543 Analytische Chemie
Refine
Has Fulltext
- yes (19)
Is part of the Bibliography
- yes (19)
Year of publication
Document Type
- Doctoral Thesis (17)
- Journal article (2)
Keywords
- HPLC (7)
- Charged Aerosol Detection (4)
- Chemische Reinheit (4)
- Instrumentelle Analytik (4)
- Impurity Profiling (3)
- Pharmaceutical Analysis (3)
- Aminosäuren (2)
- Chromatographie (2)
- Malaria (2)
- Reinheitsanalytik (2)
Institute
- Institut für Pharmazie und Lebensmittelchemie (13)
- Institut für Funktionsmaterialien und Biofabrikation (2)
- Graduate School of Life Sciences (1)
- Institut für Anorganische Chemie (1)
- Institut für Organische Chemie (1)
- Lehrstuhl für Tissue Engineering und Regenerative Medizin (1)
- Medizinische Klinik und Poliklinik I (1)
Sonstige beteiligte Institutionen
Non-contact, Label-free Monitoring of Cells and Extracellular Matrix using Raman Spectroscopy
(2012)
Non-destructive, non-contact and label-free technologies to monitor cell and tissue cultures are needed in the field of biomedical research.1-5 However, currently available routine methods require processing steps and alter sample integrity. Raman spectroscopy is a fast method that enables the measurement of biological samples without the need for further processing steps. This laser-based technology detects the inelastic scattering of monochromatic light.6 As every chemical vibration is assigned to a specific Raman band (wavenumber in cm-1), each biological sample features a typical spectral pattern due to their inherent biochemical composition.7-9 Within Raman spectra, the peak intensities correlate with the amount of the present molecular bonds.1 Similarities and differences of the spectral data sets can be detected by employing a multivariate analysis (e.g. principal component analysis (PCA)).10
Here, we perform Raman spectroscopy of living cells and native tissues. Cells are either seeded on glass bottom dishes or kept in suspension under normal cell culture conditions (37 °C, 5% CO2) before measurement. Native tissues are dissected and stored in phosphate buffered saline (PBS) at 4 °C prior measurements. Depending on our experimental set up, we then either focused on the cell nucleus or extracellular matrix (ECM) proteins such as elastin and collagen. For all studies, a minimum of 30 cells or 30 random points of interest within the ECM are measured. Data processing steps included background subtraction and normalization.
Mass spectrometry-based quantification of steroids for the diagnostic workup of adrenal tumors
(2023)
Tumors of the adrenal gland belong to the most frequent neoplasms in humans with a prevalence of 3–10 % in adults. The aim of the diagnostic workup is the identification of potentially hormone-secreting and / or malignant tumors, because most of these tumors will require surgical resection. Malignant adrenocortical carcinomas (ACC) are very rare and associated with a poor prognosis in advanced stages, therefore, an early and accurate diagnosis is crucial.
Within this thesis, two liquid chromatography tandem mass spectrometry (LC-MS/MS) methods for the quantification of steroids in different biomaterials were developed to improve the diagnostic workup of adrenal tumors.
First, an LC-MS/MS method for the simultaneous quantification of cortisol and dexamethasone in serum samples after dexamethasone suppression test (DST) was developed, validated, and applied to 400 clinical samples. Newly established method-specific threshold concentrations for cortisol and dexamethasone increased DST specificity from 67.5 % to 92.4 % while preserving 100 % sensitivity.
Second, an LC-MS/MS method for the quantification of eleven urinary steroids was developed and validated to improve the differentiation between ACC and adrenocortical adenomas (ACA). A decision tree requiring only two steroids was trained for classification and tested based on 24 h urine samples from 268 patients with adrenal tumor. Malignancy was excluded with a negative predictive value of 100 % in an independent validation cohort of 84 samples of 24-h urine. A newly proposed simplified diagnostic workflow with urinary steroid profiling as first tier test could obviate additional adrenal-specific imaging in 42 of 64 patients with ACA.
The new DST method is already in clinical use at the University Hospital Würzburg, whereas the classification model based on urinary steroid profiling will require prospective validation in a larger cohort.
