DSA might thus be a suitable complementary tool in the assessment of complex vascular patterns prevailing in aortic dissections, especially when blood pressure measurements are not conclusive or feasible. KW - angiography KW - fenestration KW - color-coding KW - aortic dissection KW - DSA KW - endovascular Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301107 VL - 22 IS - 1 ER - TY - JOUR A1 - Diessner, Joachim A1 - Anders, Laura A1 - Herbert, Saskia A1 - Kiesel, Matthias A1 - Bley, Thorsten A1 - Schlaiss, Tanja A1 - Sauer, Stephanie A1 - Wöckel, Achim A1 - Bartmann, Catharina T1 - Evaluation of different imaging modalities for axillary lymph node staging in breast cancer patients to provide a personalized and optimized therapy algorithm JF - Journal of Cancer Research and Clinical Oncology N2 - Purpose The reliable detection of tumor-infiltrated axillary lymph nodes for breast cancer [BC] patients plays a decisive role in further therapy. We aimed to find out whether cross-sectional imaging techniques could improve sensitivity for pretherapeutic axillary staging in nodal-positive BC patients compared to conventional imaging such as mammography and sonography. Methods Data for breast cancer patients with tumor-infiltrated axillary lymph nodes having received surgery between 2014 and 2020 were included in this study. All examinations (sonography, mammography, computed tomography [CT] and magnetic resonance imaging [MRI]) were interpreted by board-certified specialists in radiology. The sensitivity of different imaging modalities was calculated, and binary logistic regression analyses were performed to detect variables influencing the detection of positive lymph nodes. Results All included 382 breast cancer patients had received conventional imaging, while 52.61% of the patients had received cross-sectional imaging. The sensitivity of the combination of all imaging modalities was 68.89%. The combination of MRI and CT showed 63.83% and the combination of sonography and mammography showed 36.11% sensitivity. Conclusion We could demonstrate that cross-sectional imaging can improve the sensitivity of the detection of tumor-infiltrated axillary lymph nodes in breast cancer patients. Only the safe detection of these lymph nodes at the time of diagnosis enables the evaluation of the response to neoadjuvant therapy, thereby allowing access to prognosis and improving new post-neoadjuvant therapies. KW - breast cancer imaging KW - positive nodal status KW - cross-sectional imaging KW - conventional imaging KW - post-neoadjuvant therapies KW - neoadjuvant therapies Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-324047 VL - 149 IS - 7 ER - TY - JOUR A1 - Laqua, Fabian Christopher A1 - Woznicki, Piotr A1 - Bley, Thorsten A. A1 - Schöneck, Mirjam A1 - Rinneburger, Miriam A1 - Weisthoff, Mathilda A1 - Schmidt, Matthias A1 - Persigehl, Thorsten A1 - Iuga, Andra-Iza A1 - Baeßler, Bettina T1 - Transfer-learning deep radiomics and hand-crafted radiomics for classifying lymph nodes from contrast-enhanced computed tomography in lung cancer JF - Cancers N2 - Objectives: Positron emission tomography (PET) is currently considered the non-invasive reference standard for lymph node (N-)staging in lung cancer. However, not all patients can undergo this diagnostic procedure due to high costs, limited availability, and additional radiation exposure. The purpose of this study was to predict the PET result from traditional contrast-enhanced computed tomography (CT) and to test different feature extraction strategies. Methods: In this study, 100 lung cancer patients underwent a contrast-enhanced \(^{18}\)F-fluorodeoxyglucose (FDG) PET/CT scan between August 2012 and December 2019. We trained machine learning models to predict FDG uptake in the subsequent PET scan. Model inputs were composed of (i) traditional “hand-crafted” radiomics features from the segmented lymph nodes, (ii) deep features derived from a pretrained EfficientNet-CNN, and (iii) a hybrid approach combining (i) and (ii). Results: In total, 2734 lymph nodes [555 (20.3%) PET-positive] from 100 patients [49% female; mean age 65, SD: 14] with lung cancer (60% adenocarcinoma, 21% plate epithelial carcinoma, 8% small-cell lung cancer) were included in this study. The area under the receiver operating characteristic curve (AUC) ranged from 0.79 to 0.87, and the scaled Brier score (SBS) ranged from 16 to 36%. The random forest model (iii) yielded the best results [AUC 0.871 (0.865–0.878), SBS 35.8 (34.2–37.2)] and had significantly higher model performance than both approaches alone (AUC: p < 0.001, z = 8.8 and z = 22.4; SBS: p < 0.001, z = 11.4 and z = 26.6, against (i) and (ii), respectively). Conclusion: Both traditional radiomics features and transfer-learning deep radiomics features provide relevant and complementary information for non-invasive N-staging in lung cancer. KW - computed tomography KW - computational neural networks KW - lymphatic