Blood oxygen saturation is an important clinical parameter, especially in postoperative hospitalized patients, monitored in clinical practice by arterial blood gas (ABG) and/or pulse oximetry that both are not suitable for a long-term continuous monitoring of patients during the entire hospital stay, or beyond. Technological advances developed recently for consumer-grade fitness trackers could—at least in theory—help to fill in this gap, but benchmarks on the applicability and accuracy of these technologies in hospitalized patients are currently lacking. We therefore conducted at the postanaesthesia care unit under controlled settings a prospective clinical trial with 201 patients, comparing in total >1,000 oxygen blood saturation measurements by fitness trackers of three brands with the ABG gold standard and with pulse oximetry. Our results suggest that, despite of an overall still tolerable measuring accuracy, comparatively high dropout rates severely limit the possibilities of employing fitness trackers, particularly during the immediate postoperative period of hospitalized patients.
Highlights
•The accuracy of O2 measurements by fitness trackers is tolerable (RMSE ≲4%)
•Correlation with arterial blood gas measurements is fair to moderate (PCC = [0.46; 0.64])
•Dropout rates of fitness trackers during O2 monitoring are high (∼1/3 values missing)
•Fitness trackers cannot be recommended for O2 measuring during critical monitoring
Physical and mental well-being during the COVID-19 pandemic is typically assessed via surveys, which might make it difficult to conduct longitudinal studies and might lead to data suffering from recall bias. Ecological momentary assessment (EMA) driven smartphone apps can help alleviate such issues, allowing for in situ recordings. Implementing such an app is not trivial, necessitates strict regulatory and legal requirements, and requires short development cycles to appropriately react to abrupt changes in the pandemic. Based on an existing app framework, we developed Corona Health, an app that serves as a platform for deploying questionnaire-based studies in combination with recordings of mobile sensors. In this paper, we present the technical details of Corona Health and provide first insights into the collected data. Through collaborative efforts from experts from public health, medicine, psychology, and computer science, we released Corona Health publicly on Google Play and the Apple App Store (in July 2020) in eight languages and attracted 7290 installations so far. Currently, five studies related to physical and mental well-being are deployed and 17,241 questionnaires have been filled out. Corona Health proves to be a viable tool for conducting research related to the COVID-19 pandemic and can serve as a blueprint for future EMA-based studies. The data we collected will substantially improve our knowledge on mental and physical health states, traits and trajectories as well as its risk and protective factors over the course of the COVID-19 pandemic and its diverse prevention measures.
Background: Over the recent years, technological advances of wrist-worn fitness trackers heralded a new era in the continuous monitoring of vital signs. So far, these devices have primarily been used for sports.
Objective: However, for using these technologies in health care, further validations of the measurement accuracy in hospitalized patients are essential but lacking to date.
Methods: We conducted a prospective validation study with 201 patients after moderate to major surgery in a controlled setting to benchmark the accuracy of heart rate measurements in 4 consumer-grade fitness trackers (Apple Watch 7, Garmin Fenix 6 Pro, Withings ScanWatch, and Fitbit Sense) against the clinical gold standard (electrocardiography).
Results: All devices exhibited high correlation (r≥0.95; P<.001) and concordance (rc≥0.94) coefficients, with a relative error as low as mean absolute percentage error <5% based on 1630 valid measurements. We identified confounders significantly biasing the measurement accuracy, although not at clinically relevant levels (mean absolute error<5 beats per minute).
Conclusions: Consumer-grade fitness trackers appear promising in hospitalized patients for monitoring heart rate.