Experimental evidence has emerged that local platelet activation contributes to inflammation and infarct formation in acute ischemic stroke (AIS) which awaits confirmation in human studies. We conducted a prospective observational study on 258 consecutive patients undergoing mechanical thrombectomy (MT) due to large-vessel-occlusion stroke of the anterior circulation (08/2018-05/2020). Intraprocedural microcatheter aspiration of 1 ml of local (occlusion condition) and systemic arterial blood samples (self-control) was performed according to a prespecified protocol. The samples were analyzed for differential leukocyte counts, platelet counts, and plasma levels of the platelet-derived neutrophil-activating chemokine C-X-C-motif ligand (CXCL) 4 (PF-4), the neutrophil attractant CXCL7 (NAP-2), and myeloperoxidase (MPO). The clinical-biological relevance of these variables was corroborated by specific associations with molecular-cellular, structural-radiological, hemodynamic, and clinical-functional parameters. Seventy consecutive patients fulfilling all predefined criteria entered analysis. Mean local CXCL4 (+ 39%: 571 vs 410 ng/ml, P = .0095) and CXCL7 (+ 9%: 693 vs 636 ng/ml, P = .013) concentrations were higher compared with self-controls. Local platelet counts were lower (- 10%: 347,582 vs 383,284/µl, P = .0052), whereas neutrophil counts were elevated (+ 10%: 6022 vs 5485/µl, P = 0.0027). Correlation analyses revealed associations between local platelet and neutrophil counts (r = 0.27, P = .034), and between CXCL7 and MPO (r = 0.24, P = .048). Local CXCL4 was associated with the angiographic degree of reperfusion following recanalization (r = - 0.2523, P = .0479). Functional outcome at discharge correlated with local MPO concentrations (r = 0.3832, P = .0014) and platelet counts (r = 0.288, P = .0181). This study provides human evidence of cerebral platelet activation and platelet-neutrophil interactions during AIS and points to the relevance of per-ischemic thrombo-inflammatory mechanisms to impaired reperfusion and worse functional outcome following recanalization.
Background
Mobile 3-dimensional fluoroscopes are an integral part of modern neurosurgical operating theatres and can also be used in combination with free available image post processing to depict cerebral vessels. In preparation of stereotactic surgery, preoperative Computed Tomography (CT) may be required for image fusion. Contrast CT may be of further advantage for image fusion as it regards the vessel anatomy in trajectory planning. Time-consuming in-hospital transports are necessary for this purpose. Mobile 3D-fluoroscopes may be used to generate a CT equal preoperative data set without an in-hospital transport. This study was performed to determine the feasibility and image quality of intraoperative 3-dimensional fluoroscopy with intravenous contrast administration in combination with stereotactical procedures.
Methods
6 patients were included in this feasibility study. After fixation in a radiolucent Mayfield clamp a rotational fluoroscopy scan was performed with 50 mL iodine contrast agent. The image data sets were merged with the existing MRI images at a planning station and visually evaluated by two observer. The operation times were compared between the frame-based and frameless systems (“skin-to-skin” and “OR entry to exit”).
Results
The procedure proves to be safe. The entire procedure from fluoroscope positioning to the transfer to the planning station took 5–6 min with an image acquisition time of 24 s. In 5 of 6 cases, the fused imaging was able to reproduce the vascular anatomy accurately and in good quality. Both time end-points were significantly shorter compared to frame-based interventions.
Conclusion
The images could easily be transferred to the planning and navigation system and were successfully merged with the MRI data set. The procedure can be completely integrated into the surgical workflow. Preoperative CT imaging or transport under anaesthesia may even be replaced by this technique in the future. Furthermore, hemorrhages can be successfully visualized intraoperatively and might prevent time delays in emergencies.
Mobile 3D fluoroscopes have become increasingly available in neurosurgical operating rooms. We recently reported its use for imaging cerebral vascular malformations and aneurysms. This study was conducted to evaluate various radiation settings for the imaging of cerebral aneurysms before and after surgical occlusion. Eighteen patients with cerebral aneurysms with the indication for surgical clipping were included in this prospective analysis. Before surgery the patients were randomized into one of three different scan protocols according (default settings of the 3D fluoroscope): Group 1: 110 kV, 80 mA (enhanced cranial mode), group 2: 120 kV, 64 mA (lumbar spine mode), group 3: 120 kV, 25 mA (head/neck settings). Prior to surgery, a rotational fluoroscopy scan (duration 24 s) was performed without contrast agent followed by another scan with 50 ml of intravenous iodine contrast agent. The image files of both scans were transferred to an Apple PowerMac(R) workstation, subtracted and reconstructed using OsiriX(R) MD 10.0 software. The procedure was repeated after clip placement. The image quality regarding preoperative aneurysm configuration and postoperative assessment of aneurysm occlusion and vessel patency was analyzed by 2 independent reviewers using a 6-grade scale. This technique quickly supplies images of adequate quality to depict intracranial aneurysms and distal vessel patency after aneurysm clipping. Regarding these features, a further optimization to our previous protocol seems possible lowering the voltage and increasing tube current. For quick intraoperative assessment, image subtraction seems not necessary. Thus, a native scan without a contrast agent is not necessary. Further optimization may be possible using a different contrast injection protocol.