TY - JOUR A1 - Reiter, Theresa A1 - Gensler, Daniel A1 - Ritter, Oliver A1 - Weiss, Ingo A1 - Geistert, Wolfgang A1 - Kaufmann, Ralf A1 - Hoffmeister, Sabine A1 - Friedrich, Michael T. A1 - Wintzheimer, Stefan A1 - Düring, Markus A1 - Nordbeck, Peter A1 - Jakob, Peter M. A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Bauer, Wolfgang R. T1 - Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures JF - Journal of Cardiovascular Magnetic Resonance N2 - Background: One of the safety concerns when performing electrophysiological (EP) procedures under magnetic resonance (MR) guidance is the risk of passive tissue heating due to the EP catheter being exposed to the radiofrequency (RF) field of the RF transmitting body coil. Ablation procedures that use catheters with irrigated tips are well established therapeutic options for the treatment of cardiac arrhythmias and when used in a modified mode might offer an additional system for suppressing passive catheter heating. Methods: A two-step approach was chosen. Firstly, tests on passive catheter heating were performed in a 1.5 T Avanto system (Siemens Healthcare Sector, Erlangen, Germany) using a ASTM Phantom in order to determine a possible maximum temperature rise. Secondly, a phantom was designed for simulation of the interface between blood and the vascular wall. The MR-RF induced temperature rise was simulated by catheter tip heating via a standard ablation generator. Power levels from 1 to 6 W were selected. Ablation duration was 120 s with no tip irrigation during the first 60 s and irrigation at rates from 2 ml/min to 35 ml/min for the remaining 60 s (Biotronik Qiona Pump, Berlin, Germany). The temperature was measured with fluoroscopic sensors (Luxtron, Santa Barbara, CA, USA) at a distance of 0 mm, 2 mm, 4 mm, and 6 mm from the catheter tip. Results: A maximum temperature rise of 22.4 degrees C at the catheter tip was documented in the MR scanner. This temperature rise is equivalent to the heating effect of an ablator's power output of 6 W at a contact force of the weight of 90 g (0.883 N). The catheter tip irrigation was able to limit the temperature rise to less than 2 degrees C for the majority of examined power levels, and for all examined power levels the residual temperature rise was less than 8 degrees C. Conclusion: Up to a maximum of 22.4 degrees C, the temperature rise at the tissue surface can be entirely suppressed by using the catheter's own irrigation system. The irrigated tip system can be used to increase MR safety of EP catheters by suppressing the effects of unwanted passive catheter heating due to RF exposure from the MR scanner. KW - EP Procedures KW - radiofrequency ablation KW - contact force KW - lesion size KW - MRI KW - temperature KW - tissue KW - wires KW - model KW - ablation KW - safety KW - catheter tip KW - MR guidance Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-134927 VL - 14 IS - 12 ER - TY - JOUR A1 - Lapa, Constantin A1 - Reiter, Theresa A1 - Kircher, Malte A1 - Schirbel, Andreas A1 - Werner, Rudolf A. A1 - Pelzer, Theo A1 - Pizarro, Carmen A1 - Skowasch, Dirk A1 - Thomas, Lena A1 - Schlesinger-Irsch, Ulrike A1 - Thomas, Daniel A1 - Bundschuh, Ralph A. A1 - Bauer, Wolfgang R. A1 - Gartner, Florian C. T1 - Somatostatin receptor based PET/CT in patients with the suspicion of cardiac sarcoidosis: an initial comparison to cardiac MRI JF - Oncotarget N2 - Diagnosis of cardiac sarcoidosis is often challenging. Whereas cardiac magnetic resonance imaging (CMR) and positron emission tomography/computed tomography (PET/CT) with \(^{18}\)F-fluorodeoxyglucose (FDG) are most commonly used to evaluate patients, PET/CT using radiolabeled somatostatin receptor (SSTR) ligands for visualization of inflammation might represent a more specific alternative. This study aimed to investigate the feasibility of SSTR–PET/CT for detecting cardiac sarcoidosis in comparison to CMR. 15 patients (6 males, 9 females) with sarcoidosis and suspicion on cardiac involvement underwent SSTR-PET/CT imaging and CMR. Images were visually scored. The AHA 17-segment model of the left myocardium was used for localization and comparison of inflamed myocardium for both imaging modalities. In semi-quantitative analysis, mean (SUV\(_{mean}\)) and maximum standardized uptake values (SUV\(_{max}\)) of affected myocardium were calculated and compared with both remote myocardium and left ventricular (LV) cavity. SSTR-PET was positive in 7/15, CMR in 10/15 patients. Of the 3 CMR+/PET- subjects, one patient with minor involvement (<25% of