TY - JOUR A1 - Tamihardja, Jörg A1 - Razinskas, Gary A1 - Exner, Florian A1 - Richter, Anne A1 - Kessler, Patrick A1 - Weick, Stefan A1 - Kraft, Johannes A1 - Mantel, Frederick A1 - Flentje, Michael A1 - Polat, Bülent T1 - Comparison of treatment plans for hypofractionated high-dose prostate cancer radiotherapy using the Varian Halcyon and the Elekta Synergy platforms JF - Journal of Applied Clinical Medical Physics N2 - Purpose To compare radiotherapy plans between an O-ring and a conventional C-arm linac for hypofractionated high-dose prostate radiotherapy in terms of plan quality, dose distribution, and quality assurance in a multi-vendor environment. Methods Twenty prostate cancer treatment plans were irradiated on the O-ring Varian Halcyon linac and were re-optimized for the C-arm Elekta Synergy Agility linac. Dose-volume histogram metrics for target coverage and organ at risk dose, quality assurance, and monitor units were retrospectively compared. Patient-specific quality assurance with ion chamber measurements, gamma index analysis, and portal dosimetry was performed using the Varian Portal Dosimetry system and the ArcCHECK® phantom (Sun Nuclear Corporation). Prostate-only radiotherapy was delivered with simultaneous integrated boost (SIB) volumetric modulated arc therapy (VMAT) in 20 fractions of 2.5/3.0 Gy each. Results For both linacs, target coverage was excellent and plan quality comparable. Homogeneity in PTVBoost was high for Synergy as well as Halcyon with a mean homogeneity index of 0.07 ± 0.01 and 0.05 ± 0.01, respectively. Mean dose for the organs at risk rectum and bladder differed not significantly between the linacs but were higher for the femoral heads and penile bulb for Halcyon. Quality assurance showed no significant differences in terms of ArcCHECK gamma pass rates. Median pass rate for 3%/2 mm was 99.3% (96.7 to 99.8%) for Synergy and 99.8% (95.6 to 100%) for Halcyon. Agreement between calculated and measured dose was high with a median deviation of −0.6% (−1.7 to 0.8%) for Synergy and 0.2% (−0.6 to 2.3%) for Halcyon. Monitor units were higher for the Halcyon by approximately 20% (p < 0.001). Conclusion Hypofractionated high-dose prostate cancer SIB VMAT on the Halcyon system is feasible with comparable plan quality in reference to a standard C-arm Elekta Synergy linac. KW - acute toxicity KW - dose evaluation KW - Halcyon KW - hypofractionation KW - prostate cancer KW - Synergy Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-260722 VL - 22 IS - 9 ER - TY - JOUR A1 - Kraft, Johannes A1 - Weick, Stefan A1 - Breuer, Kathrin A1 - Lutyj, Paul A1 - Bratengeier, Klaus A1 - Exner, Florian A1 - Richter, Anne A1 - Tamihardja, Jörg A1 - Lisowski, Dominik A1 - Polat, Bülent A1 - Flentje, Michael T1 - Treatment plan comparison for irradiation of multiple brain metastases with hippocampal avoidance whole brain radiotherapy and simultaneous integrated boost using the Varian Halcyon and the Elekta Synergy platforms JF - Radiation Oncology N2 - No abstract available. KW - treatment plan KW - multiple brain metastases Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301221 VL - 17 ER - TY - JOUR A1 - Richter, Anne A1 - Weick, Stefan A1 - Krieger, Thomas A1 - Exner, Florian A1 - Kellner, Sonja A1 - Polat, Bülent A1 - Flentje, Michael T1 - Evaluation of a software module for adaptive treatment planning and re-irradiation JF - Radiation Oncology N2 - Background: The aim of this work is to validate the Dynamic Planning Module in terms of usability and acceptance in the treatment planning workflow. Methods: The Dynamic Planning Module was used for decision making whether a plan adaptation was necessary within one course of radiation therapy. The Module was also used for patients scheduled for re-irradiation to estimate the dose in the pretreated region and calculate the accumulated dose to critical organs at risk. During one year, 370 patients were scheduled for plan adaptation or re-irradiation. All patient cases were classified according to their treated body region. For a sub-group of 20 patients treated