Der Gruppe der Macrogole sowie den darauf basierenden Abkömmlingen, den Macrogolfettalkoholethern, Macrogolfettsäureestern und Polysorbaten, kommt in der modernen Galenik eine wichtige Rolle zu. Dienten sie vormals nur als gewöhnliche Emulgatoren, so finden sie heutzutage vor allem im Bereich der gezielten Wirkstofffreisetzung, der Erhöhung der Bioverfügbarkeit sowie als Löslichkeitsvermittler komplexer Systeme Anwendung. Diese vielschichtigen Anwendungsgebiete erfordern, auch aufgrund der polydispersen Strukturen der Macrogole, eine reproduzierbare und aussagekräftige Analytik.
Das Europäische Arzneibuch (Ph. Eur.) bietet zur Charakterisierung der Hilfsstoffe eine Handvoll Messgrößen, die sog. Fettkennzahlen, die eine Größenordnung vorhandener funktioneller Gruppen liefern. Zu diesen gehören Werte wie Hydroxylzahl, Iodzahl, Peroxidzahl oder Säurezahl. Diese bieten zwar einen Überblick über den Größenbereich der mittleren Kettenlängen oder einen möglichen Abbau der Strukturen, beispielsweise durch Autoxidation, jedoch geben sie keine Auskunft über die Polymerverteilung. Insbesondere diese kann jedoch, je nach Herstellungsweise, stark variieren. Außerdem ist die Methodik der Fettkennzahlenbestimmungen aufgrund der strikten Reaktionsabläufe und zahlreicher Reaktionsschritte einerseits sehr zeitaufwändig und andererseits anfällig für Fehler.
Die HPLC hat, insbesondere aufgrund der Automation, bereits seit Jahren den Status des Goldstandards in der pharmazeutischen Analytik inne. Gekoppelt mit der UV-Detektion bietet sie für zahlreiche Wirkstoffe die Möglichkeit zur schnellen, einfachen und robusten Analyse. Im Bereich der Hilfsstoffe verbreitet sich die HPLC-Analytik langsamer, da viele Hilfsstoffe keinen Chromophor aufweisen. Eine Anwendung der hochsensitiven Massenspektrometrie wäre zwar zur Detektion geeignet, würde sich für die Routineanwendung jedoch als zu komplex und kostenintensiv gestalten. Doch mit der Entwicklung der Aerosol-basierten Detektoren wie dem ELSD (evaporative light scattering detector), dem CAD (charged aerosol detector) und dem NQADTM (nano quantity aerosol detector) wurde auch für nicht-chromophore Substanzen ein Einsatz der HPLC möglich.
Die vorliegende Arbeit befasste sich mit der Entwicklung einer HPLC-CAD-Methode, die eine möglichst große Bandbreite der Macrogole und der darauf basierenden Hilfsstoffe erfassen kann. Die Trennung erfolgte an einer C18-Trennsäule. Es wurde eine Gradienten-Methode entwickelt, die aus mehreren linearen Gradientenstufen zusammengesetzt wurde, um verschiedene Kettenlängen der Polymere besser voneinander zu trennen. Als mobile Phasen dienten Wasser und Acetonitril, denen jeweils 0.1 % Ameisensäure zugesetzt wurden.
Es konnten Macrogole im Bereich PEG 300 bis PEG 3000 mit akzeptabler Auflösung aufgetrennt werden. Diese Ergebnisse wurden für PEG 300 – 1500 mittels Massenspektrometrie verifiziert. Es konnten fünf gesättigte und zwei ungesättigte Fettsäuren, sowie zwei Fettalkohole verschiedener Kettenlängen voneinander getrennt werden. Es wurden 13 Macrogol-basierte Hilfsstoffe mit der entwickelten Methode untersucht und erfolgreich getrennt. Die Macrogolfettalkoholether, -stearate und Polysorbate wurden insoweit aufgetrennt, dass die Polymerverteilung beobachtet werden konnte.
