Background
The role of cement-augmented screw fixation for calcaneal fracture treatment remains unclear. Therefore, this study was performed to biomechanically analyze screw osteosynthesis by reinforcement with either a calcium phosphate (CP)-based or polymethylmethacrylate (PMMA)-based injectable bone cement.
Methods
A calcaneal fracture (Sanders type IIA) including a central cancellous bone defect was generated in 27 synthetic bones, and the specimens were assigned to 3 groups. The first group was fixed with four screws (3.5 mm and 6.5 mm), the second group with screws and CP-based cement (Graftys (R) QuickSet; Graftys, Aix-en-Provence, France), and the third group with screws and PMMA-based cement (Traumacem (TM) V+; DePuy Synthes, Warsaw, IN, USA). Biomechanical testing was conducted to analyze peak-to-peak displacement, total displacement, and stiffness in following a standardized protocol.
Results
The peak-to-peak displacement under a 200-N load was not significantly different among the groups; however, peak-to-peak displacement under a 600- and 1000-N load as well as total displacement exhibited better stability in PMMA-augmented screw osteosynthesis compared to screw fixation without augmentation. The stiffness of the construct was increased by both CP- and PMMA-based cements.
Conclusion
Addition of an injectable bone cement to screw osteosynthesis is able to increase fixation strength in a biomechanical calcaneal fracture model with synthetic bones. In such cases, PMMA-based cements are more effective than CP-based cements because of their inherently higher compressive strength. However, whether this high strength is required in the clinical setting for early weight-bearing remains controversial, and the non-degradable properties of PMMA might cause difficulties during subsequent interventions in younger patients.
Intra- and interobserver reliability of glenoid fracture classifications by Ideberg, Euler and AO
(2018)
Background:
Representing 3%-5% of shoulder girdle injuries scapula fractures are rare. Furthermore, approximately 1% of scapula fractures are intraarticularfractures of the glenoid fossa. Because of uncertain fracture morphology and limited experience, the treatment of glenoid fossa fractures is difficult. The glenoid fracture classification by Ideberg (1984) and Euler (1996) is still commonly used in literature. In 2013 a new glenoid fracture classification was introduced by the AO. The purpose of this study was to examine the new AO classification in clinical practice in comparison with the classifications by Ideberg and Euler.
Methods:
In total CT images of 84 patients with glenoid fossa fractures from 2005 to 2018 were included. Parasagittal, paracoronary and axial reconstructions were examined according to the classifications of Ideberg, Euler and the AO by 3 investigators (orthopedic surgeon, radiologist, student of medicine) at three individual time settings. Inter- and intraobserver reliability of the three classification systems were ascertained by computing Inter- and Intraclass (ICCs) correlation coefficients using Spearman's rank correlation coefficient, 95%-confidence intervals as well as F-tests for correlation coefficients.
Results:
Inter- and intraobserver reliability for the AO classification showed a perspicuous coherence (R = 0.74 and R = 0.79). Low to moderate intraobserver reliability for Ideberg (R = 0.46) and Euler classification (R = 0.41) was found. Furthermore, data show a low Interobserver reliability for both Ideberg and Euler classification (R < 0.2). Both the Inter- and Intraclass reliability using AO is significantly higher than those using Ideberg and Euler (p < 0.05). Using the new AO classification, it was possible to find a proper class for every glenoid fossa fracture. On average, according to Euler classification 10 of 84 fractures were not classifiable whereas to Ideberg classification 21 of 84 fractures were not classifiable.
Conclusion:
The new AO classification system introduced 2013 facilitates reliable grading of glenoid fossa fractures with high inter- and intraobserver reliability in 84 patients using CT images. It should possibly be applied in order to enable a valid, reliable and consistent academic description of glenoid fossa fractures. The established classifications by Euler and Ideberg are not capable of providing a similar reliability.
Introduction
Proximal ulna fractures are common in orthopaedic surgery. Comminuted fractures require a high primary stability by the osteosynthesis, to allow an early functional rehabilitation as fast as possible, to reduce long-term limitations of range of motion. Classical dorsal plating is related to wound healing problems due to the prominence of the implant. New low-profile double plates are available addressing the soft tissue problems by positioning the plates at the medial and lateral side. This study analysed whether, under high loading conditions, these new double plates provide an equivalent stability as compared to the rigid olecranon locking compression plate (LCP).
