Direct cooling of the catheter tip increases safety for CMR-guided electrophysiological procedures
(2012)
Background: One of the safety concerns when performing electrophysiological (EP) procedures under magnetic resonance (MR) guidance is the risk of passive tissue heating due to the EP catheter being exposed to the radiofrequency (RF) field of the RF transmitting body coil. Ablation procedures that use catheters with irrigated tips are well established therapeutic options for the treatment of cardiac arrhythmias and when used in a modified mode might offer an additional system for suppressing passive catheter heating.
Methods: A two-step approach was chosen. Firstly, tests on passive catheter heating were performed in a 1.5 T Avanto system (Siemens Healthcare Sector, Erlangen, Germany) using a ASTM Phantom in order to determine a possible maximum temperature rise. Secondly, a phantom was designed for simulation of the interface between blood and the vascular wall. The MR-RF induced temperature rise was simulated by catheter tip heating via a standard ablation generator. Power levels from 1 to 6 W were selected. Ablation duration was 120 s with no tip irrigation during the first 60 s and irrigation at rates from 2 ml/min to 35 ml/min for the remaining 60 s (Biotronik Qiona Pump, Berlin, Germany). The temperature was measured with fluoroscopic sensors (Luxtron, Santa Barbara, CA, USA) at a distance of 0 mm, 2 mm, 4 mm, and 6 mm from the catheter tip. Results: A maximum temperature rise of 22.4 degrees C at the catheter tip was documented in the MR scanner. This temperature rise is equivalent to the heating effect of an ablator's power output of 6 W at a contact force of the weight of 90 g (0.883 N). The catheter tip irrigation was able to limit the temperature rise to less than 2 degrees C for the majority of examined power levels, and for all examined power levels the residual temperature rise was less than 8 degrees C.
Conclusion: Up to a maximum of 22.4 degrees C, the temperature rise at the tissue surface can be entirely suppressed by using the catheter's own irrigation system. The irrigated tip system can be used to increase MR safety of EP catheters by suppressing the effects of unwanted passive catheter heating due to RF exposure from the MR scanner.
Background: Cystic fibrosis (CF) patients would benefit from a safe and effective tool to detect early-stage, regional lung disease to allow for early intervention. Magnetic Resonance Imaging (MRI) is a safe, non-invasive procedure capable of providing quantitative assessments of disease without ionizing radiation. We developed a rapid normalized T1 MRI technique to detect regional lung disease in early-stage CF patients.
Materials and Methods: Conventional multislice, pulmonary T1 relaxation time maps were obtained for 10 adult CF patients with normal spirometry and 5 healthy non-CF control subjects using a rapid Look-Locker MRI acquisition (5 seconds/imaging slice). Each lung absolute T1 map was separated into six regions of interest (ROI) by manually selecting upper, central, and lower lung regions in the left and right lungs. In order to reduce the effects of subject-to-subject variation, normalized T1 maps were calculated by dividing each pixel in the absolute T1 maps by the mean T1 time in the central lung region. The primary outcome was the differences in mean normalized T1 values in the upper lung regions between CF patients with normal spirometry and healthy volunteers.
Results: Normalized T1 (nT1) maps showed visibly reduced subject-to-subject variation in comparison to conventional absolute T1 maps for healthy volunteers. An ROI analysis showed that the variation in the nT1 values in all regions was <= 2% of the mean. The primary outcome, the mean (SD) of the normalized T1 values in the upper right lung regions, was significantly lower in the CF subjects [.914 (.037)] compared to the upper right lung regions of the healthy subjects [.983 (.003)] [difference of .069 (95% confidence interval .032-.105); p=.001). Similar results were seen in the upper left lung region.
Conclusion: Rapid normalized T1 MRI relaxometry obtained in 5 seconds/imaging slice may be used to detect regional early-stage lung disease in CF patients.