Background
Small-animal single-photon emission computed tomography (SPECT) systems with multi-pinhole collimation and large stationary detectors have advantages compared to systems with moving small detectors. These systems benefit from less labour-intensive maintenance and quality control as fewer prone parts are moving, higher accuracy for focused scans and maintaining high resolution with increased sensitivity due to focused pinholes on the field of view. This study aims to investigate the performance of a novel ultra-high-resolution scanner with two-detector configuration (U-SPECT5-E) and to compare its image quality to a conventional micro-SPECT system with three stationary detectors (U-SPECT\(^+\)).
Methods
The new U-SPECT5-E with two stationary detectors was used for acquiring data with \(^{99m}\)Tc-filled point source, hot-rod and uniformity phantoms to analyse sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR). Three dedicated multi-pinhole mouse collimators with 75 pinholes each and 0.25-, 0.60- and 1.00-mm pinholes for extra ultra-high resolution (XUHR-M), general-purpose (GP-M) and ultra-high sensitivity (UHS-M) imaging were examined. For CNR analysis, four different activity ranges representing low- and high-count settings were investigated for all three collimators. The experiments for the performance assessment were repeated with the same GP-M collimator in the three-detector U-SPECT\(^+\) for comparison.
Results
Peak sensitivity was 237 cps/MBq (XUHR-M), 847 cps/MBq (GP-M), 2054 cps/MBq (UHS-M) for U-SPECT5-E and 1710 cps/MBq (GP-M) for U-SPECT\(^+\). In the visually analysed sections of the reconstructed mini Derenzo phantoms, rods as small as 0.35 mm (XUHR-M), 0.50 mm (GP-M) for the two-detector as well as the three-detector SPECT and 0.75 mm (UHS-M) were resolved. Uniformity for maximum resolution recorded 40.7% (XUHR-M), 29.1% (GP-M, U-SPECT5-E), 16.3% (GP-M, U-SPECT\(^+\)) and 23.0% (UHS-M), respectively. UHS-M reached highest CNR values for low-count images; for rods smaller than 0.45 mm, acceptable CNR was only achieved by XUHR-M. GP-M was superior for imaging rods sized from 0.60 to 1.50 mm for intermediate activity concentrations. U-SPECT5-E and U-SPECT+ both provided comparable CNR.
Conclusions
While uniformity and sensitivity are negatively affected by the absence of a third detector, the investigated U-SPECT5-E system with two stationary detectors delivers excellent spatial resolution and CNR comparable to the performance of an established three-detector-setup.
This longitudinal study was performed to evaluate the feasibility of detecting the interaction between wall shear stress (WSS) and plaque development. 20 ApoE\(^{-/-}\)mice were separated in 12 mice with Western Diet and 8 mice with Chow Diet. Magnetic resonance (MR) scans at 17.6 Tesla and histological analysis were performed after one week, eight and twelve weeks. Allin vivoMR measurements were acquired using a flow sensitive phase contrast method for determining vectorial flow. Histological sections were stained with Hematoxylin and Eosin, Elastica van Gieson and CD68 staining. Data analysis was performed using Ensight and a Matlab-based "Flow Tool". The body weight of ApoE\(^{-/-}\)mice increased significantly over 12 weeks. WSS values increased in the Western Diet group over the time period; in contrast, in the Chow Diet group the values decreased from the first to the second measurement point. Western Diet mice showed small plaque formations with elastin fragmentations after 8 weeks and big plaque formations after 12 weeks; Chow Diet mice showed a few elastin fragmentations after 8 weeks and small plaque formations after 12 weeks. Favored by high-fat diet, plaque formation results in higher values of WSS. With wall shear stress being a known predictor for atherosclerotic plaque development, ultra highfield MRI can serve as a tool for studying the causes and beginnings of atherosclerosis.