Freie PEGs in den Hilfsstoffen wurden getrennt und identifiziert. Anhand dieser konnten unterschiedliche Herstellungsweisen zugeordnet werden. Abhängig von der mittleren Kettenlänge der verarbeiteten PEGs konnten teilweise die freien Fettsäuren bzw. -alkohole von den Estern bzw. Ethern getrennt und identifiziert werden. Im Bereich der kürzeren mittleren Kettenlängen wurden die freien Fettsäuren und -alkohole von den Estern und Ethern überlagert.
Macrogolglycerolhydroxystearat (Cremophor® RH40) wurde in seine Komponenten aufgetrennt, mit Ausnahme der linearen Monoester, die mit den freien PEGs partiell koeluierten und die Glyceroltriester, die Größenausschlusseffekte zeigten.
Die Methode wurde für Stabilitätsuntersuchungen der ungesättigten Fettsäuren, Öl- und Linolsäure, eingesetzt. Hierzu wurden diese Säuren in Lösung chemisch (Wasserstoffperoxid) und thermisch (60 °C) gestresst und in bestimmten Zeitabständen analysiert. Es zeigte sich ein zeit- und temperaturabhängiger Abbau. Die teilweise Zuordnung der Abbauprodukte erfolgte durch Bestimmung des m/z mittels Massenspektrometrie. Die Methode war geeignet, um das Ausmaß eines oxidativen Abbaus von der Hauptsubstanz zu trennen und strukturell einzuordnen.
Generell bietet die Methode eine gute Basis, die eine Vielzahl an Substanzgruppen erfassen und charakterisieren kann. Sie bietet eine Ergänzung der Fettkennzahlen, die einen verringerten Arbeitsaufwand mit sich bringt. Für spezifischere Betrachtungen (Langzeitstabilität, verwandte Substanzgruppen) stellt sie einen guten Ausgangspunkt dar.
The aim of this work was to synthesize and functionalize different bio-relevant nanomaterials like silica-coated superparamagnetic iron oxide nanoparticles (SPIONs) as contrast agents for T2 magnetic resonance imaging (MRI) and detonation nanodiamond (DND) with the neurohormone peptide allatostatin 1 (ALST1) and a fluorescent dye. Analytical techniques for the determination and quantification of surface functional groups like amines, azides, and peptides were also developed and established.
Thus, in the first part of the work, a TGF-1 binding peptide and allatostatin 1 (ALST1), both supposed to act as active tumour targeting vectors, were synthesized by solid-phase peptide synthesis (SPPS) and characterized by high pressure liquid chromatography (HPLC) and mass spectrometry. Then, azide-functionalized silica nanoparticles were synthesized by the Stöber process and characterized by transmission electron microscopy (TEM) and infrared spectroscopy (IR). The surface loading of amine and azide groups was determined by a new protocol. The azide groups were reduced with sodium boronhydride to amine and then functionalized with Fmoc-Rink Amide linker according to a standard SPPS protocol. Upon cleavage of Fmoc by piperidine, the resulting dibenzofulvene and its piperidine adduct were quantified by UV/Vis spectroscopy and used to determine the amount of amine groups on the nanoparticle surface. Then, ALST1 and related tyrosine- and phenylalanine substituted model peptides were conjugated to the azide-functionalized silica nanoparticles by copper(I)-catalyzed azide-alkyne dipolar cycloaddition (CuAAC). The successful peptide conjugation was demonstrated by the Pauly reaction, which however is only sensitive to histidine- and tyrosine-containing peptides. As a more general alternative, the acid hydrolysis of the peptides to their individual amino acid building blocks followed by derivatization with phenyl isothiocyanate (PITC) allowed the separation, determination, and quantification of the constituent amino acids by HPLC.
