• Treffer 2 von 6
Zurück zur Trefferliste

Pumpless extracorporeal hemadsorption technique (pEHAT): a proof-of-concept animal study

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-297347
  • Background: Extracorporeal hemadsorption eliminates proinflammatory mediators in critically ill patients with hyperinflammation. The use of a pumpless extracorporeal hemadsorption technique allows its early usage prior to organ failure and the need for an additional medical device. In our animal model, we investigated the feasibility of pumpless extracorporeal hemadsorption over a wide range of mean arterial pressures (MAP). Methods: An arteriovenous shunt between the femoral artery and femoral vein was established in eight pigs. TheBackground: Extracorporeal hemadsorption eliminates proinflammatory mediators in critically ill patients with hyperinflammation. The use of a pumpless extracorporeal hemadsorption technique allows its early usage prior to organ failure and the need for an additional medical device. In our animal model, we investigated the feasibility of pumpless extracorporeal hemadsorption over a wide range of mean arterial pressures (MAP). Methods: An arteriovenous shunt between the femoral artery and femoral vein was established in eight pigs. The hemadsorption devices were inserted into the shunt circulation; four pigs received CytoSorb\(^®\) and four Oxiris\(^®\) hemadsorbers. Extracorporeal blood flow was measured in a range between mean arterial pressures of 45–85 mmHg. Mean arterial pressures were preset using intravenous infusions of noradrenaline, urapidil, or increased sedatives. Results: Extracorporeal blood flows remained well above the minimum flows recommended by the manufacturers throughout all MAP steps for both devices. Linear regression resulted in CytoSorb\(^®\) blood flow [mL/min] = 4.226 × MAP [mmHg] − 3.496 (R-square 0.8133) and Oxiris\(^®\) blood flow [mL/min] = 3.267 × MAP [mmHg] + 57.63 (R-square 0.8708), respectively. Conclusion: Arteriovenous pumpless extracorporeal hemadsorption resulted in sufficient blood flows through both the CytoSorb\(^®\) and Oxiris\(^®\) devices over a wide range of mean arterial blood pressures and is likely an intriguing therapeutic option in the early phase of septic shock or hyperinflammatory syndromes.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Mascha O. Fiedler, Ralf M. Muellenbach, Caroline Rolfes, Christopher Lotz, Felix Nickel, Beat P. Müller-Stich, Alexander Supady, Philipp M. Lepper, Markus A. Weigand, Patrick MeybohmORCiD, Armin Kalenka, Christian Reyher
URN:urn:nbn:de:bvb:20-opus-297347
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Medizinische Fakultät / Klinik und Poliklinik für Anästhesiologie (ab 2004)
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Journal of Clinical Medicine
ISSN:2077-0383
Erscheinungsjahr:2022
Band / Jahrgang:11
Heft / Ausgabe:22
Aufsatznummer:6815
Originalveröffentlichung / Quelle:Journal of Clinical Medicine (2022) 11:22, 6815. https://doi.org/10.3390/jcm11226815
DOI:https://doi.org/10.3390/jcm11226815
Allgemeine fachliche Zuordnung (DDC-Klassifikation):6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):adsorption; animal model; arteriovenous extracorporeal hemadsorption technique; blood purification; cytokines; endotoxin; extracorporeal hemadsorption; extracorporeal techniques in hemadsorption therapy; immunosorbents; septic shock
Datum der Freischaltung:13.11.2023
Datum der Erstveröffentlichung:18.11.2022
Lizenz (Deutsch):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International