In total knee arthroplasty (TKA), functional knee phenotypes are of interest regarding surgical alignment strategies. Functional knee phenotypes were introduced in 2019 and consist of limb, femoral, and tibial phenotypes. The hypothesis of this study was that mechanically aligned (MA) TKA changes preoperative functional phenotypes, which decreases the 1-year Forgotten Joint (FJS) and Oxford Knee Score (OKS) and increases the 1-year WOMAC. All patients included in this study had end-stage osteoarthritis and were treated with a primary MA TKA, which was supervised by four academic knee arthroplasty specialists. To determine the limb, femoral, and tibial phenotype, a long-leg radiograph (LLR) was imaged preoperatively and two to three days after TKA. FJS, OKS, and WOMAC were obtained 1 year after TKA. Patients were categorized using the change in functional limb, femoral, and tibial phenotype measured on LLR, and the scores were compared between the different categories. A complete dataset of preoperative and postoperative scores and radiographic images could be obtained for 59 patients. 42% of these patients had a change of limb phenotype, 41% a change of femoral phenotype, and 24% a change of tibial phenotype of more than ±1 relative to the preoperative phenotype. Patients with more than ±1 change of limb phenotype had significantly lower median FJS (27 points) and OKS (31 points) and higher WOMAC scores (30 points) relative to the 59-, 41-, and 4-point scores of those with a 0 ± 1 change (p < 0.0001 to 0.0048). Patients with a more than ±1 change of femoral phenotype had significantly lower median FJS (28 points) and OKS (32 points) and higher WOMAC scores (24 points) relative to the 69-, 40-, and 8-point scores of those with a 0 ± 1 change (p < 0.0001). A change in tibial phenotype had no effect on the FJS, OKS, and WOMAC scores. Surgeons performing MA TKA could consider limiting coronal alignment corrections of the limb and femoral joint line to within one phenotype to reduce the risk of low patient-reported satisfaction and function at 1-year.
Introduction
The goals of successful bi-compartmental knee arthroplasty are to achieve correct fit and positioning of the implant, while appropriately correcting the mechanical alignment of the leg after surgery. As these requirements are not always reliably fulfilled using off-the-shelf implant systems, newer approaches for bi-compartmental resurfacing have been explored.
Material and methods
In this article we report the radiographic results of 30 patients with anteromedial osteoarthritis (OA) who were treated with a novel patient-specific fixed-bearing bi-compartmental knee resurfacing system using custom-made implants and instruments. Utilizing standardized pre- and postoperative radiographic analyses (based on anterior-posterior and lateral, anterior-posterior weight-bearing full-length radiographs, patella skyline views and preoperative computed tomography (CT) scanning) implant fit and positioning as well as correction of the mechanical axis (hip-knee-ankle angle, HKA) were determined.
Results
On average, HKA was corrected from 173.4 ±3.47° preoperatively to 179.4 ±2.85° postoperatively. The coronal femoro-tibial angle was corrected on average 5.61°. The preoperative tibial slope measured on lateral views was 6.38 ±2.4°, while the average slope in the CT-based planning protocol (iView) was 6.14 ±2.40°. Postoperative lateral tibial slope was determined to be 5.77 ±1.97°. The thickness of the posterior femoral cuts was measured intraoperatively and, in all cases, corresponded well to the targeted thickness of the cuts provided by the iView. The joint line was preserved in all cases and the average Insall-Salvati index was 1.078 ±0.11 pre- and 1.072 ±0.11 postoperatively. The fit of the implant components measured by over- or underhang was excellent throughout (< 1.01 mm).
Conclusions
Custom-made bicompartmental knee arthroplasty can ensure optimized fitting and positioning of the implant with restoration of the leg axis. These implants could be considered as an alternative primary solution for knee surgeons treating bi-compartmental disease.