• Treffer 12 von 41
Zurück zur Trefferliste

Influence of contrast agent dispersion on bolus-based MRI myocardial perfusion measurements: A computational fluid dynamics study

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-208698
  • Purpose: Bolus‐based dynamic contrast agent (CA) perfusion measurements of the heart are subject to systematic errors due to CA bolus dispersion in the coronary arteries. To better understand these effects on quantification of myocardial blood flow and myocardial perfusion reserve (MPR), an in‐silico model of the coronary arteries down to the pre‐arteriolar vessels has been developed. Methods: In this work, a computational fluid dynamics analysis is performed to investigate these errors on the basis of realistic 3D models of the left andPurpose: Bolus‐based dynamic contrast agent (CA) perfusion measurements of the heart are subject to systematic errors due to CA bolus dispersion in the coronary arteries. To better understand these effects on quantification of myocardial blood flow and myocardial perfusion reserve (MPR), an in‐silico model of the coronary arteries down to the pre‐arteriolar vessels has been developed. Methods: In this work, a computational fluid dynamics analysis is performed to investigate these errors on the basis of realistic 3D models of the left and right porcine coronary artery trees, including vessels at the pre‐arteriolar level. Using advanced boundary conditions, simulations of blood flow and CA transport are conducted at rest and under stress. These are evaluated with regard to dispersion (assessed by the width of CA concentration time curves and associated vascular transport functions) and errors of myocardial blood flow and myocardial perfusion reserve quantification. Results: Contrast agent dispersion increases with traveled distance as well as vessel diameter, and decreases with higher flow velocities. Overall, the average myocardial blood flow errors are −28% ± 16% and −8.5% ± 3.3% at rest and stress, respectively, and the average myocardial perfusion reserve error is 26% ± 22%. The calculated values are different in the left and right coronary tree. Conclusion: Contrast agent dispersion is dependent on a complex interplay of several different factors characterizing the cardiovascular bed, including vessel size and integrated vascular length. Quantification errors evoked by the observed CA dispersion show nonnegligible distortion in dynamic CA bolus‐based perfusion measurements. We expect future improvements of quantitative perfusion measurements to make the systematic errors described here more apparent.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Johannes MartensORCiD, Sabine Panzer, Jeroen van den Wijngaard, Maria SiebesORCiD, Laura M. Schreiber
URN:urn:nbn:de:bvb:20-opus-208698
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Deutsches Zentrum für Herzinsuffizienz (DZHI)
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Magnetic Resonance in Medicine
Erscheinungsjahr:2020
Band / Jahrgang:84
Heft / Ausgabe:1
Seitenangabe:467-483
Originalveröffentlichung / Quelle:Magnetic Resonance in Medicine 2020;84(1):467–483. DOI: 10.1002/mrm.28125
DOI:https://doi.org/10.1002/mrm.28125
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):bolus-based perfusion measurement; computational fluid dynamics; magnetic resonance imaging; myocardial blood flow; myocardial perfusion reserve
Datum der Freischaltung:22.09.2020
Lizenz (Deutsch):License LogoCC BY-NC: Creative-Commons-Lizenz: Namensnennung, Nicht kommerziell 4.0 International