• Treffer 3 von 13
Zurück zur Trefferliste

A dosimetric comparison of real-time adaptive and non-adaptive radiotherapy: a multi-institutional study encompassing robotic, gimbaled, multileaf collimator and couch tracking

Zitieren Sie bitte immer diese URN: urn:nbn:de:bvb:20-opus-189605
  • Purpose: A study of real-time adaptive radiotherapy systems was performed to test the hypothesis that, across delivery systems and institutions, the dosimetric accuracy is improved with adaptive treatments over non-adaptive radiotherapy in the presence of patient-measured tumor motion. Methods and materials: Ten institutions with robotic(2), gimbaled(2), MLC(4) or couch tracking(2) used common materials including CT and structure sets, motion traces and planning protocols to create a lung and a prostate plan. For each motion trace, the plan wasPurpose: A study of real-time adaptive radiotherapy systems was performed to test the hypothesis that, across delivery systems and institutions, the dosimetric accuracy is improved with adaptive treatments over non-adaptive radiotherapy in the presence of patient-measured tumor motion. Methods and materials: Ten institutions with robotic(2), gimbaled(2), MLC(4) or couch tracking(2) used common materials including CT and structure sets, motion traces and planning protocols to create a lung and a prostate plan. For each motion trace, the plan was delivered twice to a moving dosimeter; with and without real-time adaptation. Each measurement was compared to a static measurement and the percentage of failed points for gamma-tests recorded. Results: For all lung traces all measurement sets show improved dose accuracy with a mean 2%/2 mm gamma-fail rate of 1.6% with adaptation and 15.2% without adaptation (p < 0.001). For all prostate the mean 2%/2 mm gamma-fail rate was 1.4% with adaptation and 17.3% without adaptation (p < 0.001). The difference between the four systems was small with an average 2%/2 mm gamma-fail rate of <3% for all systems with adaptation for lung and prostate. Conclusions: The investigated systems all accounted for realistic tumor motion accurately and performed to a similar high standard, with real-time adaptation significantly outperforming non-adaptive delivery methods.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Statistik - Anzahl der Zugriffe auf das Dokument
Metadaten
Autor(en): Emma Colvill, Jeremy Booth, Simeon Nill, Martin Fast, James Bedford, Uwe Oelfke, Mitsuhiro Nakamura, Per Poulsen, Esben Worm, Rune Hansen, Thomas Ravkilde, Jonas Scherman Rydhög, Tobias Pommer, Per Munck af Rosenschold, Stephanie Lang, Matthias Guckenberger, Christian Groh, Christian Herrmann, Dirk Verellen, Kenneth Poels, Lei Wang, Michael Hadsell, Thilo Sothmann, Oliver Blanck, Paul Keall
URN:urn:nbn:de:bvb:20-opus-189605
Dokumentart:Artikel / Aufsatz in einer Zeitschrift
Institute der Universität:Fakultät für Mathematik und Informatik / Institut für Informatik
Medizinische Fakultät / Klinik und Poliklinik für Strahlentherapie
Sprache der Veröffentlichung:Englisch
Titel des übergeordneten Werkes / der Zeitschrift (Englisch):Radiotherapy and Oncology
Erscheinungsjahr:2016
Band / Jahrgang:119
Heft / Ausgabe:1
Seitenangabe:159-165
Originalveröffentlichung / Quelle:Radiotherapy and Oncology (2016) 119:1, S. 159-165. https://doi.org/10.1016/j.radonc.2016.03.006
DOI:https://doi.org/10.1016/j.radonc.2016.03.006
Allgemeine fachliche Zuordnung (DDC-Klassifikation):0 Informatik, Informationswissenschaft, allgemeine Werke / 00 Informatik, Wissen, Systeme / 000 Informatik, Informationswissenschaft, allgemeine Werke
6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Freie Schlagwort(e):Couch tracking; Gimbaled tracking; MLC tracking; Organ motion; Robotic tracking
Datum der Freischaltung:17.12.2020
Lizenz (Deutsch):License LogoCC BY-NC-ND: Creative-Commons-Lizenz: Namensnennung, Nicht kommerziell, Keine Bearbeitungen 4.0 International