wall thickness in CMR) was missed by PET. The remaining two CMR+/PET- patients displayed no adverse cardiac events during follow-up. In the 17-segment model, PET/CT yielded 27 and CMR 29 positive segments. Overall concordance of the 2 modalities was 96.1% (245/255 segments analyzed). SUV\(_{mean}\) and SUV\(_{max}\) in inflamed areas were 2.0±1.2 and 2.6±1.2, respectively. The lesion-to-remote myocardium and lesion-to-LV cavity ratios were 1.8±0.2 and 1.9±0.2 for SUV\(_{mean}\) and 2.0±0.3 and 1.7±0.3 for SUV\(_{max}\), respectively. Detection of cardiac sarcoidosis by SSTR-PET/CT is feasible. Our data warrant further analysis in larger prospective series. KW - sarcoidosis KW - PET KW - SSTR KW - somatostatin receptor KW - DOTATOC Y1 - 2016 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-175423 VL - 7 IS - 47 ER - TY - JOUR A1 - Reiter, Theresa A1 - Ritter, Oliver A1 - Prince, Martin R. A1 - Nordbeck, Peter A1 - Wanner, Christoph A1 - Nagel, Eike A1 - Bauer, Wolfgang R. T1 - Minimizing Risk of Nephrogenic systemic fibrosis in Cardiovascular Magnetic Resonance N2 - Nephrogenic Systemic Fibrosis is a rare condition appearing only in patients with severe renal impairment or failure and presents with dermal lesions and involvement of internal organs. Although many cases are mild, an estimated 5 % have a progressive debilitating course. To date, there is no known effective treatment thus stressing the necessity of ample prevention measures. An association with the use of Gadolinium based contrast agents (GBCA) makes Nephrogenic Systemic Fibrosis a potential side effect of contrast enhanced magnetic resonance imaging and offers the opportunity for prevention by limiting use of gadolinium based contrast agents in renal failure patients. In itself toxic, Gadolinium is embedded into chelates that allow its safe use as a contrast agent. One NSF theory is that Gadolinium chelates distribute into the extracellular fluid compartment and set Gadolinium ions free, depending on multiple factors among which the duration of chelates exposure is directly related to the renal function. Major medical societies both in Europe and in North America have developed guidelines for the usage of GBCA. Since the establishment of these guidelines and the increased general awareness of this condition, the occurrence of NSF has been nearly eliminated. Giving an overview over the current knowledge of NSF pathobiochemistry, pathogenesis and treatment options this review focuses on the guidelines of the European Medicines Agency, the European Society of Urogenital Radiology, the FDA and the American College of Radiology from 2008 up to 2011 and the transfer of this knowledge into every day practice. KW - CMR Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-75068 ER - TY - JOUR A1 - Reiter, Theresa A1 - Ritter, Oliver A1 - Nordbeck, Peter A1 - Beer, Meinrad A1 - Bauer, Wolfgang Rudolf T1 - MRI-guided ablation of wide complex tachycardia in a univentricular heart JF - World Journal of Cardiology N2 - Magnetic resonance imaging can be used for preprocedural assessment of complex anatomy for radiofrequency (RF) ablations, e.g., in a univentricular heart. This case report features the treatment of a young patient with a functionally univentricular heart who suffered from persistent sudden onset tachycardia with wide complexes that required RF ablation as treatment. KW - magnetic resonsance imaging KW - ablation KW - Fontan’s Operation KW - ventricular tachycardia KW - univentricular heart Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-123165 VL - 4 IS - 8 ER - TY - JOUR A1 - Reiter, Theresa A1 - Weiss, Ingo A1 - Weber, Oliver M. A1 - Bauer, Wolfgang R. T1 - Signal voids of active cardiac implants at 3.0 T CMR JF - Scientific Reports N2 - Recent technical advancements allow cardiac MRI (CMR) examinations in the presence of so-called MRI conditional active cardiac implants at 3.0 T. However, the artifact burden caused by susceptibility effects remain an obstacle. All measurements were obtained at a clinical 3.0 T scanner using an in-house designed cubic phantom and optimized sequences for artifact evaluation (3D gradient echo sequence, multi-slice 2D turbo spin echo sequence). Reference sequences according to the American Society for Testing and Materials (ASTM) were additionally applied. Four representative active cardiac devices and a generic setup were analyzed regarding volume and shape of the signal void. For analysis, a threshold operation was applied to the grey value profile of each data set. The presented