Refine
Has Fulltext
- yes (24)
Is part of the Bibliography
- yes (24)
Year of publication
Document Type
- Journal article (24) (remove)
Language
- English (24) (remove)
Keywords
- dosimetry (6)
- PRRT (5)
- 53BP1 (3)
- γ-H2AX (3)
- 177Lu (2)
- CXCR4 (2)
- PET/CT (2)
- biodosimetry (2)
- chemokine receptor (2)
- neuroendocrine tumor (2)
- positron emission tomography (2)
- quantitative SPECT/CT (2)
- repair (2)
- 131I (1)
- 133Ba (1)
- 177Lu SPECT/CT imaging (1)
- 223Ra (1)
- 224Ra (1)
- 3D printing (1)
- 5IA-SPECT (1)
- BSS directive (1)
- Barium-133 (1)
- Biokinetics (1)
- Combination (1)
- DNA Breaks (1)
- DNA damage (1)
- DNA double-strand breaks (1)
- DNA repair (1)
- DOTATOC (1)
- DSB damage (1)
- DSB focus substructure (1)
- Deep learning (1)
- Denoising (1)
- Diagnostic radiopharmaceuticals (1)
- Dosimetry (1)
- EANM dosage card (1)
- EBRT (1)
- Effective dose (1)
- Ga-68 (1)
- JR11 (1)
- Lu-177 (1)
- Lutetium (1)
- MAG3 (1)
- Medizin (1)
- Meningioma (1)
- Monte Carlo (1)
- OPS201 (1)
- PET (1)
- PET/MR systems (1)
- PSMA (1)
- Parkinson disease (1)
- Peptide receptor radionuclide therapy (1)
- Positronen-Emissions-Tomografie (1)
- Ra-224 (1)
- Radiotherapy (1)
- Radium (1)
- SPECT (1)
- SPECT/CT (1)
- SSTR (1)
- Single Molecule Localization Microscopy (SMLM) (1)
- [\(^{68}\)Ga]Pentixafor (1)
- \(^{177}\)Lu-OPS201 (1)
- absorbed dose to the blood (1)
- acute myeloid leukemia (1)
- alpha particles (1)
- alpha-emitters (1)
- antagonist (1)
- biokinetics (1)
- biological dosimetry (1)
- biomarkers (1)
- blood (1)
- bone-targeting radiopharmaceuticals (1)
- calibration (1)
- cancer treatment (1)
- chromatin mobility (1)
- cognitive decline (1)
- comparison exercise (1)
- complex DNA damage (1)
- contrast agent (1)
- damage (1)
- dopamine acetylcholine (1)
- dose response (1)
- double-stranded (1)
- epidemiology (1)
- gamma rays (1)
- harmonization of SPECT/CT imaging (1)
- health care (1)
- healthy volunteers (1)
- high LET irradiation (1)
- histone H2AX (1)
- humans (1)
- in vivo formation (1)
- in vivo imaging (1)
- international multicenter comparison exercise (1)
- irradiation (1)
- isotopes (1)
- leukocytes (1)
- lutetium-177 (1)
- lymphoma (1)
- medicine (1)
- molecular radiotherapy (1)
- molecular radiotherapy (MRT) (1)
- multi-centre (1)
- neuroendocrine tumor (NET) (1)
- neuroendocrine tumors (1)
- nicotinic receptors (1)
- nuclear medicine (1)
- nuclear medicine therapy (1)
- optimization (1)
- peptide receptor radionuclide therapy (1)
- phantom (1)
- phosphorylation (1)
- photons (1)
- pig model (1)
- prognostic value (1)
- prostate cancer (1)
- quantitative imaging (1)
- radiation effects (1)
- radiobiology (1)
- radioiodine (1)
- radioiodine therapy (1)
- radionuclide therapy (1)
- radium (1)
- renal scintigraphy (1)
- signal to noise ratio (1)
- solid surrogate source (1)
- standardization of SPECT/CT imaging (1)
- super ultraviolet (1)
- therapeutic target (1)
- thyroid cancer (1)
- thyroid carcinomas (1)
- traceability of SPECT/CT imaging (1)
- α-Particle (1)
- α-emitter (1)
Institute
EU-Project number / Contract (GA) number
- 701983 (1)
Introduction
Commissioning, calibration, and quality control procedures for nuclear medicine imaging systems are typically performed using hollow containers filled with radionuclide solutions. This leads to multiple sources of uncertainty, many of which can be overcome by using traceable, sealed, long-lived surrogate sources containing a radionuclide of comparable energies and emission probabilities. This study presents the results of a quantitative SPECT/CT imaging comparison exercise performed within the MRTDosimetry consortium to assess the feasibility of using 133Ba as a surrogate for 131I imaging.
