TY - JOUR A1 - Reiter, Theresa A1 - Gensler, Daniel A1 - Ritter, Oliver A1 - Weiss, Ingo A1 - Geistert, Wolfgang A1 - Kaufmann, Ralf A1 - Hoffmeister, Sabine A1 - Friedrich, Michael T. A1 - Wintzheimer, Stefan A1 - Düring, Markus A1 - Nordbeck, Peter A1 - Jakob, Peter M. A1 - Ladd, Mark E. A1 - Quick, Harald H. A1 - Bauer, Wolfgang R. T1 - Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures JF - Journal of Cardiovascular Magnetic Resonance N2 - Background: One of the safety concerns when performing electrophysiological (EP) procedures under magnetic resonance (MR) guidance is the risk of passive tissue heating due to the EP catheter being exposed to the radiofrequency (RF) field of the RF transmitting body coil. Ablation procedures that use catheters with irrigated tips are well established therapeutic options for the treatment of cardiac arrhythmias and when used in a modified mode might offer an additional system for suppressing passive catheter heating. Methods: A two-step approach was chosen. Firstly, tests on passive catheter heating were performed in a 1.5 T Avanto system (Siemens Healthcare Sector, Erlangen, Germany) using a ASTM Phantom in order to determine a possible maximum temperature rise. Secondly, a phantom was designed for simulation of the interface between blood and the vascular wall. The MR-RF induced temperature rise was simulated by catheter tip heating via a standard ablation generator. Power levels from 1 to 6 W were selected. Ablation duration was 120 s with no tip irrigation during the first 60 s and irrigation at rates from 2 ml/min to 35 ml/min for the remaining 60 s (Biotronik Qiona Pump, Berlin, Germany). The temperature was measured with fluoroscopic sensors (Luxtron, Santa Barbara, CA, USA) at a distance of 0 mm, 2 mm, 4 mm, and 6 mm from the catheter tip. Results: A maximum temperature rise of 22.4 degrees C at the catheter tip was documented in the MR scanner. This temperature rise is equivalent to the heating effect of an ablator's power output of 6 W at a contact force of the weight of 90 g (0.883 N). The catheter tip irrigation was able to limit the temperature rise to less than 2 degrees C for the majority of examined power levels, and for all examined power levels the residual temperature rise was less than 8 degrees C. Conclusion: Up to a maximum of 22.4 degrees C, the temperature rise at the tissue surface can be entirely suppressed by using the catheter's own irrigation system. The irrigated tip system can be used to increase MR safety of EP catheters by suppressing the effects of unwanted passive catheter heating due to RF exposure from the MR scanner. KW - EP Procedures KW - radiofrequency ablation KW - contact force KW - lesion size KW - MRI KW - temperature KW - tissue KW - wires KW - model KW - ablation KW - safety KW - catheter tip KW - MR guidance Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-134927 VL - 14 IS - 12 ER - TY - JOUR A1 - Parczyk, Marco A1 - Herold, Volker A1 - Klug, Gert A1 - Bauer, Wolfgang R. A1 - Rommel, Eberhard A1 - Jakob, Peter M. T1 - Regional in vivo transit time measurements of aortic pulse wave velocity in mice with high-field CMR at 17.6 Tesla N2 - Background: Transgenic mouse models are increasingly used to study the pathophysiology of human cardiovascular diseases. The aortic pulse wave velocity (PWV) is an indirect measure for vascular stiffness and a marker for cardiovascular risk. Results: This study presents a cardiovascular magnetic resonance (CMR) transit time (TT) method that allows the determination of the PWV in the descending murine aorta by analyzing blood flow waveforms. Systolic flow pulses were recorded with a temporal resolution of 1 ms applying phase velocity encoding. In a first step, the CMR method was validated by pressure waveform measurements on a pulsatile elastic vessel phantom. In a second step, the CMR method was applied to measure PWVs in a group of five eight-month-old apolipoprotein E deficient (ApoE(-/-)) mice and an age matched group of four C57Bl/6J mice. The ApoE(-/-) group had a higher mean PWV (PWV = 3.0 ± 0.6 m/s) than the C57Bl/6J group (PWV = 2.4 ± 0.4 m/s). The difference was statistically significant (p = 0.014). Conclusions: The