In the second part of the work, amine- and azide-functionalized silica-coated superparamagnetic iron oxide nanoparticles (SPIONs) were synthesized by co-precipitation and subsequent silica-coated based on the Stöber process and characterized by TEM and IR. The amine surface loading was determined by the method already established for the pure silica systems. The azide surface loading could also be quantified by reduction with sodium boronhydride to amine groups and then conjugation to Fmoc-Rink amide linker. Upon cleavage of Fmoc with piperidine, the total amine surface loading was obtained. The amount of azide surface groups was then determined from the difference of the total amine surface loading and the amine surface loading. Thus, it was possible to quantify both amine and azide surface groups on a single nanoparticle system. Superparamagnetic iron oxide nanoparticles (SPIONs) are potent T2 contrast agents for magnetic resonance imaging (MRI). Due to their natural metabolism after injection into the blood stream, SPIONs mostly end up inside macrophages, liver, spleen or kidneys. To generate a potential target-specific SPION-based T2 contrast agent for MRI, the neurohormone peptide ALST1 was conjugated by CuAAC to the azide- and amine functionalized superparamagnetic iron oxide nanoparticles, since ALST1 is supposed to target difficult-to-treat neuroendocrinic tumours due to its analogy to galanin and somastatin receptor ligands. The organic fluorescent dye cyanine 5 (Cy5) was also conjugated to the silica-coated superparamagnetic iron oxide nanoparticles (SPIONs) via a NHS-ester to the amines to enable cell uptake studies by fluorescence microscopy. These constructs were characterized by TEM, dynamic light scattering (DLS), and IR. The amino acids of the conjugated ALST1 were determined by the HPLC method as described before for peptide-modified silica nanoparticle surfaces. Then, the relaxivity r2 was measured at 7 T. However, a r2 value of 27 L/mmolFe·s for the dual ALST1-/Cy5-functionalized silica-coated SPIONs was not comparable to T2 contrast agents in clinical use, since their relaxivity is commonly determined at 1.5 T, and no such instrument was available. However, it can be assumed that the synthesized dual
ALST1-/Cy5-functionalized silica-coated SPION would show a lower r2 at 1.5 T than at 7T. Commercial T2 MRI contrast agents like VSOP-C184 from Ferropharm show at r2 values of about 30 L/mmolFe·s at 1.5 T. Still, the relaxivity of the new material has some potential for application as a T2 contrast agent. Then, the material was used in cell uptake studies by fluorescence microscopy with the conjugated Cy5 dye as a probe. The dual
ALST1-/Cy5-functionalized silica-coated SPION showed a high degree of agglomeration with no cellular uptake unlike described for ALST1-functionalized nanoparticles in literature. It is assumed that upon agglomeration of the particles, constructs form which are unable to be internalized by the cellular endocytotic pathways anymore. As a future perspective, the tendency of the particle to agglomerate should be reduced by changing the coating material to polyethylene glycol (PEG) or chitosan, which are known to be bio-compatible, bio-degradable and prevent agglomeration.
In the third part of the work, the rhenium compound [ReBr(CO)3(L)] with L = 2-phenyl-1H-imidazo[4,5-f][1,10]phenanthroline and its manganese analogue were synthesized by heating the ligand and rhenium pentacarbonyl bromide or and manganese pentacarbonyl bromide respectively, in toluene. However, [MnBr(CO)3(L)] was unstable upon illumination by UV light at 365 nm. Thus, it was dismissed for further application. The photophysical properties of [ReBr(CO)3(L)] were explored, by determination of the excited-state life time by the time-correlated single-photon counting (TCSPC) method and the quantum yield by a fluorescence spectrometer equipped with an integration sphere. A value of = 455 ns, a Stokes shift of 197 nm and a rather low quantum yield =were found. Metal complexes are supposed to have superior properties compared to organic dyes due to their large Stokes shifts, long excited-state life times, and high quantum yields. Thus, amine- and azide-functionalized detonation nanodiamond (DND) as an alternative biological inert carrier system was functionalized with ALST1 to enhance its cell uptake properties. A luminescent probe for cell uptake studies using fluorescence microscopy was also attached, either based on the new rhenium complex or the commercially available organic dye Cy5, respectively. The aldehyde-functionalized rhenium complex was conjugated to the DND via oxime ligation, which is known to be a mild and catalyst-free conjugation method. The amount of peptide ALST1 on the DND was analyzed and quantified after acid hydrolysis and PITC derivatization by HPLC as described before. Then, the ALST1-/luminescent probe-functionalized DND was investigated for its photophysical properties by fluorescence spectroscopy. The Cy5-functionalized material showed a slightly lower fluorescence performance in aqueous solution than reported in literature and commercial suppliers with a life time < 0.4 ns and quantum yields not determinable by integration sphere due to the week signal intensity. The rhenium complex-functionalized material had a very low signal intensity in only aqueous medium, and thus determination of life times and quantum yield by fluorescence spectroscopy was not possible. After incubation with MDA-MB 231 cells, the Cy5-functionalized DND could easily be detected due to its red fluorescence. However, it was not possible to visualize the rhenium complex-functionalized DND with fluorescence microscopy due to the low fluorescence intensity of the complex in aqueous medium and the lack of proper filters for the fluorescence microscope. Cy5-functionalized DND did not show any cellular uptake in fluorescence microscopy after conjugation with ALST1. Since the nanodiamond surface is known to strongly adsorb peptides and proteins, it is assumed that the peptide chain is oriented perpendicular to the nanoparticle surface and thus not able to interact with cell membrane receptors to promote cell uptake of the particles. As a future perspective, the ALST1-promoted cellular uptake of the DND should be improved by using different linker systems for peptide conjugation to prevent adsorption of the peptide chain on the particle surface.
