Introduction
The goals of successful bi-compartmental knee arthroplasty are to achieve correct fit and positioning of the implant, while appropriately correcting the mechanical alignment of the leg after surgery. As these requirements are not always reliably fulfilled using off-the-shelf implant systems, newer approaches for bi-compartmental resurfacing have been explored.
Material and methods
In this article we report the radiographic results of 30 patients with anteromedial osteoarthritis (OA) who were treated with a novel patient-specific fixed-bearing bi-compartmental knee resurfacing system using custom-made implants and instruments. Utilizing standardized pre- and postoperative radiographic analyses (based on anterior-posterior and lateral, anterior-posterior weight-bearing full-length radiographs, patella skyline views and preoperative computed tomography (CT) scanning) implant fit and positioning as well as correction of the mechanical axis (hip-knee-ankle angle, HKA) were determined.
Results
On average, HKA was corrected from 173.4 ±3.47° preoperatively to 179.4 ±2.85° postoperatively. The coronal femoro-tibial angle was corrected on average 5.61°. The preoperative tibial slope measured on lateral views was 6.38 ±2.4°, while the average slope in the CT-based planning protocol (iView) was 6.14 ±2.40°. Postoperative lateral tibial slope was determined to be 5.77 ±1.97°. The thickness of the posterior femoral cuts was measured intraoperatively and, in all cases, corresponded well to the targeted thickness of the cuts provided by the iView. The joint line was preserved in all cases and the average Insall-Salvati index was 1.078 ±0.11 pre- and 1.072 ±0.11 postoperatively. The fit of the implant components measured by over- or underhang was excellent throughout (< 1.01 mm).
Conclusions
Custom-made bicompartmental knee arthroplasty can ensure optimized fitting and positioning of the implant with restoration of the leg axis. These implants could be considered as an alternative primary solution for knee surgeons treating bi-compartmental disease.
Purpose: Patient-specific, flanged acetabular components are used for the treatment of Paprosky type III defects during revision total hip arthroplasty (THA). This monocentric retrospective cohort study analyzes the outcome of patients treated with custom made monoflanged acetabular components (CMACs) with intra- and extramedullary iliac fixation. Methods: 14 patients were included who underwent revision THA with CMACs for the treatment of Paprosky type III defects. Mechanism of THA failure was infection in 4 and aseptic loosening in 10 patients. Seven patients underwent no previous revision, the other seven patients underwent three or more previous revisions. Results: At a mean follow-up of 35.4 months (14–94), the revision rate of the implant was 28.3%. Additionally, one perioperative dislocation and one superficial wound infection occurred. At one year postoperatively, we found a significant improvement of the Western Ontario and McMaster Universities Arthritis Index (WOMAC) score (p = 0.015). Postoperative radiographic analysis revealed good hip joint reconstruction with a mean leg length discrepancy of 3 mm (−8–20), a mean lateralization of the horizontal hip center of rotation of 8 mm (−8–35), and a mean proximalization of the vertical hip center of rotation of 6 mm (13–26). Radiolucency lines were present in 30%. Conclusion: CMACs can be considered an option for the treatment of acetabular bone loss in revision THA. Iliac intra- and extramedullary fixation allows soft tissue-adjusted hip joint reconstruction and improves hip function. However, failure rates are high, with periprosthetic infection being the main threat to successful outcome.