Background
Boluses are routinely used in radiotherapy to modify surface doses. Nevertheless, considerable dose discrepancies may occur in some cases due to fit inaccuracy of commercially available standard flat boluses. Moreover, due to the simple geometric design of conventional boluses, also surrounding healthy skin areas may be unintentionally covered, resulting in the unwanted dose buildup. With the fused deposition modeling (FDM) technique, there is a simple and possibly cost-effective way to solve these problems in routine clinical practice. This paper presents a procedure of self-manufacturing bespoke patient-specific silicone boluses and the evaluation of buildup and fit accuracy in comparison to standard rectangular commercially available silicone boluses.
Methods
3D-conformal silicone boluses were custom-built to cover the surgical scar region of 25 patients who received adjuvant radiotherapy of head and neck cancer at the University Hospital Würzburg. During a standard CT-based planning procedure, a 5-mm-thick 3D bolus contour was generated to cover the radiopaque marked surgical scar with an additional safety margin. From these digital contours, molds were 3D printed and poured with silicone. Dose measurements for both types of boluses were performed with radiochromic films (EBT3) at three points per patient—at least one aimed to be in the high-dose area (scar) and one in the lower-dose area (spared healthy skin). Surface–bolus distance, which ideally should not be present, was determined from cone-beam CT performed for positioning control. The dosimetric influence of surface–bolus distance was also determined on slab phantom for different field sizes. The trial was performed with hardware that may be routinely available in every radiotherapy department, with the exception of the 3D printer. The required number of patients was determined based on the results of preparatory measurements with the help of the statistical consultancy of the University of Würzburg. The number of measuring points represents the total number of patients.
Results
In the high-dose area of the scar, there was a significantly better intended dose buildup of 2.45% (95%CI 0.0014–0.0477, p = 0.038, N = 30) in favor of a 3D-conformal bolus. Median distances between the body surface and bolus differed significantly between 3D-conformal and commercially available boluses (3.5 vs. 7.9 mm, p = 0.001). The surface dose at the slab phantom did not differ between commercially available and 3D-conformal boluses. Increasing the surface–bolus distance from 5 to 10 mm decreased the surface dose by approximately 2% and 11% in the 6 × 6- and 3 × 3-cm2 fields, respectively. In comparison to the commercially available bolus, an unintended dose buildup in the healthy skin areas was reduced by 25.9% (95%CI 19.5–32.3, p < 0.01, N = 37) using the 3D-conformal bolus limited to the region surrounding the surgical scar.
Conclusions
Using 3D-conformal boluses allows a comparison to the commercially available boluses’ dose buildup in the covered areas. Smaller field size is prone to a larger surface–bolus distance effect. Higher conformity of 3D-conformal boluses reduces this effect. This may be especially relevant for volumetric modulated arc therapy (VMAT) and intensity-modulated radiotherapy (IMRT) techniques with a huge number of smaller fields. High conformity of 3D-conformal boluses reduces an unintended dose buildup in healthy skin. The limiting factor in the conformity of 3D-conformal boluses in our setting was the immobilization mask, which was produced primarily for the 3D boluses. The mask itself limited tight contact of subsequently produced 3D-conformal boluses to the mask-covered body areas. In this respect, bolus adjustment before mask fabrication will be done in the future setting.
Simple Summary
Prostate cancer often relapses after initial radical prostatectomy, and salvage radiotherapy offers a second chance of cure for relapsed patients. Modern imaging techniques, especially prostate-specific membrane antigen positron emission tomography/computed tomography (PSMA PET/CT), enable radiation oncologists to target radiotherapy at the involved sites of disease. In a group of patients, PSMA PET/CT imaging can detect a macroscopic local recurrence with or without locoregional lymph node metastasis. In these cases, an escalation of the radiotherapy dose is often considered for controlling the visible tumor mass. As the evidence for dose-escalated salvage radiotherapy for macroscopic recurrent prostate cancer after PSMA PET/CT imaging is still limited, we address this topic in the current analysis. We found that the outcome of patients with dose-escalated salvage radiotherapy for macroscopic prostate cancer recurrence is encouragingly favorable, while the toxicity is very limited.
Abstract
Background: The purpose of this study was to access the oncological outcome of prostate-specific membrane antigen positron emission tomography (PSMA PET/CT)-guided salvage radiotherapy (SRT) for localized macroscopic prostate cancer recurrence. Methods: Between February 2010 and June 2021, 367 patients received SRT after radical prostatectomy. Out of the 367 screened patients, 111 patients were staged by PSMA PET/CT before SRT. A total of 59 out of these 111 (53.2%) patients were treated for PSMA PET-positive macroscopic prostatic fossa recurrence. Dose-escalated SRT was applied with a simultaneous integrated boost at a median prescribed dose of 69.3 Gy (IQR 69.3–72.6 Gy). The oncological outcome was investigated using Kaplan-Meier and Cox regression analyses. The genitourinary (GU)/gastrointestinal (GI) toxicity evaluation utilized Common Toxicity Criteria for Adverse Events (version 5.0). Results: The median follow-up was 38.2 months. The three-year biochemical progression-free survival rate was 89.1% (95% CI: 81.1–97.8%) and the three-year metastasis-free survival rate reached 96.2% (95% CI: 91.2–100.0%). The cumulative three-year late grade 3 GU toxicity rate was 3.4%. No late grade 3 GI toxicity occurred. Conclusions: Dose-escalated PSMA PET/CT-guided salvage radiotherapy for macroscopic prostatic fossa recurrence resulted in favorable survival and toxicity rates.
Purpose: Any Linac will show geometric imprecisions, including non-ideal alignment of the gantry, collimator and couch axes, and gantry sag or wobble. Their angular dependence can be quantified and resulting changes of the dose distribution predicted (Wack, JACMP 20(5), 2020). We analyzed whether it is feasible to correct geometric shifts during treatment planning. The successful implementation of such a correction procedure was verified by measurements of different stereotactic treatment plans.
Methods: Isocentric shifts were quantified for two Elekta Synergy Agility Linacs using the QualiForMed ISO-CBCT+ module, yielding the shift between kV and MV isocenters, the gantry flex and wobble as well as the positions of couch and collimator rotation axes. Next, the position of each field's isocenter in the Pinnacle treatment planning system was adjusted accordingly using a script. Fifteen stereotactic treatment plans of cerebral metastases (0.34 to 26.53 cm3) comprising 9–11 beams were investigated; 54 gantry and couch combinations in total. Unmodified plans and corrected plans were measured using the Sun Nuclear SRS-MapCHECK with the Stereophan phantom and evaluated using gamma analysis.
Results: Geometric imprecisions, such as shifts of up to 0.8 mm between kV and MV isocenter, a couch rotation axis 0.9 mm off the kV isocente,r and gantry flex with an amplitude of 1.1 mm, were found. For eight, mostly small PTVs D98 values declined more than 5% by simulating these shifts. The average gamma (2%/2 mm, absolute, global, 20% threshold) was reduced from 0.53 to 0.31 (0.32 to 0.30) for Linac 1 (Linac 2) when including the isocentric corrections. Thus, Linac 1 reached the accuracy level of Linac 2 after correction.
Conclusion: Correcting for Linac geometric deviations during the planning process is feasible and was dosimetrically validated. The dosimetric impact of the geometric imperfections can vary between Linacs and should be assessed and corrected where necessary.