TY - JOUR A1 - Guckenberger, Matthias A1 - Roesch, Johannes A1 - Baier, Kurt A1 - Sweeney, Reinhart A. A1 - Flentje, Michael T1 - Dosimetric consequences of translational and rotational errors in frame-less image-guided radiosurgery N2 - Background: To investigate geometric and dosimetric accuracy of frame-less image-guided radiosurgery (IG-RS) for brain metastases. Methods and materials: Single fraction IG-RS was practiced in 72 patients with 98 brain metastases. Patient positioning and immobilization used either double- (n = 71) or single-layer (n = 27) thermoplastic masks. Pre-treatment set-up errors (n = 98) were evaluated with cone-beam CT (CBCT) based image-guidance (IG) and were corrected in six degrees of freedom without an action level. CBCT imaging after treatment measured intra-fractional errors (n = 64). Pre- and posttreatment errors were simulated in the treatment planning system and target coverage and dose conformity were evaluated. Three scenarios of 0 mm, 1 mm and 2 mm GTV-to-PTV (gross tumor volume, planning target volume) safety margins (SM) were simulated. Results: Errors prior to IG were 3.9 mm± 1.7 mm (3D vector) and the maximum rotational error was 1.7° ± 0.8° on average. The post-treatment 3D error was 0.9 mm± 0.6 mm. No differences between double- and single-layer masks were observed. Intra-fractional errors were significantly correlated with the total treatment time with 0.7mm±0.5mm and 1.2mm±0.7mm for treatment times ≤23 minutes and >23 minutes (p<0.01), respectively. Simulation of RS without image-guidance reduced target coverage and conformity to 75% ± 19% and 60% ± 25% of planned values. Each 3D set-up error of 1 mm decreased target coverage and dose conformity by 6% and 10% on average, respectively, with a large inter-patient variability. Pre-treatment correction of translations only but not rotations did not affect target coverage and conformity. Post-treatment errors reduced target coverage by >5% in 14% of the patients. A 1 mm safety margin fully compensated intra-fractional patient motion. Conclusions: IG-RS with online correction of translational errors achieves high geometric and dosimetric accuracy. Intra-fractional errors decrease target coverage and conformity unless compensated with appropriate safety margins. KW - Medizin KW - Radiosurgery KW - Frame-less KW - Frame-based KW - Stereotactic KW - Image-guidance Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-75669 ER - TY - JOUR A1 - Guckenberger, Matthias A1 - Hawkins, Maria A1 - Flentje, Michael A1 - Sweeney, Reinhart A. T1 - Fractionated radiosurgery for painful spinal metastases: DOSIS - a phase II trial N2 - Background One third of all cancer patients will develop bone metastases and the vertebral column is involved in approximately 70 % of these patients. Conventional radiotherapy with of 1–10 fractions and total doses of 8-30 Gy is the current standard for painful vertebral metastases; however, the median pain response is short with 3–6 months and local tumor control is limited with these rather low irradiation doses. Recent advances in radiotherapy technology – intensity modulated radiotherapy for generation of highly conformal dose distributions and image-guidance for precise treatment delivery – have made dose-escalated radiosurgery of spinal metastases possible and early results of pain and local tumor control are promising. The current study will investigate efficacy and safety of radiosurgery for painful vertebral metastases and three characteristics will distinguish this study. 1) A prognostic score for overall survival will be used for selection of patients with longer life expectancy to allow for analysis of long-term efficacy and safety. 2) Fractionated radiosurgery will be performed with the number of treatment fractions adjusted to either good (10 fractions) or intermediate (5 fractions) life expectancy. Fractionation will allow inclusion of tumors immediately abutting the spinal cord due to higher biological effective doses at the tumor - spinal cord interface compared to single fraction treatment. 3) Dose intensification will be performed in the involved parts of the vertebrae only, while uninvolved parts are treated with conventional doses using the simultaneous integrated boost concept. Methods / Design It is the study hypothesis that hypo-fractionated image-guided radiosurgery significantly improves pain relief compared to historic data of conventionally fractionated radiotherapy. Primary endpoint is pain response 3 months after radiosurgery, which is defined as pain reduction of ≥2 points at the treated vertebral site on the 0 to 10 Visual Analogue Scale. 60 patients will be included into this two-centre phase II trial. Conclusions Results of this study will refine the methods of patient selection, target volume definition, treatment planning and delivery as well as quality assurance for radiosurgery. It is the intention of this study to form the basis for a future randomized controlled trial comparing conventional radiotherapy with fractionated radiosurgery for palliation of painful vertebral metastases. Trial registration ClinicalTrials.gov Identifier: NCT01594892 KW - Medizin KW - Phase II trial KW - Spinal metastasis KW - Pain KW - Radiosurgery KW - Stereotactic body radiotherapy Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-75853 ER -