• Treffer 5 von 11
Zurück zur Trefferliste

Measurement of tibial orientation helps select the optimal insert thickness to personalize PCL tension in a medial ball-in-socket TKA

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-286232
  • As the conformity of a medial ball-in-socket total knee arthroplasty (TKA) provides intrinsic anterior-posterior (A-P) stability, surgeons cannot rely on the manual examination of sagittal laxity to identify the optimal insert thickness. Instead, the present study determined whether measuring tibial axial orientation in extension and 90° flexion with an insert goniometer could identify the optimal thickness that, when implanted, provides high postoperative function. In twenty-two patients that underwent unrestricted caliper-verified kinematicAs the conformity of a medial ball-in-socket total knee arthroplasty (TKA) provides intrinsic anterior-posterior (A-P) stability, surgeons cannot rely on the manual examination of sagittal laxity to identify the optimal insert thickness. Instead, the present study determined whether measuring tibial axial orientation in extension and 90° flexion with an insert goniometer could identify the optimal thickness that, when implanted, provides high postoperative function. In twenty-two patients that underwent unrestricted caliper-verified kinematic alignment (KA) with a PCL retaining implant, two surgeons measured tibial orientation in extension and 90° flexion with 10, 11, 12, and 13 mm thick insert goniometers. Each TKA had one insert thickness that restored either the maximum external tibial orientation in extension, the maximum internal tibial orientation at 90° flexion, or both relative to 1 mm thinner and thicker inserts. In addition, the 6-month median [interquartile range] Forgotten Joint Score of 73 (54–87) and Oxford Knee Score of 42 (38–45) indicated high satisfaction and function. In conclusion, surgeons using a medial ball-in-socket TKA design can measure external tibial orientation in extension and internal tibial orientation at 90° flexion with an insert goniometer. Furthermore, implanting an insert with the thickness that provided the maximum orientation values resulted in high postoperative function, thereby personalizing PCL tension.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Alexander J. Nedopil, Stephen M. Howell, Maury L. Hull
URN:urn:nbn:de:bvb:20-opus-286232
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Lehrstuhl für Orthopädie
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Journal of Personalized Medicine
ISSN:2075-4426
Erscheinungsjahr:2022
Band / Jahrgang:12
Heft / Ausgabe:9
Aufsatznummer:1427
Originalveröffentlichung / Quelle:Journal of Personalized Medicine (2022) 12:9, 1427. https://doi.org/10.3390/jpm12091427
DOI:https://doi.org/10.3390/jpm12091427
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):kinematic alignment; medial pivot; posterior cruciate ligament; tibial rotation; total knee arthroplasty
Datum der Freischaltung:24.08.2023
Datum der Erstveröffentlichung:31.08.2022
Lizenz (Deutsch):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International