The new analytical methods for amino-, azide-, and peptide-functionalized nanoparticles have great potential to assist in the quantification of nanoparticle surface modifications by UV/Vis spectroscopy and HPLC. The determination of surface amine and azide groups based on the cleavage of conjugated Fmoc-Rink amide linker and detected by UV/Vis spectroscopy is applicable to all amine-/azide-functionalized nanomaterials. However, particles which form very stable suspension with the cleavage mixture can cause quantification problems due to scattering, making an accurate quantification of dibenzofulvene and its piperidine adduct impossible. The detection of tyrosine- and histidine-containing peptides based on the Pauly reaction is well-suited as a fast and easy-to-perform qualitative demonstration of successful peptide surface conjugation. However, its major drawback as a colourimetric approach is that coloured particles cannot be evaluated by this method. The amino acid analysis based on HPLC after acid hydrolysis of peptides conjugated to nanoparticle surfaces to its individual building blocks and subsequent derivatization with PITC, can be used on all nanomaterials with peptide or protein surface modification. It allows detection of amino acids down to picomolar concentrations and even enables analysis of very small peptide surface loadings. However, the resulting HPLC traces are difficult to analyze.
Three new analytical methods based on UV/Vis and HPLC techniques have been developed and established. They assisted in the characterization of the synthesized DND and SPIONs with dual functionalization by ALST1 and Cy5 or [ReBr(CO)3(L)], respectively. However, the nanomaterials showed no cellular uptake due to a high tendency to agglomerate. The cellular uptake should be improved and the tendency to agglomerate of the SPIONs should be reduced by changing the surface coating from silica to either PEG or chitosan. Furthermore, different linker systems for connecting peptides to DND surfaces should be synthesized and evaluated to reduce potential peptide chain adsorption.
A closer look at long-established drugs: enantioselective protein binding and stability studies
(2023)
The aim of this work was to investigate older, established drugs. The extent of the protein binding of chiral ephedra alkaloids to AGP and of ketamine to albumin was determined. Since enantiomers of these drugs are individual available, the focus was on possible enantioselective binding and structural moieties involved in the binding.
Previously published work suggested that ephedrine and pseudoephedrine can bind stereoselectively to proteins other than albumin in serum. For the determination of the extent of protein binding, the established ultrafiltration with subsequent chiral CE analysis was used. To determine the influence of basicity on binding, the drugs methylephedrine and norephedrine were also analyzed. Drug binding to AGP increased with increasing basicity as follows: norephedrine < methylephedrine < ephedrine < pseudoephedrine. pKaff was determined both graphically using the Klotz plot and mathematical indicating a low affinity of the ephedra alkaloids to AGP. Using STD-NMR spectroscopy experiments the aromatic protons and the C-CH3 side chain were shown to be most strongly involved in binding, which could be confirmed by molecular docking experiments in more detail. For all drugs, van der Waals-, π π , cationic interactions, hydrogen bonds, and a formation of a salt bridge were observed. The individual enantiomers showed no significant differences and thus the binding of ephedra alkaloids to AGP is not significant.
