The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets — Quantification and possible corrections

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-218146
  • Purpose To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D\(_{98}\), target volume coverage (TVC), conformity (CI), and gradient index (GI). Methods Winston‐Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assurance (QA) plan was applied to the ArcCHECK phantom to assess the impact of isocenter shift corrections on Gamma pass rates. These corrections included gantry sag, distance of collimator and couch axes to the gantry axis,Purpose To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D\(_{98}\), target volume coverage (TVC), conformity (CI), and gradient index (GI). Methods Winston‐Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assurance (QA) plan was applied to the ArcCHECK phantom to assess the impact of isocenter shift corrections on Gamma pass rates. These corrections included gantry sag, distance of collimator and couch axes to the gantry axis, and distance between cone‐beam computed tomography (CBCT) isocenter and treatment beam (MV) isocenter. We applied the shifts via script to the treatment plan in Pinnacle 16.2. In a planning study, isocenter and mechanical rotation axis shifts of 0.25 to 2 mm were applied to stereotactic plans of spherical planning target volumes (PTVs) of various volumes. The shifts determined via WL measurements were applied to 16 patient plans with PTV sizes between 0.22 and 10.4 cm3. Results ArcCHECK measurements of a stereotactic treatment showed significant increases in Gamma pass rate for all three measurements (up to 3.8 percentage points) after correction of measured isocenter deviations. For spherical targets of 1 cm3, CI was most severely affected by increasing the distance of the CBCT isocenter (1.22 to 1.62). Gradient index increased with an isocenter‐collimator axis distance of 1.5 mm (3.84 vs 4.62). D98 (normalized to reference) dropped to 0.85 (CBCT), 0.92 (table axis), 0.95 (collimator axis), and 0.98 (gantry sag), with similar but smaller changes for larger targets. Applying measured shifts to patient plans lead to relevant drops in D\(_{98}\) and TVC (7%) for targets below 2 cm\(^3\) treated on linac 1. Conclusion Mechanical deviations during gantry, collimator, and table rotation may adversely affect the treatment of small stereotactic lesions. Adjustments of beam isocenters in the treatment planning system (TPS) can be used to both quantify their impact and for prospective correction of treatment plans.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Linda J. Wack, Florian Exner, Sonja Wegener, Otto A. Sauer
URN:urn:nbn:de:bvb:20-opus-218146
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Klinik und Poliklinik für Strahlentherapie
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Journal of Applied Clinical Medical Physics
Erscheinungsjahr:2020
Band / Jahrgang:21
Heft / Ausgabe:5
Seitenangabe:56-64
Originalveröffentlichung / Quelle:Journal of Applied Clinical Medical Physics 2020, 21: 56-64. https://doi.org/10.1002/acm2.12854
DOI:https://doi.org/10.1002/acm2.12854
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):Winston‐Lutz test; isocenter; quality assurance; stereotactic radiotherapy