The search result changed since you submitted your search request. Documents might be displayed in a different sort order.
  • search hit 15 of 252
Back to Result List

Highly precise navigation at the lateral skull base by the combination of flat-panel volume CT and electromagnetic navigation

Please always quote using this URN: urn:nbn:de:bvb:20-opus-250268
  • This study aimed to evaluate the feasibility and accuracy of electromagnetic navigation at the lateral skull base in combination with flat panel volume computed tomography (fpVCT) datasets. A mastoidectomy and a posterior tympanotomy were performed on 10 samples of fresh frozen temporal bones. For registration, four self-drilling titanium screws were applied as fiducial markers. Multi-slice computed tomography (MSCT; 600 µm), conventional flat panel volume computed tomography (fpVCT; 466 µm), micro-fpVCT (197 µm) and secondary reconstructedThis study aimed to evaluate the feasibility and accuracy of electromagnetic navigation at the lateral skull base in combination with flat panel volume computed tomography (fpVCT) datasets. A mastoidectomy and a posterior tympanotomy were performed on 10 samples of fresh frozen temporal bones. For registration, four self-drilling titanium screws were applied as fiducial markers. Multi-slice computed tomography (MSCT; 600 µm), conventional flat panel volume computed tomography (fpVCT; 466 µm), micro-fpVCT (197 µm) and secondary reconstructed fpVCT (100 µM) scans were performed and data were loaded into the navigation system. The resulting fiducial registration error (FRE) was analysed, and control of the navigation accuracy was performed. The registration process was very quick and reliable with the screws as fiducials. Compared to using the MSCT data, the micro-fpVCT data led to significantly lower FRE values, whereas conventional fpVCT and secondary reconstructed fpVCT data had no advantage in terms of accuracy. For all imaging modalities, there was no relevant visual deviation when targeting defined anatomical points with a navigation probe. fpVCT data are very well suited for electromagnetic navigation at the lateral skull base. The use of titanium screws as fiducial markers turned out to be ideal for comparing different imaging methods. A further evaluation of this approach by a clinical trial is required.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Johannes TaegerORCiD, Franz-Tassilo Müller-GraffORCiD, Tilmann Neun, Maria Köping, Philipp Schendzielorz, Rudolf Hagen, Kristen Rak
URN:urn:nbn:de:bvb:20-opus-250268
Document Type:Journal article
Faculties:Medizinische Fakultät / Klinik und Poliklinik für Hals-, Nasen- und Ohrenkrankheiten, plastische und ästhetische Operationen
Medizinische Fakultät / Institut für diagnostische und interventionelle Neuroradiologie (ehem. Abteilung für Neuroradiologie)
Language:English
Parent Title (English):Science Progress
ISSN:2047-7163
Year of Completion:2021
Volume:104
Issue:3
Source:Science Progress (2021) 104:3. https://doi.org/10.1177/00368504211032090
DOI:https://doi.org/10.1177/00368504211032090
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Tag:cochlear implantation; electromagnetic navigation; fiducial registration error; fpVCT; lateral skull base; otology
Release Date:2022/04/07
Date of first Publication:2021/08/16
Open-Access-Publikationsfonds / Förderzeitraum 2021
Licence (German):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International