In contrast to the ephedra alkaloids, the possible enantioselective binding to albumin was investigated for R and S ketamine. Again, ultrafiltration followed by CE analysis was performed. The binding of ketamine to one main binding site could be identified. A non-linear fit was used for the determination of pKaff. Using the NMR methods STD-NMR, waterLOGSY-NMR, and CPMG-NMRspectroscopy: the aromatic protons as well as the protons of the NCH3 methyl group showed the largest signal intensity changes, while the cyclohexanone protons showed the smallest changes. pKaff was also determined by the change in the chemical shift at different drug-protein ratios. These obtained values confirm the values obtained from ultrafiltration. Based on this, ketamine is classified as a low-affinity ligand to albumin. There were no significant differences between the individual enantiomers and thus the binding of ketamine to albumin is not a stereoselective process.
Using statistical design of experiments an efficient chiral CE method for determining the extent of protein binding of R and S ketamine to albumin was developed and validated according to ICH Q2 (R1) guideline.
The stability of ketamine was also investigated because a yellowish discoloration of an aqueous solution of ketamine developed under heat. XRPD investigations showed the same crystal structure for all batches examined. An untargeted screening using LC HRMS as well as LC UV measurements showed no degradation of ketamine or the presence of impurities in stress and non-stressed ketamine solutions, confirming the stability of ketamine under the stress conditions investigated. The lower the quality of the water used in the stress tests, the more intense the yellow discoloration occurred. The impurity or the mechanism that causes the yellow discoloration could not be identified.
Liquid chromatography has become the gold standard for modern quality control and purity analytics since its establishment in the 1930s. However, some analytical questions remain very challenging even today. Several molecules and impurities do not possess a suitable chromophore for the application of UV detection or cannot be retained well on regular RP columns. Possible solutions are found in derivatization procedures, but they are time consuming and can be prone to errors. In order to detect non chromophore molecules underivatized, the concept of aerosol based universal detection was established with the introduction of the evaporative light scattering detector (ELSD) in the 1970s and the charged aerosol detector (CAD) followed in 2002. These two challenging fields – polar and non chromophore molecules – are tackled in this thesis.
An overview of applications of the CAD in the literature and a comparison to its aerosol based competitors and MS is presented, emphasizing on its high sensitivity and robustness. Parameters and techniques to overcome the drawbacks of CAD, such as the use of gradient compensation or adjusted evaporation temperatures are discussed. A consideration of aspects and drawbacks of data transformation such as the integrated power function value (PFV) in the GMP environment is performed.
A method for the fatty acid analysis in polysorbate 80 that was developed on HPLC CAD was transferred to UHPLC CAD. Time and eluent savings of over 75% and 40%, respectively, as well as ways to determine the optimal CAD parameters resulted from this investigation. The evaporation temperature was determined as the most crucial setting, which has to be adjusted with care. Optimal signal to noise ratios are found at a compromise between maintaining analyte signal and reducing background noise. The incorporation of semi volatile short chain fatty acids enabled the observation of differences based on volatility of the analyte. E.g. for semi volatiles, an improved linearity by means of adjusting the PFV is achieved at values below 1.0 instead of at elevated PFVs.
Using sugars and sugar related antibiotics, a proof-of-concept was given that artificial neural networks can describe correlations between the structure and physicochemical properties of molecules and their response in CAD. Quantitative structure property relationships obtained by design of experiment approaches were able to predict the response of unseen substances and yielded insights on the response generation of the detector, which heavily relies on the formed surface area of the dried particle. Further work can substantiate upon these findings, eventually building a library of diverse eluent compositions, analytes and settings.
In order to cope with a chromatographically challenging substances, the application of ion pairing reversed phase chromatography coupled to low wavelength UV detection has been shown as a possible approach for the amino acid L asparagine. A method capable of compendial purity analysis in one single HPLC approach, thus making the utilization of the semi quantitative TLC-ninhydrin analysis obsolete, resulted from this. One cyclic dipeptide impurity (diketoasparagine) that was formerly not assessed, could be identified in several batches and added to the monograph of the Ph.Eur.