Materials and methods
In Sawbones, Mayo Type IIB fractures were simulated and stabilized by plate osteosyntheses: In group one, two low-profile plates were placed. In group two, a single dorsal plate (LCP) was used. The bones was than cyclically loaded simulating flexion grades of 0°, 30°, 60° and 90° of the elbow joint with increasing tension forces (150 , 150 , 300 and 500 N). The displacement and fracture gap movement were recorded. In the end, in load-to-failure tests, load at failure and mode of failure were determined.
Results
No significant differences were found for the displacement and fracture gap widening during cyclic loading. Under maximum loading, the double plates revealed a comparable load at failure like the single dorsal plate (LCP). The double plates failed with a proximal screw pull-out of the plate, whereas in the LCP group, in 10 out of 12 specimens the mode of failure was a diaphyseal shaft fracture at the distal plate peak.
Conclusion
Biomechanically, the double plates are a good alternative to the dorsal LCP providing a high stability under high loading conditions and, at the same, time reducing the soft tissue irritation by a lateral plate position.
Background
Morphology and glenoid involvement determine the necessity of surgical management in scapula fractures. While being present in only a small share of patients with shoulder trauma, numerous classification systems have been in use over the years for categorization of scapula fractures. The purpose of this study was to evaluate the established AO/OTA classification in comparison to the classification system of Euler and Rüedi (ER) with regard to interobserver reliability and confidence in clinical practice.
Methods
Based on CT imaging, 149 patients with scapula fractures were retrospectively categorized by two trauma surgeons and two radiologists using the classification systems of ER and AO/OTA. To measure the interrater reliability, Fleiss kappa (κ) was calculated independently for both fracture classifications. Rater confidence was stated subjectively on a five-point scale and compared with Wilcoxon signed rank tests. Additionally, we computed the intraclass correlation coefficient (ICC) based on absolute agreement in a two-way random effects model to assess the diagnostic confidence agreement between observers.
Results
In scapula fractures involving the glenoid fossa, interrater reliability was substantial (κ = 0.722; 95% confidence interval [CI] 0.676–0.769) for the AO/OTA classification in contrast to moderate agreement (κ = 0.579; 95% CI 0.525–0.634) for the ER classification system. Diagnostic confidence for intra-articular fracture patterns was superior using the AO/OTA classification compared to ER (p < 0.001) with higher confidence agreement (ICC: 0.882 versus 0.831). For extra-articular fractures, ER (κ = 0.817; 95% CI 0.771–0.863) provided better interrater reliability compared to AO/OTA (κ = 0.734; 95% CI 0.692–0.776) with higher diagnostic confidence (p < 0.001) and superior agreement between confidence ratings (ICC: 0.881 versus 0.912).
Conclusions
The AO/OTA classification is most suitable to categorize intra-articular scapula fractures with glenoid involvement, whereas the classification system of Euler and Rüedi appears to be superior in extra-articular injury patterns with fractures involving only the scapula body, spine, acromion and coracoid process.
The coronal unicondylar fracture of the distal femur (AO 33-B3) is a rare intraarticular injury within the weight bearing area of the knee, initially described by Albert Hoffa in 1904. We report an unusual combination of a Hoffa fracture with lateral patellar dislocation in a young adult. Our patient sustained the injury by a sudden twist of his leg during sports. He presented clinically with knee swelling, dislocation of the patella, and localized tenderness; unable to bare weight. After plane radiograph confirmed the injury, manual reduction of the patella was done by hyperextension of the knee and medialward pressure. Afterwards, a CT scan and MRI were conducted. The injury was surgically treated with lag-screws, locking-plate and MPFL-reconstruction.