findings of this study demonstrate that high field CMR is applicable to non-invasively determine and distinguish PWVs in the arterial system of healthy and diseased groups of mice. KW - Medizin Y1 - 2010 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-68219 ER - TY - JOUR A1 - Nordbeck, Peter A1 - Bönhof, Leoni A1 - Hiller, Karl-Heinz A1 - Voll, Sabine A1 - Arias-Loza, Paula A1 - Seidlmaier, Lea A1 - Williams, Tatjana A1 - Ye, Yu-Xiang A1 - Gensler, Daniel A1 - Pelzer, Theo A1 - Ertl, Georg A1 - Jakob, Peter M. A1 - Bauer, Wolfgang R. A1 - Ritter, Oliver T1 - Impact of Thoracic Surgery on Cardiac Morphology and Function in Small Animal Models of Heart Disease: A Cardiac MRI Study in Rats JF - PLoS ONE N2 - Background Surgical procedures in small animal models of heart disease might evoke alterations in cardiac morphology and function. The aim of this study was to reveal and quantify such potential artificial early or long term effects in vivo, which might account for a significant bias in basic cardiovascular research, and, therefore, could potentially question the meaning of respective studies. Methods Female Wistar rats (n = 6 per group) were matched for weight and assorted for sham left coronary artery ligation or control. Cardiac morphology and function was then investigated in vivo by cine magnetic resonance imaging at 7 Tesla 1 and 8 weeks after the surgical procedure. The time course of metabolic and inflammatory blood parameters was determined in addition. Results Compared to healthy controls, rats after sham surgery showed a lower body weight both 1 week (267.5±10.6 vs. 317.0±11.3 g, n<0.05) and 8 weeks (317.0±21.1 vs. 358.7±22.4 g, n<0.05) after the intervention. Left and right ventricular morphology and function were not different in absolute measures in both groups 1 week after surgery. However, there was a confined difference in several cardiac parameters normalized to the body weight (bw), such as myocardial mass (2.19±0.30/0.83±0.13 vs. 1.85±0.22/0.70±0.07 mg left/right per g bw, p<0.05), or enddiastolic ventricular volume (1.31±0.36/1.21±0.31 vs. 1.14±0.20/1.07±0.17 µl left/right per g bw, p<0.05). Vice versa, after 8 weeks, cardiac masses, volumes, and output showed a trend for lower values in sham operated rats compared to controls in absolute measures (782.2±57.2/260.2±33.2 vs. 805.9±84.8/310.4±48.5 mg, p<0.05 for left/right ventricular mass), but not normalized to body weight. Matching these findings, blood testing revealed only minor inflammatory but prolonged metabolic changes after surgery not related to cardiac disease. Conclusion Cardio-thoracic surgical procedures in experimental myocardial infarction cause distinct alterations upon the global integrity of the organism, which in the long term also induce circumscribed repercussions on cardiac morphology and function. This impact has to be considered when analyzing data from respective animal studies and transferring these findings to conditions in patients. KW - heart rate KW - body weight KW - surgical and invasive medical procedures KW - magnetic resonance imaging KW - blood KW - vascular surgery KW - myocardial infarction KW - cardiac ventricles Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-130064 VL - 8 IS - 8 ER - TY - JOUR A1 - Winter, Patrick A1 - Kampf, Thomas A1 - Helluy, Xavier A1 - Gutjahr, Fabian T. A1 - Meyer, Cord B. A1 - Rommel, Eberhard A1 - Bauer, Wolfgang R. A1 - Jakob, Peter M. A1 - Herold, Volker T1 - Fast retrospectively triggered local pulse-wave velocity measurements in mice with CMR-microscopy using a radial trajectory JF - Journal of Cardiovascular Magnetic Resonance N2 - Background The aortic pulse-wave velocity (PWV) is an important indicator of cardiovascular risk. In recent studies MRI methods have been developed to measure this parameter noninvasively in mice. Present techniques require additional hardware for cardiac and respiratory gating. In this work a robust self-gated measurement of the local PWV in mice without the need of triggering probes is proposed. Methods The local PWV of 6-months-old wild-type C57BL/6J mice (n=6) was measured in the abdominal aorta with a retrospectively triggered radial Phase Contrast (PC) MR sequence using