TY - JOUR A1 - Müller, Joachim A1 - Brill, Stefan A1 - Hagen, Rudolf A1 - Moeltner, Alexander A1 - Brockmeier, Steffi-Johanna A1 - Stark, Thomas A1 - Helbig, Silke A1 - Maurer, Jan A1 - Zahnert, Thomas A1 - Zierhofer, Clemens A1 - Nopp, Peter A1 - Anderson, Ilona T1 - Clinical Trial Results with the MED-EL Fine Structure Processing Coding Strategy in Experienced Cochlear Implant Users JF - ORL N2 - Objectives: To assess the subjective and objective performance of the new fine structure processing strategy (FSP) compared to the previous generation coding strategies CIS+ and HDCIS. Methods: Forty-six adults with a minimum of 6 months of cochlear implant experience were included. CIS+, HDCIS and FSP were compared in speech perception tests in noise, pitch scaling and questionnaires. The randomized tests were performed acutely (interval 1) and again after 3 months of FSP experience (interval 3). The subjective evaluation included questionnaire 1 at intervals 1 and 3, and questionnaire 2 at interval 2, 1 month after interval 1. Results: Comparison between FSP and CIS+ showed that FSP performed at least as well as CIS+ in all speech perception tests, and outperformed CIS+ in vowel and monosyllabic word discrimination. Comparison between FSP and HDCIS showed that both performed equally well in all speech perception tests. Pitch scaling showed that FSP performed at least as well as HDCIS. With FSP, sound quality was at least as good and often better than with HDCIS. Conclusions: Results indicate that FSP performs better than CIS+ in vowel and monosyllabic word understanding. Subjective evaluation demonstrates strong user preferences for FSP when listening to speech and music. KW - pitch KW - CIS+ KW - OPUS KW - fine structure processing KW - cochlear implant KW - coding strategy KW - speech perception KW - music Y1 - 2012 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:20-opus-196396 SN - 0301-1569 SN - 1423-0275 N1 - This publication is with permission of the rights owner freely accessible due to an Alliance licence and a national licence (funded by the DFG, German Research Foundation) respectively. VL - 74 IS - 4 ER -