The calcaneal beak fracture is a rare avulsion fracture of the tuber calcanei characterized by a solid bony fragment at the Achilles tendon insertion. Treatment usually requires osteosynthesis. However, lack of biomechanical understanding of the ideal fixation technique persists. A beak fracture was simulated in synthetic bones and assigned to five different groups of fixation: A) 6.5-mm partial threaded cannulated screws, B) 4.0-mm partial threaded cannulated screws, C) 5.0-mm headless cannulated compression screws, D) 2.3-mm locking plate, and E) 2.8-mm locking plate. Different traction force levels were applied through an Achilles tendon surrogate in a material-testing machine on all stabilized synthetic bones. Outcome measures were peak-to-peak displacement, total displacement, plastic deformation, stiffness, visual-fracture-line displacement, and mode of implant failure. The 2.3- and 2.8-mm plating groups showed a high drop-out rate at 100 N tension force and failed under higher tension levels of 200 N. The fracture fixation using 4.0-mm partial threaded screws showed a significantly higher repair strength and was able to withhold cyclic loading up to 300 N. The lowest peak-to-peak displacement and the highest load-to-failure and stiffness were provided by fracture fixation using 6.5-mm partial threaded cannulated screws or 5.0-mm headless cannulated compression screws. As anticipated, large 6.5-mm screw diameters provide the best biomechanical fixation. Surprisingly, the 5.0-mm headless cannulated compression screws yield reliable stability despite the absent screw head and washer. When such large screws cannot be applied, 4.0-mm screws also allow reasonable fixation strength. Plate fixation should be implemented with precaution and in combination with a restrictive postoperative motion protocol. Finally, clinical cases about the surgical application and recovery are included.
Die Ellenbogenfraktur ist eine häufige Fraktur der oberen Extremität. Um die Funktion des Ellenbogengelenks zu erhalten, sind eine gute Reposition und eine stabile Osteosynthese, sowie eine frühfunktionelle Nachbehandlung der Olekranonfraktur erforderlich.
Zur Versorgung von komplexen Frakturen des Olekranons mit mehreren Fragmenten stehen prinzipiell zwei Osteosyntheseverfahren zur Verfügung. Die dorsale Plattenosteosynthese ist ein schon länger zur Verfügung stehendes Verfahren. Bei der klinischen Anwendung treten jedoch immer wieder Weichteilirritationen durch die dorsale Lage der recht massiven Platte auf. Seit einigen Jahren existiert eine weitere Osteosynthese, bei der zwei flache laterale Olekranonplatten verwendet werden. Durch die Positionierung sollen Wundheilungsstörungen reduziert werden. Ziel dieser Arbeit war die Entwicklung eines biomechanischen Testmodells und die Stabilitätsprüfung der zwei Verfahren zur Plattenosteosynthese bei mehrfragmentären Olekranonfrakturen in einem klinisch relevanten Belastungsumfang.
Die zwei seitlich angebrachten Olekranon Zwillingsplatten stellen eine gute Alternative zu der singulären dorsalen LCP- Olekranonplatte dar, da sie weniger Weichteilirritationen verursachen und im biomechanischen Model unter hoher Belastung eine vergleichbare Stabilität gewährleisten.
In this study, the impact of reconstruction sharpness on the visualization of the appendicular skeleton in ultrahigh-resolution (UHR) photon-counting detector (PCD) CT was investigated. Sixteen cadaveric extremities (eight fractured) were examined with a standardized 120 kVp scan protocol (CTDI\(_{vol}\) 10 mGy). Images were reconstructed with the sharpest non-UHR kernel (Br76) and all available UHR kernels (Br80 to Br96). Seven radiologists evaluated image quality and fracture assessability. Interrater agreement was assessed with the intraclass correlation coefficient. For quantitative comparisons, signal-to-noise-ratios (SNRs) were calculated. Subjective image quality was best for Br84 (median 1, interquartile range 1–3; p ≤ 0.003). Regarding fracture assessability, no significant difference was ascertained between Br76, Br80 and Br84 (p > 0.999), with inferior ratings for all sharper kernels (p < 0.001). Interrater agreement for image quality (0.795, 0.732–0.848; p < 0.001) and fracture assessability (0.880; 0.842–0.911; p < 0.001) was good. SNR was highest for Br76 (3.4, 3.0–3.9) with no significant difference to Br80 and Br84 (p > 0.999). Br76 and Br80 produced higher SNRs than all kernels sharper than Br84 (p ≤ 0.026). In conclusion, PCD-CT reconstructions with a moderate UHR kernel offer superior image quality for visualizing the appendicular skeleton. Fracture assessability benefits from sharp non-UHR and moderate UHR kernels, while ultra-sharp reconstructions incur augmented image noise.