Materials and methods
Two sets of four traceable 133Ba sources were produced at two National Metrology Institutes and encapsulated in 3D-printed cylinders (volume range 1.68–107.4 mL). Corresponding hollow cylinders to be filled with liquid 131I and a mounting baseplate for repeatable positioning within a Jaszczak phantom were also produced. A quantitative SPECT/CT imaging comparison exercise was conducted between seven members of the consortium (eight SPECT/CT systems from two major vendors) based on a standardised protocol. Each site had to perform three measurements with the two sets of 133Ba sources and liquid 131I.
Results
As anticipated, the 131I pseudo-image calibration factors (cps/MBq) were higher than those for 133Ba for all reconstructions and systems. A site-specific cross-calibration reduced the performance differences between both radionuclides with respect to a cross-calibration based on the ratio of emission probabilities from a median of 12–1.5%. The site-specific cross-calibration method also showed agreement between 133Ba and 131I for all cylinder volumes, which highlights the potential use of 133Ba sources to calculate recovery coefficients for partial volume correction.
Conclusion
This comparison exercise demonstrated that traceable solid 133Ba sources can be used as surrogate for liquid 131I imaging. The use of solid surrogate sources could solve the radiation protection problem inherent in the preparation of phantoms with 131I liquid activity solutions as well as reduce the measurement uncertainties in the activity. This is particularly relevant for stability measurements, which have to be carried out at regular intervals.
[\(^{223}\)Ra]RaCl\(_2\) and [\(^{224}\)Ra]RaCl\(_2\) are bone seekers, emitting high LET, and short range (< 100 μm) alpha-particles. Both radionuclides show similar decay properties; the total alpha energies are comparable (\(^{223}\)Ra: ≈28 MeV, \(^{224}\)Ra: ≈26 MeV). [\(^{224}\)Ra]RaCl\(_2\) has been used from the mid-1940s until 1990 for treating different bone and joint diseases with activities of up to approximately 50 MBq [\(^{224}\)Ra]RaCl\(_2\). In 2013 [\(^{223}\)Ra]RaCl\(_2\) obtained marketing authorization by the FDA and by the European Union for the treatment of metastatic prostate cancer with an activity to administer of 0.055 MBq per kg body weight for six cycles. For intravenous injections in humans a model calculation using the biokinetic model of ICRP67 shows a ratio of organ absorbed dose coefficients (\(^{224}\)Ra:\(^{223}\)Ra) between 0.37 (liver) and 0.97 except for the kidneys (2.27) and blood (1.57). For the red marrow as primary organ-at-risk, the ratio is 0.57. The differences are mainly caused be the differing half-lives of the decay products of both radium isotopes. Both radionuclides show comparable DNA damage patterns in peripheral blood mononuclear cells after internal ex-vivo irradiation. Data on the long-term radiation-associated side effects are only available for treatment with [\(^{224}\)Ra]RaCl\(_2\). Two epidemiological studies followed two patient groups treated with [\(^{224}\)Ra]RaCl\(_2\) for more than 25 years. One of them was the “Spiess study”, a cohort of 899 juvenile patients who received several injections of [\(^{224}\)Ra]RaCl\(_2\) with a mean specific activity of 0.66 MBq/kg. Another patient group of ankylosing spondylitis patients was treated with 10 repeated intravenous injections of [\(^{224}\)Ra]RaCl\(_2\), 1 MBq each, 1 week apart. In total 1,471 of these patients were followed-up in the “Wick study”. In both studies, an increased cancer mortality by leukemia and solid cancers was observed. Similar considerations on long-term effects likely apply to [\(^{223}\)Ra]RaCl\(_2\) as well since the biokinetics are similar and the absorbed doses in the same range. However, this increased risk will most likely not be observed due to the much shorter life expectancy of prostate cancer patients treated with [\(^{223}\)Ra]RaCl\(_2\).
Aim
Recent advancements in PET technology have brought with it significant improvements in PET performance and image quality. In particular, the extension of the axial field of view of PET systems, and the introduction of semiconductor technology into the PET detector, initially for PET/MR, and more recently available long-field-of-view PET/CT systems (≥ 25 cm) have brought a step change improvement in the sensitivity of PET scanners. Given the requirement to limit paediatric doses, this increase in sensitivity is extremely welcome for the imaging of children and young people. This is even more relevant with PET/MR, where the lack of CT exposures brings further dose reduction benefits to this population. In this short article, we give some details around the benefits around new PET technology including PET/MR and its implications on the EANM paediatric dosage card.