metastasis KW - carcinoma KW - non-small-cell lung KW - small-cell lung Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-319231 SN - 2072-6694 VL - 15 IS - 10 ER - TY - JOUR A1 - Huflage, Henner A1 - Kunz, Andreas Steven A1 - Hendel, Robin A1 - Kraft, Johannes A1 - Weick, Stefan A1 - Razinskas, Gary A1 - Sauer, Stephanie Tina A1 - Pennig, Lenhard A1 - Bley, Thorsten Alexander A1 - Grunz, Jan-Peter T1 - Obesity-related pitfalls of virtual versus true non-contrast imaging — an intraindividual comparison in 253 oncologic patients JF - Diagnostics N2 - Objectives: Dual-source dual-energy CT (DECT) facilitates reconstruction of virtual non-contrast images from contrast-enhanced scans within a limited field of view. This study evaluates the replacement of true non-contrast acquisition with virtual non-contrast reconstructions and investigates the limitations of dual-source DECT in obese patients. Materials and Methods: A total of 253 oncologic patients (153 women; age 64.5 ± 16.2 years; BMI 26.6 ± 5.1 kg/m\(^2\)) received both multi-phase single-energy CT (SECT) and DECT in sequential staging examinations with a third-generation dual-source scanner. Patients were allocated to one of three BMI clusters: non-obese: <25 kg/m\(^2\) (n = 110), pre-obese: 25–29.9 kg/m\(^2\) (n = 73), and obese: >30 kg/m\(^2\) (n = 70). Radiation dose and image quality were compared for each scan. DECT examinations were evaluated regarding liver coverage within the dual-energy field of view. Results: While arterial contrast phases in DECT were associated with a higher CTDI\(_{vol}\) than in SECT (11.1 vs. 8.1 mGy; p < 0.001), replacement of true with virtual non-contrast imaging resulted in a considerably lower overall dose-length product (312.6 vs. 475.3 mGy·cm; p < 0.001). The proportion of DLP variance predictable from patient BMI was substantial in DECT (R\(^2\) = 0.738) and SECT (R\(^2\) = 0.620); however, DLP of SECT showed a stronger increase in obese patients (p < 0.001). Incomplete coverage of the liver within the dual-energy field of view was most common in the obese subgroup (17.1%) compared with non-obese (0%) and pre-obese patients (4.1%). Conclusion: DECT facilitates a 30.8% dose reduction over SECT in abdominal oncologic staging examinations. Employing dual-source scanner architecture, the risk for incomplete liver coverage increases in obese patients. KW - dual-energy CT KW - dual-source CT KW - virtual non-contrast KW - radiation dose KW - spectral CT KW - obesity Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-313519 SN - 2075-4418 VL - 13 IS - 9 ER - TY - JOUR A1 - Huflage, Henner A1 - Grunz, Jan-Peter A1 - Patzer, Theresa Sophie A1 - Pannenbecker, Pauline A1 - Feldle, Philipp A1 - Sauer, Stephanie Tina A1 - Petritsch, Bernhard A1 - Ergün, Süleyman A1 - Bley, Thorsten Alexander A1 - Kunz, Andreas Steven T1 - Potential of unenhanced ultra-low-dose abdominal photon-counting CT with tin filtration: a cadaveric study JF - Diagnostics N2 - Objectives: This study investigated the feasibility and image quality of ultra-low-dose unenhanced abdominal CT using photon-counting detector technology and tin prefiltration. Materials and Methods: Employing a first-generation photon-counting CT scanner, eight cadaveric specimens were examined both with tin prefiltration (Sn 100 kVp) and polychromatic (120 kVp) scan protocols matched for radiation dose at three different levels: standard-dose (3 mGy), low-dose (1 mGy) and ultra-low-dose (0.5 mGy). Image quality was evaluated quantitatively by means of contrast-to-noise-ratios (CNR) with regions of interest placed in the renal cortex and subcutaneous fat. Additionally, three independent radiologists performed subjective evaluation of image quality. The intraclass correlation coefficient was calculated as a measure of interrater reliability. Results: Irrespective of scan mode, CNR in the renal cortex decreased with lower radiation dose. Despite similar mean energy of the applied x-ray spectrum, CNR was superior for Sn 100 kVp over 120 kVp at standard-dose (17.75 ± 3.51 vs. 14.13 ± 4.02), low-dose (13.99 ± 2.6 vs. 10.68 ± 2.17) and ultra-low-dose levels (8.88 ± 2.01 vs. 11.06 ± 1.74) (all p ≤ 0.05). Subjective image quality was highest for both standard-dose protocols (score 5; interquartile range 5–5). While no difference was ascertained between Sn 100 kVp and 120 kVp examinations at standard and low-dose levels, the subjective image quality of tin-filtered scans was superior to 120 kVp with ultra-low radiation dose (p < 0.05). An intraclass correlation coefficient of 0.844 (95% confidence interval 0.763–0.906; p < 0.001) indicated good interrater reliability. Conclusions: Photon-counting detector CT permits excellent image quality in unenhanced abdominal CT with very low radiation dose. Employment of tin prefiltration at 100 kVp instead of polychromatic imaging at 120 kVp increases the image quality even further in the ultra-low-dose range of 0.5 