the flow-area (QA) method. A navigator signal was extracted from the CMR data of highly asymmetric radial projections with short repetition time (TR=3 ms) and post-processed with high-pass and low-pass filters for retrospective cardiac and respiratory gating. The self-gating signal was used for a reconstruction of high-resolution Cine frames of the aortic motion. To assess the local PWV the volume flow Q and the cross-sectional area A of the aorta were determined. The results were compared with the values measured with a triggered Cartesian and an undersampled triggered radial PC-Cine sequence. Results In all examined animals a self-gating signal could be extracted and used for retrospective breath-gating and PC-Cine reconstruction. With the non-triggered measurement PWV values of 2.3±0.2 m/s were determined. These values are in agreement with those measured with the triggered Cartesian (2.4±0.2 m/s) and the triggered radial (2.3±0.2 m/s) measurement. Due to the strong robustness of the radial trajectory against undersampling an acceleration of more than two relative to the prospectively triggered Cartesian sampling could be achieved with the retrospective method. Conclusion With the radial flow-encoding sequence the extraction of a self-gating signal is feasible. The retrospective method enables a robust and fast measurement of the local PWV without the need of additional trigger hardware. KW - pulse-wave velocity KW - mouse KW - self-gating KW - phase-contrast CMR KW - non-triggered KW - retrospective KW - radial KW - aorta Y1 - 2013 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-96602 UR - http://jcmr-online.com/content/15/1/88 ER - TY - THES A1 - Bauer, Wolfgang Rudolf T1 - Analytische Näherungsverfahren zur Beschreibung der nuklearen Spin-Dephasierung T1 - Analytical Approaches for the Description of Nuclear Spin Dephasing N2 - Die Dynamik der Kernspindephasierung in lebenden Systemen enhält relevante Informationen über biologisch wichtige Parameter, wie Sauerstoffversorgung, Mikrozirkulation, Diffusion etc.. Ursächlich für die Dephasierung sind Interaktionen des Spins mit fluktuierenden Magnetfeldern. Notwendig sind also Modelle, welche diese Interaktionen mit den biologisch relevanten Parametern in Beziehung setzen. Problematisch ist, daß fast alle analytische Ansätze nur in extremen Dynamikbereichen der Störfeldfluktuationen (motional narrowing - , static dephasing limit) gültig sind. In dieser Arbeit zeigen wir einen Ansatz, mit dem man die Dynamik der Störfeldfluktuationen erheblich vereinfachen und trotzdem noch deren wesentliche Eigenschaften beibehalten kann. Dieser Ansatz ist nicht auf einen speziellen Dynamikbereich festgelegt. Angewendet wird dieses Näherungsverfahren zur Beschreibung der Spin Dephasierung im Herzmuskel. Die Relaxationszeiten erhält man als Funktion der Kapillardichte und Blutoxygenierung. Vergleiche mit numerisch errechneten Daten anderer, eigenen Messungen am menschlichen Herzen und experimentellen Befunden in der Literatur, bestätigen die theoretischen Vorhersagen. N2 - The dynamics of nuclear spin dephasing in living objects contains relevant information about important parameters as oxygen supply, microcirculation, and diffusion. Spin dephasing is induced by interaction of spins with fluctuating perturbation fields. Desirable are models which relate these interactions to the biological parameters. Unfortunately most analytical approaches are restricted to certain motion regimes, e.g. the motional narrowing or static dephasing limit. In this work we present an analytical approach, which simplifies the perturbation field dynamics but still conserves its relevant properties. This approach is not restricted to any motion regime. As an application we describe spin dephasing in the cardiac muscle. We obtain the relaxation time as a function of capillary density and blood oxygenation. Data are in excellent agreement with numerical simulations of others, own measurements in humans, and experimental data in the literature. KW - Biologisches System KW - NMR-Bildgebung KW - Spinrelaxation KW - Näherungsverfahren KW - Spin Relaxation KW - Magnetresonanz KW - MR KW - Herz KW - Durchblutung KW - spin dephasing KW - magnetic