Diagnostische Relevanz der Impuls-Oszillometrie im Vergleich zur Bodyplethysmographie im Kindesalter

Diagnostic relevance of impulse oscillometry in comparison with bodyplethysmography in childhood

Please always quote using this URN: urn:nbn:de:bvb:20-opus-2906
  • Die Impuls-Oszillometrie (IOS), die Ganzkörperplethysmographie (BP) und teilweise die Spirometrie wurden in der vorliegenden Arbeit umfassend beurteilt. Die Reliabilität der drei Methoden wurde durch Wiederholungsmessungen an zwölf gesunden Probanden überprüft. Ergebnisse der IOS wurden mit der BP korreliert. Anhand 87 atemwegsgesunder Kinder zwischen neun und elf Jahren erfolgte eine Normwerterstellung für die IOS und die BP. Die gewonnenen Referenzwerte wurden auf 38 Kinder mit Asthma bronchiale im gleichen Altersbereich angewandt. DieDie Impuls-Oszillometrie (IOS), die Ganzkörperplethysmographie (BP) und teilweise die Spirometrie wurden in der vorliegenden Arbeit umfassend beurteilt. Die Reliabilität der drei Methoden wurde durch Wiederholungsmessungen an zwölf gesunden Probanden überprüft. Ergebnisse der IOS wurden mit der BP korreliert. Anhand 87 atemwegsgesunder Kinder zwischen neun und elf Jahren erfolgte eine Normwerterstellung für die IOS und die BP. Die gewonnenen Referenzwerte wurden auf 38 Kinder mit Asthma bronchiale im gleichen Altersbereich angewandt. Die Validität des IOS und der BP wurde mittels Sensitivitätsindices und Diskriminanzanalyse (Sensitivität, Spezifität) miteinander verglichen. In der Wiederholbarkeit (intraindividuelle Variabilität) lag die Spirometrie mit Variationskoeffizienten (CV) von z. B. 2,0% für VCmax und 3,7% für FEV1 an der Spitze. Die Parameter der Fluss - Volumen - Messung erreichten CV um 10%. Für die BP errechneten sich 5,0% für das ITGV und 11,6% bzw. 13,8% für den absoluten bzw. spezifischen effektiven Atemwegswiderstand. Sämtliche anderen untersuchten Bestimmungsmethoden (tot, Mitte, 0,5, peak, in, ex) erreichen schlechtere Variationskoeffizienten bis zu 23,7% für R0,5. Die Widerstände der IOS erzielten mit einem CV von 10,4% für die Resistance bei 5 und 10 Hz und 10,6% für die Impedanz geringfügig bessere Ergebnisse als die Widerstände der BP. Der CV der Resonanzfrequenz war 9,5%. Für die Reaktanzen ist eine Berechnung des Variationskoeffizienten auf Grund des Betrags der Reaktanz um Null nicht sinnvoll. Die Standardabweichungen für die Reaktanzen lagen jedoch etwas unter den entsprechenden Resistanzen. Die in dieser Arbeit eingeführten Differenzparameter Zrespir-Z15 und R5-R10 erreichten mit 21,9% und 30,9% eine hohe intraindividuelle Variabilität. Oszillatorische und bodyplethysmographische Widerstände korrelierten insgesamt gut miteinander. Insbesondere R5, X10 und R5-R10 erreichten mit sämtlichen effektiven Atemwegswiderständen hohe Korrelationskoeffizienten um 0,8. Die Referenzwerterstellung mittels eines multiplen, linearen Modells erfolgte für beide Geschlechter getrennt anhand von Größe, Gewicht und Alter, soweit diese Faktoren Einfluss auf die Regressionsfunktion hatten. Eine Normwertfunktion für die Resonanzfrequenz war aufgrund der hohen interindividuellen Variabilität und die Festlegung der Resonanzfrequenz durch die Reaktanz nicht sinnvoll. Sowohl für Wiederholungsmessungen (Verlaufsbeobachtung) als auch für die Primärdiagnostik anhand von Normwerten wurden für die hier eingeführte Differenz Zrespir-Z15 die besten Ergebnisse erzielt. Der Vorsprung gegenüber Zrespir und sReff war gering. Die Verwendung der Referenzwerte für die spezifischen Widerstände ist jedoch trotz der guten Ergebnisse durch die hohe interindividuelle Variabilität eingeschränkt. Für sämtliche effektiven Atemwegswiderstände, R5 und dessen Differenzparameter R5-R15 wurden ebenfalls gute Ergebnisse erzielt. In der Verlaufsbeobachtung verliert die Variabilität der Normwerte gegenüber der intraindividuellen Variabilität an Bedeutung. Bei den Wiederholungsmessungen wurden die spezifischen Atemwegswiderstände und die Reaktanzen (v. a. X10) aussagekräftiger. Somit ergeben sich geringe Vorteile der IOS gegenüber der BP in der Validität und Objektivität. In der Objektivität deshalb, da das plethysmographische Messergebnis durch eine BTPS - Korrektur beeinflussbar ist. Die IOS ist für Kinder angenehmer und einfacher durchzuführen. Aufgrund des günstigeren Anschaffungspreises und der kurzen Untersuchungszeit ist die IOS wirtschaftlicher. Ebenbürtig sind sich beide Verfahren in der Reliabilität (Zuverlässigkeit). Mittels BP und Heliumeinwaschmethode kann jedoch im Gegensatz zur IOS „trapped air“ exakt bestimmt werden.show moreshow less