with RT for lung cancer, the dosimetric effect of plan adaptation during the main treatment course was evaluated in detail. Changes in tumor volume, frequency of re-planning and the time interval between treatment start and plan adaptation were assessed. Results: The Dynamic Planning Tool was used in 20% of treated patients per year for both approaches nearly equally (42% plan adaptation and 58% re-irradiation). Most cases were assessed for the thoracic body region (51%) followed by pelvis (21%) and head and neck cases (10%). The sub-group evaluation showed that unintended plan adaptation was performed in 38% of the scheduled cases. A median time span between first day of treatment and necessity of adaptation of 17 days (range 4–35 days) was observed. PTV changed by 12 ± 12% on average (maximum change 42%). PTV decreased in 18 of 20 cases due to tumor shrinkage and increased in 2 of 20 cases. Re-planning resulted in a reduction of the mean lung dose of the ipsilateral side in 15 of 20 cases. Conclusion: The experience of one year showed high acceptance of the Dynamic Planning Module in our department for both physicians and medical physicists. The re-planning can potentially reduce the accumulated dose to the organs at risk and ensure a better target volume coverage. In the re-irradiation situation, the Dynamic Planning Tool was used to consider the pretreatment dose, to adapt the actual treatment schema more specifically and to review the accumulated dose. KW - re-irradiation KW - lung cancer KW - adaptation KW - re-planning Y1 - 2017 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-158711 VL - 12 IS - 205 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Schortmann, Max A1 - Lawrenz, Ingulf A1 - Weick, Stefan A1 - Bratengeier, Klaus A1 - Flentje, Michael A1 - Guckenberger, Matthias A1 - Polat, Bülent T1 - Moderately hypofractionated radiotherapy for localized prostate cancer: updated long-term outcome and toxicity analysis JF - Strahlentherapie und Onkologie N2 - Purpose Evaluation of long-term outcome and toxicity of moderately hypofractionated radiotherapy using intensity-modulated radiotherapy (IMRT) with simultaneous integrated boost treatment planning and cone beam CT-based image guidance for localized prostate cancer. Methods Between 2005 and 2015, 346 consecutive patients with localized prostate cancer received primary radiotherapy using cone beam CT-based image-guided intensity-modulated radiotherapy (IG-IMRT) and volumetric modulated arc therapy (IG-VMAT) with a simultaneous integrated boost (SIB). Total doses of 73.9 Gy (n = 44) and 76.2 Gy (n = 302) to the high-dose PTV were delivered in 32 and 33 fractions, respectively. The low-dose PTV received a dose (D95) of 60.06 Gy in single doses of 1.82 Gy. The pelvic lymph nodes were treated in 91 high-risk patients to 45.5 Gy (D95). Results Median follow-up was 61.8 months. The 5‑year biochemical relapse-free survival (bRFS) was 85.4% for all patients and 93.3, 87.4, and 79.4% for low-, intermediate-, and high-risk disease, respectively. The 5‑year prostate cancer-specific survival (PSS) was 94.8% for all patients and 98.7, 98.9, 89.3% for low-, intermediate-, and high-risk disease, respectively. The 5‑year and 10-year overall survival rates were 83.8 and 66.3% and the 5‑year and 10-year freedom from distant metastasis rates were 92.2 and 88.0%, respectively. Cumulative 5‑year late GU toxicity and late GI toxicity grade ≥2 was observed in 26.3 and 12.1% of the patients, respectively. Cumulative 5‑year late grade 3 GU/GI toxicity occurred in 4.0/1.2%. Conclusion Moderately hypofractionated radiotherapy using SIB treatment planning and cone beam CT image guidance resulted in high biochemical control and survival with low rates of late toxicity. KW - simultaneous integrated boost KW - cone beam CT KW - hypofractionation KW - intensity-modulated radiation therapy KW - image-guided radiation therapy Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-232509 SN - 0179-7158 VL - 197 ER - TY - JOUR A1 - Richter, Anne A1 - Wegener, Sonja A1 - Breuer, Kathrin A1 - Razinskas, Gary A1 - Weick, Stefan A1 - Exner, Florian A1 - Bratengeier, Klaus A1 - Flentje, Michael A1 - Sauer, Otto A1 - Polat, Bülent T1 - Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems JF - Radiation Oncology N2 - Background To implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans. Methods For 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom. Results The mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans. Conclusions For the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system. KW - whole breast irradiation KW - Halcyon KW - IGRT KW - dose to OARs Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-265704 VL - 16 ER - TY - JOUR A1 - Richter, Anne A1 - Exner, Florian A1 - Bratengeier, Klaus A1 - Polat, Bülent A1 - Flentje, Michael A1 - Weick, Stefan T1 - Impact of beam configuration on VMAT plan quality for Pinnacle\(^3\)Auto-Planning for head and neck cases JF - Radiation Oncology N2 - Background The purpose of this study was to compare automatically generated VMAT plans to find the superior beam configurations for Pinnacle3 Auto-Planning and share “best practices”. Methods VMAT plans for 20 patients with head and neck cancer were generated using Pinnacle3 Auto-Planning Module (Pinnacle3 Version 9.10) with different beam setup parameters. VMAT plans for single (V1) or double arc (V2) and partial or full gantry rotation were optimized. Beam configurations with different collimator positions were defined. Target coverage and sparing of organs at risk were evaluated based on scoring of an evaluation parameter set. Furthermore, dosimetric evaluation was performed based on the composite objective value (COV) and a new cross comparison method was applied using the COVs. Results The evaluation showed a superior plan quality for double arcs compared to one single arc or two single arcs for all cases. Plan quality was superior if a full gantry rotation was allowed during optimization for unilateral target volumes. A double arc technique with collimator setting of 15° was superior to a double arc with collimator 60° and a two single arcs with collimator setting of 15° and 345°. Conclusion The evaluation showed that double and full arcs are superior to single and partial arcs in terms of organs at risk sparing even for unilateral target volumes. The collimator position was found as an additional setup parameter, which can further improve the target coverage and sparing of organs at risk. KW - auto-planning KW - VMAT KW - single arc KW - double arc KW - full arc KW - partial arc KW - plan comparison Y1 - 2019 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-200301 VL - 14 ER - TY - JOUR A1 - Weick, Stefan A1 - Breuer, Kathrin A1 - Richter, Anne A1 - Exner, Florian A1 - Ströhle, Serge-Peer A1 - Lutyj, Paul A1 - Tamihardja, Jörg A1 - Veldhoen, Simon A1 - Flentje, Michael A1 - Polat, Bülent T1 - Non-rigid image registration of 4D-MRI data for improved delineation of moving tumors JF - BMC Medical Imaging N2 - Background To increase the image quality of end-expiratory and end-inspiratory phases of retrospective respiratory self-gated 4D MRI data sets using non-rigid image registration for improved target delineation of moving tumors. Methods End-expiratory and end-inspiratory phases of volunteer and patient 4D MRI data sets are used as targets for non-rigid image registration of all other phases using two different registration schemes: In the first, all phases are registered directly (dir-Reg) while next neighbors are successively registered until the target is reached in the second (nn-Reg). Resulting data sets are quantitatively compared using diaphragm and tumor sharpness and the coefficient of variation of regions of interest in the lung, liver, and heart. Qualitative assessment of the patient data regarding noise level, tumor delineation, and overall image quality was performed by blinded reading based on a 4 point Likert scale. Results The median coefficient of variation was lower for both registration schemes compared to the target. Median dir-Reg coefficient of variation of all ROIs