Material and methods
Reflecting on EANM adult guidance on injected activities, and making reference to bed overlap and the concept of MBq.min bed\(^{-1}\) kg\(^{-1}\), we use published data on image quality from PET/MR systems to update the paediatric dosage card for PET/MR and extended axial field of view (≥ 25 cm) PET/CT systems. However, this communication does not cover the expansion of paediatric dosing for the half-body and total-body scanners that have recently come to market.
Results
In analogy to the existing EANM dosage card, new parameters for the EANM paediatric dosage card were developed (class B, baseline value: 10.7 MBq, minimum recommended activity 10 MBq). The recommended administered activities for the systems considered in this communication range from 11 MBq [\(^{18}\)F]FDG for a child with a weight of 3 kg to 149 MBq [\(^{18}\)F]FDG for a paediatric patient weight of 68 kg, assuming a scan of 3 min per bed position. The mean effective dose over all ages (1 year and older) is 2.85 mSv.
Conclusion
With this, recommendations for paediatric dosing are given for systems that have not been considered previously.
Purpose
As α-emitters for radiopharmaceutical therapies are administered systemically by intravenous injection, blood will be irradiated by α-particles that induce clustered DNA double-strand breaks (DSBs). Here, we investigated the induction and repair of DSB damage in peripheral blood mononuclear cells (PBMCs) as a function of the absorbed dose to the blood following internal ex vivo irradiation with [\(^{223}\)Ra]RaCl2.
Methods
Blood samples of ten volunteers were irradiated by adding [\(^{223}\)Ra]RaCl2 solution with different activity concentrations resulting in absorbed doses to the blood of 3 mGy, 25 mGy, 50 mGy and 100 mGy. PBMCs were isolated, divided in three parts and either fixed directly (d-samples) or after 4 h or 24 h culture. After immunostaining, the induced γ-H2AX α-tracks were counted. The time-dependent decrease in α-track frequency was described with a model assuming a repair rate R and a fraction of non-repairable damage Q.
Results
For 25 mGy, 50 mGy and 100 mGy, the numbers of α-tracks were significantly increased compared to baseline at all time points. Compared to the corresponding d-samples, the α-track frequency decreased significantly after 4 h and after 24 h. The repair rates R were (0.24 ± 0.05) h−1 for 25 mGy, (0.16 ± 0.04) h−1 for 50 mGy and (0.13 ± 0.02) h−1 for 100 mGy, suggesting faster repair at lower absorbed doses, while Q-values were similar.
Conclusion
The results obtained suggest that induction and repair of the DSB damage depend on the absorbed dose to the blood. Repair rates were similar to what has been observed for irradiation with low linear energy transfer.
Background
In recent years, a lot of effort has been put in the enhancement of medical imaging using artificial intelligence. However, limited patient data in combination with the unavailability of a ground truth often pose a challenge to a systematic validation of such methodologies. The goal of this work was to investigate a recently proposed method for an artificial intelligence-based generation of synthetic SPECT projections, for acceleration of the image acquisition process based on a large dataset of realistic SPECT simulations.
Methods
A database of 10,000 SPECT projection datasets of heterogeneous activity distributions of randomly placed random shapes was simulated for a clinical SPECT/CT system using the SIMIND Monte Carlo program. Synthetic projections at fixed angular increments from a set of input projections at evenly distributed angles were generated by different u-shaped convolutional neural networks (u-nets). These u-nets differed in noise realization used for the training data, number of input projections, projection angle increment, and number of training/validation datasets. Synthetic projections were generated for 500 test projection datasets for each u-net, and a quantitative analysis was performed using statistical hypothesis tests based on structural similarity index measure and normalized root-mean-squared error. Additional simulations with varying detector orbits were performed on a subset of the dataset to study the effect of the detector orbit on the performance of the methodology. For verification of the results, the u-nets were applied to Jaszczak and NEMA physical phantom data obtained on a clinical SPECT/CT system.
Results
No statistically significant differences were observed between u-nets trained with different noise realizations. In contrast, a statistically significant deterioration was found for training with a small subset (400 datasets) of the 10,000 simulated projection datasets in comparison with using a large subset (9500 datasets) for training. A good agreement between synthetic (i.e., u-net generated) and simulated projections before adding noise demonstrates a denoising effect. Finally, the physical phantom measurements show that our findings also apply for projections measured on a clinical SPECT/CT system.
