The use of non-invasive continuous blood pressure measuring (ClearSight\(^®\)) during central neuraxial anaesthesia for caesarean section — a retrospective validation study

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-286042
  • Background: The close monitoring of blood pressure during a caesarean section performed under central neuraxial anaesthesia should be the standard of safe anaesthesia. As classical oscillometric and invasive blood pressure measuring have intrinsic disadvantages, we investigated a novel, non-invasive technique for continuous blood pressure measuring. Methods: In this monocentric, retrospective data analysis, the reliability of continuous non-invasive blood pressure measuring using ClearSight\(^®\) (Edwards Lifesciences Corporation) is validatedBackground: The close monitoring of blood pressure during a caesarean section performed under central neuraxial anaesthesia should be the standard of safe anaesthesia. As classical oscillometric and invasive blood pressure measuring have intrinsic disadvantages, we investigated a novel, non-invasive technique for continuous blood pressure measuring. Methods: In this monocentric, retrospective data analysis, the reliability of continuous non-invasive blood pressure measuring using ClearSight\(^®\) (Edwards Lifesciences Corporation) is validated in 31 women undergoing central neuraxial anaesthesia for caesarean section. In addition, patients and professionals evaluated ClearSight\(^®\) through questioning. Results: 139 measurements from 11 patients were included in the final analysis. Employing Bland–Altman analyses, we identified a bias of −10.8 mmHg for systolic, of −0.45 mmHg for diastolic and of +0.68 mmHg for mean arterial blood pressure measurements. Pooling all paired measurements resulted in a Pearson correlation coefficient of 0.7 for systolic, of 0.67 for diastolic and of 0.75 for mean arterial blood pressure. Compensating the interindividual differences in linear regressions of the paired measurements provided improved correlation coefficients of 0.73 for systolic, of 0.9 for diastolic and of 0.89 for mean arterial blood pressure measurements. Discussion: Diastolic and mean arterial blood pressure are within an acceptable range of deviation from the reference method, according to the Association for the Advancement of Medical Instrumentation (AAMI) in the patient collective under study. Both patients and professionals prefer ClearSight\(^®\) to oscillometric blood pressure measurement in regard of comfort and handling.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Philipp Helmer, Daniel Helf, Michael Sammeth, Bernd Winkler, Sebastian Hottenrott, Patrick MeybohmORCiD, Peter Kranke
URN:urn:nbn:de:bvb:20-opus-286042
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Klinik und Poliklinik für Anästhesiologie (ab 2004)
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Journal of Clinical Medicine
ISSN:2077-0383
Erscheinungsjahr:2022
Band / Jahrgang:11
Heft / Ausgabe:15
Aufsatznummer:4498
Originalveröffentlichung / Quelle:Journal of Clinical Medicine (2022) 11:15, 4498. https://doi.org/10.3390/jcm11154498
DOI:https://doi.org/10.3390/jcm11154498
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):Bland–Altman; ClearSight\(^®\); Fisher Z-score transformation; Pearson correlation coefficient; epidural anaesthesia; supine hypotensive syndrome; volume clamp
Datum der Freischaltung:21.08.2023
Datum der Erstveröffentlichung:02.08.2022
Lizenz (Deutsch):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International