Motion detectors as additional monitoring devices in the intensive care unit — a proof-of-concept study

Please always quote using this URN: urn:nbn:de:bvb:20-opus-362404
  • Background: Monitoring the vital signs of delirious patients in an intensive care unit (ICU) is challenging, as they might (un-)intentionally remove devices attached to their bodies. In mock-up scenarios, we systematically assessed whether a motion detector (MD) attached to the bed may help in identifying emergencies. Methods: We recruited 15 employees of the ICU and equipped an ICU bed with an MD (IRON Software GmbH, Grünwald, Germany). Participants were asked to replay 22 mock-up scenes of one-minute duration each: 12 scenes with movementsBackground: Monitoring the vital signs of delirious patients in an intensive care unit (ICU) is challenging, as they might (un-)intentionally remove devices attached to their bodies. In mock-up scenarios, we systematically assessed whether a motion detector (MD) attached to the bed may help in identifying emergencies. Methods: We recruited 15 employees of the ICU and equipped an ICU bed with an MD (IRON Software GmbH, Grünwald, Germany). Participants were asked to replay 22 mock-up scenes of one-minute duration each: 12 scenes with movements and 10 without movements, of which 5 were emergency scenes (“lying dead-still, with no or very shallow breathing”). Blinded recordings were presented to an evaluation panel consisting of an experienced ICU nurse and a physician, who was asked to assess and rate the presence of motions. Results: Fifteen participants (nine women; 173 ± 7.0 cm; 78 ± 19 kg) joined the study. In total, 286 out of 330 scenes (86.7%) were rated correctly. Ratings were false negative (FN: “no movements detected, but recorded”) in 7 out of 180 motion scenes (3.9%). Ratings were false positive (FP: “movements detected, but not recorded”) in 37 out of 150 scenes (24.7%), more often in men than women (26 out of 60 vs. 11 out of 90, respectively; p < 0.001). Of note, in 16 of these 37 FP-rated scenes, a vibrating mobile phone was identified as a potential confounder. The emergency scenes were correctly rated in 64 of the 75 runs (85.3%); 10 of the 11 FP-rated scenes occurred in male subjects. Conclusions: The MD allowed for identifying motions of test subjects with high sensitivity (96%) and acceptable specificity (75%). Accuracy might increase further if activities are recorded continuously under real-world conditions.show moreshow less

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author: Gülmisal GüderORCiD, Eva von Rein, Thomas Flohr, Dirk Weismann, Dominik SchmittORCiD, Stefan StörkORCiD, Stefan Frantz, Vincent Kratzer, Christian Kendi
URN:urn:nbn:de:bvb:20-opus-362404
Document Type:Journal article
Faculties:Medizinische Fakultät / Medizinische Klinik und Poliklinik I
Medizinische Fakultät / Deutsches Zentrum für Herzinsuffizienz (DZHI)
Language:English
Parent Title (English):Applied Sciences
ISSN:2076-3417
Year of Completion:2023
Volume:13
Issue:16
Article Number:9319
Source:Applied Sciences (2023) 13:16, 9319. https://doi.org/10.3390/app13169319
DOI:https://doi.org/10.3390/app13169319
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Tag:Internet of Things devices; motion detector; noncontact monitoring
Release Date:2024/06/10
Licence (German):License LogoCC BY: Creative-Commons-Lizenz: Namensnennung 4.0 International