Conclusion
Our study shows the large potential of u-nets for accelerating SPECT/CT imaging. In addition, our analysis numerically reveals a denoising effect when generating synthetic projections with a u-net. Clinically interesting, the methodology has proven robust against camera orbit deviations in a clinically realistic range. Lastly, we found that a small number of training samples (e.g., ~ 400 datasets) may not be sufficient for reliable generalization of the u-net.
With an increasing variety of radiopharmaceuticals for diagnostic or therapeutic nuclear medicine as valuable diagnostic or treatment option, radiobiology plays an important role in supporting optimizations. This comprises particularly safety and efficacy of radionuclide therapies, specifically tailored to each patient. As absorbed dose rates and absorbed dose distributions in space and time are very different between external irradiation and systemic radionuclide exposure, distinct radiation-induced biological responses are expected in nuclear medicine, which need to be explored. This calls for a dedicated nuclear medicine radiobiology. Radiobiology findings and absorbed dose measurements will enable an improved estimation and prediction of efficacy and adverse effects. Moreover, a better understanding on the fundamental biological mechanisms underlying tumor and normal tissue responses will help to identify predictive and prognostic biomarkers as well as biomarkers for treatment follow-up. In addition, radiobiology can form the basis for the development of radiosensitizing strategies and radioprotectant agents. Thus, EANM believes that, beyond in vitro and preclinical evaluations, radiobiology will bring important added value to clinical studies and to clinical teams. Therefore, EANM strongly supports active collaboration between radiochemists, radiopharmacists, radiobiologists, medical physicists, and physicians to foster research toward precision nuclear medicine.
Purpose
Patient-specific dosimetry is required to ensure the safety of molecular radiotherapy and to predict response. Dosimetry involves several steps, the first of which is the determination of the activity of the radiopharmaceutical taken up by an organ/lesion over time. As uncertainties propagate along each of the subsequent steps (integration of the time–activity curve, absorbed dose calculation), establishing a reliable activity quantification is essential. The MRTDosimetry project was a European initiative to bring together expertise in metrology and nuclear medicine research, with one main goal of standardizing quantitative \(^{177}\)Lu SPECT/CT imaging based on a calibration protocol developed and tested in a multicentre inter-comparison. This study presents the setup and results of this comparison exercise.
Methods
The inter-comparison included nine SPECT/CT systems. Each site performed a set of three measurements with the same setup (system, acquisition and reconstruction): (1) Determination of an image calibration for conversion from counts to activity concentration (large cylinder phantom), (2) determination of recovery coefficients for partial volume correction (IEC NEMA PET body phantom with sphere inserts), (3) validation of the established quantitative imaging setup using a 3D printed two-organ phantom (ICRP110-based kidney and spleen). In contrast to previous efforts, traceability of the activity measurement was required for each participant, and all participants were asked to calculate uncertainties for their SPECT-based activities.
Results
Similar combinations of imaging system and reconstruction lead to similar image calibration factors. The activity ratio results of the anthropomorphic phantom validation demonstrate significant harmonization of quantitative imaging performance between the sites with all sites falling within one standard deviation of the mean values for all inserts. Activity recovery was underestimated for total kidney, spleen, and kidney cortex, while it was overestimated for the medulla.
Conclusion
This international comparison exercise demonstrates that harmonization of quantitative SPECT/CT is feasible when following very specific instructions of a dedicated calibration protocol, as developed within the MRTDosimetry project. While quantitative imaging performance demonstrates significant harmonization, an over- and underestimation of the activity recovery highlights the limitations of any partial volume correction in the presence of spill-in and spill-out between two adjacent volumes of interests.
The EC Directive 2013/59/Euratom states in article 56 that exposures of target volumes in nuclear medicine treatments shall be individually planned and their delivery appropriately verified. The Directive also mentions that medical physics experts should always be appropriately involved in those treatments. Although it is obvious that, in nuclear medicine practice, every nuclear medicine physician and physicist should follow national rules and legislation, the EANM considered it necessary to provide guidance on how to interpret the Directive statements for nuclear medicine treatments.
For this purpose, the EANM proposes to distinguish three levels in compliance to the optimization principle in the directive, inspired by the indication of levels in prescribing, recording and reporting of absorbed doses after radiotherapy defined by the International Commission on Radiation Units and Measurements (ICRU):
Most nuclear medicine treatments currently applied in Europe are standardized. The minimum requirement for those treatments is ICRU level 1 (“activity-based prescription and patient-averaged dosimetry”), which is defined by administering the activity within 10% of the intended activity, typically according to the package insert or to the respective EANM guidelines, followed by verification of the therapy delivery, if applicable.