resonance KW - perfusion KW - T2 KW - cardiac imaging Y1 - 2002 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-4674 ER - TY - JOUR A1 - Gotschy, Alexander A1 - Bauer, Wolfgang R. A1 - Winter, Patrick A1 - Nordbeck, Peter A1 - Rommel, Eberhard A1 - Jakob, Peter M. A1 - Herold, Volker T1 - Local versus global aortic pulse wave velocity in early atherosclerosis: An animal study in ApoE\(^{-/-}\) mice using ultrahigh field MRI JF - PLoS ONE N2 - Increased aortic stiffness is known to be associated with atherosclerosis and has a predictive value for cardiovascular events. This study aims to investigate the local distribution of early arterial stiffening due to initial atherosclerotic lesions. Therefore, global and local pulse wave velocity (PWV) were measured in ApoE\(^{-/-}\) and wild type (WT) mice using ultrahigh field MRI. For quantification of global aortic stiffness, a new multi-point transit-time (TT) method was implemented and validated to determine the global PWV in the murine aorta. Local aortic stiffness was measured by assessing the local PWV in the upper abdominal aorta, using the flow/area (QA) method. Significant differences between age matched ApoE\(^{-/-}\) and WT mice were determined for global and local PWV measurements (global PWV: ApoE\(^{-/-}\): 2.7 ±0.2m/s vs WT: 2.1±0.2m/s, P<0.03; local PWV: ApoE\(^{-/-}\): 2.9±0.2m/s vs WT: 2.2±0.2m/s, P<0.03). Within the WT mouse group, the global PWV correlated well with the local PWV in the upper abdominal aorta (R\(^2\) = 0.75, P<0.01), implying a widely uniform arterial elasticity. In ApoE\(^{-/-}\) animals, however, no significant correlation between individual local and global PWV was present (R\(^2\) = 0.07, P = 0.53), implying a heterogeneous distribution of vascular stiffening in early atherosclerosis. The assessment of global PWV using the new multi-point TT measurement technique was validated against a pressure wire measurement in a vessel phantom and showed excellent agreement. The experimental results demonstrate that vascular stiffening caused by early atherosclerosis is unequally distributed over the length of large vessels. This finding implies that assessing heterogeneity of arterial stiffness by multiple local measurements of PWV might be more sensitive than global PWV to identify early atherosclerotic lesions. KW - MRI KW - Atherosclerosis KW - Aorta KW - Stiffness KW - Measurement KW - Time measurement KW - Magnetic resonance imaging KW - Mouse models KW - Systole Y1 - 2017 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-171824 VL - 12 IS - 2 ER - TY - JOUR A1 - Winter, Patrick M. A1 - Andelovic, Kristina A1 - Kampf, Thomas A1 - Hansmann, Jan A1 - Jakob, Peter Michael A1 - Bauer, Wolfgang Rudolf A1 - Zernecke, Alma A1 - Herold, Volker T1 - Simultaneous measurements of 3D wall shear stress and pulse wave velocity in the murine aortic arch JF - Journal of Cardiovascular Magnetic Resonance N2 - Purpose Wall shear stress (WSS) and pulse wave velocity (PWV) are important parameters to characterize blood flow in the vessel wall. Their quantification with flow-sensitive phase-contrast (PC) cardiovascular magnetic resonance (CMR), however, is time-consuming. Furthermore, the measurement of WSS requires high spatial resolution, whereas high temporal resolution is necessary for PWV measurements. For these reasons, PWV and WSS are challenging to measure in one CMR session, making it difficult to directly compare these parameters. By using a retrospective approach with a flexible reconstruction framework, we here aimed to simultaneously assess both PWV and WSS in the murine aortic arch from the same 4D flow measurement. Methods Flow was measured in the aortic arch of 18-week-old wildtype (n = 5) and ApoE\(^{−/−}\) mice (n = 5) with a self-navigated radial 4D-PC-CMR sequence. Retrospective data analysis was used to reconstruct the same dataset either at low spatial and high temporal resolution (PWV analysis) or high spatial and low temporal resolution (WSS analysis). To assess WSS, the aortic lumen was labeled by semi-automatically segmenting the reconstruction with high spatial resolution. WSS was determined from the spatial velocity gradients at the lumen surface. For calculation