mGy. KW - spectral shaping KW - tin prefiltration KW - abdominal imaging KW - ultra-low-dose CT KW - urinary calculi KW - photon-counting Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-304122 SN - 2075-4418 VL - 13 IS - 4 ER - TY - JOUR A1 - Patzer, Theresa Sophie A1 - Kunz, Andreas Steven A1 - Huflage, Henner A1 - Conrads, Nora A1 - Luetkens, Karsten Sebastian A1 - Pannenbecker, Pauline A1 - Paul, Mila Marie A1 - Ergün, Süleyman A1 - Bley, Thorsten Alexander A1 - Grunz, Jan-Peter T1 - Ultrahigh-resolution photon-counting CT in cadaveric fracture models: spatial frequency is not everything JF - Diagnostics N2 - In this study, the impact of reconstruction sharpness on the visualization of the appendicular skeleton in ultrahigh-resolution (UHR) photon-counting detector (PCD) CT was investigated. Sixteen cadaveric extremities (eight fractured) were examined with a standardized 120 kVp scan protocol (CTDI\(_{vol}\) 10 mGy). Images were reconstructed with the sharpest non-UHR kernel (Br76) and all available UHR kernels (Br80 to Br96). Seven radiologists evaluated image quality and fracture assessability. Interrater agreement was assessed with the intraclass correlation coefficient. For quantitative comparisons, signal-to-noise-ratios (SNRs) were calculated. Subjective image quality was best for Br84 (median 1, interquartile range 1–3; p ≤ 0.003). Regarding fracture assessability, no significant difference was ascertained between Br76, Br80 and Br84 (p > 0.999), with inferior ratings for all sharper kernels (p < 0.001). Interrater agreement for image quality (0.795, 0.732–0.848; p < 0.001) and fracture assessability (0.880; 0.842–0.911; p < 0.001) was good. SNR was highest for Br76 (3.4, 3.0–3.9) with no significant difference to Br80 and Br84 (p > 0.999). Br76 and Br80 produced higher SNRs than all kernels sharper than Br84 (p ≤ 0.026). In conclusion, PCD-CT reconstructions with a moderate UHR kernel offer superior image quality for visualizing the appendicular skeleton. Fracture assessability benefits from sharp non-UHR and moderate UHR kernels, while ultra-sharp reconstructions incur augmented image noise. KW - photon-counting KW - tomography KW - X-ray computed KW - fracture KW - cancellous bone KW - convolution kernel Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-319281 SN - 2075-4418 VL - 13 IS - 10 ER - TY - JOUR A1 - Guggenberger, Konstanze V. A1 - Vogt, Marius L. A1 - Song, Jae W. A1 - Weng, Andreas M. A1 - Fröhlich, Matthias A1 - Schmalzing, Marc A1 - Venhoff, Nils A1 - Hillenkamp, Jost A1 - Pham, Mirko A1 - Meckel, Stephan A1 - Bley, Thorsten A. T1 - Intraorbital findings in giant cell arteritis on black blood MRI JF - European Radiology N2 - Objective Blindness is a feared complication of giant cell arteritis (GCA). However, the spectrum of pathologic orbital imaging findings on magnetic resonance imaging (MRI) in GCA is not well understood. In this study, we assess inflammatory changes of intraorbital structures on black blood MRI (BB-MRI) in patients with GCA compared to age-matched controls. Methods In this multicenter case-control study, 106 subjects underwent BB-MRI. Fifty-six patients with clinically or histologically diagnosed GCA and 50 age-matched controls without clinical or laboratory evidence of vasculitis were included. All individuals were imaged on a 3-T MR scanner with a post-contrast compressed-sensing (CS) T1-weighted sampling perfection with application-optimized contrasts using different flip angle evolution (SPACE) BB-MRI sequence. Imaging results were correlated with available clinical symptoms. Results Eighteen of 56 GCA patients (32%) showed inflammatory changes of at least one of the intraorbital structures. The most common finding was enhancement of at least one of the optic nerve sheaths (N = 13, 72%). Vessel wall enhancement of the ophthalmic artery was unilateral in 8 and bilateral in 3 patients. Enhancement of the optic nerve was observed in one patient. There was no significant correlation between imaging features of inflammation and clinically reported orbital symptoms (p = 0.10). None of the age-matched control patients showed any inflammatory changes of intraorbital structures. Conclusions BB-MRI revealed inflammatory findings in the orbits in up to 32% of patients with GCA. Optic nerve sheath enhancement was the most common intraorbital inflammatory change on BB-MRI. MRI findings were independent of clinically reported orbital symptoms. Key Points • Up to 32% of GCA patients shows signs of inflammation of intraorbital structures on BB-MRI. • Enhancement of the optic nerve sheath is the most common intraorbital finding in GCA patients on BB-MRI. • Features of inflammation of intraorbital structures are independent of clinically reported symptoms. KW - giant cell arteritis KW - magnetic resonance imaging KW - orbit KW - ophthalmic artery KW - optic nerve Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-324978 VL - 33 IS - 4 ER -