Non-standardized treatments are essentially those in developmental phase or approved radiopharmaceuticals being used off-label with significantly (> 25% more than in the label) higher activities. These treatments should comply with ICRU level 2 (“activity-based prescription and patient-specific dosimetry”), which implies recording and reporting of the absorbed dose to organs at risk and optionally the absorbed dose to treatment regions.
The EANM strongly encourages to foster research that eventually leads to treatment planning according to ICRU level 3 (“dosimetry-guided patient-specific prescription and verification”), whenever possible and relevant.
Evidence for superiority of therapy prescription on basis of patient-specific dosimetry has not been obtained. However, the authors believe that a better understanding of therapy dosimetry, i.e. how much and where the energy is delivered, and radiobiology, i.e. radiation-related processes in tissues, are keys to the long-term improvement of our treatments.
The aim was to investigate the induction and repair of radiation-induced DNA double-strand breaks (DSBs) as a function of the absorbed dose to the blood of patients undergoing PET/CT examinations with [68Ga]Ga-PSMA. Blood samples were collected from 15 patients before and at four time points after [68Ga]Ga-PSMA administration, both before and after the PET/CT scan. Absorbed doses to the blood were calculated. In addition, blood samples with/without contrast agent from five volunteers were irradiated ex vivo by CT while measuring the absorbed dose. Leukocytes were isolated, fixed, and stained for co-localizing γ-H2AX+53BP1 DSB foci that were enumerated manually. In vivo, a significant increase in γ-H2AX+53BP1 foci compared to baseline was observed at all time points after administration, although the absorbed dose to the blood by 68Ga was below 4 mGy. Ex vivo, the increase in radiation-induced foci depended on the absorbed dose and the presence of contrast agent, which could have caused a dose enhancement. The CT-dose contribution for the patients was estimated at about 12 mGy using the ex vivo calibration. The additional number of DSB foci induced by CT, however, was comparable to the one induced by 68Ga. The significantly increased foci numbers after [68Ga]Ga-PSMA administration may suggest a possible low-dose hypersensitivity.
Introduction. \(^{177}\)Lu-OPS201 is a high-affinity somatostatin receptor subtype 2 antagonist for PRRT in patients with neuroendocrine tumors. The aim is to find the optimal scaling for dosimetry and to compare the biokinetics of \(^{177}\)Lu-OPS201 in animals and humans. Methods. Data on biokinetics of \(^{177}\)Lu-OPS201 were analyzed in athymic nude Foxn1\(^{nu}\) mice (28 F, weight: 26 ± 1 g), Danish Landrace pigs (3 F-1 M, weight: 28 ± 2 kg), and patients (3 F-1 M, weight: 61 ± 17 kg) with administered activities of 0.19–0.27 MBq (mice), 97–113 MBq (pigs), and 850–1086 MBq (patients). After euthanizing mice (up to 168 h), the organ-specific activity contents (including blood) were measured. Multiple planar and SPECT/CT scans were performed until 250 h (pigs) and 72 h (patients) to quantify the uptake in the kidneys and liver. Blood samples were taken up to 23 h (patients) and 300 h (pigs). In pigs and patients, kidney protection was applied. Time-dependent uptake data sets were created for each species and organ/tissue. Biexponential fits were applied to compare the biokinetics in the kidneys, liver, and blood of each species. The time-integrated activity coefficients (TIACs) were calculated by using NUKFIT. To determine the optimal scaling, several methods (relative mass scaling, time scaling, combined mass and time scaling, and allometric scaling) were compared. Results. A fast blood clearance of the compound was observed in the first phase (<56 h) for all species. In comparison with patients, pigs showed higher liver retention. Based on the direct comparison of the TIACs, an underestimation in mice (liver and kidneys) and an overestimation in pigs’ kidneys compared to the patient data (kidney TIAC: mice = 1.4 h, pigs = 7.7 h, and patients = 5.8 h; liver TIAC: mice = 0.7 h, pigs = 4.1 h, and patients = 5.3 h) were observed. Most similar TIACs were obtained by applying time scaling (mice) and combined scaling (pigs) (kidney TIAC: mice = 3.9 h, pigs = 4.8 h, and patients = 5.8 h; liver TIAC: mice = 0.9 h, pigs = 4.7 h, and patients = 5.3 h). Conclusion. If the organ mass ratios between the species are high, the combined mass and time scaling method is optimal to minimize the interspecies differences. The analysis of the fit functions and the TIACs shows that pigs are better mimicking human biokinetics.