TY - JOUR A1 - Arnholdt, Jörg A1 - Kamawal, Yama A1 - Holzapfel, Boris Michael A1 - Ripp, Axel A1 - Rudert, Maximilian A1 - Steinert, Andre Friedrich T1 - Evaluation of implant fit and frontal plane alignment after bi-compartmental knee arthroplasty using patient-specific instruments and implants JF - Archives of Medical Science N2 - Introduction The goals of successful bi-compartmental knee arthroplasty are to achieve correct fit and positioning of the implant, while appropriately correcting the mechanical alignment of the leg after surgery. As these requirements are not always reliably fulfilled using off-the-shelf implant systems, newer approaches for bi-compartmental resurfacing have been explored. Material and methods In this article we report the radiographic results of 30 patients with anteromedial osteoarthritis (OA) who were treated with a novel patient-specific fixed-bearing bi-compartmental knee resurfacing system using custom-made implants and instruments. Utilizing standardized pre- and postoperative radiographic analyses (based on anterior-posterior and lateral, anterior-posterior weight-bearing full-length radiographs, patella skyline views and preoperative computed tomography (CT) scanning) implant fit and positioning as well as correction of the mechanical axis (hip-knee-ankle angle, HKA) were determined. Results On average, HKA was corrected from 173.4 ±3.47° preoperatively to 179.4 ±2.85° postoperatively. The coronal femoro-tibial angle was corrected on average 5.61°. The preoperative tibial slope measured on lateral views was 6.38 ±2.4°, while the average slope in the CT-based planning protocol (iView) was 6.14 ±2.40°. Postoperative lateral tibial slope was determined to be 5.77 ±1.97°. The thickness of the posterior femoral cuts was measured intraoperatively and, in all cases, corresponded well to the targeted thickness of the cuts provided by the iView. The joint line was preserved in all cases and the average Insall-Salvati index was 1.078 ±0.11 pre- and 1.072 ±0.11 postoperatively. The fit of the implant components measured by over- or underhang was excellent throughout (< 1.01 mm). Conclusions Custom-made bicompartmental knee arthroplasty can ensure optimized fitting and positioning of the implant with restoration of the leg axis. These implants could be considered as an alternative primary solution for knee surgeons treating bi-compartmental disease. KW - implant fit KW - bi-compartmental knee arthoplasty KW - bi-compartmental KW - implant positioning KW - knee osteoarthritis KW - knee arthroplasty KW - patient-specific KW - knee alignment Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-159668 VL - 14 IS - 6 ER - TY - JOUR A1 - Boelch, Sebastian P. A1 - Rueckl, Kilian A1 - Fuchs, Clara A1 - Jordan, Martin A1 - Knauer, Markus A1 - Steinert, Andre A1 - Rudert, Maximilian A1 - Luedemann, Martin T1 - Comparison of elution characteristics and compressive strength of biantibiotic-loaded PMMA bone cement for spacers: Copal\(^®\) spacem with gentamicin and vancomycin versus Palacos\(^®\) R+G with vancomycin JF - BioMed Research International N2 - Purpose. Copal\(^®\) spacem is a new PMMA bone cement for fabricating spacers. This study compares elution of gentamicin, elution of vancomycin, and compressive strength of Copal\(^®\) spacem and of Palacos\(^®\) R+G at different vancomycin loadings in the powder of the cements. We hypothesized that antibiotic elution of Copal\(^®\) spacem is superior at comparable compressive strength. Methods. Compression test specimens were fabricated using Copal\(^®\) spacem manually loaded with 0.5 g gentamicin and additionally 2 g, 4 g, and 6 g of vancomycin per 40 g of cement powder (COP specimens) and using 0.5 g gentamicin premixed Palacos\(^®\) R+G manually loaded with 2 g, 4 g, and 6 g of vancomycin per 40 g of cement powder (PAL specimens). These specimens were used for determination of gentamicin and vancomycin elution (in fetal calf serum, at 22°C) and for determination of compressive strength both prior and following the elution tests. Results. Cumulative gentamicin concentrations (p < 0.005) and gentamicin concentration after 28 days (p ≤ 0.043) were significantly lower for COP specimens compared to PAL specimens. Cumulative vancomycin concentrations were significantly higher (p ≤ 0.043) for COP specimens after the second day. Vancomycin concentrations after 28 days were not significantly higher for the Copal specimens loaded with 2 g and 4 g of vancomycin. Compressive strength was not significantly different between COP specimens and PAL specimens before elution tests. Compressive strength after the elution tests was significantly lower (p = 0.005) for COP specimens loaded with 2 g of vancomycin. Conclusion. We could not demonstrate consistent superior antibiotic elution from Copal\(^®\) spacem compared to Palacos\(^®\) R+G for fabricating gentamicin and vancomycin loaded spacers. The results do not favor Copal\(^®\) spacem over Palacos\(^®\) R+G for the use as a gentamicin and vancomycin biantibiotic-loaded spacer. KW - Copal\(^®\) spacem KW - PMMA bone cement KW - elution KW - compressive strength Y1 - 2018 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-177435 VL - 2018 IS - 4323518 ER -