of the PWV, segmentation data was interpolated along the temporal dimension. Subsequently, PWV was quantified from the through-plane flow data using the multiple-points transit-time method. Reconstructions with varying frame rates and spatial resolutions were performed to investigate the influence of spatiotemporal resolution on the PWV and WSS quantification. Results 4D flow measurements were conducted in an acquisition time of only 35 min. Increased peak flow and peak WSS values and lower errors in PWV estimation were observed in the reconstructions with high temporal resolution. Aortic PWV was significantly increased in ApoE\(^{−/−}\) mice compared to the control group (1.7 ± 0.2 versus 2.6 ± 0.2 m/s, p < 0.001). Mean WSS magnitude values averaged over the aortic arch were (1.17 ± 0.07) N/m\(^2\) in wildtype mice and (1.27 ± 0.10) N/m\(^2\) in ApoE\(^{−/−}\) mice. Conclusion The post processing algorithm using the flexible reconstruction framework developed in this study permitted quantification of global PWV and 3D-WSS in a single acquisition. The possibility to assess both parameters in only 35 min will markedly improve the analyses and information content of in vivo measurements. KW - 4D flow KW - pulse wave velocity KW - wall shear stress KW - radial KW - self-navigation KW - mouse KW - aortic arch KW - atherosclerosis KW - mice KW - flow KW - plaque KW - CMR KW - quantification KW - microscopy Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-259152 VL - 23 IS - 1 ER - TY - JOUR A1 - Andelovic, Kristina A1 - Winter, Patrick A1 - Kampf, Thomas A1 - Xu, Anton A1 - Jakob, Peter Michael A1 - Herold, Volker A1 - Bauer, Wolfgang Rudolf A1 - Zernecke, Alma T1 - 2D Projection Maps of WSS and OSI Reveal Distinct Spatiotemporal Changes in Hemodynamics in the Murine Aorta during Ageing and Atherosclerosis JF - Biomedicines N2 - Growth, ageing and atherosclerotic plaque development alter the biomechanical forces acting on the vessel wall. However, monitoring the detailed local changes in wall shear stress (WSS) at distinct sites of the murine aortic arch over time has been challenging. Here, we studied the temporal and spatial changes in flow, WSS, oscillatory shear index (OSI) and elastic properties of healthy wildtype (WT, n = 5) and atherosclerotic apolipoprotein E-deficient (Apoe\(^{−/−}\), n = 6) mice during ageing and atherosclerosis using high-resolution 4D flow magnetic resonance imaging (MRI). Spatially resolved 2D projection maps of WSS and OSI of the complete aortic arch were generated, allowing the pixel-wise statistical analysis of inter- and intragroup hemodynamic changes over time and local correlations between WSS, pulse wave velocity (PWV), plaque and vessel wall characteristics. The study revealed converse differences of local hemodynamic profiles in healthy WT and atherosclerotic Apoe\(^{−/−}\) mice, and we identified the circumferential WSS as potential marker of plaque size and composition in advanced atherosclerosis and the radial strain as a potential marker for vascular elasticity. Two-dimensional (2D) projection maps of WSS and OSI, including statistical analysis provide a powerful tool to monitor local aortic hemodynamics during ageing and atherosclerosis. The correlation of spatially resolved hemodynamics and plaque characteristics could significantly improve our understanding of the impact of hemodynamics on atherosclerosis, which may be key to understand plaque progression towards vulnerability. KW - atherosclerosis KW - mouse KW - 4D flow MRI KW - aortic arch KW - flow dynamics KW - WSS KW - mapping KW - PWV KW - plaque characteristics Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-252164 SN - 2227-9059 VL - 9 IS - 12 ER - TY - JOUR A1 - Riedl, Katharina A. A1 - Kampf, Thomas A1 - Herold, Volker A1 - Behr, Volker C. A1 - Bauer, Wolfgang R. T1 - Wall shear stress analysis using 17.6 Tesla MRI: A longitudinal study in ApoE\(^{-/-}\)mice with histological analysis JF - PLoS One N2 - This longitudinal study was performed to evaluate the feasibility of detecting the interaction between wall shear stress (WSS) and plaque development. 