approach allows the evaluation of the signal void and shape even for larger implants such as ICDs. The void shape is influenced by the orientation of the B0-field and by the chosen sequence type. The distribution of ferromagnetic material within the implants also matters. The void volume depends both on the device itself, and on the sequence type. Disturbances in the B0 and B1 fields exceed the visual signal void. This work presents a reproducible and highly defined approach to characterize both signal void artifacts at 3.0 T and their influencing factors. KW - cardiac MRI KW - cardiac implants KW - signal voids Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-300502 VL - 12 IS - 1 ER - TY - JOUR A1 - Reiter, Theresa A1 - Demirbas, Senem A1 - Schmalzing, Marc A1 - Voelker, Wolfram A1 - Bauer, Wolfgang R. A1 - Güder, Gülmisal T1 - CMR detects extensive intracavitary thrombi as solitary clinical presentation of Antiphospholipid Syndrome: A case report JF - Clinical Case Reports N2 - Intracavitary thrombi are an important differential diagnosis of cardiac masses. Cardiac magnetic resonance imaging (CMR) allows their non-invasive characterization. This case highlights extensive cardiac thrombi detected by CMR as solitary presentation of antiphospholipid syndrome. KW - antiphospholipid syndrome KW - cardiac thrombi KW - CMR Y1 - 2022 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-312766 SN - 2050-0904 VL - 10 IS - 11 ER - TY - JOUR A1 - Schreiber, Laura M. A1 - Lohr, David A1 - Baltes, Steffen A1 - Vogel, Ulrich A1 - Elabyad, Ibrahim A. A1 - Bille, Maya A1 - Reiter, Theresa A1 - Kosmala, Aleksander A1 - Gassenmaier, Tobias A1 - Stefanescu, Maria R. A1 - Kollmann, Alena A1 - Aures, Julia A1 - Schnitter, Florian A1 - Pali, Mihaela A1 - Ueda, Yuichiro A1 - Williams, Tatiana A1 - Christa, Martin A1 - Hofmann, Ulrich A1 - Bauer, Wolfgang A1 - Gerull, Brenda A1 - Zernecke, Alma A1 - Ergün, Süleyman A1 - Terekhov, Maxim T1 - Ultra-high field cardiac MRI in large animals and humans for translational cardiovascular research JF - Frontiers in Cardiovascular Medicine N2 - A key step in translational cardiovascular research is the use of large animal models to better understand normal and abnormal physiology, to test drugs or interventions, or to perform studies which would be considered unethical in human subjects. Ultrahigh field magnetic resonance imaging (UHF-MRI) at 7 T field strength is becoming increasingly available for imaging of the heart and, when compared to clinically established field strengths, promises better image quality and image information content, more precise functional analysis, potentially new image contrasts, and as all in-vivo imaging techniques, a reduction of the number of animals per study because of the possibility to scan every animal repeatedly. We present here a solution to the dual use problem of whole-body UHF-MRI systems, which are typically installed in clinical environments, to both UHF-MRI in large animals and humans. Moreover, we provide evidence that in such a research infrastructure UHF-MRI, and ideally combined with a standard small-bore UHF-MRI system, can contribute to a variety of spatial scales in translational cardiovascular research: from cardiac organoids, Zebra fish and rodent hearts to large animal models such as pigs and humans. We present pilot data from serial CINE, late gadolinium enhancement, and susceptibility weighted UHF-MRI in a myocardial infarction model over eight weeks. In 14 pigs which were delivered from a breeding facility in a national SARS-CoV-2 hotspot, we found no infection in the incoming pigs. Human scanning using CINE and phase contrast flow measurements provided good image quality of the left and right ventricle. Agreement of functional analysis between CINE and phase contrast MRI was excellent. MRI in arrested hearts or excised vascular tissue for MRI-based histologic imaging, structural imaging of myofiber and vascular smooth muscle cell architecture using high-resolution diffusion tensor imaging, and UHF-MRI for monitoring free radicals as a surrogate for MRI of reactive oxygen species in studies of oxidative stress are demonstrated. We conclude that UHF-MRI has the potential to become an important precision imaging modality in translational cardiovascular research. KW - ultrahigh-field MRI KW - large animal models KW - translational research KW - research infrastructure KW - heart KW - organoid KW - pig KW - cardiovascular MRI Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-317398 SN - 2297-055X VL - 10 ER -