Studying ibandronate sodium with CAD and ELSD, it was found that randomly occurring spike peaks represent a major flaw of the ELSD when high sample load is present. The research with this non chromophore bisphosphonate drug furthermore shed light on possible drawbacks of mixed mode chromatography methods and ways to overcome these issues. Due to strong adsorption of the analyte onto the column, over ten injections of the highly concentrated test solution were found to be necessary to ensure reproducible peak areas. Preconditioning steps should thus be evaluated for mixed mode approaches during method development and validation.
Last, using a ternary mixed mode stationary phase coupled to CAD, a method for the impurity profiling of pamidronate disodium, also applicable to the assessment of phosphate and phosphite in four other bisphosphonate drugs, has been developed. This represents a major advantage over the Ph.Eur. impurity profiling of pamidronate, which requires two different methods, one of which is only a semi quantitative TLC approach.
Among external stimuli used to trigger release of a drug from a polymeric carrier, ultrasound has gained increasing attention due to its non-invasive nature, safety and low cost. Despite this attention, there is only limited knowledge about how materials available for the preparation of drug carriers respond to ultrasound. This study investigates the effect of ultrasound on the release of a hydrophobic drug, dexamethasone, from poly(2-oxazoline)-based micelles. Spontaneous and ultrasound-mediated release of dexamethasone from five types of micelles made of poly(2-oxazoline) block copolymers, composed of hydrophilic poly(2-methyl-2-oxazoline) and hydrophobic poly(2-n-propyl-2-oxazoline) or poly(2-butyl-2-oxazoline-co-2-(3-butenyl)-2-oxazoline), was studied. The release profiles were fitted by zeroorder and Ritger-Peppas models. The ultrasound increased the amount of released dexamethasone by 6% to 105% depending on the type of copolymer, the amount of loaded dexamethasone, and the stimulation time point. This study investigates for the first time the interaction between different poly(2-oxazoline)-based micelle formulations and ultrasound waves, quantifying the efficacy of such stimulation in modulating dexamethasone release from these nanocarriers.
The charged aerosol detector (CAD) is an aerosol-based detector employed in liquid chromatography which has become established in the field of pharmaceutical analysis due to its outstanding performance characteristics, e.g. the almost uniform response for nonvolatile analytes. Owing to its principle of detection, the response of the CAD depends on the volatility of a compound and is inherently nonlinear. However, the newly implemented instrumental settings evaporation temperature and power function value (PFV) are valuable tools to overcome some of these drawbacks and can even enhance the detector’s capabilities when adjusted properly.
This thesis aimed to evaluate the impact of the new instrumental settings on the CAD performance. Additionally, the influence of modern separation techniques for small polar compounds on the CAD was assessed and the applicability of hyphenated UV-CAD techniques explored. The optimization strategies derived from the evaluation procedures and the conjunction of the instrumental and chromatographic techniques investigated were utilized for the challenging impurity profiling of amino acids and amino acid-like drugs.
The results of the method validation procedures confirmed the broad applicability of the CAD in the pharmaceutical analysis of nonvolatile compounds, supported by satisfactory sensitivity and reproducibility for meeting the regulatory requirements with respect to the ICH guidelines Q2(R1) and Q3A(R2). The limits of applicability include the analysis of semivolatile compounds, and the method transfer between current and legacy CAD models. Further advances in the definition and standardization of allowed ranges for the instrumental settings and the establishment of general optimization procedures in the method development could lead to a more widespread use of the detection technique in compendial methods.
Nowadays, the management of infectious diseases is especially threatened by the rapid emergence of drug resistance. It has been suggested that the medicine quality assurance combined with good medication adherence may help to reduce this impendence. Moreover, the search for new antimicrobial agents from medicinal plants is strongly encouraged for the exploration of alternatives to existing therapies. In this context, the present work focused on both the quality evaluation of commercialized antimalarial medicines from the Democratic Republic of the Congo and on the phytochemical investigations of a Congolese Ancistrocladus species.