  • Impulse oscillometry (IOS), bodyplethysmography (BP) and partly the spirometry were compared in the present study. Reliability of the three methods was checked by repetition measurements of twelve healthy children. Results of the IOS were correlated with BP. On the basis of 87 breath-way-healthy children between nine and eleven years reference values were calculated for IOS and BP. The reference values were applied to 38 children with asthma within the same age. The validity of the IOS and the BP was compared by means of sensitivity indices,Impulse oscillometry (IOS), bodyplethysmography (BP) and partly the spirometry were compared in the present study. Reliability of the three methods was checked by repetition measurements of twelve healthy children. Results of the IOS were correlated with BP. On the basis of 87 breath-way-healthy children between nine and eleven years reference values were calculated for IOS and BP. The reference values were applied to 38 children with asthma within the same age. The validity of the IOS and the BP was compared by means of sensitivity indices, sensitivity and specificity. In the repeatability (intraindividual variability) the spirometry with coefficient of variation (CV) of e.g. 2.0% for VCmax and 3.7% for FEV1 were at the top. The parameters of the flow - volume - measurement achieved CV around 10%. For the BP 5.0% for the ITGV and 11.6% / 13,8% for the absolute / specific effective airway resistance were calculated. All other examined methods (Rtot, Rmid, R0.5, Rpeak, Rin, Rex) achieved worse CV up to 23,7% for R0.5. Resistances of the IOS (RIOS) obtained with a CV of 10.6% for impedance Z and 10,4% for RIOS at 5 and 10 Hz slightly better results as for the BP. The CV of the resonant frequency was 9,5%. The standard deviations for the reactances X were slightly under the appropriate resistances. The difference parameters Zrespir-Z15 and R5-R10 (introduced in this study) achieved a high intraindividual variability with 21,9% and 30,9%. Oscillatory and bodyplethysmographic resistances correlated altogether high. In particular R5, X10 and R5-R10 achieved high coefficients of correlation with all effective airway resistances (Reff) around 0,8. The reference value creation by means of a multiple, linear model took place for both sexes separately on the basis from size, weight and age, as far as these factors had influence on the regression function. A standard value function for the resonant frequency was not meaningful due to the high interindividual variability and the definition of the resonant frequency by the reactance. Both for repetition measurements (process observation) and for the primary diagnostics on the basis of reference values the difference impedance Zrespir-Z15 (introduced in this study) obtained the best results. The lead to Zrespir and specific airway resistance was small. The use of the reference values for specific airway resistance is however reduced by its high interindividual variability. For all effective airway resistances, R5 and its difference parameter R5-R15 were obtained likewise good results. In the process observation the variability of the reference values loses importance in relation to the intraindividuel variability. With the repetition measurements specific airway resistances and the reactances (especially X10) became more important. Thus small advantages result for impulse oscillometry in validity and objectivity. In the objectivity, because the plethysmographic result is influenced by a BTPS – correction. The IOS is more pleasantly and more simply to execute and thus even suitable for younger children. Due to the lower price and the short investigation time the IOS is more economical. Both procedures are equal in the reliability. By means of BP and helium dilution method however trapped air can be determined accurately in contrast to IOS.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Alexander Kraus
URN:urn:nbn:de:bvb:20-opus-2906
Document Type:Doctoral Thesis
Granting Institution:Universität Würzburg, Medizinische Fakultät
Faculties:Medizinische Fakultät / Kinderklinik und Poliklinik
Date of final exam:2002/06/04
Language:German
Year of Completion:1999
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Tag:Atemwegswiderstand; Bodyplethysmographie; IOS; Impedanz; Impuls-Oszillometrie; Normwerte für Kinder; Reaktanz; Resistanz
IOS; airway resistance; bodyplethysmography; impedance; impulse oscillometry; reactance; reference values for children; resistance
Release Date:2002/09/02
Advisor:Prof. Dr. med. H. W. Kreth