• Treffer 2 von 7
Zurück zur Trefferliste

Custom Made Monoflange Acetabular Components for the Treatment of Paprosky Type III Defects

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-236513
  • Purpose: Patient-specific, flanged acetabular components are used for the treatment of Paprosky type III defects during revision total hip arthroplasty (THA). This monocentric retrospective cohort study analyzes the outcome of patients treated with custom made monoflanged acetabular components (CMACs) with intra- and extramedullary iliac fixation. Methods: 14 patients were included who underwent revision THA with CMACs for the treatment of Paprosky type III defects. Mechanism of THA failure was infection in 4 and aseptic loosening in 10Purpose: Patient-specific, flanged acetabular components are used for the treatment of Paprosky type III defects during revision total hip arthroplasty (THA). This monocentric retrospective cohort study analyzes the outcome of patients treated with custom made monoflanged acetabular components (CMACs) with intra- and extramedullary iliac fixation. Methods: 14 patients were included who underwent revision THA with CMACs for the treatment of Paprosky type III defects. Mechanism of THA failure was infection in 4 and aseptic loosening in 10 patients. Seven patients underwent no previous revision, the other seven patients underwent three or more previous revisions. Results: At a mean follow-up of 35.4 months (14–94), the revision rate of the implant was 28.3%. Additionally, one perioperative dislocation and one superficial wound infection occurred. At one year postoperatively, we found a significant improvement of the Western Ontario and McMaster Universities Arthritis Index (WOMAC) score (p = 0.015). Postoperative radiographic analysis revealed good hip joint reconstruction with a mean leg length discrepancy of 3 mm (−8–20), a mean lateralization of the horizontal hip center of rotation of 8 mm (−8–35), and a mean proximalization of the vertical hip center of rotation of 6 mm (13–26). Radiolucency lines were present in 30%. Conclusion: CMACs can be considered an option for the treatment of acetabular bone loss in revision THA. Iliac intra- and extramedullary fixation allows soft tissue-adjusted hip joint reconstruction and improves hip function. However, failure rates are high, with periprosthetic infection being the main threat to successful outcome.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Sebastian Philipp von Hertzberg-Boelch, Mike Wagenbrenner, Jörg Arnholdt, Stephan Frenzel, Boris Michael Holzapfel, Maximilian Rudert
URN:urn:nbn:de:bvb:20-opus-236513
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Lehrstuhl für Orthopädie
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Journal of Personalized Medicine
ISSN:2075-4426
Erscheinungsjahr:2021
Band / Jahrgang:11
Heft / Ausgabe:4
Aufsatznummer:283
Originalveröffentlichung / Quelle:Journal of Personalized Medicine 2021, 11(4), 283; https://doi.org/10.3390/jpm11040283
DOI:https://doi.org/10.3390/jpm11040283
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):Paprosky; custom made implant; patient specific implant; pelvic discontinuity; revision hip
Datum der Freischaltung:05.01.2022
Datum der Erstveröffentlichung:08.04.2021
Open-Access-Publikationsfonds / Förderzeitraum 2021
Lizenz (Deutsch):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International