All presented studies aimed on the improvement of the quality analysis of already monographed drugs. Thereby different LC methods were applied and coupled to i.e., the UV/VIS detector, the CAD or a hyphenation of these detectors, respectively. The choice of the chromatographic system including the detector was largely dependent on the physicochemical properties of the respective analytes.
With the risk-assessment report on the API cetirizine we presented an exemplary tool, that can help to minimize the risk of the occurrence of unexpected impurities. An in- deep analysis of each step within synthesis pathway by means of reaction matrices of all compounds was performed. It is essential to understand the complete impurity profile of all reactants, solvents, and catalysts and to include them in the matrix. Finally, the API of this synthesis was checked if all impurities are identified by this tool. Of note, a shortcoming of such a targeted approach is that impurities can still occur, but they are not captured. This disadvantage can be partially compensated by non-targeted approaches if they are performed in parallel with the other studies that represent most of the impurities. However, this work also shows that even in a supposedly simple synthesis, potentially hundreds of by-products can be formed. For each of them, it must be decided individually whether their formation is probable or how their quantity can be minimized in order to obtain APIs, that are as pure as possible.
In the dapsone project it was aimed to replace the existing old Ph. Eur. TLC method with a modern RP-HPLC method. This was successful and since Ph. Eur. 10.6, the method developed in this work, became a valid monograph. Within the revision process of the monograph, the individual limits for impurities were tightened. However, this new method needs HPLC instrumentation, suitable to perform gradients. As this is not always available in all control laboratories, we also developed an alternative, more simple method using two different isocratic runs for the impurity analysis. The obtained batch results of both, the new pharmacopoeial method and the more simple one, were in a comparable order of magnitude. Furthermore, within the method development stage of the Ph. Eur. method, we could identify one unknown impurity of the impurity reference by high-resolution MS/MS analysis.
Also, in the baclofen project it was aimed to replace the existing Ph. Eur. method with the introduction of an additional impurity to be quantified. A corresponding method was developed and validated. However, due to the harmonization process of the pharmacopoeias, it is currently not used. In addition, we tried to find further, non- 116
SUMMARY
chromophoric impurities by means of the CAD. However, except for one counterion of an impurity, no further impurities were found. Also, the aforementioned new impurity could not be detected above the reporting threshold in the batches analyzed. As the only individually specified impurity A is also present at a low level, it can be concluded that the examined batches of baclofen are very pure.
The use of universal detectors, such as the CAD can be particularly interesting for compounds with no chromophore or those with only a weak chromophore. Therefore, we decided to take a closer look at the impurity profile of acarbose. Currently, acarbose and its impurities are being studied by low wavelength UV detection at 210 nm. Therefore, the question arose whether there are no other impurities in the API that do not show absorption at this wavelength. CAD, which offers consistent detection properties for all non-volatile compounds, is ideally suited for this purpose. However, it was not so easy to use the CAD together with the UV detector, for example, as a hyphenated detection technique, because the Ph. Eur. method uses phosphate buffers. However, this is non-volatile and therefore inappropriate for the CAD. Therefore, an attempt was made to replace the buffer with a volatile one. However, since this did not lead to satisfactory results and rather the self-degradation process of the stationary phase used could be observed by means of the CAD, it was decided to switch to alternative stationary phases. A column screening also revealed further difficulties with acarbose and its impurities: they show an epimerization reaction at the end of the sugar chain. However, since one wanted to have uniform peaks in the corresponding chromatograms, one had to accelerate this reaction significantly to obtain only one peak for each component. This was best achieved by using two stationary phases: PGC and Amide-HILIC. Impurity-profiling methods could be developed on each of the two phases. In addition, as expected, new impurities could be detected, albeit at a low level. Two of them could even be identified by spiking experiments as the sugar fragments maltose and maltotriose.
Taken together, it can be concluded, that this work has contributed significantly to the improvement of the quality analysis of monographed drugs. In addition to the presented general tool for the identification of potential impurities, one of the methods developed, had already been implemented to the Ph. Eur. In an effort to improve the CAD's universal detection capabilities, additional methods have also been developed. Further, new improved methods for the impurity profiling are ready to use.