Das Suchergebnis hat sich seit Ihrer Suchanfrage verändert. Eventuell werden Dokumente in anderer Reihenfolge angezeigt.
  • Treffer 3 von 5
Zurück zur Trefferliste

Spin echo based cardiac diffusion imaging at 7T: An ex vivo study of the porcine heart at 7T and 3T

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-201376
  • Purpose of this work was to assess feasibility of cardiac diffusion tensor imaging (cDTI) at 7 T in a set of healthy, unfixed, porcine hearts using various parallel imaging acceleration factors and to compare SNR and derived cDTI metrics to a reference measured at 3 T. Magnetic resonance imaging was performed on 7T and 3T whole body systems using a spin echo diffusion encoding sequence with echo planar imaging readout. Five reference (b = 0 s/mm\(^2\)) images and 30 diffusion directions (b = 700 s/mm\(^2\)) were acquired at both 7 T and 3 TPurpose of this work was to assess feasibility of cardiac diffusion tensor imaging (cDTI) at 7 T in a set of healthy, unfixed, porcine hearts using various parallel imaging acceleration factors and to compare SNR and derived cDTI metrics to a reference measured at 3 T. Magnetic resonance imaging was performed on 7T and 3T whole body systems using a spin echo diffusion encoding sequence with echo planar imaging readout. Five reference (b = 0 s/mm\(^2\)) images and 30 diffusion directions (b = 700 s/mm\(^2\)) were acquired at both 7 T and 3 T using a GRAPPA acceleration factor R = 1. Scans at 7 T were repeated using R = 2, R = 3, and R = 4. SNR evaluation was based on 30 reference (b = 0 s/mm\(^2\)) images of 30 slices of the left ventricle and cardiac DTI metrics were compared within AHA segmentation. The number of hearts scanned at 7 T and 3 T was n = 11. No statistically significant differences were found for evaluated helix angle, secondary eigenvector angle, fractional anisotropy and apparent diffusion coefficient at the different field strengths, given sufficiently high SNR and geometrically undistorted images. R≥3 was needed to reduce susceptibility induced geometric distortions to an acceptable amount. On average SNR in myocardium of the left ventricle was increased from 29±3 to 44±6 in the reference image (b = 0 s/mm\(^2\)) when switching from 3 T to 7 T. Our study demonstrates that high resolution, ex vivo cDTI is feasible at 7 T using commercial hardware.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): David Lohr, Maxim Terekhov, Andreas Max Weng, Anja Schroeder, Heike Walles, Laura Maria Schreiber
URN:urn:nbn:de:bvb:20-opus-201376
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Institut für diagnostische und interventionelle Radiologie (Institut für Röntgendiagnostik)
Medizinische Fakultät / Lehrstuhl für Tissue Engineering und Regenerative Medizin
Medizinische Fakultät / Deutsches Zentrum für Herzinsuffizienz (DZHI)
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):PLoS ONE
Erscheinungsjahr:2019
Band / Jahrgang:14
Heft / Ausgabe:3
Seitenangabe:e0213994
Originalveröffentlichung / Quelle:PLoS ONE (2019) 14:3, e0213994. https://doi.org/10.1371/journal.pone.0213994
DOI:https://doi.org/10.1371/journal.pone.0213994
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):Cardiac ventricles; Data acquisition; Diffusion tensor imaging; Eigenvectors; Heart; Magnetic resonance imaging; Swine; Tractography
Datum der Freischaltung:11.05.2020
Sammlungen:Open-Access-Publikationsfonds / Förderzeitraum 2019
Lizenz (Deutsch):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International