was 5.6% lower for expiration and 7.0% lower for inspiration compared with nn-Reg. Statistical significant differences between the two schemes were found in all comparisons. Median sharpness in inspiration is lower compared to expiration sharpness in all cases. Registered data sets were rated better compared to the targets in all categories. Over all categories, mean expiration scores were 2.92 +/- 0.18 for the target, 3.19 +/- 0.22 for nn-Reg and 3.56 +/- 0.14 for dir-Reg and mean inspiration scores 2.25 +/- 0.12 for the target, 2.72 +/- 215 0.04 for nn-Reg and 3.78 +/- 0.04 for dir-Reg. Conclusions In this work, end-expiratory and inspiratory phases of a 4D MRI data sets are used as targets for non-rigid image registration of all other phases. It is qualitatively and quantitatively shown that image quality of the targets can be significantly enhanced leading to improved target delineation of moving tumors. KW - 4D-MRI KW - Non-rigid image registration KW - Radiotherapy treatment planning KW - Respiratory induced tumor motion Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-229271 VL - 20 ER - TY - JOUR A1 - Richter, Julian A. J. A1 - Wech, Tobias A1 - Weng, Andreas M. A1 - Stich, Manuel A1 - Weick, Stefan A1 - Breuer, Kathrin A1 - Bley, Thorsten A. A1 - Köstler, Herbert T1 - Free‐breathing self‐gated 4D lung MRI using wave‐CAIPI JF - Magnetic Resonance in Medicine N2 - Purpose The aim of this study was to compare the wave‐CAIPI (controlled aliasing in parallel imaging) trajectory to the Cartesian sampling for accelerated free‐breathing 4D lung MRI. Methods The wave‐CAIPI k‐space trajectory was implemented in a respiratory self‐gated 3D spoiled gradient echo pulse sequence. Trajectory correction applying the gradient system transfer function was used, and images were reconstructed using an iterative conjugate gradient SENSE (CG SENSE) algorithm. Five healthy volunteers and one patient with squamous cell carcinoma in the lung were examined on a clinical 3T scanner, using both sampling schemes. For quantitative comparison of wave‐CAIPI and standard Cartesian imaging, the normalized mutual information and the RMS error between retrospectively accelerated acquisitions and their respective references were calculated. The SNR ratios were investigated in a phantom study. Results The obtained normalized mutual information values indicate a lower information loss due to acceleration for the wave‐CAIPI approach. Average normalized mutual information values of the wave‐CAIPI acquisitions were 10% higher, compared with Cartesian sampling. Furthermore, the RMS error of the wave‐CAIPI technique was lower by 19% and the SNR was higher by 14%. Especially for short acquisition times (down to 1 minute), the undersampled Cartesian images showed an increased artifact level, compared with wave‐CAIPI. Conclusion The application of the wave‐CAIPI technique to 4D lung MRI reduces undersampling artifacts, in comparison to a Cartesian acquisition of the same scan time. The benefit of wave‐CAIPI sampling can therefore be traded for shorter examinations, or enhancing image quality of undersampled 4D lung acquisitions, keeping the scan time constant. KW - free‐breathing KW - lung KW - self‐gated KW - wave‐CAIPI Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-218075 VL - 84 IS - 6 SP - 3223 EP - 3233 ER - TY - JOUR A1 - Zimmermann, Marcus A1 - Richter, Anne A1 - Weick, Stefan A1 - Exner, Florian A1 - Mantel, Frederick A1 - Diefenhardt, Markus A1 - Fokas, Emmanouil A1 - Kosmala, Rebekka A1 - Flentje, Michael A1 - Polat, Bülent T1 - Acute toxicities of patients with locally advanced rectal cancer treated with intensified chemoradiotherapy within the CAO/ARO/AIO-12 trial: comparing conventional versus VMAT planning at a single center JF - Scientific Reports N2 - In locally advanced rectal cancer (LARC) neoadjuvant chemoradiotherapy is regarded as standard treatment. We assessed acute toxicities in patients receiving conventional 3D-conformal radiotherapy (3D-RT) and correlated them with dosimetric parameters after re-planning with volumetric