20 ApoE\(^{-/-}\)mice were separated in 12 mice with Western Diet and 8 mice with Chow Diet. Magnetic resonance (MR) scans at 17.6 Tesla and histological analysis were performed after one week, eight and twelve weeks. Allin vivoMR measurements were acquired using a flow sensitive phase contrast method for determining vectorial flow. Histological sections were stained with Hematoxylin and Eosin, Elastica van Gieson and CD68 staining. Data analysis was performed using Ensight and a Matlab-based "Flow Tool". The body weight of ApoE\(^{-/-}\)mice increased significantly over 12 weeks. WSS values increased in the Western Diet group over the time period; in contrast, in the Chow Diet group the values decreased from the first to the second measurement point. Western Diet mice showed small plaque formations with elastin fragmentations after 8 weeks and big plaque formations after 12 weeks; Chow Diet mice showed a few elastin fragmentations after 8 weeks and small plaque formations after 12 weeks. Favored by high-fat diet, plaque formation results in higher values of WSS. With wall shear stress being a known predictor for atherosclerotic plaque development, ultra highfield MRI can serve as a tool for studying the causes and beginnings of atherosclerosis. KW - phase-contrast MRI KW - flow patterns KW - blood flow KW - apolipoprotein-E KW - atheriosclerosis KW - mouse KW - mice KW - quantification KW - association KW - lesions Y1 - 2020 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-229318 VL - 15 IS - 8 ER - TY - JOUR A1 - Andelovic, Kristina A1 - Winter, Patrick A1 - Jakob, Peter Michael A1 - Bauer, Wolfgang Rudolf A1 - Herold, Volker A1 - Zernecke, Alma T1 - Evaluation of plaque characteristics and inflammation using magnetic resonance imaging JF - Biomedicines N2 - Atherosclerosis is an inflammatory disease of large and medium-sized arteries, characterized by the growth of atherosclerotic lesions (plaques). These plaques often develop at inner curvatures of arteries, branchpoints, and bifurcations, where the endothelial wall shear stress is low and oscillatory. In conjunction with other processes such as lipid deposition, biomechanical factors lead to local vascular inflammation and plaque growth. There is also evidence that low and oscillatory shear stress contribute to arterial remodeling, entailing a loss in arterial elasticity and, therefore, an increased pulse-wave velocity. Although altered shear stress profiles, elasticity and inflammation are closely intertwined and critical for plaque growth, preclinical and clinical investigations for atherosclerosis mostly focus on the investigation of one of these parameters only due to the experimental limitations. However, cardiovascular magnetic resonance imaging (MRI) has been demonstrated to be a potent tool which can be used to provide insights into a large range of biological parameters in one experimental session. It enables the evaluation of the dynamic process of atherosclerotic lesion formation without the need for harmful radiation. Flow-sensitive MRI provides the assessment of hemodynamic parameters such as wall shear stress and pulse wave velocity which may replace invasive and radiation-based techniques for imaging of the vascular function and the characterization of early plaque development. In combination with inflammation imaging, the analyses and correlations of these parameters could not only significantly advance basic preclinical investigations of atherosclerotic lesion formation and progression, but also the diagnostic clinical evaluation for early identification of high-risk plaques, which are prone to rupture. In this review, we summarize the key applications of magnetic resonance imaging for the evaluation of plaque characteristics through flow sensitive and morphological measurements. The simultaneous measurements of functional and structural parameters will further preclinical research on atherosclerosis and has the potential to fundamentally improve the detection of inflammation and vulnerable plaques in patients. KW - atherosclerosis KW - mouse models KW - wall shear stress KW - pulse wave velocity KW - arterial elasticity KW - inflammation KW - magnetic resonance imaging Y1 - 2021 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-228839 SN - 2227-9059 VL - 9 IS - 2 ER -