Refine
Has Fulltext
- yes (1)
Is part of the Bibliography
- yes (1)
Year of publication
- 2023 (1)
Document Type
- Doctoral Thesis (1) (remove)
Language
- English (1) (remove)
Keywords
- cine loop (1) (remove)
This work deals with the acceleration of cardiovascular MRI for the assessment
of functional information in steady-state contrast and for viability assessment
during the inversion recovery of the magnetization. Two approaches
are introduced and discussed in detail. MOCO-MAP uses an exponential
model to recover dynamic image data, IR-CRISPI, with its low-rank plus
sparse reconstruction, is related to compressed sensing.
MOCO-MAP is a successor to model-based acceleration of parametermapping
(MAP) for the application in the myocardial region. To this end, it
was augmented with a motion correction (MOCO) step to allow exponential
fitting the signal of a still object in temporal direction. Iteratively, this
introduction of prior physical knowledge together with the enforcement of
consistency with the measured data can be used to reconstruct an image
series from distinctly shorter sampling time than the standard exam (< 3 s
opposed to about 10 s). Results show feasibility of the method as well as
detectability of delayed enhancement in the myocardium, but also significant
discrepancies when imaging cardiac function and artifacts caused already by
minor inaccuracy of the motion correction.
IR-CRISPI was developed from CRISPI, which is a real-time protocol
specifically designed for functional evaluation of image data in steady-state
contrast. With a reconstruction based on the separate calculation of low-rank
and sparse part, it employs a softer constraint than the strict exponential
model, which was possible due to sufficient temporal sampling density via
spiral acquisition. The low-rank plus sparse reconstruction is fit for the use on
dynamic and on inversion recovery data. Thus, motion correction is rendered
unnecessary with it.
IR-CRISPI was equipped with noise suppression via spatial wavelet filtering.
A study comprising 10 patients with cardiac disease show medical
applicability. A comparison with performed traditional reference exams offer
insight into diagnostic benefits. Especially regarding patients with difficulty
to hold their breath, the real-time manner of the IR-CRISPI acquisition provides
a valuable alternative and an increase in robustness.
In conclusion, especially with IR-CRISPI in free breathing, a major acceleration
of the cardiovascular MR exam could be realized. In an acquisition
of less than 100 s, it not only includes the information of two traditional
protocols (cine and LGE), which take up more than 9.6 min, but also allows
adjustment of TI in retrospect and yields lower artifact level with similar
image quality.