Analysis of Impact of Humidity and Temperature on Excimer Laser Ablation of Polyethylene Terephthalate, Polymethylmethacrylate, and Porcine Corneal Tissue

Please always quote using this URN: urn:nbn:de:bvb:20-opus-213395
  • Background and Objectives To analyze the impact of humidity and temperature on excimer laser ablation of polyethylene terephthalate (PET), polymethylmethacrylate (PMMA) and porcine corneal tissue, and an ablation model to compensate for the temperature and humidity changes on ablation efficiency. Study Design/Materials and Methods The study was conducted using an AMARIS 1050RS (Schwind eye‐tech‐solutions) placed inside a climate chamber at ACTS. Ablations were performed on PET, PMMA, and porcine cornea. The impact of a wide range ofBackground and Objectives To analyze the impact of humidity and temperature on excimer laser ablation of polyethylene terephthalate (PET), polymethylmethacrylate (PMMA) and porcine corneal tissue, and an ablation model to compensate for the temperature and humidity changes on ablation efficiency. Study Design/Materials and Methods The study was conducted using an AMARIS 1050RS (Schwind eye‐tech‐solutions) placed inside a climate chamber at ACTS. Ablations were performed on PET, PMMA, and porcine cornea. The impact of a wide range of temperature (~18°C to ~30°C) and relative humidity (~25% to ~80%) on laser ablation outcomes was tested using nine climate test settings. For porcine eyes, change in defocus was calculated from the difference of post‐ablation to pre‐ablation average keratometry readings. Laser scanning deflectometry was performed to measure refractive change achieved in PMMA. Multiple linear regression was performed using the least square method with predictive factors: temperature, relative humidity, time stamp. Influence of climate settings was modeled for pulse energy, pulse fluence, ablation efficiency on PMMA and porcine cornea tissue. Results Temperature changes did not affect laser pulse energy, pulse fluence (PET), and ablation efficiency (on PMMA or porcine corneal tissue) significantly. Changes in relative humidity were critical and significantly affected laser pulse energy, high fluence and low fluence. The opposite trend was observed between the ablation performance on PMMA and porcine cornea. Conclusions The proposed well‐fitting multi‐linear model can be utilized for compensation of temperature and humidity changes on ablation efficiency. Based on this model, a working window for optimum operation has been found (temperature 18°C to 28°C and relative humidity 25% to 65%) for a maximum deviation of ±2.5% in ablation efficiency in PMMA and porcine corneal tissue.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Shwetabh Verma, Tobias Kehrer, Jürgen Hesser, Samuel Arba Mosquera
URN:urn:nbn:de:bvb:20-opus-213395
Document Type:Journal article
Faculties:Fakultät für Physik und Astronomie / Institut für Theoretische Physik und Astrophysik
Language:English
Parent Title (English):Lasers in Surgery and Medicine
Year of Completion:2020
Volume:52
Issue:7
First Page:627
Last Page:638
Source:Lasers in Surgery and Medicine 2020, 52(7):627-638. DOI: 10.1002/lsm.23190
DOI:https://doi.org/10.1002/lsm.23190
Dewey Decimal Classification:5 Naturwissenschaften und Mathematik / 53 Physik / 530 Physik
6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Tag:PET; PMMA; cornea; excimer laser ablation; impact of humidity and temperature; refractive surgery
Release Date:2021/04/13
Licence (German):License LogoCC BY-NC: Creative-Commons-Lizenz: Namensnennung, Nicht kommerziell 4.0 International