modulated arc therapy (VMAT). Patients were randomized within the multicenter CAO/ARO/AIO-12 trial and received 50.4 Gy in 28 fractions and simultaneous chemotherapy with fluorouracil and oxaliplatin. Organs at risk (OAR) were contoured in a standardized approach. Acute toxicities and dose volume histogram parameters of 3D-RT plans were compared to retrospectively calculated VMAT plans. From 08/2015 to 01/2018, 35 patients with LARC were treated at one study center. Thirty-four patients were analyzed of whom 1 (3%) was UICC stage II and 33 (97%) patients were UICC stage III. Grade 3 acute toxicities occurred in 5 patients (15%). Patients with acute grade 1 cystitis (n = 9) had significantly higher D\(_{mean}\) values for bladder (29.4 Gy vs. 25.2 Gy, p < 0.01) compared to patients without bladder toxicities. Acute diarrhea was associated with small bowel volume (grade 2: 870.1 ccm vs. grade 0–1: 647.3 ccm; p < 0.01) and with the irradiated volumes V5 to V50. Using VMAT planning, we could reduce mean doses and irradiated volumes for all OAR: D\(_{mean}\) bladder (21.9 Gy vs. 26.3 Gy, p < 0.01), small bowel volumes V5–V45 (p < 0.01), D\(_{mean}\) anal sphincter (34.6 Gy vs. 35.6 Gy, p < 0.01) and D\(_{mean}\) femoral heads (right 11.4 Gy vs. 25.9 Gy, left 12.5 Gy vs. 26.6 Gy, p < 0.01). Acute small bowel and bladder toxicities were dose and volume dependent. Dose and volume sparing for all OAR could be achieved through VMAT planning and might result in less acute toxicities. KW - radiotherapy KW - rectal cancer Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-301255 VL - 12 ER - TY - JOUR A1 - Tamihardja, Jörg A1 - Lawrenz, Ingulf A1 - Lutyj, Paul A1 - Weick, Stefan A1 - Guckenberger, Matthias A1 - Polat, Bülent A1 - Flentje, Michael T1 - Propensity score-matched analysis comparing dose-escalated intensity-modulated radiation therapy versus external beam radiation therapy plus high-dose-rate brachytherapy for localized prostate cancer JF - Strahlentherapie und Onkologie N2 - Purpose Dose-escalated external beam radiation therapy (EBRT) and EBRT + high-dose-rate brachytherapy (HDR-BT) boost are guideline-recommended treatment options for localized prostate cancer. The purpose of this study was to compare long-term outcome and toxicity of dose-escalated EBRT versus EBRT + HDR-BT boost. Methods From 2002 to 2019, 744 consecutive patients received either EBRT or EBRT + HDR-BT boost, of whom 516 patients were propensity score matched. Median follow-up was 95.3 months. Cone beam CT image-guided EBRT consisted of 33 fractions of intensity-modulated radiation therapy with simultaneous integrated boost up to 76.23 Gy (D\(_{Mean}\)). Combined treatment was delivered as 46 Gy (D\(_{Mean}\)) EBRT, followed by two fractions HDR-BT boost with 9 Gy (D\(_{90\%}\)). Propensity score matching was applied before analysis of the primary endpoint, estimated 10-year biochemical relapse-free survival (bRFS), and the secondary endpoints metastasis-free survival (MFS) and overall survival (OS). Prognostic parameters were analyzed by Cox proportional hazard modelling. Genitourinary (GU)/gastrointestinal (GI) toxicity evaluation used the Common Toxicity Criteria for Adverse Events (v5.0). Results The estimated 10-year bRFS was 82.0% vs. 76.4% (p = 0.075) for EBRT alone versus combined treatment, respectively. The estimated 10-year MFS was 82.9% vs. 87.0% (p = 0.195) and the 10-year OS was 65.7% vs. 68.9% (p = 0.303), respectively. Cumulative 5‑year late GU ≥ grade 2 toxicities were seen in 23.6% vs. 19.2% (p = 0.086) and 5‑year late GI ≥ grade 2 toxicities in 11.1% vs. 5.0% of the patients (p = 0.002); cumulative 5‑year late grade 3 GU toxicity occurred in 4.2% vs. 3.6% (p = 0.401) and GI toxicity in 1.0% vs. 0.3% (p = 0.249), respectively. Conclusion Both treatment groups showed excellent long-term outcomes with low rates of severe toxicity. KW - long-term outcome KW - dose escalation KW - high-dose-rate brachytherapy boost KW - propensity score matching KW - toxicity Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-325055 VL - 198 IS - 8 ER -