Background
To investigate the effects of B\(_1\)-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI).
Methods
Stenosis phantoms with different grades of stenosis (0%, 20%, 40%, 60%, 80%, and 100%; 5 mm inner vessel diameter) were produced using 3D printing (clear resin). Phantoms were imaged with four different concentrations of diluted Gd-DOTA representing established arterial concentrations after intravenous injection in humans. Samples were centrally positioned in a thorax phantom of 30 cm diameter filled with a custom-made liquid featuring dielectric properties of muscle tissue. MRI was performed on a 7 T whole-body system. 2D-gradient-echo sequences were acquired with an 8-channel transmit 16-channel receive (8 Tx / 16 Rx) cardiac array prototype coil with and without pTX mode. Measurements were compared to those obtained with identical scan parameters using a commercially available 1 Tx / 16 Rx single transmit coil (sTX). To assess reproducibility, measurements (n = 15) were repeated at different horizontal angles with respect to the B0-field.
Results
B\(_1\)-shimming and pTX markedly improved flip angle homogeneity across the thorax phantom yielding a distinctly increased signal-to-noise ratio (SNR) averaged over a whole slice relative to non-manipulated RF fields. Images without B\(_1\)-shimming showed shading artifacts due to local B\(_1\)\(^+\)-field inhomogeneities, which hampered stenosis quantification in severe cases. In contrast, B\(_1\)-shimming and pTX provided superior image homogeneity. Compared with a conventional sTX coil higher grade stenoses (60% and 80%) were graded significantly (p<0.01) more precise. Mild to moderate grade stenoses did not show significant differences. Overall, SNR was distinctly higher with B\(_1\)-shimming and pTX than with the conventional sTX coil (inside the stenosis phantoms 14%, outside the phantoms 32%). Both full and half concentration (10.2 mM and 5.1 mM) of a conventional Gd-DOTA dose for humans were equally suitable for stenosis evaluation in this phantom study.
Conclusions
B\(_1\)-shimming and pTX at 7 T can distinctly improve image homogeneity and therefore provide considerably more accurate MR image analysis, which is beneficial for imaging of small vessel structures.
Background:
Accurate preoperative assessment of the aortic annulus dimension is crucial for successful transcatheter aortic valve implantation (TAVI). In this study we validated a new method using two-dimensional transesophageal echocardiography (2D-TEE) for measurement of the aortic annulus prior to TAVI.
Methods:
We analysed 124 patients who underwent successful TAVI using a self-expandable prosthesis, divided equally into two groups; in the study group we used the cross sectional short axis 2D-TEE for measurement of the aortic annulus and in the control group we used the long axis 2D-TEE.
Results:
Both groups were comparable regarding the clinical parameters. On the other hand, patients in the study group had less left ventricular ejection fraction (38.9 % versus 45.6 %, p = 0.01). The aortic valve annulus was, although not statistically significant, smaller in the study group (21.58 versus 23.28 mm, p = 0.25). Post procedural quantification of the aortic regurgitation revealed that only one patient in both groups had severe aortic regurgitation (AR), in this patient the valve was implanted deep. The incidence of significant AR was higher in the control group (29.0 % versus 12.9 %, p = 0.027).
Conclusions:
Sizing of the aortic valve annulus using cross-sectional 2D-TEE offers a safe and plausible method for patients undergoing TAVI using the self-expandable prosthesis and is significantly superior to using long axis 2D-TEE.