The search result changed since you submitted your search request. Documents might be displayed in a different sort order.
  • search hit 3 of 7
Back to Result List

Wertigkeit der Power-Doppler Sonographie bei der Erkennung von periprothetischen Infektionen

Value of the Power Doppler Sonography in the detection of periprosthetic infection

Please always quote using this URN: urn:nbn:de:bvb:20-opus-66194
  • Ziel dieser Studie war die Beurteilung der diagnostischen Wertigkeit der PDS bei endoprothetischen Wechseloperationen von Hüft- und Kniegelenk. Dabei stand im Vordergrund die vergleichende Betrachtung der Darstellung der periprothetischen Synovialitis mittels PDS, der histologischen Beurteilung entsprechend der Konsensus-Klassifikation nach Morawietz und der Hygiene. Ein weiterer Aspekt war die Prüfung der Reliabilität der neu etablierten Konsensus- Klassifikation nach Morawietz. Es wurden dazu 83 Patienten, 33 Männer und 50 Frauen, mit einemZiel dieser Studie war die Beurteilung der diagnostischen Wertigkeit der PDS bei endoprothetischen Wechseloperationen von Hüft- und Kniegelenk. Dabei stand im Vordergrund die vergleichende Betrachtung der Darstellung der periprothetischen Synovialitis mittels PDS, der histologischen Beurteilung entsprechend der Konsensus-Klassifikation nach Morawietz und der Hygiene. Ein weiterer Aspekt war die Prüfung der Reliabilität der neu etablierten Konsensus- Klassifikation nach Morawietz. Es wurden dazu 83 Patienten, 33 Männer und 50 Frauen, mit einem Durchschnittsalter von 70 Jahren (43 – 88 Jahre) jeweils vor endoprothetischer Revision mittels PDS untersucht sowie eine präoperative Labordiagnostik und intraoperative Probenentnahme zur Bestimmung der Hygiene und histologischen Beurteilung nach Morawietz durchgeführt. Es folgte ein Vergleich der PDS mit den Ergebnissen aus der Histologie und den intraoperativ gewonnenen Hygienebefunden. In dieser Studie zeigte sich keine signifikante Korrelation zwischen der PDS und den histologischen Ergebnissen. Auch konnte kein signifikanter Zusammenhang zwischen der PDS und Infektionen mit septischer Genese (intraoperativer Keimnachweis, erhöhte Entzündungsparameter) ermittelt werden. Somit erweist sich in dieser Studie, dass die PDS keinen geeigneten Beitrag im Verfahren zur präoperativen Diagnostik von endoprothetischen Revisionen liefert. Am höchsten fiel der positive Prädiktivwert in unserer Studie beim Vergleich von histologischen Ergebnissen und mikrobiologischen Ergebnissen aus. Eine Untersuchung zum Einsatz von kontrastmittel-verstärkter PDS in der Beurteilung von endoprothetischen Revisionen wäre sinnvoll. Nahezu identische Ergebnisse brachte der Vergleich unserer histopathologischen Daten der Konsensus-Klassifikation mit den von Morawietz et al. publizierten Daten. Diese Ergebnisse sprechen für eine gute Festlegung der Kriterien von Morawietz et al. und damit einer reliablen Methode zur histopathologischen Auswertung der Synovialmembran.show moreshow less
  • The aim of this study was to evaluate the diagnostic value of the PDS at endoprosthetic replacement surgery of hip and knee. The main focus was a comparative study of the representation of the periprosthetic synovitis by PDS, the histological assessment according to the consensus classification by Morawietz and hygiene. Another aspect was to examine the reliability of the newly established consensus Classification by Morawietz. We had 83 patients, 33 men and 50 women, with an average age of 70 years (43 - 88 years) examined both beforeThe aim of this study was to evaluate the diagnostic value of the PDS at endoprosthetic replacement surgery of hip and knee. The main focus was a comparative study of the representation of the periprosthetic synovitis by PDS, the histological assessment according to the consensus classification by Morawietz and hygiene. Another aspect was to examine the reliability of the newly established consensus Classification by Morawietz. We had 83 patients, 33 men and 50 women, with an average age of 70 years (43 - 88 years) examined both before endoprosthetic revision by PDS, and preoperative laboratory and diagnostic, intraoperative sampling to determine the hygienic and histological assessment conducted by Morawietz. It was followed by a comparison of PDS with the results of the histology and the intra-operatively acquired Hygiene findings. In this study we couldn't find a significant correlation between the PDS and the histological results. Also, no significant correlation between the PDS and infections with septic causes (intraoperative pathogen detection, elevated inflammatory parameters to be determined). So we found out in this study, the PDS no suitable post in the process of preoperative diagnosis of endoprosthetic revisions supplies. Most significant was the positive predictive value in our study when comparing results of histological and microbiological findings. An investigation on the use of contrast medium-enhanced PDS in the assessment revisions of total hip and knee replacement would be useful. Nearly identical results brought the comparison of our histopathological data of the consensus classification of Morawietz et al. with published data. These results indicate a good definition of the criteria by Morawietz et al. and thus a reliable method for histopathological evaluation of the synovium.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Karoline Emge
URN:urn:nbn:de:bvb:20-opus-66194
Document Type:Doctoral Thesis
Granting Institution:Universität Würzburg, Medizinische Fakultät
Faculties:Medizinische Fakultät / Lehrstuhl für Orthopädie
Date of final exam:2011/11/11
Language:German
Year of Completion:2011
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
GND Keyword:Doppler-Sonographie; Infektion; Bakterielle Infektion; Ultraschalldiagnostik
Tag:Konsensus-Klassfikation; periprothetische Infektion
Morawietz; Power Doppler Sonography; consensus classification; periprosthetic infection; value
Release Date:2011/12/05
Advisor:PD Dr. Bernd Baumann
Licence (German):License LogoDeutsches Urheberrecht